FDA-Cleared Sterilants and High Level Disinfectants with General Claims for Processing Reusable Medical and Dental Devices - March 2009

See Related

Manufacturer

Active Ingredient(s)

Sterilant Contact Conditions

High Level
Disinfectant Contact Conditions

K051305  TD-5 High-level Disinfectant

CS Medical, L.L.C.

2.65% glutaraldehyde

No indication for device sterilization.

Passes the AOAC Sporicidal Activity Test in 10 hrs at 22 °C.

5 min at 37.8ºC
Single use to be used exclusively with the TD-100 Transesophageal Probe Disinfector.
Contact conditions established by simulated use testing with endoscopes.

K041984 Acecide™ High Level Disinfectant and Sterilant
Minntech Corp 8.3% hydrogen peroxide
7.0% peracetic acid

Indication for device sterilization.

5 hrs at 25°C
5 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.

5 min at 25°C
5 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes
K041360 Aldahol III High Level Disinfectant
Healthpoint LTD 3.4% glutaraldehyde
26% isopropanol

Indication for device sterilization.

10 hrs at 20° C
Contact conditions based on AOAC Sporicidal Activity Test only.

10 min at 20ºC
14 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.
K012889 Banicide Advanced for Sterilization and High Level Disinfection
Pascal Company, Inc. 3.5% glutaraldehyde Indication for device sterilization.

10 hrs at 25°C
30 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.

45 min at 25°C
30 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.
K032959 Cidex® OPA Concentrate
Advanced Sterilization Products 5.75% ortho-phthaldehyde (OPA)

No indication for device sterilization.

In-Use solution (0.05% OPA) passes the AOAC Sporicidal Activity Test in 32 hrs at 50°C.

5 min at 50ºC
Single use – diluted by system
For use in the EvoTech Integrated Endoscope Disinfection System only.
Contact conditions established by simulated use testing with endoscopes.
K030004 Cidex OPA Solution

Advanced Sterilization Products

0.55% ortho-phthaldehyde

No indication for device sterilization. Passes the AOAC Sporicidal Activity Test in 32 hrs at 20°C and 25°C.

Manual Processing
12 min at 20°C
14 days Maximum Reuse

Automated Endoscope Reprocessor (AER)
5 min at 25°C
14 days Maximum Reuse
(For processing in an AER only with FDA-cleared capability to maintain solution temperature at 25°C.) Contact conditions established by simulated use testing with endoscopes.

K013280 Sterilox Liquid High Level Disinfectant System
Sterilox, Technologies, Inc. Hypochlorite and
Hypochlorous acid 650-675 ppm Active free chlorine
No indication for device sterilization. Passes the Modified AOAC Sporicidal Activity Test in 24 hrs at 25°C. 10 min at 25°C
Single use - generated on site
Contact conditions established by simulated use testing with endoscopes.
K003087 Sporicidin Sterilizing and Disinfecting Solution

Sporicidin International

1.12% glutaraldehyde
1.93% phenol/phenate

Indication for device sterilization.
12 hrs at 25°C
14 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.

20 min at 25°C
14 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.

K993042 Rapicide High Level Disinfectant and Sterilant

MediVators, Inc.

2.5% glutaraldehyde

Indication for device sterilization.
7 hrs 40 min at 35°C
28 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes and additional supporting information.

Automated Endoscope Reprocessor
5.0 min at 35°C
28 days Maximum Reuse
(For processing in an AER only with FDA-cleared capability to maintain the solution temperature at 35°C.)
Contact conditions established by simulated use testing with endoscopes.

K991487 °Cidex®OPA Solution High Level Disinfectant

Advanced Sterilization Products

0.55% ortho-phthaldehyde

No indication for device sterilization. Passes the AOAC Sporicidal Activity Test in 32 hrs at 20°C.

12 min at 20°C
14 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.

K974188 °Cetylcide-G Concentrate and Diluent Concentrate

Cetylite Industries, Inc.

3.2% glutaraldehyde

Indication for device sterilization.
10 hrs at 20°C
28 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.

40 min at 20°C
28 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.

K974062 MedSci 3% Glutaraldehyde

MedSci, Inc.

3% glutaraldehyde

Indication for device sterilization. 10 hrs at 25°C
28 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.

25 min at 25°C
28 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.

K972708 EndoSpor Plus Sterilizing and Disinfecting Solution
Note: Due to the lack of test strips for monitoring the concentrations of the active ingredients, the reuse period is limited to 14 days.

Cottrell Limited

7.35% hydrogen peroxide
0.23% peracetic acid

Indication for device sterilization.
180 min at 20°C
14 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.

15 min at 20°C
14 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.

K970230 SporoxTM Sterilizing & Disinfection Solution

Reckitt & Colman Inc.

7.5% hydrogen peroxide

Indication for device sterilization.
6 hrs at 20°C
21 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.

30 min at 20°C
21 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.

K960513 PeractTM 20 Liquid Sterilant/Disinfectant

Minntech Corporation

1.0% hydrogen peroxide
0.08% peracetic acid

Indication for device sterilization.
8 hrs at 20°C
14 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.

25 min at 20°C
14 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.

K932922 Procide 14 N.S.

Cottrell Limited

2.4% glutaraldehyde

Indication for device sterilization.
10 hrs at 20°C
14 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.

45 min at 20°C
14 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.

K932922 OmnicideTM Long Life Activated Dialdehyde Solution

Cottrell Limited

2.4% glutaraldehyde

Indication for device sterilization.
10 hrs at 20°C
28 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.

45 min at 20°C
28 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.

K932922 OmnicideTM Plus

Cottrell Limited

3.4% glutaraldehyde

Indication for device sterilization.
10 hrs at 20°C
28 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.

45 min at 20°C
28 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.

K931592 Metricide Plus 30 Long-Life Activated Dialdehyde Solution

Metrex Research, Inc.

3.4% glutaraldehyde

Indication for device sterilization.
10 hrs at 25°C
28 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only. FDA accepted Metricide Plus as identical to Cidex Plus.

90 min at 25°C
28 days Maximum Reuse
FDA accepted Metricide Plus as identical to Cidex Plus.

K931052 Metricide 28 Long-Life Activated Dialdehyde Solution

Metrex Research, Inc.

2.5% glutaraldehyde

Indication for device sterilization.
10 hrs at 25°C
28 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only. FDA accepted Metricide 28 as identical to Cidex Formula 7.

90 min at 25°C
28 days Maximum Reuse
FDA accepted Metricide 28 as identical to Cidex Formula 7.

K930284 Metricide Activated Dialdehyde Solution

Metrex Research, Inc.

2.6% glutaraldehyde

Indication for device sterilization.
10 hrs at 25°C
14 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only. FDA accepted Metricide as identical to Cidex.

45 min at 25°C
14 days Maximum Reuse
FDA accepted Metricide as identical to Cidex.

K924434 Cidex™ Activated Dialdehyde Solution

Johnson & Johnson Medical Products

2.4% glutaraldehyde

Indication for device sterilization.
10 hrs at 25°C
14 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.

45 min at 25°C
14 days Maximum Reuse Contact conditions based on literature references.

K924334 Cidex Formula 7™ Long-Life Activated Dialdehyde Solution

Johnson & Johnson Medical Products

2.5% glutaraldehyde

Indication for device sterilization.
10 hrs at 20-25°C
28 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.

90 min at 25°C
28 days Maximum Reuse
Contact conditions based on literature references.

K923744 Cidex PlusTM 28 Day Solution

Johnson & Johnson Medical Products

3.4% glutaraldehyde

Indication for device sterilization.
10 hrs at 20-25°C
28 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.

20 min at 25°C
28 days Maximum Reuse Contact conditions based on literature references.

K914749 Wavicide - 01

Wave Energy Systems

2.5% glutaraldehyde

Indication for device sterilization.
10 hrs at 22°C
30 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.

45 min at 22°C
30 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.

K875280 STERIS 20TM Sterilant
Note: Cleared for use with the STERIS System 1TM Processor only.

STERIS Corporation

0.2% peracetic acid

Indication for device sterilization.
12 min at 50-56°C
Single use only.
Contact conditions established by simulated use testing with endoscopes and passing a modified AOAC Sporicidal Activity Test.

No indication for high level disinfection.

K080420 Resert XL HLD High Level Disnfectant

Steris Corporation

2.0% Hydrogen peroxide

No indication for device sterilization.
In-Use solution passes the AOAC Sporicidal Activity Test in 6 hrs at 20 °C

8 min at 20ºC
21 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.

K070627 Opaciden Solution

Ciden Technologies, LLC

0.60% ortho-phthalaldehyde (OPA)

 

No indication for device sterilization.
In-Use solution passes the AOAC Sporicidal Activity Test in 32 hrs at 21 ° C and 32 hrs at 25-27°C.

For Manual Processing:
12 min at 20ºC
14 days Maximum Reuse
For processing in a legally marketed Automated Endoscope Reprocessor capable of maintaining solution temperature at 25ºC:
5 min at 25ºC
14 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.

K063159 Sterilox High Level Disinfectant System

Sterilox, Technologies, Inc.

Hypochlorous acid/hypochlorite
400-450 ppm Active free chlorine

No indication for device sterilization.
Passes the Modified AOAC Sporicidal Activity Test in 32 hrs at 30 ° C.

10 min at 30ºC
Single use - generated on site
Contact conditions established by simulated use testing with endoscopes.
K060618 Cidex Activated Dialdehyde Solution
Advanced Sterilization Products

2.4% glutaraldehyde

10 hrs at 25 °C
14 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test and by simulated use testing with endoscopes.

45 min at 25 °C
14 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.


Updated March 30, 2009