ISSUE DATE: November 2, 2007

PROGRAM INFORMATION BULLETIN NO. P07-27

FROM:          MARK E. SKILES  Linda Zeiler for
                       Director of Technical Support

                       KEVIN G. STRICKLIN  KEVIN G. STRICKLIN
                       Administrator for
                       Coal Mine Safety and Health

                      FELIX A. QUINTANA  FELIX A. QUINTANA
                       Administrator for Metal and Nonmetal
                       Mine Safety and Health

SUBJECT:    General Monitors Model 420, 420d, S500, and S800 Machine Mounted
                        Methane Monitoring Systems

Who needs this information?
Mine Safety and Health Administration (MSHA) personnel, underground mine operators, miners' representatives, and mine equipment repair/rebuild facilities should have this information.

Why is MSHA issuing this Program Information Bulletin?
This Program Information Bulletin (PIB) is to inform the mining community of a retrofit program for the calibration cup for the General Monitors Machine Mounted Methane Monitoring Systems listed in the table below. In addition, General Monitors added a verification ("bump test") to the calibration procedure to ensure the unit is properly calibrated.

What equipment does this retrofit program affect?
This retrofit program affects the following General Monitors Machine Mounted Methane Monitoring Systems:

MODELMSHA CERTIFICATION NUMBER
42032A-6/MS-0
420d32A-16/MS(-0)(-1)(-2)(-3), 32A-01/MD-0
S50032A-21/MS-0
S80032A-23/MS-0, 32A-03/MD-0


Calibration cup specifications were added to the MSHA certifications listed in the table above.

Why do the General Monitors Machine Mounted Methane Monitoring System calibration cups require a retrofit?
The General Monitors Machine Mounted Methane Monitoring System calibration cups require a retrofit for the following reasons:

   (1) During performance testing at MSHA's Approval and Certification Center (A&CC) a calibration problem was identified. The monitor appeared to be properly calibrated; however, when the monitor was "bump tested" with the specified calibration kit, inaccurate readings were observed.

   (2) Previously there was no control over the wide variety of calibration cups that were used for calibrating the system. Calibration cup documentation was added to the MSHA certification to ensure the proper calibration cup is used.

What modifications were made to the calibration cup?
The following modifications were made to the calibration cup:

   1) The calibration cup inside diameter was modified to allow use with the sensor splash guard or dust guard.

   2) A label was added to the calibration cup to designate the part number, revision, and the Machine Mounted Methane Monitoring System models on which it is used.

   3) A clear heat-shrink tube was added to the outside of the calibration cup to protect the label and to ensure the calibration cup body and spout remains intact.

How can the proper calibration cup be identified?
The calibration cup can be identified by the label and clear heat shrink tube as shown below. The label contains the part number and revision (1400154, Rev. J) and the models on which it is used (420, 420d, S500, and S800).

calibration cup


What modifications were made to the calibration instructions?
A "bump test" was added to the instructions to verify the unit is properly calibrated.

What is the recommended action concerning this retrofit program?
General Monitors' distributor is providing modified calibration cups and a revised calibration manual to all known customers that use any of the models listed in the table above. If you use any of the models listed in the table above and have not received a modified calibration cup and revised calibration manual, you should contact General Monitors' distributor. See contact information below.

All previous calibration cup designs must be removed from service. All previous calibration cup designs should be disposed of to prevent future use.

Calibration Cup part number 1400154, Rev. J is the only calibration cup that can be used with the models listed in the table above.

Contact Information:
Mr. Steven F. Tomalewski
United Central Industrial Supply Company
649 Blue Ridge Drive
Apollo, PA 15613-1939

Phone & Fax: 1 (724) 334-7147
Cell: (724) 875-5955
E-mail: stomalewski@verizon.net

What happens to the Machine Mounted Methane Monitoring Systems that are not calibrated with the modified calibration cup?
At the conclusion of the retrofit program, ONLY General Monitors' Models 420, 420d, S500, and S800 Machine Mounted Methane Monitoring Systems calibrated with the modified calibration cup will be considered as meeting the applicable certification requirements of Title 30 Code of Federal Regulations (30 C.F.R.). The retrofit program will conclude on February 2, 2008.

What is the background of this PIB?
During testing of the Model S800 at the A&CC, the monitor failed to provide an impulse to actuate a power shutoff component when a gas mixture concentration of 2.1% methane in air was applied to the sensor. An investigation revealed that the system was not properly calibrated. A "bump test" was added to the calibration procedure to ensure the system is properly calibrated. In addition, the calibration cup was modified and added to the MSHA certifications to ensure that the proper calibration cup is used.

What is MSHA's authority for this PIB?
The Federal Mine Safety and Health Act of 1977, as amended, 30 U.S.C. § 801 et seq.; 30 C.F.R. §27.24(b)(1).

Who are the MSHA contact persons for this PIB?
Coal Mine Safety and Health, Safety Division
Steve Gigliotti, (202) 693-9479
E-mail: Gigliotti.Stephen@dol.gov

Metal and Nonmetal Safety and Health
Neal Merrifield, (202) 693-9640
E-mail: Merrifield.Neal@dol.gov

Technical Support, Approval and Certification Center
Steve Cole, (304) 547-2304
E-mail: Cole.Stephen@dol.gov

Is this PIB available on the internet?
This PIB may be viewed on the World Wide Web by accessing MSHA's home page at (http://www.msha.gov) and choosing "Compliance Info " and "Program Information Bulletins. "

Who will receive this PIB?
Program Policy Manual Holders
MSHA Enforcement
Underground Mine Operators
Mine Equipment Repair and Rebuild Facilities
Miners' Representatives
Special Interest Groups




Back to Top   Back to Top www.msha.gov www.dol.gov

Frequently Asked Questions | Freedom of Information Act | Customer Survey
Accessibility | Privacy & Security Statement | Disclaimers

Mine Safety and Health Administration (MSHA)
1100 Wilson Boulevard, 21st Floor
Arlington, VA 22209-3939
Phone:    (202) 693-9400
Fax-on-demand: (202) 693-9401
Technical (web) questions: Webmaster
On-line Filing Help: MSHAhelpdesk@dol.gov
or call (877) 778-6055

Contact Us