Absolute cryogenic infrared radiometry at the LBIR facility

Summary:

In support of defense programs in remote sensing, NIST maintains the scale for infrared radiant flux by using the Absolute Cryogenic radiometer (ACR) in a low background environment. The program activities include developing methods and standards for calibrating sensors for missile defense or remote sensing in general as well as performing calibrations of cryogenic blackbodies and detectors for the user community.

Description:

The Strategic Defense Initiative (SDI), and subsequently the Ballistic Missile Defense Organization (BMDO) of the U.S. Department of Defense (DoD), has motivated NIST’s work developing calibration methods and standards for space-based sensors used in missile defense. These on-board sensors must detect and measure, on parity with the radiation from the Earth, faint signals from small targets viewed against the cold background of space.  Because the flux levels at the detectors are so low, the sensors are often cooled to 77 K and below to achieve the required sensitivity.  Thus, NIST supports the pre-launch calibration activities of the missile defense projects by calibrating their blackbodies and infrared test chambers at low temperatures using NIST standards.  These calibrated blackbodies and test chambers are then used as transfer standards for the aerospace industry when performing pre-launch radiometric calibrations for DoD missile defense programs.

We have developed absolute cryogenic radiometers (ACRs) to serve as the primary standards for measuring optical power in a cold-space background environment.  These instruments are trap detectors where the power incident on a receiver cone is measured by electrical power substitution.  Due to the trapping nature of the ACR design, it has uniform sensitivity, close to unity, for all wavelengths of radiation from soft x-rays to microwaves.  The first ACR was designed and built in1989, having a noise floor of approximately 50 pW and a time constant of approximately 20 s. Since then, we have continued to improve our capabilities for power measurement. Currently, we are working on an ACR that can measure power (on the order of a picowatt, pW) with a noise floor at femtowatt (fW) levels.

Calibration services for the user community are carried out at the Low background infrared facility.  For more details about this facility, see Low background infrared (LBIR) facility.

Orginial absolute cryogenic radiometer (ACR)

Lead Organizational Unit:

pml

Customers/Contributors/Collaborators:

US Air Force - Arnold Engineering Development Center (AEDC), Tullahoma, TN

Raytheon Missile Systems Company, Tucson, AZ

Space Dynamics Laboratory, Logan, UT

Johns Hopkins Applied Physics Laboratory, Laurel, MD

Contractors/Guest Researchers:

Adriaan Carter
Timothy Jung
James Proctor
Greg Ly
Dale Sears
Tomoya Ohno

Contact

Absolute cryogenic infrared radiometry:
Simon Kaplan, Project Leader
301-975-2336 Telephone
301-840-8557 Facsimile

100 Bureau Drive, M/S 8441
Gaithersburg, MD 20899-8441