The Center for Devices and Radiological Health (CDRH) is accepting and encouraging sponsors of premarket medical device submissions to include a copy of their submission in the prescribed electronic form along with the required paper copies.
The answers to the following questions explain how to create a complete electronic copy. Please read all of the information carefully before submitting an electronic copy. To quickly navigate to a particular question and answer, click on the specific link below.
A: An electronic copy is an exact duplicate of a paper submission, created and submitted on a CD or DVD, accompanied by a copy of the signed cover letter and the complete original paper submission. An electronic copy is not an electronic submission. Electronic copies of premarket submissions accompanying the paper submission should be submitted to the CDRH Document Mail Center.
For more information on electronic submissions, refer to:
A:Any submission type (510(k), IDE, PMA, 513(g), Pre-IDE, HDE, PDP, Emergency Use Authorization, and subsequent amendments or supplements) can be submitted as an electronic copy. Just indicate in the cover letter that an electronic copy of the submission is being provided and that it is an exact duplicate of the paper copy.
A:The electronic copy should be sent in an Acrobat Portable Document Format (PDF) because Center staff will use Adobe’s Acrobat 8.0 to view the submission. This will assure that what a reviewer sees on the screen is the same as what has been submitted on paper. It is also important to follow the correct naming conventions for files and folders when submitting an electronic copy. For documents that are scanned, Optical Character Recognition (OCR) should be performed prior to submission.
Note:At this time, we cannot support the new features in Adobe Acrobat 8.0. Attachments, embedded files or use of the bundled files feature within Adobe Acrobat are not processed by FDA systems. Electronic copies containing files with attachments will not be processed and a Replacement CD will be requested (see Q14).
Many Acrobat Portable Document Format (PDF) files created by a third party vendor such as journal articles, user guides, product labels etc. contain attachments which are embedded during the creation process. Such attachments may be named “…joboptions”, for example. To detect these attachments, open the Acrobat Portable Document Format (PDF) and click the paper-clip icon in the lower left-hand corner of the Navigation Panel as shown below.
Figure 1: Attachments
Attachments must be removed before submittal. To remove the attachment:
Contact the vendor directly and ask that they create the Acrobat Portable Document Format (PDF) without appending any attachments like the one shown above.
OR
Print the file selecting Adobe PDF as the Printer Name (see Figure 11). The file will be printed to a new Acrobat Portable Document Format (PDF) file without the attachment.
Manually deleting this file will not delete the attachment attribute from the file entirely. Attachments will still be detected and cannot be processed by FDA systems.
Additional Files (non-PDF)
In addition to the PDF files, the submission may contain other data files typically provided electronically (e.g., clinical data) in support of the review. The following files types will be accepted: SAS, XPORT, XML, SGML, and Molfile. For specific instructions on electronic clinical data, please visit: Clinical Data for Premarket Submissions.
Microsoft Word (or comparable products) is not an accepted format for electronic copies.
Statistical Data
Files submitted under the “Clinical Data for Premarket Submissions” instructions or any statistical data provided as SAS, XPORT, XML, SGML and Molfiles can be placed in a zip file(s) and placed within a STATISTICAL DATA folder as shown below.
Figure 2: Statistical Data Folder
Media Files
Files that do not require additional software to view can be sent for the following media types: .gif, .tif, .jpg, .avi, .wmv
Place all files in a zip file and place it within a MEDIA FILES folder as shown below.
Figure 3: Media File Folder
Note: If you want to submit a file type that is not listed above, please send your request to: cdrhesub@cdrh.fda.gov and we will contact you to discuss the alternatives.
A: Electronic copies should be sent on CD-ROM or DVD. We encourage applicants to select media that allows the submission to be contained on one CD-ROM or DVD, whenever possible.
A: All CDRH submissions for ODE and OIVD, whether or not they contain an electronic copy, are sent to the same location:
FDA Center for Devices and Radiological Health
Document Mail Center, HFZ-401
9200 Corporate Boulevard
Rockville, MD 20850
A: At present, documents that require signatures cannot be submitted with an electronic signature. This includes, but is not limited to: IDE; PMA; 510(k); Master Access File (MAF) cross-reference letters; financial disclosure documents; 510(k) truth and accuracy statements; 510(k) statements; 510(k) class III certifications; declarations of conformity to consensus standards; and third party documents.
To ensure the electronic copy is complete, it is essential that you:
After the Question & Answer section, a checklist is provided listing the essential components of an electronic copy. Before submitting your electronic copy, please ensure you are submitting a formatted electronic copy.
Important Note:A scanned PDF is not a substitute for the original paper copy. The submission will not be complete without the paper copy of the cover letter with an original handwritten signature. The electronic copy (CD/DVD) and the accompanying cover letter should be identified as the electronic copy. (See Figure 4 in the question below.)
A:Currently, an electronic copy that follows this document will serve as one of the required number of copies for the various premarket submission types. For the submission to be complete, the electronic copy should be considered one unit of the regular paper submission requirements. Figure 2 depicts each unit that is required for the submission.
Figure 4: Submission contents and its discrete units
As per this document, copies of a submission are required as indicated below. In the case of an electronic copy (which includes the paper original), the electronic media counts as:
Premarket Notifications - 510(K)
513(g)
Pre - Investigational Device Exemption (PIDE)
Investigational Device Exemption (IDE)
Premaket Applications (PMA), Humanitarian Device Exemptions (HDE)
Master AccessFiles (MAF)
A: The structure of an electronic copy is highly dependent on the overall file size of the submission and can be organized using files and folders in one of the following ways:
Note: File size should not exceed 50MB.
Figures are presented below to illustrate how to name and structure a submission. Figure 5 provides a sample PMA submission that contains volumes. Figure 6 provides an example of a PMA submission that contains multiple volumes as PDF files. Figure 7 provides an example of a submission that contains multiple PDF files, as in the case of a 510(k) submission. Figure 8 provides an example of an IDE with volumes. Figure 9 provides an example of a 510(k) structured in one PDF document with bookmarks.
Figure 5: Example of a PMA submission with Volumes
Figure 6: Example of a PMA with PDF Volumes
Figure 7: 510(k) submission without Volumes
Figure 8: Investigational Device Exemption (IDE) with Volumes
Figure 9: Sample of a 510(k) structured in one PDF file with Bookmarks
Note: In Figure 9 above, the example illustrates how to consolidate the multiple sections of a submission into one file with the use of bookmarks. This is in contrast to the multi-file submission displayed in Figure 7.
If you have questions about any one of the structures above, please contact CDRH at cdrhesub@cdrh.fda.gov to ensure that the format is acceptable prior to developing your submission.
A: There are several ways to format an electronic copy. However, at a minimum, it is required that the following instructions are used to create a compliant electronic copy of the submission, to include: naming of folders and files; bookmarks and hypertext links; and Optical Character Recognition (OCR). The following instructions are provided with regard to (1) naming of folders and files; (2) bookmarks and hypertext links; (3) optical character recognition; and (4) file size.
1. Naming of folders and files
Submissions may be sent in any of the following structures:
- Volume-based submission (generally recommended for large or complex submissions), create either multiple volumes organized into folders containing each volume's associated files or create one PDF document for each volume
- Non-Volume based submission (generally recommended for small submissions), create either one PDF document for each table of contents entry or create a single PDF document with bookmarks corresponding to the table of contents.
The following section outlines the required naming convention for all folders and files:
- Folder Naming Convention for Content Volumes: If the submission is large it should contain volumes, each volume should be a folder with the following naming convention for the folder: VOL_XXX Name or VOL_XXX (e.g., Vol_005 Patient Study Data). The XXX is the sequential volume number (i.e., 001, 002, 003, etc). The volume numbering in a submission must be non-repeating consecutive numbers. The limit for volumes is 999. For example, a folder file can be named:
- VOL_001_Name or VOL_001
- If a descriptive name is used for a folder, the folder name should be descriptive of its content and meaningful to the reviewer. The descriptive name can be up to 250 characters but must not contain special characters (tilde (~), asterisk (*), forward slash ( /), backward slash (\), colon (:), question mark (?), single quotation marks (‘), double quotation marks (“), less than sign (<), greater than sign (>) or vertical bar (|)).
Note: Content Volume Folders can only be named: Vol_001 thru Vol_999 or Vol_001 Description thru Vol_999 Description. It is important to follow the folder naming convention because it is used to upload the submission into CDRH’s document repository. If this naming convention is not used, the submission will fail the loading process. Volumes should be used when the submission has a large quantity of files.
An example of a file name is as follows:
- XXX_Name.file extension (e.g., 013_Proposed Labeling.pdf).
For files that are organized as volumes, the file name is as follows:
- XXX_Vol X Name.file extension (i.e., 001_Vol 001 Administrative.pdf).
Note: Files size should determine when to create more than one file in a submission. If the file size is greater than 50MB, consider splitting the contents into multiple files and use the next number in the sequence. It is important to follow the file naming convention because it is used to order the submission in the correct sequence in CDRH’s document repository. Therefore, a file name should not be repeated or deviate from the sequential order by using decimals. When possible, the file name should be descriptive of its content and meaningful to the reviewers.
If you have questions about complex submissions, please contact cdrhesub@cdrh.fda.gov prior to submission to ensure the electronic copy structure is acceptable.
Bookmarks and hyperlinks should be used to assist the reviewers in navigating through the content of the submission. If you used either bookmarks or hypertext links, consider the following:
Important Note:Hyperlinks between individual PDF document files are not currently supported and any absolute links that reference across files will not work.
In general, for documents with a table of contents, provide bookmarks and hypertext links for each item listed in the table of contents including all tables, figures, publications, other references, and appendices. These bookmarks and hypertext links are essential for the efficient navigation through documents.
3. Optical Character Recognition
PDF documents produced by scanning paper documents are usually inferior to those produced from an electronic source document such as MS Word. Scanned documents are more difficult to read and do not allow the reviewers to search or copy and paste text for editing. The use of scanned documents should be avoided if at all possible. If scanning cannot be avoided, the following is highly recommended:
Quick Tip: Once a document has been scanned and completed optical character recognition, you may take the following steps to ensure that the text is searchable:
Figure 10: Selecting text in a PDF Document
If the source document is only available on paper, it should be scanned at resolutions that will ensure the pages are legible both on the computer screen and when printed. At the same time, remember to limit the file size to be less than 50MB. We recommend scanning at a resolution of 300 dots per inch (dpi) to balance legibility and file size. We discourage the use of grayscale or color because of file size. After scanning, avoid re-sampling to a lower resolution.
For files with images and photographs:
Also, when creating PDF files containing images, you should not resample images. Re-sampling does not preserve all of the pixels in the original. For photographs, the image should be obtained with a resolution of 600 dpi. If black and white photos are submitted, consider 8-bit gray scale images. If color photos are submitted, consider 24-bit RGB Color Model images. A captured image should not be subjected to non-uniform scaling (i.e., sizing).
Files with scanned images and photographs tend to be large in file size. Please do not exceed 50MB for a single file. Consider multiple files for these types of documents.
Note: Scanned tables and graphs cannot be extracted easily if scanned. Most OCR programs will distort the data in tables and graphs. Convert MS Word documents to PDF, as this method usually retains the formatting.
For a paper document with handwritten notes:
Paper documents containing handwritten notes should be scanned at 300 dpi. These handwritten notes should be made in black ink for clarity.
If you have questions about creating electronic copies, please contact CDRH at cdrhesub@cdrh.fda.gov to ensure that the format is acceptable prior to developing your submission.
An electronic copy CD/DVD can have one or multiple files. There is not a limitation of the total size of the submission, but each file should be limited to 50MB in file size. There are several ways to compress file size, including but not limited to: performing Optical Character Recognition, reducing file size in Adobe and creating logical section breaks.
If you have questions about file size, please contact CDRH at cdrhesub@cdrh.fda.gov to ensure the file size is acceptable.
A:Choose a method for creating PDF documents that produces the best replication of a paper document. You can ensure that the paper copy is an exact duplicate of the electronic copy by using a content rendering software to print the Word document to a PDF version of the file. Once you have the final PDF version, simply print the PDF version for submission as the original paper copy.
The PDF files must be compatible with Adobe Acrobat 6.0, 7.0 and 8.0 without the use of additional plug-ins other than those provide by Adobe as part of Acrobat. There are features in Adobe Acrobat 8.0 that cannot be supported. If you would like to submit Adobe Acrobat 8.0 files, please contact CDRH at cdrhesub@cdrh.fda.gov to ensure you do not use a feature that is unsupported (Also see the common mistakes section below).
1. Converting a MS Word Document to PDF using Adobe Acrobat 6.0 Professional
Converting a MS Word Document to PDF using Adobe Acrobat 6.0 Professional
After you have completed your Word document (.doc format), you can easily create a portable document format (.pdf) file using the same process you use to print documents. In Microsoft Word, instead of printing your document to a printer, you can select the Adobe PDF printer option from the drop down print menu, and create a new file. (See sample in Figure 11.) Note: All figures in this section are using Adobe Acrobat 6.0. If you are using another version of Adobe Acrobat, the figures may be slightly different.
Figure 11: Selecting the Adobe PDF option under the Print Function
After you select Adobe PDF and press OK, you are prompted to save your document as a .pdf file and to select a location to save it. (See Figure 12.) Remember to use the correct naming convention. The PDF document will be stored in the location in which it was saved.
Figure 12: Saving your Word Document to a specified location as a PDF file
2. Fonts
PDF viewing software automatically substitutes a font to display text if the font used to create the text is unavailable on the reviewer’s computer. Font substitution can affect a document’s appearance and structure, and in some cases it can affect the information conveyed by a document. We cannot guarantee the availability of any one font. Therefore, you should embed all fonts you are using in the PDF files to ensure that those fonts will always be available to the reviewer. When embedding fonts, all characters for the font should be embedded (not just a subset of the fonts being used in the document). One problem associated with embedding fonts is that embedding requires additional computer storage space. Three techniques to help limit the storage space taken by embedding fonts:
3. Page Orientation
Pages should be properly oriented. For example, you should set the page orientation of landscape pages to landscape prior to saving the PDF document in final form to ensure correct page presentation. Landscape pages (including tables) should be oriented such that the header of the document aligns with the left edge of the page and the footer of the document aligns with the right edge of the page.
4. Page Size and Margins
The print area for pages should fit on a sheet of paper that is 8.5 inches by 11 inches. You should allow a margin of at least 1 inch on all sides to avoid obscuring information if the pages are subsequently printed and bound.
If a submission includes files divided into more than one PDF document, such files should have contiguous pagination. To simplify navigation, please make page numbers for the paper document and the PDF file the same. In order to ensure that the reviewer can locate information by using the page numbers as a reference (see Figure 13), consider naming the files something that reflects the pages involved (e.g., 001_Manufacturing_1-100.pdf should precede a file named 002_Manufacturing_101-200.pdf).
This will also ensure that the pagination is consistent throughout the submission. For submissions organized in volumes, it is suggested that each volume start its pagination with page 1.
Figure 13: Example of a PMA submission with Volumes and using page numbers in the file names
7. Document Properties - Description Tab
Document properties are used to search for individual documents and to identify the document when found. To modify document properties, from the tool bar navigate to File | Document Properties. Figure 14shows the Document Properties Description tab, the keywords can be added to the Keyword field.
Figure 14: Sample of Document Properties
8. Document Properties – Initial View Tab
In the Document Properties box, the Initial View tab can also be found. (See Figure 15) In the Document Options section, all PDF files should set the Show field as Bookmarks Panel and Page. If there are no bookmarks, set the initial view as Page only. Set the Page Layout and Magnification fields to default.
Figure 15: Set Document Properties to always open with Bookmarks and page view
9. Document Properties – Security
PDF files are stored as original documents and will not be altered from their original form. Remove any security settings, read-only, or password protection used on the files. Electronic Copies submitted with security settings will not be accepted.
10. Indexing PDF Documents
Full text indexes are used to help find specific documents and/or to search for text within documents. When a document or group of documents is indexed, all words and numbers in the file and all information stored in the Document Information fields are stored in special index files that are functionally accessible using the search tools available in Acrobat. Portions of a document that are imaged are not indexed. Even if the document only contains images, the text in the Document Information fields of the file will be indexed. All PDF files should be full-text searchable prior to submitting to the FDA.
11. Plug Ins
It is acceptable to use plug-ins to assist in the creation of a submission. However, the review of the submission should not require the use of any plug-ins other than any provided by Adobe as part of Acrobat, and must be compatible with Adobe Acrobat 5 or greater.
A: Submitting an electronic copy of your application makes that copy available to review staff more quickly, enables additional editing, saves resources as well as provides navigational and searching tools. However, it should be understood that submission of an electronic copy does not provide preferential treatment or change the order in which applications are reviewed.
A: An EUA request may be submitted with an electronic copy. To obtain further information regarding EUA requests, we encourage you to refer to the Guidance entitled “Emergency Use Authorization of Medical Products,” which may be found on the FDA website at http://www.fda.gov/oc/guidance/emergencyuse.html.
A: Many of the electronic copies submissions received to date have not followed all of the instructions presented on this webpage.
Common mistakes | Suggested solution |
---|---|
Cover letter does not state that the electronic copy is an exact duplicate of the original paper submission. | Put the following statement in your cover letter, “Per the instructions accessed at http://www.fda.gov/cdrh/elecsub.html, an electronic copy is being provided with this submission and it is an exact duplicate of the original paper submission.” |
Files do not follow the numbering convention. For example:
|
Files should be named with the following convention: Files : XXX_ plus name extension. The XXX represents a numeric sequence (i.e., 001, 002, 003, etc) for which the files should be ordered in the table of contents. For example:
Keep all files in sequential order; do not use decimals to organize the content. Create a new file and use the next number in the sequence. |
Files have security settings | Remove all Security settings on the PDF document(s). |
nappropriate use and naming of folders. For example:
|
Folders should be placed at the root (in the example below that is “IDE Sample”) of the CD/DVD. Content Volume Folders: VOL_XXX. The XXX represents a numeric sequence (i.e., 001, 002, 003, etc.) for which the folders should be ordered in the table of contents. For example:If only one file exists for a volume, it is suggested that you do not use volumes to organize your submission. Statistical Data Folders: STATISTICAL DATA. There should be no additional labels placed on this folder. Media File Folders: MEDIA FILES. There should be no additional labels placed on this folder. |
Subfolders within the Volume folders. | Subfolders are not to be placed within a volume for any reason. If a section needs to be organized further, consider creating a new volume for that section. |
File sizes greater than 50MB (i.e., 250MB) | The submission content should be organized in volumes or by logical headings to reduce the overall size of each file. Always check file size before submitting your electronic copy. To reduce the file size on a PDF document, complete the following steps:
The reduction in file size varies document to document. If this does not reduce the file size to less than 50MB, then the content in the document should be split into multiple PDF files. |
Use of unsupported Adobe Acrobat functionality, specifically the packaging of multiple files in one PDF |
Our systems do not support attachments or embedded files within an Adobe Acrobat document. With Adobe Acrobat 8.0 there is a function that allows files to be packaged together, which we do not support. Please refrain from using this type of document packaging. The CD/DVD should contain individual files at the root level of the CD/DVD unless you are using the Volume folders. |
Index Files are provided on the CD/DVD and linked to PDF documents, see below: |
CDRH systems do not support the storage of index files and linkages within PDF documents. Please remove all Adobe Acrobat index folders and files, which are packaged as follows:
|
Unallocated Space Files are placed on the CD/DVD | CDRH systems do not support the processing of electronic copies that have “unallocated space” files. This may be a byproduct of CD burning software. Please do not configure your CD burning software to consume the additional disk space on the CD/DVD. |
Please contact the CDRH Office of Information Technology at cdrhesub@cdrh.fda.gov regarding any technical details of the construction of the PDF file.
Any questions regarding FDA regulations or guidance should be directed to the appropriate division point of contact listed below.
A: If an electronic copy fails during processing, the submitter will be contacted via telephone or e-mail explaining the reason for failure and asked to submit a Replacement CD. In most cases, review of the submission will not begin until a Replacement CD is provided and can be processed successfully.
Any questions regarding FDA regulations or guidance should be directed to the appropriate division point of contact listed below.
ODE Divisions and Contacts
ODE Program Operations Staff
(240) 276-4040
Division of General, Restorative, and Neurological Devices
Assistant Director of Program Operations (ADPO): (240) 276-3737
Division of Cardiovascular Devices
Project Manager : (240) 276-4080
Division of Ophthalmic and ENT Devices
ADPO: (240) 276-4200
Division of Reproductive, Abdominal, and Radiological Devices
Project Manager: (240) 276-3636
Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices
Project Manager: (240) 276-3700
OIVD Contacts
Sousan Altaie: (240) 276-0378
Patricia Beverly: (240) 276-0379Updated April 29, 2009
CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page
Center for Devices and Radiological Health / CDRH