Zeiss LSM 510 Inverted Meta Specifications
This point scanning instrument is available for either fixed slides or live specimen with temperature control, perfusion and solution changes. Photoactivation (405 nm), FRAP and FRET are fully supported. A Meta detector allows for spectral imaging and online emission fingerprinting.
 
Microscope
  Zeiss Axiovert 200 M microscope (inverted)  
     
Objectives
  10x Zeiss plan-neofluar dry, 0.3 NA  
  20x Zeiss plan-apochromat dry, 0.75 NA  
  40x Zeiss plan-neofluar oil, 1.3 NA, DIC  
  63X Zeiss plan-apochromat oil, 1.4 NA, DIC  
  100x Zeiss alpha plan fluar oil, 1.45 NA, DIC  
     
Fluorescence Filters Cubes (visual inspection only)
  FITC / GFP: 450-490 nm excitation, 515-586 nm emission.  
  Rhodamin: 540-552 nm excitation, emission beyond 590 nm.  
  DAPI: 350-400 nm excitation, 417-477 nm emission.  
  Cy5: 610-650 nm excitation, 670-720 nm emission.  
     
Laser Lines (scanning only)
  405 nm: solid-state laser, 25 mW  
  458, 488 and 514 nm: Argon ion laser, 30 mW  
  543 nm: Helium-Neon laser, 1mW  
  633 nm: Helium-Neon laser, 15mW  
  AOTF switched: photoactivation, FRAP, FRET and uncagging fully supported  
     
Filter Blocks (scanning only, must be used in certain combinations)
  Excitation:  
  NT 80/20: beam splitter, 20% excitation reflected, 80% emission transmitted  
  HFT 405/488/543/633: dichroic 405, 488, 543 and 633 nm  
  HFT 405/488/543: dichroic 405, 488 and 543 nm  
  HFT 405/514: dichroic 405, 514 nm  
  HFT 458/514: dichroic 458, 514 nm  
  HFT 458/543: dichroic 458, 543 nm  
  HFT 458: dichroic 458 nm  
  HFT 488: dichroic 488 nm  
     
  Steering Meta vs PMT2/3:  
  NFT 490: long pass 490 nm  
  NFT 515: long pass 515 nm  
  NFT 545: long pass 545 nm  
  NFT 635 vis: long pass 635 nm  
  NFT KP 545: short pass 545 nm  
  Plate: glass plate correcting for misalignment, automatically inserted by software  
     
  Steering PMT 2 vs. PMT 3:  
  NFT 490: long pass 490 nm  
  NFT 515: long pass 515 nm  
  NFT 545: long pass 545 nm  
     
Emission Filters
  Meta detector  
  The Meta is a spectrometer spreading fluorescence emission among an array of detectors. It is, in itself, a configurable emission filter.  
     
  PMT 2  
  LP 420: long pass 420 nm  
  LP 475: long pass 475 nm  
  LP 505: long pass 505 nm  
  BP 420-480 IR: band pass 420 to 480 nm, IR-blocking  
  BP 470-500: band pass 470 to 500 nm  
  BP 475-525: band pass 475 to 525 nm  
  BP 505-530: band pass 505 to 530 nm  
     
  PMT 3  
  LP 505: long pass 505 nm  
  LP 560: long pass 560 nm  
  LP 650: long pass 650 nm  
  BP 500/20 IR: band pass 490 to 510 nm, IR-blocking  
  BP 505-530: band pass 505 to 530 nm  
  BP 505-550: band pass 505 to 550 nm  
  BP 530-600: band pass 530 to 600 nm  
  BP 560-615: band pass 560 to 615 nm  
     
Detectors
  Two photo-multiplier tubes fluorescence, 8 or 12 bit sampling (256 or 4096 gray levels, respectively)  
  One transmission detector for transmission / DIC imaging  
  One Meta detector for spectral imaging.. Up to 8 channels may be defined, each grouping any of the 32 detectors with a spectral width of 10 nm each (overlap between channels is not allowed). The meta detector is almost as sensitive as a PMT except at higher wavelength. However, its intrinsic noise is twice as high as a PMT.  
     
Computer & Software
  Fujitsu Siemens 2.4 GHz Xenon CPU running Windows 2000, 4 GB RAM, dual 114 GB storage, 2.4x DVD+/- recorder, dual 19" 1280 x 1024 flat panels  
  LSM 5 acquisition software with Physiology and Time Series modules  

Top of Page