U.S. Food and Drug Administration - CDRH Mobile
Skip NavigationFDA Logo links to FDA home pageCenter for Devices and Radiological Health, U.S. Food and Drug AdministrationHHS Logo links to Department of Health and Human Services website
FDA Home Page | CDRH Home Page | Search | A-Z Index U.S. Food and Drug AdministrationCenter for Devices and Radiological Health Questions?
horizonal rule

The Quality System Regulation


Return to Quality System Manual Table of Contents
1 THE QUALITY SYSTEM REGULATION

INTRODUCTION
FLEXIBILITY OF THE GMP
MANUAL CONTENTS
GMP APPLICATIONS AND EXEMPTIONS
     Exemptions
     Component Manufacturers
     Remanufacturers
     Custom Device Manufacturers
     Contract Manufacturers
     Contract Testing
     Repackagers, Relabelers, and Specification Developers
     Initial Distributors of Imported Devices


INTRODUCTION

The current Good Manufacturing Practices (GMP) requirements set forth in the Quality System (QS) regulation are promulgated under section 520 of the Food, Drug and Cosmetic (FD&C) Act. They require that domestic or foreign manufacturers have a quality system for the design and production of medical devices intended for commercial distribution in the United States. The regulation requires that various specifications and controls be established for devices; that devices be designed under a quality system to meet these specifications; that devices be manufactured under a quality system; that finished devices meet these specifications; that devices be correctly installed, checked and serviced; that quality data be analyzed to identify and correct quality problems; and that complaints be processed. Thus, the QS regulation helps assure that medical devices are safe and effective for their intended use. The Food and Drug Administration (FDA) monitors device problem data and inspects the operations and records of device developers and manufacturers to determine compliance with the GMP requirements in the QS regulation.

The QS regulation is in Part 820 of Title 21 of the Code of Federal Regulations (CFR). This regulation covers quality management and organization, device design, buildings, equipment, purchase and handling of components, production and process controls, packaging and labeling control, device evaluation, distribution, installation, complaint handling, servicing, and records. The preamble describes the public comments received during the development of the QS regulation and describes the FDA Commissioner's resolution of the comments. Thus, the preamble contains valuable insight into the meaning and intent of the QS regulation.

The QS regulation is reprinted in the appendix of this manual.

FLEXIBILITY OF THE GMP

Manufacturers should use good judgment when developing their quality system and apply those sections of the QS regulation that are applicable to their specific products and operations. Section 820.5 of the QS regulation requires that, "Each manufacturer shall establish and maintain a quality system that is appropriate for the specific device(s) designed or manufactured, and that meets the requirements of this part." The word "appropriate" means that the rule is a flexible regulation. However, if manufacturers decide to not implement certain GMP requirements which are qualified by the term "where appropriate," they should document their justification for nonimplementation. The justification should show that not implementing a requirement is not reasonably expected to result in product that does not meet specifications or failureto carry out any necessary corrective action [820.1(a)(30]. Operating within this flexibility, it is the responsibility of each manufacturer to establish requirements for each type or family of devices that will result in devices that are safe and effective, and to establish methods and procedures to design, produce, and distribute devices that meet the quality system requirements. FDA has identified in the QS regulation the essential elements that a quality system shall embody for design, production and distribution, without prescribing specific ways to establish these elements. Because the QS regulation covers a broad spectrum of devices and production processes, it allows some leeway in the details of quality system elements. It is left to manufacturers to determine the necessity for, or extent of some quality elements and to develop and implement specific procedures tailored to their particular processes and devices. For example, if it is impossible to mix up labels at a manufacturer because there is only one label or one product, then there is no necessity for the manufacturer to comply with all of the GMP requirements under device labeling.

The medical device QS regulation requires an "umbrella" quality system intended to cover the design, production, and distribution of all medical devices from simple surgical hand tools to very complex computerized axial tomography (CAT) scanners. It is not practical for a regulation to specify details of quality system elements for such a wide range of products. Rather, the QS regulation specifies general objectives such as use of trained employees, design reviews, design validation, calibrated equipment, process controls, etc., rather than methods, because a specific method would not be appropriate to all operations.

In most cases, it is left to the manufacturer to determine the best methods to attain quality objectives. In some cases, however, the QS regulation does specify the particular type of method to be used, such as written procedures or written instructions. This does not mean, however, that manufacturers cannot vary from the method specified if the intent of the GMP requirement can be met by another method such as using an engineering drawing plus a model device as manufacturing instructions. Written procedures are not restricted to paper copies. Written procedures may be filed and distributed by automated data processing equipment. This flexibility is allowed by section 820.180.

Typically, large manufacturers will have a quality system that exceeds the medical device QS regulation. Small manufacturers will typically have a proportionally simpler system. FDA recognizes: that a small manufacturer may not need the same amount of documentation that a large manufacturer does in order to achieve a state-of-control; and, that some of records maintained to fulfill the GMP requirements for written procedures may not be as long and complex for a small manufacturer.

After a manufacturer establishes a quality system, it should be maintained. Each manufacturer should assure that with growth and process or product changes their quality system is still adequate. This assurance is obtained through change control, day­to­day observance of operations, and by periodic audits of the quality system. The auditor should first identify the elements of the company's quality system. Next the audit should determine how well each element is functioning, and then determine its adequacy with respect to the intent of the device GMP requirements and meeting the company's quality claims.

MANUAL CONTENTS

To aid auditors, QA managers, and others, this manual provides guidance in the interpretation of the GMP requirements, and demonstrates the flexibility of the QS regulation in its application to diverse devices, manufacturing processes, and manufacturers. In the absence of guidance from FDA, manufacturers may rely on industry, national, and international consensus standards or guidances to meet GMP requirements.

This manual was also developed to aid manufacturers in completing, maintaining, or expanding their quality system. Contents include educational materials, aids, and examples of how to implement elements of a quality system, together with detailed examples of procedures, control forms, and associated data. The examples of typical procedures, drawings, and forms found in this manual were derived from quality systems in the device industry. These materials are not meant to describe universally applicable elements of a quality system that can be used unchanged by any manufacturer. Of course, a form or aid as presented in this manual may be suitable for direct use for a specific device and operation; however, in general, manufacturers will need to use care in adopting and modifying a selected form or procedure to meet the specific quality system needs of their devices and operations.

This manual is arranged as if the reader were starting a new business. That is, as if an entrepreneur were sequentially:

1. obtaining information on GMP requirements;
2. determining the appropriate quality system needed to control the design, production and distribution of the proposed device;
3. designing products and processes;
4. training employees;
5. acquiring adequate facilities;
6. purchasing and installing processing equipment;
7. drafting the device master record;
8. noting how to change the device master records;
9. procuring components and materials;
10. producing devices;
11. labeling devices;
12. evaluating finished devices;
13. packaging devices;
14. distributing devices;
15. processing complaints and analyzing service and repair data;
16. servicing devices;
17. auditing and correcting deficiencies in the quality system; and,
18. preparing for an FDA inspection.

If manufacturers perform these activities as required by the QS regulation and as expounded in this manual, they should be prepared for a GMP inspection of their operations by an FDA investigator.

Manufacturers and importers of medical devices shall also comply with the Medical Device Reporting (MDR) regulation, 21 CFR Part 803, which requires that serious complaints be reported to FDA. MDR is related to the GMP complaint and failure investigation requirements, which are covered in Chapter 15. If manufacturers comply with the QS regulation and guidance in this manual and in other sources, there is a high probability that they will reduce the frequency of reportable events.

GMP APPLICATIONS AND EXEMPTIONS

The QS regulation applies to finished devices intended to be commercially distributed for human use unless there is an approved exemption in effect. GMP exemptions are codified in the classification regulations 21 CFR 862 to 892. The exemption of most Class I devices from design controls is in section 820.30(a).

Certain components such as blood tubing and major diagnostic x­ray components are considered by FDA to be finished devices because they are accessories to finished devices. The manufacturer of such accessories is subject to the QS regulation when the accessory device is labeled and sold separately from the primary device for a health­related purpose to a hospital, physician, or other user.

The designation of a device as a "custom" or "customized" device does not confer a GMP exemption.

Contract manufacturers and specification developers shall comply with the sections of the QS regulation that apply to the functions they perform.

Contract test laboratories are considered an extension of a manufacturer's quality system and presently are not routinely scheduled for GMP inspections. The finished device manufacturer shall meet the requirement of the QS regulation, particularly 820.50, Purchasing, when they obtain products or services. Internal test laboratories, however, that are part of a corporate manufacturer that provides services to individual corporation factories should meet GMP requirements. Internal laboratories are inspected as part of the FDA GMP inspection of the member factories.

Situations are discussed in the remainder of this chapter where various manufacturers are exempt from the QS regulation or are not routinely inspected. However, these manufacturers are still subject to the FD&C Act. If these manufacturers or any manufacturer render devices unsafe or ineffective, the devices are adulterated and/or misbranded and the manufacturers are subject to the penalties of the FD&C Act.

Exemptions

FDA has determined that certain types of establishments are exempt from GMP requirements; and

FDA has defined GMP responsibilities for others. Exemption from the GMP requirements does not exempt manufacturers of finished devices from keeping complaint files (820.198) or from general requirements concerning records (820.180). Sterile devices are never exempted from GMP requirements. A device that normally would be subject to GMP requirements may be exempt under the following conditions:


Manufacturers should be aware of the GMP exemption status of their devices. In addition, manufacturers should keep on file records of any specific GMP exemption granted to them by FDA. Upon request during a factory visit, the exemption records need to be shown during normal business hours to the FDA investigator in order to verify that an exemption has been granted.

Component Manufacturers

A "component" is defined by 820.3(c) as "any raw material, substance, piece, part, software, firmware, labeling, or assembly which is intended to be included as part of the finished, packaged, and labeled device." Component manufacturers are excluded from the QS regulation by 820.1(a)(i). Current FDA policy is to rely upon the finished device manufacturer to assure that components are acceptable for use. Component manufacturers are not routinely scheduled for GMP inspections; however, FDA encourages them to use the QS regulation as guidance for their quality system.

When finished device manufacturers produce components specifically for use in medical devices they produce, whether in the same building or another location, such production of components is considered part of the device manufacturing operations, and the production should comply with the QS regulation.

Accessory devices [807.20(a)(5)] such as hemodialysis tubing or major diagnostic x­ray components, that are packaged, labeled, and distributed separately to a hospital, physician, etc., for health­related purposes are sometimes inappropriately referred to as components. However, FDA considers them finished devices because they are suitable for use or capable of functioning and are distributed for health-related purposes; and the QS regulation applies to their manufacture. Similarly, a device or component including software that is sold as an addition to a finished medical device to augment or supplement its performance is also termed an accessory. An accessory to a medical device is considered a finished device and, therefore, is subject to the QS regulation.

Remanufacturers

A remanufacturer is any person who processes, conditions, renovates, repackages restores or does any other act to a finished device which has been previously distributed to significantly change the finished device's performance or safety specifications or intended use from that established by the original finished device manufacturer. Remanufacturers are considered manufacturers. As such, these manufacturers are subject to inspection by FDA and shall meet the applicable requirements of the medical device QS regulation. These manufacturers shall establish and implement quality systems to assure the safety and effectiveness of the devices that are distributed. Such activities include drafting of master records, rebuilding per the master records, inspection and testing, calibration of measurement equipment, control of components, updating of labeling, processing of complaints, and any other GMP requirement applicable to the activities being performed.

Remanufacturers are also required to comply with the labeling requirements of 21 CFR 801.1(c). This labeling regulation requires that where the person or manufacturer named on the label of the device is not the original manufacturer, the name shall be qualified by an appropriate phrase which reveals the connection that person has with the device, e.g., remanufactured by XYZ Company.

Custom Device Manufacturers

Section 520(b) of the FD&C Act and the IDE regulation (21 CFR Part 812) define a custom device. Custom devices are exempt from certain statutory requirements. For example, manufacturers of custom devices are not required to comply with premarket approval requirements (Section 515) and are exempt from premarket notification requirements [Section 510(k)]. Custom devices are NOT exempt from the GMP requirements. Current FDA policy, however, is to not inspect manufacturers of custom devices. Manufacturers of custom devices should comply with the GMP requirements while considering the flexibility allowed.

Contract Manufacturers

A person(s) that manufactures a finished device under the terms of a contract with another manufacturer is a contract manufacturer. The agreement between the manufacturers should be documented in a written contract. Contract manufacturers of finished devices shall comply with applicable requirements of the quality system and shall register their establishment with FDA. Depending on the circumstances, both the contractor and manufacturer may be held jointly responsible by FDA for the activities performed.

Contract Testing Laboratories

Contract laboratories that designs or test components or finished devices for a manufacturer are considered an extension of the manufacturer's quality system. These laboratories may provide services to a number of customers, many of which are not medical device manufacturers. These contract laboratories are not subject to routine GMP inspections. Through the conduct of quality audits or other means, the finished device manufacturer is responsible for assuring that equipment and procedures used by a lab are adequate and appropriate (820.50). However, an internal test laboratory, if part of a manufacturer that does testing for various facilities within the corporation, is subject to inspection when FDA GMP inspections are conducted at the individual manufacturing facilities. That is, the test laboratory is simply a part of a medical device manufacturer of which all device-related divisions shall comply with the QS regulation.

Repackagers, Relabelers, and Specification Developers

Repackaging and relabeling of a device and specification development are defined as manufacturing in 21 CFR Part 807, Establishment Registration and Device Listing for Manufacturers of Devices. Some definitions from 807.3(d) are reprinted below because they affect the applications of the QS regulation.

(d) "Manufacture, preparation, propagating, compounding, assembly, or processing" of a device means the making by chemical, physical, biological, or other procedures of any article that meets the definition of a device in section 201(h) of the Act.

These terms include the following activities:

(1) Repackaging or otherwise changing the container, wrapper, or labeling of any device package in furtherance of the distribution of the device from the original place of manufacture to the person who makes final delivery or sale to the ultimate consumer;

(2) Initial distribution of imported devices; or

(3) Initiation of specifications for devices that are manufactured by a second party for subsequent commercial distribution by the person initiating specifications.

As defined above, repackaging and relabeling are manufacturing operations. Further, a repacker, repackager or relabeler is a manufacturer per 820.3(o) and subject to the applicable requirements of the QS regulation. Individuals are repackers or relabelers if they:

Individuals are not considered repackers or relabelers or a manufacturer for purposes of applying the QS regulation if they pack only previously packaged and labeled individual devices into packages for the convenience of the user. (Note that this activity is essentially the same as a drug store employee placing packaged items into a bag labeled with the name of the drug store.)

A distributor who only adds a label bearing their name and address is exempt from the GMP requirements. A manufacturer simply affixing a sticker label bearing the distributor's name and address would not require record keeping demonstrating compliance with labeling controls requirements.

Specification developers provide specifications to contract manufacturers, who produce devices to meet the specifications. The contract manufacturer may package and label the device, or the finished device may be shipped to the specification developer for packaging and labeling.

Specification developers are manufacturers and are subject to the GMP requirements that apply to the activities they conduct, such as various design controls including correct transfer of the design information to a contract manufacturer [820.30(h)]. This activity, in turn, requires an adequate device master record (820.181) and adequate change control [820.40(b)]. Further, if the product carries the specification developer's label, the developer is responsible for maintaining a complaint file and processing complaints, plus maintaining the device specifications and other appropriate documents in the device master record.

Initial Distributors of Imported Devices

The initial distributor is the foreign manufacturer's official correspondent with the FDA. With regards to the GMP, this initial distributor is responsible for maintaining complaint files and general record keeping requirements. A procedure shall be established and maintained for receiving, reviewing, and evaluating complaints. All complaints, including oral complaints, are to be processed in a uniform and timely manner. These complaints shall be evaluated to determine whether or not they require reporting to FDA under 21 CFR part 804 or 803, Medical Device Reporting. The initial distributor is also required to evaluate all complaints to determine whether an investigation is necessary, as well as complying with all other requirements in 820.198, Complaint Files. See Chapter 15 in this manual for more complete guidance on handling complaints.

Updated January 1, 1997

horizonal rule

CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page

Center for Devices and Radiological Health / CDRH