Automounter Equipment Requirements

Puck Set / Tools
Bases
Pins
Automounter Spreadsheets

The automounter requires a complete puck tool set and specially sized bases and pins. Crystals will need to be pre-mounted in the pucks prior to arrival for beamtime.

 Puck Set / Tools

Pucks Sets can be ordered from:

Crystal Positioning Systems, www.crystalpositioningsystems.com or MiTeGen, www.mitegen.com.

Qty. Item Price/Each
1 10" bent cryo tong
1 10" straight cryo tong
1 Puck pusher
1 Puck wand
7 Cryo puck
7 Magnetic cryo puck lid
1 Cryo puck shipping cane*
1 Puck separator tool
Total ~ approximately $3100

*Shipping cane for Taylor Wharton CP100 shipping dewar or 34HC storage dewar.

 Bases

We advise users to use ONLY "ALS style" magnetic bases: either Hampton part HR4-779, CrystalCap Magnetic (ALS) or MiTeGen, Part Numbers B1A, B3A or B3S.

Hampton Research CrystalCap Magnetic
Part Number HR4-779 (ALS)
http://hamptonresearch.com/product_detail.aspx?cid=24&sid=135&pid=378

or

MiTeGen
Part Numbers B1A, B3A or B3S
http://www.mitegen.com/products/bases/bases.shtml

 Pins

The only pins which can be used with the robot are the "Hampton 18 mm" pins. This corresponds to breaking off the two end segments of a Hampton pin.

The image below details the assembled ALS-style base+pin ("18 mm") which is the preferred design for the automounter. The dimensions are fairly critical here, and the most important one from the automounter's standpoint is the distance  from the inside mounting surface (which corresponds with the top of the goniometer magnetic mount) to the center  of the crystal = 20.3mm ± 1mm.

It appears that the SSRL-style pins with the copper stem do not work well.

 Automounter Spreadsheets

JBluIce requires a Microsoft Excel spreadsheet file with samples descriptions to be loaded prior to using the automounter. The automounter templates are avalible for downloading through this link or after logging onto GM/CA computers, site files can be copied from the /px/data/ directory. The file names are Rigaku_template.xls for Rigaku pucks and ALS_template.xls for ALS pucks or SSRL UniPucks. The number of rows is used to determine which style puck is loaded; so do not alter the number of rows. The spreadsheet may be populated with information or left blank.

Only three of the columns present in the spreadsheet are used by the screening system as described here:

Port: refers to a port in the puck. The robot will use this port to retrieve the sample. Don't change this column.
CrystalID: this label is used by the screening system to generate filenames. It is recommended that a unique ID be used for each sample.
Directory: If provided, this field will be used to generate subdirectories for each sample under the root screening directory. The default template Excel spreadsheet uses the Port as the subdirectory name.
   
ATTENTION: Please do not put any spaces into the spreadsheet filename or in Port, CrystalID and Directory columns. Instead, use underscores or dashes when a separator is needed.

 

 

 