On September 14, 2006, FDA issued an initial statement related to concerns about adverse events related to coronary drug-eluting stents (DES). The statement noted that new data suggested a small but significant increased risk of stent thrombosis in patients who have been treated with the currently approved DES (the CYPHER stent and the TAXUS stent). FDA has made detection of DES thrombosis signals a priority because of the potential for serious harm to patients—even though stent thrombosis occurs at low rates. While the new data raised important questions, the Agency did not have enough information to draw conclusions. FDA announced plans to convene a public panel meeting of outside scientific experts to assist us in a thorough review of all available data and make recommendations about what actions may be appropriate, such as possible labeling changes or additional studies.
On December 7 and 8, 2006, the Circulatory System Devices Advisory Panel met in an effort to fully characterize the risks, timing and incidence of DES thrombosis. The purposes of this meeting were: (1) to provide a forum for the presentation of clinical data relevant to the issue of DES thrombosis (both when DES are used according to their label and in more complex patients beyond their labeled indication) and (2) to address the appropriate duration of antiplatelet therapy (aspirin plus clopidogrel) in DES patients. Panel members and public speakers represented a broad spectrum of interest and expertise including interventional cardiologists, non-interventional cardiologists, cardiovascular surgeons, biostatisticians, and the DES manufacturers.
In response to specific questions posed by FDA, the Panel had the following recommendations regarding DES when they are used in accordance with their approved indications:
The Panel was also asked to address the broader use of DES in patients with more complex patients and coronary lesions compared to those patients studied to support initial marketing approval. The use of a drug or device outside the FDA-approved indications is known as “off-label use” Although FDA regulates the manufacture, labeling, and promotion of devices, we do not regulate how they are used by individual clinicians in the practice of medicine. However, FDA may take action if safety issues with any use of a device become a public health concern. We felt that DES safety associated with off-label use should be included in the Panel’s deliberations given observations that at least 60% of current DES use is off-label. The Panel had the following comments and recommendations:
Regarding the duration of antiplatelet therapy:
Following this meeting, FDA has been carefully considering the new data presented at the meeting, the opinions from public speakers, and the Panel’s deliberations and recommendations. We will be working closely with the manufacturers of both approved DES and other DES still under study to incorporate appropriate modifications to labeling and changes to pre- and post-approval studies. Additionally, we will continue to work with professional societies, consumer organizations, and health care providers to provide physicians and patients with the most updated information as quickly as possible.
Updated January 4, 2007
CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page
Center for Devices and Radiological Health / CDRH