Bookmark and Share

Calibration of Cryogenic Resistance Thermometers

The NIST Low Temperature Calibration Facility (LTCF) can accommodate most capsule and miniature type resistance thermometers suitable for use in vacuum within the range 0.65 K to 83.8 K, with special tests possible as high as 165 K. Calibrations of cryogenic resistance thermometers at NIST are performed by comparison to standard thermometers on the International Temperature Scale of 1990 (ITS-90) for the following thermometers:

  • capsule standard platinum resistance thermometers (SPRTs)
  • rhodium-iron
  • germanium
  • platinum cobalt
  • zirconium nitride, ruthenia

Additional special tests are available for non-standard cryogenic thermometers.

The Standard Platinum Resistance Thermometer Calibration Laboratory (SPRTL) calibrates thermometers from 84 K to 1235 K (Ar TP to Ag FP). Information for these calibrations are found on the ITS-90 Defining Artifacts webpage.

How to Arrange a Calibration

  1. Contact the staff member in charge of the calibration service for your artifact with any technical or calibration cost questions.
  2. Calibrations are performed on a first come first serve basis.
  3. Ship your artifact with a copy of the purchase order. Use a sturdy box in which we can return the calibrated instrument.
  4. The purchase order may be sent to contact by e-mail, fax, or mail. Please include the following information:
  • service ID for the calibration
  • model and serial number of your artifact
  • customer contact name, address, phone, and e-mail
  • return shipping method (e.g., shipping account number)  
Mailing Address   Shipping Address
NIST   NIST
Attn: Wes Tew   Attn: Wes Tew
100 Bureau Drive   100 Bureau Drive
MS 8363   Bldg. 221, Rm. B308
Gaithersburg, MD 20899   Gaithersburg, MD 20899



Description and Cost of Services

Service ID

Fee, US$

Description of Service

33010C

11,151

Capsule SPRT (13.8 K to 30 °C)

33020C

11,447

Capsule SPRT (13.8 K to 157 °C)

33030C

11,742

Capsule SPRT (13.8 K to 232 °C)

33031C

 8,295

Capsule SPRT (24.5 K to 30 °C)

33032C

 8,589

Capsule SPRT (24.5 K to 157 °C)

33033C

 8,883

Capsule SPRT (24.5 K to 232 °C)

33040C

 7,429

Capsule SPRT (54 K to 30 °C)

33050C

 7,723

Capsule SPRT (54 K to 157 °C)

33060S

 8,018

Capsule SPRT (54 K to 232 °C)

33140C

12,273

Rhodium-Iron or Platinum-Cobalt Resistance Thermometer (0.65 K to 24.6 K)

33141C

15,077

Rhodium-Iron or Platinum-Cobalt Resistance Thermometer (0.65 K to 83.8 K)

33142C

12,011

n-Type Germanium Resistance Thermometer (0.65 K to 24.6 K)

33320C

      68

Additional copy of Calibration Report at time of test

33330C

    433

Additional copy of Calibration Report at a later date

33340C

    904

Minimum charge for an unsuitable thermometer

33355S

at cost

Special tests of cryogenic resistance thermometers




Calibration Measurement Uncertainties

ITS-90 Defining Fixed-Points for Capsule SPRTs

Fixed-Point Cell

U (k=2), mK

e-H2 TP

0.41

e-H2 VP

0.89

e-H2-VP

0.23

Ne TP

0.40

O2 TP

0.19

Ar TP

0.32

Hg TP

0.49

TPW

0.20




ITS-90 Subrange for Capsule SPRTs

Temperature
Range, K

Maximum
U (k=2), mK† 

13.8033 to 273.16

1.36


† The ITS-90 temperature subrange uncertainties include the propagated fixed-point uncertainties and the contribution from non-uniqueness types I and III.



Rhodium-Iron Resistance Thermometers

Temperature
Range, K

Maximum
U (k=2), mK

0.65 to 2

0.50

2 to 5

0.32

5 to 9

0.23

9 to 14

0.30

14 to 17

0.34

17 to 20.5

0.40

20.5 to 24.5

0.50

24.5 to 54

0.58

54 to 84

0.40



General information regarding NIST calibration services may be found on the NIST Technology Services Calibration webpage