Primary Outcome Measures:
- primary patency detectable by duplex ultrasound through the index lesion [ Time Frame: 12 months ] [ Designated as safety issue: Yes ]
Secondary Outcome Measures:
- technical success [ Time Frame: time of deployment ] [ Designated as safety issue: No ]
- procedural success [ Time Frame: up to removal of catheter sheath ] [ Designated as safety issue: No ]
- procedural complications [ Time Frame: up to removal of catheter sheath ] [ Designated as safety issue: Yes ]
- Ankle Brachial Index [ Time Frame: discharge and 12 months ] [ Designated as safety issue: Yes ]
- primary patency [ Time Frame: 6 months ] [ Designated as safety issue: Yes ]
- binary restenosis [ Time Frame: 6 and 12 months ] [ Designated as safety issue: Yes ]
- stent fractures [ Time Frame: 6 and 12 months ] [ Designated as safety issue: Yes ]
- target lesion revascularisation [ Time Frame: 6 and 12 months ] [ Designated as safety issue: Yes ]
- target vessel revascularisation [ Time Frame: 6 and 12 months ] [ Designated as safety issue: Yes ]
- adverse events [ Time Frame: baseline, discharge, 6 and 12 months post procedure ] [ Designated as safety issue: Yes ]
- clinical categorization of chronic limb ischemia by means of the Rutherford classification [ Time Frame: discharge, 6 and 12 months post procedure ] [ Designated as safety issue: Yes ]
This is a German multi-center prospective, randomized, two-arm study evaluating performance of the Cordis S.M.A.R.T.™ CONTROL ™ Nitinol Stent System as compared to the C.R. Bard® Luminexx™ 6F Vascular Stent. Patients will be randomized on a 1:1 basis. It is anticipated that a total of 200 patients will be entered into the study.
The study population will consist of approximately 200 symptomatic peripheral vascular disease patients with TASC C & D SFA lesions. The disease will consist of symptomatic, long de novo or restenotic lesions (> 70% stenosis) and occlusions (5 - 22 cm) on diagnostic imaging. The lesion must not extend into the distal SFA and the popliteal artery. At least one distal popliteal artery must be patent as well as one calf vessel. Reference vessel diameter must be >= 4.0 to <= 6.0 mm.
Trial participants will be randomized to the Cordis S.M.A.R.T.™ CONTROL™ Nitinol Stent System or to the Bard® Luminexx™ 6F Vascular Stent.
Patients will be followed for twelve months post-procedure. Study examinations will be done at screening, procedure time, discharge, six, and twelve months post procedure.
This study will be conducted over up to 15 investigational sites in Germany.