Guidance for Third Parties and FDA Staff
Third Party Review of Premarket Notifications
Appendix 3B: Review Memorandum Template and Instructions for
Assay Only Submissions
A. 510(k) Number:
Indicate the 510(k) tracking number assigned to the document.
B. Purpose for Submission:
Indicate the reason for submission of this 510(k), e.g., reagent formulation
changes, labeling changes, software changes.
C. Measurand:
Identify the specific analyte to be tested.
D. Type of Test:
State whether the test is either quantitative or qualitative. Indicate the
technology, e.g., ELISA, RT-PCR.
E. Applicant:
Provide name of the sponsor.
F. Proprietary and Established Names:
List all the proprietary and established names.
G. Regulatory Information:
- Regulation section:
Provide the relevant Code of Federal Regulations sections for the device,
e.g., 21 CFR section 866.XXX, Tumor Associated Antigen Immunological Test
System.
-
Classification:
State whether the device is class I, II, or III.
-
Product code:
Indicate the specific analyte code for the device.
-
Panel:
State the regulatory panel for the device: Chemistry (75), Hematology (81),
Immunology (82), Microbiology (83), Pathology (88), Toxicology (91).
H. Intended Use:
State the final version of the intended use statement, i.e., as it will appear
in the labeling of the device.
- Intended use(s):
State the specific intended use(s) for which the test will be used.
- Indication(s) for use:
State the specific indication(s) for which the test will be used.
- Special conditions for use statement(s):
Describe any special applications of the device (see 809.10) or specific
contraindications or indications for use not addressed in the Intended Use
Statement, if applicable, e.g., prescription use, OTC, waived, POC, home
use.
- Special instrument requirements:
Briefly describe any special instrument requirements for the device.
I. Device Description:
Provide a brief description of the characteristics of the device, that is,
the design, model, components, etc., (i.e., the make-up of the device).
J. Substantial Equivalence Information:
- Predicate device name(s):
In a list provide the name(s) of the legally marketed device(s) with the same
intended use to which substantial equivalence is (are) claimed.
- Predicate K number(s):
In a list (in the same order as in J. 1. above) provide the 510(k) submission
number(s) for the predicate device(s) for which substantial equivalence is
(are) claimed.
- Comparison with predicate:
Provide a side-by-side comparison of the technological and analytical characteristics,
for the new device and the predicate. For example, see table below:
Similarities
|
Item |
Device |
Predicate |
|
|
|
Differences |
Item |
Device |
Predicate |
|
|
|
K. Standard/Guidance Document Referenced (if applicable):
List any recognized standards or relevant guidance referenced. Describe modification(s)
to recognized standards (if any).
L. Test Principle:
Provide a description of the technology/ methodology utilized in the device.
Discuss the principles of the device methodology, and indicate whether it
is well established or new and unproven.
M. Performance Characteristics (if/when applicable):
The type of data to be provided in this section depends on the intended use,
technological characteristics of the new device, and on claims made by the
manufacturer.
- Analytical performance:
- Precision/Reproducibility:
- Study design: Describe test methods. Include information such as
sample types tested, number of samples or measurements, target concentrations.
Indicate n for the parameters varied (e.g., observations, days, runs,
operators, sites, lots, instruments, calibration). Describe statistical
methods.
- Results/Acceptance criteria: Include percent CV /Standard deviation
for total and within-run precision (and other parameters, e.g., between-site,
between-instrument, where applicable).
- Linearity/assay reportable range:
- Study design: Describe test methods. Include information such as
sample types tested (e.g., unaltered patient sample, spiked or diluted
patient sample, spiked patient pools, control material), target concentrations
or range, number of samples, calculations/statistical methods. Include
description of test for high dose hook effect where appropriate.
- Results/Acceptance criteria: Identify the linear range and acceptable
or maximum difference from linearity within the range.
- Traceability, Stability, Expected values (controls, calibrators,
or methods):
Where applicable, briefly summarize the information about traceability
of calibrators and method. Include, for example, methods and acceptance
criteria for stability studies, traceability to reference material and
description of value assignment, validation, and acceptance criteria.
- Detection limit:
- Study Design: Describe the test methods. Include information such
as sample types tested, target concentrations/range, number of measurements,
statistical methods. Describe which aspects of sensitivity were measured,
for example: Limit of blank: usually mean of blank +2 to 3 SD, Limit
of detection: lowest level detected > 95% of the time, Functional
lower limit: Lowest analyte concentration where % CV is acceptable.
- Results/Acceptance criteria: Describe the sensitivity level for
each measure of sensitivity assessed.
- Analytical specificity:
- Study design: Describe test methods. Include information such as
sample types tested (e.g., samples spiked with interferent, samples
naturally high in interferent etc.), concentrations or intervals of
potential interferents or cross-reactants tested, and the concentration
of intended use analyte (if present) in the sample. Define the comparator/control
samples used (e.g. spiked samples containing analyte without interferent),
statistical methods, definitions/calculations of cross-reactivity
or recovery.
- Results/Acceptance criteria: Describe any interference observed
(e.g., recoveries for interference levels tested). Describe any trends
(e.g., "high levels of metabolite x may cause depressed results").
- Assay cut-off:
Describe whether the cutoff was established previously in the literature
(and whether it was established with this specific device) or in this
submission. If the latter, describe the study design, including methods
for determining the cutoff, such as, including the population(s) studied
(demographics/ selection/ inclusion and exclusion criteria, number of
individuals included), reference method, or method of clinical diagnosis
and statistical methods used to generate results. Define “gray zone,”
if applicable.
Results: sensitivity and specificity for chosen cutoffs values.
- Comparison studies:
- Method comparison with predicate device:
- Study design: Describe test methods. Identify the comparator(s)
(e.g., predicate device, reference method). Describe the sample type(s)
(e.g., unaltered patient specimens, spiked or diluted patient specimens,
spiked patient pools, and control material), matrix, number of samples,
sample range. When appropriate include information such as number/types
of sites, sample selection methods, inclusion/exclusion criteria,
overall demographic description of patients represented by the samples
(e.g., age, gender, race, how/whether samples represent the intended
use population), number of individuals represented. Describe statistical
methods used to generate results (e.g., regression methods, data exclusion,
number of observations represented by each data point).
- Results: Describe overall results and/or results for specific sites
and patient groups, as appropriate. For quantitative tests, include
information such as slope and intercept (with confidence intervals),
correlation coefficient, measure of scatter around the regression
line (e.g., sy..x), measure of bias at medical decision levels. In
some cases, it may be helpful to include a graph (x-y graph or bias
plot). For qualitative or semi-quantitative tests, include percent
agreement with comparator for positive/negative samples, confidence
intervals.
- Matrix comparison:
- Study design: For each matrix in the intended use, describe the
method for comparison or determination of accuracy. Include information
such as sample types tested, number of samples, sample range or target
concentrations tested, calculations/statistical methods.
- Results/Acceptance criteria: Describe the accuracy of the new matrix
or results of the matrix comparison.
- Clinical studies:
- Clinical Sensitivity:
- Study design: Explain how “Clinical Truth” determined.
Include information such as method(s) of diagnosis, specimen types,
patient demographics, inclusion/exclusion criteria, number of specimens
collected, (any discarded), number of individuals represented, number/types
of sites, statistical methods. When appropriate include collection
methods, other pertinent information.
- Results: Sensitivity, specificity, positive and negative predictive
value with confidence intervals. Describe results for overall study
and/or for specific patient groups or sites, as appropriate.
- Clinical specificity:
See Section 3.a.i.
- Other clinical supportive data (when a. and b. are not applicable):
Provide other clinical supporting data such as clinical monitoring data,
etc.
- Clinical cut-off:
Describe whether the cutoff was established previously in the literature
(and whether it was established with this specific device) or in this submission.
If the latter, describe the study design including methods for determining
the cutoff including the population(s) studied (demographics/ selection/
inclusion and exclusion criteria, number of individuals included), reference
method, or method of clinical diagnosis and statistical methods used to
generate results. Define “gray zone,” if applicable.
Results: sensitivity and specificity for chosen cutoffs values.
-
Expected values/Reference range:
Describe whether the expected range was established in the literature (and
whether it was established with this specific device), or in this submission.
If the latter, describe the methods for determining the expected range including
the population(s) studied (demographics/ selection/ inclusion and exclusion
criteria, number of individuals), reference method or method of clinical
diagnosis, statistical methods.
Results: Describe and define (e.g., 95th percentile) normal range values
for relevant populations.
N. Proposed Labeling
Indicate whether the labeling is sufficient and it satisfies the requirements
of 21 CFR section 809.10.
O. Conclusion:
State the rationale (i.e., justification) for substantial equivalence (SE)
decision.
P. Other Supportive Information:
Describe any additional supporting information provided by the applicant
that was evaluated as part of the review.
Q. Administrative Information:
- Applicant contact information:
Record the mailing address and contact information of the applicant, and identify
the contact person responsible for the submission.
- Review documentation:
Record all applicable information for the administrative record (e.g., the
submission chronology, telephone conversations, faxes, e-mails, meeting minutes,
discussions regarding labeling issues, etc.). If, necessary, attach copies
of these documents to the completed review.
REVIEW MEMORANDUM
ASSAY ONLY TEMPLATE
A. 510(k) Number:
B. Purpose for Submission:
C. Measurand:
D. Type of Test:
E. Applicant:
F. Proprietary and Established Names:
G. Regulatory Information:
- Regulation section:
- Classification:
- Product code:
- Panel:
H. Intended Use:
- Intended use(s):
- Indication(s) for use:
- Special conditions for use statement(s):
- Special instrument requirements:
I. Device Description:
J. Substantial Equivalence Information:
- Predicate device name(s):
- Predicate 510(k) number(s):
- Comparison with predicate:
Similarities |
Item |
Device |
Predicate |
|
|
|
Differences |
Item |
Device |
Predicate |
|
|
|
K. Standard/Guidance Document Referenced (if applicable):
L. Test Principle:
M. Performance Characteristics (if/when applicable):
- Analytical performance:
- Precision/Reproducibility:
- Linearity/assay reportable range:
- Traceability, Stability, Expected values (controls, calibrators,
or methods):
- Detection limit:
- Analytical specificity:
- Assay cut-off:
- Comparison studies:
- Method comparison with predicate device:
- Matrix comparison:
- Clinical studies:
- Clinical Sensitivity:
- Clinical specificity:
- Other clinical supportive data (when a. and b. are not applicable):
- Clinical cut-off:
- Expected values/Reference range:
N. Proposed Labeling:
O. Conclusion:
P. Other Supportive Information:
Q. Administrative Information:
- Applicant Contact Information:
- Name of applicant:
- Mailing address:
- Phone #:
- Fax #:
- E-mail address (optional):
- Contact:
- Review Documentation:
Updated September 30, 2004
CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page
Center for Devices and Radiological Health / CDRH