Automounter Equipment Requirements
Puck Set / Tools
Bases
Pins
Automounter Spreadsheets
The automounter requires a complete puck tool set and specially
sized bases and pins. Crystals will need to be pre-mounted in the
pucks prior to arrival for beamtime.
Puck Set / Tools
Pucks Sets can be ordered from:
Crystal Positioning Systems,
www.crystalpositioningsystems.com
or
MiTeGen, www.mitegen.com.
Qty. Item Price/Each
1 10" bent cryo tong
1 10" straight cryo tong
1 Puck pusher
1 Puck wand
7 Cryo puck
7 Magnetic cryo puck lid
1 Cryo puck shipping cane*
1 Puck separator tool
Total ~ approximately $3100
*Shipping cane for Taylor Wharton CP100 shipping dewar or 34HC storage dewar.
Bases
We advise users to use ONLY "ALS style" magnetic bases: either
Hampton part HR4-779, CrystalCap Magnetic (ALS) or MiTeGen, Part Numbers B1A, B3A or B3S.
Hampton Research CrystalCap Magnetic
Part Number HR4-779 (ALS)
http://hamptonresearch.com/product_detail.aspx?cid=24&sid=135&pid=378
or
MiTeGen
Part Numbers B1A, B3A or B3S
http://www.mitegen.com/products/bases/bases.shtml
Pins
The only pins which can be used with the robot are the "Hampton 18
mm" pins. This corresponds to breaking off the two end segments of a
Hampton pin.
The image below details the assembled ALS-style base+pin ("18 mm")
which is the preferred design for the automounter. The dimensions are fairly
critical here, and the most important one from the automounter's standpoint
is the distance from the inside mounting surface (which corresponds
with the top of the goniometer magnetic mount) to the center of the
crystal = 20.3mm ± 1mm.
It appears that the SSRL-style pins with the copper stem do not work well.
Automounter Spreadsheets
JBluIce requires a Microsoft Excel spreadsheet file with samples descriptions
to be loaded prior to using the automounter. The automounter templates are avalible
for downloading through this link or
after logging onto GM/CA computers, site files can be copied from the /px/data/
directory. The file names are Rigaku_template.xls
for Rigaku pucks and
ALS_template.xls for ALS pucks or
SSRL UniPucks. The number of rows is used to determine which style puck is
loaded; so do not alter the number of rows. The spreadsheet may be populated
with information or left blank.
Only three of the columns present in the spreadsheet are used by the
screening system as described here:
Port: |
refers to a port in the puck. The robot will use this port
to retrieve the sample. Don't change this column. |
CrystalID: |
this label is used by the screening system to generate filenames. It is
recommended that a unique ID be used for each sample. |
Directory: |
If provided, this field will be used to generate subdirectories for each
sample under the root screening directory. The default template Excel
spreadsheet uses the Port as the subdirectory name. |
|
|
ATTENTION: |
Please do not put any spaces into the spreadsheet
filename or in Port, CrystalID and Directory columns. Instead, use
underscores or dashes when a separator is needed. |
|