Date Recall Initiated |
January 16, 2008 |
---|---|
Product: |
Medtronic Neuromodulation Implantable Infusion Pumps:
|
Use: |
These devices administer drugs to a specific site in the body to treat pain, spasticity (continuous muscle contraction), and cancer. The pump is implanted in the patient, either with or without a side catheter access port, catheters, and catheter accessories. |
Recalling Firm: |
Medtronic Neuromodulation 800 53rd Avenue NE Minneapolis, Minnesota 55440-1250 |
Reason for Recall: |
The company updated the labeling for the devices to include current patient management and treatment recommendations. The company also received reports of inflammatory mass formations at or near the distal tip of intrathecal catheters which infuse opioids, baclofen, or chemotherapy drugs into patients. For more detailed information, please see the link below to the company’s press release. |
Public Contact: |
Patients or customers may contact Medtronic Neuromodulation at 1-800-707-0933 from Monday - Friday, 7:00 a.m. – 6:00 p.m. (Central Standard Time) or by emailing the company at: http://www.medtronic.com/corporate/contact.jsp |
FDA District: |
Minneapolis |
FDA Comments: |
On January 16, 2008, Medtronic sent a letter to doctors who implant these devices and/or provide care to patients with the implanted device. The letter described the problem, patient risks, patient management, recommendations and next steps. For more information about this recall, including Medtronic's letter to physicians, please see the company’s press release at: http://wwwp.medtronic.com/Newsroom/ NewsReleaseDetails.do?itemId=1205958103681 Class 1 recalls are the most serious type of recall and involve situations in which there is a reasonable probability that use of the product will cause serious injury or death. Health care professionals and consumers may report adverse reactions or quality problems experienced with the use of this product to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by FAX.
|
Updated March 28, 2008
CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page
Center for Devices and Radiological Health / CDRH