Analytical Methods  Approved for Drinking  Water  Compliance
\^ ^
'r0 Monitoring under the 1 otal Colilorm Rule
S Analysis for the following contaminants shall be conducted in accordance with the methods in the following table or their equivalent as determined by
^T EPA. The methods and monitoring requirements for these contaminants are specified in 40 CFR 141.21. Additional methods are listed in Appendix A to
^ SubpartCofPartl41.
The CFR is the legal reference for approved methods and takes precedent over this table. The table should accurately reflect the analytical methods
information published in 40 CFR 141. If you find discrepancies, please notify The Safe Drinking Water Hotline (800-426-4791) so that EPA can
correct the table.
Contaminant EPA Publication
OUIIldll Illdlll Mothis\ri „ 	 ^ .
Method Publication Order
Method Organization ReferenceTitle Date Number Number Source of Method
Total Coliforms
The time from sample collection to initiation of analysis may not exceed 30 hours. Systems are encouraged but not required to hold samples below 10°C during
  9221 A
  9221 A
                       transit.
                       EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms.  EPA also
                       encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
                       intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used.  The Agency suggests that laboratories perform
                       these studies on a minimum of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
                       Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test.  Methods for establishing false-
                       positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase and cytochrome oxidase, multi-test identification systems, or
                       equivalent confirmation tests. False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 18th Edition
1992
Standard Methods
                             Lactose broth, as commercially available, may be used in lieu of lauryl tryptose broth, if the system conducts at least 25 parallel tests between this medium
                             and lauryl tryptose broth using the water normally tested, and this comparison demonstrates that the false-positive rate and false-negative rate for total
                             coliform, using lactose broth, is less than 10 percent.
                             If inverted tubes are used to detect gas production, the media should cover these tubes at least one-half to two-thirds after the sample is added.
                             No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed tubes.
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 19th Edition
1995
Standard Methods
                             Lactose broth, as commercially available, may be used in lieu of lauryl tryptose broth, if the system conducts at least 25 parallel tests between this medium
                             and lauryl tryptose broth using the water normally tested, and this comparison demonstrates that the false-positive rate and false-negative rate for total
                             coliform, using lactose broth, is less than 10 percent.
                             If inverted tubes are used to detect gas production, the media should cover these tubes at least one-half to two-thirds after the sample is added.
                             No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed tubes.
Revised June 2008
                          Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                                                                                  Page 1 of 18

-------
Contaminant
  Method          Organization      ReferenceTitle
                                                              Method
                                                              Date
                                                          EPA
                                                          Publication
                                                          Number
                                   Publication
                                   Order
                                   Number
                                                                                                                  Source of Method
Total Coliforms
  9221 A
  9221 A
   The time from sample collection to initiation of analysis may not exceed 30 hours. Systems are encouraged but not required to hold samples below 10°C during
   transit.
   EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms.  EPA also
   encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
   intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used. The Agency suggests that laboratories perform
   these studies on a minimum  of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
   Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test. Methods for establishing false-
   positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase and cytochrome oxidase, multi-test identification systems, or
   equivalent confirmation tests. False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 20th Edition
1998
Standard Methods
                               Lactose broth, as commercially available, may be used in lieu of lauryl tryptose broth, if the system conducts at least 25 parallel tests between this medium
                               and lauryl tryptose broth using the water normally tested, and this comparison demonstrates that the false-positive rate and false-negative rate for total
                               coliform, using lactose broth, is less than 10 percent.
                               If inverted tubes are used to detect gas production, the media should cover these tubes at least one-half to two-thirds after the sample is added.
                               No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed tubes.
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 21 st Edition
2005
Standard Methods
                               Lactose broth, as commercially available, may be used in lieu of lauryl tryptose broth, if the system conducts at least 25 parallel tests between this medium
                               and lauryl tryptose broth using the water normally tested, and this comparison demonstrates that the false-positive rate and false-negative rate for total
                               coliform, using lactose broth, is less than 10 percent.
                               If inverted tubes are used to detect gas production, the media should cover these tubes at least one-half to two-thirds after the sample is added.
                               No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed tubes.

  9221A-99          Standard           Online version of Standard Methods for the                                                        http://www.standardmethods.org/
                      Methods           Examination of Water and Wastewater.
                      Online             Approval year by Standard Methods
                                         Committee is designated by  last 2 digits.
                                         This is the only online version that is
                                         approved.
                               Lactose broth, as commercially available, may be used in lieu of lauryl tryptose broth, if the system conducts at least 25 parallel tests between this medium
                               and lauryl tryptose broth using the water normally tested, and this comparison demonstrates that the false-positive rate and false-negative rate for total
                               coliform, using lactose broth, is less than 10 percent.
                               If inverted tubes are used to detect gas production, the media should cover these tubes at least one-half to two-thirds after the sample is added.
                               No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed tubes.
 Revised June 2008
                           Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                                                                                        Page 2 of 18

-------
Contaminant
  Method          Organization      ReferenceTitle
                                                              Method
                                                              Date
                                                          EPA
                                                          Publication
                                                          Number
                                   Publication
                                   Order
                                   Number
                                                                                                                 Source of Method
Total Coliforms
  9221 B
  9221 B
  9221 B
   The time from sample collection to initiation of analysis may not exceed 30 hours. Systems are encouraged but not required to hold samples below 10°C during
   transit.
   EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms. EPA also
   encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
   intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used. The Agency suggests that laboratories perform
   these studies on a minimum  of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
   Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test. Methods for establishing false-
   positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase and cytochrome oxidase, multi-test identification systems, or
   equivalent confirmation tests. False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 18th Edition
1992
Standard Methods
                               Lactose broth, as commercially available, may be used in lieu of lauryl tryptose broth, if the system conducts at least 25 parallel tests between this medium
                               and lauryl tryptose broth using the water normally tested, and this comparison demonstrates that the false-positive rate and false-negative rate for total
                               coliform, using lactose broth, is less than 10 percent.
                               If inverted tubes are used to detect gas production, the media should cover these tubes at least one-half to two-thirds after the sample is added.
                               No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed tubes.
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 19th Edition
1995
Standard Methods
                               Lactose broth, as commercially available, may be used in lieu of lauryl tryptose broth, if the system conducts at least 25 parallel tests between this medium
                               and lauryl tryptose broth using the water normally tested, and this comparison demonstrates that the false-positive rate and false-negative rate for total
                               coliform, using lactose broth, is less than 10 percent.
                               If inverted tubes are used to detect gas production, the media should cover these tubes at least one-half to two-thirds after the sample is added.
                               No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed tubes.
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 20th Edition
1998
Standard Methods
                               Lactose broth, as commercially available, may be used in lieu of lauryl tryptose broth, if the system conducts at least 25 parallel tests between this medium
                               and lauryl tryptose broth using the water normally tested, and this comparison demonstrates that the false-positive rate and false-negative rate for total
                               coliform, using lactose broth, is less than 10 percent.
                               If inverted tubes are used to detect gas production, the media should cover these tubes at least one-half to two-thirds after the sample is added.
                               No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed tubes.
Revised June 2008
                           Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                                                                                        Page 3 of 18

-------
Contaminant
  Method          Organization      ReferenceTitle
                                                              Method
                                                              Date
                                                         EPA
                                                         Publication
                                                         Number
                                   Publication
                                   Order
                                   Number
                                                                                                                 Source of Method
Total Coliforms
  9221 B
  9221 D
   The time from sample collection to initiation of analysis may not exceed 30 hours. Systems are encouraged but not required to hold samples below 10°C during
   transit.
   EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms.  EPA also
   encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
   intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used.  The Agency suggests that laboratories perform
   these studies on a minimum of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
   Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test. Methods for establishing false-
   positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase  and cytochrome oxidase, multi-test identification systems, or
   equivalent confirmation tests. False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 21 st Edition
2005
Standard Methods
                               Lactose broth, as commercially available, may be used in lieu of lauryl tryptose broth, if the system conducts at least 25 parallel tests between this medium
                               and lauryl tryptose broth using the water normally tested, and this comparison demonstrates that the false-positive rate and false-negative rate for total
                               coliform, using lactose broth, is less than 10 percent.
                               If inverted tubes are used to detect gas production, the media should cover these tubes at least one-half to two-thirds after the sample is added.
                               No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed tubes.

  9221 B-99           Standard           Online version of Standard Methods for the                                                        http://www.standardmethods.org/
                      Methods           Examination of Water and Wastewater.
                      Online             Approval year by Standard Methods
                                         Committee is designated by last 2 digits.
                                         This is the only online version that is
                                         approved.
                               Lactose broth, as commercially available, may be used in lieu of lauryl tryptose broth, if the system conducts at least 25 parallel tests between this medium
                               and lauryl tryptose broth using the water normally tested, and this comparison demonstrates that the false-positive rate and false-negative rate for total
                               coliform, using lactose broth, is less than 10 percent.
                               If inverted tubes are used to detect gas production, the media should cover these tubes at least one-half to two-thirds after the sample is added.
                               No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed tubes.
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 18th Edition
1992
Standard Methods
                               No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed tubes.
                               Six-times formulation strength may be used if the medium is filter-sterilized rather than autoclaved.
  9221 D
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 19th Edition
1995
Standard Methods
                               No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed tubes.
                               Six-times formulation strength may be used if the medium is filter-sterilized rather than autoclaved.
Revised June 2008
                           Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                                                                                        Page 4 of 18

-------
Contaminant
  Method         Organization      ReferenceTitle
                                                             Method
                                                             Date
                                                         EPA
                                                         Publication
                                                         Number
                                  Publication
                                  Order
                                  Number
                                                                                                                Source of Method
Total Coliforms
  9221 D
   The time from sample collection to initiation of analysis may not exceed 30 hours. Systems are encouraged but not required to hold samples below 10°C during
   transit.
   EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms.  EPA also
   encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
   intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used.  The Agency suggests that laboratories perform
   these studies on a minimum of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
   Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test. Methods for establishing false-
   positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase  and cytochrome oxidase, multi-test identification systems, or
   equivalent confirmation tests. False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 20th Edition
1998
Standard Methods
                              No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed tubes.
                              Six-times formulation strength may be used if the medium is filter-sterilized rather than autoclaved.
  9221 D
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 21 st Edition
2005
Standard Methods
                              No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed tubes.
                              Six-times formulation strength may be used if the medium is filter-sterilized rather than autoclaved.
  9221 D-99           Standard            Online version of Standard Methods for the
                      Methods            Examination of Water and Wastewater.
                      Online             Approval year by Standard Methods
                                         Committee is designated by last 2 digits.
                                         This is the only online version that is
                                         approved.
                              No requirement exists to run the completed phase on 10 percent of all total coliform-positive confirmed tubes.
                              Six-times formulation strength may be used if the medium is filter-sterilized rather than autoclaved.
                                                                                                                http://www.standardmethods.org/
  9222 A
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 18th Edition
1992
Standard Methods
                              MI agar also may be used.  Preparation and use of MI agar is set forth in the article, "New medium for the simultaneous detection of total coliform and
                              Escherichia coli in water" by Brenner, K.P., etal, 1993, Appl. Environ. Microbiol. 59:3534-3544. EPA/600/J-99/225. Available at:
                              http://www.epa.gov/nerlcwww/online.htm. Verification of colonies is not required.
                              Coliscan® is approved as a modification of MI under the ATP program. It is available from Micro logy Laboratories, P.O. Box 340, Goshen, IN 46527-0340.
Revised June 2008
                           Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                                                                                      Page 5 of 18

-------
Contaminant
  Method         Organization      ReferenceTitle
                                                             Method
                                                             Date
                                                         EPA
                                                         Publication
                                                         Number
                                  Publication
                                  Order
                                  Number
                                                                                                                Source of Method
Total Coliforms
  9222 A
  9222 A
  9222 A
   The time from sample collection to initiation of analysis may not exceed 30 hours. Systems are encouraged but not required to hold samples below 10°C during
   transit.
   EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms.  EPA also
   encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
   intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used.  The Agency suggests that laboratories perform
   these studies on a minimum of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
   Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test. Methods for establishing false-
   positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase  and cytochrome oxidase, multi-test identification systems, or
   equivalent confirmation tests. False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 19th Edition
1995
Standard Methods
                              MI agar also may be used.  Preparation and use of MI agar is set forth in the article, "New medium for the simultaneous detection of total coliform and
                              Escherichia coli in water" by Brenner, K.P., et al., 1993, Appl. Environ. Microbiol. 59:3534-3544. EPA/600/J-99/225. Available at:
                              http://www.epa.gov/nerlcwww/online.htm. Verification of colonies is not required.
                              Coliscan® is approved as a modification of MI under the ATP program. It is available from Micro logy Laboratories, P.O. Box 340, Goshen, IN 46527-0340.
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 20th Edition
1998
Standard Methods
                              MI agar also may be used.  Preparation and use of MI agar is set forth in the article, "New medium for the simultaneous detection of total coliform and
                              Escherichia coli in water" by Brenner, K.P., et al., 1993, Appl. Environ. Microbiol. 59:3534-3544. EPA/600/J-99/225. Available at:
                              http://www.epa.gov/nerlcwww/online.htm. Verification of colonies is not required.
                              Coliscan® is approved as a modification of MI under the ATP program. It is available from Micro logy Laboratories, P.O. Box 340, Goshen, IN 46527-0340.
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 21 st Edition
2005
Standard Methods
                              MI agar also may be used.  Preparation and use of MI agar is set forth in the article, "New medium for the simultaneous detection of total coliform and
                              Escherichia coli in water" by Brenner, K.P., et al., 1993, Appl. Environ. Microbiol. 59:3534-3544. EPA/600/J-99/225. Available at:
                              http://www.epa.gov/nerlcwww/online.htm. Verification of colonies is not required.
                              Coliscan® is approved as a modification of MI under the ATP program. It is available from Micro logy Laboratories, P.O. Box 340, Goshen, IN 46527-0340.
Revised June 2008
                           Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                                                                                      Page 6 of 18

-------
Contaminant
  Method         Organization      ReferenceTitle
                                                                                 Method
                                                                                 Date
                                                         EPA
                                                         Publication
                                                         Number
                                  Publication
                                  Order
                                  Number
                                                                                                                                    Source of Method
Total Coliforms
                      The time from sample collection to initiation of analysis may not exceed 30 hours.  Systems are encouraged but not required to hold samples below 10°C during
                      transit.
                      EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms. EPA also
                      encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
                      intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used. The Agency suggests that laboratories perform
                      these studies on a minimum of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
                      Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test. Methods for establishing false-
                      positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase and cytochrome oxidase, multi-test identification systems, or
                      equivalent confirmation tests.  False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).

9222 A-97           Standard           Online version of Standard Methods for the                                                      http://www.standardmethods.org/
                    Methods           Examination of Water and Wastewater.
                    Online            Approval year by Standard Methods
                                      Committee is designated by last 2 digits.
                                      This is the only online version that is
                                      approved.
                            MI agar also may be used.  Preparation and use of MI agar is set forth in the article, "New medium for the simultaneous detection of total coliform and
                            Escherichia coli in water" by Brenner, K.P., et al., 1993, Appl. Environ. Microbiol. 59:3534-3544. EPA/600/J-99/225.  Available at:
                            http://www.epa.gov/nerlcwww/online.htm. Verification of colonies is not required.
                            Coliscan® is approved as a modification of MI under the ATP program. It is available from Micro logy Laboratories, P.O. Box 340, Goshen, IN 46527-0340.
  9222 B
  9222 B
                    Standard
                    Methods
Standard Methods for the Examination of
Water and Wastewater, 18th Edition
1992
Standard Methods
                              MI agar also may be used.  Preparation and use of MI agar is set forth in the article, "New medium for the simultaneous detection of total coliform and
                              Escherichia coli in water" by Brenner, K.P., et al., 1993, Appl. Environ. Microbiol. 59:3534-3544. EPA/600/J-99/225. Available at:
                              http://www.epa.gov/nerlcwww/online.htm. Verification of colonies is not required.
                              Coliscan® is approved as a modification of MI under the ATP program. It is available from Micro logy Laboratories, P.O. Box 340, Goshen, IN 46527-0340.
                    Standard
                    Methods
Standard Methods for the Examination of
Water and Wastewater, 19th Edition
1995
Standard Methods
                              MI agar also may be used.  Preparation and use of MI agar is set forth in the article, "New medium for the simultaneous detection of total coliform and
                              Escherichia coli in water" by Brenner, K.P., et al., 1993, Appl. Environ. Microbiol. 59:3534-3544. EPA/600/J-99/225. Available at:
                              http://www.epa.gov/nerlcwww/online.htm. Verification of colonies is not required.
                              Coliscan® is approved as a modification of MI under the ATP program. It is available from Micro logy Laboratories, P.O. Box 340, Goshen, IN 46527-0340.
Revised June 2008
                                               Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                                                                                      Page 7 of 18

-------
Contaminant
  Method         Organization      ReferenceTitle
                                                             Method
                                                             Date
                                                         EPA
                                                         Publication
                                                         Number
                                  Publication
                                  Order
                                  Number
                                                                                                                Source of Method
Total Coliforms
  9222 B
  9222 B
   The time from sample collection to initiation of analysis may not exceed 30 hours. Systems are encouraged but not required to hold samples below 10°C during
   transit.
   EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms.  EPA also
   encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
   intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used.  The Agency suggests that laboratories perform
   these studies on a minimum of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
   Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test. Methods for establishing false-
   positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase and cytochrome oxidase, multi-test identification systems, or
   equivalent confirmation tests. False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 20th Edition
1998
Standard Methods
                              MI agar also may be used.  Preparation and use of MI agar is set forth in the article, "New medium for the simultaneous detection of total coliform and
                              Escherichia coli in water" by Brenner, K.P., et al., 1993, Appl. Environ. Microbiol. 59:3534-3544. EPA/600/J-99/225. Available at:
                              http://www.epa.gov/nerlcwww/online.htm. Verification of colonies is not required.
                              Coliscan® is approved as a modification of MI under the ATP program. It is available from Micro logy Laboratories, P.O. Box 340, Goshen, IN 46527-0340.
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 21 st Edition
2005
Standard Methods
                              MI agar also may be used.  Preparation and use of MI agar is set forth in the article, "New medium for the simultaneous detection of total coliform and
                              Escherichia coli in water" by Brenner, K.P., et al., 1993, Appl. Environ. Microbiol. 59:3534-3544. EPA/600/J-99/225. Available at:
                              http://www.epa.gov/nerlcwww/online.htm. Verification of colonies is not required.
                              Coliscan® is approved as a modification of MI under the ATP program. It is available from Micro logy Laboratories, P.O. Box 340, Goshen, IN 46527-0340.


  9222 B-97           Standard            Online version of Standard Methods for the                                                       http://www.standardmethods.org/
                      Methods            Examination of Water and Wastewater.
                      Online             Approval year by Standard Methods
                                         Committee is designated by last 2 digits.
                                         This is the only online version that is
                                         approved.
                              MI agar also may be used.  Preparation and use of MI agar is set forth in the article, "New medium for the simultaneous detection of total coliform and
                              Escherichia coli in water" by Brenner, K.P., et al., 1993, Appl. Environ. Microbiol. 59:3534-3544. EPA/600/J-99/225. Available at:
                              http://www.epa.gov/nerlcwww/online.htm. Verification of colonies is not required.
                              Coliscan® is approved as a modification of MI under the ATP program. It is available from Micro logy Laboratories, P.O. Box 340, Goshen, IN 46527-0340.
 Revised June 2008
                           Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                                                                                      Page 8 of 18

-------
Contaminant
  Method         Organization      ReferenceTitle
                                                             Method
                                                             Date
                                                         EPA
                                                         Publication
                                                         Number
                                  Publication
                                  Order
                                  Number
                                                                                                                Source of Method
Total Coliforms
  9222 C
  9222 C
  9222 C
  9222 C
   The time from sample collection to initiation of analysis may not exceed 30 hours. Systems are encouraged but not required to hold samples below 10°C during
   transit.
   EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms.  EPA also
   encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
   intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used.  The Agency suggests that laboratories perform
   these studies on a minimum of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
   Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test. Methods for establishing false-
   positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase and cytochrome oxidase, multi-test identification systems, or
   equivalent confirmation tests. False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 18th Edition
1992
Standard Methods
                              MI agar also may be used.  Preparation and use of MI agar is set forth in the article, "New medium for the simultaneous detection of total coliform and
                              Escherichia coli in water" by Brenner, K.P., et al., 1993, Appl. Environ. Microbiol. 59:3534-3544. EPA/600/J-99/225. Available at:
                              http://www.epa.gov/nerlcwww/online.htm. Verification of colonies is not required.
                              Coliscan® is approved as a modification of MI under the ATP program. It is available from Micro logy Laboratories, P.O. Box 340, Goshen, IN 46527-0340.
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 19th Edition
1995
Standard Methods
                              MI agar also may be used.  Preparation and use of MI agar is set forth in the article, "New medium for the simultaneous detection of total coliform and
                              Escherichia coli in water" by Brenner, K.P., et al., 1993, Appl. Environ. Microbiol. 59:3534-3544. EPA/600/J-99/225. Available at:
                              http://www.epa.gov/nerlcwww/online.htm. Verification of colonies is not required.
                              Coliscan® is approved as a modification of MI under the ATP program. It is available from Micro logy Laboratories, P.O. Box 340, Goshen, IN 46527-0340.
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 20th Edition
1998
Standard Methods
                              MI agar also may be used.  Preparation and use of MI agar is set forth in the article, "New medium for the simultaneous detection of total coliform and
                              Escherichia coli in water" by Brenner, K.P., et al., 1993, Appl. Environ. Microbiol. 59:3534-3544. EPA/600/J-99/225. Available at:
                              http://www.epa.gov/nerlcwww/online.htm. Verification of colonies is not required.
                              Coliscan® is approved as a modification of MI under the ATP program. It is available from Micro logy Laboratories, P.O. Box 340, Goshen, IN 46527-0340.
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 21 st Edition
2005
Standard Methods
                              MI agar also may be used.  Preparation and use of MI agar is set forth in the article, "New medium for the simultaneous detection of total coliform and
                              Escherichia coli in water" by Brenner, K.P., et al., 1993, Appl. Environ. Microbiol. 59:3534-3544. EPA/600/J-99/225. Available at:
                              http://www.epa.gov/nerlcwww/online.htm. Verification of colonies is not required.
                              Coliscan® is approved as a modification of MI under the ATP program. It is available from Micro logy Laboratories, P.O. Box 340, Goshen, IN 46527-0340.
Revised June 2008
                           Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                                                                                       Page 9 of 18

-------
Contaminant
  Method          Organization      ReferenceTitle
                                                                                Method
                                                                                Date
                                                        EPA
                                                        Publication
                                                        Number
                                  Publication
                                  Order
                                  Number
                                                                                                                                  Source of Method
Total Coliforms
                      The time from sample collection to initiation of analysis may not exceed 30 hours.  Systems are encouraged but not required to hold samples below 10°C during
                      transit.
                      EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms. EPA also
                      encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
                      intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used.  The Agency suggests that laboratories perform
                      these studies on a minimum of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
                      Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test. Methods for establishing false-
                      positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase and cytochrome oxidase, multi-test identification systems, or
                      equivalent confirmation tests. False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).

9222 C-97          Standard           Online version of Standard Methods for the                                                      http://www.standardmethods.org/
                   Methods           Examination of Water and Wastewater.
                   Online             Approval year by Standard Methods
                                      Committee is designated by last 2 digits.
                                      This is the only online version that is
                                      approved.
                            MI agar also may be used.  Preparation and use of MI agar is set forth in the article, "New medium for the simultaneous detection of total coliform and
                            Escherichia coli in water" by Brenner, K.P., et al., 1993, Appl. Environ. Microbiol. 59:3534-3544. EPA/600/J-99/225. Available at:
                            http://www.epa.gov/nerlcwww/online.htm. Verification of colonies is not required.
                            Coliscan® is approved as a modification of MI under the ATP program. It is available from Micro logy Laboratories, P.O. Box 340, Goshen, IN 46527-0340.
9223
Standard Standard Methods for the Examination of 1 992
Methods Water and Wastewater, 18th Edition
The ONPG-MUG Test is also known as the Autoanalysis Colilert System.
Standard Methods
  9223
                   Standard
                   Methods
Standard Methods for the Examination of
Water and Wastewater, 19th Edition
1995
Standard Methods
                              The ONPG-MUG Test is also known as the Autoanalysis Colilert System.
  9223
                   Standard
                   Methods
Standard Methods for the Examination of
Water and Wastewater, 20th Edition
1998
Standard Methods
                              The ONPG-MUG Test is also known as the Autoanalysis Colilert System.
  9223
                   Standard
                   Methods
Standard Methods for the Examination of
Water and Wastewater, 21 st Edition
2005
Standard Methods
                              The ONPG-MUG Test is also known as the Autoanalysis Colilert System.
Revised June 2008
                                              Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                                                                                    Page 10 of 18

-------
Contaminant
  Method          Organization      ReferenceTitle
                                                         Method
                                                         Date
EPA
Publication
Number
Publication
Order
Number
                                                                                                           Source of Method
Total Coliforms
The time from sample collection to initiation of analysis may not exceed 30 hours. Systems are encouraged but not required to hold samples below 10°C during
transit.
EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms.  EPA also
encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used. The Agency suggests that laboratories perform
these studies on a minimum of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test. Methods for establishing false-
positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase and cytochrome oxidase, multi-test identification systems, or
equivalent confirmation tests. False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).
  9223 B-97          Standard           Online version of Standard Methods for the
                     Methods           Examination of Water and Wastewater.
                     Online             Approval year by Standard Methods
                                        Committee is designated by last 2 digits.
                                        This is the only online version that is
                                        approved.
                              The ONPG-MUG Test is also known as the Autoanalysis Colilert System.
                                                                                                           http: //www. standardmethods.org/
  Chromocult®       EM Science        Chromocult® Coliform Agar
  Coliform Agar                         Presence/Absence Membrane Filter Test
                                        Method for Detection and Identification of
                                        Coliform Bacteria and Escherichia coli in
                                        Finished Waters, Version 1.0
                                                         November
                                                         2000
                                    EMD Chemicals
Colisure Test


Colitag®




IDEXX
Laboratories,
Inc.
CPI
International,
Inc.


Colisure Test


Colitag® Product as a Test for Detection and
Identification of Coliforms and E. coli
Bacteria in Drinking Water and Source Water
as Required in National Primary Drinking
Water Regulations
February IDEXX Laboratories, Inc.
28, 1994

August CPI International, Inc.
2001



  E*Colite® Test     Charm             Presence/Absence for Coliforms and E. coli   December
                     Sciences, Inc.       in Water                                  21,1997
                                                                                                           Charm Sciences, Inc
Revised June 2008
                        Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                            Page 11 of 18

-------
Contaminant
  Method          Organization      ReferenceTitle
                                                            Method
                                                            Date
                                                        EPA
                                                        Publication
                                                        Number
Publication
Order
Number
                                                                                                              Source of Method
Total Coliforms
  m-ColiBlue24®
  Test
   The time from sample collection to initiation of analysis may not exceed 30 hours. Systems are encouraged but not required to hold samples below 10°C during
   transit.
   EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms.  EPA also
   encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
   intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used. The Agency suggests that laboratories perform
   these studies on a minimum of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
   Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test. Methods for establishing false-
   positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase and cytochrome oxidase, multi-test identification systems, or
   equivalent confirmation tests.  False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).
Hach Co.
m-ColiBlue 24 Test, "Total Coliforms and£.  August 17,
coli Membrane Filtration Method with m-      1999
ColiBlue 24 Broth," Method No. 10029,
Revision 2.
                Hach Company
  Ready cult®
  Coliforms 100
  Presence/Absence
   Test
EM Science         Ready cult® Coliforms 100 Presence/Absence  November                                         EMD Chemicals
                   Test for Detection and Identification of       2000
                   Coliform Bacteria and Escherichia coli in
                   Finished Waters," Versonl.O
         Fluorocult® is approved as an acceptable version of Ready cult® under the ATP program. It is available from EMD Chemicals (formerly EM Science).
Revised June 2008
                           Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                                                                                    Page 12 of 18

-------
                                                                                                   EPA               Publication
                                                                                                   n  . •-              «.  .
                                                                                                   Publication      Order
  Method          Organization      ReferenceTitle                           Date         Number          Number       Source of Method
Escherichia coli        The time from sample collection to initiation of analysis may not exceed 30 hours.  Systems are encouraged but not required to hold samples below 10°C during
                         transit.
                         EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms.  EPA also
                         encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
                         intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used. The Agency suggests that laboratories perform
                         these studies on a minimum of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
                         Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test. Methods for establishing false-
                         positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase and cytochrome oxidase, multi-test identification systems, or
                         equivalent confirmation tests. False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).

  9222 G             Standard           Standard Methods for the Examination of      1995                                                Standard Methods
                      Methods           Water and Wastewater, 19th Edition
                               Alternatively, the 18th edition of Standard Methods may be used forE. coli detection if the following protocol is used: at least 10 mL of EC medium is
                               supplemented with 50 ug/mL of 4-methylumbelliferyl-beta-D-glucuronide (MUG) (EC-MUG) before autoclaving.  The inner inverted fermentation tube may
                               be omitted. Transfer the total coliform-positive culture by one of the following methods:
                               •remove the membrane containing the total coliform colonies from the substrate with a sterile forceps and carefully curl and insert the membrane into a tube
                               of EC-MUG (the laboratory may first remove a small portion of selected colonies for verification),
                               •swab entire membrane filter surface with a sterile cotton swab and transfer the inoculum to the EC-MUG (do not leave the cotton swab in the EC-MUG), or
                               •inoculate individual total coliform-positive colonies into EC-MUG.
                               Gently shake the inoculated tubes of EC-MUG to insure adequate mixing and incubate in a waterbath at 44.5 ± 0.2°C for 24 ± 2 hours.  Following incubation,
                               observe fluorescence with an ultraviolet light (366 nm) in the dark.  If fluorescence is visible, E. coli are present.
                               Alternatively, the 18th edition (1992) may be used if the membrane filter containing a total coliform-positive colony(ies) is transferred to nutrient agar, as
                               described in Standard Method 9221 B.3  (18th edition), supplemented with 100 ug/mL ofMUG.  If the 18th edition is used, incubate the agar plate at 35°C for
                               4 hours and then observe the  colony(ies) under ultraviolet light (366 nm) in the dark for fluorescence.  If fluorescence is visible, E. coli are present.


  9222 G             Standard           Standard Methods for the Examination of      1998                                                Standard Methods
                      Methods           Water and Wastewater, 20th Edition
                               Alternatively, the 18th edition of Standard Methods may be used forE. coli detection if the following protocol is used: at least 10 mL of EC medium is
                               supplemented with 50 ug/mL of 4-methylumbelliferyl-beta-D-glucuronide (MUG) (EC-MUG) before autoclaving.  The inner inverted fermentation tube may
                               be omitted. Transfer the total coliform-positive culture by one of the following methods:
                               •remove the membrane containing the total coliform colonies from the substrate with a sterile forceps and carefully curl and insert the membrane into a tube
                               of EC-MUG (the laboratory may first remove a small portion of selected colonies for verification),
                               •swab entire membrane filter surface with a sterile cotton swab and transfer the inoculum to the EC-MUG (do not leave the cotton swab in the EC-MUG), or
                               •inoculate individual total coliform-positive colonies into EC-MUG.
                               Gently shake the inoculated tubes of EC-MUG to insure adequate mixing and incubate in a waterbath at 44.5 ± 0.2°C for 24 ± 2 hours.  Following incubation,
                               observe fluorescence with an ultraviolet light (366 nm) in the dark.  If fluorescence is visible, E. coli are present.
                               Alternatively, the 18th edition (1992) may be used if the membrane filter containing a total coliform-positive colony(ies) is transferred to nutrient agar, as
                               described in Standard Method 9221 B.3  (18th edition), supplemented with 100 ug/mL ofMUG.  If the 18th edition is used, incubate the agar plate at 35°C for
                               4 hours and then observe the  colony(ies) under ultraviolet light (366 nm) in the dark for fluorescence.  If fluorescence is visible, E. coli are present.
 Revised June 2008                                 Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations                         Page 13 of 18

-------
Contaminant
  Method         Organization     ReferenceTitle
                                                            Method
                                                            Date
                                                        EPA
                                                        Publication
                                                        Number
                                  Publication
                                  Order
                                  Number
                                                                                                              Source of Method
Escherichia coli        The time from sample collection to initiation of analysis may not exceed 30 hours.  Systems are encouraged but not required to hold samples below 10°C during
                        transit.
                        EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms. EPA also
                        encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
                        intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used.  The Agency suggests that laboratories perform
                        these studies on a minimum of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
                        Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test. Methods for establishing false-
                        positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase and cytochrome oxidase, multi-test identification systems, or
                        equivalent confirmation tests. False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).
  9223/ONPG-
  MUG (Colilert)
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 18th Edition
1992
Standard Methods
                              Edberg, et.al, Applied and Environmental Microbiology, 55:1003-1008, 1989.
                              The Autoanalysis Colilert System is an MMO-MUG test.
  9223/ONPG-
  MUG (Colilert)
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 19th Edition
1995
Standard Methods
                              Edberg, et.al., Applied and Environmental Microbiology, 55:1003-1008, 1989.
                              The Autoanalysis Colilert System is an MMO-MUG test.
  9223/ONPG-
  MUG (Colilert)
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 20th Edition
1998
Standard Methods
                              Edberg, et.al., Applied and Environmental Microbiology, 55:1003-1008, 1989.
                              The Autoanalysis Colilert System is an MMO-MUG test.
  9223/ONPG-
  MUG (Colilert)
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 21 st Edition
2005
Standard Methods
                              Edberg, et.al., Applied and Environmental Microbiology, 55:1003-1008, 1989.
                              The Autoanalysis Colilert System is an MMO-MUG test.
  9223-97/ONPG-      Standard           Online version of Standard Methods for the
  MUG (Colilert)      Methods           Examination of Water and Wastewater.
                      Online             Approval year by Standard Methods
                                        Committee is designated by last 2 digits.
                                        This is the only online version that is
                                        approved.
                              Edberg, et.al., Applied and Environmental Microbiology, 55:1003-1008, 1989.
                              The Autoanalysis Colilert System is an MMO-MUG test.
                                                                                                              http://www.standardmethods.org/
 Revised June 2008
                           Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                                                                                    Page 14 of 18

-------
Contaminant
  Method         Organization     ReferenceTitle
                                                            Method
                                                            Date
                                                        EPA
                                                        Publication
                                                        Number
                                  Publication
                                  Order
                                  Number
                                                                                                              Source of Method
Escherichia coli        The time from sample collection to initiation of analysis may not exceed 30 hours.  Systems are encouraged but not required to hold samples below 10°C during
                        transit.
                        EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms. EPA also
                        encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
                        intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used.  The Agency suggests that laboratories perform
                        these studies on a minimum of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
                        Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test. Methods for establishing false-
                        positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase and cytochrome oxidase, multi-test identification systems, or
                        equivalent confirmation tests. False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).
  Chromocult®        EM Science        Chromocult® Coliform Agar
  Coliform Agar                         Presence/Absence Membrane Filter Test
                                        Method for Detection and Identification of
                                        Coliform Bacteria and Escherichia coli in
                                        Finished Waters, Version 1.0
                                                            November
                                                            2000
                                                                                            EMD Chemicals
Colisure Test


Colitag®

IDEXX
Laboratories,
Inc.
CPI
International,
Inc.
Colisure Test


Colitag® Product as a Test for Detection and
Identification of Coliforms and E. coli
Bacteria in Drinking Water and Source Water
as Required in National Primary Drinking
Water Regulations
February IDEXX Laboratories, Inc.
28, 1994

August CPI International, Inc.
2001

  E*Colite® Test      Charm             Presence/Absence for Coliforms and E. coli    December
                      Sciences, Inc.       in Water                                  21,1997
                                                                                                              Charm Sciences, Inc
  Filter Membrane
  Method with MI
  Medium
EPA
Brenner, K.P., et.al, Applied and
Environmental Microbiology, 59:3534-3544
1993
EPA/600/J-99/225
http://www.epa.gov/nerlcwww/online.htm
                              Coliscan® is approved as a modification of MI under the ATP program. It is available from Micro logy Laboratories, P.O. Box 340, Goshen, IN 46527-0340.
 Revised June 2008
                           Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                                                                                    Page 15 of 18

-------
Contaminant
  Method         Organization     ReferenceTitle
                                                            Method
                                                            Date
                                                        EPA
                                                        Publication
                                                        Number
Publication
Order
Number
                                                                                                              Source of Method
Escherichia coli        The time from sample collection to initiation of analysis may not exceed 30 hours.  Systems are encouraged but not required to hold samples below 10°C during
                        transit.
                        EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms. EPA also
                        encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
                        intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used.  The Agency suggests that laboratories perform
                        these studies on a minimum of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
                        Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test. Methods for establishing false-
                        positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase and cytochrome oxidase, multi-test identification systems, or
                        equivalent confirmation tests. False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).
  m-ColiBlue24®
  Test
Hach Co.
m-ColiBlue 24 Test, "Total Coliforms and£.   August 17,
coli Membrane Filtration Method with m-      1999
ColiBlue 24 Broth," Method No. 10029,
Revision 2.
                Hach Company
  Ready cult®
  Coliforms 100
  Presence/Absence
   Test
EM Science         Ready cult® Coliforms 100 Presence/Absence  November                                         EMD Chemicals
                   Test for Detection and Identification of       2000
                   Coliform Bacteria and Escherichia coli in
                   Finished Waters," Versonl.O
         Fluorocult® is approved as an acceptable version of Ready cult® under the ATP program. It is available from EMD Chemicals (formerly EM Science).
 Revised June 2008
                           Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                                                                                    Page 16 of 18

-------
Contaminant
  Method         Organization      ReferenceTitle
                                                             Method
                                                             Date
                                                         EPA
                                                         Publication
                                                         Number
                                  Publication
                                  Order
                                  Number
                                                                                                                Source of Method
Fecal Coliforms
  9221 E
   Public water systems need only determine the presence or absence of fecal coliforms; a determination of fecal coliform density is not required.
   When the MTF Technique or Presence-Absence (PA) Coliform Test is used to test for total coliforms, shake the lactose-positive presumptive tube or P-A vigorously
   and transfer the growth with a sterile 3-mm loop or sterile applicator stick into brilliant green lactose bile broth and EC medium to determine the presence of total
   and fecal coliforms, respectively.
   When EPA-approved total coliform membrane filtration methods are used, transfer the total coliform-positive culture by one of the following methods:
   •remove the membrane containing the total  coliform colonies from the substrate with a steril forceps and carefully curl and insert the membrane into a tube of EC
   medium (the laboratory may first remove a  small portion of selected colonies for verification),
   •swab the entire membrane filter surface with a sterile cotton swab and transfer the inoculum to the EC medium (do not leave the cotton swab in the EC medium), or
   •inoculate individual total coliform-positive colonies into EC medium.
   Gently  shake the inoculated tubes of EC medium to insure adequate mixing and incubate in a waterbath at 44.5 ± 0.2 °C for 24 ± 2 hours.  Gas production of any
   amount in the inner fermentation tube of the EC medium indicates a positive fecal coliform test.
   The time from sample collection to initiation of analysis may not exceed 30 hours. Systems are encouraged but not required to hold samples below 10°C during
   transit.
   EPA strongly recommends that laboratories evaluate the false-positive and negative rates for the methods(s) they use for monitoring total coliforms. EPA also
   encourages laboratories to establish false-positive and false-negative rates within their own laboratory and sample matrix (drinking water or source water) with the
   intent that if the method they choose has an unacceptable false-positive or negative rate, another method can be used.  The Agency suggests that laboratories perform
   these studies on a minimum of 5% of all total coliform-positive samples, except for those methods where verification/confirmation is already required, e.g., theM-
   Endo and LES Endo Membrane Filter Tests, Standard Total Coliform Fermentation Technique, and Presence-Absence Coliform Test. Methods for establishing false-
   positive and negative-rates may be based on lactose fermentation, the rapid test for B-galactosidase and cytochrome oxidase, multi-test identification systems, or
   equivalent confirmation tests. False-positive and false-negative information is often available in published studies and/or from the manufacturer(s).
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 18th Edition
1992
Standard Methods
                               The preparation of EC medium is described in paragraph la.
  9221 E
Standard           Standard Methods for the Examination of     1995
Methods           Water and Wastewater, 19th Edition
         The preparation of EC medium is described in paragraph la.
                                                                                             Standard Methods
  9221 E
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 20th Edition
1998
Standard Methods
                               The preparation of EC medium is described in paragraph la.
 Revised June 2008
                           Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                                                                                     Page 17 of 18

-------
Contaminant
  Method         Organization     ReferenceTitle
                                                      Method
                                                      Date
                                                  EPA
                                                  Publication
                                                  Number
                              Publication
                              Order
                              Number
                                                                                                  Source of Method
Heterotrophic Bacteria The time from sample collection to initiation of analysis may not exceed 30 hours. Systems are encouraged but not required to hold samples below 10°C during
                      transit.
  9215withR2A
  Medium
  9215withR2A
  Medium
  9215withR2A
  Medium
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 18th Edition
1992
Standard Methods
                           Any method in Standard Methods Section 9215, Heterotrophic Plate Count, may be used with R2A medium, for enumerating heterotrophic bacteria in
                           drinking water.
                           May be used if public water system operates under a variance to the Total Coliform Rule.
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 19th Edition
1995
Standard Methods
                           Any method in Standard Methods Section 9215, Heterotrophic Plate Count, may be used with R2A medium, for enumerating hetero trophic bacteria in
                           drinking water.
                           May be used if public water system operates under a variance to the Total Coliform Rule.
Standard
Methods
Standard Methods for the Examination of
Water and Wastewater, 20th Edition
1998
Standard Methods
                           Any method in Standard Methods Section 9215, Heterotrophic Plate Count, may be used with R2A medium, for enumerating hetero trophic bacteria in
                           drinking water.
                           May be used if public water system operates under a variance to the Total Coliform Rule.
Contact information for methods that are not available on the Internet are summarized in the report titled
"Sources of Approved Analytical Methods for National Drinking Water Regulations."
Revised June 2008
                        Approved Methods for the Total Coliform Rule - National Primary Drinking Water Regulations
                                                                                                       Page 18 of 18

-------