ICP Analysis of Metal/Metalloid Particulates From Solder Operations
[92 KB
PDF,
13 pages]
Related Information: Chemical Sampling -
Antimony & Compounds (as Sb),
Beryllium and Beryllium Compounds (as Be),
Cadmium,
Copper Dusts & Mists (as Cu), Copper
Fume (as Cu),
Lead, Inorganic (as Pb),
Silver, Metal & Soluble Compounds (as Ag),
Tin, inorganic compounds (except oxides) (as Sn),
Zinc Oxide Fume
|
Method no.: |
ID-206 |
|
|
Matrix: |
Air, Wipe (Smear Tab), or Bulk |
|
|
OSHA Time Weighted Average (TWA)
Permissible Exposure Limits: |
0.01 mg/m3 Silver (Ag)
0.002 mg/m3 Beryllium (Be)*
0.1 mg/m3 Cadmium (Cd) fume
0.1 mg/m3 Copper (Cu) fume
0.05 mg/m3 Lead (Pb)
0.5 mg/m3 Antimony (Sb)
2.0 mg/m3 Tin (Sn)
5.0 mg/m3 Zinc Oxide (ZnO) fume |
|
|
Collection Procedure: |
A personal sampling pump is used to draw a known volume of air through a mixed cellulose ester membrane filter contained in a polystyrene cassette. Wipe (smear tab) and bulk material are collected by grab sampling techniques. |
|
|
Recommended Sampling Rate: |
2 liters per minute |
|
|
Recommended Air Volumes
Be, Cd, Cu, Pb, Sb, Sn, ZnO:
If Ag is determined: |
480 L
960 L |
|
|
Analytical Procedure: |
Filters are digested with hydrochloric and nitric acids. Analysis is performed using Inductively Coupled Plasma-Atomic Emission Spectroscopy. |
|
|
Detection Limits: |
See Section 2 |
|
|
Validation Range: |
See Section 3 |
|
|
Method Classification: |
Validated analytical method |
|
|
Chemist: |
Dixon C. Cook |
|
|
Date: |
May, 1991 |
|
|
* This method may be used to determine STEL (0.005 mg/m3) or Ceiling (0.025 mg/m3) exposures to Be. At least a 30-min sample (2 L/min) should be taken for STEL determinations; at least 5-min for Ceiling monitoring. |
|
|
Commercial manufacturers and products mentioned in this method
are for descriptive use only and do not constitute endorsements by USDOL-OSHA. Similar products from
other sources can be substituted.
|
|
|
Division of Physical Measurements and Inorganic Analyses
OSHA Technical Center
Sandy, Utah
|
1. Introduction
1.1. Scope
1.1.1. This method describes the collection and analysis of airborne metal and metalloid particulates
from solder operations in industry. Time Weighted Average (TWA) air samples are collected using personal sampling pumps
and mixed-cellulose ester (MCE) filters. Analysis is by Inductively Coupled Plasma-Atomic
Emission Spectroscopy (ICP-AES). Bulk and wipe samples can also be collected and analyzed by this
method.
1.1.2. This method provides rapid simultaneous analysis and data reduction for a wide range of elements, eliminating the
necessity of separate analyses by conventional atomic absorption techniques.
1.1.3. This method was validated for 8 elements (Ag, Be, Cd, Cu, Pb, Sb, Sn, and Zn). A total of 21 elements are
analyzed, but 13 of these are determined for screening purposes only (Al, As, Ca, Co, Cr, Fe, Ni, Mg, Mn, Mo, Se, Si,
and V). Other elements can be added to or subtracted from the method. The capability for expanding the analysis to
other elements is dependent on laboratory instrumentation, and element solubility and stability in the acid matrix used
for digestion.
1.1.4. The elements validated are those commonly found in solders or soldering operations. The acid matrices used for
sample digestion (4:1 ratio of HCl:HNO3) and dilution (32% HCl/4%
HNO3) were selected to ensure solubility of the eight elements.
1.2. History
1.2.1. Previous to the introduction of ICP-AES, samples containing metallic particulates were digested in a
variety of ways and analyzed by Atomic Absorption Spectroscopy (AAS) at the OSHA Salt Lake City Analytical Laboratory
(SLCAL).
1.2.2. An ICP simultaneous spectrometer [Instruments SA (ISA), Model JY-32, Edison, NJ] was used to
validate this method.
This method is applicable to any simultaneous spectrometer. The validation of the JY-32 is described in the
backup report (8.1).
2. Detection Limits and Working Ranges (8.1)
2.1. Detection limits are listed below:
|
|
Qualitative Detection Limit |
Quantitative Detection Limit |
|
|
Element |
(µg/mL) |
(µg*) |
(µg/mL) |
(µg*) |
|
|
Ag |
0.018 |
0.45 |
0.061 |
1.5 |
Be |
0.00029 |
0.0072 |
0.00086 |
0.022 |
Cd |
0.0062 |
0.15 |
0.0205 |
0.51 |
Cu |
0.0078 |
0.19 |
0.026 |
0.65 |
Pb |
0.071 |
1.8 |
0.237 |
5.9 |
Sb |
0.14 |
3.5 |
0.47 |
12. |
Sn |
0.074 |
1.8 |
0.246 |
6.1 |
Zn |
0.0075 |
0.19 |
0.025 |
0.63 |
|
|
* 25-mL solution volume |
|
The quantitative limit is considered as the lower working range limit. These values are based on
instrument performance and will normally change over time.
2.2. The upper working range limit is 100 µg/mL for all validated elements except Be. Due to the high
sensitivity resulting from instrument settings and choice of wavelength, the upper limit for Be is
10 µg/mL.
3. Precision and Accuracy (8.1)
3.1. The precision and accuracy data for the 8 validated elements are listed below:
|
Element |
Range
(µg) |
Bias |
CV1
(pooled) |
Overall Error (±%)
(Total) |
|
|
Ag |
4.6 |
- |
19. |
0.006 |
0.054 |
11.5 |
Be |
0.47 |
- |
2.0 |
-0.075 |
0.037 |
14.9 |
Cd |
23. |
- |
100. |
0.048 |
0.041 |
13.1 |
Cu |
23. |
- |
94. |
0.065 |
0.038 |
14.2 |
Pb |
11. |
- |
48. |
-0.038 |
0.060 |
15.7 |
Sb |
126. |
- |
510. |
0.014 |
0.043 |
10.1 |
Sn |
476. |
- |
2,000. |
-0.008 |
0.043 |
9.4 |
Zn |
973. |
- |
2,900. |
0.031 |
0.039 |
10.9 |
|
These values are based on six samples at each of three concentration levels tested. Solutions of the
eight elements were spiked on MCE filters. These samples were then digested and analyzed as in Section 6. of this
method.
3.2. The eight elements reported above were spiked at 0.5, 1, and 2 times the PEL, assuming a 480-L air volume
for all elements except Ag where a 960-L air volume was used. Overall analytical errors for the elements
tested were within ±25%; the greatest value was 15.7% for Pb. The large overall error may be attributed to the fact that
Pb was validated near its detection limit.
4. Interferences (8.2)
High temperatures present in the plasma (5,000 to 8,000 °C) minimize most chemical and matrix interferences.
Interferences do
exist, however, and can be categorized as follows:
4.1. Physical interferences such as nebulization and transport effects are influences that determine
the rate and particle size in which analytes are delivered to the plasma. These effects are minimized by matching the acid
concentrations of samples and standards.
4.2. Chemical interferences are characterized by molecular compound formation, ionization effects, and solute
volatilization effects. These effects are not severe in ICP analysis and are minimized by matrix matching and careful
selection of operating conditions such as: Incident plasma source power, sample uptake rate, and plasma observation
height.
4.3. Spectral interferences include:
a. Unresolved overlap of molecular band spectra.
b. Overlap of a spectral line from another element.
c. Background from continuous or recombination phenomena.
d. Background from stray light.
The first effect (a) can be minimized by a careful selection of wavelengths for the reported elements.
The wavelengths selected for analysis using the JY-32 are shown in
Table 1. The other types of spectral
interferences (spectral overlap and elevated background) are minimized by software which performs interelement corrections.
This software assumes a linear relationship between the analyte concentration and interference within the working range
limits. The spectral interference correction equation typically used by ICP manufacturers is:
Corrected Concn = Calculated Concn -
Ai × CPi |
Where:
Ai = Correction factor
CPi = Concentration of the interfering element
Samples having analyte concentrations above the working range limits listed in Section 2.2. should be
diluted into range; interelement corrections may not be accurate above the working range. Further information and
experimentally determined interelement corrections for the validated elements are listed in the backup report
(8.1).
4.4. If necessary, supplemental background correction can be performed with additional software supplied by the instrument
manufacturer.
5. Sampling
5.1. Safety Precautions
5.1.1. Attach the sampling equipment to the worker such that it will not interfere with work performance or
safety.
5.1.2. Follow all safety practices that apply to the work area being sampled.
5.1.3. Wear impermeable gloves when taking wipe or bulk samples.
5.2. Equipment
5.2.1. Mixed cellulose ester (MCE) filters (0.8-µm pore size), cellulose backup pads, and twoor
three-piece cassettes, 37-mm diameter, (part no. MAWP 037 A0, Millipore Corp.,
Bedford, MA).
5.2.2. Gel bands (Omega Specialty Instrument Co., Chelmsford, MA) for sealing cassettes.
5.2.3. Sampling pumps capable of sampling at 2 liters per minute (L/min).
5.2.4. Assorted flexible tubing.
5.2.5. Stopwatch and bubble tube or meter for pump calibration.
5.2.6. Scintillation vials, 20 mL, (part no. 74515 or 58515, Kimble, Div. of Owens-Illinois Inc.,
Toledo, OH) with polypropylene or Teflon cap liners. If possible, submit bulk or wipe samples in these vials for
ICP analysis.
5.2.7. Smear tabs, (part no. 225-24, SKC Inc., Eighty Four, PA) for wipe sampling.
5.2.8. Gloves, disposable (for wipe sampling).
5.3. Sampling Procedure - Air Samples
5.3.1. Place a MCE filter and a cellulose backup pad in each twoor three-piece
cassette. Seal each cassette with a gel band.
5.3.2. Calibrate each personal sampling pump with a prepared cassette in-line to within ±10% of the
recommended flow rate of 2 L/min.
5.3.3. Attach prepared cassettes to calibrated sampling pumps (the backup pad should face the pump) and place in
appropriate positions on the employees or the workplace areas. Whenever possible, collect samples using the air volumes
listed below:
Suspected Analyte |
Recommended Air Volume |
|
|
All analytes except Ag |
480 L |
If Ag is also suspected |
960 L |
STEL or Ceiling for Be |
60 L (STEL), 10 L (Ceiling) |
5.3.4. If the filter becomes overloaded with particulate while sampling, another filter
cassette should be prepared. Consecutive samples using shorter sampling periods should be taken if overloading
occurs.
5.3.5. Place plastic end caps on each cassette after sampling.
5.3.6. Attach an OSHA-21 seal around each cassette in such a way as to secure the end caps.
5.4. Sampling Procedure - Wipe Samples
5.4.1. Wear clean, impervious, disposable gloves when taking each wipe sample to prevent sample contamination.
5.4.2. Moisten the wipe filters with deionized water prior to use.
5.4.3. If possible, wipe a surface area covering 100 cm2.
5.4.4. Fold the wipe sample with the exposed side in.
5.4.5. Transfer the wipe sample into a 20-mL scintillation vial and seal with vinyl or electrical tape. Securely
wrap an OSHA-21 seal length-wise from vial top to bottom.
5.5. Sampling Procedure - Bulk Samples
In order of laboratory preference, bulk samples may be one of the following:
1. a high-volume filter sample,
2. a representative settled dust (i.e. rafter) sample,
3. a sample of the bulk material in the workplace.
Transfer the bulk material into a 20-mL scintillation vial and seal with vinyl or electrical tape.
Securely wrap an OSHA-21 seal length-wise from vial top to bottom.
5.6. Shipment
5.6.1. When other compounds or elements are known or suspected to be present in the sampled air,
such information should be transmitted with the sample.
5.6.2. Request ICP/Solder analysis. If silver is suspected in any of the samples, request ICP/Solder-Silver
analysis.
5.6.3. Submit at least one blank sample with each set of air or wipe samples. Blank filter samples should be handled in
the same manner as other samples, except an actual sample is not taken.
5.6.4. Send the samples to the laboratory with the OSHA-91A paperwork.
5.6.5. Ship bulk samples separately from air samples. They should be accompanied by Material Safety Data Sheets, if
available. Check current shipping restrictions and ship to the laboratory by the appropriate method. The type of bulk
sample should be stated on the OSHA-91A and cross-referenced to the appropriate air
sample(s).
6. Analysis
6.1. Safety Precautions
6.1.1. Handle reagents and samples carefully. Use protective equipment such as: gloves, laboratory coats, safety glasses,
and an exhaust hood. Wear a fit-tested respirator if necessary. Clean up spills immediately.
6.1.2. Carefully prepare any solutions or samples (especially those containing acid) within a suitable ventilated
exhaust hood. The acids used to digest the samples are very corrosive and volatile.
6.1.3. Do not view the plasma directly.
6.1.4. Do not override the radio-frequency generator or torch box safety interlocks.
6.2. Equipment
6.2.1. Inductively coupled plasma/atomic emission direct-reading spectrometer, nebulizer, cooling unit for torch
assembly, computer, and radio-frequency (rf) generator.
6.2.2. Automatic sampler.
6.2.3. Peristaltic pumps (optional). Use one pump for automatic sampler rinse. Use the other pump for sample introduction
into the nebulizer.
6.2.4. Mass flow controller (optional). Use the controller to regulate nebulizer argon flow and sample uptake rate.
6.2.5. Borosilicate glass Phillips beakers, 125 and 250-mL.
6.2.6. Borosilicate glass volumetric flasks, 10-, 25-, 100-mL, and 1-L or 2-L. Use the larger flasks to prepare standard
solutions.
6.2.7. Hot plate capable of reaching 300 °C.
6.2.8. Mixed cellulose ester filters (0.45-µm pore size) and a filtering apparatus.
6.2.9. Amber colored bottle (1-L) with screw-top cap and Teflon cap liner.
6.2.10. Volumetric pipets, glass. Various sizes for sample dilutions and standard preparation.
6.2.11. Analytical balance (0.01 mg).
6.3. Reagents (Reagent grade or better)
Please see precautions in Sections 6.1.1.-6.1.2. before preparing any reagents having an acid matrix.
6.3.1. Deionized water (DI H2O).
6.3.2. Hydrochloric acid (HCl), concentrated (36.5 to 38% w/w).
6.3.3. Nitric acid (HNO3), concentrated (69 to 71% w/w).
6.3.4. Reagent blank - sample diluting solution (32% HCl/4% HNO3 by volume). Prepare as described
below:
1. Slowly and carefully add 320 mL concentrated HCl to 500 mL DI H2O in a 1-L
volumetric flask. Gently swirl the solution and let cool.
2. Slowly and very carefully add 40 mL concentrated HNO3 to the flask, gently swirl, and let
cool.
3. Dilute to volume with DI H2O. Mix thoroughly.
6.3.5. Stock solutions of 1,000 µg/mL for the various elements.
6.3.6. Argon - grade or quality specified by the ICP instrument manufacturer.
6.4. Standard Preparation
Prepare multielement working and control standard solutions using 1,000 or 10,000 µg/mL stock solutions. Multielement
standard concentrations recommended for the JY-32 are shown in
Table 1. A suggested control standard is shown in
Table 2. Whenever possible, prepare the control standard from different stock solutions than those used for calibration standards.
The control standard should contain elements and concentrations reflecting what is expected in the majority of the samples, or
problem elements. Any standards containing silver should be prepared in amber colored containers to protect the silver from exposure
to light.
The final acid concentration of the working and control standards is 32% HCl/4% HNO3. These standards
should be stable for at least a year. Also prepare a reagent blank (32% HCl/4% HNO3).
6.5. Sample Preparation
The final acid concentration for the different sample matrices should be 32% HCl/4% HNO3. All of the
elements validated are soluble when using the following acid digestion procedures. Filters, backup pads, wipes, and bulks are
prepared by the following procedures:
6.5.1. Mixed cellulose ester membrane filters, wipe samples (smear tabs), and contaminated
back-up pads.
1. Acid-wash the insides of the 125-mL Phillips beakers by refluxing 1:1 HNO3 using a hot plate
in a ventilated hood. Remove the beakers from the hotplate and carefully pour the used 1:1 HNO3
into an appropriate labeled container. Allow beakers to cool, then rinse several times with DI
H2O and allow to dry.
2. Carefully transfer any loose dust from the cassette into a separately labeled and acid-washed beaker. Using forceps transfer the
sample filter into the same digestion beaker. If the backup pad appears contaminated, include it with the sample filter. If there is
loose dust present, rinse the cassette top (and ring, if present) with a small amount of DI H2O and pour the water into the beaker with
the sample filter. Wipe out the cassette top (and ring, if present) interior surface with a clean Smear Tab that has been moistened with
DI H2O and place it in the same digestion beaker with the rinse and sample filter. Similarly wipe out the cassette bottom interior
surface if the cassette contains loose dust or if the backup pad is contaminated. Do not use Ghost Wipes to wipe out cassettes for this
method. Ensure that blank samples are prepared and analyzed using the same materials and procedures as used for air samples.
Note: |
Do not combine sample filters (i.e. if an industrial hygienist has taken two
consecutive samples from an employee). Always prepare each filter separately. If necessary, combine the results of
consecutive samples. |
3. If the backup pad appears to be discolored, it may be due to leakage of air around the filter during sampling.
4. Slowly add 8 mL HCl to each beaker and allow to sit for 5 minutes while swirling gently. Add 2 mL
HNO3.
Note: |
Always add the HCl first to prevent conversion of Sn to
SnO2 or Sb to Sb2O3.
Always slowly add the HCl and allow the filter to become saturated before adding the HNO3.
Loss of Sb or Sn is possible if the procedure is not strictly followed. See
Reference 8.1 for further details. |
5. Digest the filters in the beakers on a hot plate until the solution volume is approximately:
Solution volume |
Air volume |
|
|
1 mL |
> 400 L |
0.5 mL |
< 400 L or if Ag is requested |
Note: |
Do not allow the solutions to evaporate to dryness. Loss of analyte may occur. |
Remove the beakers from the hotplate and allow to cool in the exhaust hood.
6. For samples with air volumes > 400 L:
Carefully add 8 mL HCl to the beaker, and gently swirl. Quantitatively transfer the sample to a clean
25-mL volumetric flask with DI H2O rinses. Be sure to rinse the sides of the
beaker. Dilute to volume with DI H2O.
7. For samples with air volumes < 400 L or if Ag is a requested analyte:
Add 3 mL of HCl to the beaker and gently swirl. Quantitatively transfer the sample with DI
H2O rinses to a 10-mL volumetric flask and then dilute to volume with DI
H2O.
8. If particulate matter is present after dilution, filter the solution through a 0.45-µm
MCE filter contained in a filtering apparatus. Save the filtrate. Repeat the digestion procedure above for the filter
containing the particulate.
6.5.2. Polyvinyl Chloride (PVC) Filter Samples
Note: |
Polyvinyl chloride filters are not routinely used for solder sample collection and
analysis by ICP. In some cases an industrial hygienist will sample for total or respirable dust using PVC filters and
also submit these samples for ICP analysis. The PVC filter will not be completely digested using this procedure. |
For PVC filters, proceed as in Section 6.5.1., Step 1. Since the PVC filter will not completely digest,
thoroughly rinse the filter residue with DI H2O during quantitative transfer of the sample
solution.
6.5.3. Bulk samples
1. Weigh a representative portion of the bulk sample and transfer to an acid-washed 250-mL
Phillips beaker.
Note: |
Aliquot amounts of bulks are dependent on the analytical sensitivity, detection
limit, and solubility of the material used. If uncertain, a 20 to 50-mg aliquot of a solid material can
be taken as a starting point. Make sure the aliquot taken is representative of the entire bulk sample. If necessary,
use a mortar and pestle to grind any nonhomogenous particulate bulk samples in an exhaust hood. |
2. Slowly add 16 mL HCl and allow to sit for 5 minutes while swirling gently. Then add 4 mL
HNO3.
3. Digest the bulk until approximately 4 mL of sample solution remains. Remove the sample from the hot plate and allow
to cool.
4. Slowly and carefully add 32 mL of HCl.
5. Quantitatively transfer the sample to a 100-mL volumetric flask with DI H2O rinses, and
gently swirl. Dilute to volume with DI H2O.
6. If particulate matter is present after dilution, filter the solution through a 0.45-µm MCE
filter contained in a filtering apparatus. Save the filtrate. Repeat the digestion procedure above for the filter
containing the particulate.
6.6. Instrument Startup and Calibration
Follow the instrument manufacturer's instructions for instrument start-up and calibration. An example of ICP operating
parameters for the JY-32 with a Meinhard-type nebulizer is shown below. These settings may vary from
instrument to instrument:
Gas used |
Argon |
|
|
Gas flow |
Plasma
Nebulizer
Auxiliary plasma |
12 to 16 L/min
*
0.14 to 0.18 L/min |
|
|
Power (rf) |
Incident
Reflected |
1.1 kW
< 5 W |
|
|
Observation height |
Plasma |
15 mm above work coil |
|
|
Wash time |
With automatic sampler
Without automatic sampler |
60 s
10 s |
|
|
Number of exposures |
Standards & samples |
5 |
|
|
Integration time |
Peak signal |
10 s |
|
|
Nebulizer |
Solution uptake rate
Pressure |
0.8 to 1.6 mL/min
~ 30 psig |
|
|
Mass flow controller |
Flow rate range |
varies* |
|
|
* |
Nebulizer is controlled by a mass flow controller. The nebulizer flow will vary
depending on the type in use. |
6.6.1. Profile the instrument before beginning the calibration and analysis. Follow the standard operating
procedure (SOP) (8.3) or manufacturer's instructions for computer initialization and profiling.
6.6.2. A two-point calibration curve is normally obtained by nebulizing the working standards into the plasma and measuring
atomic emission intensities. For calibrations using the JY-32, a first-order linear fit of the data
is computer calculated and slope and intercept coefficients are obtained. Calibrations should be performed by following the
instrument manufacturer's guidelines.
6.7. Analytical Procedure
For more details regarding analytical procedures, refer to the SOP (8.3) or instrument manufacturer's software manual
(s).
Note: |
The samples and standards are made up in 32% HCl/4% HNO3.
These solutions are very caustic. Care should be exercised during analysis to prevent physical contact with these
solutions. |
6.7.1. If necessary, calculate detection limits. The manufacturer's software (if available) should be
used for determining detection limits. These limits normally do not significantly change during short time spans. A general rule
would be to calculate detection limits every 2 to 4 months or when an integral component (nebulizer, torch, mass flow
controller, etc.) of the ICP has been replaced or adjusted. A typical calculation of detection limit (DL) in µg is
shown:
DL = |
(K × SDI × C)
(I - Io) |
× S |
Where: |
K |
= |
Degree of confidence (sigma value)* |
SDI |
= |
Standard deviation of reagent blank intensity (Io) |
C |
= |
Concentration of the calibration standard in µg/mL |
S |
= |
Solution volume in mL |
I |
= |
Total intensity of standard containing concentration C |
Io |
= |
Background intensity (reagent blank) |
|
|
* |
In most cases K=2 or 3 for qualitative and K=10 for quantitative determinations. |
6.7.2. Analysis using an automatic sampler is described:
1. Fill automatic sample vials with the minimum sample volume needed for one analysis and a potential rerun.
2. Load the automatic sampler with labeled standard and sample vials. A multielement working standard should be analyzed
after every 5 to 6 samples. A control standard should be occasionally analyzed to ensure proper instrument operation. If an
element or elements contained in the control standard are not within specification, the analyst should recalibrate before
proceeding with the analysis. A general rule is to use a value less than ±10 or 15% of the known concentration. See
Table 2
for additional details.
3. Aspirate each sample or standard for approximately 1 minute prior to initiating the exposure cycle. This ensures
equilibration in the plasma and minimizes carry-over effects.
4. Dilute and reanalyze any samples containing elements (both screened and validated) exceeding the working range.
Interelement corrections may not be accurate above the working range. Prepare the dilutions by pipetting an appropriate
aliquot from the original solution and dilute with 32% HCl/4% HNO3.
5. Based on the calibration curve initially obtained, convert the sample intensities to concentrations. Then, using the
air volume, solution volume, dilution factor, and sample weight, calculate the concentration for each element analyzed
as mg/m3 (air samples), total micrograms (wipes), or percentage of total weight (bulks) using
the equations shown below. Such calculations are usually performed by software programs within the instrument's dedicated
computer or by an external data reduction method.
6.8. Calculations
6.8.1. The amount of analyte in each sample or blank is calculated as:
µg A |
= |
(µg/mL A) × (mL S) × (DF) |
(1) |
|
|
Where: |
µg A |
= |
Total µg of analyte in the sample or blank |
|
|
µg/mL A |
= |
Measured concentration of analyte in solution (derived from calibration curve) |
|
|
mL S |
= |
total volume of the solution analyzed |
|
|
DF |
= |
amount of dilution applied to an aliquot of the original solution (ratio of final volume
divided by the aliquot volume) |
6.8.2. The blank value, if any, is subtracted from each sample:
µgc A |
= |
µg A - µgb A |
(2) |
|
|
Where: |
µgc A |
= |
µg of analyte, blank corrected |
|
|
µg A |
= |
µg of analyte from equation (1) above |
|
|
µgb A |
= |
µg of analyte in blank |
6.8.3. For air samples, the concentration of analyte in the sample is expressed in
mg/m3 analyte for each element or compound analyzed.
mg A/m3 = |
(µgc A) × (GF)
air volume, L |
|
(3) |
|
|
Where: |
GF |
= |
Gravimetric Factor |
For those elements having a PEL listed as an oxide, the gravimetric factors are:
1.2447 for ZnO, 1.4298 for Fe2O3, 1.7852 for
V2O5
6.8.4. Convert bulk sample analytes to % composition using:
analyte % (w/w) = |
(µgc A) (100%)
(sample weight) (1,000 µg/mg) |
|
(4) |
Where: |
µgc A |
= |
analyte amount (µg) |
|
Sample wt |
= |
aliquot (in mg) of bulk taken in Section 6.5.3. |
|
7. Reporting Results
7.1. Wipe sample concentrations are calculated and reported as total micrograms (µgc A) or
milligrams for each element.
7.2. Bulk sample results are calculated and reported as elemental percent by weight. Although the acid matrix is
constant for samples and standards, the differences in the overall sample matrix between a bulk and standard can
be large. Therefore, bulk results are reported as approximate values for each element determined.
7.3. Air sample results are reported in units of mg/m3. Results for analytes having an
oxide PEL are reported as mg/m3 of the oxide.
7.4. Determinations of the screened elements or compounds are not routinely reported. Spectral interference
corrections for these analytes are not included, and validations have not been performed. If a sample has a
screened analyte over the PEL, the analyst should contact her/his supervisor. Additional sampling or, if possible,
additional analysis of the original sample should be performed to quantitate the potential overexposure.
8. References
8.1. Occupational Safety and Health Administration Technical Center:
ICP Backup Date Report
for Soldering and Brazing Matrices (Arl 3560) by D.C. Cook. Salt Lake City, UT. 1991.
8.2. National Institute for Occupational Safety and Health: NIOSH Manual of Analytical Methods,
2nd ed., Vol. 7 (DHEW/NIOSH Pub. No. 82-100). Cincinnati, OH, 1981. Method No. 351.
8.3. Occupational Safety and Health Administration Technical Center: ICP Standard Operating
Procedure by J.C. Septon. Salt Lake City, UT. 1988 (unpublished).
Table 1
Calibration Standards and Wavelengths |
|
|
|
|
|
|
|
|
|
|
|
|
NO |
|
ELE |
STD SOLN |
CONCN |
WAVELENGTH (nm) |
|
|
1 |
Ag |
1 |
1.00 |
328.0 |
2 |
Be |
1 |
1.00 |
313.0 |
3 |
Cd |
1 |
10.00 |
226.5 |
4 |
Pb |
1 |
10.00 |
220.3 |
|
|
5 |
Cu |
2 |
1.00 |
324.7 |
6 |
Sb |
2 |
10.00 |
217.5 |
7 |
Sn |
2 |
10.00 |
189.9 |
8 |
Zn |
2 |
10.00 |
213.8 |
|
|
9 |
Al |
3 |
10.00 |
396.1 |
10 |
As |
3 |
10.00 |
189.0 |
11 |
Ca |
3 |
10.00 |
393.3 |
12 |
Co |
3 |
10.00 |
237.9 |
13 |
Cr |
3 |
10.00 |
267.7 |
14 |
Fe |
3 |
100.00 |
271.4 |
15 |
Mg |
3 |
10.00 |
279.5 |
16 |
Mn |
3 |
10.00 |
257.6 |
17 |
Mo |
3 |
10.00 |
202.0 |
18 |
Ni |
3 |
10.00 |
231.6 |
19 |
Se |
3 |
10.00 |
196.0 |
20 |
Si |
3 |
10.00 |
212.4 |
21 |
V |
3 |
10.00 |
310.2 |
Where:
ELE
STD SOLN
CONCN |
Element
Number of Calibration standard
Concentration (µg/mL) of calibration standard |
Notes: STD SOLN 1 - Prepare in an amber-colored glass bottle to protect the Ag from photo-
decomposition.
STD SOLN 3 - This standard is used only for calibration of the screened elements. The digestion and analysis are not
optimized for these elements.
For the ISA JY-32 ICP, calibration is accomplished using a two-point calibration curve with the concentration
for each element listed above. A reagent blank was used as the low standard. Each element calibrated is contained in one of
three separate calibration standards (STD SOLN). For example, STD SOLN 1 contains Ag, Be, Cd, and Pb.
The three mixed calibration standards were selected because of chemical compatibility and potential interferences. Other
combinations of elements or concentrations can be used; however, compatibility and possible interferences have to be
considered when combining elements other than the mixtures listed above.
Table 2
Control Standard |
|
|
|
|
|
|
|
|
|
|
NO. |
ELE |
UPPER CONCN |
LOWER CONCN |
STD CONCN |
|
|
1 |
Pb |
5.75 |
4.25 |
5.0 |
2 |
Cd |
5.75 |
4.25 |
5.0 |
3 |
Ag |
0.575 |
0.425 |
0.5 |
4 |
Be |
0.575 |
0.425 |
0.5 |
Where:
ELE
UPPER CONCN
LOWER CONCN
STD CONCN |
Element
The upper concentration limit in µg/mL
The lower concentration limit in µg/mL
The calibration concentration in µg/mL |
A control standard should be prepared and analyzed as an analytical monitor of ICP performance.
Some instrument manufacturers have instituted analytical software routines which will evaluate control standard results
during the analysis. The control standard example listed above is used for the JY-32 ICP. An alternate control
mixture can be prepared.
To illustrate the control standard concept, 5 µg/mL Pb is used. For this control standard, Pb has a limit of ±
15% (upper and lower concentration limits of 5.75 µg/mL and 4.25 µg/mL, respectively). If a
calculated value greater than ±15% is obtained while analyzing this standard, the analysis will automatically halt. The
ICP operator should then re-calibrate the instrument. |
|
|
|