Skip page top navigationFDA Logo--links to FDA home page Logo of and Link to start page of Office of Regulatory Affairs, U.S. Food and Drug Administration U.S. Food and Drug Administration Office of Regulatory Affairs HHS Logo and link to Department of Health and Human Services website

FDA Home Page | Compliance | Federal-State | Inspection | Import Program | Science Ref | ORA Search  Vignette Managed Content: Click here for Metadata

Science References
ORA Laboratory Manual
Previous Page| Document TOC| Next Page

DFS Pyramid Logo

ORA Laboratory Procedure

Food and Drug Administration

DOCUMENT NO.:

ORA-LAB.4.9

VERSION NO.:1.2

Control of Non-conforming Work ORA-LAB.4.9

EFFECTIVE DATE: 10/01/2003 REVISED: 06/27/2008

Sections Included in this Document and Document History

  1. Purpose 
  2. Scope
  3. Responsibilities
  4. Background
  5. References
  6. Procedure
  7. Definitions
  8. Records
  9. Supporting Documents
  10. Attachments
  11. Document History

1. Purpose

To establish a procedure and to assign responsibilities for identification, documentation and disposition of non-conforming work products.

2. Scope

This procedure applies to the Office of Regulatory Affairs (ORA) laboratories and laboratory work products. This procedure directly concerns the laboratory’s quality control program.

3. Responsibilities

ORA laboratories ensure that the authority and the responsibility for controlling non-conformances are delegated to the appropriate management authorities. These authorities are authorized to halt any analytical testing or procedures related to the nonconformance and to invalidate test results that are affected. The responsible personnel authorized to resume work after a nonconformance is identified in the laboratory’s corrective action procedure.

4. Background

None

5. References

None

6. Procedure

  1. Non-conformances can occur at various places within the quality system and technical operations, examples include customer complaints, unacceptable quality control samples, instrument problems, samples, environmental problems that affect results, purchased materials for laboratory use, staff observations, management reviews and audits.
  2. Identified non-conformances with any procedure, quality control parameter or customer requirement are documented on the laboratory’s corrective action form and initiate the laboratory’s corrective action process. This process involves the evaluation of the impact on quality and operations.
  3. When a non-conformance is detected, laboratory data is held and not released until the problem is resolved and verified by laboratory management in accordance with the corrective action process. Resumption of work is performed after the corrective action has been taken and approved.
  4. Non-releasable data is not approved by management until product disposition has been made and documented. Dispositions or actions taken on a non-conforming work product are:
    • Rework – action taken on non-conforming product so that it will fulfill the specified requirements;
    • Redone – action taken  to re-collect sample or  reanalyze (redo) sample to bring the product into conformance;
    • Use as is – approving the use of non-conforming product without rework or redoing, a disclaimer is made that the product was accepted and the quality requirements that the product did not meet are specified; and
    • Unable to use – action taken if unable to resolve the problem. The receiver is notified that the data cannot be reported. 
  5. Reworked or redone products are reviewed to verify that they comply with specifications.
  6. When necessary, the customer is notified of the non-conformance and specifications may be changed depending on the usage of the data, for example, informational purposes only.
  7. A customer supplied product (sample) which is lost, damaged or otherwise unsuitable will be annotated in the Field Accomplishment and Compliance Tracking System (FACTS) and reported to the customer verbally or electronically.
  8. If properly executed, quality control parameters can monitor the various aspects of data quality on a routine basis. In instances where performance falls outside acceptable limits, the data produced can be questioned and, after investigation, a determination made as to its validity. The laboratory’s internal quality control program is the principal recourse available for ensuring that only a quality product is released. Quality control parameters and quality assurance elements are defined in the laboratory’s quality control program. These identified quality objectives are the critical elements that would cause a non-conforming product if not met.

7. Definitions

Non-conformance – This is a departure of a quality characteristic from its intended level or state that occurs with a severity sufficient to cause an associated product or service not to meet a specification requirement.

8. Records

Corrective Action form.

9. Supporting Documents

Local corrective action procedure.

10. Attachments

None

11. Document History

Version
No.

Status
(I, R, C)

Date
Approved

Location of Change History 

Name & Title

Author

Approving Official

           

 

 

 

 

 

 

 

 

 

 

 

 

           
           
           

 

 

 

 

 


 
Approving Official’s Signature: ______________________________________ Date: _______________

 