![]() ![]() ![]() ![]() |
FDA Home Page | CDRH Home Page | Search | A-Z Index | ![]() |
|
![]() |
![]() | ||||
|
||||
|
||||
Introduction
General Information
Table of Contents (recommended)
510(k) Screening Checklist (recommended)
Statement of Indications for Use
510(k) Statement or Summary
Truthful and Accuracy Statement
Proposed Labeling
Specifications
Substantial Equivalence Comparison
Performance
Additional requirements, as appropriate
The 510(k) regulation is found in 21 CFR 807 Subpart E and includes information required in a 510(k). The 510(k) is not a form. The information should be provided in an organized, tabulated document. The 510(k) should provide sufficient detail for FDA to be able to determine that the device is substantially equivalent (SE) to another similar legally marketed device(s). Some sections will contain only one page; others may contain 50 or more pages. The average 510(k) is about 35 pages; others may run to 100 or more depending on the complexity of the device. For any device, the 510(k) is formatted essentially the same way and contains the same basic information (required elements).
This section is a general guide for all 510(k)'s. Please review the appropriate information for the type of 510(k) you will submit: Traditional, Special, or Abbreviated 510(k)s.
The 510(k) submitter should prepare and submit a complete application in order to obtain marketing clearance. To facilitate FDA review of the data, analysis, and conclusions in the application, the manufacturer should check for the:
A description of the tests and the results obtained are essential. Reasonable and sufficient details of all test procedures and results should be submitted to FDA. The following suggestions will help assure that your application is complete.
Important Note: In order to address the required elements, you will need the following information to assure that your 510(k) is complete. If you do not have this information, you should obtain it prior to proceeding.
- Classification of your device
- Predicate device(s)
- Final draft labeling
- Specifications including engineering drawings, photos, etc.
- Performance data such as bench, animal, or clinical testing (if applicable)
- Sterilization information (if applicable)
- Guidance document(s) specific to your device type, if it exists
It is not a requirement that you organize your 510(k) into the following labeled sections, you may make modifications in order to meet the specific needs of your device. The following is a brief discussion of the required elements.
Medical Device User Fee Cover Sheet (Form FDA 3601)
The first page of a 510(k) should be a printed copy of the Medical Device User Fee Cover Sheet (Form FDA 3601). The Medical Device User Fee Cover Sheet and instructions are available at http://www.fda.gov/oc/mdufma/coversheet.html.
CDRH Premarket Review Submission Cover Sheet
A cover letter and/or the CDRH Premarket Review Submission Cover Sheet should follow the User Fee Cover Sheet. If you choose NOT to use the CDRH Premarket Review Submission Cover Sheet, the cover letter should include all the elements contained in the coversheet. The use of the 510(k) coversheet may help expedite the processing of your 510(k).
Cover Letter
You may use a cover letter to provide the information covered on the CDRH Premarket Review Submission Cover Sheet or provide this information elsewhere in the 510(k). The information provided should include the following.
Prepare a draft Table of Contents and update it as you prepare your 510(k).
The table of contents should list each required item with page numbers, including
a list of attachments and appendices. It is usually easier to number pages by
section, e.g. 1.1, 1.2 (or use dashes, 1-1), 2.2, 2.3, etc. If additional pages
need to be inserted into a section, it is easier to renumber that section than
having to renumber all the pages that follow the inserted material if
you used sequential numbering.
The 510(k) Screening Checklist is should be used to assure that the 510(k) is complete. It is helpful to attach the 510(k) screening checklist following the Table of Contents. It should include page numbers where each of the elements in the 510(k) can be found. This will allow the FDA reviewer to easily find each required element. Secondly, by writing page numbers on the checklist, the preparer of the 510(k) can be certain the 510(k) is complete. The 510(k) may not be accepted for review if any of the required elements are not provided.
Screening Checklist for all Premarket Notification 510(k) Submissions [PDF]
Prepare a Statement of Indications for Use as a separate page.
We recommend that you use the Indications for Use format. Use the ODE recommended format for submissions to ODE and the OIVD recommended format for submissions to OIVD. The statement should include specific indications, clinical settings,
define the target population, anatomical sites, etc. This statement must be
consistent with your labeling, advertising and instructions for use. Once the
review is complete, FDA will include the Indications for Use Statement with
the Substantial Equivalence (SE) letter to the applicant and make it available
to the public on the Internet.
Prepare either a 510(k) Summary or a 510(k) Statement. The Summary or Statement is required for all 510(k)s whether the device is Class I, II, or III. A 510(k) Summary is a summary of information upon which you based your claim of substantial equivalence. The 510(k) Statement is a certification that the 510(k) owner will provide safety and effectiveness information supporting the FDA finding of substantial equivalence to ANY person within 30 days of a written request.
A 510(k) Summary or 510(k) Statement must be included in your 510(k) submission in order for FDA to begin its scientific review of the 510(k) submission. The choice between the 510(k) Summary and 510(k) Statement should be made before the 510(k) is submitted. You may elect to change your choice between the summary or statement BEFORE the substantial equivalence determination is reached. However, after this determination is made, you cannot change your choice of a 510(k) Summary or 510(k) Statement.
If you choose to meet the conditions for a 510(k) summary (21 CFR 807.92), it must be in sufficient detail to provide an understanding of the basis for a determination of substantial equivalence. As required by section 807.92(a), FDA will accept summaries or amended summaries until FDA issues a determination regarding substantial equivalence.
The 510(k) summary must contain the information described below. Please make a copy of the following to use as a checklist and check off each item to make sure your summary is adequate and complete.
[ ] |
The summary should be in a separate section of the submission. It should begin on a new page and end on a page not shared with any other part of the 510(k) submission. It is clearly identified as "510(k) Summary" as required by section 807.92(c). |
[ ] |
The summary contains on the first page, preferably on your letterhead paper, the 510(k) owner's name, address, phone and fax numbers, name of contact person, and date the summary was prepared [807.92(a)(1)]. |
[ ] |
The summary includes the name of the device, including the
trade or proprietary name, if applicable, the common or usual name, and
the classification name, if known [807.92(a)(2)].
Example:
|
[ ] |
The summary identifies the legally marketed device to which your firm is claiming equivalence [807.92(a)(3)]. |
[ ] |
The summary includes a description of the device such as might be found in the labeling or promotional material for the device, including an explanation of how the device functions, the scientific concepts that form the basis for the device, and the significant physical and performance characteristics of the device, such as device design, material used, and physical properties; [807.92(a)(4)]. |
[ ] |
The summary provides the intended use of the device including a general description of the diseases or conditions that the device will diagnose, treat, prevent, cure, or mitigate, including a description, where appropriate, of the patient population for which the device is intended. If the indication statements are different from those of the predicate device, the 510(k) summary shall contain an explanation as to why the differences are not critical to the intended therapeutic, diagnostic, prosthetic, or surgical use of the device, and why the differences do not affect the safety and effectiveness of the device when used as labeled [807.92(a)(5)]. |
[ ] |
The 510(k) summary contains a summary of the technological characteristics of your device compared to the predicate device. If the device has the same technological characteristics (i.e., design, material, chemical composition, energy source) as the predicate device, a summary of the technological characteristics of the new device in comparison to those of the predicate device should be included. If your device has different technological characteristics from the predicate device, provide a summary of how the technological characteristics of your device compare to the predicate device. [807.92(a)(6)] |
[ ] |
If the determination of substantial equivalence is also based on an assessment of non-clinical performance data, the summary includes a brief discussion of the nonclinical tests submitted, referenced, or relied on in the premarket notification submission for a determination of substantial equivalence and how their results support a determination of substantial equivalence [807.92(b)(1)]. |
[ ] |
If the determination of substantial equivalence is also based on an assessment of clinical performance data, the summary includes a brief discussion of clinical tests submitted, referenced, or relied on in the premarket notification submission for a determination of substantial equivalence and how their results support a determination of substantial equivalence. This discussion shall include, where applicable, a description of the subjects upon whom the device was tested, a discussion of the safety or effectiveness data obtained from the testing, with specific reference to adverse effects and complications, and any other information from the clinical testing relevant to a determination of substantial equivalence [807.92(b)(2)]. Please note: Clinical data is not needed for most devices cleared by the 510(k) process. |
[ ] |
The summary includes the conclusions drawn from the nonclinical and clinical tests (discussed above) that demonstrate that the device is as safe, as effective, and performs as well as or better than the predicate device . 807.92(b)(3) |
[ ] |
The summary includes any other information reasonably deemed necessary by FDA. Such requests will be made directly to the applicant by FDA or the requirements will be published in guidance documents. Additional information requested by FDA during review of the 510(k) may include additional safety and effectiveness information which may necessitate an update of your summary if requested by FDA. 807.92(d) |
[ ] |
Please make sure you have included all of the information
listed above and verify that the following criteria have been met.
|
FDA will place the summary on the Internet 30 days following the substantial equivalent decision.
510(k) Statement
If you choose to submit a 510(k) Statement, the regulation requires the specific statement as provided in 21 CFR 807.93. The statement should be clearly identified as "510(k) Statement" and must be signed by the certifier, not a consultant to the 510(k) submitter.
For a new submission, leave the space for the 510(k) number blank. You will receive your 510(k) number in your 510(k) acknowledgment letter. The 510(k) document control number begins with the letter K followed by 6 digits.
If you elect to prepare a signed 510(k) Statement, anyone may request a copy of the 510(k) [with patient identifiers, trade secret and confidential information deleted] from the applicant of record. These written requests must be filled within 30 days. 510(k) owner's may not charge requesters for compiling and disseminating this data. Noncompliance with the 510(k) statement will be deemed a prohibited act under section 301(p) of the FD&C Act and FDA may choose to use its enforcement powers to obtain compliance.
All 510(k) submitters must include a statement certifying that all information submitted in the 510(k) is truthful and accurate and that no material fact has been omitted. The statement may be included in the 510(k) Cover Letter or may be on a separate page identified in the table of contents. If the CDRH Premarket Review Submission Cover Sheet is used, the statement should attached as a separate page. Truthful and Accurate Statement
Prepare a labeling section to include copies of all proposed labels, labeling, package inserts, service manuals, instructions for use, advertising and/or promotional materials. The directions for use should include a specific intended use statement and any warnings, contraindications, or limitations. The material does not have to be glossy, finished labeling or promotional material, draft is sufficient. However, the labeling you submit should be final draft. Copies of labeling for the predicate device(s) is recommended. Labeling guidance is provided below:
Labeling
Blue Book Memorandum #G91-1: Device Labeling Guidance
This section should include both a narrative description
of the device and a physical or technical description.
The narrative description of the "new" device
should include the indications for use, principles of operation, power source,
composition and other information necessary to understand the device. If the
510(k) is for an accessory or component sold to an end-user, describe a typical
device with which the accessory or component will be used. List all variations
of the "new" device which you intend to market.
The physical description of the "new" device
may include labeled diagrams, photographs or pictures, engineering drawings,
schematics, etc. These may include all internal and external, assembled and
unassembled, interchangeable, etc., parts of the device and should address their
name and function. In addition, the description should include the length, width,
height, diameter, weight, etc., of the device and identify any parts which are
intended for single use.
Device specific guidance documents, if available, usually provide extensive information on the level of detail which should be included in the specifications list.
The device specifications are the basis for the comparison of features between the new and the legally marketed device to which compared (predicate device). Substantial equivalence is to be established with respect, but not limited to, intended use, design, energy used/delivered, materials, performance, safety, effectiveness, labeling and other applicable characteristics, such as sterility. You should include a comparison table AND discussion of the similarities and differences of your device compared to one or more predicate devices to which you are claiming equivalency.
The equivalence information should be provided in a clear and comprehensive format. A chart, table or other side-by-side comparision is a systematic way to compare the devices. Side-by-side comparisions, wherever possible, are desirable. For some devices a simple table of comparision which lists characteristics will be sufficient to establish equivalence. Often, information is necessary to resolve questions of safety or effectiveness, especially where differences in technologies exists between the predicate and the new device. It must be shown that technological differences do not adversely affect safety and effectiveness. Supporting information can be obtained from bench testing, animal studies or clinical studies (information gathered from medical literature) or clinical trials.
State whether the legally marketed device for comparision is a preamendments device, or a device which has been granted marketing clearance by FDA following the submission of a 510(k). Provide the 510(k) document control number (i.e., K followed by 6 digits) for the device to which you are claiming equivalency, if known.
The comparison table should identify relevant similarities and differences in areas such as:
|
|
The discussion of the similarities and differences should elaborate on the similarities identified in the table of comparisons and justify the differences with supporting rationale and/or data. It is recommended to submit labeling for the device to which you are claiming equivalency.
Most 510(k)s will include some type of performance data. The extent of performance data will depend on the complexity of the device and its intended use and indications. Performance data should be provided to help demonstrate SE of your device to one or more legally marketed devices (predicate device). The data may include test results from engineering, bench, design verification, human factors, and animal testing, and clinical studies and clinical trials. Tests should be conducted on all sizes and models of the device in a manner as similar as possible to how the device will be used. The results of testing and methodology / parameters used for testing should be included. Information on the extent of performance testing as well as the extent of specification documentation can be found in the product specific guidance documents, if one has been prepared for your device by FDA. Search the Guidance Document database to determine if guidance documents are available for your type of device.
Additional information may be required under certain conditions, such as if your device contains software or a color additive, is labeled sterile, or emits electronic radiation. See "Special Considerations" under Device Advice Premarket Notification 510(k) for additional guidance.
Updated July 10, 2008
CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page
Center for Devices and Radiological Health / CDRH