Reopening Dialysis Clinics

FDA Advice to Follow when Reopening Hemodialysis Clinics after Restoration of Power and Water

See Related Information

PDF

This document gives recommendations to follow before reopening dialysis clinics after a period of time without power or water.

NOTE: If you perform dialysis at home, and you have lost power or water, see V. Home Hemodialysis Systems.

Before you reopen your dialysis clinic, check each component using the guidelines in each section below.

I. Water Treatment Systems
II. Dialysis Machines
III. Filters (Dialyzers) and Tubing
IV. Dialysate and Dialysate Mixing Systems
V. Home Hemodialysis Systems
VI. Additional Information

I. Water Treatment Systems

All dialysis units require purified, dialysis-quality water. Most dialysis units use on-site water treatment systems to purify the municipal water supply. Note that the loss of power to a water treatment system is just as serious as the lost of a potable water supply. Both can result in the potential contamination of water.

WARNING! Upon resumption of power and a potable water supply, DO NOT attempt to use water from your water treatment system for treating dialysis patients until you have fully assessed your water treatment system, replaced any contaminated system components, and disinfected and thoroughly rinsed the entire system.

Flooded Facilities:
If your facility flooded, even for a short time, the water treatment system and distribution loops may be contaminated. Flooding can also lead to electrical shorting of your power supply and loss of power, which compounds the problem.

Contaminated Water Supply:
If either your municipal water supply or your water treatment system is or was contaminated, it may be difficult to produce dialysis-quality water. Follow the steps below to determine whether or not you may resume producing water for dialysis.

Back to top

II. Dialysis Machines

Dialysis machines must be clean, dry, and disinfected prior to use.

WARNING! Do NOT use your dialysis machine if the outside or any internal components are wet, if your facility flooded, or if your facility was exposed to corrosive chemicals or excessive heat (above the limits listed in the manufacturer’s specifications).

Flooded Facilities:
If your facility flooded, even for a short time, the dialysis machines may be contaminated.

Power Loss:
If your clinic lost power, your dialysis machines could be contaminated. Follow the steps below before using your dialysis machines.

Back to top

III. Filters (Dialyzers) and Tubing

Hemodialysis systems use disposable dialyzers and tubing. Some dialyzers can be reused for the same patient.

WARNING! Do NOT reuse a dialyzer that is intended for single use. Follow instructions on dialyzer label to determine if it can be reprocessed for reuse.

Flooded Facilities:
If your facility flooded, even for a short time, the seals and packaging on disposable components may be compromised.

Power Loss:
If your clinic lost power, your supplies may be compromised. Follow the steps below to make sure your dialyzers and tubing are safe to use.

Back to top

IV. Dialysate and Dialysate Mixing Systems

Dialysate powder must be kept dry before mixing with purified water.

WARNING! Do NOT use dialysate concentrate that is standing in water, even if the container appears to be sealed, because it is difficult to determine if there has been seepage into the container.

Flooded Facilities:
If your facility flooded, even for a short time, the dialysate may have become wet.

Power Loss:
If your clinic lost power, your dialysate or dialysate mixing system may have been compromised. Follow the steps below before using your dialysate.

Back to top

V. Home Hemodialysis Systems

If you use a home hemodialysis system, and you lost power or water, your dialysis machine may not work properly. Check with your clinic or kidney doctor (nephrologist) if you are unsure about resuming treatment.

WARNING! Do NOT do home dialysis if you are unsure whether or not your water supply is clean or if your dialysis machine was in a room that flooded.

Back to top

VI. Questions and Answers

Q. Can you use your water treatment system on water that has been trucked in from another location instead of using the municipal water supply?
A. Yes, if the water complies with EPA standards for potable water. For more information on trucked water considerations, see: http://www.cdc.gov/ncidod/hip/Dialysis/dialysis.htm.

Q. Can you do dialysis with sterile, bagged dialysate?
A. Yes, if you have a dialysis system that uses sterile bagged dialysate that does not require dilution, and you have all necessary components.

Q. Does FDA need to inspect my dialysis machines before I am ready to restart them?
A. No.

Q. Can you use a reverse osmosis water treatment system that FDA has not cleared for dialysis if that is your only way to get purified water?
A. Yes, if an unapproved reverse osmosis water treatment system is the only device you have available, you may use it to prepare dialysis-quality water. Under Section 561(a) of the Food, Drug, and Cosmetic Act, (21 U.S.C. 360bbb), and FDA’s Guidance on Investigational Device Exemption (IDE) Policies and Procedures, devices not yet approved or cleared for marketing may be used in an emergency situation. As with any water treatment system, you must first test the water output to make sure it complies with AAMI RD62:2001.

Within 5 days of using the system or learning of its use, you should send the following information to the FDA address below

The Food and Drug Administration
Center for Devices and Radiological Health
Office of Device Evaluation