Phosphor X-ray Detectors

The physics of X-ray conversion for phosphors is essentially the same as described for scintillators. Also, the key functional elements of any phosphor-based detector are the same as for a scintillator. For astronomy, the similarities end there. While large area single crystals of NaI and CsI are valued for their stopping power in hard X-ray and gamma-ray astronomy, thin layers of phosphors are used to do high resolution, soft X-ray imaging. In fact, theoretically, phosphors can have the highest spatial resolution of all photon counting X-ray imagers -- limited to a few times the phosphor grain size (about a micron) in films only a few grain layers thick. This extraordinary resolution has yet to be achieved in any practical sense.

Decay constants vary enormously with phosphor composition. A commonly used material, P43 (the name for Gd2 O2 S(Tb)), has a light decay time of a few hundred microseconds -- which severely limits its maximum counting rate.

X-ray energy conversion efficiencies are comparable with those of NaI, about 10 - 15 The same is true of energy resolution.

The X-ray absorption efficiency is extremely dependent on what kind of phosphor you have and how thick it is. At 6 keV, P43 exceeds 85absorption efficiency.

If words seem to be missing from the articles, please read this.

Imagine the Universe! is a service of the High Energy Astrophysics Science Archive Research Center (HEASARC), Dr. Alan Smale (Director), within the Astrophysics Science Division (ASD) at NASA's Goddard Space Flight Center.

The Imagine Team
Project Leader: Dr. Jim Lochner
Curator:Meredith Gibb
Responsible NASA Official:Phil Newman
All material on this site has been created and updated between 1997-2008.
Last Updated: Monday, 27-Sep-2004 11:26:10 EDT