Chapter 3. How To Mistake-Proof the Design
The Design Change Imperative
Donald Berwick of the Institute for Healthcare
Improvement (IHI) argues that improving patient safety
requires changes in the design of health care systems:
...We are human and humans err. Despite outrage,
despite grief, despite experience, despite our best
efforts, despite our deepest wishes, we are born fallible
and will remain so. Being careful helps, but it brings
us nowhere near perfection... The remedy is in
changing systems of work. The remedy is in design.
The goal should be extreme safety. I believe we should
be as safe in our hospitals as we are in our homes. But
we cannot reach that goal through exhortation,
censure, outrage, and shame. We can reach it only by
commitment to change, so that normal, human errors
can be made irrelevant to outcome, continually found,
and skillfully mitigated.1
Berwick is not the only proponent of employing design as
the chief approach to improving patient safety. The British
Department of Health and The Design Council issued a
joint report in the early stages of their patient safety
program, calling for "A system-wide design-led approach
to tackling patient safety in the National Health Service."2,a
a. The British Department of Health and The Design Council's
book, Designing for Patient Safety, is available at
http://www.designcouncil.org.uk/resources/assets/assets/pdf/Publications/Design_for_Patient_Safety.pdf.
FMEA Implicitly Requires Changes
in Design
In Chapter 2, we reviewed the basic steps involved in
performing failure modes and effects analysis (FMEA).
With the exception of JCAHO,3 the various versions of
FMEA do not explicitly state that design changes are
required. JCAHO's FMEA Step 6 is "redesign the
process." In versions of FMEA that do not explicitly
require them, design changes generally, and mistake-proofing
particularly, are implicit requirements of FMEA.
Figure 3.1 illustrates the FMEA form used by the auto
industry.4 Failure modes and effects are prioritized
according to three 1- to 10-point scales: severity,
likelihood, and detectability (a term used in the
automotive version of FMEA that serves the purpose of
"ensuring that the hazard is obvious"). The results are
multiplied to create an overall assessment called the Risk
Priority Number, or RPN. In Figure 3.1, two columns are
labeled RPN. The first is an initial priority. The second is
a recalculation after taking action. The idea is that any
worthwhile action should improve the second RPN.
In the health care version of FMEA (HFMEA™),
proposed by the Department of Veterans Affairs (VA)
National Center for Patient Safety,5 the assessments of
severity, likelihood, and detectability are accomplished by
employing a decision flowchart instead of merely rating
them on a 1- to 10-point scale.
The decision flowchart is shown in Figure 3.2.
The flowchart determines if action is required (proceed) or
if existing systems are adequate and action is not required (stop). HFMEA™ does not explicitly state the followup
step that is explicit in the automotive version of FMEA.
After the recommended actions have been taken, the
improvements should eliminate the need for action the
next time that an HFMEA™ analysis is revisited.
Otherwise, further action is called for. The decision after
taking action should be "stop (no action required)."
This implies that the actions must accomplish at least
one of the following three objectives in order to be
considered effective and to avoid the implied iterations
of the FMEA process:
- Remove a single-point weakness.
- Create one or more effective control measures.
- Make the hazard so obvious that control measures are not needed.
Removing Single-Point
Weaknesses
The term single-point weakness refers to the creation of
redundancies in the system. Fault trees explicitly
recommend redundancy as an approach to improving
system reliability. Redundancy is created by increasing
the number of systems or individuals involved in the
process. There are many examples from everyday life of
how processes can be made more understandable. One
example is shown in Figure 3.3. Having more back-up
systems or more checks (double, triple, or more)
increases the probability that the process will proceed
correctly. Too often in health care, this redundancy is
created when a second trained health care professional
double-checks the first to ensure that the process has
been performed correctly. However, at the time of this
writing, many health care organizations are experiencing
a nursing shortage; utilizing several nurses to repeat a task for the sake of redundancy is an approach that is
too costly, if not impossible, to implement.
An alternative that would not require more staff would be
to involve patients themselves, where possible, or a
concerned family member or friend. In order for this
alternative to work, medical processes must be rendered
transparent to untrained individuals; errors must be made
obvious. Creating transparency in health care's jargon-rich,
complex processes can be very challenging. Implementing
visual systems (5Ss, go to Chapter 1, Figure 1.7) and
providing the clear process cues suggested by Norman6
could increase transparency significantly. Another option
would be to employ mistake-proofing devices or design
features in the error-detection process.
Effective Control Measures
In their article explaining HFMEA™, DeRosier, et al,5
cite the pin indexing system in medical gases as an
example of an effective control measure. They state:
If your hospital does not use universal adaptors (for
regulators), and all the connectors in the building have
the correct pin index, the pin indexing would be an
effective control measure; it would prevent the
incorrect gas from being connected to the regulator.
An example of pin indexing is shown in Figure 3.4. The
implication is that an effective control measure stops the
process when an error occurs. This is mistake-proofing.
Effective control measures are design changes that prevent
or stop processes from continuing when an error has
occurred by introducing a process failure (Figure 3.5).
Make Hazards Obvious
If hazards are made obvious, HFMEA™ does not require
further action. Color-coding is a common approach to
making mistakes more obvious. The face of the gauge in
Figure 3.6A uses color changes to indicate the range of
correct settings.
Other everyday examples of making aspects of a system
more obvious are shown in Figure 3.6B-F. Figure 3.7A
provides an example of a design to make errors more
obvious and Figure 3.7B a secondary safeguard—the
warning label.
Design changes are required actions in response to FMEA.
The answer to the question of what the design changes
should look like is an odd one. To create patient safety
through mistake-proofing, the thing to design into
processes is failure. The design changes should be carefully
designed process failures.
Experts Agree on Designing Failures
Findings from engineering, cognitive psychology, quality
management, and medicine all agree that to avoid human
error or its impact, it is necessary to create a process that
ensures failure.
Henry Petroski,7 a noted engineering author, states:
We rely on failure of all kinds being designed into
many of the products we use every day, and we have
come to depend upon things failing at the right time
to protect our health and safety... Failure is a relative
concept, and we encounter it daily in more frequent
and broad ranging ways than is generally realized. And
that is a good thing, for certain types of desirable
failures, those designed to happen, are ones that
engineers want to succeed at effecting. We often thus
encourage one mode of failure to obviate a less
desirable mode.
This approach is supported by recommendations from
psychology as well. Norman6 recommends the installation
of "forcing functions." Forcing functions create "situations
in which the actions are constrained so that failure at one
stage prevents the next step from happening." Forcing
functions are attractive because they rely "upon properties
of the physical world for their operation; no special
training is necessary."6
From a quality management perspective, Shigeo Shingo
recommends that "when abnormalities occur, shut down
the machines or lock clamps to halt operations, thereby
preventing the occurrence of serial defects." Source
inspections "are based on the idea of discovering errors in
conditions that give rise to defects and performing
feedback and action at the error stage so as to keep those
errors from turning into defects."8
This approach is not unheard of in medicine. Croteau and
Schyve9 discuss this approach:
A process that is designed to detect failure and to
interrupt the process flow is preferable to a process
that continues on in spite of the failure...We should
favor a process that can, by design, respond
automatically to a failure by reverting to a
predetermined (usually safe) default mode.
Petroski7, Norman6, Shingo8, and Croteau and Schyve9
each approach the problem from a different perspective
and discipline, yet their prescriptions are identical. To
reduce human error, make design changes that prevent or
stop processes from continuing when an error has
occurred. When a process stops, it is a process failure.
Under other circumstances, having the process stop would
be undesirable. However, a process stoppage can be a far
more benign failure than allowing a medical error to
progress. Stopping the process will not always be the
appropriate action. Some errors can be more benign, and
stopping the process may not be necessary.
The term "benign" is used here in a relative sense to mean
favorable or propitious.10 It is possible that what is
perceived to be a benign failure in some circumstances
might actually be perceived to be very undesirable in
others. An example will illustrate this point.
In the decade prior to the U.S. Civil War (1853), Elisha
Otis demonstrated his safety elevator at the New York
World's Fair.11 The novel feature of the safety elevator was
the Otis Elevator Brake (Chapter 1, Figure 1.4). This
device would prevent the elevator from falling when the
cable broke. What was the failure? The brake was designed
to get the elevator stuck, usually between floors. Being
stuck in an elevator between floors is a very undesirable
failure. However, when compared with falling to one's
death at the bottom of an elevator shaft, the option of
being stuck between floors becomes a much more
palatable alternative.
Return to Contents
Multiple Fault Trees
Multiple fault trees can be seen in the flowchart of
mistake-proofing tools (Chapter 2, Figure 2.7). Multiple
fault trees are used to help design benign failures. The
traditional use of fault trees is to carefully define the
current situation, determine causes of undesirable failure,
and identify the resources required to generate that
undesirable failure. The new, second use of multiple fault
trees is to determine ways to cause or create benign failures
and use them as preventive measures. The use of multiple
fault trees provides insights into the causes of desired
failures and identifies the "resources" required to generate
them. The design of the process must be changed so that
the failure associated with the undesirable event causes the
more benign event (desired failure) to occur instead. The
objective is to move failures from the harmful event fault
tree to the benign failure fault tree.
Figures 3.8 and 3.9 show both fault trees before and after
the failures are moved.
In the "before" picture (Figure 3.8), the harmful event has
three minimal cut sets. The first set, containing failures 1
and 2 in a redundant system (AND), has an overall
probability of 0.01. The other two sets, containing single
failures 3 and 4, respectively (OR), have no redundancy
and have a probability of occurrence of 0.05 each.
Suppose that Cause 4 is selected as the first target for
improvement. Assume that a benign failure that would
adequately safeguard patient safety, because it would
prevent the harmful event from occurring, has been
identified. The fault tree for the benign failure is shown in
its initial state on the right side of Figure 3.8. The fault
tree shows substantial redundancy. Three failures must
occur simultaneously in order for a failure to occur, a
situation that has a 0.001 chance of happening.
Figure 3.9 shows the fault trees after the mistake-proofing
has been accomplished. Cause 4 now appears in the fault
tree of the benign failure. The probability of the harmful
event occurring has been reduced by approximately 45
percent. However, as the diagram shows, more can be
done. Mistake-proofing Cause 3 by moving it to another
fault tree should also be considered. If it can be moved
successfully, the probability of the harmful event would be
reduced to 0.01. Perhaps 0.01 is still unacceptably high. It
is, however, a substantial improvement. To further reduce
the probability of the harmful event would require that
either Cause 1 or Cause 2 be moved to a more benign
fault tree.
The fault trees in Figures 3.8 and 3.9 illustrate the logic of
the changes sought through mistake-proofing. Go to Chapter 4 for a more technically precise version of these fault trees
that model mistake-proofing devices but which may be
less than perfectly reliable.
Fault trees enable process designers to anticipate how the
process will behave after mistake-proofing has been
implemented. After mistake-proofing, the benign failure is
far more likely to occur than before. When the benign
failure occurs, the staff member using the process must
troubleshoot it to determine the reason for the failure. The
benign failure in Figure 3.9 is nearly ideal for use with one
of the causes. If the benign failure shown occurs, the most
likely cause is Cause 4. That cause can be confirmed
quickly, and the process can be reset and restarted as
needed.
It is important to ensure that a single benign event does
not become the failure mode for too many mistake-proofing
devices. For example, Causes 1 through 4 should
not all be moved to the same benign failure. When
multiple causes can stop the process in the same way, the
process can become difficult to troubleshoot. The result is
that team members may become uncertain about how to
re-start the process after a failure.
Effective mistake-proofing should involve a diverse set of fault
trees that result in a variety of benign failures.
|
While designing a mistake-proofed cooktop (Figure 3.10),
the designer could consider that the presence of a cooking
pan could be detected by the mass or the weight of the
pan. The burner could be deactivated if there is no pan
sitting on it. Additional features could include a small light
near each burner to indicate that the burner is on, and
arranging the burner control knobs to correspond to the
physical arrangement of the burners. This is a natural
mapping.
Although fault trees are central to this discussion, other
failure analysis tools can be employed and will yield
similar insights (relatively detailed FMEA,4,5 anticipatory
failure determination,12 current reality trees,13 and causal
trees,14 for example).
Fault trees have two advantages:
- The application of the method to a situation is straightforward.
- The failure is displayed simply and in substantial detail.
In addition to understanding fault trees, team members
need information, resources, and creativity:
- Team members must be privy to all information normally generated by the enabling tools.
- Team members must have access to detailed knowledge of the medical processes involved.
Only then will the team be able to link the causes of
undesirable failures to the outcomes of benign failures in
ways that:
- Are inexpensive.
- Have minimal impact on the existing process.
Return to Contents
Designing Mistake-Proofing Devices that Cause Benign Failures
There are eight primary steps involved in designing mistake-proofing devices.
Step 1. Select an undesirable failure mode for further analysis. In order to make an informed decision about which
failure mode to analyze, the RPN or the criticality number of the failure mode must have been determined in the course
of performing FMEA or FMECA.
Step 2. Review FMEA findings and brainstorm solutions (Figure 3.11). Most existing mistake-proofing has been done
without the aid of a formal process. This is also where designers should search for existing solutions in medicine or
elsewhere. The examples in Chapter 6 include comparisons of solutions from medical, industrial, and everyday life. Many
exploit the same ideas (Go to Chapter 6, Examples 6.7 and 6.13). Common sense, creativity, and adapting existing examples
are often enough to solve the problem. If not, continue to Step 3.
Step 3. Create a detailed fault tree of the undesirable failure mode (Figure 3.12). This step involves the traditional use
of fault tree analysis. Detailed knowledge regarding the process and its cause-and-effect relationships discovered during
root cause analysis and FMEA provide a thorough understanding of how and why the failure mode occurs. The result of
this step is a list and contents of minimal cut sets. Since severity and detectability of the failure mode could be the same
for all of the minimal cut sets, the probability of occurrence will most likely be the deciding factor in a determination of
which causes to focus on initially.
Step 4. Select a benign failure mode(s) that would be preferred to the undesirable failure. The tools that precede
multiple fault trees in Figure 2.7 once again provide information about other failure modes and their severity. Ideally, the
benign failure alone should be sufficient to stop the process; the failure, which would normally lead to the undesirable
event, causes the benign failure instead.
Step 5. Using a detailed fault tree, identify "resources" available to create the benign failure (Figure 3.13). These
resources, basic events at the bottom of the benign fault tree, can be employed deliberately to cause the benign failure to
occur.
Step 6. Generate alternative mistake-proofing device designs that will create the benign failure (Figure 3.14). This
step requires individual creativity and problem-solving skills. Creativity is not always valued by organizations and may be
scarce. If necessary, employ creativity training, methodologies, and facilitation tools like TRIZ (described in Chapter 2) if
brainstorming alone does not result in solutions.
Step 7. Consider alternative approaches to designed failures (Figure 3.15). Some processes have very few resources. If
creativity tools do not provide adequate options for causing benign process failures, consider using cues to increase the
likelihood of correct process execution. Changing focus is another option to consider when benign failures are not
available.
If you cannot solve the problem, change it into one that is
solvable. Changing focus means, essentially, exploring the
changes to the larger system or smaller subsystem that
change the nature of the problem so that it is more easily
solved. For example, change to a computerized physician
order entry (CPOE) system instead of trying to mistakeproof
handwritten prescriptions. There are very few
resources available to stop the processes associated with
handwritten paper documents. Software, on the other
hand, can thoroughly check inputs and easily stop the
process.
Often, responses to events appropriately include multiple
actions, thereby creating their own redundancy.
|
Of course, there are no guarantees that a failure mode will
be solved. If little or no progress is made after completing
Steps 1 through 7, it may be prudent to just give up.
Some inventions needed to solve patient safety concerns
are not, at present, technically or financially feasible.
Although it may be necessary to give up, the worst-case
scenario is the continuation of the current process with no
change at all.
Step 8. Implement a solution. A full discussion of
solution implementation is beyond the scope of this book.
A rigorous look at implementation issues is available in
the change management literature. Some basic tasks
usually required as part of the implementation are listed
below:
- Select a design from among the solution alternatives:
- Forecast or model the device's effectiveness.
- Estimate implementation costs.
- Assess the training needs and possible cultural resistance.
- Calculate the solution priority number (SPN) as described in Chapter 2.
- Assess any negative impact on the process.
- Explore and identify secondary problems (side effects or new concerns raised by the device).
- Assess device reliability.
- Create and test the prototype design:
- Find sources who can fabricate, assemble, and install custom devices, or find manufacturers willing to make design changes (more in Chapter 9).
- Resolve technical issues of implementation.
- Undertake clinical trials (because of the stakes involved, medical implementations will need to be much more deliberate than those in other industries).
- Trial implementation:
- Resolve nontechnical and organizational issues of implementation.
- Draft a maintenance plan.
- Draft process documentation.
- Broad implementation leads to:
- Consensus building.
- Organizational change.
The eight steps to creating mistake-proofing devices can
be initiated by a root cause analysis or FMEA team, an
organization executive, a quality manager, or a risk
manager. An interdisciplinary team of 6 to 10 individuals
should execute the process steps. An existing FMEA or
root cause analysis team is ideal because its members
would already be familiar with the failure mode. Help and
support from others with creative, inventive, or technical
abilities may be required during the later stages of the
process. A mistake-proofing device is designed using the
eight steps just discussed in the application example that
follows.
Return to Contents
An Application Example
Step 1. Determine the undesirable failure mode. For this
example, consider the undesirable event of a patient
injured by a fall during a transfer to or from a standard
(nonpowered) wheelchair. Berg, Hines, and Allen15 report
that 37.9 percent of wheelchair users fell at least once in
the past 12 months. Of those who fell, 46.7 percent were
injured as a result of their fall. Tideiksaar,16 Calder and
Kirby,17 and Ummat and Kirby,18 confirm that transfers to
and from wheelchairs are common causes of injuries.
Step 2. FMEA is well-known in health care, and detailed
instruction on its implementation is available. It may be
that a well-done FMEA would be enough to generate
ideas for how to change the process design so that falls are
prevented during transfers to or from wheelchairs. In order
to demonstrate subsequent steps, though, assume that
FMEA and brainstorming did not generate an adequate
number of possible solutions, so the process continues to
Step 3.
Step 3. This step calls for the creation of a detailed fault
tree used to understand and make sense of how failures
occur. In this case, the information from the literature on
wheelchair injuries (cited in Step 1) and information from
previously created FMEAs would inform the creation of
the fault tree. The fault tree in Figure 3.16 shows the
undesirable event—patient falls during transfer to or from
a standard wheelchair. Each level of the tree provides
additional detail into why the event occurred. The fault
trees in this example have purposely been kept small and
simplified. They are intended only to illustrate how using
fault trees can assist in making sense of the causes of
undesirable failures and how more benign failures can be
designed into the process instead.
Given the fault tree for the undesirable event in Figure 3.16, there are several possible alternative approaches to
preventing the failure. One or more of the causes (often
called basic failures) shown in the bold-lined boxes in
Figure 3.16 need to be addressed. To avoid "hand brake
not engaged," for example, it is necessary to find ways to
ensure that the patient does not forget (Box A) and to
provide training (Box B). To avoid "footplate present when
it should not be," requires actions to prevent patients from
failing to move the footplate and to prevent the footplate
from moving back into position for use (caused by Boxes
2 and 3). To prevent "Patient falls during transfer..." (the
top event), preventive actions must be taken on both the
left and right branches of the fault tree, "failure to land on
seat..." and "trip on footplate," respectively. The next
several paragraphs show how benign failures might be used
to think through design changes that will prevent either
branch from resulting in patient falls.
Step 4. Selecting a benign failure mode in Step 4 requires
asking, "What failure would be preferable to having a
patient fall"? Separating this question into sections, we
arrive at:
- What failure would be more benign than failing to land on the wheelchair's seat?
- What failure would be more benign than tripping on the footplate?
The answer to the first of these questions might be to
prevent the wheelchair from rolling (Figure 3.17).
Although assuring that the wheelchair does not roll is a
failure that completely defeats the purpose of having a
wheelchair, this outcome could be better than that of
wheelchair users injuring themselves, especially if the
failure is temporary. The white box in Figure 3.17 shows
one of the causes of the wheelchair rolling away, "armrest
used for support, seat vacant," being moved into the fault
tree for "wheelchair will not roll." This move generates
creative or inventive questions. Can a mechanism be
invented in which the brake is always engaged when the
seat is vacant? Alternatively, is it possible to develop a
brake that is activated when most of a patient's weight is
on the armrests instead of on the seat? These creative or
inventive questions are the starting places for changing the
design of the process. It might be necessary to explore
several, perhaps many, possible solutions to find the best
one.
Proposing a benign failure converts a problem into a
question of creativity: Can we invent a mechanism so
that the brake is always engaged when the seat is vacant?
|
Figure 3.17 also provides the information required by Step
5. This tree identifies the resources necessary for creating
the benign failure. In many cases, the basic failure (cause)
of the undesirable fault tree can be moved directly onto an
existing branch of the benign fault tree, thereby employing
an already-existing basic failure as a starting point for
creating the benign failure. In this case, the existing basic
failure "hand brake engaged" is suggestive of a device that
solves the problem: an automatically engaged brake.
It turns out that Steps 6 and 7 are unneeded since an
automatic locking device that creates the benign failure
suggested in Figure 3.17 is already commercially available.
It is a braking system that uses a spring to engage the
brake whenever the wheelchair is vacant (Figure 3.18).
The brake is disengaged by the weight of the wheelchair
occupant, which depresses a lever beneath the seat. The
device moves the basic failure, "armrest used for support,
seat vacant," from the undesirable fault tree to the benign
fault tree. Now, the brakes are automatically engaged when
the armrest is used for support (Figure 3.19).
This device comes with a significant secondary problem: it
is much more difficult to move empty wheelchairs around
because their brakes are always engaged. However, this
problem is not difficult to resolve. A hand-activated brake
release enables attendants to override the automatic brake
system.
Is this a "good" solution (part of Step 8)? The locking
device is a very effective solution. The probability of falls is
reduced dramatically. The device provides a control
measure capable of preventing the error of not engaging
the hand brake. The cost of the product is moderate.
While not affordable out of daily operating funds, funds
allocated from a unit-level budget would probably be
adequate. The implementation would most likely be
considered easy, depending on the culture of the organization. The only training required would be to
point out the brake release to the staff. If more than
minimal resistance is expected, the organization's problems
go well beyond those that mistake-proofing is likely to
help. Consequently, the SPN equals 18, the second
highest possible score.
Calculating the SPN indicates that this device is a
promising direction for improvement. It does not provide
a definitive answer to the question of whether this device
should or should not be implemented. The remaining
tasks in Step 8 must be performed, including an
assessment of device reliability, device trials, maintenance
planning, drafting process documentation, etc.
Will the automatic brake eliminate the possibility of the
top event, "patient falls during transfers?" No. The other
branch of the tree, "trip on footplate," must also be
addressed. Returning to Step 4, the second question is:
What failure would be preferable to having a patient trip
on a footplate? One possible response would be the failure
of having the footplate absent or completely unavailable.
Step 5. What would cause the footplate of a wheelchair
(Figure 3.20) to be absent or unavailable? Figure 3.21
indicates a few possibilities. It could have broken off due
to an impact or other excessive force. This failure suggests
solutions involving parts that break away, like the
attachment of ski bindings to a boot. Under the correct
amount and direction of forces, the ski breaks away from
the boot and prevents injuries to the skier's legs. Perhaps
the footplate should be designed so that it snaps off when
the patient's entire weight is put on it; or, if the footplate
is bumped from the rear, it could easily detach from the
chair frame. The footplate might also be absent or
unavailable because it has been disassembled and removed
intentionally.
This failure suggests that the footplate should be present
to hold the patient's feet while the chair is in use, but the
footplate would, ideally, be absent at the time of entry or
exit from the wheelchair. A situation in which the
footplate is ideally both present and absent is an example
of what TRIZ users call "an inherent contradiction," or a
"physical contradiction."
Step 6. Step 6 recommends using creative or inventive
tools to find directions for further exploration. TRIZ
offers ready approaches for resolving contradictions.
A contradiction is defined as:
Opposition between things or properties of things. There
are two kinds of contradictions:
- Tradeoffs—a situation in which if something good happens, something bad also happens; or, if something good gets better, something undesirable gets worse.
- Inherent contradictions—a situation in which one thing has two opposite properties.20
Once a contradiction is identified in the language of
TRIZ, approaches used by other inventors to solve that
contradiction can be looked up on a large matrix.
TRIZ approaches for resolving the contradiction of
incompatible requirements at different times include the
following eight approaches. The number following each
approach is the TRIZ "principle number." These numbers
are consistent throughout the TRIZ literature, although
their descriptive label will vary among authors (as shown
in parentheses below). A brief description of each follows.
More detailed information is available from other
sources.21
- Segmentation (fragmentation) (1): "Divide an object
into independent parts."21 "Make an object easy to
disassemble."20 Example: flexible poles of dome-shaped
tents can be folded compactly when not in use (Figure 3.22).
- Preliminary counteraction (9): "Preload countertension
to an object to compensate excessive and
undesirable stress."21
- Preliminary action (10): "Perform required changes to
an object completely or partially in advance."21 For
example (Figure 3.23): Glue a strong cord inside a
shipping box and connect a "pull here" tab before use to
make it easier to open the box later.
- Beforehand compensation (cushion in advance) (11):
"Compensate for relatively low reliability of an object with
emergency measures prepared in advance." For example:
Automotive airbags or guard rails, especially the ends that
are designed to attenuate impact (Figure 3.24).
- Dynamic parts (dynamicity) (15): Allow (or design) the
characteristics of an object ...to change to be optimal..."20
"Divide an object into elements capable of changing their
position relative to each other."21 For example: Flaps on
airplane wings (Figure 3.25).
- Periodic action (19): "Instead of continuous actions,
use periodic or pulsating actions."20 For example: Sprinkler
does not damage soil by applying water in droplets instead
of a steady stream (Figure 3.26).
- Hurrying (rushing through) (21): "Perform harmful and
hazardous operations at very high speed" (Figure 3.27). For
example: For a given surgical procedure, the more rapidly it
can be done, the better.21
- Discarding and recovery (Rejecting and regenerating
parts) (34): "Make portions of an object that have fulfilled
their functions go away (discard by dissolving,
evaporating, etc.), or modify them directly during the
operation. Conversely, restore consumable parts of an
object directly in operation."20 For example (Figure 3.28):
Dissolvable polylactides screws and pins are used in
surgery to mend broken bones, which makes the second
operation for their removal unnecessary.20
Clearly, not all of these approaches seem promising for the
solution to this particular problem of having the footplate
present some of the time but not at others. However,
preliminary counteraction, preliminary action, and the
related approaches of segmentation, dynamic parts, and
discarding and recovery seem promising.
The eight approaches suggested using multiple fault trees
fit nicely into these TRIZ recommendations. Having
breakaway parts is an example of discarding and recovery.
Approaches suggested by the TRIZ principle of
"preliminary counteraction" involve putting a barrier in
place that will keep the foot from getting behind the
footplate, and creating a spring-loaded footplate assembly
that is raised whenever the foot is not resting on it. The
barrier could take the form of a heel strap or leg strap
(Figures 3.29 and 3.30).
Another approach that TRIZ
suggests is dynamic parts. Figure 3.31 shows that the use
of dynamic parts is already available. Some wheelchairs
have footplates that can be released and swung to the side.
What this solution lacks is a system for causing it to
happen automatically, a simple means of detecting
whether the footplates should swing out of the way.
Because patients could conceivably reposition themselves
by putting most of their weight on the armrests,
footplates, and chair back, having a seat lever mechanism
similar to the previous example could be problematic.
Perhaps in this case the setting function (or detection
system) would need to test for total weight exerted on the
wheels instead of on the seat itself. What should the final
solution be? That remains an open question; but the
engineering task does not appear so daunting that it would
be difficult to develop if someone were so inclined.
Return to Contents
Conclusion
Changing the design of processes is a critical task in
reducing the human errors that plague health care. Of
course, the term design can mean different things in
different contexts. In the contexts of mistake-proofing and
FMEA, it means physical changes in the design of
processes. Only these design changes are adequate to
escape the repetitive revisiting, followup, and preventive
actions that FMEA otherwise requires.
The goal is to rapidly find inexpensive, effective solutions
that are easy to implement. The process is only an aid in
accomplishing this objective.
Adequate design changes will remove a single point of
weakness, create one or more effective control measures, or
make the hazard so obvious that control measures are not
needed. All of these required actions suggest forms of
mistake-proofing. Moreover, experts from the disciplines
of engineering, cognitive psychology, quality management,
and medicine all agree that these process design changes
should not bring processes to a stop. They should
introduce benign failures into a process.
A series of tools and some novel applications of those tools
were presented in an eight-step process designed to
generate and evaluate several potential solutions. An
application example was presented. While it was presented in a relatively linear step-by-step fashion, real life
circumstances will occasionally prove to be less linear. While
all the steps need to be considered eventually, the designer of
mistake-proofing devices can opportunistically skip a step.
At other times, the designer may repeatedly return to some
steps on an iterative basis. The goal is not to complete the
eight steps. The goal is to rapidly find inexpensive, effective
solutions that are easy to implement. The process is only an
aid in accomplishing this objective.
The approach presented here represents an incremental step
in thinking about mistake-proofing medical processes.
Future research and experience will provide additional tools,
techniques, and enhanced approaches for designing effective
medical processes that will have the ability to prevent
specific, undesirable failure modes. Some of the limitations,
drawbacks, and design issues of mistake-proofing are
presented in the next chapter.
Return to Contents
References
1. Berwick DM. Not again! BMJ 2001;22:247-8.
2. Department of Health, The Design Council. Design for
patient safety. London: Department of Health and The
Design Council; 2003.
3. Joint Commission on Accreditation of Healthcare
Organizations. Patient safety: essentials for health care, 2nd
ed. Oakbrook Terrace, IL: Joint Commission Resources;
2004.
4. Automotive Industry Action Group. Process failure mode
effect analysis. Southfield, MI: Automotive Industry Action
Group; 1995.
5. DeRosier J, et al. Using health care failure mode and effect
analysis: The VA National Center for Patient Safety's
prospective risk analysis system. Jt Comm J Qual Improv
2002; 28:248-67.
6. Norman DA. The design of everyday things. New York:
Doubleday; 1989.
7. Petroski H. Designed to fail. American Scientist 1997;
85:412-46.
8. Shingo S. Zero quality control: Source inspection and the
poka-yoke system. New York: Productivity Press; 1986.
9. Croteau RJ, Schyve PM. Proactively error-proofing health
care processes, in Spath PL, ed. Error reduction in health
care. Chicago: AHA Press; 2000.
10. Webster's new universal unabridged dictionary. Definition
3. New York: Barnes and Noble Books; 1996.
11. Goodwin O. Giving rise to the modern city. Chicago: Ivan
R. Dee; 2001.
12. Kaplan S. New tools for failure and risk
analysis/anticipatory failure determination™ (AFD™) and
the theory of scenario structuring. Southfield, MI: Ideation
International Inc; 1999.
13. Goldratt EM. It's not luck. Great Barrington, MA: North
River Press; 1994.
14. Gano DL. Apollo root cause analysis: a new way of
thinking. Yakima, WA: Apollonian Publications; 1999.
15. Berg K, Hines M, Allen S. Wheelchair users at home: few
home modifications and many injurious falls. Am J Public
Health 2002; 92(1):48.
16 Tideiksaar R. Risk management: Falls and wheelchair safety.
RN+ Fall Prevention and Restraint Reduction Newsletter.
http://www.rnplus.com/pdf.newsletter.1/april.04.pdf.
Accessed July 2005.
17. Calder C, Kirby R. Fatal wheelchair-related accidents in the
United States. Am J Phys Med Rehab 1990;69:184-90.
18. Ummat S, Kirby R. Nonfatal wheelchair-related accidents
reported to the national electronic injury surveillance
system. Am J Physical Medicine & Rehabilitation 1994;
73:163-7.
20. Rantanen K, Domb E. Simplified TRIZ: New problem-solving
applications for engineers and manufacturing
professionals. Boca Raton, FL: CRC Press; 2002.
21. Altshuller GS, Shulyak L. 40 Principles: TRIZ keys to
technical innovation. Trans. Rodmans, Shulyak L.
Worcester, MA: Technical Innovation Center; 2001.