Method no.: |
ID-211 |
|
|
Matrix: |
Air |
|
|
OSHA Permissible Exposure Limits:
Final Rule Limits:
Transitional Limit: |
0.3 mg/m3 as sodium azide (Ceiling) 0.1 ppm as hydrazoic acid (Ceiling) Also Skin Designation
None |
|
|
Collection Device: |
An air sample is collected using a calibrated sampling pump and a glass tube containing impregnated silica gel (ISG). A pre-filter is used to collect particulate azide. Wipe samples can be taken to determine work surface contamination. |
|
|
Recommended Sampling Rate: |
1 liter per minute (L/min) |
|
|
Recommended Minimum
Sampling Time: |
5 minutes |
|
|
Analytical Procedure: |
The sampling medium is desorbed using an aqueous solution which contains a mixture of 0.9 mM sodium carbonate (Na2CO3) and 0.9 mM sodium bicarbonate (NaHCO3). An aliquot of this solution is analyzed as azide (N3-) by an ion chromatograph equipped with a UV detector. |
|
|
Special Precautions: |
Ship samples to the laboratory as soon as possible after collection. Store samples under refrigeration when not in transit. Samples stored at room temperature need to be analyzed within 10 days. |
|
|
Detection Limit: Qualitative: Quantitative: |
0.001 ppm as HN3 or 0.003 mg/m3 as NaN3 (5-L air sample) 0.004 ppm as HN3 or 0.011 mg/m3 as NaN3 (5-L air sample) |
|
|
Precision and Accuracy: Validation Range: Bias: Overall Error: CVT (pooled): |
0.057 to 2.63 ppm -0.022 ±12.6% 0.052 |
|
|
Method Classification: |
Validated Method |
|
|
Chemist: |
James C. Ku |
|
|
Date: |
September, 1992 |
|
|
Please note: For problems with accessibility in using
figures and illustrations in this method,
please contact the author at (801) 233-4900. |
|
|
Commercial manufacturers and products mentioned in this method are for descriptive use only and do not constitute
endorsements by USDOL-OSHA. Similar products from other sources can be substituted.
|
Branch of Inorganic Methods Development OSHA Salt Lake Technical Center Sandy, Utah
|
1. Introduction
This method describes the sample collection and analysis of airborne azides [as sodium azide
(NaN3) and hydrazoic acid (HN3)]. Air samples are taken in the breathing zone
of workplace personnel, and analysis is performed by ion chromatography (IC) with a UV detector.
Note: Hydrazoic acid vapor may coexist with NaN3 in the workplace when NaN3 is
handled in the presence of moisture. This method addresses the potential exposure to both substances
(NaN3/HN3), and may be extended to include other azide compounds, provided
they are soluble in the desorbing solution and collected using the procedure described below. Wipe or bulk samples can also be
collected and analyzed using this method.
1.1. History
1.1.1. Various sampling and analysis methods have been proposed in the literature
(5.1. -
5.5.) for monitoring azide exposures; however, most
lack the sensitivity needed to meet the 0.3 mg/m3 (as NaN3) or 0.1 ppm
(as HN3) Ceiling PEL when using short sampling periods. Some of these methods are subject to interferences
from many compounds. The ion chromatographic method has interferences from nitrates or bromides. The National Institute of Occupational
Safety and Health (NIOSH) had proposed a method for inorganic azide particulates using polyvinyl chloride (PVC) filter collection
followed by water extraction and IC determination using sodium bicarbonate/sodium hydroxide eluant
(5.6.). To trap any HN3, NIOSH
further recommended using a solid sorbent tube containing chromosorb coated with sodium carbonate. The NIOSH method is also
subject to interferences and the conductivity detector used lacks sufficient sensitivity for short-term samples.
1.1.2. The OSHA Salt Lake Technical Center (SLTC) previously used a stopgap method for NaN3
(5.7.). Samples were collected with impingers which were inconvenient
to use as personal samplers due to possible spillage of the liquid collection solutions or breakage. Other disadvantages are similar to
those mentioned above: 1) low sensitivity due to the conductivity detector used; 2) interferences from ions such as bromide, adipic
acid, and nitrate.
1.1.3. It was desirable to develop a solid-sorbent sampling and analytical method capable of measuring azide for OSHA compliance
purposes. A method was evaluated using a base-impregnated silica gel (ISG) as the collection media. The media is similar to that
found in reference 5.5.
1.2. Principle
Particulate NaN3 is collected on a PVC filter or in the glass wool plug of the sampling tube. Gaseous
HN3 is collected and converted to NaN3 by the ISG sorbent within the sampling tube.
The collected azide on either media is desorbed in a weak buffer solution. The resultant anion, N3-,
is analyzed by IC using a variable wavelength UV detector at 210 nm. A gravimetric conversion is used to calculate the amount of
NaN3 or HN3 collected.
1.3. Advantages and Disadvantages
1.3.1. This method has adequate sensitivity to determine compliance with the OSHA Ceiling PEL azide exposures.
1.3.2. The method is simple, rapid, and easily automated.
1.3.3. The potential for sample contamination is minimal. The azide anion, N3-, is normally not
detected in sorbent blanks.
1.3.4. One disadvantage is sample storage stability. Samples should be refrigerated after collection to improve stability. Samples need
not be refrigerated during shipment provided they are shipped as soon as possible.
1.3.5. Another disadvantage is the method does not distinguish azide compounds. If other azide compounds are present during sampling,
and are soluble in the desorbing solution, positive interferences could occur. However, most industrial operations do not mix different
azide-containing compounds in their processes.
1.4. Method Performance
A synopsis of the method performance is presented below. Further information can be found in Section
4.
1.4.1. This method was validated over the concentration range of 0.057 to 0.263 ppm as HN3. An air volume
of 5 L and a flow rate of approximately 1 L/min were used.
1.4.2. The qualitative detection limit was 0.00347 µg/mL or 0.0104 µg (as N3-)
when using a 3-mL solution volume. This corresponds to 0.001 ppm HN3 or 0.004 mg/m3
NaN3 for a 5-L air volume.
1.4.3. The quantitative detection limit was 0.0116 µg/mL or 0.0348 µg (as N3-)
when using a 3-mL solution volume. This corresponds to 0.004 ppm HN3 or 0.011 mg/m3
NaN3 for a 5-L air volume. A 50-µL sample loop and a detector setting of 0.01 absorbance unit (AU)
full-scale output were used.
1.4.4. The sensitivity of the analytical method, when using the instrumental parameters listed in Section
3.7., was calculated from the slope of a linear working range curve
(0.1 to 1.0 µg/mL N3-). The sensitivity was 2.1 × 107 area
units per 1 µg/mL. A Dionex Series 4500i ion chromatograph with AI450 computer software was used (Dionex, Sunnyvale, CA).
1.4.5. The precision and accuracy results are shown below (OE = Overall Error):
|
Ceiling |
CV |
0.052 |
Bias |
-2.2% |
OE |
±12.6% |
1.4.6. The collection efficiency at 2 times the PEL was 100%. Samples were collected from a generated test atmosphere of 0.26
ppm HN3 for 5 min.
1.4.7. A breakthrough test was performed at a concentration of 0.9 ppm HN3. Breakthrough was not found when
using a sampling time of 30 min and an average sample flow rate of 1 L/min.
1.4.8. Tests indicated the recovery for samples stored at room temperature (20 to 25°C) gradually decreases to between 75 and 80% after
30 days. Slight losses (6%) were observed for samples
stored 30 days in a refrigerator or freezer.
1.5. Interferences
1.5.1. Other azide compounds will interfere in the analysis of N3- if they are collected by the ISG,
glass wool, or on the PVC pre-filter. These compounds should normally not exist in industrial operations which specifically
use NaN3 or HN3 (i.e. manufacture of air bags, analytical laboratories, etc.).
1.5.2. Any substance that has the same retention time and absorbs UV at 210 nm, when using the ion chromatographic operating conditions
described in this method, may be an interference. If the possibility of an interference exists, changing the separation conditions
(column, eluant flow rate, eluant concentration, analytical wavelength, etc.) may circumvent the problem.
1.6. Sources of Exposure
Note: Because NaN3 is rapidly converted to HN3 on contact with
moisture, HN3 is believed to be the ultimate toxic agent in humans exposed to NaN3
(5.8.).
Sodium azide has been used for a wide variety of military, laboratory, medical, and commercial purposes. While it is not explosive under
normal conditions, NaN3 is commonly used in detonators and other explosives. Sodium azide is used extensively as
an intermediate in the production of lead azide. The biological uses of azides include inhibition of respiration, differential selection
procedures for bacteria, and bacteriocidal agents in diagnostic products
(5.9.,
5.10.).
Sodium azide is also the chief chemical used to inflate safety airbags in automobiles. Nitrogen gas is produced upon
NaN3 detonation. After inflation, a small residue of sodium hydroxide may be left, in addition to lubricants
such as corn starch or talc.
1.7. Physical and Chemical Properties (5.11.,
5.12.)
Hydrazoic acid (CAS No. 7782-79-8) is a colorless, volatile liquid which is soluble in water. It has a pungent obnoxious odor.
Chemical name |
Hydrozoic acid |
Synonym name |
Hydrogen azide |
Chemical formula |
HN3 |
Structural formula |
H-N=NN |
Formula weight |
43.03 |
Freezing point |
-80°C |
Boiling point |
37°C |
Sodium azide (CAS No. 26628-22-8) is a colorless, hexagonal crystalline solid. It is soluble in water or liquid ammonia, slightly
soluble in alcohol, and insoluble in ether. It is highly toxic and presents a severe explosion risk when shocked or heated. When
heated to 275 to 330°C in dry air, the solid crystals decompose with the evolution of nitrogen gas, leaving a residue of sodium oxide.
Sodium hydroxide forms in moist air.
Chemical name |
Sodium azide |
Synonym name |
Sodium azoimide |
Chemical formula |
NaN3 |
Structural formula |
Na-N=NN |
Formula weight |
65.01 |
Decomposition temperature |
300°C |
Specific gravity |
1.846 (@ 20°C) |
1.8. Toxicology (5.13.)
Information listed within this section is a synopsis of current knowledge of the physiological effects of
NaN3 and is not intended to be used as a basis for OSHA policy.
Sodium azide/hydrazoic acid is known to produce hypotension (low blood pressure) in laboratory animals and humans, and to form
strong complexes with hemoglobin, and consequently block oxygen transport in the blood.
Acute inhalation of HN3 vapor by humans (which forms when NaN3 contacts water) results
in lowered blood pressure, eye irritation, bronchitis, headache, weakness, and collapse. A skin designation has been assigned to the OSHA
PEL due to the ability of NaN3 to readily penetrate intact skin, and any dermal exposure can significantly
contribute to the overall exposure to azide.
2. Sampling
2.1. Equipment - Air Samples
2.1.1. Calibrated personal sampling pumps capable of sampling within ±5% of the recommended flow rate of 1 L/min are used.
2.1.2. Solid sorbent sampling tubes containing ISG are prepared by using clean silica gel impregnated with a base.
The sampling tube is proprietary and is composed of a glass jacket containing a 150-mg ISG front and 75-mg ISG backup section
(Cat. No. 226-55, SKC Inc., Eighty Four, PA). The dimensions of the tube are 7-mm o.d., 5-mm i.d., and 70-mm long. The ISG is held in
place with glass wool and a stainless steel retainer clip. A pre-filter/cassette sampling assembly should be used with
this tube. See Section 2.1.5. for more details regarding
the pre-filter.
2.1.3. A stopwatch and bubble tube or meter are used to calibrate pumps.
2.1.4. Various lengths of polyvinyl chloride tubing are used to connect sampling tubes to pumps.
2.1.5. Anytime the workplace air being sampled is suspected of containing NaN3, use
the pre-filter/cassette assembly listed below.
a. PVC membrane filter, 37-mm, 5-mm pore size, [part no. 625413, Mine Safety Appliances (MSA), Pittsburgh, PA or cat. no. P-503700, Omega
Specialty Instrument Co., Chelmsford, MA] b. Polystyrene cassette, 37-mm diameter. c. Spacer support pad (cat. no. 225-23, SKC Inc.) (use a spacer in place of a backup pad to hold the PVC filter securely in the
cassette.)
Assemble the pre-filter and sampling tube such that sampled air enters the cassette first. Use a minimum amount of
tubing to connect the sampling tube to the cassette.
2.1.6. Optional: Desorbing solution (0.9 mM Na2CO3 + 0.9 mM
NaHCO3):
Dissolve 0.191 g Na2CO3 and 0.151 g NaHCO3 in 2.0 L
deionized water.
Note: This solution is only used if a delay in sample shipment is expected.
2.2. Equipment - Wipe Samples
Note: Do not use wipe materials such as smear tabs or those composed of cellulose; preliminary tests indicate azide
is unstable on this media (recovery was about 50%). Recoveries of NaN3 spiked on glass fiber or PVC filters
were adequate.
Use either a polyvinyl chloride (PVC) membrane filter, 37-mm, 5-mm pore size, [part no. 625413, Mine Safety Appliances (MSA),
Pittsburgh, PA or cat. no. P-503700, Omega Specialty Instrument Co., Chelmsford, MA] or a glass fiber filter, 37-mm, (part no. 61715,
Gelman Instrument Company, Ann Arbor, MI). Also see the scintillation vial specification in Section
2.3.
2.3. Equipment - Bulk Samples
Scintillation vials, 20-mL (part no. 74515 or 58515, Kimble, Div. of Owens-Illinois Inc., Toledo, OH) with polypropylene or
Teflon® cap liners. If possible, submit bulk or wipe samples in these vials. Tin or other metal cap
liners should not be used because the metal and azide may react.
2.4. Sampling Procedure - Air Samples
Very few industrial operations are conducted where HN3 exists and
NaN3 does not. The tube is used to capture the HN3 while the filter will capture
NaN3. Particulate NaN3 can be captured in the glass wool plug of the tube; however,
a pre-filter is more effective in capturing the particulate.
2.4.1. Connect the cassette/tube assembly to the calibrated sampling pump. Ensure that sampled air will
enter the tube following the direction of the arrow sign (--->) stamped on the outer glass. Place the sampling device on the
employee such that air is sampled from the breathing zone.
2.4.2. Use a flow rate of 1 L/min and a minimum sampling time of 5 min. Take additional samples as necessary.
2.4.3. After sampling, place plastic end caps tightly on both ends of the tube and the filter cassette. Apply OSHA Form 21 seals.
Record the sampling conditions such as sampling time, air volume, etc. on the OSHA 91A form. When other compounds are known or suspected
to be present in the air, record such information and transmit with the samples. See note in Section
2.7., regarding sample shipment.
2.4.4. Use the same lot of ISG tubes and PVC filters for blank and collected samples. Prepare and handle the blank sorbent tube(s) and
filter cassette(s) in exactly the same manner as the sample tubes except that no air is drawn through blanks.
2.5. Sampling Procedure - Wipe Samples for Sodium Azide Particulate
A skin designation has been assigned by OSHA to these azide-containing compounds.
2.5.1. Wear clean, impervious, disposable glove when taking each wipe sample.
2.5.2. DO NOT moisten the wipe PVC or glass fiber filters with deionized water prior to use. Use a dry filter to wipe
for surface contamination of azide compounds.
2.5.3. If possible, wipe a surface area covering 100 cm2.
2.5.4. Fold the wipe filter sample with exposed side in. See note in Section
2.7., regarding sample shipment.
2.5.5. Transfer the wipe sample into a 20-mL scintillation vial and seal with vinyl tape. Securely wrap an OSHA-21
seal length-wise from vial top to bottom.
2.5.6. Prepare a blank wipe sample by placing an unused wipe filter sample in a scintillation vial. Seal the vial as discussed in
Section 2.5.5.
2.6. Sampling Procedure - Bulk Samples
2.6.1. Take a representative sample of the bulk material in the workplace. Transfer the bulk material into a 20-mL scintillation vial and
seal with vinyl or electrical tape. Securely wrap an OSHA-21 seal length-wise from vial top to bottom.
2.6.2. The type of bulk sample should be stated on the OSHA 91A and cross-referenced to the appropriate air sample(s).
2.7. Shipment
Note: If a delay in shipment is anticipated (> 2 days after taking samples), remove the PVC filters from the cassettes and place in
individual scintillation vials. Add 5.0 mL of desorbing solution (Section
2.1.6.) to each scintillation vial containing a PVC filter. Add 10
mL of desorbing solution to each scintillation vial containing a wipe filter sample. Refrigerate any tube samples until shipment.
2.7.1. Submit at least one blank sample with each set of air or wipe samples.
2.7.2. Send the samples to the laboratory as soon as possible with the OSHA 91A paperwork requesting total azide analysis.
2.7.3. Bulk samples should be shipped separately from air samples. They should be accompanied by Material Safety Data Sheets if
possible. Check current shipping restrictions and ship to the laboratory by the appropriate method.
3. Analysis
Note: Upon receipt by the laboratory, all samples are stored under refrigeration (~4°C) until analysis. This includes wipe, filter,
sorbent, and bulk samples. Samples inadvertently stored at room temperature need to be analyzed within 10 days.
3.1. Safety Precautions
3.1.1. Refer to appropriate IC instrument manuals and the Standard Operating Procedure (SOP) for proper instrument operation
(5.14.).
3.1.2. Observe laboratory safety regulations and practices.
3.1.3. Sodium azide is highly toxic and presents a severe explosion hazard if shocked or heated. Use appropriate personal
protective equipment such as safety glasses, goggles, gloves, and lab coat when handling this chemical. Prepare solutions in an
exhaust hood. Store unused solutions in a refrigerator or dispose of properly.
3.2. Equipment
Chromatographic equipment which allows for analyte contact with metal surfaces MAY reduce the amount of azide
present. It is recommended to use equipment in which samples have minimal or no contact with metal surfaces. Analysts should avoid using
metal spatulas when weighing azide compounds, or IC pre-column or columns contaminated with heavy metals.
3.2.1. Ion chromatograph (Model 4000i or 4500i Dionex, Sunnyvale, CA) equipped with a variable UV detector.
3.2.2. Automatic sampler (Dionex Model AS-1) and sample vials (0.5 mL).
3.2.3. Laboratory automation system: Ion chromatograph interfaced to a data reduction system (AutoIon 450, Dionex).
3.2.4. Separator and guard columns, anion (Model HPIC-AS9 and AG9, Dionex).
Note: The pH of the eluant must be maintained between 2-11 and hydroxide ion must not be present in
significant amounts if Dionex AS9 and AG9 columns are used. Irreversible damage to the columns (guard and separator column) will
result.
3.2.5. Disposable syringes (1 mL).
3.2.6. Plastic or Teflon®-coated spatulas used for weighing NaN3.
3.2.7. Miscellaneous volumetric glassware: Micropipettes, 10-mL volumetric flasks, 25-mL Erlenmeyer flasks, graduated cylinders,
and beakers.
3.2.8. Scintillation vials, glass, 20-mL, with polypropylene- or Teflon®-lined caps.
3.2.9. Equipment for eluant degassing (vacuum pump, ultrasonic bath).
3.2.10. Analytical balance (0.01 mg).
3.2.11. Exhaust hood.
3.3. Reagents - All chemicals should be at least reagent grade.
3.3.1. Principal reagents:
CAUTION: |
NaN3 can be a dangerous chemical, and can cause an explosion when shocked or
heated. It is also a skin irritant and a hypotensive agent. Avoid skin contact and handle this chemical and any solutions with care.
Do not dry NaN3 in a drying oven! |
Sodium carbonate (Na2CO3)
Sodium bicarbonate (NaHCO3)
Sodium azide (NaN3)
Hydrochloric acid (HCl)
Deionized water (DI H2O) - specific conductance <10 µS.
3.3.2. Eluent and desorbing solution (0.9 mM Na2CO3 +
0.9 mM NaHCO3):
Dissolve 0.191 g Na2CO3 and 0.151 g NaHCO3 in 2.0 L
DI H2O. Sonicate this solution and degas under vacuum for 15 min. Prepare weekly.
3.3.3. Azide (N3-) stock standard (1,000 µg/mL):
Prepare the azide stock standard in an exhaust hood. Carefully weigh 1.5476 g of NaN3 (Aldrich Chemical Company,
Inc., Milwaukee, WI). Dissolve and dilute to 1.0 L with DI H2O. Prepare monthly.
3.3.4. Azide (N3-) standard solutions (100, 10, and 1 µg/mL):
Perform serial dilutions of the 1,000 µg/mL N3- stock standard using volumetric pipets
and flasks. Dilute to the mark with eluant. Prepare every two weeks. The larger standards (100 and 10 µg/mL) can be used as
working standards, if necessary.
3.3.5. Dispose of azide or azide solutions according to the chemical manufacturer, and local or federal waste disposal guidelines. A
method for disposal of aqueous azide solutions recommended by the Royal Society of Chemistry
(5.15.) is to dilute the solution greatly with water and then run to
waste.
CAUTION: |
Do not dispose of untreated azides or concentrated azide solutions by pouring down sink drains. |
3.4. Working Standard Preparation
3.4.1. Prepare N3- working standards in the ranges specified below:
Working Std (µg/mL)
|
Std Solution (µg/mL)
|
Aliquot (mL)
|
Eluant Added (mL)
|
|
0.05 |
1.0 |
0.5 |
9.5 |
0.10 |
1.0 |
1.0 |
9.0 |
0.20 |
1.0 |
2.0 |
8.0 |
0.50 |
1.0 |
5.0 |
5.0 |
0.75 |
1.0 |
7.5 |
2.5 |
1.00 |
1.0 |
* |
* |
|
* Already prepared in Section
3.3.4. |
3.4.2. To prepare 10 mL of each working standard, pipette an appropriate aliquot (Aliquot column listed above)
of the 1.0 µg/mL standard solution into a scintillation vial or Erlenmeyer flask. Add the specified amount of eluant
(Eluant Added column). As an alternative, pipet each aliquot into a 10-mL volumetric flask and dilute to volume with eluant.
3.5. Sample Preparation - Air Samples
Note: |
Samples desorbed in the field (Section
2.7.)
are ready for analysis (Section 3.7.). |
3.5.1. Remove filter cassette and tube samples from the refrigerator and allow them to warm to room temperature.
3.5.2. Tube Samples:
Carefully remove the end glass wool plug. The sorbent should always be removed from the glass tube via the opposite end of collection
(i.e. backup section is removed first). This will minimize the possibility of contamination from any collected particulate.
3.5.3. Transfer each section of the ISG and glass wool plugs and place in separate 25-mL Erlenmeyer flasks or
20-mL scintillation vials. Place the front glass wool plug and front ISG section (150 mg) in one container and place the middle and
end glass wool plug in another container with the backup ISG section (75 mg).
3.5.4. Pipette 3.0 mL of desorbing solution into each container. Cap each flask tightly and allow the solution to sit for at least 60 min.
Swirl the solution occasionally.
3.5.5. Filter Samples:
Carefully remove each filter from the cassette and place into individual 20-mL scintillation vials. Add 5.0 mL of desorbing solution to
each vial. Cap each vial tightly and allow the solution to sit for at least 60 min. Swirl the solution occasionally.
3.6. Sample Preparation - Wipe and Bulk Samples
Note: Samples desorbed in the field (Section
2.7.) are ready for analysis (Section
3.7.).
3.6.1. Remove wipe and bulk samples from the refrigerator and allow them to warm to room temperature.
3.6.2. Weigh out representational aliquots of bulks.
3.6.3. Carefully transfer wipe samples, and previously weighed aliquots of bulk samples to separate labeled 20-mL scintillation vials and
add 10.0 mL of desorbing solution into each vial. Cap each vial tightly and allow the solution to sit for at least 60 min. Swirl the
solution occasionally.
3.7. Sample Analysis
3.7.1. Pipette a 0.5- to 0.6-mL portion of each standard or sample solution into separate automatic sampler vials. Place a filtercap
into each vial. The large filter portion of the cap should face the solution.
3.7.2. Load the automatic sampler with labeled samples, standards, and blanks.
3.7.3. Set up the ion chromatograph in accordance with the SOP (5.14.)
Typical operating conditions for a Dionex 4500i with a variable wavelength UV detector and an automated sampler are listed below:
Ion Chromatograph |
|
Eluant: |
0.9 mM Na2CO3/0.9 mM NaHCO3 |
Column temperature: |
ambient |
Anion pre-column: |
AG9 |
Anion separator column: |
AS9 |
Variable UV wave length: |
210 nm |
Variable UV output range: |
0.01 AU |
Sample injection loop: |
50 µL |
|
Pump |
|
Pump pressure: |
500 psi |
Flow rate: |
1 mL/min |
|
Chromatogram |
|
Run time: |
12 min |
Peak retention time: |
8 to 9 min for N3- |
3.7.4. Follow the SOP for further instructions regarding analysis (5.14.).
3.8. Calculations
3.8.1. After the analysis is completed, retrieve the peak areas or heights for the azide anion. Obtain hard copies of chromatograms
from a printer. An example chromatogram of a solid sorbent sample collected at an hydrozoic acid concentration of approximately 2 ×
PEL is shown below:
Example chromatogram of a solid sorbent sample collected at an hydrozoic acid concentration of approximately 2 ×
PEL |
|
Sample Name: AZS-124 |
Detector: VDM-1 |
|
REPORT |
VOLUME |
DILUTION |
POINTS |
RATE |
START |
STOP |
AREA |
REJECT |
|
External |
1 |
1 |
3605 |
5Hz |
0.00 |
12.02 |
100000 |
|
Peak No. |
Retention Time (Min) |
|
Component Name |
|
|
--------Peak--------- |
|
1 |
0.77 |
|
28,132 |
1,527,608 |
2 |
1.67 |
|
1,193,355 |
26,810,606 |
3 |
3.48 |
|
612,318 |
19,049,090 |
4 |
5.28 |
|
53,446 |
791,025 |
5 |
5.72 |
|
20,583 |
286,812 |
6 |
8.42 |
|
Azide |
|
196,012 |
4,072,104 |
7 |
9.50 |
|
Nitrate |
|
550,223 |
12,829,040 |
3.8.2. Prepare a concentration-response curve by plotting the peak areas or peak heights versus the concentration of the
N3- standards in µg/mL.
3.8.3. Calculate the air concentration of NaN3 (in mg/m3) for each filter or
sorbent sample:
WSA = (µg/mL N3-)(SV)(GF)SA |
|
Where:
WSA |
= |
Total µg of NaN3 in the sampleWB = Total µg of NaN3 in the blank sample |
µg/mL N3- |
= |
Amount found (from curve) |
SV |
= |
Solution volume (mL) from Section
3.5.3.
|
(GF)SA, NaN3/N3-
|
= |
Gravimetric factor = 1.5476 |
AV |
= |
Air volume (L) |
3.8.4. Calculate the total concentration of NaN3 (in µg) in each wipe or
bulk sample using the appropriate equation:
µg NaN3 = WSA - WB |
(Wipe Samples) |
|
NaN3 % (w/w) = |
WSA × 100% S × F |
|
(Bulk Samples) |
Where:
S |
= |
Sample wt, mg = Aliquot of bulk taken in Section
3.6.2. |
F |
= |
1,000 µg/mg |
3.9. Reporting Results
3.9.1. Add the PVC filter and sorbent results together for each sample. Report this sum result to the industrial hygienist as
mg/m3 NaN3 (total).
Note: |
Vapor phase and particulate results should be combined to determine compliance and to minimize
confusion. Although the vapor phase is a ppm value, the OSHA regulation stipulates "sodium azide" as sodium azide or as
hydrozoic acid (5.13.). The total exposure to both phases needs
to be considered for compliance and the results need to be reported as either total mg/m3
NaN3 or total ppm HN3 to minimize confusion. If it is necessary to determine the
ppm amount of HN3, see the Appendix. |
3.9.2. Wipe sample concentrations are reported as total micrograms or milligrams of NaN3.
3.9.3. Bulk sample results are reported as approximate percent by weight sodium azide. Due to differences in sample matrices between
bulks and analytical standards, bulk results are approximate.
4. Backup Data
This method has been validated for a 5-L, 5-min sample taken at a flow rate of 1 L/min. The method validation was conducted near the
OSHA Ceiling PEL. The sampling media used during the validation consisted of two-section tubes packed with 150-mg of ISG
for the front and also 150 mg for the backup sections. Tubes were obtained commercially from SKC (Lot no. 782, Cat. no. 226-55, SKC Inc.,
Eighty Four, PA).
Note: |
After the validation was completed, the manufacturer reduced the amount of sorbent in the backup
section to 75 mg, and reduced the length of the sampling tube from 110 mm to 70 mm. This change produces a smaller, more convenient
sampling train (pre-filter cassette/sampling tube) and should not affect results. |
The validation consisted of the following experiments and discussion:
- An analysis of 24 spiked samples was performed (8 samples each at 2 ×, 1 ×, and 0.5 × the Ceiling PEL) to evaluate desorption
efficiency (DE).
- A sampling and analysis of 18 samples (6 samples each at 2 ×, 1 ×, and 0.5 × Ceiling PEL) collected from dynamically generated
test atmospheres at 50% RH. to determine bias and overall error.
- A determination of the sampling media collection efficiency at approximately 0.26 ppm
(2 × Ceiling PEL).
- A determination of potential breakthrough.
- An evaluation of storage stability at room (20 to 25°C), refrigerator (0 to 4°C), and freezer (-10 to -14°C) temperatures for
64 collected samples.
- A determination of any significant effects on results when sampling at different humidities.
- A determination of the qualitative and quantitative detection limits.
- Evaluation of a pre-filter/cassette assembly or foam for use during sampling.
- Determination of stability of NaN3 on wipe sampling media.
- Summary.
A generation system was assembled, as shown in
Figure 1, and used for
all experiments except detection limit determinations. All samples were analyzed by IC. All known concentrations of generated test
atmospheres were calculated from impinger samples which contained 1.0 mM Na2CO3/1.0
mM NaHCO3 solutions. These impinger samples were taken side-by-side with any ISG samples.
All results were calculated from concentration-response curves and statistically examined for outliers. In addition, the analysis
(Section 4.1.) and sampling and analysis results
(Section 4.2.) were tested for homogeneity of variance. Possible
outliers were determined using the Treatment of Outliers test (5.16.).
Homogeneity of variance was determined using Bartlett's test (5.17.).
Statistical evaluation was conducted according to the Inorganic Methods Evaluation Protocol
(5.18.). The overall error (OE)
(5.18.) was calculated using the equation:
OEi% = ±(|biasi| + 2CV
i) × 100% (95% confidence level) |
Where i is the respective sample pool being examined.
4.1. Analysis
Twenty-four samples were prepared by adding known amounts of NaN3 (as N3-)
stock solution to the ISG tubes to determine desorption efficiencies (DEs) for the analytical portion of the method.
4.1.1. Procedure: Sampling tubes containing ISG were spiked using a 25-µL syringe (Hamilton Microliter
®/Gastight® Syringe, Hamilton Co., Reno, NV). Spikes were 0.5, 1.0, and 2.0
µg N3-. These levels correspond approximately to 0.5, 1, and 2 times the Ceiling PEL for a 5-L
air sample at a 1-L/min flow rate.
4.2. Sampling and Analysis
To determine the precision and accuracy of the method, known concentrations of HN3 were generated, samples
were collected and then analyzed. A block diagram of the generation system used is shown in
Figure 1.
4.2.1. Procedure:
Test atmospheres of HN3 were generated using a syringe pump (Model 355 syringe pump, Sage Instruments,
Cambridge, MA) and a dynamic generation system. To prepare the atmospheric concentrations, a 1,000 µg/mL azide solution (prepared from
NaN3, EM Science, Cherry Hill, NJ) was used. For each HN3 atmosphere generated, 100
µL of concentrated HCl was added to 10 mL of the azide solution to drive the reaction of NaN3 to
HN3. The HCl/NaN3/H2O solution was immediately loaded into a
10-mL disposable syringe driven between 0.13 and 0.52 mL/h through 60 cm of a Teflon® needle (KF30TF needle,
Hamilton Co., Reno, NV) into a glass mixing chamber. The mixing chamber was connected to a filtered and humidified airstream.
Dynamic generation system
A Miller-Nelson Research Inc. flow, temperature, and humidity control system (Model HCS-301, Monterey, CA) was used to control
and condition the airstream. All generation system fittings and connections were Teflon®. The HN
3 concentrations were varied by either adjusting the dilution airstream volume or the speed of the syringe pump
delivering the azide. The dilution airstream was adjusted using the mass flow controller of the Miller-Nelson system.
The system was set to generate test atmospheres at 50% RH and 25°C.
The total flow rate of the generation system was measured using a dry test meter.
Side-by-side solid-sorbent and impinger samples were taken from the sampling manifold using constant-flow pumps. Alpha 1 pumps
(E.I. Du Pont de Nemours & Co., Wilmington, DE) and Gilian® Gil-Air SC pumps (Gilian Instrument Corp.,
W. Caldwell, NJ) were used for impinger and ISG samples, respectively. For the ISG samples, pump flow rates were approximately 1 L/min
and sampling time was 5 min. For impinger samples, a 1 L/min sampling rate for 15 min was used. Generation system concentrations were
approximately 0.5, 1, and 2 times the OSHA Ceiling PEL.
4.2.2. An independent source was used for NaN3 analytical standard preparations (Aldrich
Chemical Company, Inc., Milwaukee, WI). All samples and standards were analyzed in accordance with Section
3 of this method.
4.2.3. Results: The results are shown in
Table 2)
and spiked sample
(Table 1) results each passed the Bartlett's test and were pooled to
determine a total CV (CVT) for the sampling and analytical method. For the experiments, the pooled coefficients
of variation, bias, and OE are as follows:
CV1 (pooled) = 0.023; |
CV2 (pooled) = 0.051 |
|
CVT (pooled) = 0.052; |
bias = -0.022; |
OE = ±12.6% |
4.3. Collection Efficiency
Procedure: Six commercially-prepared sampling tubes were used for collection at a concentration of approximately 2 times the OSHA
Ceiling PEL for 5 min at 1 L/min (50% RH and 25°C). The amounts of HN3 vapor collected in the first section
(150 mg of sorbent) and second section (150 mg) were determined. The collection efficiency (CE) was calculated by dividing the amount of
HN3 collected in the first section by the total amount of HN3 collected in the first
and second sections.
Results: The results in
Table 3 show a CE of 100%. No
HN3 was found in the second sorbent section for the CE experiment.
4.4. Breakthrough
(Note: Breakthrough is defined as >5% loss of analyte through the sampling media at 50% RH)
Procedure: The same procedure as the CE experiment (Section
4.3.)
was used with two exceptions: The generation concentration was increased to a level approximately 9 times the Ceiling PEL, and samples were
taken at 1 L/min for 30 min.
The amount of breakthrough for each sampling tube was calculated by dividing the amount collected in the second section by the total
amount of HN3 collected in the first and second sections.
Results: No breakthrough of HN3 into the second section was found. Results are shown in
Table 4.
4.5. Storage Stability
Procedure: Two tests were conducted to assess storage stability. The first was a preliminary study of storage at room temperature
(20 to 25°C) after HN3 collection. Twenty-four samples were taken near the OSHA Ceiling PEL of 0.1 ppm. After
collection, all samples were stored under normal laboratory conditions (20 to 25°C) on a lab bench and were not protected from light.
Six samples were initially desorbed and analyzed, then six samples were desorbed and analyzed after various periods of storage (5, 15,
and 32 days).
An additional test was conducted by generating 40 samples (4 room-temperature samples at day 15 were discarded due to analytical problems)
for a temperature-dependent storage stability test, including 4 control samples (used for day 0). The samples were separated into 3 groups
and each group consisted of 4 samples per storage period. A group was stored at either room, refrigerated, or freezer temperature. The same
analytical procedure as the previous storage test was used. Samples were analyzed after 0, 7, 15, and 30 days.
Results: As shown in
Table 5a and the graph below, the results of
the first test show the mean of samples analyzed after 32 days was only 77% of the value of day 0.
Table 5b and the graph below show the results of the second study at
different temperatures. The recovery is only 77% of the value of day 0 after a 30-day storage at room temperature. This drastic change
was not noted for samples stored at refrigerated or freezer temperatures; however, a slight decrease in sample recoveries (93 - 94%)
after 30 days was apparent.
Storage Stability Results |
4.6. Humidity Study
Procedure: A study was conducted to determine any effect on results when samples are collected at different humidities. Samples were
taken using the generation system and procedure described in Section
4.2.
Test atmospheres were generated at 25°C and at approximately 0.5, 1, and 2 times the OSHA Ceiling PEL. Relative humidities of 30%, 50%, and
80% were used at each concentration level tested.
Results: Results of the humidity tests are listed in
Table 6. An
F test was used to determine if any significant effect occurred when sampling at different humidities. As shown, the calculated F values
are less than critical F values (5.19.) for all the concentrations tested
and no significant difference in results occurred across the humidity ranges tested.
4.7. Qualitative and Quantitative Detection Limit Study
Procedure: Low concentration samples were prepared by spiking desorbing solutions
(Section 3.3.2.) with aliquots of aqueous standards prepared from
NaN3 (Section 3.3.4.). These samples were
analyzed using a 50-µL sample injection loop and a variable wavelength UV detector setting of 0.01 absorbance unit (AU). A
derivation of the International Union of Pure and Applied Chemistry (IUPAC) detection limit equation
(5.20.) was used to calculate detection limits.
4.8. Pre-filter Evaluation
Procedure: Past research regarding aerosols (5.21.) has
indicated that particulate in the air sampled may penetrate any glass wool plugs and the sorbent when using conventional sampling tubes.
A pre-filter can be used to assist in capturing particulate before entry into the sampling tube. A study was conducted to
evaluate the possibility of HN3 reacting with a pre-filter/cassette or foam sampling device to
capture NaN3 or other particulate. Evaluations were performed using either ISG sampling tubes with
pre-filter sampling assemblies (PVC filter/spacer support/cassette), or with polyurethane foam [foam used in the combination
sampling device for SO2 (Type II tube), OSHA Method No. ID-200] for particulate collection.
A test was conducted by taking four ISG samples without pre-filters side-by-side with four ISG samples connected with
pre-filters. This test was repeated with foam plugs instead of the pre-filters. Samples were taken such that
the test atmosphere entered the pre-filter or foam first and then entered the ISG. Short pieces of
Tygon® tubing were used to connect the cassettes and ISG sampling tubes. All samples were taken at a flow rate
of about 0.8 L/min for 5 min. The generation system concentration was approximately 1.5 times the Ceiling PEL.
Results:
The results of the comparison of ISG samples taken, with and without pre-filter or foam, are shown in
Tables 8a and
8b, respectively. As shown, a difference in the amount of
HN3 collected was not noted between the pre-filter/ISG and ISG, or foam/ISG. The PVC
pre-filter/cassette or the foam does not appear to inhibit the collection of HN3 when using
the sampling conditions stated in Section 2. The PVC filter/cassette
assembly is recommended for particulate sample collection. The foam may be validated for use in the future to develop a combination
sampling device for collection of both NaN3 and HN3. The ability of the foam to
effectively capture NaN3 needs to be further examined.
4.9. Stability of NaN3 on Wipe Sampling Media
Procedure: A determination of the stability of NaN3 was conducted using 37-mm glass fiber filters
(Cat. no. 61715, Gelman Instrument Company, Ann Arbor, MI) and smear tabs (Lot. no. 3034, Whatman LabSales Inc., Hillsboro, OR). The
stability of sodium azide on PVC membranes has been previously reported as stable up to 10 days of storage
(5.6.). Glass fiber filters or smear tabs were spiked using a
25-µL syringe (Hamilton Microliter®/Gastight® Syringe, Hamilton Co.,
Reno, NV). Solution spikes contained between 7 and 15 µg NaN3. Filters were allowed to dry and
were stored for 3 days on a lab bench, then refrigerated until analysis.
Results: The precision and accuracy results for glass fiber filters and smear tabs are shown below (F/T = Found/Theoretical
recovery):
Collection Media
|
|
N
|
|
Mean (F/T)
|
|
Std Dev
|
|
CV
|
Glass Fiber Filters |
|
5 |
|
1.001 |
|
0.036 |
|
0.036 |
Smear Tabs |
|
5 |
|
0.412 |
|
0.134 |
|
0.325 |
The recovery data shows that azide is unstable on cellulose media and stable on glass fiber filters.
4.10. Summary
The validation results indicate the method meets both the NIOSH and OSHA criteria for accuracy and precision
(5.17.,
5.18.). Performance during collection efficiency, breakthrough, and
humidity tests is adequate. Although it appears that the recovery dramatically decreases when storing collected samples at room
temperature after 15 days, no losses were found when storing the sampling tubes after sample collection in a refrigerator or freezer.
It is recommended to analyze samples within 10 days if samples are stored without refrigeration and within 30 days if refrigeration is
used. Detection limits are adequate when samples are taken for 5 min at 1 L/min. The method is adequate for monitoring occupational
exposures to the OSHA Ceiling PEL.
5. References
5.1. Westwood, L.C. and E.L. Stockes: Determination of Azide in Environmental Samples by Ion Chromatography. Ion Chromatographic Analysis of Environmental Pollutants, Vol. 2, edited by J.J. Mulick and E. Sawicki. Ann Arbor, MI. Ann Arbor Science, 1979. p.141.
5.2. Roberson, C.E. and C.M. Austin: Colorimetric Estimation of Milligram Quantities of Inorganic Azides. Anal. Chem. 29:854-855 (1957).
5.3. Williams, K.E., G.G. Esposito, and D.S. Rinehart: Sampling Tubes for the Collection of Selected Acid Vapors in Air. Am. Ind. Hyg. Assoc. J. 42:476-478 (1981).
5.4. Zehner, J.M. and R.A. Simonaitis: Gas Chromatographic Determination of Hydrazoic Acid. J. Chromatogr. Sci. 14:493-494 (1976).
5.5. Puskar, M.A., S.M. Fergon and L.H. Hecker: A Short-Term Solid Sorbent Determination of Hydrazoic Acid in Air. Am. Ind. Hyg. Assoc. J. 52(1):14-19 (1991).
5.6. National Institute for Occupational Safety and Health: Azide particulates (Method No. P&CAM 369). Cincinnati, OH: National Institute for Occupational Safety and Health, 1982.
5.7. Occupational Safety and Health Administration Salt Lake Technical Center: Azide by Ion Chromatography (Stopgap Method - Unpublished) by J. Germ, Salt Lake City, UT, 1985.
5.8. National Research Council, Committee on Hazardous Substances in the Laboratory: Prudent Practice for Handling Hazardous Chemical in Laboratories, Washington D.C.: National Academy Press, 1981. pp.145-147.
5.9. Kleinhofs, A., W.M. Owais, and R.A. Nilan: Azide. Mutat. Res. 55:165-195 (1978).
5.10. Owais, W.M., A. Kleinhofs, and R.A. Nilan: In Vivo Conversion of Sodium Azide to a Stable Mutagenic Metabolite in Salmonella Typhimuriun. Mutat. Res. 68:15-22 (1979).
5.11. Hawley, G.G., ed. The Condensed Chemical Dictionary, 8th rev. ed. New York: Van Nostrand Reinhold Co., 1971.
5.12. Yost, D.M., and H. Russell, Jr.: Systematic Inorganic Chemistry, New York: Prentice-Hall, Inc., 1946. Ch. 3, pp.122-131.
5.13. "Sodium azide" Federal Register 54:12 (19 Jan. 1989). pp 2540.
5.14. Occupational Safety and Health Administration Salt Lake Technical Center: Ion Chromatography Standard Operating Procedure (Ion Chromatographic Committee). Salt Lake City, UT. In progress.
5.15. Bretherick, L, ed.: Hazards in the Chemical Laboratory, 4th ed. London: Royal Society of Chemistry, 1986. pp. 491-492.
5.16. Mandel, J.: Accuracy and Precision, Evaluation and Interpretation of Analytical Results, The Treatment of Outliers. In Treatise On Analytical Chemistry. 2nd ed., Vol. 1, edited by I. M. Kolthoff and P. J. Elving. New York: John Wiley and Sons, 1978. pp. 282-285.
5.17. National Institute for Occupational Safety and Health: Documentation of the NIOSH Validation Tests by D. Taylor, R. Kupel, and J. Bryant (DHEW/NIOSH Pub. No. 77-185). Cincinnati, OH: National Institute for Occupational Safety and Health, 1977. pp. 1-12.
5.18. Occupational Safety and Health Administration Salt Lake Technical Center: Evaluation Guidelines of the Inorganic Methods Branch. In OSHA Analytical Methods Manual. 2nd ed. Cincinnati, OH: American Conference of Governmental Industrial Hygienists, 1991.
5.19. Dowdy, S. and S. Wearden: Statistics for Research. New York: John Wiley and Sons, 1983. Chapter 8.
5.20. Long, G.L. and J.D. Winefordner: Limit of Detection A Closer Look at the IUPAC Definition. Anal. Chem. 55:712A-724A (1983).
5.21. Fairchild, C.I., and M.I. Tillery: The Filtration Efficiency of Organic Vapor Sampling Tubes against Particulates. Am. Ind. Hyg. Assoc. J. 38:277-283 (1977).
Table 1
Azide (as N3-) Analysis - Desorption Efficiency (DE)
|
(OSHA-PEL) |
Taken (µg N3-) |
Found (µg N3-) |
DE (F/T) |
N |
Mean |
Std Dev |
CV |
|
(0.5 × PEL) |
0.500 0.500 0.500 0.500 0.500 0.500 0.500 0.500
|
0.500 0.515 0.520 0.500 0.505 0.510 0.465 0.525 |
1.000 1.030 1.040 1.000 1.010 1.020 0.930 1.050 |
|
|
8 |
1.010 |
0.037 |
0.037 |
(1 × PEL) |
1.000 1.000 1.000 1.000 1.000 1.000 1.000 1.000 |
0.995 0.985 0.995 1.000 1.015 0.995 0.995 1.000 |
0.995 0.985 0.995 1.000 1.015 0.995 0.995 1.000 |
|
|
8 |
0.998 |
0.008 |
0.008 |
(2 × PEL) |
2.000 2.000 2.000 2.000 2.000 2.000 2.000 2.000 |
1.980 1.960 2.000 1.990 1.975 2.010 2.035 1.970 |
0.990 0.980 1.000 0.995 0.988 1.005 1.018 0.985 |
|
|
8 |
0.995 |
0.012 |
0.012 |
|
F/T = Found/Taken |
|
DE = Desorption Efficiency |
CV1 (Pooled) = 0.023 |
|
Average DE = 1.001 |
|
The average DE is very close to 1.0; therefore, a DE correction is not necessary. |
Table 2
Hydrazoic Acid Sampling and Analysis - Ceiling PEL Determination
* (25°C and 50% RH)
|
(OSHA-PEL) |
Taken (ppm HN3) |
Found (ppm HN3) |
Recovery (F/T) |
N |
Mean |
Std Dev |
CV |
OE (±%) |
|
(0.5 × PEL) |
0.057 0.057 0.057 0.057 0.057 0.057 |
0.053 0.058 0.053 0.061 0.056 0.054 |
0.930 1.018 0.930 1.070 0.982 0.947 |
|
|
6 |
0.980 |
0.056 |
0.057 |
13.4 |
(1 × PEL) |
1.130 1.130 1.130 1.130 1.130 1.130 |
0.129 0.121 0.135 0.117 0.122 0.121 |
0.992 0.931 1.038 0.900 0.938 0.931 |
|
|
6 |
0.955 |
0.050 |
0.053 |
15.1 |
(2 × PEL) |
0.263 0.263 0.263 0.263 0.263 0.263 |
0.259 0.267 0.252 0.251 0.276 0.272 |
0.985 1.015 0.958 0.954 1.049 1.034 |
|
|
6 |
0.999 |
0.040 |
0.040 |
8.01 |
|
F/T |
= |
Found/Taken |
Bias |
= |
-0.022 |
CV2 (Pooled) |
= |
0.051 |
CVT (Pooled) |
= |
0.052 |
Overall Error (Total) |
= |
±12.6% |
|
*Samples were taken for 5 min. |
Table 3
Collection Efficiency (2 × PEL, 25°C & 50% RH)
|
|
------------------ppm HN3 ------------------- |
|
Sample No. |
First Section |
Second Section |
% Collection Efficiency |
|
1 2 3 4 5 6 |
0.259 0.267 0.252 0.251 0.273 0.273 |
ND ND ND ND ND ND |
100.0 100.0 100.0 100.0 100.0 100.0 |
Notes: |
- Sampled at 1 L/min for 5 min.
- Samples desorbed using a sample solution volume of 3.0 mL
- ND = None detectable (<0.001 ppm HN3)
|
|
|
Table 4
Breakthrough Study (25°C and 50% RH)
|
|
------------------ppm HN3 ------------------- |
|
Sample No. |
First Section |
Second Section |
% Breakthrough |
|
1 2 3 4 5 6 7 8 |
0.933 0.940 0.897 0.889 0.938 0.891 0.913 0.925 |
ND ND ND ND ND ND ND ND |
0 0 0 0 0 0 0 |
Notes: |
- Sampled at 1 L/min for 30 min
- Due to the large concentration generated and the analytical sensitivity, the front ISE sections of sampling tubes were desorbed using
larger sample solution volumes of 10.0 mL.
- ND = None detectable (<0.001 ppm HN3)
|
|
|
Table 5a
Preliminary Test Storage Stability - HN3 (1 × PEL, 25°C, and 50% RH)
|
Day |
Air Vol |
Found |
Taken |
--------------Statistical Analysis------------- |
|
(L) |
(--ppm HN3--) |
N |
Mean |
Std Dev |
CV |
Recovery (%) |
|
0 |
4.21 4.21 4.21 4.21 4.21 4.21 |
0.129 0.121 0.135 0.117 0.122 0.121 |
0.130 0.130 0.130 0.130 0.130 0.130 |
|
|
6 |
0.124 |
0.007 |
0.053 |
95.4 |
5 |
4.21 4.21 4.21 4.21 4.21 4.21 |
0.116 0.124 0.116 0.121 0.117 0.114 |
0.130 0.130 0.130 0.130 0.130 0.130 |
|
|
6 |
0.118 |
0.004 |
0.032 |
90.8 |
15 |
3.92 3.92 3.92 3.92 3.92 3.92 |
0.106 0.106 0.107 0.113 0.101 0.101 |
0.130 0.130 0.130 0.130 0.130 0.130 |
|
|
6 |
0.106 |
0.004 |
0.042 |
81.3 |
32 |
4.21 4.21 4.21 4.21 4.21 4.21 |
0.096 0.099 0.104 0.100 0.108 0.092 |
0.130 0.130 0.130 0.130 0.130 0.130 |
|
|
6 |
0.100 |
0.006 |
0.057 |
76.8 |
|
Table 5b
Storage Stability - HN3 at Room, Refrigerated and Freezer Temperatures
(Known HN3 Concentration = 0.108 ppm at 50% RH)
|
Temperature: |
Room
|
Refrigerated
|
Freezer
|
Day |
Air Vol (L) |
|
Found (ppm) |
Air Vol (L) |
Found (ppm) |
Air Vol (L) |
Found (ppm) |
|
0 |
3.05 3.05 3.05 3.05 |
|
0.116 0.120 0.097 0.099 |
* * * * |
* * * * |
* * * * |
* * * * |
|
N Mean Std Dev CV2 Recovery |
= = = = = |
4 0.108 0.012 0.108 100% |
|
* * * * |
|
* * * * |
7 |
3.54 3.54 3.54 3.54 |
|
0.121 0.115 0.095 0.087 |
3.05 3.05 3.05 3.05 |
0.098 0.126 0.105 0.116 |
3.54 3.54 3.54 3.54 |
0.127 0.108 0.098 0.104 |
|
N Mean Std Dev CV2 Recovery |
= = = = = |
4 0.105 0.016 0.154 96.8% |
|
4 0.111 0.012 0.111 103% |
|
4 0.109 0.013 0.115 101% |
15 |
+ + + + |
|
+ + + + |
3.54 3.54 3.54 3.54 |
0.120 0.102 0.100 0.106 |
3.05 3.05 3.05 3.05 |
0.129 0.101 0.102 0.110 |
|
N Mean Std Dev CV2 Recovery |
= = = = = |
+ + + + + |
|
4 0.107 0.009 0.084 99.1% |
|
4 0.111 0.013 0.117 102% |
30 |
3.54 3.54 3.54 3.54 |
|
0.089 0.076 0.085 0.082 |
3.05 3.05 3.05 3.05 |
0.104 0.103 0.102 0.097 |
3.54 3.54 3.54 3.54 |
0.104 0.091 0.104 0.103 |
|
N Mean Std Dev CV2 Recovery |
= = = = = |
4 0.083 0.0055 0.066 76.9% |
|
4 0.102 0.003 0.031 94.0% |
|
4 0.101 0.006 0.063 93.1% |
* Same as day 0 for room temperature + Due to poor precision and analytical difficulties, data are deleted from statistical analysis and are not presented graphically in Section 4.5. |
Table 6
Humidity Test - HN3 (0.5 × PEL & 25°C)
|
% RH
|
30
|
50
|
80
|
ppm HN3 Taken |
0.061 |
0.057 |
0.057 |
ppm HN3 Found |
0.068 0.058 0.063 0.057 0.062 0.058 0.062 0.059 |
0.053 0.058 0.053 0.061 0.056 0.054 |
0.057 0.065 0.047 0.067 0.046 0.051 |
N Mean (ppm) Std Dev (ppm) CV Ave Recovery |
8 0.061 0.004 0.060 100% |
6 0.056 0.003 0.057 98.0% |
6 0.056 0.009 0.163 97.4% |
|
At the 99% confidence level: |
|
|
Fcrit = 6.11 |
Fcalc = 0.12 (2, 17 degrees of freedom) |
|
Fcrit > Fcalc; therefore,
no significant difference in results was noted across the humidity levels tested. |
Humidity Test - HN3 (1 × PEL & 25°C)
|
% RH
|
30
|
50
|
80
|
ppm HN3 Taken |
0.124 |
0.130 |
0.121 |
ppm HN3 Found |
0.122 0.129 0.131 0.125 |
0.129 0.121 0.135 0.117 0.122 0.121 |
0.119 0.124 0.118 0.115 0.115 0.122 |
N Mean (ppm) Std Dev (ppm) CV Ave Recovery |
4 0.127 0.004 0.032 102% |
6 0.124 0.007 0.053 95.5% |
6 0.119 0.004 0.031 98.2% |
|
At the 99% confidence level: |
|
|
Fcrit = 6.70 |
Fcalc = 3.42 (2, 13 degrees of freedom) |
|
Fcrit > Fcalc; therefore, no significant difference in results was noted across the humidity levels tested. |
Humidity Test - HN3 (2 × PEL & 25°C)
|
% RH
|
30
|
50
|
80
|
ppm HN3 Taken |
0.202 |
0.263 |
0.206 |
ppm HN3 Found |
0.203 0.214 0.202 0.210 0.200 0.203 0.215 |
0.259 0.267 0.252 0.251 0.276 0.272 |
0.218 0.199 0.213 0.191 0.182 0.196 |
N Mean (ppm) Std Dev (ppm) CV Ave Recovery |
7 0.207 0.006 0.030 102% |
6 0.263 0.010 0.040 99.9% |
6 0.200 0.014 0.068 97.0% |
|
At the 99% confidence level: |
|
|
Fcrit = 6.23 |
Fcalc = 2.11 (2, 16 degrees of freedom) |
|
Fcrit > Fcalc; therefore, no significant difference in results was noted across the humidity levels tested. |
Table 7
Qualitative and Quantitative Detection Limits (IUPAC Method)
|
|
---------HN3 (as N3-) Level----------- |
|
Sample No.
|
0.02 µg/mL PA
|
0.05 µg/mL PA
|
0.10 µg/mL PA
|
|
1 2 3 4 5 6 7
|
5.07 5.31 4.65 4.78 4.16 4.42 5.61 |
15.19 17.03 14.10 14.87 12.35 14.41 13.44 |
42.23 36.72 40.02 36.29 39.73 40.21 38.18 |
|
N Mean Std Dev CV |
7 4.86 0.507 0.104 |
7 14.48 1.468 0.101 |
7 39.05 2.108 0.054 |
PA = Integrated Peak Area (N3-)/100,000 The blank and 0.01 mg/mL integrated peak areas, and their standard deviations (Std Dev) were all equal to zero. |
|
Using the equation: Cld = k(sd)/m |
Where: |
Cld |
= |
the smallest reliable detectable concentration an analytical instrument can determine at a given confidence level. |
k |
= |
3 (Qualitative Detection Limit, 99.86% Confidence) |
|
= |
10 (Quantitative Detection Limit, 99.99% Confidence) |
sd |
= |
standard deviation of the reagent blank (Rbl) readings. |
m |
= |
analytical sensitivity or slope as calculated by linear regression. |
Cld |
= |
3(0.507)/438.6 = 0.00347 µg/mL as N3- for the qualitative limit. |
Cld |
= |
10(0.507)/438.6 = 0.01156 µg/mL as N3- for the quantitative limit. |
Qualitative detection limit = 0.0104 µg N3- (3-mL sample volume) or 0.001 ppm HN3 (5-L air volume).
Quantitative detection limit = 0.0348 µg N3- (3-mL sample volume) or 0.004 ppm HN3 (5-L air volume). |
|
Table 8a
Comparison Study - With/Without Pre-filter (Known Concentration = 0.131 ppm HN3) (25°C, and 50% RH)
|
|
With Pre-filter
|
Without Pre-filter
|
Sample Set #
|
Air Vol, L
|
ppm HN3 Found
|
Air Vol, L
|
ppm HN3 Found
|
1 2 3 4 |
3.41 3.41 3.41 3.41 |
0.124 0.140 0.134 0.098* |
3.38 3.38 3.38 3.38 |
0.141 0.138 0.131 0.114 |
N Mean Std Dev CV Recovery |
|
3 0.133 0.008 0.061 101% |
|
4 0.131 0.012 0.092 100% |
* Outlier |
|
Notes: |
- A 37-mm PVC filter was used as the pre-filter in a polystyrene cassette.
- Sampling Time = 5 min
- Flow Rate = 0.825 to 0.829 L/min
- Sample Solution Volume for Desorption = 3.0 mL
|
|
|
At the 99% confidence level: |
|
tcrit = 9.92 tcalc = -0.615 (2 degrees of freedom) |
|
tcrit > tcalc; therefore, no significant difference in results was noted across the two sets tested. |
|
Table 8b
Comparison Study - With/Without Foam (Known Concentration = 0.141 ppm HN3) (25°C, and 50% RH)
|
|
With Foam
|
Without Foam
|
Sample Set #
|
Air Vol, L
|
ppm HN3 Found
|
Air Vol, L
|
ppm HN3 Found
|
1 2 3 4 |
3.35 3.35 3.35 3.35 |
0.129 0.140 0.138 0.155 |
3.43 3.43 3.43 3.43 |
0.155 0.135 0.140 0.154 |
N Mean Std Dev CV Recovery |
|
4 0.141 0.011 0.077 100% |
|
4 0.146 0.010 0.069 104% |
Notes: |
- Type II containing 150 mg-ISG glass jacket was used. The dimensions of the front portion of the glass jacket are 12-mm o.d., 10-mm i.d., and 25-mm long and is used for collecting azide particulate. The second part of the glass tube contains ISG and is used for collecting HN3. The dimensions of the second part are 6-mm o.d., 4-mm i.d., and 50-mm long. Both ends of the sampling tube are sealed with plastic caps (see Method No. ID-200 for a graphic description of the Type II glass jacket used).
- Foam analyzed after sampling contained 0.004 ppm as HN3 (average).
- Sampling Time = 5 min
- Flow Rate = 0.627 to 0.787 L/min
- Sample Solution Volume for Desorption = 3.0 mL
|
|
|
At the 99% confidence level: |
|
tcrit = 5.84 tcalc = -0.0055 (3 degrees of freedom) |
|
tcrit > tcalc; therefore, no significant difference in results was noted across the two sets tested. |
|
The system shown above was used to generate dynamic test atmospheres. The system consists of four essential
elements:
- A flow-temperature-humidity control system,
- An HN3 vapor generating system,
- A mixing chamber, and
- An active sampling manifold.
Appendix
WHA = (µg/mL N3-)(SV)(GF)HA
ppm HN3 = |
WHA × MV
AV × MW |
|
Where:
AV |
= |
Air volume |
WHA |
= |
µg HN3 |
(GF)HA, HN3/N3- |
= |
Gravimetric factor = 1.0238 |
MV |
= |
Molar volume (L/mol) = 24.45 (25°C and 760 mmHg) |
MW |
= |
Molecular weight for HN3 = 43.0 (g/mol) |
SV |
= |
Solution volume |
µg/mL N3- |
= |
Sample result taken from concentration-response curve |
|
|
|
|