Medical Device Recalls
Class I Recall: Baxter Healthcare Corp. COLLEAGUE® Volumetric Infusion Pumps

Date Recall
Initiated:

July 19, 2005

Product:
All models of Baxter Healthcare Corp. COLLEAGUE® Volumetric Infusion Pumps.
Affected Models are: 2M8151, 2M8151R, 2M8161, 2M8161R, 2M8153, 2M8153R, 2M8163, and 2M8163R.
Use:
These electronic infusion pumps are used to give controlled amounts of medications or other fluids to patients through an intravenous (IV), intra-arterial (IA), epidural or other direct line into the bloodstream.
Recalling Firm:
Baxter Healthcare Corporation
Rt. 120 & Wilson Road
Round Lake, IL 60073
Reason for Recall:
The pumps can shut down while supplying critical medication and fluids to patients. Critically ill patients needing continuous delivery of life-sustaining medications at the time of the pump’s failure could be seriously harmed.
Public Contact:
Center for One Baxter
Baxter Healthcare Corporation
1 Deerfield Parkway
Deerfield, IL 60015
800-422-9837
FDA District:
Chicago
Advice to Users:
  • Baxter advised customers on March 15, 2005, to stop using any pumps that exhibit a failure code beginning with 402, 403, 533, 535, or 599, related to these electronic problems.
  • Baxter also advised customers to take out of service any pumps that exhibit failure codes 810:04 and 810:11 related to air-in-line sensor problems, until they are inspected by authorized service personnel.
  • These failures may occur during the infusion of therapy, so it is imperative that health care institutions have a contingency plan to mitigate any disruptions of infusions of life-sustaining drugs or fluids.

Class I recalls are the most serious type of recall and involve situations in which there is a reasonable probability that use of the product will cause serious injury or death.

 

Consumers who have questions about the recall may contact Baxter Healthcare at 1-800-422-9837. Those who have technical questions may contact Baxter Healthcare at 1-800-THE-PUMP (800-843-7867). Baxter's letters to customers are available on its web site at www.Baxter.com.

The FDA press release announcing the Baxter COLLEAGUE® Volumetric Infusion Pumps recall is available at http://www.fda.gov/bbs/topics/NEWS/2005/NEW01209.html.

Updated July 22, 2005