(You are encouraged to copy and distribute this information)
October 11, 2000
Dear Colleague:
This is to notify you of the potential for serious thermal injury and related complications associated with the use of microwave energy to treat benign prostatic hyperplasia (BPH), and to provide information that can help avoid these complications. Although the use of microwave thermotherapy for the treatment of BPH has been demonstrated to be safe and effective, and more than 25,000 procedures have been performed, we are concerned about some unexpected procedure-related complications that have occurred since the marketing of these devices.
Currently marketed devices include the Prostatron (Edap Technomed, Inc.) and the Targis System (Urologix, Inc.). Dornier Medical Systems, Inc. has received approval to market their UroWave System but is not yet marketing it. We are working with the manufacturers to ensure that labeling and training programs address these complications.
Nature of the Problem
Since 1996, we have received reports of 16 thermal injuries related to microwave thermotherapy systems. Of these, 10 resulted in fistula formation and 6 resulted in clinically significant tissue damage to the penis or urethra. These injuries may not be apparent at the time of treatment, and may take hours or days to develop. (Note that the original labeling for these devices did not list fistula formation as a procedure-related complication.) The reported injuries have required colostomies, partial amputation of the penis, and/or other therapeutic interventions.
We have identified several factors that may have contributed to the injuries noted:
Recommendations
Background
Microwave thermotherapy systems are intended to heat the prostate, resulting in the necrosis of periurethral prostatic tissue, to provide relief of urinary symptoms in patients with obstructive BPH. These devices heat the prostate to therapeutic levels using microwave energy delivered by an antenna contained within a specially designed urethral catheter. The catheter is designed so that when the balloon is seated at the neck of the bladder, the active portion of the antenna is positioned within the prostate. To prevent overheating, the systems circulate cooling fluid through the urethral catheter to protect the urethral tissue from excessive heat and automatically vary microwave energy output during treatment based on information supplied by temperature sensors placed posterior to the prostate within the rectum. Treatment may last from 30 to 60 minutes.
Because the catheter and/or the rectal temperature sensors can migrate during treatment, and because the correct placement of both of these components is critical for safe and effective treatment, the labeling for all these devices instructs the treating physician to: (1) verify that the urethral catheter (and rectal temperature sensor probe, if applicable) has a working retention balloon prior to placement, and (2) verify the proper position of both the urethral catheter and the rectal temperature sensors prior to and at specified time intervals consistent with the manufacturer’s recommendation for treatment. These requirements are intended to help ensure that catheter or rectal temperature sensor migration does not occur in a manner which would cause undetected excessive heating of surrounding tissues or the delivery of therapeutic heating levels to areas of the body that are not intended to receive treatment. The labeling for microwave thermotherapy devices also instructs the treating physician to monitor the equipment and patient during treatment, and manually reduce or pause the microwave power if the patient experiences excessive pain or extreme heating is observed.
Reporting Adverse Events to FDA
The Safe Medical Devices Act of 1990 (SMDA) requires hospitals and other user facilities to report deaths and serious injuries associated with the use of medical devices. FDA is interested in additional data on adverse events involving the use of microwave thermotherapy systems. When submitting a report, please identify the treatment protocol and catheter type, if known. Healthcare providers that are employed by facilities that are subject to FDA’s user facility reporting requirements should follow the reporting procedures established by their facilities. All other providers may submit their reports to MedWatch, FDA’s voluntary reporting program. The reports can be submitted by phone at 1-800-FDA-1088; by fax at 1-800-FDA-0178; by mail to MedWatch, Food and Drug Administration, HF-2, 5600 Fishers Lane, Rockville, Maryland 20857, or online at www.accessdata.fda.gov/scripts/medwatch.
Getting More Information
If you have questions regarding this letter, please contact the Issues Management Staff, Office of Surveillance and Biometrics (HFZ-510), 1350 Piccard Drive, Rockville, Maryland, 20850, by fax at 240-276-3356, or by e-mail at phann@cdrh.fda.gov. Additionally, a voice mail message may be left at 240-276-3357and your call will be returned as soon as possible.
All of the FDA medical device postmarket safety notifications can be found on the World Wide Web at http://www.fda.gov/cdrh/safety.html. Postmarket Safety notifications can also be obtained through e-mail on the day they are released by subscribing to our list server. Subscribe at: http://service.govdelivery.com/service/subscribe.html?code=USFDACDRH_10
Sincerely yours, | |
David W. Feigal, Jr., MD, MPH Director Center for Devices and Radiological Health Food and Drug Administration |
Updated October 11, 2000
CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page
Center for Devices and Radiological Health / CDRH