Assay Guidance  //  NCGC Assay Guidance Criteria
>>  Assay Guidance
Assay Guidance Manual
NCGC Assay Guidance Criteria
NCGC High Throughput Screening Assay Guidance Criteria
Criteria Biochemical Cell-based
Plate Format* 96-well or higher density plate

NCGC: 1536-well format

Assay volume 2-6 ul

96-well or higher density plate

NCGC: 1536-well format

Assay volume 4-6 ul

Assay Steps =10 steps with 96-well plate.

Steps include, reagent additions, timed incubations, plate transfers to incubator, reading, etc.

=10 steps with 96-well plate.

Steps include, reagent additions, timed incubations, plate transfers to incubator, reading, etc.

Minimum time increments and maximum assay duration Minimum assay window is 5 min. (i.e., earliest time point after last reagent addition) <24 hr is ideal; max 48 hrs.

Minimum assay window is 5 min.

Reagent Addition Steps 4 maximum (4 unique reagents max; more if pre-mixed) 4 maximum (4 unique reagents including cells max; more if pre-mixed)
Reagent removal steps* No plate coating steps Aspiration steps*
Temperature Between RT and 37°C Between RT and 37°C
Demonstrated DMSO Tolerance* 0.5 – 1% DMSO 0.5 – 1% DMSO
Signal : Background Ratio =3-fold =3-fold
Day-to-Day variation of control (e.g., IC50, EC50) <3-fold <3-fold
Reagent stability @ final working concentration =8 hrs @ RT or on ice bath;

No on-line thawing

=8 hrs @ RT or on ice bath;

No on-line thawing

Validation run reagent supply 10 – 96-well plate equivalents 10 – 96-well plate equivalents
Protocol Complete detailed protocol. All steps, equipment used, all vendor & catalog # for reagents. Data from 96-well or high density plate tests. Complete detailed protocol. All steps, equipment used, all vendor & catalog # for reagents. Detailed cell culture procedure, passage # .Data from 96-well or high density plate tests.
Detectors PE ViewLux (Top reading only: FI, TRF, FP, Abs, Luminescence)

PE Envision (bottom reading FI, ALPHA)

Acumen Explorer (fluorescent laser cytometry)

PE ViewLux (Top reading only: FI, TRF, FP, Abs, Luminescence)

PE Envision (bottom reading FI, ALPHA)

Acumen Explorer (fluorescent laser cytometry)

Special For unique reagents either investigator prepares sufficient quantity for HTS or identifies a reliable 3rd party vendor. Cells must be certified micoplasma-free by direct culture assay and cell-DNA fluorochrome staining.

Table Notes:

Plate Formats: 96-well plates contain 8 rows x 12 columns with volumes ranging between 50-200 ul; 384-well plates contain 16 rows x 24 columns with volumes ranging between 30-50 ul; 1536-well plates contain 32 rows by 48 columns with volumes ranging between 2-8 ul. NCGC will convert assays to 1536-well from 96- or 384-well format.

Reagent removal steps: Are any step that requires the removal of material from the well of a microtiter plate. While such steps may be routine with 96-well plates they are not recommended on robotic systems using 1536-well plates.

Demonstrated DMSO Tolerance: Because all compounds screened are stored in ~100% DMSO and delivered as a 1 to 100 dilution to the assay the assay sensitivity to between 0.5 and 1% DMSO must be determined.