PREPARATION AND STABILITY OF STANDARD REFERENCE AIR MIXTURES

Duane Kitzis

NOAA Earth System Research Laboratory, R/GMD, 325 Broadway, Boulder, CO 80305-3328
Carbon Cycle Greenhouse Gases group
and
Cooperative Institute for Research in Environmental Sciences
University of Colorado at Boulder

Updated August, 2009

Abstract

The ESRL/Carbon Cycle Greenhouse Gases group S group methods and materials for the preparation of standard reference mixtures of trace gases in natural air are presented in detail. The standards are prepared in high pressure aluminum cylinders to calibrate trace gas measurement systems in use by ESRL and others participating in the Global Atmospheric Watch Program of the World Meteorological Organization (WMO/GAW). The trace gas CO2 is primarily discussed, with some mention of common processes for CO and CH4 where applicable. Cylinders are conditioned with natural air and then pressurized to 135 atmospheres using an oil free compressor. The mole fractions of CO2, CH4, and CO as well as the isotopic ratios of CO2 can be altered by spiking the cylinder with low or high concentration gas prior to filling with a balance of natural air. The air is dried to a dew point of -80°C by passing the air through a very strong desiccant. The calculations and spiking methods for CO2 are presented for targeting CO2 mole fractions to within 2 µmol/mol (often written as “ppm”). The stability of CO2 and CH4 in aluminum high-pressure cylinders is better than the present analytical precision.

Introduction

High precision measurements of the trace gases CO2, CH4, and CO are being undertaken by many laboratories of various nations in order to better understand the biogeochemical cycles of these gases. There can be no integration of all these various measurement projects unless the data are all referenced to common, well defined calibration scales. The WMO has stated that the consistency required for optimally usable atmospheric CO2 measurements and comparison must be no worse than 0.1 µmol/mol for the northern hemisphere and 0.05 µmol/mol in the southern hemisphere [WMO, 1993]. Because of stringent quality controls necessary for the measurement of long-lived trace gases, stable reference gas standards are an integral part of any long-term measurement program. The Carbon Cycle Greenhouse Gases group (CCGG) of the ESRL Global Monitoring Division (formerly the Climate Monitoring and Diagnostics Laboratory) has been involved in maintaining CO2-in-air standards since the early 1970s and over the years has expanded to include methane, carbon monoxide, the stable isotopes of CO2 and methane, and many other trace gases. Today the standards preparation facility makes more than 300 standards per year and has enabled many laboratories world wide to be on the same WMO CO2, CH4, and CO [Novelli et al., 1991, 1994; Dlugokencky, 2005] scales. Most of the discussion in this paper is focused on CO2.

The need for natural air standards was recognized when standards were measured using different non-dispersive infrared analyzers (NDIR). The comparative work done by Komhyr et al. [1985] was prompted by calculations of CO2 pressure broadening in nitrogen and natural air presented by others [e.g., Bischof, 1975; Griffith et al., 1982]. Because of the difference in pressure broadening of the CO2 absorption lines by nitrogen and by natural air, considerable corrections need to be applied to NDIR measurements when using CO2-in-nitrogen reference standards to calibrate CO2-in-air measurements. These corrections are strongly analyzer dependent and may not be constant in time. From these comparisons it became apparent that standards of dried whole natural air needed to be created. Efforts started in 1979 [Thoning et al, 1987] and have continued through the present. All standards are calibrated at ESRL in Boulder Colorado. The NOAA/ESRL secondary standards, to which all other NOAA standard reference mixtures are tied, were measured approximately every 3 years against the WMO designated primary standards maintained by C. D. Keeling at the Scripps Institution of Oceanography. In 1995 the WMO designated the Carbon Cycle group as the CO2 Central Calibration Laboratory. The world network is presently tied to the 15 CCGG primary standards, now designated as the new WMO Primaries. The WMO Primary Standards are calibrated at regular intervals on our own manometric system [Zhao, 1997] and have been compared with the previous WMO calibration scale maintained at Scripps, and with a set of gravimetrically prepared primary standards at the National Institute for Environmental Sciences in Tsukuba, Japan. As of today, there appears to be no significant offset between the WMO scale and the other two scales in the range of ambient concentrations.

Site Description

Figure 1
Figure 1. The air standard preparation facility at Niwot Ridge, Colorado.
Click on image to view full size figure.

The air standard preparation facility is located in a biosphere preserve at 3040m altitude, 40° 02' N., 105° 32' W (Figure 1). This sub alpine-forested area has very limited vehicle access and usually provides non-urban continental air. The biosphere preserve is maintained by the National Forest Service and the University of Colorado, Mountain Research Station. The nearest paved road is approximately 10 km away at a much lower elevation.

Figure 2
Aerial view of Niwot Ridge

The CO2 mole fraction and its seasonal variations at the site can be approximately seen at the ESRL/CCGG flask sampling network site Niwot Ridge (NWR) (Figure 2). The flask sampling site is located just to the west of the air standard preparation facility and is at 40° 03' N., 105° 35' W. at 3475 m.

Figure 2
Figure 2. Niwot Ridge CO2 Flask Sample data. The red points are individual flask values, and the blue curve represents the long term trend of the data with the seasonal cycle removed.
Click on image to view full size figure.

Summer and winter transportation to the standards preparation site over the last 3 km varies due to snow coverage and conditions. The site is unreachable in winter except via snow-tracked vehicles or skis. To the East is the urban front range of Colorado, 25 km away. The densely populated Denver metropolitan area is 45 km to the southeast. To the south is the Arapahoe watershed, which is closed to all human traffic. To the north and west is the Arapahoe national forest and wilderness area. Basically south, west and north is very sparsely populated for 100 km or more.

Transporting compressed gas cylinders to the air standard preparation facility at Niwot Ridge.

Weather and wind trajectories

Figure 3
Figure 3. Average daily wind speed and direction at the filling facility at Niwot Ridge, CO for 8 years.

In general the site experiences clean well mixed air from western North America, and beyond that from the North Pacific Ocean. The daily averaged wind speed and direction, from 1990 to 1997 (Figure 3) shows that wind direction is predominantly from the west with more variable direction and weaker winds during the summer months. Some spring and summer days have short periods of local up slope easterly sourced air with a component of relatively polluted metropolitan air. This is very evident from high concentrations of correlated CO and CO2. It is common for winter winds to average 5-10 meters per second.

Figure 4
Figure 4. Wind trajectory clusters by season. Five day back trajectories averaged from 1985 to 2001 and grouped by the six most likely trajectories.
Click on image to view full size figure.

Figure 4 shows back trajectories at the filling site. These trajectories are averages from 1993 to 1997 and are divided into 6 of the most common patterns. All are predominantly from the west with spring and summer showing more continental air as compared to the longer winter trajectories. At present, trajectories can be observed on the ESRL web site.

Cylinder and Pumping Facility Description

  • Calibrations
  • Figure 4
    Figure 5. Primary standards are used to calibrate a smaller secondary set which in turn is used as reference for all other calibrations. A subset of the secondaries is chosen to bracket the standards to be calibrated.
    Click on image to view full size figure.

    A hierarchy of the CCGG CO2 standards is shown in Figure 5. The Central Calibration Laboratory for CO2 uses 15 Primary Standards, with a range of 246 to 520 µmol/mol. The Primaries have been measured 4 times at Scripps and 7 times on the CCGG manometric system [Zhao, 1997]. The calibration of these standards is presently used to provide the WMO [CO2-in-air] mole fraction scale. At present, 5 new primary standards are being added at CO2 mole fractions of approximately 700, 1000, 1500, 2000, 2500, and 3000 µmol/mol to provide reference standards for the oceanographic community. The ten secondaries in the range of 290 to 420 µmol/mol are calibrated relative tothe Primaries. All other standards, internal and external are calibrated on a single NDIR, which is referenced to these secondaries. The secondaries are typically used up in 2-5 years, whereas the Primaries are designed to last for many decades.

    With respect to CO2 stability, the calibration histories of the various internal ESRL standards were checked for calibration reproducibility and long term drift. The mean difference of repeated measurements of CO2 mole fractions between 325 and 425 µmol/mol is 0.002 +/-0 .014 (Histogram of distribution, Figure 6). This is for 381 cases since 1988, where individual cylinders were recalibrated in less than 6 months. Thus we can say that the precision of our measurement is 0.014 µmol/mol (one standard deviation).

    Figure 6
    Figure 6. Histogram of calibration differences of 381 standards, 325-345 micromole/mole mixing ratio. The bars represent the absolute value of the differences for repeat calibrations of the same standard within 6 months of each other. The number of cylinders in each subset is above the bars.

    Outside of this range the precision was found to be 0.1 for mole fractions from 200 to 325 µmol/mol and .25 µmol/mol for mole fractions from 425 to 500 µmol/mol (Histogram of distributions, Figure 7). The loss in precision of relatively low and high CO2 mole fractions is due to the limitations of measuring toward the end of the range of the WMO scale and the slight degradation in measurement precision at high mole fractions. This estimate of reproducibility takes all cylinders into account and may contain some cases of real drift and so is a worst-case scenario.

    The uncertainty stated by the CO2 calibration lab is 0.07 µmol/mol (one sigma). The reproducibilityand internal consistency of the standards in the range between 325 and 425 µmol/mol is 0.04 µmol/mol over a two-year period at the 2-sigma confidence level. If a cylinder is recalibrated at the end of its useful volume, the estimate of the stability of the calibrated value is limited by the reproducibility of our measurements (0.028 at the two sigma level), and necessary assumptions about the history of the drift, if any, between calibrations should include this uncertainty.

    Figure 7
    Figure 7. Histogram of calibration differences of the same standards, divided into 25 micro-mole/mole groups.

  • Cylinders
  • We use brass, pack-less, taper threaded valves in almost all cylinders with Teflon tape as the thread sealant. The Ceodeux[TM] valves have a soft seat, which does not become difficult to turn or seal over years of use. We have tested several types of cylinders, and surface treatments, and have found no evidence of specific coatings improving CO, CO2 or CH4 stability over cleaned, bare, uncoated aluminum. Luxfer aluminum, alloy 6061, cylinders with their standard cleaning preparation are the most used cylinder in CCGG/ESRL. Luxfer, at present, provides high heat treatment, proprietary acid wash (with no etching of the surface), steam cleaning and forced air drying. The steel cylinders used in the past exhibited too much drift in CO2 mole fractions. Half of these drifts were on the order of 0.05 µmol/mol CO2 per year or more different rates or of different sign in each cylinder. Aluminum cylinders have much better stability than steel for CO and CO2. CH4 seems to be stable in any type of cylinder we use, be it aluminum, surface treated or steel, as long as the air has been dried to a low dew point. The same cylinders are used for stable CO2 isotopes and have exhibited no measurable drift [Trolier, 1996]. There have been some cases of major drift (.05 per mil 13C per week) of the CO2 stable isotopes, which were attributed to Teflon paste used as the cylinder valve sealant and also to unknown regulator contamination. There may be some evidence for long term drift of CO in aluminum cylinders as exhibited in smaller high pressure cylinders [Paul Novelli, ESRL, personal communication]. This drift however is on the order of the measurement precision and has not yet been well quantified.

    Figure 8
    Figure 8. Distribution of drifting cylinders. Based on 125 cylinders which have a calibration history greater than 2 years. One standard deviation of this distribution is 0.016 ppm. Number of cylinders in each subset is shown above the bars.

    A histogram of CO2 mole fraction drift is shown in Figure 8 for 125 cylinders, which have a calibration history greater than 2 years. This shows drift in both directions. There are some cases (8%) with evidence of drift greater than 0.045 µmol/mol/year. For this reason we recommend that standards be recalibrated during long use and at the end of their useful volume. The drifts, as shown in the histogram, cannot presently be predicted. Drift in most of our standards is often undetectable for histories less than 2.5 years. In some documented cases, the drift has been attributed to H2O mole fractions greater than 5 µmol/mol, stainless steel valves, or valves with packing materials that preferentially absorb trace gases. There may be an increasing trend in the CO2 with decreasing cylinder pressure below 20 atmospheres. This is different in each cylinder and may not be present at all.

    All cylinders are hydro-tested after manufacturing and need to be retested periodically. All cylinders are stamped with the month and year of the last hydro-test and, if out of date, must be retested before they can be shipped. Currently the period for aluminum cylinders is five years after the stamped date. In the United States hydro-testing can only be done by companies certified by the U. S. Department of Transportation (DOT). The process consists of removing the valve and attaching the cylinder to a water/hydraulic pump system,which displaces all the air in the cylinder. The cylinder is immersed in a water bath of measured volume and the water in the cylinder is pressurized to 1.67 times the working pressure of the cylinder. The elastic expansion is measured via the outside water displacement, pressure is released in the cylinder and the permanent expansion is measured. The typical cylinder we use (Size AL-150) elastically expands by about 190 cc and is permanently expanded 0-2 cc after the pressure is released. The cylinder is deemed unusable if it permanently expands by more than 10 percent of the elastic expansion. We have never had a cylinder fail the test.

    Hydro-testing is followed by a drying /passivation process by the Scott Marrin company. After this process we have them filled with 20 atmospheres of ultra pure(TM) air. Ultrapure air is whole air compressed at Scott Marrin and is highly filtered with mole sieves to trap most trace gases and water (Jack Marrin, Scott Marrin Inc., personal communication).

  • Pump
  • The compressor presently in service is a RIX SA6B. It is a three-stage oil-less, piston compressor commonly used for recharging Scuba tanks and is used as sold with no alterations. The three stages of compression are to 6 atm, 34 atm and 238 atmospheres respectively. The compression rings are made of filled PTFE plastic and require no lubrication. There are moisture separators between the second and third stage and at the output of the compressor. The third stage is designed with a 102 atmosphere back pressure regulator to maintain this minimum pressure against the piston. The flow at 3040 m altitude for this compressor is 126 liters per minute at ambient pressure. The temperature of the compressed air in the output-cooling coil can reach 121° Celsius. These pumps are not airtight and some alterations of trace gases may be observed. Others have made some modifications to this pump to lower the air temperature from compression and improve its use for ambient air sampling (John Mak, 1994). Some of the improvements may result in some loss of pump performance however. I am presently trying some of the methods to make the pump run cooler which do not affect the air stream.

  • Drying System
  • The stability of many trace gases is sensitive to water vapor and especially sensitive to liquid water. Therefore the drying step is important in the preparation of any gas standards. The air exiting the compressor is passed through a Balston, 1 micron coalescent filter with a manual drain. The air then passes through 2 stainless steel tubes of 50 cm length in series. These have a 2.5 cm cross section and are filled with Anhydrous Magnesium Perchlorate ( Mg(ClO4)2 ) (Anhydrone, J T Baker). This system reduces the water vapor in the air stream to typically less than 1 µmol/mol. After approximately 52,000 liters of air has passed through the traps, the upstream trap is replaced with the downstream trap, which is replaced with a fresh one. The second moisture trap is for complete redundancy as one will dry the air stream to less than 1 µmol/mol. Water accumulated in the compressor moisture separators and the coalescent filter is blown out before and after each cylinder pressurization. Another possible desiccant is Aquasorb (Mallinckrodt). Aquasorb is phosphorus pentoxide (P2O5). However CO2 was found to increase by 2-5 µmol/mol with new drying material. This problem and the safety issue of cycling phosphoric acid in a high pressure system prompted the decision to discontinue use.

  • Pressure Measurement and Compressor Control
  • The measurement, display and control of the compressor is done with an Omega series DP205-S controller with a PX302 pressure transducer. The output solid state relay, controlled by the alarm setting, activates a set of Crydom DC control, solid state opto-isolated relays. These relays control the 220 V compressor power and is normally off unless turned on via the Omega. The setting of the control is 135 atmospheres for a full cylinder.

  • Complete System
  • Figure 9
    Figure 9. Pumping System
    The pumping system is shown in Figure 9. The air is drawn through a 10 m tall, 2.5 cm diameter stainless steel intake. All high-pressure flex hoses are stainless steel braided, PTFE lined with 1/4" VCR connections. All connections are 1/4" NPT or 1/4" VCR with silver coated nickel gaskets. A 1/2 meter length between the compressor and the Balston filter allows for some extra cooling of the air and therefore better condensation, before going through a coalescent filter. This was found to improve the performance of the filter significantly. All parts are secured to a rack for safety. The pop off valves and the check valve are for safety. Pop-off valves are set to 170 atmospheres on either end in case of a trap clogging, and the check valve will prevent the cylinder from dumping all the air volume in the case of a line break. 7 micron filters are in line at the end to prevent dirt from getting to the safety valve or the pressure transducer. This system can either fill 5 steel cylinders as a ballast system or directly fill the target standards. The 43.3 liter volume steel tanks can be pressurized to 238 Atm. and used to pressurize the usual 29.5 liter aluminum standards to 134 Atm. We fill aluminum cylinders of various sizes, 5.9, 15.8, 29.9, or 46.4 liters.

  • Air/Cylinder Measurement System
  • New standards are tested for H2O content with a MEECO Waterboy, which uses a P2O5 conductivity method and is set up to measure from 1 to 1000 µmol/mol H2O. The maximum allowable H2O mole fraction in a filled cylinder is 5 µmol/mol. Most cylinders test between 0.5-1.0 µmol/mol. This corresponds to dew points less than -75°C and usually less than -80°C when expanded, and less than -30°C under full pressure.

    Figure 7

Air Standard Preparation

  • Conditioning
  • When a cylinder is to be refilled and contains air with the trace gases near ambient, conditioning is not usually necessary. New cylinders have ultra pure air and cylinders with undesirable concentrations left in them must be purged of the old air to have the walls conditioned to ambient air. Drifts have been observed due to evacuating the cylinder down to vacuum before filling, which have persisted for months. Drifts due to spiking, discussed below, have been eliminated by filling the cylinder immediately after spiking and filling the cylinder laid down. There is no evidence to show drifting due to the filling process described below. As a result of these experiences and to remain conservative about long-term effects we do not know about yet, the following steps have been adopted by CCGG as procedure;

    1. Blow off the residual air down to ambient pressure. This must be slow enough to avoid water condensation at the valve.
    2. Fill to 20 atmospheres and blow down to ambient pressure two times to purge old air. This uses mass flow to flush out the original air.
    3. Fill to 33 atmospheres and let sit for at least a week for cylinder wall equilibration.

    From this flushing procedure the original air left in the cylinder is less than .01 % of the volume at the time of spiking and less than 5E-5 % after the final filling. Drifts in the stable isotopes of CO2 have been observed, even after this procedure, and have stabilized over time.

    Conditioning and Filling cylinder
    Conditioning and filling cyliners.

  • Filling
  • The 33 atmospheres of air is blown down to ambient pressure. The standard can now be filled with ambient air or the trace gas can be adjusted via one of the next two steps. The cylinders are rated to 153 atmospheres and are filled to 149 atmospheres in about 35 to 45 minutes. When the air cools the pressure in the tank decreases to 136 atmospheres.

  • Targeting Above Ambient Levels
  • For increased standards above ambient we use one of several cylinders with very high mole fractions of the trace gas of interest. A small amount of the spiking gas is put in a 1 liter transfer volume at higher than ambient pressure. This over pressure is allowed to flow into the cylinder until pressures are equilibrated. The formula for increased trace gas values is as follows;

    Trace gas increase = K * (Target mole fraction - Ambient mole fraction)

    Where K is a constant for trace gas increase/ pressure of the spiking gas in the transfer volume.

    The over pressures we use in the transfer volume are on the order of 0.1 atm. of over pressure per increased µmol/mol CO2 in a standard 29.5 l. cylinder. Because the composition of the spiking gas varies, this value varies and must be found empirically with each cylinder of spike gas. Once a few cylinders are spiked and calibrated the constant K is the same over a wide range of increased trace gases. Examples of spike gas mole fractions are 10 % CO2 in air, 1.5 % CO in air and 2.5% CH4 in air.

  • Targeting Below Ambient Levels
  • Lowering a trace gas below ambient levels requires that some volume of air with none of the trace gas be put in the cylinder.

    Pf * Vf = Pz * Vz + Pa * Va

    Where
    Pf is the final filling pressure of the standard at room temperature
    Vf is the final mole fraction of the trace gas
    Pz is the pressure of the zero gas put in the standard first, to lower the mole fraction below ambient.
    Vz = 0; the mole fraction of the zero gas.
    Pa = Pf - Pz; the balance pressure of ambient air necessary to fill the standard to 149 atmospheres.
    Va is the ambient mole fraction of the trace gas.

    The zero air is created with specific chemical traps for each trace gas. CO2 can be removed with sodium hydroxide, Ascarite (Thomas). CO is removed with Schutze reagent. The zero air from both trapping agents has been tested and found to be less than 5 µmol/mol for CO2 and immeasurable trace amounts for CO. The isotopes of CO2 have been manipulated with mixtures composed of light or heavy isotopes. The same type of trap holder is used as for the desiccant traps. Specific trace gas traps can be installed between the Mg(ClO4)2 filled tubes. The second trap will catch any residual moisture released by the trace gas traps. The zero air is put directly into the cylinder if very low final mole fractions are targeted. The more precise method requires the filling of an intermediate ballast with the zero air and then transferring the partial pressure necessary to the standard cylinder. The necessary partial pressure of the zero gas is calculated from the above formula, re-written as follows;

    Pf * Vf = [Pf - Pz] * Va
    Pf * [Va - Vf] / Va = Pz

    The standard cylinder is filled to the pressure P(z) from the ballast whole air (with the specific trace gas removed) and then filled to the final pressure with dry natural air of ambient mole fractions.

  • Adjusting Trace Gas Levels in a Cylinder
  • Occasionally, the targeted concentration is missed by an amount, which requires adjustment. If the mole fraction of the standard is closer to ambient than the target concentration, it is easiest to blow this off and start over. Experimenting with introducing small volumes of high pressure (high or low mole fractions) into the air stream have proven very time consuming due to the time necessary for the cylinder to become evenly mixed before obtaining a usable measurement from the standard.

    If, however the final mole fraction turns out to be farther away from ambient than the desired targeted trace gas mole fraction, it is possible to adjust the standard toward the desired concentration. The same partial pressure formula above can be used, adding a term for the gas already in the cylinder and blowing the cylinder down to the necessary partial pressure. When the cylinder is filled to the final pressure again, with ambient levels of the trace gas the mole fraction will be adjusted toward the targeted value fairly precisely.

    In the case where the mole fraction in the cylinder is too low and must be increased;

    Pf * [Va - Vf] / [Va - Vm] = Pm

    where Vm is the mole fraction of the gas in the cylinder. and Pm is the pressure to lower the standard to, before refilling with ambient air.

    In the case where the mole fraction in the cylinder is too high and can be lowered;

    Pf * [Vf - Va] / [Vm - Va] = Pm

  • Initial Calibration and Use of Trace Gases
  • After the standard has been filled, checked for low water content, and checked for targeting of the trace gas, it is transported to the WMO Central Calibration Laboratory in Boulder, CO. The standard is then calibrated three times with approximately one week intervals between each calibration. For CO2 these measurements are done with a non-dispersive infrared analyzer, comparing them to secondary ESRL standards that are closely tied to the primary WMO standards. The trace gas mole fraction reported includes measurement repeatability as well as a check on cylinder stability. These standards are used at our base line observatories, by international collaborative projects and by many national laboratories as tertiary standards. There is evidence that as the volume of pressurized air is used up, the trace gas can come off the walls in disproportionate amounts. This has sometimes been seen in cylinders with CO2, when a standard is used at pressures below 20 atm. The CO2 mole fraction may increase with pressure loss. This effect is variable and may not be measurable in all tanks. Also, when air is used from the cylinder at a high flow rate, the cylinder head may cool which could produce drift. To make sure we can get a representative final calibration we discontinue use when the cylinder pressure becomes lower than 34 atm.

References

Bischof, W., The influence of the carrier gas on the infrared gas analysis of atmospheric CO2, Tellus, No. 27, pp. 59-61, 1975.

Dlugokencky, E., R.C. Myers, P.M. Lang, K.A. Masarie, A.M. Crotwell, K.W. Thoning, B.D. Hall, J.W. Elkins, and L.P. Steele, Conversion of NOAA atmospheric dry air CH4 mole fractions to a gravimetrically prepared standard scale, J. Geophys Res. 110, D18306, doi: 10.1029/2005/JD006035.

Geophysical Monitoring for Climatic Change, Summary Report 1972, No. 1, U.S. Department of Commerce, National Oceanic and Atmospheric Administration, Environmental Research Labs, Boulder Colorado, January 1974.

Griffith, D. W. T., Calculations of carrier gas effects in nondispersive infrared analyzers, I. Thoery., Tellus, No. 34, pp. 376-384, 1982.

Komhyr, W.D., T.B. Harris, L.S. Waterman, Calibration of Non-dispersive Infrared CO2 Analyzers with CO2-in-Air Reference Gases, Journal of Atmospheric and Ocean Technology, Vol.2., No. 1,March 1985.

Mak, John, C.M. Brenninkmeijer, Compressed Air Technology for Isotopic Analysis of Atmospheric Carbon Monoxide, Journal of Atmospheric and Oceanic Technology, Vol. 11, No. 2, pp 425-431, April 1994.

Novelli, P. C., L. P. Steel, and J. W. Elkins, The development and evaluation of a gravimetric reference scale for measurements of atmospheric carbon monoxide, J. Geophys. Res., 96, No. D7, 22,461-22, 476, 1991.

Novelli, P. C., J. E. Collins, Jr., R. C. Myers, G. W. Sachse, and H. E. Scheel, Reevaluation of the NOAA/ESRL carbon monoxide reference scale and comparisons with CO reference gasses at NASA-Langley and the Fraunhofer Institut, J. Geophys. Res., 99, No. D6, 12, 833-12, 839, 1994.

Thoning, K.W., P. Tans, T.J. Conway, and L.S. Waterman, NOAA/GMCC calibrations of CO2-in-air reference gases:1979-1985, NOAA Tech. Memo. ERL ARL-150, NOAA Environmental Research Laboratories, Boulder, CO., 1987.

Trolier, M., J.W. White, P.P. Tans, K.A. Masarie, and P.A. Gemery, Monitoring the isotopic composition of atmospheric CO2: Measurements from the NOAA Global Air Sampling Network, J. of Geophys. Res., Vol. 101, No. D20, pp. 25,897-25,916, Nov. 20, 1996.

World Meteorological Organization Global Atmosphere Watch, Report of the Seventh WMO Meeting of Experts on Carbon Dioxide Concentration and Isotopic Measurement Techniques, WMO/TD-No. 669, No. 88 Rome, Italy, 1993.

World Meteorological Organization Global Atmosphere Watch, Report of the Fourth WMO Meeting of Experts on the Quality Assurance/Science Activity Centers (QA/SACs) of the Global Atmosphere Watch, WMO/TD No. 689, No. 104, Garmisch-Partenkirchen, Germany, 1995.

World Meteorological Organization (WMO), Report of the experts meeting on carbon dioxide monitoring, WMO network for monitoring background air pollution, Pub. EP-1/Doc. 5, Geneva, 1975.

Zhao, C. L., P. Tans, K. W. Thoning, A high precision manometric system for absolute calibrations of CO2 in dry air, J. of Geophys. Res., Vol. 102, No. D5, pp. 5885-5894, March 20, 1997.