EPA-821-R-01-009
                                    January 2001
                 Method 1690

Ammonia-N in Water and Biosolids by Automated
    Colorimetry with Preliminary Distillation
                   DRAFT
                 January 2001
     U.S. Environmental Protection Agency
                Office of Water
       Office of Science and Technology
    Engineering and Analysis Division (4303)
          1200 Pennsylvania Ave. NW
            Washington, D.C. 20460

-------
Method 1690
                                    Acknowledgments

This method was prepared under the direction of William A. Telliard of the U.S. Environmental Protection
Agency's (EPA's) Office of Water (OW), Engineering and Analysis Division (EAD). The method was
prepared under EPA Contract 68-C-98-139 by DynCorp Information and Enterprise Technology.
                                         Disclaimer

This draft method has been reviewed and approved for publication by the Analytical Methods Staff within
the Engineering and Analysis Division of the U.S. Environmental Protection Agency. Mention of trade
names or commercial products does not constitute endorsement or recommendation for use. EPA plans
further validation of this draft method. The method may be revised following validation to reflect results of
the study. This method version contains minor editorial changes to the February 1999 version.

EPA welcomes suggestions for improvement of this method.  Suggestions and questions concerning this
method or its application should be addressed to:

Maria Gomez-Taylor
Engineering and Analysis Division (4303)
U.S. Environmental Protection Agency
1200 Pennsylvania Ave., NW
Washington, D.C. 20460
Phone: (202)260-7134
Fax: (202)260-7185
                                                                             Draft, January 2001

-------
                                                                                     Method 1690
     Note: This method is performance based. The laboratory is permitted to omit any step or modify any
     procedure provided that all performance requirements in this method are met.  The laboratory may
     not omit any quality control analyses. The terms "shall," "must," and "may not" define procedures
     required for producing reliable results.  The terms "should" and "may"  indicate optional steps that
     may be modified or omitted if the laboratory can demonstrate that the  modified method produces
     results equivalent or superior to results produced by this method.
Draft, January 2001                                                                              Hi

-------
                                    Method 1690


        Ammonia-N  in Water and Biosolids  by Automated
                Colorimetry with Preliminary Distillation

1.0  Scope and Application

1.1  This method describes procedures for the determination of ammonia-N in drinking, ground, and
     surface water; domestic and industrial waste; and biosolids (municipal sewage sludge).  Distillation
     of ammonia from the sample is followed by analysis using an automated colorimetric analyzer. This
     method is based on U.S.  Environmental Protection Agency (EPA) Methods 350.1: Nitrogen,
     Ammonia (Colorimetric, AutomatedPhenate) and 350.2: Nitrogen, Ammonia (Colorimetric;
     Titrimetric; Potentiometric Distillation Procedure) (Reference 16.1). This method is associated
     with Method 1691: Municipal Biosolids Sampling Guidance (Reference 16.2).

1.2  This method is to be used in EPA's data gathering and monitoring programs under the Clean Water
     Act, the Resource Conservation and Recovery Act, the Comprehensive Environmental Response,
     Compensation, and Liability Act, the Solid Waste Disposal Act, the Solid Waste Disposal Act, and
     the Safe Drinking Water Act.

1.3  Method detection limits and minimum levels for ammonia-N have not been formally established for
     this draft method. These values will be determined during the validation of the method.

1.4  This method is performance based.  The laboratory is permitted to omit any step or modify any
     procedure, provided that  all performance requirements in this method are met. Requirements for
     establishing method equivalency are given in Section 9.1.2.

1.5  Each laboratory that uses this method must demonstrate the ability to generate acceptable results
     using the procedures in Section 9.2.

2.0  Summary of Method

2.1  The ammonia in the sample is first separated by distillation, eliminating many potential interferences.
     The ammonia released is captured in a dilute sulfuric acid solution.

2.2  The ammonia concentration is determined by colorimetric measurement of indophenol blue, which is
     formed when ammonia reacts with alkaline phenol and hypochlorite.

2.3  Quality is assured through calibration and testing of the analytical instruments and testing of the
     sample preparation.

3.0  Definitions

     Definitions for terms used in this method are given in Section 18.
Draft, January 2001

-------
Method 1690
4.0  Interferences

4.1   Cyanate, which may be encountered in certain industrial effluents, will hydrolyze to some extent
      even at the pH of 9.5 at which distillation is carried out, creating a positive bias.

4.2   Residual chlorine, if present, must be removed by pretreatment of the sample with sodium thiosulfate
      before distillation. Typically, this will be necessary if the sample contains free water or is aqueous.

5.0  Safety

5.1   The toxicity or carcinogenicity of reagents used in this method has not been fully established.  Each
      chemical and environmental sample should be regarded as a potential health hazard and exposure
      should be minimized.  Each laboratory is responsible for maintaining a current awareness file of
      OSHA regulations regarding the safe handling of the chemicals specified in this method. A reference
      file of material safety data sheets (MSDS)  should be available to all personnel involved in the
      chemical analysis. Additional information on laboratory safety can be found in Reference 16.3.

5.2   If samples originate from a highly contaminated area, appropriate sample handling procedures must
      be followed to minimize worker exposure.

5.3   All personnel handling environmental samples known to contain or to have been in contact with
      human waste should be immunized against known disease causative agents.

6.0  Apparatus and Equipment

      NOTE: Brand names, suppliers, and part numbers are for illustration only, and no endorsement
      is implied. Equivalent performance may be achieved using apparatus and materials other than
      those specified here. Meeting the performance  requirements of this method is the responsibility of
	the sampling team and laboratory.	

6.1   Equipment for Distillation

      6.1.1 An all-glass distillation apparatus with a 1-L flask.

      6.1.2 Erlenmeyer flasks, 500-mL.

6.2   Colorimetric analysis—Automated analytical equipment with the capability of delivering the
      reagents specified in Figure 1 at the specified rates,  and with a colorimeter capable of measuring at
      630-660 nm.

6.3   General equipment

      6.3.1 Analytical balance, capable of weighing to 0.001 g (1 mg).

      6.3.2 Hot plate or other heating apparatus.

      6.3.3 Glass beads or boiling chips.

2                                                                             Draft, January 2001

-------
                                                                                    Method 1690
7.0  Reagents and Consumable Materials

7.1   All water and reagents used in this method must be free of ammonia.

      7.1.1 Ammonia-free reagent water—Ammonia-free water can be prepared by passing reagent water
            through an ion-exchange column containing a mixture of both strongly acidic cation and
            strongly basic anion exchange resins. When necessary, regenerate the column according to
            the manufacturer's instructions.

      7.1.2 All reagents should be ACS Reagent Grade or better.

7.2   0.04N Sulfuric acid—Dilute 1 mL concentrated sulfuric acid (H2SO4) to 1L with reagent water.

7.3   Sodium hydroxide solutions

            CAUTION: Considerable heat is generated upon dissolution of sodium hydroxide in water.
            It may be advisable to cool the container in an ice bath when preparing sodium hydroxide
	solutions.	


      7.3.1 0.1N Sodium hydroxide (NaOH)—Dissolve 4.0 g NaOH in reagent water, cool, and dilute to
            1L.

      7.3.2 IN Sodium hydroxide—Dissolve 40 g NaOH in reagent water, cool, and dilute to 1 L.

7.4   Borate buffer

      7.4.1 0.025M Sodium tetraborate—Dissolve 5 g anhydrous sodium tetraborate (Na2B4O7) or 9.5 g
            sodium tetraborate decahydrate (Na2B4O7»10H2O) in reagent water and dilute to 1 L.

      7.4.2 Add 88 mL of 0.1N NaOH solution (Section 7.3.1) to 500 mL 0.025M Na2B4O7 (Section
            7.4.1) and dilute to 1 L with reagent water.

7.5   Sodium hypochlorite solution—Dilute 250 mL of a bleach solution containing 5.25% NaOCL (such
      as "Clorox") to 500 mL. Available chlorine level should approximate 2-3%.  Since "Clorox" is a
      proprietary product, its formulation is subject to change.  The analyst must remain alert to detecting
      any variation in this product significant to its use in this procedure. Due to the instability of bleach,
      storage over an extended period should be avoided.

7.6   Dechlorinating reagent—Dissolve  0.35 g sodium thiosulfate (Na2S2O3»5H2O) in reagent water and
      dilute to 100 mL. One milliliter of this reagent will neutralize 1 mg/L of residual chlorine in a 500
      mL sample aliquot.

7.7   Blank sand—Bake 500 g diatomaceous earth  or clean sand at 400°C for eight hours.  Cool and store
      in a glass container with a sealing lid.

7.8   Ammonia standards


Draft, January 2001                                                                              3

-------
Method 1690
      7.8.1 Ammonia stock solution (1000 mg/L NH3-N)—Dissolve 0.382 g of anhydrous ammonium
            chloride (NH4C1) in reagent water and dilute to 100 mL in a volumetric flask. NH4C1 should
            be dried at 105 °C prior to use.

      7.8.2 Ammonia working standard (10 mg/L NH3-N)—In a volumetric flask, dilute 10 mL of
            ammonia stock solution (Section 7.11.1) to 1 L.

8.0  Sample Collection, Preservation  and Storage

8.1   Ammonia is a pervasive contaminant. Minimize  exposure of samples to air as much as possible.

8.2   A sufficient volume of sample for analysis must be collected using the procedures found in
      Reference  16.2 for biosolids samples and Reference 16.4 for water and wastewater samples.

8.3   Ammonia can be formed or lost during storage due to biological activity or lost by volatilization or
      oxidation.  The following preservation procedures will help prevent significant changes in the analyte
      concentration.

      8.3.1 Samples should be collected in wide mouth jars with a minimum of air space above them.

      8.3.2 Aqueous samples or samples that contain free water

            8.3.2.1    Samples should be checked for residual chlorine and treated with sodium
                       thiosulfate, if necessary, during collection.

            8.3.2.2    Samples should be preserved  with 2 mL concentrated H2SO4 and cooled to 4°C
                       as soon as possible after collection. The holding time for samples should not
                       exceed 28 days from sampling.

      8.3.3 If the sample contains no free water or is solid, the sample should be cooled to 4°C as soon as
            possible after collection. The holding time for samples should not exceed 28 days from
            sampling.

8.4   Collect a separate sample for total solids determination (Appendix A).  The holding time for total
      solids determination is seven days.

9.0  Quality Control

9.1   Each laboratory using this method is required to operate a formal quality control  (QC) program.
      The minimum requirements of this program consist of an initial demonstration of laboratory
      capability, the ongoing analysis of laboratory reagent blanks, precision and recovery standards, and
      matrix-spiked samples as a continuing check on performance. The laboratory is required to maintain
      performance records that define the quality of data thus generated. Laboratory performance is
      compared to established performance criteria to determine if the results of analyses meet the
      performance characteristics of the method.

      9.1.1 The analyst shall make an initial demonstration of the ability to generate acceptable accuracy
            and precision with this method. This ability is established as described in Section 9.2.

4                                                                              Draft, January 2001

-------
                                                                                       Method 1690
      9.1.2 In recognition of advances that are occurring in analytical technology, the analyst is permitted
            certain options to improve separations or lower the costs of measurements, provided that all
            performance specifications are met.  Changes that degrade method performance are not
            allowed. If an analytical technique other than the techniques specified in this method is used,
            that technique must have a specificity equal to or better than the specificity of the techniques
            in this method for NH3-N in the sample of interest.  Specificity is defined as producing results
            equivalent to the results produced by this method for analytical standards (Section 9.4) and,
            where applicable, environmental samples (Section 9.5), and as meeting all of the QC criteria
            stated in this method.

            9.1.2.1    Each time a modification is made to this method, the analyst is required to repeat
                        the IPRtest in Section 9.2.2 to demonstrate that the modification produces
                        results equivalent to or better than results produced by this method.  If the
                        detection limit of the method will be affected by the modification, the analyst
                        must demonstrate that the MDL (40 CFR part 136, appendix B) is less than or
                        equal to the MDL in this method or one-third the regulatory compliance level,
                        whichever is higher. The tests required for this equivalency demonstration are
                        given in Section 9.1.2.2.4.

            9.1.2.2    The laboratory is required to maintain  records of modifications made to this
                        method.  These records include the following, at a minimum:

                        9.1.2.2.1  The names, titles, addresses, and telephone numbers of the
                                    analyst(s) who performed the analyses and modification, and of the
                                    quality control officer who witnessed and will verify the analyses
                                    and modification.

                        9.1.2.2.2  A listing of pollutant(s) measured (NH3-N).

                        9.1.2.2.3  A narrative stating reason(s) for the modification.

                        9.1.2.2.4  Results from all quality control (QC) tests comparing the modified
                                    method to this method, including:

                                           (a)    Calibration  (Section 10).
                                           (b)    Calibration verification (Section 9.6).
                                           (c)    Initial precision and recovery (Section 9.2.2).
                                           (d)    Analysis of blanks (Section 9.3).
                                           (e)    Accuracy assessment (Sections 9.5 and 9.7).
                                           (f)     Ongoing precision and recovery (Section 9.4).
                                           (g)    Method detection limit (Section 9.2.1)

                        9.1.2.2.5  Data that will allow an independent reviewer to validate each
                                    determination by tracing the instrument output (weight, absorbance,
                                    or other signal) to the final result. These data are to  include:

                                           (a)    Sample numbers and other identifiers.
                                           (b)    Sample preparation dates.
                                           (c)    Analysis dates and times.
                                           (d)    Analysis sequence/run chronology.

Draft, January 2001                                                                                5

-------
Method 1690
                                          (e)    Sample weight or volume.
                                          (f)    Dry weight ratio (solid or semisolid samples only;
                                                Appendix A).
                                          (g)    Distillate solution volume.
                                          (h)    Make and model of analytical balance and weights
                                                traceable to NIST.
                                          (i)    Copies of logbooks, printer tapes, and other
                                                recordings of raw data.
                                          (j)    Data system outputs, and other data to link the raw
                                                data to the results reported.

      9.1.3 Analyses of laboratory blanks are required to demonstrate freedom from contamination. The
            procedures and criteria for blank analyses are described in Section 9.3

      9.1.4 Analyses of ongoing precision and recovery samples are required to demonstrate that the
            sample preparation and analysis are within the specified limitations.  The procedure and
            criteria for OPR sample analysis are described in Section 9.4.

      9.1.5 Analyses of matrix spike and matrix spike duplicate samples are required to demonstrate
            method accuracy and precision, and to monitor interferences caused by the sample matrix.
            The procedure and criteria for spiking are described in Section 9.5.

      9.1.6 Analyses of calibration verification standards are required to demonstrate accuracy and
            stability of the initial  calibration.  The procedure and criteria for calibration verification
            analyses are described in Section 9.6.

      9.1.7 Analyses of quality control samples (QCS) are required to demonstrate the accuracy of the
            calibration standards  and the analytical system.  The procedure and criteria for the QCS
            sample analyses are described in Section 9.7.

9.2   Initial demonstration of laboratory capability—The initial demonstration  of laboratory capability is
      used to characterize  laboratory performance and method detection limits.  The MDL and IPR for
      solid samples should be determined using blank sand as a reference matrix. The MDL and IPR for
      aqueous samples should be determined using reagent water as a reference matrix.

      9.2.1 Method detection limit (MDL)—The MDL should be established for NH3-N according to the
            procedures at 40 CFR part 136, appendix B (Reference 16.5). First, spike a reference matrix
            with ammonia working standard (Section 7.8.2) to produce a concentration one to five times
            the estimated  detection limit. To determine the MDL, take  seven replicate aliquots of the
            spiked reference matrix and process each aliquot through each step of the analytical method.
            Perform all calculations and report the concentration values in the appropriate units. Aqueous
            and/or solid method detection limits should be determined every year or whenever a
            modification to the method or analytical system  is made that will affect the MDL.

      9.2.2 Initial precision and recovery (IPR)—To establish the ability to generate acceptable precision
            and accuracy, the analyst shall perform the following operations:

            9.2.2.1    Prepare  four spiked samples as detailed in Section 9.4.  Using the procedures in
                        Section  11, prepare and analyze these spiked samples for NH3-N.
                                                                                 Draft, January 2001

-------
                                                                                      Method 1690
            9.2.2.2     Using the results of the set of four analyses, compute the average percent
                        recovery (X) and the standard deviation (s) of the percent recovery for NH3-N.

            9.2.2.3     Compare s and X with the corresponding limits for initial precision and recovery
                        in Table 1.  If s and X meet the acceptance criteria, system performance is
                        acceptable and analysis of samples may begin.  If, however, s exceeds the
                        precision limit or X falls outside the range for recovery, system performance is
                        unacceptable. In this event, correct the problem, and repeat the test.

9.3   Laboratory blanks—Laboratory blanks are analyzed to demonstrate freedom from contamination.
      Aqueous samples should be run with an aqueous blank, and solid samples should be run with a solid
      blank.

      9.3.1 Prepare and analyze a laboratory blank initially (i.e., with the tests in Section 9.2) and with
            each analytical batch. The blank must be subjected to the same procedural steps as a sample,
            and will consist of 250 mL of ammonia-free reagent water (aqueous blank) or 5 g aliquot of
            blank sand in 250 mL of ammonia-free reagent water (solid blank).

      9.3.2 If material is detected in the blank at a concentration greater than the MDL (Section 1.3),
            analysis of samples is halted until the source of contamination is eliminated and a blank shows
            no evidence of contamination. All samples must be associated with an uncontaminated
            laboratory blank before the results may be reported for regulatory compliance purposes.

9.4   Ongoing precision and recovery (OPR)—The laboratory must analyze at least one ongoing precision
      and recovery sample with each analytical batch.  A solid OPR should be run with solid samples, and
      an aqueous OPR should be run with aqueous samples. An aqueous OPR is prepared by spiking
      reagent water with ammonia working standard (Section  7.8.2) so that the concentration of NH3-N in
      the OPR is one to five times the ML. A solid OPR is prepared by mixing 5 g of blank sand with 250
      mL reagent water and spiking with ammonia working standard (Section 7.8.2) so that the
      concentration of NH3-N in the OPR is one to five times the ML.  The spiked aliquot is carried
      through the entire analytical process (Section 11).  Calculate accuracy as percent recovery. If the
      recovery of the analyte falls outside the control limits in Table  1, the system performance is
      unacceptable, and the source of the problem should be identified and resolved before continuing
      analyses.

9.5   Matrix spike and matrix spike duplicates (MS/MSD)—To assess the performance of the method on
      a given sample matrix, the laboratory must spike, in duplicate, a minimum of 10% (one sample in
      10) of the samples from a given sampling site or, if for compliance monitoring, from a given
      discharge.  Blanks may not be used for MS/MSD analysis.

      9.5.1 The concentration of the MS and MSB shall be determined as follows:

            9.5.1.1     If, as in compliance monitoring, the concentration of analytes in the sample is
                        being checked against a regulatory concentration limit, the spiking level shall  be
                        at that limit or at 1-5 times the background concentration of the sample,
                        whichever is greater.

            9.5.1.2     If the concentration of NH3-N in a sample is not being checked against a
                        regulatory concentration limit, the spike shall be at 1-5 times the background
                        concentration.

Draft, January 2001                                                                                7

-------
Method 1690
           9.5.1.3    For solid and sludge samples, the concentration added should be expressed as
                       mg/kg and is calculated for a one gram aliquot by multiplying the added analyte
                       concentration (mg/L) in solution by the conversion factor 100 (mg/L x
                       0.1L/0.001kg= 100).

      9.5.2 Assessing spike recovery

           9.5.2.1    To determine the background concentration, analyze one sample aliquot from
                       each set of 10 samples from each site or discharge according to the procedure in
                       Section 11.  If the expected background concentration is known from previous
                       experience or other knowledge, the spiking level may be established a priori.

           9.5.2.2    Prepare the MS/MSD samples by spiking two sample aliquots with ammonia
                       working standard (Section 7.8.2). Analyze the MS/MSD aliquots as described
                       in Section 11 to determine the concentration of the  samples after spiking.

      9.5.3 Calculate the percent recovery (P) and relative percent difference (RPD) of the two matrix
           spike samples for the analyte, corrected for the background concentration measured in the
           sample, and compare these values to the control limits given in Table 1.  Percent recovery is
           calculated in units appropriate to the matrix, using Equation  1. RPD is calculated using
           Equation 2.
                                         Equation 1

                                                 (c  -c]
                           percent recovery=^-1	— * 100
           where:      Cs = Measured sample concentration after spiking
                       Cb=Measured sample backgound concentration
                       S=known concentration of the spike

                                         Equation 2
                                          (lA-Al)
                                        =^ - ^f
                                          (A+A)
           where:      D, concentration NH3-N of MS sample
                       D2=concentration NH3-N ofMSD sample
      9.5.4 If the percent recovery or the RPD of the analyte in the MS/MSD samples falls outside the
           designated range, and the laboratory performance on the OPR for the analyte is within the
           specified limits (Section 9.4), the recovery problem encountered with the MS/MSD sample is
           judged to be matrix-related instead of method-related.
                                                                             Draft, January 2001

-------
                                                                                      Method 1690
      9.5.5 Recovery for samples should be assessed and records maintained.

            9.5.5.1     After the analysis of five samples of a given matrix type (river water, lake water,
                        etc.) for which the results pass the tests in Section 9.5.3, compute the average
                        percent recovery (R) and the standard deviation of the percent recovery (SR) for
                        the analyte(s).  Express the accuracy assessment as  a percent recovery interval
                        from R-2SR to R+2SR for each matrix. For example, if R=90% and SR=10%
                        for five analyses of river water, the accuracy interval is expressed as 70-110%.

            9.5.5.2     Update the accuracy assessment for each matrix regularly (e.g., after each five to
                        ten new measurements).

9.6   Calibration verification (CV)—The laboratory must analyze a calibration verification standard
      before running any samples and once per ten analyses thereafter. The CV should be prepared at a
      concentration that is at or near the midpoint of the calibration curve.  The source of the CV standard
      should be different from the source used to prepare the calibration standards. If a different
      ammonium compound is used for the CV stock, the amount weighed will have to be adjusted
      according to the ratio of NH3-N atomic weight to the molecular weight. Results of the CV analysis
      should be evaluated according to the specifications in Table 1.  If the CV does not meet acceptance
      criteria, the problem must be identified and corrected, including possible recalibration of the
      instrument.

9.7   Quality control sample (QCS)—It is suggested that the laboratory obtain a prepared quality control
      sample from a standards vendor (ERA catalog  # 545, or equivalent) and analyze it with each day's
      distillations, or every twelve hours, whichever is more frequent.  The results of the QCS analysis
      should be evaluated according to the manufacturer's specifications.

10.0  Calibration and Standardization

10.1  Calibrate the colorimeter with a minimum  often standards that cover the expected range of the
      samples and a blank.  If the autoanalyzer is linked to a data system, follow the manufacturer's
      instructions for calibration.  Alternatively, prepare a calibration graph relating absorbance to the
      concentration of NH3-N, or develop a linear regression formula from the calibration data using
      concentration versus absorbance. If the correlation coefficient falls below 0.995, check the system
      for faults, correct them if found, and reanalyze the calibration standards.

10.2  Preparation of calibration curve—Prepare the standards (Table 2) and analyze them according to the
      procedure in Section 11.2, beginning with the lowest standard. It is not necessary to distill the
      calibration standards.

10.3  An OPR run after the calibration must be within the specified limits (Table 1).  Unacceptable OPR
      results require corrective action before analysis of samples may continue; they may indicate
      problems with the distillation procedure.

10.4  Balance calibration—Calibrate the analytical balance at 2 mg and 1000 mg using class "S" weights.
      The calibration shall be within ± 10% at 2 mg and ± 0.5% at 1000 mg.  If values are not within
      these limits, recalibrate the balance.
Draft, January 2001

-------
Method 1690
11.0 Procedure

11.1  Distillation

      11.1.1      Preparation of glassware—Add 500 mL reagent water and 20 mL borate buffer
                  (Section 7.4) to the distillation flask. Place a few glass beads or boiling chips in the
                  flask.

      11.1.2      Steam out the distillation apparatus until the distillate shows no traces of ammonia.
                  Leave the apparatus assembled until just before beginning the sample distillation.

      11.1.3      Sample preparation

                  11.1.3.1    Aqueous samples—Add 250 mL sample and 25 mL borate buffer to a
                              clean distillation flask. Adjust the sample to a pH of 9.5 with IN NaOH
                              (Section 7.3.2) using a pH meter.

                  11.1.3.2   Solid samples—Add 5 g of the sample to a clean distillation flask.  Add
                              250 mL reagent water and 25 mL borate buffer.  Adjust the sample to a
                              pH of 9.5 with IN NaOH using a pH meter.

      11.1.4      Remove the steaming out flask from the distillation apparatus and immediately replace
                  it with the sample distillation flask.  Place the tip of the condenser below the surface of
                  solution in the receiving flask.  The receiving flask should contain 50 mL of 0.04N
                  H2SO4 (Section 7.2)

      11.1.5      Distill at a rate of 6-10 mL/min. until 100 mL of distillate is collected.

      11.1.6      Dilute the distillate to 250 mL with reagent water.

11.2  Colorimetric analysis

      11.2.1      Since the intensity of the color used to quantify the concentration is pH dependent, the
                  acid concentration of the wash water and the standard ammonia solutions should
                  approximate that of the samples.

      11.2.2      Set up the autoanalyzer according to the manufacturer's  instructions, using the
                  information in Figure 1 as a guide.

      11.2.3      Allow both colorimeter and recorder (if applicable) to warm up for 30 minutes.  Obtain
                  a stable baseline with all reagents, feeding reagent water through sample line.

      11.2.4      Switch the sample line  from reagent water to the first sample and begin analysis.

11.3  For solid samples, the dry weight/wet weight ratio must be determined separately (Appendix A).
10                                                                              Draft, January 2001

-------
                                                                                   Method 1690
12.0 Calculations

12.1  Aqueous samples—Compare the absorbance reading for each sample to the calibration curve and
      determine the sample concentration. Report all values in mg/L to three significant figures.

12.2  Solid samples—Compare the absorbance reading for each sample to the calibration curve and
      determine the sample concentration using Equations 3 and 4. Report all values in mg/kg to three
      significant figures.

                                         Equation 3

                                             C *V *F*1000 sike
                         NH3-N(mg/kg}=   d    d      iuuug//cg
                                                 Ws*WQQmL/L
           where:      Cd=Concentration ofNH3-Nin distillate (mg/L)
                       Vd=Volume of distillate collected (250 mL)
                       F=Dilution factor (1 if no dilution)
                       Ws=Weight of sample distilled (g)

                                         Equation 4
           where:      Cs=Amount ofNH3-Nin sample (mg/kg, Equation 3)
                       W=Dry weight ratio (Appendix A)
12.3 Report all results below the ML as "less than the ML."

12.4 The QC data obtained during the analysis provides an indication of the quality of the sample data
      and should be provided with the sample results.

13.0 Method  Performance

      This is a draft method, and is currently undergoing validation.  Method performance criteria will be
      set following the validation of the method.

14.0 Pollution Prevention

14.1  Pollution prevention encompasses any technique that reduces or eliminates the quantity or toxicity of
      waste  at the point of generation. Many opportunities for pollution prevention exist in laboratory
      operation.  The EPA has established a preferred hierarchy of environmental management techniques
      that places pollution prevention as the management option of first choice. Whenever feasible,
      laboratory personnel should use pollution prevention techniques to address their waste generation.
      When wastes cannot be feasiblely reduced at the source, EPA recommends recycling as the next best
      option. The acids used in this method should be reused as practicable by purifying by
      electrochemical techniques.  The only other chemicals used in this method are the neat materials used
      in preparing standards. These standards are  used in extremely small amounts and  pose little threat

Draft, January 2001                                                                           11

-------
Method 1690
      to the environment when managed properly.  Standards should be prepared in volumes consistent
      with laboratory use to minimize the volume of expired standards to be disposed.

14.2  For information about pollution prevention that may be applicable to laboratories and research
      institutions, consult "Less is Better: Laboratory Chemical Management for Waste Reduction,"
      available from the American Chemical Society's Department of Government Relations and Science
      Policy, 1155 16th Street NW, Washington, D.C.  20036, 202/872-4477.

15.0 Waste Management

15.1  The laboratory is responsible for complying with all Federal, State, and local regulations governing
      waste management, particularly hazardous waste identification rules and land disposal restrictions,
      and for protecting the air, water, and land by minimizing and controlling all releases from fume
      hoods and bench operations. Compliance with all sewage discharge permits and regulations is also
      required. An overview of requirements can be found in Environmental Management Guide for
      Small Laboratories (EPA 233-B-98-001).

15.2  Samples containing strong acids or bases are  hazardous and must be either neutralized before being
      disposed or handled as hazardous waste.

16.0  References
16.1  U.S. Environmental Protection Agency, 1979. Methods for Chemical Analysis of Water and Wastes.
      Publ. 600/4-79-020, rev. March 1983.  Environmental Monitoring and Support Lab., U.S.
      Environmental Protection Agency, Cincinnati, Ohio.

16.2  U.S. Environmental Protection Agency, 1998. Method 1691: Municipal Biosolids Sampling
      Guidance. Draft, September 1998. Office of Water, Washington, DC.

16.3  Sax, N.I. and Lewis, Sr., R.I., Dangerous Properties of Industrial Materials. 5th. ed.. Van Nostrand
      Reinhold, New York, 1989.

16.4  U.S. Environmental Protection Agency, 1982. Handbook for Sampling and Sample Preservation of
      Water and Wastewater. Publ. 600/4-82-029, Environmental Monitoring and Support Lab., U.S.
      Environmental Protection Agency, Cincinnati, Ohio.

16.5  Code of Federal Regulations 40, Ch. 1, Part 136, Appendix B.

17.0 Tables, Diagrams, and Validation Data

                       Table 1. Quality Control Sample Acceptance Criteria

                          (To be determined during laboratory validation)
Analyte
NH3-N
Blank
limit

IPR recovery
(x) limit

IPR precision
(s) limit

OPR
recovery limit

CV recovery
limit

QCS recovery
limit


12
Draft, January 2001

-------
                                                                                        Method 1690
    Table 2.  The volume of working standard necessary to make the calibration standards in 100-mL
                                          volumetric flasks


                           Volume of standard         Concentration of
                         (Section 7.8.2) added (mL)     calibration standard
                                                        (mg/L)

                                  0.1                     0.01

                                  0.2                     0.02

                                  0.5                     0.05

                                  1.0                     0.10

                                  2.0                     0.20

                                  5.0                     0.50

                                  8.0                     0.80

                                 10.0                     1.00

                                 15.0                     1.50

                                 20.0                    2.00
Draft, January 2001                                                                                13

-------
Method 1690
                                              Peristaltic Pump Manifold
                                                                         O.23 mL/min Air*
                      Mixing ooil
                                   •4—r
          Mixing ooil

          SOdeg C
           heating
            bath
          Colorimeter w/15mm flow cell
                           to Waste
 Recorder or data system
                                                                         O.42 mL/min Sample
                                                                         O.8O mL/min
                                                                         EPTA
                                                                         O.42 Sodium Phenolate
O.32 mL/min Hypochlorite
                                                                         O.42 mL/min Nitroprusside
                                                                         1.6 mL/min  Waste
                                                                         2.O mL/min  Sampler Wash
                                                               *scrubbed through 5N sulfuric acid
                              Figure 1.  Continuous Flow Diagram
14
                Draft, January 2001

-------
                                                                                     Method 1690
18.0 Glossary of Terms

      The definitions and purposes below are specific to this method, but have been conformed to common
      usage as much as possible.

18.1    Ammonia-N—The analyte NH3 quantitated based on the trivalent nitrogen concentration.

18.2    Analyte—A compound or element tested for by the methods referenced in this method.

18.3    Analytical Batch—The set of samples analyzed at the  same time, to a maximum of 10 samples.

18.4    Biosolids—The treated residuals from wastewater treatment that can be used beneficially.

18.5    Apparatus—The sample container and other containers, filters, filter holders, labware, tubing,
        pipettes, and other materials and devices used for sample collection or sample preparation, and
        that will contact  samples, blanks, or analytical standards.

18.6    Calibration Standard—A solution prepared from a dilute mixed standard and/or stock solution
        and used to calibrate the response of the instrument with respect to analyte concentration.

18.7    Calibration Verification Standard (CV)—A solution prepared from a different source than the
        calibration standards that is used to confirm the accuracy of the instrument's calibration.

18.8    Initial Precision and Recovery (IPR)—Four aliquots  of the OPR standard analyzed to establish
        the ability to generate acceptable precision and accuracy. IPR is performed  before a method is
        used for the first  time and any time the method or instrumentation is modified.

18.9    Laboratory Blank—An aliquot of reagent water that is treated exactly as a sample including
        exposure to all glassware, equipment, solvents, reagents, internal standards, and surrogates that
        are used with samples.  The laboratory blank is used to determine if method analytes or
        interferences are  present in the laboratory environment, the reagents, or the apparatus (Section
        9.3).

18.10  Matrix Spike  (MS) and Matrix Spike Duplicate (MSD)—Aliquots of an environmental sample
        to which known  quantities of the method analytes are added in the laboratory. The MS and MSD
        are analyzed exactly like a sample.  Their purpose is to quantify the bias and precision caused by
        the sample matrix. The background concentrations of the analytes in the sample matrix must be
        determined in a separate aliquot and the measured values in the MS and MSD corrected for
        background concentrations (Section 9.5).

18.11  May—This action, activity, or procedural step is optional.

18.12  May Not—This  action, activity, or procedural  step is prohibited.

18.13  Method Detection Limit (MDL)—The minimum concentration of an analyte that can be
        identified, measured, and reported with 99% confidence that the analyte concentration is greater
        than zero (Section 9.2.1).


Draft, January 2001                                                                              15

-------
Method 1690
18.14  Minimum Level (ML)—The lowest level at which the entire analytical system gives a
        recognizable signal and acceptable calibration point.

18.15  Must—This action, activity, or procedural step is required.

18.16  Ongoing Precision and Recovery (OPR) Standard—A laboratory blank spiked with known
        quantities of the method analytes. The OPR is analyzed exactly like a sample. Its purpose is to
        determine whether the methodology is in control and to assure that the results produced by the
        laboratory remain within the method-specified limits for precision and accuracy (Section 9.4).

18.17  Reagent Water—Water demonstrated to be free from the method analytes and potentially
        interfering substances at the MDL for the method.

18.18  Sewage Sludge—Sewage sludge is solid, semi-solid, or liquid residue generated during the
        treatment process  of domestic sewage in a treatment works.  Sewage sludge includes but is not
        limited to, domestic septage; scum or solids removed in primary, secondary, or advanced
        wastewater treatment processes; and a material derived from sewage sludge.   Sewage sludge does
        not include ash generated during the firing of sewage sludge in a sewage sludge incinerator or grit
        and screenings generated during preliminary treatment of domestic sewage in a treatment works.

18.19  Shall—This action, activity, or procedural step is required.

18.20  Should—This action, activity, or procedural step is suggested but not required.

18.21  Stock Standard Solution—A solution containing one or more method analytes that is prepared
        using a reference material traceable to EPA, the National Institute of Science  and Technology
        (NIST), or a source that will attest to the purity and authenticity of the reference material.
16                                                                              Draft, January 2001

-------
                                                                                Method 1690
    Appendix A:  Total Solids  in  Solids and Biosolids


1.0   Scope and Application

1.1   This procedure is applicable to the determination of total solids in such solid and semisolid samples
      as soils, sediments, biosolids (municipal sewage sludge) separated from water and wastewater
      treatment processes, and sludge cakes from vacuum filtration, centrifugation, or other sludge
      dewatering processes.

1.2   This procedure is taken from EPA Method 1684: Total, Fixed, and Volatile Solids in Water, Solids
      and Biosolids.

1.3   Method detection limits (MDLs) and minimum levels (MLs) have not been formally established for
      this draft procedure.  These values will be determined during the validation of Method 1684.

1.4   This procedure is performance based. The laboratory is permitted to omit any step or modify any
      procedure (e.g. to overcome interferences, to lower the cost of measurement), provided that all
      performance requirements in this procedure are met. Requirements for establishing equivalency are
      given in Section 9.1.2 of Method 1690.

1.5   Each laboratory that uses this procedure must demonstrate the ability to generate acceptable results
      using the procedure in Section 9.2 of this appendix.

2.0   Summary  of Method

2.1   Sample aliquots of 25-50 g are dried at 103 °C to 105 °C to drive off water in the sample.

2.2   The mass of total solids in the sample is  determined by comparing the mass of the sample before and
      after each drying step.

3.0   Definitions

3.1   Analytical batch—The set of samples analyzed at the same time, to a maximum of 10 samples.
      Each analytical batch of 10 or fewer samples must be accompanied by a laboratory blank, an
      ongoing precision and recovery sample, and a set of duplicate samples, resulting in a minimum of
      five analyses (1 sample, 1 blank, 1 OPR, and 2 duplicates) and a maximum of 14 samples.

3.2   Total Solids—The residue left in the vessel after evaporation of liquid from a sample and subsequent
      drying in an oven at 103°C to 105 °C.

3.3   Additional definitions are given in Sections 3.0 and 18.0 of Method 1690.
Draft, January 2001                                                                        A-l

-------
Method 1690
4.0  Interferences

4.1   Sampling, subsampling, and pipeting multi-phase samples may introduce serious errors (Reference
      16.1).  Make and keep such samples homogeneous during transfer.  Use special handling to ensure
      sample integrity when subsampling. Mix small samples with a magnetic stirrer.  If visible suspended
      solids are present, pipet with wide-bore pipets.  If part of a sample adheres to the sample container,
      intensive homogenization is required to ensure accurate results. When dried, some samples form a
      crust that prevents evaporation; special handling such as extended drying times are required to deal
      with this. Avoid using a magnetic stirrer with samples containing magnetic particles.

4.2   The temperature and time of residue drying has an important bearing on results (Reference 16.1).
      Problems such as weight losses due to volatilization of organic matter, and evolution of gases from
      heat-induced chemical decomposition, weight gains due to oxidation, and confounding factors like
      mechanical occlusion of water and water of crystallization depend on temperature and time of
      heating.  It is therefore essential that samples be dried  at a uniform temperature, and for no longer
      than specified.  Each sample requires close attention to desiccation after drying. Minimize the time
      the desiccator is open because moist air may enter and be absorbed by the samples.  Some samples
      may be stronger desiccants than those used in the desiccator and may take on water. If uptake of
      water by a sample is suspected, the operator should weigh the sample to see if it gains weight while
      in the desiccator. If the sample is indeed taking on water, then a vacuum desiccator should be used.

4.3   Residues dried at 103 °C to  105 °C may retain some bound water as water of crystallization or as
      water occluded in the interstices of crystals.  They lose CO2 in the conversion of bicarbonate to
      carbonate. The residues usually lose only slight amounts of organic matter by volatilization at this
      temperature. Because removal of occluded water is marginal at this temperature, attainment of
      constant  weight may be very slow.

4.4   Results for residues high in oil or grease may be questionable because of the difficulty of drying to
      constant  weight in a reasonable time.

4.5   The determination of total solids is subject to negative error due to loss of ammonium carbonate and
      volatile organic matter during the drying  step at 103 °C to 105 °C. Carefully observe specified
      ignition time and temperature to control losses of volatile inorganic salts if these are a problem.

5.0  Safety

5.1   Refer to  Section 5.0 of Method 1690 for  safety precautions

6.0  Equipment and Supplies
      NOTE:   Brand names, suppliers, and part numbers are cited for illustrative purposes only.  No endorsement
      is implied. Equivalent performance may be achieved using equipment and materials other than those specified
      here, but demonstration ofequivalent performance that meets the requirements of this method is the responsibility
      of the laboratory.
A-2                                                                              Draft, January 2001

-------
                                                                                     Method 1690
6.1   Evaporating Dishes—Dishes of 100-mL capacity.  The dishes may be made of porcelain (90-mm
      diameter), platinum, or high-silica glass.

6.2   Watch glass—Capable of covering the evaporating dishes (Section 6.1).

6.3   Steam bath for evaporation of liquid samples.

6.4   Desiccator—Moisture concentration in the desiccator should be monitored by an instrumental
      indicator or with a color-indicator desiccant.

6.5   Drying oven—Thermostatically-controlled, capable of maintaining a uniform temperature of 103 °C
      to 105 °C throughout the drying chamber.

6.6   Analytical balance—Capable of weighing to 0.1 mg for samples having a mass up to 200 g.

6.7   Reference weights—2 mg, 1000 mg, and 50 g class "S" weights.

6.8   Container handling apparatus—Gloves, tongs, or a suitable holder for moving and handling hot
      containers after drying.

6.9   Sample handling apparatus—Spatulas, spoonulas, funnels, or other equipment for transfer and
      manipulation of samples.

6.10 Bottles—Glass or plastic bottles of a suitable size for sample collection.

6.11 Rubber gloves (Optional).

6.12 No. 7 Cork borer (Optional).

6.13 Desiccant (Optional).

7.0   Reagents and  Standards

7.1   Reagent water—Deionized, distilled, or otherwise purified water.

7.2   Quality control spiking solution—If a  commercially available standard can be purchased that
      contains standard total solids, the laboratory may use that standard.  The laboratory may also
      prepare a spiking solution. One possible recipe is given below for a NaCl-KHP solution.

      7.2.1 Dissolve 0.10 g sodium chloride (NaCl) in 500 mL reagent water. Mix to dissolve.

      7.2.2 Add 0.10 g potassium hydrogen phthalate (KHP) to the NaCl solution (Section 7.2.1) and
            mix.  If the KHP does not dissolve readily, warm the solution while mixing.  Dilute to 1 L
            with reagent water.  Store at 4°C. Assuming 100% volatility of the acid phthalate ion, this
            solution contains 200 mg/L total solids, 81.0 mg/L volatile solids, and 119 mg/L fixed solids.
Draft, January 2001                                                                             A-3

-------
Method 1690
8.0  Sample Collection, Preservation, and Storage

8.1   Use resistant-glass or plastic bottles to collect sample for solids analysis, provided that the material
      in suspension does not adhere to container walls.  Sampling should be done in accordance with
      Reference 16.2. Begin analysis as soon as possible after collection because of the impracticality of
      preserving the sample. Refrigerate the sample at 4 °C up to the time of analysis to minimize
      microbiological decomposition of solids. Preferably do not hold samples more than 24 hours. Under
      no circumstances should the sample be held more than seven days. Bring samples to room
      temperature before analysis.

9.0  Quality  Control

9.1   Quality control requirements and requirements for performance-based methods are given in Section
      9.1 of Method 1690.

9.2   Initial demonstration of laboratory capability—The initial demonstration of laboratory capability is
      used to characterize laboratory performance and method detection limits.

      9.2.1 Method detection limit (MDL)—The method detection limit should be established for total
            solids using the QC spiking solution (Section 7.2).  To determine MDL values, take seven
            replicate aliquots of the diluted QC spiking solution and process each aliquot through each
            step of the analytical method. Perform all calculations and report the concentration values in
            the appropriate units.  MDLs should be determined every year or whenever a modification to
            the method or analytical system is made that will affect the method detection limit.

      9.2.2 Initial precision and recovery (IPR)—To establish the ability to generate acceptable precision
            and accuracy, the analyst shall perform the following operations:

            9.2.2.1    Prepare four samples by diluting the QC spiking  solution (Section 7.2) to 1-5
                       times the MDL. Using the procedures in Section 11, analyze these samples  for
                       total solids.

            9.2.2.2    Using the results of the four analyses, compute the average percent recovery (x)
                       and the standard deviation (s, Equation 1) of the percent recovery for total solids.
                                           Equation 1
                                                n-l
            Where:
                  n = number of samples
                  x = % recovery in each sample
                  s = standard deviation
A-4                                                                            Draft, January 2001

-------
                                                                                      Method 1690
            9.2.2.3     Compare s and x with the corresponding limits for initial precision and recovery
                        in Table 2 (to be determined in validation study). If s and x meet the acceptance
                        criteria, system performance is acceptable and analysis of samples may begin.
                        If, however, s exceeds the precision limit or x falls outside the range for
                        recovery, system performance is unacceptable.  In this event, correct the
                        problem, and repeat the test.

9.3   Laboratory blanks

      9.3.1 Prepare and analyze a laboratory blank initially (i.e.  with the tests in Section 9.2) and with
            each analytical batch.  The blank must be subjected to the same procedural steps as a sample,
            and will consist of approximately 25  g of reagent water.

      9.3.2 If material is detected in the blank at a concentration greater than the MDL (Section 1.3),
            analysis of samples must be halted until the source of contamination is eliminated and a new
            blank shows no evidence of contamination. All samples must be associated with an
            uncontaminated laboratory blank before the results may be reported for regulatory compliance
            purposes.

9.4   Ongoing precision and recovery

      9.4.1 Prepare an ongoing precision and recovery (OPR) solution identical to the IPR  solution
            described in Section 9.2.2.1.

      9.4.2 An aliquot of the OPR solution must be analyzed with each sample batch (samples started
            through the sample preparation process (Section 11) on the same 12-hour shift, to a maximum
            of 10 samples).

      9.4.3 Compute the percent recovery of total solids in the OPR sample.

      9.4.4 Compare the results to the limits for ongoing recovery in Table 2 (to be determined in
            validation study).  If the results meet the acceptance criteria, system performance is
            acceptable and analysis of blanks and samples may proceed.  If, however, the recovery of
            total solids falls outside of the range given, the analytical processes are not being performed
            properly. Correct the problem, reprepare the sample batch, and repeat the OPR test. All
            samples must be associated with an OPR analysis that passes acceptance criteria before the
            sample results can be reported for regulatory compliance purposes.

      9.4.5 Add results that pass the specifications in Section 9.4.4 to IPR and previous OPR data.
            Update QC charts to form a graphic representation of continued laboratory performance.
            Develop a statement of laboratory accuracy for each analyte by calculating the  average
            percent recovery (R) and the standard deviation of percent recovery (SR). Express the
            accuracy as a recovery interval from R-2SRto R+2SR. For example, if R=05% and SR=5%,
            the accuracy is 85-115%.

9.5   Duplicate analyses
Draft, January 2001                                                                             A-5

-------
Method 1690
      9.5.1 Ten percent of samples must be analyzed in duplicate. The duplicate analyses must be
            performed within the same sample batch (samples whose analysis is started within the same
            12-hour period).

      9.5.2 The total solids of the duplicate samples must be within 10% of the solids determination.

10.0 Calibration  and Standardization

10.1 Calibrate the analytical balance at 2 mg and 1000 mg using class "S" weights.

10.2 Calibration shall be within ± 10% (i.e. ±0.2 mg) at 2 mg and ± 0.5% (i.e.  ±5 mg) at 1000 mg.  If
      values are not within these limits, recalibrate the balance.

10.3 Place a 50 g weight and a 2 mg weight on the balance.  Verify that the balance reads 50.002 ±10%
      (i.e.  ±0.2 mg)

11.0 Procedure

11.1 Preparation of evaporating dishes—Heat dishes and watch glasses at 103 °C to 105 °C for 1 hour in
      an oven.  Cool and store the dried equipment in a desiccator. Weigh each dish and watch glass prior
      to use (record combined weight as "Wdlsh").

11.2 Preparation of samples

      11.2.1      Fluid samples—If the sample contains enough moisture to flow readily, stir to
                  homogenize, place a 25 to 50 g sample aliquot on the prepared evaporating dish. If the
                  sample is to be analyzed in duplicate, the mass of the two aliquots may not differ by
                  more than 10%. Spread each sample so that it is evenly distributed over the
                  evaporating dish.  Cover each sample with a watch glass, and weigh (record weight as
                  "Wsample").  Evaporate the samples to dryness  on a steam bath.
      NOTE:   Weigh wet samples quickly because wet samples tend to lose weight by evaporation.  Samples should
      be weighed immediately after aliquots are prepared.
      11.2.2     Solid samples—If the sample consists of discrete pieces of solid material (dewatered
                  sludges, for example), take cores from each piece with a No.  7 cork borer or pulverize
                  the entire sample coarsely on a clean surface by hand, using rubber gloves. Place a 25
                  to 50 g sample aliquot of the pulverized sample on the prepared evaporating dish. If
                  the sample is to be analyzed in duplicate, the mass of the two  aliquots may not differ by
                  more than 10%.  Spread each sample so that it is evenly distributed over the
                  evaporating dish. Cover each sample with a watch glass, and weigh (record weight as
                   ^* sample / •
A-6                                                                            Draft, January 2001

-------
                                                                                     Method 1690
11.3 Dry the samples at 103 °C to 105 °C for a minimum of 12 hours, cool to balance temperature in an
      individual desiccator containing fresh desiccant, and weigh.  Heat the residue again for 1 hour, cool
      it to balance temperature in a desiccator, and weigh. Repeat this heating, cooling, desiccating, and
      weighing procedure until the weight change is less than 5% or 50 mg, whichever is less.  Record the
      final weight as "Wtotal."
      NOTE:  It is imperative that dried samples be weighed quickly since residues often are very hygroscopic and
      rapidly absorb moisture from the air. Samples must remain in the desiccator until the analyst is ready to weigh
      them.
12.0 Data Analysis and Calculations

12.1  Calculate the % solids or the mg solids/kg sample for total solids (Equation 2).

                                          Equation 2
                              % total solids = —^	— * 100
                                            "sample ~ "dish
                              or
                              mg total solids    W. ., - W,.,
                              —	= —^	^-* 1 000 000
                                kg sludge     Wsample - Wdish

                  Where:
                        Wdlsh=Weightofdish (mg)
                        Wsample=Weight of wet sample and dish (mg)
	Wtotal=Weight of dried residue and dish (mg)	

12.2 Sample results should be reported as % solids or mg/kg to three significant figures.  Report results
      below the ML as < the ML, or as required by the permitting authority or in the permit.

13.0 Method Performance

13.1 Method performance (MDL and quality control acceptance criteria) will be determined during the
      multi-lab validation of this method.

13.2 Total solids duplicate determinations must agree within 10% to be reported for permitting purposes.
      If duplicate samples do not meet this criteria, the problem must be discovered and the sample must
      be run over.

14.0 Pollution Prevention

14.1 Pollution prevention details are given in Section 14 of Method 1690.
Draft, January 2001                                                                            A-7

-------
Method 1690
15.0 Waste Management

15.1 Waste management details are given in Section 15 of Method 1690.

16.0 References

16.1 "Standard Methods for the Examination of Water and Wastewater," 18th ed. and later revisions,
     American Public Health Association, 1015 15th Street NW, Washington, DC 20005. 2-59: Section
     2540 G (Total, Fixed, and Volatile Solids in Solid and Semisolid Matrices), 1992.

16.2 U.S. Environmental Protection Agency, 1992. Control of Pathogens and Vector Attraction in
     Sewage Sludges. Publ 625/R-92/013. Office of Research and Development, Washington, DC.

17.0 Tables, Diagrams, Flowcharts, and Validation Data

17.1 Tables containing method requirements for QA/QC will be added after the validation study has been
     performed.
A-8                                                                     Draft, January 2001

-------