BMBL Appendix A - Table1

Comparison of Biological Safety Cabinets

cdcgo.gif (1118 bytes) o_intern.gif (1760 bytes) Biosafety Documents BMBL Table of Contents

 

Type Face velocity (lfpm) Airflow Pattern Radionuclides/ Toxic Chemicals Biosafety Level(s) Product Protection
Class I* open front 75 In at front; rear and top through HEPA filter No 2,3 No
Class II Type A 75 70% recirculated through HEPA; exhaust through HEPA No 2,3 Yes
Type B1 100 30% recirculated through HEPA; exhaust via HEPA and hard ducted Yes (Low levels/volatility) 2,3 Yes
Type B2 100 No recirculation; total exhaust via HEPA and hard ducted Yes 2,3 Yes
Type B3 100 Same as IIA, but plena under negative pressure to room and exhaust air is ducted Yes 2,3 Yes
Class III NA Supply air inlets and exhaust through 2 HEPA filters Yes 3,4 Yes

* Glove panels may be added and will increase face velocity to 150 lfpm; gloves may be added with an inlet air pressure release that will allow work with chemicals/radionuclides.

 

 