Usefulness of the Tsui Test in Combined Spinal-Epidural Analgesia in Labour
This study is ongoing, but not recruiting participants.
First Received: April 23, 2007   Last Updated: December 12, 2007   History of Changes
Sponsored by: Samuel Lunenfeld Research Institute, Mount Sinai Hospital
Information provided by: Samuel Lunenfeld Research Institute, Mount Sinai Hospital
ClinicalTrials.gov Identifier: NCT00464841
  Purpose

Combined spinal epidural anesthesia (CSE) is a very effective technique to provide labor analgesia. One of the disadvantages of this technique is the delay in recognizing an error in the position of the epidural catheter because of the effects of the spinal component. Eventually in case of a misplaced catheter, the patient will experience pain or discomfort requiring a repeat procedure after the effect of drug given during the spinal wears off. Low current electrical stimulation test, or the Tsui test, has been used successfully to confirm catheter location in the epidural space. The investigators' objective in this study is to test the usefulness of the Tsui test to confirm the correct placement of the epidural catheter during CSE in laboring patients.


Condition Intervention
Labor Pain
Procedure: Combined spinal epidural
Procedure: Tsui Test for intrathecal catheter

U.S. FDA Resources
Study Type: Interventional
Study Design: Diagnostic, Non-Randomized, Open Label, Active Control, Parallel Assignment, Safety/Efficacy Study
Official Title: Can the Tsui Test be Used to Test the Placement of Epidural Catheters Following the Intrathecal Component of a Combined Spinal-Epidural Technique Used for Labor Analgesia?

Further study details as provided by Samuel Lunenfeld Research Institute, Mount Sinai Hospital:

Primary Outcome Measures:
  • change in current intensity required to produce a positive Tsui test after an intrathecal injection of bupivacaine [ Time Frame: Within 2 hours of catheter placement ] [ Designated as safety issue: No ]

Estimated Enrollment: 30
Study Start Date: February 2007
Estimated Study Completion Date: December 2008
Arms Assigned Interventions
1: Experimental
Tsui test for combined spinal-epidural
Procedure: Combined spinal epidural
0.7 mL of 0.25% isobaric bupivacaine and 15mcg of fentanyl
2: Experimental
Tsui test for intrathecal catheter
Procedure: Tsui Test for intrathecal catheter
0.7 mL of 0.25% isobaric bupivacaine and 15mcg of fentanyl

Detailed Description:

CSE is a widely used technique to provide analgesia in laboring patients. One of the disadvantages of the CSE is the delay in recognizing the mal-position of the epidural catheter, which is initially masked due to the effects of the spinal component. A mal-positioned epidural catheter can cause unnecessary maternal discomfort and dissatisfaction, as it requires a repeat procedure when the effect of spinal drug wears off. Additionally, in the event of emergency caesarean delivery, it would be advantageous to know that the catheter is properly placed. Low current electrical stimulation test, or the Tsui test, has been used successfully to confirm catheter location in the epidural space. Our objective in this study is to test the usefulness of the Tsui test to confirm the correct placement of the epidural catheter during CSE in laboring patients.

Two groups of patients will be studied. In the first group, after the placement of the epidural catheter and noting the electric current required to elicit a motor response (maximum current used is 20 mA), a spinal needle is introduced one inter vertebral space below the space where the epidural catheter is inserted and 0.7 ml of 0.25% isobaric bupivacaine and 15 mcg of fentanyl is injected intrathecally. The epidural catheter is then re-stimulated and the magnitude of the current required to reproduce the motor response is noted. The stimulation is repeated after 5, 10, and 15 minutes. The second group consists of patients with accidental dural puncture. An intrathecal catheter is inserted and the Tsui test performed (maximum current used is 2 mA), followed by the injection of same mixture of local anesthetic and opioid as used in the first group. The Tsui test is repeated at 5, 10 and 15 minute intervals, similar to what is done in the epidural group. After 15 minutes, patients in both groups are managed as per departmental policy. In both the groups, the Tsui test will be repeated after 120 minutes during the maintenance of labor analgesia to determine the change in intensity of current required to elicit the motor response.

  Eligibility

Ages Eligible for Study:   18 Years and older
Genders Eligible for Study:   Female
Accepts Healthy Volunteers:   Yes
Criteria

Inclusion Criteria:

  • American Society of Anesthesiologists (ASA) 1 or 2
  • Full term pregnancy
  • Patients requesting epidural in early labor

Exclusion Criteria:

  • Any contraindication to regional anesthesia
  • Allergy or hypersensitivity to lidocaine, bupivacaine, or fentanyl
  • Received prior sedatives or opioids
  • Abnormal vertebral anatomy
  • Neurological disorders with lumbar involvement
  • Patients with implanted electronic devices
  Contacts and Locations
Please refer to this study by its ClinicalTrials.gov identifier: NCT00464841

Locations
Canada, Ontario
Mount Sinai Hospital
Toronto, Ontario, Canada, M5G 1X5
Sponsors and Collaborators
Samuel Lunenfeld Research Institute, Mount Sinai Hospital
Investigators
Principal Investigator: Jose CA Carvalho, MD, PhD MOUNT SINAI HOSPITAL
  More Information

No publications provided

Responsible Party: Mount Sinai Hospital ( Dr. Jose Carvalho )
Study ID Numbers: 07-02, 06-0205-A
Study First Received: April 23, 2007
Last Updated: December 12, 2007
ClinicalTrials.gov Identifier: NCT00464841     History of Changes
Health Authority: Canada: Ethics Review Committee

Keywords provided by Samuel Lunenfeld Research Institute, Mount Sinai Hospital:
Tsui Test
Combined Spinal Epidural
Labour

Study placed in the following topic categories:
Signs and Symptoms
Fentanyl
Labor Pain
Neurologic Manifestations
Bupivacaine
Pain

Additional relevant MeSH terms:
Signs and Symptoms
Nervous System Diseases
Labor Pain
Neurologic Manifestations
Pain

ClinicalTrials.gov processed this record on May 06, 2009