Organic Laboratory Team
The Organic Laboratory Team analyzes samples of water, soils, and oils for pesticides, herbicides, polychlorinated biphenols (PCBs), and semi-volatile compounds. These samples come from a variety of places such as wells that provide drinking water to the public, waste treatment plants, chemical plants, landfills, or abandoned sites where hazardous waste has been disposed. The organic laboratory consists of an extraction laboratory, a pesticide/PCB laboratory, a volatile laboratory, and a semi-volatile laboratory.
Extraction Laboratory
In
the organic extraction laboratory, the analytes of interest are extracted
from the sample by adding a solvent such as methylene chloride, and then
reducing the extract to concentrate the analytes. It is then sent to the
analytical labs (gas chromatographs or mass spectrometers) to be analyzed.
Some of the extraction processes used for extracting water samples include Separatory Funnel, Accelerated One Step, Continuous Liquid-Liquid, and Excell. The extraction is performed using these methods by allowing a solvent such as methylene chloride to drip slowly through the water sample so the hazardous compounds detach from the water and attach to the methylene chloride. The methylene chloride compounds are then collected and reduced to a small amount of extract for analysis.
The process is similar with soils however, the solvent may drip through the soil or is vibrated by a probe. The vibration breaks up the soil, and the hazardous compounds detach from the soil and attach to the solvent. This extraction process is known as sonication. Another process which allows the soil to boil directly in the solvent is known as soxtherm. During the boiling action, hazardous compounds are transferred from the soil to the solvent. Again the solvent is collected and reduced to a small amount of extract for analysis.
Pesticide/PCB Laboratory (GC LAB)![Picture of chemists discussing pesticide analysis](gif/gc.jpg)
This laboratory performs Pesticide, PCB, and Herbicide analysis
on sample extracts provided by the extraction laboratory. The concentrated
solvent is injected into a gas chromatograph (GC) which separates the
many possible components for identification. The GC uses a detector called an
electron capture detector (ECD) that is specific for the halogen groups
which are common to many pesticides (specifically chlorine groups). This
somewhat selective detector helps to eliminate compounds from being identified
as pesticides if they dont have a halogen group. Compounds are identified
by the amount of time it takes for them to travel through the GC column
and reach the detector which then makes a peak on the chart for that compound.
Analytes detected are confirmed on a second GC column (and or detector)
that is different than the first. Only compounds identified by both columns
are reported as present.
Semivolatiles Laboratory (GC/MS)
This
laboratory uses Gas Chromatography/Mass spectrometry (GC/MS) to identify
the semivolatile
compounds
sometimes referred to as Acid/Base/Neutral. The solvent extract is taken
from the extraction laboratory and injected into a GC/MS instrument.The
GC separates the many possible components in thesolvent and the mass spectrometer
detects these analytes as they leave the GC column. The GC produces a
retention time for each component (same as in the Pesticide lab) and the
MS gives a fingerprint identification. The semivolatile compounds are
compounds that typically are not volatile enough to be analyzed as a volatile
by purge and trap, but still volatile enough to be analyzed by GC.
Volatiles Laboratory (GC/MS)
This
laboratory also uses GC/MS to identify the target analytes (volatile compounds
of VOAs). The unique feature of this analysis is the use of
the purge and trap (P&T) to extract the sample (the extraction laboratory
is not typically involved with VOA samples). The purge and trap bubbles
a gas (helium) through the liquid sample to force the volatile analytes
out and into a special trapping material that is in the P&T. The trap
concentrates the analytes and injects them into the GC/MS for analysis.