Skip NavigationFDA Logo links to FDA home pageCenter for Devices and Radiological Health, U.S. Food and Drug AdministrationHHS Logo links to Department of Health and Human Services website

FDA Home Page | CDRH Home Page | Search | CDRH A-Z Index | Contact CDRH U.S. Food and Drug Administration Center for Devices and Radiological Health
horizonal rule

(See Related Information)

(Adobe PDF Printer Friendly Version)

Sterilized Convenience Kits for Clinical and Surgical Use;
Final Guidance for Industry

Document issued on January 7, 2002

This document does not supersede any previous document.

CDRH Logo

U.S. Department of Health and Human Services
Food and Drug Administration
Center for Devices and Radiological Health

Office of Compliance

 

 

Preface

Public Comment

Comments and suggestions may be submitted at any time for Agency consideration to Dockets Management Branch, Division of Management Systems and Policy, Office of Human Resources and Management Services, Food and Drug Administration, 5630 Fishers Lane, Room 1061, (HFA-305), Rockville, MD, 20852. When submitting comments, please refer to the exact title of this guidance document. Comments may not be acted upon by the Agency until the document is next revised or updated.

For questions regarding the use or interpretation of this guidance for medical devices, contact Wally A. Pellerite in the Office of Compliance, Center for Devices and Radiological Health at (301) 594-4692 or email WAP@cdrh.fda.gov.

For questions regarding prescription drug products, call the Division of Prescription Drug Compliance and Surveillance, in the Center for Drug Evaluation and Research (CDER) at (301) 594-0101. For questions regarding over the counter drugs, call the Division of Labeling & Nonprescription Drug Compliance, in CDER, at (301) 594-0063. For questions regarding manufacturing and quality control, call the Division of Manufacturing and Product Quality at (301) 594-0098.

Additional Copies

Additional copies are available from the Internet at: http://www.fda.gov/cdrh/comp/guidance/1390.pdf, or CDRH Facts-On-Demand. In order to receive this document via your fax machine, call the CDRH Facts-On-Demand system at 800-899-0381 or 301-827-0111 from a touch-tone telephone. Press 1 to enter the system. At the second voice prompt, press 1 to order a document. Enter the document number 1390 followed by the pound sign (#). Follow the remaining voice prompts to complete your request.

 


Sterilized Convenience Kits for Clinical and Surgical Use


This document is intended to provide guidance. It represents the Agency's current thinking on this topic. It does not create or confer any rights for or on any person and does not operate to bind the Food and Drug Administration (FDA) or the public. An alternative approach may be used if such approach satisfies the requirements of the applicable statute and regulations.

 


The Least Burdensome Approach

We believe we should consider the least burdensome approach in all areas of medical device regulation. This guidance reflects our careful review of the relevant scientific and legal requirements and what we believe is the least burdensome way for you to comply with those requirements. However, if you believe that the information requested in the guidance is not relevant to the decision-making process or that an alternative approach would be less burdensome, please contact us so we can consider your point of view. You may send your written comments to the contact person listed in the preface to this guidance or to the CDRH Ombudsman. Comprehensive information on CDRH's Ombudsman, including ways to contact him, can be found on the Internet at http://www.fda.gov/cdrh/resolvingdisputes/ombudsman.html.

Medical Device Kit Assembler / Manufacturer1

The Center for Devices and Radiological Health (CDRH) continues to have concerns about finished device components that are assembled into clinical or surgical convenience kits and the kit is then sterilized. In some cases, the finished device is actually being sterilized for a second time. Some components may be adversely affected by further processing in the sterilization process. For the purposes of this guidance, a finished device component in a convenience kit is a device in finished form held for sale to an end user that is suitable for use or capable of functioning, whether or not it is packaged, labeled or sterilized. Please ensure that your sterilization process does not adversely affect the finished components in your kits.


1For concerns regarding drug products, please review the last paragraph of this document and contact the appropriate office in the Center for Drug Evaluation and Research.

There are five areas that, at a minimum, you should cover to ensure that the sterilization process does not adversely affect the finished components in your convenience kits. Those areas include the following:

A sterilization process may pose the following issues that you need to address:

  1. Effects of the sterilization process -
  1. Labeling -
  1. Premarket Notification -
  1. General Controls -

In summary, you should address the following issues related to the manufacture of your convenience kits:

  1. How does your sterilization process affect the various device components in your kit?
  2. What are the sterilant residuals of your sterilization process?
  3. Does the labeling of the kit include validated storage instructions?; Is the expiration dating for the kits appropriate?
  4. Is a 510(k) or PMA submission needed?
  5. Do you meet the general manufacturing, labeling, and postmarket reporting requirements, i.e., adverse events and recalls (corrections and removals for devices (21 Code of Federal Regulations Parts 820, 803 and 806.))?

For questions regarding medical devices incorporated into convenience kits and their labeling requirements, please contact Wally A. Pellerite in the Center for Devices and Radiological Health, Office of Compliance at (301) 594-4692.

Drug Questions:

Contact the Center for Drug Evaluation and Research (CDER) for questions about drug products that concern labeling, new drug issues, listing and registration requirements, CGMPs, or other regulatory requirements for drugs used in convenience kits.

Prescription drugs (Rx) questions or issues should be directed to the Division of Prescription Drug Compliance and Surveillance, HFD-330 at (301) 594-0101.

Over the counter drug products questions or issues should be directed to the Division of Labeling & Nonprescription Drug Compliance, HFD-310 at (301) 594-0063.

Current Good Manufacturing Practice (CGMP) requirement issues or questions applicable to drug products in convenience kits, should be directed to the Division of Manufacturing and Product Quality at (301) 594-0093.

You may also access CDER's Internet site at http://www.fda.gov/cder/directories/reference_guide.htm for information on the offices and/or review divisions that could be contacted about the specific drug products intended for use in the sterilized kit.

Sincerely yours,


Larry D. Spears
Acting Director
Office of Compliance
Center for Devices and Radiological Health

Updated 1/11/2002

horizonal rule

CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page