Large Fixed-Point Cells

Large SRM fixed-point cells containing either high purity (>99.9999% pure) Sn and Zn and were developed for use as ITS-90 defining fixed-point cells to calibrate (HT)SPRTs. They are designated as SRM 1747 (Sn freezing-point cell) and SRM 1748 (Zn freezing-point cell). They, together with the triple point of water cell, are used to calibrate SPRTs from 0 °C to 420 °C, used for part of the calibration of SPRTs from 0 °C to 661 °C, or used for part of the calibration of HTSPRTs from 0 °C to 962 °C.

 

ITS-90 Defining Fixed-Point Cells

SRM # Fixed Point t, °C Well i.d., mm U (k=2), mK Unit Cost*,$
1747 Tin FP 231.928 8.0 0.4 10,754
1748 Zinc FP 419.527 8.0 1.1 10,754

FP = freezing point

*PRICES ARE SUBJECT TO CHANGE WITHOUT NOTICE