Compliance Program Guidance Manual

Chapter 42 - Blood and Blood Products Inspection of Source Plasma Establishments, Brokers, Testing Laboratories, and Contractors - 7342.002

 

Implementation Date: October 1, 2006
Completion Date: September 30, 2009
Product Codes: 57DY Source Plasma
57DY Source Leukocytes Human
57DI Further mfr. In Device/Device Components
57YY Miscellaneous Biologics
Programs/Assignment Codes: 42002 F Source Plasma Level 1 Inspection
    (all systems)
42002 G Source Plasma Level 2 Inspection
    (two systems)
42002A Pre-License
42832 Pre-Approval
42R825 AIDS Related Activity


FIELD REPORTING REQUIREMENTS

  1. General
  2. Districts should send Establishment Inspection Reports (EIRs) that contain issues requiring policy development or clarification or questions concerning inspection issues to the Center for Biologics Evaluation and Research (CBER) for review. Send the EIR and relevant exhibits electronically, if possible, to cberinspections@fda.hhs.gov, or by mail to the following address:

    Division of Inspections and Surveillance (HFM-650)
    Office of Compliance and Biologics Quality
    Center for Biologics Evaluation and Research
    Food and Drug Administration
    1401 Rockville Pike, Suite 200 N
    Rockville, MD 20852-1448

  3. Inspection Reporting - Endorsement Section of EIR
  4. Annotate the following information within the FACTS Endorsement (Inspection Summary field) in addition to the information specified in the Investigations Operations Manual (IOM):

    1. Inspection Level and criteria used to determine if a Level 1 or a Level 2 inspection would be performed.
    2. If a Level 2 inspection was performed, specify each system inspected.
    3. Document when an expected Level 2 inspection is changed to Level 1 based on the finding of significant objectionable conditions.

  5. Fatality Follow Up
  6. Send a copy of the relevant sections of the EIR with exhibits pertinent to the fatality to the Fatality Program Manager (HFM-650) at the general address above, or electronically to the point of contact that issued the inspection follow up assignment.

    Report a product collection-related fatality, if found during an inspection, to the Fatality Program Manager at 301-827-6220.

  7. Warning Letters to Source Plasma Establishments
  8. Send a copy of the Warning Letter and any correspondence between the firm and the district office, to the Blood and Tissue Compliance Branch, Division of Case Management (DCM) at the general address above. For electronic submission, contact DCM at (301) 827-6201.

    Copies of the Warning Letters may also be sent to an appropriate state agency. Refer to the Regulatory Procedures Manual, Chapter 4, Advisory Actions, for guidance on this issue.


TABLE OF CONTENTS

Part I Background

Part II Implementation

  1. Objective
  2. Strategy
  3. Program Management Instructions
  4. Frequency of CGMP Inspections
  5. Scheduling Inspections
  6. Assignment of Investigators and Compliance Personnel

Part III Inspection

  1. Strategy
    1. Level 1 Inspection Option
    2. Level 2 Inspection Option
    3. Standard Operating Procedures
    4. Training & Personnel
    5. Facilities
    6. Equipment
    7. Records

  2. Plasma Brokers
  3. Contractors
  4. Biologic License Applications (BLA)
  5. Inspection Instructions

Part IV Analytical

Analyzing Laboratories

Part V Regulatory / Administrative Strategy

  1. Deficiencies
    1. General
    2. Quality Assurance System
    3. Donor Eligibility System
    4. Product Testing System
    5. Quarantine / Storage / Disposition System
    6. Product Collection and Processing System

  2. Federal / State Relations

Part VI References, Attachments and Program Contacts

  1. Laws and Regulations
  2. ORA Inspection Manuals and Inspection Guides
  3. Guidance Documents and Memoranda Pertaining to Source Plasma Manufacture
    1. Donor Eligibility
    2. Source Plasma Collection and Special Programs
    3. Infectious Diseases
    4. Inspections
    5. Labeling
    6. Miscellaneous
    7. Computers

  4. Center for Biologics Evaluation and Research and Office of Regulatory Affairs Program Contact

Part VII Center for Biologics Evaluation and Research Responsibilities

Attachments

  1. Quality Assurance System
  2. Donor Eligibility System
  3. Product Testing System
  4. Quarantine / Storage / Disposition System
  5. Product Collection and Processing System
  6. Computers
  7. Lookback
  8. Physician Substitutes
  9. Source Plasma Collection Programs
  10. Types of Blood Establishments Covered Under This Program
  11. Source Leukocytes and Therapeutic Exchange Plasma

PART I - BACKGROUND

Source Plasma, Source Leukocytes, and Therapeutic Exchange Plasma are subject to the licensure provisions of Section 351 of the Public Health Service Act (PHS).

Under the provisions of Section 351 and the Federal Food, Drug, and Cosmetic Act (FD&C), FDA investigators conduct inspections of blood establishments, including those that manufacture or participate in the manufacture of Source Plasma, Source Leukocytes, and Therapeutic Exchange Plasma for human use. The inspection is to ensure that firms manufacture biological products that are safe, pure, potent and have the quality that they represent and that they manufacture these products according to Current Good Manufacturing Practice (CGMP) for Blood and Blood Components regulations and applicable standards. Source Plasma, Source Leukocytes and Therapeutic Exchange Plasma intended for further manufacture into injectable and non-injectable products are biological products. Under the FD&C Act, Source Plasma may have the legal identity of either a drug or a device depending on its intended use. Sources Plasma is subject to the full extent of regulations pursuant to all provisions of the FD&C Act.

FDA began the inspection of Source Plasma establishments in 1973. To provide more effective and efficient regulation of biological products, FDA established a cadre of trained investigators responsible for inspecting blood banks, Source Plasma establishments and firms associated with those operations. This specialized cadre has proved very effective in performing inspections since 1997. Office of Regulatory Affairs (ORA) oversight responsibility for blood and Source Plasma is assigned to the Biological Products Field Committee.

This compliance program builds upon the knowledge gained during previous FDA inspections of the Source Plasma industry and recent scientific developments. It provides a risk-based approach to the CGMP inspection of Source Plasma establishments with focus on the operating systems of those establishments.


PART II - IMPLEMENTATION

  1. Objective
  2. This program represents a continuing compliance and surveillance activity that began in 1973. The objective of this program is to ensure that Source Plasma, Source Leukocytes, and Therapeutic Exchange Plasma (TEP) for further manufacture into products for human use are safe, effective, appropriately labeled, and have the quality these products are represented to possess. FDA compliance and surveillance activities also assess compliance with donor protection standards which are intended to ensure a continuous and healthy donor population.

    The inspection instructions in this document apply to the manufacture of Source Plasma. A Source Plasma manufacturing process must:

    • Meet the standards described in applicable provisions of the regulations, including those specifically intended to protect donors. These standards include regulations in 21 CFR Parts 600, 601, 606, 607, 610, 630, and 640, process and production controls, equipment regulations and quality assurance in 21 CFR Part 211, and other applicable standards; and
    • Meet any additional conditions of licensure included in the Source Plasma manufacturer's approved Biologic License Application.

  3. Strategy
  4. This compliance program uses a systems-based approach to conduct an inspection. It identifies 5 systems in the Source Plasma establishment's operation for inspection. The inspection is a comprehensive evaluation of the critical areas in each system used by the establishment. Problems in the critical areas may adversely compromise a Source Plasma donor's safety and/or affect product quality if procedures are not performed properly or system controls are inadequate or function incorrectly. The following systems have been identified:

    1. Quality Assurance System - various planned activities that provide confidence that all procedures/processes that influence product manufacture and overall quality are monitored to ensure they are working as expected.
    2. Donor Eligibility System - the system that protects the donor's safety, determines donor eligibility for blood collection (including donor deferral resulting from either medical history screening and/or testing), and notifies the donor of ineligibility for donation.
    3. Product Testing System - the system that includes procedures to properly test products collected for further manufacture for evidence of communicable disease agents consistent with 21 CFR 610.40.
    4. Product Collection and Processing System - the system that controls the collection and processing of Source Plasma, including issues of product quality and donor safety
    5. Quarantine/Storage/Disposition System - the system that manages product quarantine, storage, and distribution (release for use or destruction).

    The inspection is based on a multi-layered set of safeguards (referred to as the "five layers of safety") related to the collection, manufacture, and distribution of blood and blood components, including Source Plasma. The five layers of safety are:

    1. Donor Screening - procedures to identify donors who have one or more defined risk factors for one or more communicable diseases or who are otherwise ineligible to donate.
    2. Donor Deferral - procedures to identify ineligible donors and prevent the distribution of blood products collected from these donors unless the donors participate in a special collection program.
    3. Product Testing - procedures to properly test products for further manufacture for evidence of infection by specific communicable disease agents.
    4. Quarantining - procedures to ensure that blood products are quarantined until all tests and control procedures are acceptable and unsuitable products are removed from inventory and destroyed or appropriately labeled and distributed; e.g., for use in research, test kit controls, etc.
    5. Monitoring and Investigating Problems - procedures to identify system problems, biologic product deviations, and donor adverse reactions, including assurance that adequate corrective action is implemented.

    Table 1 - Relationship of Layers of Safety to Source Plasma Systems

    Each system in the inspectional approach relates to one or more of the "five layers of safety" as follows:

    Layer of Safety System(s)
    Donor Screening Donor Eligibility, Quality Assurance
    Donor Deferral Donor Eligibility, Quality Assurance
    Product Testing Product Testing, Quality Assurance, Product Collection and Processing
    Quarantining Quarantine/Storage/Disposition, Quality Assurance, Product Collection and Processing
    Monitoring and Investigating Problems Quality Assurance, Product Collection and Processing

    The inspection of a Source Plasma establishment is conducted under a Level 1 or Level 2 inspection option. PAC codes document the inspection based upon the level of inspection.

    A Level 1 inspection is a comprehensive evaluation of the establishment's compliance. It includes the review of all the systems employed, regardless of the number. For example, a contract testing laboratory that performs communicable disease testing for a Source Plasma establishment usually has only two of five systems (Quality Assurance and Product Testing). All inspections of this type of establishment would be considered Level 1.

    A Level 2 inspection provides a streamlined evaluation of an establishment's compliance when the facility has met defined standards of performance during past FDA inspections. Two of the systems present will be inspected. Selection and documentation of the specific systems to inspect is described later in this compliance program.

    See Part III, Inspection, for selection criteria for Level 1 and Level 2 inspections.

  5. Program Management Instructions
  6. This program covers the following establishment types. See Attachment J for definitions and registration requirements.

    1. Source Plasma Establishments
    2. Plasma Brokers
    3. Testing Laboratories
    4. Contractors
    5. Off-site Storage Facilities
    6. Other Blood Establishments

    Investigators may query the CBER Blood Establishment Registration database on the CBER Intranet, regulatory site to review registration information.

  7. Frequency of CGMP Inspections
  8. CGMP inspections are statutory obligations that are generally conducted on a biennial schedule (based upon the previous date of inspection).

    The following are exceptions:

    1. Firms under a Consent Decree of Permanent Injunction.
    2. A new Source Plasma establishment (i.e. new applicant or new facility/location under an existing license) or an establishment re-opening under an existing license is inspected within the first year of operation.
    3. A for-cause inspection or a compliance follow-up inspection to verify a Source Plasma establishment’s implementation of corrective action after regulatory action.
    4. A facility that does not engage in manufacturing, e.g., a broker that only arranges the sale or shipment of products is inspected at the district’s discretion or for cause.
    5. An establishment that changes location should be inspected within 60-90 days of the change or as soon as district work plans permit.
    6. A Pre-License (PLI) or Pre-Approval inspection (PAI).*
    7. A firm or establishment location under a Notice of Intent to Revoke and/or other administrative action.

    * Note: The Center for Biologics Evaluation and Research (CBER) and Office of Regulatory Affairs (ORA) jointly conduct PLI and PAI inspections with CBER as the lead. On occasion, CBER may request that ORA lead a PAI inspection. PLI and PAI inspections are part of the review of a biologics license application or supplement. CBER identifies the scope and content of the inspection.

  9. Scheduling Inspections
  10. ORA and CBER jointly develop the annual inspection workplan. District office staff schedule statutory CGMP inspections of domestic establishments according to the ORA workplan.

    For the process that CBER uses to schedule pre-approval and pre-license inspections, refer to Part III, Inspection, Biologic License Applications.

  11. Assignment of Investigators and Compliance Personnel
  12. Whenever possible, only Investigators who completed the required Blood Banking and Plasmapheresis training course(s) should inspect establishments covered under this program.

    Whenever possible, only Compliance Officers who completed the required Blood Banking and Plasmapheresis training course(s) should process compliance recommendations.


PART III - INSPECTION

  1. Strategy
  2. This program provides two surveillance inspection options, Level 1 and Level 2. All levels satisfy the biennial inspection requirement.

    Each inspection of a Source Plasma establishment should extend to the required operating procedures, personnel/training, facilities, equipment, and records of each system inspected. The inspection should include actual observation of the processes applicable to the system.

    Inspect each system to the extent necessary to determine whether the Source Plasma establishment complies with applicable regulations. The systems for coverage are Quality Assurance, Donor Eligibility, Product Testing, Product Collection and Processing, and Quarantine/Storage/Disposition. Source Plasma establishment means a facility or location licensed to collect Source Plasma.

    For every inspection, select a representative sample of units with reactive viral marker test results. Track these units through the establishment from product collection, processing, donor deferral, product quarantine and disposition to determine the Source Plasma establishment's appropriate handling of these units. Determine if the test lab uses FDA approved test kits and is registered with FDA and is Clinical Laboratories Improvement Act (CLIA) certified for the level of testing.

    For every inspection, also track a few units that are unacceptable for reasons other than reactive viral marker test results (e.g.QNS, improper storage), and track these units through quarantine and product disposition to determine appropriate handling of such units.

    For every inspection, assess any major and moderate changes (e.g. significant change to manufacturing processes or manufacture of a new product) since the preceding routine inspection.

    1) Level 1 Inspection Option . The Level 1 Option is a comprehensive evaluation of a Source Plasma establishment's compliance. Select the Level 1 option for:

    • Initial inspection of a Source Plasma establishment by district investigators
    • Source Plasma establishments that have a history of fluctuating in and out of compliance
    • Compliance follow-up inspections
    • Source Plasma establishments that perform testing for communicable disease agents
    • A surveillance inspection at the district's discretion
    • After conducting two previous inspections under a Level 2 option

    The Level 1 option includes inspection of the critical areas of all the systems employed at the Source Plasma establishment regardless of their number. The inspection will cover each existing system.

    If investigators find significant objectionable conditions within one or more systems during the course of an inspection, such as those listed in Part V. A., Deficiencies, they should consult their district management to determine the scope and depth of the inspection necessary to support possible regulatory and/or administrative action.

    2) Level 2 Inspection Option. The Level 2 option is a streamlined evaluation of an establishment’s compliance. Utilize this inspection option only for a Source Plasma establishment that does not perform their own communicable disease testing on site. All systems do not need to be active during the inspection to qualify for the Level 2 inspection option.

    The Level 2 Option must include inspection of the critical areas of the Quality Assurance System and one other existing system. The district determines the additional system after reviewing the establishment's file, evaluating the inspection history and assessing biologic product deviation reports, product recalls and other available information pertaining to the establishment. District program managers and investigators should make certain that coverage of the additional system under the Level 2 option is rotated in successive inspections, unless otherwise indicated. During the course of a Level 2 inspection, verification of QA activities may require limited coverage of other systems (for example, investigation and correction of deviations). In addition, directly observe at least one donation, from donor eligibility determination through final processing, as part of the Quality Assurance verification activity.

    Select the Level 2 Option when all the following conditions are met:

    • The Source Plasma establishment has a satisfactory history of compliance (two successive NAI or VAI CGMP inspections), which may predate the implementation date of this CP.
    • One of the two previous routine inspections was a Level 1 inspection. NOTE: A comprehensive inspection under previous inspection criteria may be considered Level 1.
    • The inspection preparation revealed no specific trends that may have a significant impact on product or donor safety, as identified during reviews of product recalls, biological product deviation reports, or fatality reports,

    Finding significant objectionable conditions while conducting a Level 2 inspection may prompt the district to consider a change to Level 1 inspection, or conclude the inspection.

    The inspection should cover the following areas within each system inspected:

    1. Standard Operating Procedures (SOPs)
    2. The Source Plasma establishment shall maintain written SOPs that include all steps to be followed in the collection, processing, labeling, storage and distribution of Source Plasma products. SOPs shall be available to personnel in the areas where they perform such operations. [21 CFR 606.100]

    3. Training and Personnel
    4. The personnel responsible for the collection, processing, storage, or distribution of Source Plasma shall be adequate in number, educational background, training and experience, including professional training as necessary, or a combination thereof, to ensure competent performance of assigned functions and to ensure that the final product has the safety, purity, potency, identity, and effectiveness it purports or is represented to possess. [21 CFR 606.20]

    5. Facilities
    6. Facilities shall be maintained in a clean and orderly manner and shall be of suitable size, construction and location to facilitate adequate cleaning, maintenance and proper operation. The facility must comply with the requirements of 21 CFR 606.40.

      Conduct a walk-through of the facility to ensure that the facility complies with applicable regulations and to determine if any problem areas exist.

    7. Equipment
    8. Equipment used in the collection, processing, storage, and distribution of Source Plasma, Therapeutic Exchange Plasma and Source Leukocytes shall be maintained in a clean and orderly manner and located so as to facilitate cleaning and maintenance. The equipment shall be observed, standardized and calibrated on a regularly scheduled basis, as prescribed in the SOP manual. The equipment shall also perform in the manner for which it was designed so as to ensure compliance with the official requirements for Source Plasma manufacture. [21 CFR 606.60]

    9. Records
    10. The Source Plasma establishment must maintain records concurrently with the performance of each significant step in collecting, processing, quarantining, storing, and distributing each Source Plasma product so that all steps can be clearly traced. All records shall be legible and indelible and shall identify the person performing the work, including dates of the various entries, test results as well as interpretation of the result; the expiration date assigned to the specific product; and shall be as detailed as necessary to provide a complete history of the work performed [21 CFR 606.160]. Each donor must have a separate and complete record that is cross-referenced to the Source Plasma units collected from the donor [21 CFR 640.72].

      During the inspection,
      1. Determine if the Source Plasma establishment maintains all required records applicable to Source Plasma manufacture under 21 CFR 606.160 and the additional records as required under 21 CFR 640.72. All required records must also meet the requirements of 21 CFR 606.160 (a)(1) and (2).
      2. Review as many records as necessary in each system to determine if records are complete and maintained as required and that each record contains positive identification of the donor on all records describing the history and disposition of Source Plasma products. [21 CFR 640.64, 640.65, 640.72]

  3. Plasma Brokers
  4. Inspecting a Plasma Broker is required only if the broker takes possession of Source Plasma products and performs a manufacturing operation; e.g., culling out unsuitable Source Plasma units.

    During the inspection,
    At a minimum, evaluate the broker's compliance with the following regulations: 21 CFR 606.100(b)(10), 606.160(b)(2) and (3), 606.165, 607, and 640.70. These regulations pertain to storage, records of product receipt, product pooling, labeling, distribution and registration.

  5. Contractors
  6. A Source Plasma manufacturer may contract with another establishment to perform one or more manufacturing steps. Both the Source Plasma manufacturer and contractor are responsible for product quality. The Source Plasma establishment, as the license holder, remains responsible for compliance with applicable product and establishment standards. The contractor is responsible for complying with applicable CGMP. Inspect contractors that perform services, such as testing, pooling, culling, and/or preparing and supplying Red Blood Cell immunogen cells to a Source Plasma manufacturer. http://www.fda.gov/cber/gdlns/coopmfr.pdf

    During the inspection,
    1. Determine the extent of services provided.
    2. Determine each party's responsibility for the product or operations performed.
    3. Determine who prepared the SOPs used by the contractor.
    4. Determine who performed product quality control tests

  7. Biologic License Applications (BLAs)
  8. FDA licenses Source Plasma, Source Leukocytes, and Therapeutic Exchange Plasma, biological products, under the authority of section 351(a) of the PHS Act. A biologics license must be in effect for a biological product prior to its introduction or delivery for introduction into interstate commerce.

    CBER issues a single BLA tracking number to an applicant (the person or legal entity who submits an application to manufacture a product subject to licensure). The applicant may submit separate supplements to the BLA to manufacture individual products. CBER approves a BLA to manufacture Source Plasma and issues a U.S. license number to the applicant if the applicant can manufacture Source Plasma products that are safe, pure, potent and effective for their intended use and maintains an establishment to manufacture Source Plasma that meets CGMP and other standards designed to ensure that Source Plasma products remain safe, pure, potent and effective for their intended use. [21 CFR 601.4] The U.S. license number must appear on the product label.

    Licensed manufacturers must submit a written request, as appropriate, to CBER to implement a change in a licensed product, a production process, quality control, equipment, or to a facility previously approved in a license application. [See 21 CFR 601.12 and the Guidance for Industry, "Changes to an Approved Application: Biological Products: Human Blood and Blood Components Intended for Transfusion or for Further Manufacture." http://www.fda.gov/cber/gdlns/bldchanges.htm]

  9. Inspection Instructions
  10. Instructions for coverage within each of the five (5) systems for inspection and other relevant inspection issues are included in the following attachments:

    1. Attachment A - Quality Assurance System (QA)
    2. Attachment B - Donor Eligibility System
    3. Attachment C - Product Testing System
    4. Attachment D - Quarantine/Storage/ Disposition System
    5. Attachment E - Product Collection and Processing System
    6. Attachment F - Computers
    7. Attachment G - Lookback
    8. Attachment H - Physician Substitutes
    9. Attachment I - Source Plasma Collection Programs
    10. Attachment J - Types of Blood Establishments
    11. Attachment K - Source Leukocytes and Therapeutic Exchange Plasma

PART IV - ANALYTICAL

  1. Analyzing Laboratories
  2. No field analyses are planned under this program.

    When analyses are required, refer to the following chart for instructions.

    Individual Responsibility
    Sample Custodian Arranges for the testing of any physical sample collected on an assignment
     

    Receives samples
    Ensures staff availability for sample arrival

     

    Address: Sample Custodian (Attention: HFM-672)
    Center for Biologics Evaluation and Research
    Bldg. B., Room 113
    5516 Nicholson Lane
    Kensington , MD 20895
    Telephone: 301-594-6517
    FAX: 301-594-6924

       
    HFM-650 Returns test results to the Director, Compliance Branch
     

    Provides contact(s) to respond to questions regarding analytical results
    Telephone: 301-827-6220

       
    Investigator Obtains packaging and shipping information from Sample Custodian
      Prior to shipment, notifies the Sample Custodian of projected date of sample arrival
      Sends test samples to the Sample Custodian by courier

PART V - REGULATORY/ADMINISTRATIVE STRATEGY

When inspection findings demonstrate that a manufacturer of Source Plasma is not operating in a state of control and/or the establishment's management is either unwilling or unable to implement adequate corrections in a timely manner, districts should consider the advisory, administrative and/or judicial options currently available. A firm is considered to be operating in a state-of-control when it employs conditions and practices that ensure compliance with the intent of Section 501(a)(2)(B) of the FD&C Act, and the portions of the CGMP regulations that pertain to its systems. A firm in a state of control produces products for which there is an adequate level of assurance of quality, identity, purity and potency.

The district office should base all regulatory recommendations on significant deviations that are well documented. The quality of any action begins with the quality of evidence collected at the time of the inspection to support the observed objectionable conditions. The recognition, collection, and effective presentation of evidence are essential to successful advisory, administrative, and judicial actions. The identification of those responsible for violations is also a critical part of the inspection. Establish responsibility and identify persons to hold accountable for violations, and with whom the Agency must communicate to seek lasting corrections and/or to be the subject of enforcement actions. The decision on the type of action to recommend should be based on the seriousness of the documented deficiencies with regard to, and the most effective way to protect, public health.

A firm's written corrective action, in response to the FDA-483, should not preclude the consideration of an advisory, administrative, or judicial action. Recommend such action, if the objectionable observations represent (a) a continuing pattern of non-compliance, (b) a failure to correct significant deficiencies noted during a previous inspection, or (c) the deficiencies pose a significant threat to the public health, and voluntary action is either not appropriate or can not be readily accomplished.

For products intended for further manufacture, the advisory, administrative, and judicial options available include:

Action Among other things, consider if,
Warning Letter Violations of regulatory significance that causes one or more systems to be considered not in a state of control.

License Revocation
21 CFR 601.5

Notice of Intent to Revoke with Opportunity for Correction:

  • Unable to gain access to the manufacturing facility for inspection.
  • Licensed products are not safe or effective for their intended use, or are misbranded with respect to any such use.
  • Manufacturer fails to report a change in accordance with 21 CFR 601.12.
  • Manufacturer fails to conform to applicable standards to ensure product safety, potency, and purity.
  • Licensed products are no longer manufactured.

Direct Revocation without Opportunity for Correction

  • Demonstration of willful disregard.
License Suspension
21 CFR 601.6
Reasonable grounds for revocation and a danger to health exist. It provides immediate withdrawal of the authorization to ship a biological product in interstate commerce.
Seizure Manufacturer is unwilling or unable to retrieve violative products or products held for sale are unsuitable for safe use. U.S. Marshal takes possession of products through Court Order pursuant to Section 304 of the FD&C Act.
Injunction A current health hazard exists, the establishment has a history of uncorrected deviations despite previous warnings, suspension of the firm's license would result in an unacceptable shortage of products, and/or to halt intrastate distribution of products manufactured under violative conditions.
Prosecution Fraud; gross, flagrant, intentional violations; or a continuous or repeated course of violative conduct.

Districts should initially consider an advisory action, such as a warning letter, if there is no previous violative history at the firm.

To determine the appropriate action, consult with CBER/Office of Compliance and Biologics Quality (OCBQ)/Division of Case Management (HFM-610), early in the investigation, and consult the Regulatory Procedures Manual (RPM). This early consultation is especially critical when immediate action is indicated (e.g., license suspension or temporary restraining order (TRO)). See RPM Chapter 6, regarding TROs and an injunction to protect the public health. When inspection findings indicate the potential for fraud (e.g., falsification, counterfeiting, illegal importation, drug diversion), the investigator should notify his/her supervisor, who will notify CBER/OCBQ. District management will alert the appropriate Office of Criminal Investigations Office, as appropriate. The investigator should, however, continue to pursue any public health concerns, in coordination with CBER/OCBQ, concurrently.

Evidence of significant and /or a pattern of deficiencies (history) within a system covered could constitute failure of that system. The district should classify as Official Action Indicated, an inspection report that documents one or more systems as not in a state of control. The district should then consider a Warning Letter or other appropriate action. When deciding the type of action to recommend, follow the RPM and base the initial decision on the seriousness and /or frequency of the problem and the firm’s compliance history.

Districts may issue Warning Letters per RPM Chapter 4, to achieve voluntary compliance and establish prior notice. The RPM outlines the types of letters requiring CBER concurrence prior to issuance. All Warning and Untitled Letter recommendations require Office of Chief Counsel review and clearance.

  1. Deficiencies
  2. The following, although not all-inclusive, are examples of deficiencies that may be indicative of the firm's state-of-control:

    1. General
      1. Any practice or pattern of practices that poses a danger to public health and warrants grounds for license revocation [21 CFR 601.6 (a)
      2. ]
      3. Failure to provide adequate facilities and to maintain them in a clean and orderly manner [21 CFR 606.40]
      4. Failure of a licensed establishment to notify CBER of any change that has a substantial potential to have an adverse effect on the product as it relates to the safety or effectiveness of the product [21 CFR 601.12(b)]
      5. Falsifying, changing or altering product labels or records [42 U.S.C. 262(b) 21 CFR 606.160; 640.70; 640.72]
      6. A history of similar significant deficiencies
      7. Recurrent problems with a computer system(s), such as a donor deferral system, testing equipment interface, labeling system, and/or quarantine/distribution system that could lead to the release of unsuitable products [21 CFR 211.68(b)]
      8. Any failure to completely identify the container or laboratory samples so they can be correlated to the individual donor [21 CFR 606.140(c); 606.160(c); 640.72(a)(2)]
      9. Failure of the physician to be physically on premises during red blood cell immunizations [21 CFR 640.62]

    2. Quality Assurance System
      1. Failure to establish and implement a written quality assurance program [21 CFR 211.22(d), 211.100, and 606] (Reference this item when substantive deficiencies are noted that lead to a conclusion that system and process controls are inadequate and cannot prevent violative conditions.)
      2. Lack of computer and/or software validation or a lack of documentation associated with the performance or analysis of validation activities [21 CFR 606.160; 211.68(b)]
      3. Failure to establish and implement adequate computer security provisions (passwords, user authentication, and remote access) to ensure data integrity [21 CFR 211.68(b)]
      4. Personnel inadequately trained or supervised in the operations they perform to such an extent that a danger to the health of the donor or safety of the product exists [21 CFR 606.20 and 640.64(a)]
      5. A PS supervising an immunization program prior to receiving appropriate training and / or an individual performing the duties of a PS prior to CBER approval of a Source Plasma establishment’s PS training program [21 CFR 606.20(b) and 640.66]
      6. A pattern of personnel training deficiencies [21 CFR 606.20]
      7. Failure to maintain, standardize, calibrate, and follow established procedures for equipment used in the collection, processing, testing, storage, and distribution of Source Plasma [21 CFR 606.60]
      8. Failure to recognize, adequately respond to, and investigate adverse reactions and document and maintain appropriate records [21 CFR 606.170(a) and 606.100(b)(9)]
      9. Duplicate, discrepant, or invalid records existing in the computer's donor deferral files that could lead to the acceptance of unsuitable donors and release of unsuitable products. [21 CFR 606.160(b), 606.100(c) and 211.68]
      10. Failure to immunize donors by acceptable procedure(s) and/or use approved antigens [21 CFR 640.66]
      11. Failure to advise CBER of fatalities resulting from complications related to blood collection and failure to investigate the cause of death [21 CFR 606.170(b)]
      12. Failure to thoroughly investigate and incorporate appropriate corrections concerning any unexplained discrepancy or the failure of a lot or a unit to meet specifications that may affect the safety, purity, or potency of the product [21 CFR 606.100(c)]
      13. Failure to promptly notify CBER of biological product deviations in the manufacture of Source Plasma that may affect the safety, purity, or potency of the product [21 CFR 606.171(c) and (e)]
      14. Use of unapproved procedures pertaining to major changes and/or implementation of unapproved major changes in manufacturing methods. [21 CFR 601.12]
      15. Failure to establish procedures and/or repeated failure to follow SOPs and/or to maintain appropriate records for the proper handling of post donation information reports (CPG 230.140) [21 CFR 211.100, 211.192; 211.198; 606.100; and 606.160]
      16. Failure to establish written procedures and maintain adequate records for an immunization program [21CFR 606.160(a)(1); 640.66; CPG 250.100]

    3. Donor Eligibility System
      1. Failure to establish a donor identification system that confirms donor identity and correlates medical records, test results, and components to the donor record [21CFR 640.65(b)(3); 640.72(a)(2)]
      2. Any personnel or system failure that causes the establishment to accept or inappropriately re-enter ineligible donors [21 CFR 606.20(b), 640.63(a), and 640.640]
      3. Failure to establish or repeated failure to follow SOPs for donor eligibility determinations [21 CFR 606.100(b)(1) and (2); 640.63]
      4. Failure to explain hazards or risks of procedure by a physician or designee, to obtain a signed informed consent form, or implement appropriate safeguards to minimize the potential for transmission of unexpected infectious agents [21 CFR 606.160(b)(1)(v), 606.100(b), 640.61 and CPG 250.100]
      5. Failure to properly determine the eligibility of the donor (e.g., routine physical examinations, medical history questions, hemoglobin, blood pressure, temperature, total protein, plasma or serum protein electrophoresis or quantitative immuno-diffusion or equivalent test to determine immunoglobulin composition) [21 CFR. 640.63, 640.65(b)(1)(i), 640.65(b)(2)(i), and 606.20(b)]
      6. Failure to perform and adequately document physician's review of laboratory and collection records [21 CFR 640.65(b)(2)]
      7. Failure to have a licensed physician perform physical examinations for immunization, unless approved as an alternative procedure under 21 CFR 640.120 [21 CFR 640.63 (b)(2)(i)
      8. ]
      9. Failure to maintain accurate records that identify ineligible donors so that the establishment will not distribute products from such individuals [21 CFR 606.160 (b)(1) and (e)]
      10. Pattern of failure to make a reasonable attempt to notify donors of deferral status [21 CFR 630.6]

    4. Product Testing System
      1. Any failure to perform communicable disease testing [21 CFR 610.40 and 640.65 and 640.67]
      2. Failure to perform tests or interpret results according to manufacturers' instructions and specifications; e.g., use of outdated reagents or mixing of reagents from different master lots; failure to run the proper number of controls concurrently with the test; inappropriate invalidation of test results; calculations incorrectly determined resulting in reactive results being interpreted as nonreactive; interpreting reactive test results as nonreactive; and failure to conduct necessary retests. [21CFR 610.40, and 606.65(e)]
      3. Communicable disease tests not performed using an approved or licensed test kit or otherwise allowed in 21 CFR 610.40.
      4. Incomplete or inaccurate testing records, including all records associated with invalidated test runs [21 CFR 606.160(b)(2)(i); 640.72(a)(2)]

    5. Quarantine/Storage/Disposition System
      1. Source Plasma and/or other products for further manufacture not stored at proper temperature [21 CFR 610.53, 640.69(b), 640.74]
      2. Failure to maintain temperature records when Source Plasma is in storage or shipped. [21 CFR 606.160(b)(3)(iii); 640.72(a)(1)]
      3. Failure to store final containers under conditions that prevent exposure of the product to external factors that may cause deterioration or contamination [21 CFR 640.68(b)]
      4. Failure to establish or follow a system that prevents the distribution of any products not suitable for use [21 CFR 606.40(a)(6); 606.40(d)(2); 606.100(b); 606.160(e); and 610.40 (h)]
      5. Failure to obtain CBER approval to ship product that is not tested for communicable disease agents required in 21 CFR 610.40.
      6. Failure to establish or follow a system by which receipt and distribution of each unit of Source Plasma can be readily determined to facilitate recall, if necessary [21 CFR 606.165(a)]
      7. Failure to quarantine unsuitable products or to notify consignees in accordance with 21 CFR 610.46(a)

    6. Product Collection and Processing System
      1. Failure to collect Source Plasma by methods that protect against contamination of the final product and returned red blood cells [21 CFR 640.64(b) and (e)
      2. ]
      3. Failure to follow equipment manufacturer's instructions for plasmapheresis procedures [21 CFR 640.65, 606.60]
      4. Failure to use approved containers for the collection of Source Plasma [21 CFR 640.68(b)]
      5. Failure to maintain complete and accurate processing records [21 CFR 606.160(a)(1), (b)(2); 640.72]
      6. Failure to properly label blood components [21 CFR 610.62 and 640.70]
      7. Failure to record whole blood weights in manual collection and to recognize and/or correct overbleeding in manual or automated collection procedures [21 CFR 640.65(b)(4)(5)(6) and (8) and 606.160(a)(1) and CPG 252.100]
      8. Failure to prevent the infusion of one donor's red blood cells into another donor [21 CFR 640.65(b)(3) and 606.100(b)(17)]

  3. Federal / State Relations
  4. Currently FDA has no formal cooperative program with state or local jurisdictions to inspect or regulate Source Plasma establishments. Nonetheless, districts should cooperate with these authorities, especially if the state or local jurisdiction has a regulatory program. Whenever possible, districts should exchange information with all levels of government consistent with information disclosure procedures. Provide a copy of a warning letter to the appropriate state agency or agencies. If a state official requests a copy of the Form FDA-483, redact the document according to FOI procedures prior to release. For additional assistance, contact the Office of Regulatory Affairs/Division of Federal State Relations (HFC-150) at (301) 827-6906.


PART VI REFERENCES, ATTACHMENTS, AND PROGRAM CONTACTS

  1. Laws and Regulations
  2. Federal Food, Drug, and Cosmetic Act as Amended and Related Laws http://www.fda.gov/opacom/laws/

    Public Health Service Act, Subpart I - Biological Products. http://www.fda.gov/opacom/laws

    Title 21, Code of Federal Regulations, Parts 211, 600, 601, 606, 607, 610, 640 and 820. http://www.gpoaccess.gov/cfr/index.html

  3. ORA Inspection Manuals and Inspection Guides
  4. FDA Investigations Operations Manual, 2005 or latest edition. Sections 500-529, 560-565, 590-595, 635, 773, 921, 924, 927-928, 1026 http://www.fda.gov/ora/inspect_ref/iom

    FDA Regulatory Procedures Manual (RPM), March 2006 or latest edition. Chapter 4 – Advisory Actions, Chapter 5- Administrative Actions, Chapter 6 – Judicial Actions, Chapter 7 – Recall and Emergency Procedures, Chapter 9 – Import Operations/Actions. http://www.fda.gov/ora/compliance_ref/rpm

    FDA Compliance Policy Guides, August 2000 or latest edition, Chapter 1 - General and Chapter 2 – Biologics. http://www.fda.gov/ora/compliance_ref/cpg/default.htm

    Guide to Inspections of Infectious Disease Marker Testing Facilities, June 1996. http://www.fda.gov/ora/inspect_ref/igs/infdis.html

    Guide to Inspection of Source Plasma Establishments, June 1997 (April 2001 – Editorial Revisions) http://www.fda.gov/ora/inspect_ref/igs/Source_Plasma/default.htm

  5. Guidance Documents and Memoranda Pertaining to Source Plasma Manufacture
  6. Donor Eligibility

    FEDERAL REGISTER: Guidance for Industry: Discontinuation of Donor Deferral Related to Recent Fever with Headache as a Symptom of West Nile Virus Infection; Withdrawal of Guidance, June 30, 2005 http://www.fda.gov/cber/gdlns/wnvdnrdfr.pdf

    Guidance for Industry: Assessing Donor Suitability and Blood and Blood Product Safety in Cases of Known or Suspected West Nile Virus Infection, June 23, 2005 http://www.fda.gov/cber/gdlns/wnvguid.htm

    Guidance for Industry: Revised Recommendations for the Assessment of Donor Suitability and Blood Product Safety in Cases of Suspected Severe Acute Respiratory Syndrome (SARS) or Exposure to SARS, September 16, 2003 http://www.fda.gov/cber/gdlns/sarsbldgd.htm

    Guidance for Industry: Revised Recommendations for the Assessment of Donor Suitability and Blood and Blood Product Safety in Cases of Known or Suspected West Nile Virus Infection, May 1, 2003. http://www.fda.gov/cber/gdlns/wnvguid.pdf

    Guidance for Industry: Recommendations for the Assessment of Donor Suitability and Blood Product Safety in Cases of Suspected Severe Acute Respiratory Syndrome (SARS) or Exposure to SARS, April 2003. http://www.fda.gov/cber/gdlns/sarsbldgd.htm

    Question and Answer on FDA Guidance Entitled “Recommendations for The Assessment of Donor Suitability and Blood and Blood Product Safety in Cases of Suspected and Probable Severe Acute Respiratory Syndrome (SARS) or Exposure to SARS,” June 5, 2003. http://www.fda.gov/cber/gdlns/sarsbldq&a.htm

    Guidance for Industry: Recommendations for Deferral of Donors and Quarantine and Retrieval of Blood and Blood Products in Recent Recipients of Smallpox Vaccine (Vaccinia Virus) and Certain Contacts of Smallpox Vaccine Recipients, December 30, 2002. http://www.fda.gov/cber/gdlns/smpoxdefquar.htm

    Questions and Answers on FDA Guidance Entitled “Recommendations for Deferral of Donors and Quarantine and Retrieval of Blood and blood Products in Recent Recipients of Smallpox Vaccine (Vaccinia Virus) and Certain Contacts of Smallpox Vaccine Recipients,” http://www.fda.gov/cber/gdlns/smpoxdefquarq&a.htm

    Guidance for Industry: Recommendations for Assessment of Donor Suitability and Blood Product Safety in Cases of Possible Exposure to Anthrax, October 17, 2001. http://www.fda.gov/cber/gdlns/anthraxexp.htm

    Guidance for Industry: Revised Preventive Measures to Reduce the Possible Risk of Transmission of Creutzfeldt-Jakob Disease (CJD) and Variant Creutzfeldt-Jakob Disease (vCJD) by Blood and Blood Products, January 2002. http://www.fda.gov/cber/gdlns/cjdvcjd.htm

    Interim Recommendations for Deferral of Donors at Increased Risk for HIV-1 Group O Infection, December 11, 1996. http://www.fda.gov/cber/bldmem/mem121196a.txt

    Donor Deferral Due to Red Blood Cell Loss During Collection of Source Plasma by Automated Plasmapheresis, December 4, 1995. http://www.fda.gov/cber/bldmem/mem120495.pdf

    Recommendations for the Deferral of Current and Recent Inmates of Correctional Institutions as Donors of Whole Blood, Blood Components, Source Leukocytes, and Source Plasma, June 8, 1995. http://www.fda.gov/cber/bldmem/060895.txt

    Donor Suitability Related to Laboratory Testing for Viral Hepatitis and a History of Viral Hepatitis, December 22, 1993. http://www.fda.gov/cber/bldmem/122293.txt

    Deferral of Blood and Plasma Donors Based on Medications, July 28, 1993. http://www.fda.gov/cber/bldmem/072893.txt

    Revised Recommendations for the Prevention of Human Immunodeficiency Virus (HIV) Transmission by Blood and Blood Products, April 23, 1992. http://www.fda.gov/cber/bldmem/hiv042392.pdf

    Exemptions to Permit Persons with a History of Viral Hepatitis Before the Age of Eleven Years to Serve as Donors of Whole Blood and Plasma: Alternative Procedures, 21 CFR 640.120 April 23, 1992. http://www.fda.gov/cber/bldmem/042392ex.txt

    Deferral of Donors Who Have Received Human Pituitary-Derived Growth Hormone, November 25, 1987. http://www.fda.gov/cber/bldmem/112587.txt

    Source Plasma Collection and Special Programs

    Donor Deferral Due to Red Blood Cell Loss During Collection of Source Plasma by Automated Plasmapheresis, December 4, 1995. http://www.fda.gov/cber/bldmem/mem120495.pdf

    Revised Recommendations for Red Blood Cell Immunization Programs for Source Plasma Donors, March 14, 1995. http://www.fda.gov/cber/bldmem/031495.txt

    Revision of FDA Memorandum of August 27, 1982: Requirements for Infrequent Plasmapheresis Donors, March 10, 1995. http://www.fda.gov/cber/bldmem/031095.txt

    Volume Limits for Automated Collection of Source Plasma, November 4, 1992. http://www.fda.gov/cber/bldmem/110492.txt

    Revisions to 26 October 1989 Guideline for Collection of Blood or Blood Products from Donors with Positive Tests for Infectious Disease Markers (“High Risk” Donors), April 17, 1991. http://www.fda.gov/cber/bldmem/041791.txt

    Guideline for Collection of Blood or Blood Products from Donors with Positive Tests for Infectious Disease Markers (“High Risk” Donors), October 26, 1989. http://www.fda.gov/cber/bldmem/102689.txt

    Extension of Dating Period for Storage of Red Blood Cells, Frozen, December 4, 1984. http://www.fda.gov/cber/bldmem/120487.txt

    Plasma Derived from Therapeutic Plasma Exchange, December 14, 1984. http://www.fda.gov/cber/bldmem/121484.txt

    Guidelines for the Collection of Human Leukocytes for Further Manufacturing (Source Leukocytes), January 28, 1981. http://www.fda.gov/cber/bldmem/012881.pdf

    Guidelines for Immunization of Source Plasma (Human) Donors with Blood Substances, June 1980. http://www.fda.gov/cber/gdlns/imnztnsrcpl.pdf

    Infectious Diseases

    Guidance for Industry: Use of Nucleic Acid Tests on Pooled and Individual Samples from Donors of Whole Blood and Blood Components (including Source Plasma and Source Leukocytes) to Adequately and Appropriately Reduce the Risk of Transmission of HIV-1 and HCV, October 21, 2004. http://www.fda.gov/cber/guidelines.htm - natbld

    Guidance for Industry: Revised Preventive Measures to Reduce the Possible Risk of Transmission of Creutzfeldt-Jakob Disease (CJD) and Variant Creutzfeldt-Jakob Disease (CJD) by Blood and Blood Products, January 9, 2002. http://www.fda.gov/cber/gdlns/cjdvcjd.htm

    Questions and Answers on FDA Guidance Entitled Revised Preventive Measures to Reduce the Possible Risk of Transmission of Creutzfeldt-Jakob (CJD) Disease and Variant Creutzfeldt-Jakob Disease (vCJD) by Blood and Blood Products. Last updated 1/22/04. http://www.fda.gov/cber/gdlns/cjdvcjdq&a.htm

    Guidance for Industry: Current Good Manufacturing Practice for Blood and Blood Components: (1) Quarantine and Disposition of Units from Prior Collections from Donors with Repeatedly Reactive Screening Tests for antibody to Hepatitis C Virus (Anti-HCV); (2) Supplemental Testing and the Notification of Consignees and Blood Recipients of Donor Test Results for Anti-HCV, September 23, 1998. Obtain (Doc 0621) from CBER/OCTMA/HFM-40, 1401 Rockville Pike, Rockville, MD 20852 1-800-835-4709.

    Interim Recommendations for Deferral of Donors at Increased Risk for HIV -1 Group O Infection, December 11, 1996. http://www.fda.gov/cber/bldmem/mem121196a.txt

    Recommendations for the Quarantine and Disposition of Units from Prior Collections from Donors with Repeatedly Reactive Screening Tests for Hepatitis B Virus (HBV), Hepatitis C Virus (HCV), and Human T- Lymphotropic Virus Type I (HTLV-I), July 19, 1996. http://www.fda.gov/cber/bldmem/mem71996.txt Note: The HCV section of this document is superseded by guidance document dated September 23, 1998.

    Additional Recommendations for Testing Whole Blood, Blood Components, Source Plasma and Source Leukocytes for Antibody to Hepatitis C Virus Encoded Antigen (Anti-HCV), May 16, 1996. http://www.fda.gov/cber/bldmem/antigen1.txt

    Additional Recommendations for Donor Screening with a Licensed Test for HIV-1 Antigen, March 14, 1996 http://www.fda.gov/cber/bldmem/antigen1.txt

    Recommendation for Donor Screening with a Licensed Test for HIV-1 Antigen, August 8, 1995 http://www.fda.gov/cber/bldmem/hiv-ag.txt

    Clarification of the Use of Unlicensed Anti-HCV Supplemental Test Results in Regard to Donor Notification, August 19, 1993. http://www.fda.gov/cber/bldmem/081993.txt

    Revised Recommendations for Testing Whole Blood, Blood Components, Source Plasma and Source Leukocytes for Antibody to Hepatitis C Virus Encoded Antigen (Anti-HCV), August 5, 1993. http://www.fda.gov/cber/bldmem/080593.txt

    Revised Recommendations for the Prevention of Human Immunodeficiency Virus (HIV) Transmission by Blood and Blood Products, April 23, 1992. http://www.fda.gov/cber/bldmem/hiv042392.pdf

    Use of Fluorognost HIV-1 Immunofluorescent Assay (IFA), April 23, 1992. http://www.fda.gov/cber/bldmem/042392.txt

    Use of Genetic Systems HIV-2 EIA, June 21, 1990. http://www.fda.gov/cber/bldmem/062190.txt

    HTLV-1 Antibody Testing (to Licensed Source Plasma Manufacturers Approved for Immunization with Red Blood Cells) July 6, 1989. http://www.fda.gov/cber/bldmem/070689.pdf

    Recommendations for the Management of Donors and Units that are Initially Reactive for Hepatitis B Surface Antigen (HBSAG), December 2, 1987. http://www.fda.gov/cber/bldmem/120287.pdf

    Inspections

    Guideline for Quality Assurance in Blood Establishments, July 11, 1995. http://www.fda.gov/cber/gdlns/gde_qa.txt

    Discontinuance of Pre-License Inspection of Immunization Using Licensed Tetanus Toxoid and Hepatitis B and Rabies Vaccine, July 7, 1988. http://www.fda.gov/cber/bldmen/070788.txt

    Control of Unsuitable Blood and Blood Components, April 6, 1988. http://www.fda.gov/cber/bldmem/040688.txt

    Product Recalls, Including Removals and corrections, October 2003. http://www.fda.gov/cber/gdlns/prodrecall.pdf

    Labeling

    Guidance for Industry: Recognition and Use of a Standard for the Uniform Labeling of Blood and Blood Components, June 6, 2000. http://www.fda.gov/cber/gdlns/unilabbld.htm

    Guidance for Industry: United States Industry Consensus Standard for the Uniform Labeling of Blood and Blood Components Using ISBT 128, November 1999. http://www.fda.gov/cber/blood/bldguid.htm

    Recommendations for Labeling and Use of Units of Whole Blood, Blood Components, Source Plasma, Recovered Plasma or Source Leukocytes Obtained from Donors with Elevated Levels of Alanine Aminotransferase, (ALT), August 8, 1995. http://www.fda.gov/cber/bldmem/alt.txt

    Guideline for the Uniform Labeling of Blood and Blood Components, August 1985. [Obtain Document D0053 from CBER, OCTMA, HFM-40, 1-800-835-4709 or 301-827-1800.]

    Miscellaneous

    Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 12, 2005 http://www.fda.gov/cdrh/ode/guidance/337.html

    Guidance for Industry: Notifying FDA of Fatalities Related to Blood Collection or Transfusion, 9/22/2003 http://www.fda.gov/cber/gdlns/bldfatal.htm

    Guidance for Industry: Streamlining the Donor Interview Processes: Recommendations for Self-Administered Questionnaires, July 3, 2003. http://www.fda.gov/cber/gdlns/donorsaq.htm

    Guidance for Industry: Changes to an Approved Application: Biological Products: Human Blood and Blood Components Intended for Transfusion or for Further Manufacture, August 7, 2001. http://www.fda.gov/cber/gdlns/bldchanges.htm

    Guidance for Industry: Revised Recommendations Regarding Invalidation of Test Results of Licensed and 510(k) Cleared Bloodborne Pathogen Assays Used to Test Donors, July 11, 2001. http://www.fda.gov/cber/gdlns/bldbrn.htm

    Options for Alternative Arm Preparation – Clinipad Recall 3/9/2000; (Updated 4/25/2001). http://www.fda.gov/cber/blood/armpreprev.htm

    Guidance for Industry: Errors and Accidents Regarding Saline Dilution of Samples Used for Viral Marker Testing, June 11, 1998. http://www.fda.gov/cber/gdlns/easaline.htm

    FDA Guidance for Industry on: Exports and Imports Under the FDA Export Reform and Enhancement Act of 1996, Draft February 1998. http://www.fda.gov/opacom/fedregister/frexport.html

    Revised Recommendations for Red Blood Cell Immunization Programs for Source Plasma, March 14, 1995. http://www.fda.gov/cber/bldmem/031495.txt

    Guidance Regarding Post Donation Information Reports, December 10, 1993. http://www.fda.gov/cber/bldmem/121093.txt

    Physician Substitutes, August 15, 1988. http://www.fda.gov/cber/bldmem/081588.txt

    Guidance for Industry, For the Submission of Chemistry, Manufacturing and Controls and Establishment Description Information for Human Blood and Blood Components Intended for Transfusion or for Further Manufacture and For the Completion of the Form FDA 356h "Application to Market a New Drug, Biologic or an Antibiotic Drug for Human Use, May 10, 1999. http://www.fda.gov/cber/gdlns/cmcblood.htm

    Computers

    FEDERAL REGISTER: Draft Guideline for the Validation of Blood Establishment Computer Systems; Withdrawal of Guidance, March 9, 2005 http://www.fda.gov/cber/gdlns/bldestcompwdrl.pdf

    Guidance for Industry: Part 11, Electronic Records; Electronic Signatures – Scope and Application, September 3, 2003. http://www.fda.gov/cber/gdlns/prt11elect.pdf

    General Principles of Software Validation; Final Guidance for Industry and FDA Staff, January 11, 2002. http://www.fda.gov/cdrh/comp/guidance/938.html

    Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - 5/12/2005 http://www.fda.gov/cdrh/ode/guidance/337.html

    Glossary of Computerized System and Software Development Terminology, August 1995. http://www.fda.gov/ora/inspect_ref/igs/gloss.html

    Letter to Blood Establishment Computer Software Manufacturers, March 31, 1994. http://www.fda.gov/cber/bldmem/033194.txt

    Recommendation for Implementation of Computerization of Blood Establishments, April 6, 1988. http://www.fda.gov/cber/bldmem/040688co.txt

    Guideline on General Principles of Process Validation, May 1987. http://www.fda.gov/cder/guidance/pv.htm

    Guide to Inspection of Computerized Systems in Drug Processing, February 1983 http://www.fda.gov/ora/inspect_ref/igs/csd.html

  7. CENTER FOR BIOLOGICS EVALUATION AND RESEARCH AND OFFICE OF REGULATORY AFFAIRS PROGRAM CONTACTS
  8. OFFICE OF COMPLIANCE AND BIOLOGICS QUALITY, HFM-600

    1. Immediate Office of the Director, HFM-600

    2. Anita Richardson, 301-827-6190, FAX 301-827-3381
      General Policy Issues

    3. Division of Case Management, HFM-610

    4. Robert Sausville, Director, 301-827-6201, FAX 301-594-0940

      Advertising and Promotional Labeling; Application Integrity; Biological Product Recalls; Certificates of Export; Citations; Civil Money Penalties; Compliance Status Checks; Debarment; Import/Export Programs; Injunctions; License Suspensions; Prosecutions; Revocations and Denials; Seizures; Tissue Recall Orders; Warning Letters

      Blood and Tissue Compliance Branch, HFM-614
      Stephany Wesley, Branch Chief 301-827-6201, FAX 301-594-0940

      Helen Cowley Dana Jones
      Laura Hieronymus Sinencio Salvador
      Wendy Hively Sandra Segar
    5. Division of Inspections and Surveillance, HFM-650

    6. Gilliam Conley, Director 301-827-6220, FAX: 301-827-6748

      Program Surveillance Branch, HFM-654
      Janet Ishimoto, Chief, 301-827-9094

      Maureen Knippen MDR
      Sharon O'Callaghan
      Beth Rogerson
      Biological Product Deviations
      Susan Cannon Fatality Program
      Fatality Transfusion Reporting
      http://www.fatalities2@fda.hhs.gov
      Voice Mail: 301-827-6220
      FAX: 301-827-6748
      Kimberly Miles
      David Keith
      Blood and Plasma Programs
      & Licensing Changes

    OFFICE OF BLOOD RESEARCH AND REVIEW, HFM-300

    1. Division of Blood Applications, HFM-370

    2. Elizabeth Callaghan, Acting Director, 301-827-3524, FAX: 301-827-3535

      Blood Registration Coordinator
      Janet O'Brien 301-827-3546

      Registration, Licensing, Labeling, Variances, Approvals for Changes

      Blood and Plasma Branch, HFM-375
      Leslie Holness, Chief, 301-827-3543, FAX: 301-827-3534

      Brenda Bell Lore Fields Hoi-May Wong
      Karan Blum Diane Hall Kenneth Zemann
      Judy Ciaraldi Jennifer Jones  
      Marla Cohen Rosia Nesbitt  

    OFFICE OF REGIONAL OPERATIONS (ORO)/OFFICE OF REGULATORY AFFAIRS (ORA)/DIVISION OF FIELD INVESTIGATIONS (DFI) CONTACTS

    ORO/ORA/DFI, HFC-130
    Michael C. Rogers, Director DFI, 301-827-5658, FAX 301-443-3757
    Mary Carden, National Expert, 716-551-4461 x3152
    Joan Loreng, National Expert, 215-717-3724
    Joyce Watson, CSO (Foreign Inspections), 301-827-5664

    ORO/ORA/DFI/Biologics Group, HFC-130
    Gerald W. Miller, SCSO, 301-827-5655, FAX 301-443-3757
    Gail Katz, CSO, 301-827-3357


PART VII - CENTER FOR BIOLOGICS EVALUATION AND RESEARCH RESPONSIBILITIES

The Center for Biologics Evaluation and Research (CBER), through its Office of Compliance and Biologics Quality (OCBQ)/Division of Inspections and Surveillance (DIS), works cooperatively with the Office of Regulatory Affairs (ORA) Biological Products Field Committee to monitor the inspection and compliance accomplishments under this compliance program, and the status of the establishments inspected under this program. The ORA annual workplan, developed by CBER and ORA, provides overall resource allocations and anticipated numbers of inspections. However, current industry practices encountered during an inspection, the past compliance history of an establishment, or other compliance developments, may necessarily result in unplanned inspections or in individual CGMP inspections taking more or less time than estimated in the workplan.

ORA continues to have the primary responsibility for ensuring (1) that the program strategies, priorities, and procedures articulated in this compliance program are followed by the ORA Field staff and (2) that potential problems or needs for policy/program clarification are brought to CBER's attention. CBER and ORA jointly coordinate activities to achieve industry compliance with applicable laws, regulations, and court orders (e.g., Consent Decrees of Permanent Injunction).

CBER/OCBQ will continue to use accomplishment data from the ORA Field Accomplishment and Compliance Tracking System (FACTS), legal or administrative action recommendations, requests for policy decisions/clarification received from the public or the blood industry, and input from CBER scientific and product experts to provide overall direction to FDA's blood safety initiatives that are supported by this risk-based compliance program.

CBER/OCBQ will send to the appropriate district Director of Investigation Branch email attachments containing approved changes to biologic license applications.

The Biological Products Field Committee and the OCBQ/DIS intend to have periodic conference calls concerning this program and an annual meeting, with other ORA and CBER units (e.g., CBER Office of Blood Research and Review) participating as necessary.

CBER/OCBQ will carefully evaluate the experience with this systems-based inspection program through inspection reports and other compliance data to determine its effectiveness and to continually assess and improve the quality of the CBER products inspection program. CBER/OCBQ also will carefully review and monitor industry compliance, product developments within industry, and the safety and quality of Source Plasma.


ATTACHMENT A
QUALITY ASSURANCE SYSTEM

The Source Plasma establishment’s quality assurance (QA) program should consist of various planned activities to provide confidence that all procedures/processes that influence Source Plasma manufacture and overall quality of the final product are working as expected. The QA program should include validation of processes. This activity, referred to as process validation, is a controlled system of activities that establishes through documented evidence that the establishment’s manufacturing processes will consistently produce a product that meets predetermined specifications for quality and intended use.

The individual or unit performing QA activities coordinates, monitors, and facilitates all activities, including approval of all procedures or specifications that impact the safety, purity, and quality of Source Plasma products and those that prevent the release of unsuitable products. The QA unit ensures adequate laboratory facilities and manufacturing operations consistent with CGMP and applicable standards and ensures that staff follows them. The QA unit ensures that there is adequate quality control of routine on-line or in-process monitoring of manufacturing procedures. [21 CFR 211.22]

The guidance document, “Guideline for Quality Assurance in a Blood Establishment” contains a more complete discussion of our recommendations related to QA functions/activities relevant to blood establishment operations. http://www.fda.gov/cber/gdlns/gde_qa.txt

  1. Standard Operating Procedures (SOPs)
  2. The unit or individual performing QA activities shall have the responsibility to review and approve written SOPs, including any changes. The QA program’s various planned activities should provide confidence in the manufacture and overall quality of Source Plasma products. The Source Plasma establishment must have procedures in place to identify failures of Source Plasma to meet specifications and should correct those failures.

    The Source Plasma establishment should evaluate the effectiveness of corrective actions implemented. Source Plasma establishments use various names when referring to these systems, such as error and accident reporting systems, incident reports, problem reports or logs, in-house problem reports (non-reportable) or biologic product deviations. In addition to maintaining records of product or process deviations, the Source Plasma manufacturer must also ensure that the following incidents are documented and investigated:

    • Adverse Reactions
    • Product Collection Fatalities
    • Lookback Events
    • Biological Product Deviations (BPD)

    The manufacturer must ensure that the Source Plasma establishment complies with regulatory requirements for record maintenance, storage and required reporting.

    During the inspection,
    1. Determine if SOPs are available and followed for each system inspected.
    2. Evaluate whether the QA unit fulfills its responsibility to review and approve all procedures related to production, quality control and quality assurance, and to ensure procedures are adequate for their intended use.
    3. Review the Source Plasma establishment’s procedure for determining a donor’s eligibility on the day of donation for consistency with the requirements in 21 CFR 640.63 and current CBER recommendations regarding donor eligibility.
    4. 4Review records to determine if QA investigates, documents, and implements corrections as required for biological product deviations, unexplained discrepancies, failures of Source Plasma to meet specifications (e.g., products not stored under appropriate storage conditions or shipped at unacceptable temperatures), and adverse reactions resulting from Source Plasma collection in order to identify all factors contributing to the problem. The documentation must include conclusions of the investigation and follow-up.
    5. Determine if there are any trends in deviations (reportable or non reportable) identified by the Source Plasma establishment. Note: A pattern of recurring problems may indicate an incomplete investigation or inadequate correction.

  3. Adverse Reactions
  4. The Source Plasma establishment must maintain a record of any donor reaction that it receives and must conduct a complete investigation and document the investigation findings [21 CFR 606.170]. The extent of the investigation should be based on the seriousness of the adverse donor reaction.

    During the inspection,
    Review records of adverse donor reactions.

  5. Donor Fatalities
  6. A Source Plasma establishment must report to CBER as soon as possible any complication of blood collection that results in a donor fatality. Within 7 days, the Source Plasma establishment must submit a written report of its investigation of the fatality to the Director, Office of Compliance and Biologics Quality. The Fatality Program Manager (HFM-650) usually receives initial notice within 24 hours of the fatality or, at least, by the next business day. [21 CFR 606.170 (b)]

    If an investigator becomes aware of an unreported fatality during an inspection, contact the Fatality Program Manager, Division of Inspections and Surveillance 301-827-6220 as soon as possible to discuss the circumstances surrounding the incident and to determine if the Source Plasma establishment should report the fatality.

    Determine that the Source Plasma establishment conducted an adequate and complete investigation.

    During the inspection,

    Fatality follow up

    1. Review the Source Plasma establishment’s investigation.
    2. Review donor records and if the Source Plasma establishment does automated collection, also review the device records, maintenance log, and other relevant records.
    3. Review both the initial notification to CBER and the 7-day written report. Often the fatality investigation is not completed in 7 days, so also review any report relevant to the investigation of the case that later became available. Determine if the firm reported accurate information to the Fatality Program Manager.
    4. Review any corrective action and procedure to monitor its effectiveness.
    5. Obtain a copy of the relevant records as indicated in the guidance document, “Notifying FDA of Fatalities Related to Blood Collection or Transfusion.” http://www.fda.gov/cber/gdlns/bldfatal.pdf

    Send a copy of the pertinent section of the EIR with exhibits to the Fatality Program Manager. Refer to the cover page of this document for the address.

  7. Lookback - See Attachment G
  8. Biological Product Deviations (BPD)
  9. The investigation of all deviations, including BPDs, is an integral part of the Source Plasma establishment's QA system. Under 21 CFR 606.171, the licensed manufacturer must report to CBER any event associated with the manufacture, or any information relevant to the event, including testing, processing, packing, labeling, or storage, or with the holding or distribution of Source Plasma that may affect the safety, purity, or potency of a distributed product.

    Events are required to be reported to CBER/OCBQ/DIS as soon as possible, but no later than 45 days from the date of discovery reasonably suggesting that a reportable event occurred. Under 21 CFR 606.171, the manufacturer who holds the biologics license and who had control over the product when the deviation or unexpected or unforeseen event occurred must report a BPD.

    If a manufacturer contracts out any manufacturing step, that manufacturing step is performed under the manufacturer's control under the regulation. Thus, under 21 CFR 606.171(a), the manufacturer must establish a procedure for receiving information from that contract manufacturing facility of all deviations, complaints, and adverse events that may affect the product.

    A contract manufacturer (i.e., performs, under contract, a step in manufacturing for another facility) must conduct manufacturing in accordance with all applicable regulations.

    Ensure that any reportable events that may have occurred have been reported to CBER.

    ORA investigators have direct access to BPD information through CEARS (CBER Error and Accident Reporting System). Instructions for accessing the system are posted on the CEARS Intranet web page. Refer questions to the Email address: bp_deviations@fda.hhs.gov

    To facilitate industry reporting of BPD, CBER developed a standardized reporting format (FDA Form 3486) with both hard copy and electronic reporting. CBER encourages electronic reporting.

    Website: http://www.fda.gov/cber/biodev/biodev.htm
    Email address: bp_deviations@fda.hhs.gov

    Prior to conducting an inspection, investigators should review the establishment's BPD submissions in CEARS. Deviation codes may indicate systems that the investigator will want to examine more closely for patterns or trends. Otherwise, select a representative sample of reports to determine the adequacy of the firm's corrective action.

    During the inspection,
    1. Evaluate both reportable and non-reportable incidents or problem reports and determine the adequacy of any corrective action implemented by the Source Plasma establishment.
    2. Determine if the Source Plasma establishment filed all reportable biological product deviations. (Contact OCBQ/DIS if clarification regarding BPD reporting is required.)

    Source Plasma establishments are expected to follow-up on all reported BPDs with documentation of investigations and/or corrective actions. It is CBER's current thinking that a reportable event that has been identified, investigated, and corrected need not be included as a Form FDA 483 observation.

  10. Donor Record Files
  11. The manufacturer must ensure that the Source Plasma establishment maintains all required records applicable to Source Plasma manufacture consistent with regulatory requirements for record maintenance and storage. QA should ensure that records are accurate and provide a complete history of all work performed. [21 CFR 606.160 and 640.72]

    The Source Plasma manufacturer must have separate and complete donor records (referred to as a donor record file) of all initial and periodic examinations, tests, laboratory data, interviews, etc. related to donor eligibility, product collection, immunization, and laboratory testing. [21 CFR 640.72]

    During the inspection,

    Review a sufficient number of the records to determine if the establishment collects Source Plasma from donors with acceptable health history and screening test results. Determine if the records contain the following information:

    1. Initial and annual physical examinations and consent for plasmapheresis and immunization, as applicable [21 CFR 640.72] Note: See Attachment I - Infrequent Plasmapheresis Collection program.
    2. Donor screening test results, e.g., hematocrit, temperature, blood pressure, pulse, donor weight, total protein, and donor medical history interviews [21 CFR 640.63]
    3. Tests for communicable diseases and results of 4-month syphilis and serum protein electrophoresis testing and review [21 CFR 640.65]
    4. Records of immunization, if applicable [21 CFR 640.66]
    5. A cross-reference to unit(s) of Source Plasma collected from the donor [21 CFR 640.72]
    6. Reason(s) for donor deferral, including red blood cell loss [21 CFR 606.160]
    7. The reason Source Plasma was determined unsuitable [21 CFR 640.72] Donor reactions that occurred on or after leaving the premises [21 CFR 640.72]
    8. Collection volume and donor weight

  12. Equipment
  13. The Source Plasma establishment's QA procedures should ensure:

    1. Appropriate calibration, cleaning and preventative maintenance of equipment according to manufacturer's recommendations and/or SOPs;
    2. Qualification of equipment and process validation, as necessary, including after repairs, ensure that equipment functions properly;
    3. Computer systems used in manufacturing comply with 21 CFR 211.68 and 21 CFR 606.60
    4. Computers, software, and interfaces used in the manufacture of Source Plasma are validated prior to implementation, qualified at the location where used, and revalidated as required.

    During the inspection,
    Review records and procedures the Source Plasma manufacturer used to ensure that equipment performed as it was designed, including validation, as necessary, maintenance and periodic monitoring, if required.

  14. Validation
  15. QA should ensure that the Source Plasma establishment has procedures for conducting process validation and for assessing the need for revalidation. QA should also monitor validated processes to ensure that they continue to work as expected.

  16. Evaluation of the Quality Assurance Program
  17. A Source Plasma establishment should periodically evaluate its QA program to determine if it is effective in detecting, correcting and preventing problems. Refer to the Guideline for Quality Assurance in Blood Establishments. http://www.fda.gov/cber/gdlns/gde_qa.txt


ATTACHMENT B
DONOR ELIGIBILITY SYSTEM

This system identifies the Source Plasma establishment's procedures intended to protect the donor's health and ensure product safety. Donor eligibility requirements for manufacture of Source Plasma are found in 21 CFR 640.63. Part VI of this compliance program lists guidance documents and memoranda recommending additional donor eligibility requirements for Source Plasma donation.

  1. Medical Supervision
  2. Source Plasma regulations require a qualified licensed physician be on the premises when donor eligibility is determined, immunizations are occurring, whole blood is being collected, and red blood cells are being returned to the donor. [21 CFR 640.62] A Source Plasma establishment may have an approved PS training program and may train an individual to perform some of the duties of a physician. Attachment H gives additional information about PS.

    Duties of PS often include the following: initial medical and physical examinations of new donors; annual physical examinations of repeat donors; evaluation of donor reactions and providing appropriate therapy as prescribed by the Source Plasma establishment's SOPs; performing immunizations (except Red Blood cell immunizations); counseling donors; and reviewing collection records and accumulated laboratory data to determine a donor's continued suitability for plasmapheresis.

    During the inspection,
    1. The physician must be qualified and licensed.
    2. If the establishment uses a PS, determine if
      1. The establishment has a CBER-approved PS program.
      2. The PS meets the criteria identified in the approved program.
      3. Each PS received training for the duties performed in each Source Plasma establishment.

  3. Obtaining Informed Consent
  4. The Source Plasma establishment must obtain written informed consent for Source Plasma collection after a qualified, licensed physician or PS has clearly explained the hazards of each procedure in which the prospective donor will participate: manual collection, automated plasmapheresis, immunization with an antigen, and participation in special collection programs approved by CBER; e.g., collection of Source Plasma from coagulant factor-deficient donors. Determine if the donor is given an opportunity to ask questions and to decline participation. [21 CFR 640.61]

    During the inspection,
    1. Review the informed consent process to determine if it contains a simple explanation of the plasmapheresis procedure and the risks involved, such as infiltration, infection, and loss of red blood cells.
    2. For automated operations, the informed consent may include the risk of possible anticoagulant reactions.
    3. For manual operations, the informed consent must include the risk of inadvertently receiving another donor’s red blood cells and the possibility of experiencing a hemolytic transfusion reaction. [21 CFR 640.61]
    4. For immunization programs that use licensed vaccines, the informed consent must explain the risks associated with receiving the vaccine. [21 CFR 640.61]
    5. The Source Plasma establishment should also inform donors immunized with red blood cells that they may develop atypical or unexpected red cell antibodies that may interfere with obtaining a compatible blood, organ or tissue transplant in the future.

    See the following guidance document for additional information:
    http://www.fda.gov/cber/bldmem/031495.txt
    http://www.fda.gov/cber/gdlns/infrmdcnsntplsm.pdf

  5. Donor Screening
  6. The Source Plasma establishment must implement a system that positively identifies all donors (usually a photograph) and that relates the donor directly to the products collected and to the donor's accumulated records and laboratory data. The method used should prevent conditions that allow a prospective donor to impersonate another person or donate when not eligible, e.g. missing or poor quality photos, duplicate files, or acceptance of a deferred donor under a different name or social security number.

    Firms may present donor screening questions to the donor by several methods. These include direct oral questioning of the donor by firm personnel and self-administered donor questionnaires, using either printed forms or by a computer-assisted interactive interview.

    In the self-administered computer-assisted interactive interview procedure, the donor reviews the questions on a computer screen and enters the answers electronically into the software program managing the interview process. The computer software may or may not make decisions on the suitability of the donors depending on the responses to the questions. The computer system used in the computer-assisted interactive interview procedure includes any hardware and software needed to perform the process. It may be a stand-alone system, used solely to conduct the donor interview, or it may interface with other computer systems at the same or other locations. It may be a desktop or laptop computer or a handheld device. The software may have data storage capabilities or may send data to a printer for hardcopy printout. In addition, the computer system may be accessible from a remote location. The user interface may present both video and audio data to the user via monitors, headphones, etc. Donors and collection personnel may input data or responses via keyboard, microphone, or a pointing device such as a mouse, touch screen, or stylus. The system may use pictures or drawings to illustrate the topic of the displayed questions. Also see Attachment F, "Computers."

    The Source Plasma manufacturer must determine donor eligibility, to include physical examinations, and medical screening, according to its SOPs and applicable FDA requirements. [21 CFR 640.63 and 640.65]

    1. Identify yourself to the donor and explain that observing the screening process is part of a routine inspection. Ask the donor's permission to observe the screening process and give the donor a clear opportunity to refuse. If the donor refuses, make the request to another donor.
    2. Note: If management questions FDA's authority to observe donor screening, explain to management that observing the screening process is part of conducting the inspection of a Source Plasma establishment in a manner that is reasonable under the circumstances and, therefore, authorized by law. Follow the procedures in IOM section 514 (Inspection Refusal) if management refuses to permit observation.

    3. The Source Plasma establishment should explain the donation process in a confidential manner so donors may make an intelligent and informed decision regarding whether to participate in the donation procedure. The process must provide the donor with an opportunity to refuse the procedure. The method should ensure comprehension of the information presented and confidentiality. Source Plasma establishment must have appropriate procedures if collecting Source Plasma from hearing or vision-impaired donors, from donors who speak English as a second language, or from donors who may have a reading difficulty.
    4. A third party; e.g., a translator may assist in the interview process. To ensure confidentiality and full disclosure of information by the donor, CBER recommends that Source Plasma establishment not use the donor's friends or relatives as the third party. This third party should understand the confidential nature of the information discussed and agree not to disclose it to anyone. The third party may not complete the questionnaire.
    5. The Source Plasma establishment must incorporate all screening procedures in its SOPs, including criteria for use of a third party. Donor records should indicate participation of a third party in the donor screening process.

      During the inspection,
      1. Observe one or more physicals.
      2. Determine if medical examinations and physicals are conducted according to the SOPs and applicable FDA requirements and at the proper intervals. [21 CFR 640.63(a), (b) ]
      3. Personnel should adequately respond to donor questions or refer questions to the appropriate medical personnel, as necessary.
      4. Review the Source Plasma establishment’s procedure for determining a donor’s eligibility on the day of donation for consistency with the requirements in 21 CFR 640.63 and current CBER recommendations regarding donor eligibility.
      5. If the computer-assisted interactive screening process is used, determine if SOPs describe the process and the computer system has been adequately validated for its intended use.
      6. Determine if the physician / PS reviews a total plasma or serum protein determination and serum protein electrophoresis or equivalent test to determine immunoglobulin composition, syphilis test results, and the donor’s accumulated laboratory data within 21 days of collection of the initial and 4-month test samples.
      7. The physician / PS must certify the donor’s eligibility for immunization and /or continued plasmapheresis on a form designated by Source Plasma establishment.

      8. The Source Plasma establishment should provide AIDS educational material, including information about high-risk activities, to donorsat each visit. The Source Plasma establishment should have and follow its SOPs pertaining to any unexplained donor weight loss.
      9. Determine if the Source Plasma establishment performs all required screening tests (temperature, pulse, blood pressure, total serum or plasma protein and hematocrit or hemoglobin) in determining donor eligibility. [21 CFR 640.63] The establishment’s SOPs may require additional tests not specifically required by FDA, such as periodic drug screening.
      10. Determine if the Source Plasma establishment calibrates and maintains all equipment used in donor screening according to the device manufacturer’s instructions and its SOPs and runs proper quality control according to the equipment manufacturer’s instructions.
      11. Determine if Source Plasma donors weight at least 110 pounds.
      12. The donor’s weight determines the amount of Source Plasma the manufacturer may collect in a manual procedure. In an automated procedure, CBER must specifically approve the collection volume.
      13. Review a sufficient number of records to determine if the Source Plasma establishment only collects Source Plasma from donors with acceptable health history and screening test results, unless the establishment has amended its BLA to collect Source Plasma under a special collection program(s).
      14. Determine if the Source Plasma establishment collects Source Plasma from infrequent donors no more often than once every 4 weeks. For additional discussion of infrequent plasmapheresis, see Attachment I.
      15. Review the Source Plasma establishment’s procedure to prevent cross donation.
      16. Cross donation occurs when (a) an individual donates at more than one Source Plasma establishment or blood collection facility in a geographic area concurrently, (b) donates at one establishment although permanently or temporarily deferred at another, or (3) donates at a frequency that would be injurious to the donor’s health.

        Review sufficient records to determine if the Source Plasma establishment’s procedures are adequate to identify donors and to prevent cross donation.

  7. Donor Eligibility for Special Collection Programs - Attachment I
  8. A manufacturer must have an approved BLA supplement to collect Source Plasma from special donor populations. Special collection programs include,

    • Pre-Existing Antibody Collection
    • Disease State and "High-Risk" Donor Collection
    • Immunization
    • Infrequent Plasma Collection

  9. Donor Eligibility for Source Leukocyte / Therapeutic Exchange Plasma- Attachment K
  10. Donor Deferral
  11. The Source Plasma establishment must have specific procedures to defer donors who are determined ineligible for Source Plasma collection due to (a) medical history, (b) physical examination, or (c) a positive screening test(s) for evidence of a communicable disease agent(s) identified in 21 CFR 610.40 (HIV-1 and 2, hepatitis B virus, hepatitis C virus and syphilis). [21 CFR 606.100(b)(20), 610.41] Note: Be aware that under special collection programs, manufacturers may collect Source Plasma from ineligible donors. See Attachment I.

    The Source Plasma manufacturer must keep a record of deferred or ineligible donors so that products from such donor are not distributed. [21 CFR 606.160(e)] Most Source Plasma manufacturers determine the deferral status of new or returning donors prior to collection.

    Manufacturers may ship Source Plasma collected prior to a donor's reactive syphilis serologic test result and donations determined to have a biologic false-positive syphilis test result. A donor with a reactive serologic test for syphilis shall not be plasmapheresed again until the donor's serum is tested and found to be non-reactive. [21 CFR 640.65]. See also, Compliance Policy Guide 7134.15, Section 245.100, "Use of Units from Donors Subsequently Found to be Reactive to a Serologic Test for Syphilis."

    During the inspection,
    1. Review the Source Plasma establishment’s procedures and criteria for donor deferral for compliance with 21 CFR 606.160(e) and 610.41.
    2. Review records and observe operations to determine if the Source Plasma establishment accurately records donor screening deferrals, testing deferrals, and post donation information.
    3. Determine if the Source Plasma establishment accurately enters, either electronically or manually records, all causes of temporary or permanent deferrals.
    4. The Source Plasma establishment should have procedures/computer programs to identify discrepant and/or duplicate donor information and procedures to prevent release of unsuitable products.
    5. Determine if the establishment appropriately corrects and/or merges discrepant or duplicate records according to its SOPs.
    6. Review records to determine if the Source Plasma establishment inappropriately released unsuitable Source Plasma.

  12. Notifying Ineligible Donors
  13. A Source Plasma establishment must have procedures to notify a Source Plasma donor of a change in donation status whenever the Source Plasma establishment determines that the donor is ineligible for future donations based on the requirements in 21 CFR 640.63 or the results of required testing for communicable disease agents in 21 CFR 610.40. [21 CFR 630.6] The SOPs must include the method for notifying the donor, including follow-up if the initial attempt at notification fails. [21 CFR 606.100(b)(20)] Note: The donor may meet requirements to participate in a special Source Plasma collection program.

    During the inspection,
    Review a sampling of records with repeat reactive test results to determine if the Source Plasma establishment performs supplemental testing and notification to donors as required in 21 CFR 630.6.

  14. Donor Re-Entry Algorithms
  15. A Source Plasma establishment may re-enter donors previously deferred because of a positive test result for a required communicable disease agent after it finds the donor is otherwise eligible by a re-qualification method or process acceptable to FDA and the donor is otherwise suitable. Most Source Plasma establishments, however, do not re-enter donors.

    The following blood memoranda identify acceptable procedures to re-qualify a previously deferred donor for further Source Plasma collection.

    During the inspection,
    Identify donors that the Source Plasma establishment re-entered and determine if the establishment performed donor re-entry according to the acceptable methods or processes identified in the above documents. [21 CFR 610.41]

ATTACHMENT C
PRODUCT TESTING SYSTEM

Testing of Source Plasma for evidence of communicable disease agents must be conducted correctly and monitored closely to ensure the quality and safety of products. Manufacturers must test each donation of Source Plasma for the following communicable disease agents as required under 21 CFR 610.40:

  • Human Immunodeficiency Virus, types 1 & 2
  • Hepatitis B Virus
  • Hepatitis C Virus

The manufacturer must also test donors for syphilis, total plasma or serum protein, and immunoglobulin composition of plasma or serum, initially and every four months. [21 CFR 640.65(b)]

Further testing of each reactive donation, using screening and supplemental tests approved by FDA for such use is also required. Source Plasma establishments may contract part or all of the communicable disease testing to an outside testing laboratory. The contract, testing laboratory must register with FDA and be certified by the Centers for Medicare and Medicaid Services (CMS) or has met equivalent requirements as determined by CMS to perform infectious disease testing. The Source Plasma establishment must ensure that the contract laboratory is registered with FDA and that laboratory testing complies with 21 CFR 610.40 (a), (b), (e) and (f).

The laboratory must perform required testing for communicable disease agents using screening and supplemental test kits FDA has approved for such use. Serological tests for syphilis should be labeled for use in donor screening. A list of currently licensed HIV and hepatitis test kits is on the Internet at http://www.fda.gov/cber/products/testkits.htm

A Source Plasma establishment that does its own testing for evidence of a communicable disease agent must retain testing records as required in 21 CFR 606.160(d). A manufacturer that sends such testing to an outside laboratory must have test results; i.e., reactive, nonreactive, positive, negative or indeterminant in written form (hard copy or available electronically) prior to releasing Source Plasma for further manufacture.

  1. Communicable Disease Testing - On-Site
  2. This section applies to inspection of Source Plasma establishments that perform on-site testing for required communicable disease agents. The following inspection guidance supplements the comprehensive, “Guide to Inspections of Infectious Disease Marker Testing Facilities,” June 1996. http://www.fda.gov/ora/inspect_ref/igs/infdis.html

    During the inspection
    1. Observe actual testing practices and procedures. Choose a time when personnel and/or supervisory oversight are at a minimum level. Determine if appropriate controls are used, that samples and controls are diluted properly, that the time and temperature of incubation are accurate and that instrument and equipment settings are correct during testing.
    2. Determine if the Source Plasma establishment performs equipment maintenance according to the manufacturer’s recommendations and the firm’s SOPs.
    3. Determine if all testing problems are adequately investigated, resolved, and documented.
    4. If unable to observe infectious disease testing, then at a minimum, compare the establishment’s test procedures with the test kit inserts, test equipment user manuals, and reagent inserts.
    5. Review the package inserts for the lot of test kits and reagents in current use instead of those on file. Investigate any noncompliance noted between inserts or manuals and the firm's procedures. Discuss any questions with CBER/ Division of Emerging Transfusion Transmitted Diseases (301-827-3008).

    6. Review as many required infectious disease test records as the inspection permits, extending the review as necessary depending on findings.
    7. Consider both the size of the firm and its compliance history. If possible, select records from a time period when problems are more likely to occur, such as holidays, on evening shift, at installation of new equipment, or when there is new management or personnel. Investigate unusual test results, such as low values and invalidated test results.

    8. Select a representative number of reactive test results for each required disease agent. Track the units from donor screening, product collection, donor deferral, product quarantine, storage, and disposition to determine appropriate handling of products and required recordkeeping.
    9. Observe procedures for handling samples and labeling processing trays. Assess whether the procedures are adequate to prevent sample mix-ups. The laboratory must store samples as specified in the test kit manufacturer's directions.
    10. Ensure that the sample requirements (anticoagulant, age of sample, quantity, storage temperature, especially if testing is delayed, etc.) are met.
    11. The Source Plasma establishment must qualify automated sampling equipment and positive identification systems to ensure proper identification of samples and test results.

    12. Evaluate the Source Plasma establishment’s laboratory quality control program. Determine if all laboratory equipment is qualified, calibrated and maintained as required by user manuals, maintenance manuals and the Source Plasma establishment’s standard operating procedures. [21 CFR 606.60]

  3. Invalidation of Test Results
  4. Evaluate the Source Plasma establishment's procedures for invalidating a test result for consistency with the recommendations in the document, “Guidance for Industry: Revised Recommendations Regarding Invalidation of Test Results of Licensed and 510(k) Cleared Bloodborne Pathogen Assays Used to Test Donors,” dated July 11, 2001. http://www.fda.gov/cber/gdlns/bldbrn.htm This guidance incorporates provisions of the Clinical Laboratory Improvement Act of 1988 (CLIA) for invalidation of test results based on CLIA external control requirements.

    Laboratories or facilities that do testing may invalidate a reactive test result ONLY IF the assay run in which a sample is tested either fails to meet test kit package insert acceptance criteria OR the firm failed to do testing according to the test kit instructions; e.g., using compromised reagents or faulty equipment. If test kit package insert instructions are met, but CLIA control requirements are not met, the laboratory may invalidate only non-reactive results, but MAY NOT invalidate any reactive results. If an initially reactive specimen tests reactive on one or both of the two repeat duplicate tests, the sample is reactive and the testing laboratory should manage the results as indicated in the guidance document. When a negative or non-reactive test result is legitimately invalidated, re-test the sample singly and that result, if valid, is the test of record.

    The testing facility should document all incidents of invalidation including:

    • The basis for invalidation
    • The details of an investigation
    • The outcome of the investigation, and
    • If indicated, any corrective action taken

    During the inspection,
    Review all records of invalidation of test results for consistency with test kit manufacturer's directions for use and CBER recommendation. Notify CBER/Division of Inspection and Surveillance at 301-827-6220 if the firm uses other criteria to invalidate test results and determine donor eligibility.

ATTACHMENT D QUARANTINE/ STORAGE/DISPOSITION SYSTEM

  1. Quarantine
  2. The Source Plasma establishment must have control of the Quarantine/Storage/Disposition System to prevent the distribution of any unsuitable product. The Source Plasma establishment must quarantine Source Plasma products that test reactive for required communicable diseases agents, that are awaiting additional more specific testing (21 CFR 606.40), that were collected under a high-risk collection program(s), or that meet the requirements for HIV lookback under 21 CFR 610.46.

    The Source Plasma establishment must also quarantine other Source Plasma products (nonreactive for evidence of a communicable disease agent) that it determines unsuitable for use.

    During the inspection,
    1. Examine records to determine if the Source Plasma manufacturer quarantined Source Plasma products appropriately.
    2. Evaluate the firm’s procedure for removing products from quarantine; e.g., returning product to inventory after performing additional testing.
    3. Determine if records identify the individual who removed products from quarantine, the date removed, and the reason for the removal. [21 CFR 606.160]
    4. Determine if after collection, the establishment immediately stores and maintains Source Plasma at the appropriate temperature and that the Source Plasma establishment documents the temperature. [21 CFR 610.53, 640.69, 640.74]
    5. Injectables: -20º C or colder
      Noninjectables: temperature appropriate for the intended use
      Source Plasma Liquid: 10º C or colder, unless otherwise approved by CBER

    6. If Source Plasma for injectable use was not stored or shipped at appropriate temperatures, determine if the Source Plasma establishment re-labeled the product “Source Plasma Salvaged” consistent with 21 CFR 640.76(a)(1) or (b), unless an exception under 21 CFR 640.76(a)(2) applied or CBER determined that no re-labeling was required.
    7. Review the establishment’s distribution records to determine traceability of all Source Plasma products and maintenance of records according to 21 CFR 606.165 and 640.72.
    8. Determine if the establishment releases Source Plasma for distribution only after it receives and reviews written or computerized test results for the Source Plasma products. [21 CFR 606.100(c)]
    9. Notes:

      • CBER approval is not required to ship products under quarantine, prior to completion of PCR or NAT testing, to other locations operating under the same license (Source Plasma establishments, fractionators, or off-site storage locations).
      • CBER approval is required to ship products prior to completion of PCR or NAT testing to Source Plasma establishments, fractionators, or off-site storage facilities operating under a different license.
      • CBER approval is required to ship products under quarantine pending PCR or NAT testing, to an independently owned, unlicensed, off-site storage location.

  3. Equipment
  4. All equipment used in Source Plasma manufacture must meet the requirements of 21 CFR 606.60.

    During the inspection,
    Determine if all storage and temperature monitoring equipment is calibrated and maintained per manufacturer's instructions.
    Note: After installation and qualification of a central temperature monitoring system, CBER may permit an alternate procedure from the daily comparison of the internal thermometer to the recording chart/device.

  5. Shipment
  6. The Source Plasma establishment shall ship Source Plasma at a temperature appropriate for manufacture of the final product. [21 CFR 600.15, 640.76(b)]

    • Injectable: at –5ºC or colder
    • Noninjectable: at 10ºC or colder or as indicated in an approved BLA supplement

    Source Plasma for injectable use kept under continuous temperature monitoring requires no inspection for evidence of thawing prior to issue, provided temperature records indicate appropriate storage temperatures at –20°C or colder. [21 CFR 640.69(c)]

    Source Plasma Liquid shall meet the requirements of 21 CFR 640.74. Prior to issue, the Source Plasma establishment must immediately inspect each container for abnormal color, physical appearance or indication of microbial contamination.

    The Source Plasma manufacturer shall re-label Source Plasma stored at unacceptable temperature as “Source Plasma Salvaged,” except as provided in 21 CFR 640.76(a)(2). The manufacturer must also maintain appropriate records identifying the units involved, their disposition and an explanation of the conditions that caused the unacceptable temperature exposure. [21 CFR 640.76(a)-(c), 640.70(b)]

    During the inspection,
    Determine if the Source Plasma establishment ships products according to regulations.

  7. Imported Blood and Blood Components
  8. During the inspection,
    1. Determine if the Source Plasma establishment received plasma products identified as "import for export." Contact HFM-610 to determine if CBER approved the 'import for export' shipment pursuant to section 801(d)(4) of the FD&C Act.
    2. Determine if the establishment received from outside the United States, returns of Source Plasma or rejected Source Plasma.

    For further information, request Compliance Program 7342.007, "Examination of Blood and Blood Components Offered for Import" from CBER/OCBQ/DCM, HFM-610, (301)-827-6201.


ATTACHMENT E
PRODUCT COLLECTION AND PROCESSING SYSTEM

This system covers the Source Plasma establishment operations from collection through processing/labeling.

CBER approves a Source Plasma establishment to collect Source Plasma by manual or automated apheresis methods under various Source Plasma collection programs. The Source Plasma establishment may collect Source Plasma from donors only twice in a 7-day period, and at a 2-day interval. Donors who participate in an infrequent plasmapheresis collection program may donate no more frequently than once every four weeks. Infrequent plasmapheresis donors should meet Whole Blood donor suitability requirements and weigh at least 110 pounds. Currently, CBER approves infrequent plasmapheresis programs as a variance under 21 CFR 640.120

The collection procedures must ensure that the appropriate volume of Source Plasma is collected and that the maximum feasible volume of red blood cells is returned to the donor. [21 CFR 640.65] Overbleeding donors during manual collection and/or failure to return red blood cells due to technical difficulties during automated plasmapheresis may require temporary deferral of donors. [21 CFR 640.63]

The Source Plasma establishment must describe its collection method(s) in SOPs. Source Plasma collection must meet the requirements in 21 CFR 640.64 and 640.65.

The Source Plasma establishment must immediately store and maintain Source Plasma consistent with its intended use. It must store injectable products at - 20ºC or colder (except Source Plasma Liquid) and noninjectable products, at a temperature appropriate for the final product. [21 CFR 640.69(b), (c) and 640.70(a)(3), (11)(b)]

For additional information regarding Source Plasma collection, consult the following document at http://www.fda.gov/ora/inspect_ref/igs/Source_Plasma/default.htm.

  1. Venipuncture
  2. The Source Plasma establishment personnel must prepare the skin at the site of phlebotomy thoroughly and carefully by a method that gives maximum assurance of a sterile container of blood. [21 CFR 640.64(e)]

    During the inspection,
    Observe several phlebotomists as they prepare the venipuncture site. Determine if each phlebotomist prepares the venipuncture site consistent with the establishment's SOPs. Critical steps in preparing the skin for venipuncture include:
    • Sufficient time and vigor of scrubbing - key factors in removing superficial microbes.
    • Applying a bactericidal agent according to the reagent manufacturer's instructions.
    • No touching of the prepared area with fingers or other non-sterile object, including donors bending their arms.

    Acceptable procedures for preparing the venipuncture site are on the Internet at http:/www.fda.gov/cber/blood/armpreprev.htm

  3. Collection Methods
  4. CBER approves both manual and automated SOPs for collection of Source Plasma. Manufacturers should have a procedure to identify and prevent overbleeds. Overbleeding occurs when the amount of Source Plasma exceeds what may be collected at one time from a donor or the donor donates more frequently than twice in a 7-day period

    1. Automated Collection
    2. CBER has cleared several fully automated, stand-alone or concurrent plasma collection systems for plasma collection. Devices cleared to collect plasma products as a by-product of plateletpheresis or red blood cell apheresis are used in blood bank or blood center operations. Most Source Plasma establishments use the following stand-alone devices:

      • Baxter Healthcare Corp. - Autopheresis-C Plasmapheresis System
      • Haemonetics - PCS Plasma Collection System
      • COBE/Gambro - Trima

      During the inspection,
      1. Review the SOPs for automated collection of Source Plasma. Determine if the Source Plasma establishment collects Source Plasma according to its approved procedures and the collection device manufacturer’s instructions.
      2. Observe Source Plasma collection to assess the adequacy of employee training. The staff should be able to explain error messages and take appropriate action. Note: CBER recommends the following operator to device ratio: 1 trained operator may operate 6 devices and 1 trainee operator may operate 4 devices under the supervision of a trained operator. The trainee’s 4 devices should be included within the trained operator’s 6 devices so that the trained operator does not exceed the number of devices that the operator may safely oversee.
      3. Become familiar with device safety alarms. Determine if employees do not override or bypass the alarms without taking corrective action as indicated in the device manual(s).
      4. Review each collection device record or log to identify any problems with the device. The record or log should include all warning alarms and problems in returning red blood cells. The record or log often identifies problems with disposable collection sets.
      5. Determine if the Source Plasma establishment performs and records routine maintenance according to the device manufacturer’s instructions.
      6. Determine if the Source Plasma establishment has procedures to ensure that collection devices operate properly after software changes and following repairs. [21 CFR 606.60] Note: Computer software in collection devices can frequently be changed using manufacturer upgrades.
      7. Review donor record files to determine if the Source Plasma establishment collects the appropriate volume of plasma specifically approved for the device. [21 CFR 640.65]
      8. The establishment may use a device-specific approved nomogram or plan to determine the amount of plasma to collect.

        A nomogram may use several criteria to determine the collection volume; e.g., the donor’s gender, weight, height, and hematocrit. CBER developed a simplified nomogram that determines the maximum collection volume or weight of plasma based only on the weight of the donor. Note: CBER recommended that use of the simplified nomogram, precludes use of any other nomogram in the Source Plasma establishment, including those provided by the device manufacturer. Consult the guidance document at http://www.fda.gov/cber/bldmem/110492.txt

        The Source Plasma establishment should have a procedure to track red blood cell loss. Determine if the employees appropriately defer donors who failed to have their red blood cells returned, unless exempt under 21 CFR 640.63. Note: Any person who has donated one unit or more of Whole Blood (or who has lost the equivalent amount of red blood cells due to technical difficulties during an automated plasmapheresis procedure) must not serve as a donor of Source Plasma for 8 weeks. [21 CFR 640.63(e)]

      For additional information, consult the guidance documents at the following websites:

      http://www.fda.gov/cber/bldmem/mem120495.pdf
      http://www.fda.gov/cber/bldmem/110492.txt

    3. Manual Collection
    4. A Source Plasma establishment that uses a manual method must collect and process Whole Blood for Source Plasma according to the requirements in 640.65(b)(4) - (7), 640.68 and the establishment's approved SOPs. The SOPs should describe in detail the collection procedures, including:

      • Donors should participate in the procedure for returning red blood cells.
      • The Source Plasma establishment should adequately mix the contents of the collection bag during collection.

      During the inspection,
      1. Review donor record files to determine if the interval between donations is consistent with regulations. [21 CFR 640.65]
      2. Review the Source Plasma establishment’s procedures for returning the maximum amount of red blood cells to the donor. [21 CFR 640.66(b)(7)] Observe as many employees as possible collect Source Plasma to ensure that staff is following the establishment’s procedures.
      3. Determine if double bagging (collection of the second bag of whole blood prior to the return of the red blood cells from the first bag collected) does not occur. [21 CFR 640.65(b)(6)]
      4. Determine if the Source Plasma establishment has a procedure to identify and prevent overbleeding; e.g., monitoring scales after adjustments or repairs, as necessary. Note: Review Whole Blood weight records to determine the number of overbleeds by volume per day.
      5. Determine if appropriate procedures exist for plasma pooling to prevent cross-pooling. [21 CFR 640.69]
      6. Investigate any incidents of incorrect red blood cell infusion. Any incident of incorrect red blood cell infusion is a serious departure from the Source Plasma establishment’s SOPs.
      7. The Source Plasma establishment should have procedures to provide the donor who received the incorrect red blood cells appropriate emergency medical attention.

      8. Determine if the Source Plasma establishment’s procedures properly defer each donor involved in the incident. The manufacturer must defer donors who did not receive their own red blood cells, from further collection for eight weeks. The manufacturer must also defer donors who received incorrect cells, for 1 year because those donors received a blood transfusion.

  5. Labeling
  6. Product labels must meet the requirements of 21 CFR 640.70 and 640.74. CBER reviews most product labels prior to use. The manufacturer, however, may submit labels for collecting disease-associated Source Plasma in its annual report of minor changes.

  7. Source Plasma Collection Programs (See Attachment I)
    • Immunization Program
    • Pre-existing Antibody Collection Program
    • "High-Risk" or Disease State Donor Collection Program
    • Infrequent Plasmapheresis Collection Program

  8. Source Plasma Collection Programs (See Attachment K)
    • Source Plasma and Therapeutic Exchange Plasma

ATTACHMENT F
COMPUTERS

Blood establishments may use computer systems for a variety of purposes within the operation. They may utilize a complex, integrated computer system with a donor software module or they may use a single software module. Computerized operations may include:

  • Storing, updating, and accessing donor history information, donor deferral records and distribution records.
  • Accepting, storing, and interpreting test results. Results may be entered manually or by electronic file transmission from the test instrument or laboratory data management system.
  • Releasing Source Plasma for distribution.

Determine which operations are computerized and how the user validated the computer system on-site to demonstrate that it performs the intended functions accurately and reliably.

  1. Requirements for Source Plasma Establishment Computer Software
  2. All software, including software developed in-house, used to manufacture blood and blood components (including Source Plasma), to maintain data for making decisions about donor (suitability) eligibility, or to release products for further manufacture are medical devices under Section 201(h) of the FD&C Act. The device provisions such as: registration as a device manufacturer, product listing, medical device reporting, compliance with the quality system regulation, and pre-market notification 510(k) or application, apply to the device software manufacturer. Only blood establishment computer software that is 510(k) cleared or has pre-market approval (PMA) should enter interstate commerce. FDA has previously advised blood establishments to transition to a cleared software device.

    Source Plasma establishments that developed software for their own use and that did not ship it interstate or access or transmit any data across state lines are still medical device manufacturers and must register as a device manufacturer and list the products they manufacture. The device is subject to the Quality System Regulation requirements (21 CFR 820).

    Computer software developed by a Source Plasma establishment, but that is used by outside consignees (not under the same license number), constitutes commercial distribution and interstate commerce, and requires a 510(k). Also use of software/data by sites under the same license number may require submission of a 510(k) if the software / data crosses state lines.

    The actual use of software in Source Plasma establishments is subject to the CGMP regulations for Blood and Blood Components and the CGMP regulations for Finished Pharmaceuticals. Establishments are required to perform user validation to ensure that the software meets its intended use.

    During the inspection,
    Determine if the Source Plasma establishment uses only 510(k) cleared software. Contact CBER / OBRR / DBA / Devices Review Branch, HFM-390, for guidance regarding computer software. A list of 510(k) cleared blood bank software is posted at http://www.fda.gov/cber/products/510ksoft.htm

  3. Inspection of Source Plasma Establishments that are also Medical Device Manufacturers
  4. Use the following Compliance Program and medical device reporting codes when conducting inspections of Source Plasma establishments that also manufacture medical devices such as computer software :

    Compliance Program Guidance Manual, Inspection of Medical Device Manufacturers - 7382.845
    Establishment Type - MW
    Product Code - 81M
    PAC Codes: 42845A - Level 1; 42845B - Level 2; 42845C - Level 3 inspections

  5. Programs and Computer Functions to Include in the Inspection
  6. Criteria to consider when deciding which functions of the system to inspect:

    1. The criticality of the functions controlled by the computer, (e.g. determination of product suitability for release)
    2. Computer problems revealed by reviewing computer problem reports and biologic product deviation reports
    3. Areas suggested for inspection after reviewing computer system change control records.

    During the inspection,
    1. Review the operator’s manual.
    2. Observe the use of the computer system. Observe manual data input, screen messages, error checking, etc.
    3. Review the overall validation plan and procedures and the validation of critical programs and reports critical to the Plasmapheresis establishment’s operations. Often validation is conducted at the corporate location. User validation of Source Plasma establishment software is required by 21 CFR 211.68(b).
    4. Determine if the Source Plasma establishment validated the system prior to implementation. Determine if the establishment followed the manufacturers’ instructions regarding installation and validation of upgrades.
    5. Be alert to user customization of vendor supplied software systems. Customization is normally accomplished by a user setting certain parameters which affect how the software functions. Check the vendor’s recommended configuration and review the validation of all deviations from the vendor recommended parameters.
    6. Determine if the Source Plasma establishment includes changes to software under written change control procedures and if it documents changes to the system, including the potential impact the changes will have on the system. The establishment should document the change (who made the change, who authorized the change, and the effective date of the change).
    7. Review the use of “work-arounds. Establishments may implement work-arounds when the system does not perform exactly the way the user requires and the software vendor recommends and/or the user develops procedures to circumvent the system’s limitation. Determine the reason the work-around was created, whether it adequately addressed the situation and whether the work-around created any other problems.
    8. Determine if the Source Plasma establishment monitors the functioning of the computerized system for errors and if it documents them and assesses their impact on operations and/or records.
    9. The Source Plasma establishment should have written procedures for continuing operations when the computer system is not functional, in addition to procedures for data and system recovery in the case of system failure. It should periodically back up data and systems files and store them in a secure location.
    10. Review the Source Plasma establishment’s written policies for computer security and determine if the firm follows them. Source Plasma establishments that maintain electronic records must maintain the integrity of those records as required by 21 CFR Part 11. System access must be controlled to limit access to only authorized individuals. If a record is changed, the previous data, the person making the change, and the time the change was made must be documented electronically.

ATTACHMENT G
LOOKBACK

FDA regulations require lookback for HIV. [21 CFR 610.46] The lookback process addresses prior donations collected from donors who subsequently test HIV positive and requires that Source Plasma establishments perform the following:

  • Quarantine in-house products previously collected from the donor.

  • Note: Products may be stored at an offsite storage location or at a fractionation plant, but are not released for use. The Source Plasma establishment's procedures should define notification, quarantine, culling, and the final disposition of those products.
  • Notify consignees to quarantine products in their inventory that are not pooled or further processed.
  • Notify consignees of the results of further communicable disease testing.

During the inspection,
  1. Review the SOPs to determine that lookback procedures comply with regulations.
  2. Determine if the Source Plasma establishment notifies consignees to quarantine Source Plasma, Source Leukocytes and Therapeutic Exchange Plasma collected within 6 months prior to the positive test result if those products are not pooled or further processed, unless exempt under 21 CFR 610.46(c).
  3. Note: Source Leukocytes are processed within 24 hours of collection. Source Plasma is usually pooled within 6 months of collection.

  4. Determine if the Source Plasma establishment investigated any deviations in testing procedures or donor deferral. If the event(s) were reportable, determine if the manufacturer notified CBER.

FDA issued guidance documents describing lookback recommendations at products reactive for hepatitis B virus, hepatitis C virus, hepatitis B core and Human T-Lymphotropic virus. Refer to CBER website for current guidance documents (http://www.fda.gov/cber/guidelines.htm).


ATTACHMENT H
PHYSICIAN SUBSTITUTES

A Source Plasma establishment may supplement its biologics license to request an alternative procedure under 21 CFR 640.120 to use an adequately trained PS to perform some of the routine functions of a physician in the establishment. The educational requirements and job description or statement of responsibilities are part of the PS Training Program. After CBER approval, the manufacturer may train individuals to function as PS in the Source Plasma establishment.

In general, the PS:

  • Should be a graduate of a recognized educational program (nursing, emergency medical technician or physician assistant) and currently licensed or certified in the state where employed. The PS should maintain current certification in cardiopulmonary resuscitation.
  • Should complete the Source Plasma establishment’s training program and be determined competent by the establishment’s medical director before assuming the PS duties.
  • Should have a periodic evaluation of performance of assigned duties in the Source Plasma establishment performed by the medical director.
  • May evaluate, (1) normal healthy donors for both manual and automated apheresis procedures, and (2) donors in pre-existing disease and non-disease associated collection programs. These donors meet the eligibility criteria for Source Plasma collection.
  • With additional training and under the direction of the medical director, may administer or supervise approved immunizations, except red blood cell immunizations.
  • May administer Red Blood Cells, as an immunogen, but the physician must be on the premises.

Currently, a PS’s responsibilities may not include involvement in “high-risk,” and disease-state Source Plasma collection programs. See Attachment I for more information about disease-state and disease-associated Source Plasma collection programs.

Consult the following guidance documents for additional information.
http://www.fda.gov/cber/bldmem/081588.txt

During the inspection,
Determine if all PS in the Source Plasma establishment received training consistent with the approved training program for the duties the PS performs and meets the education and certification requirements of the program.

ATTACHMENT I
SOURCE PLASMA COLLECTION PROGRAMS

This attachment contains recommendations related to the review of BLA submissions to manufacture various Source Plasma products. To obtain additional information about these collection programs, consult the guidance documents as listed or for various program approvals, contact CBER, Division of Blood Applications, 301-827-3543.
  1. Pre-existing Antibody Collection Program
  2. CBER may approve a Source Plasma establishment to collect Source Plasma from donors who have a pre-existing antibody; e.g., antibody to Duffy red blood cell antigen (anti-Fya) or antibody to human leukocytes. Donors must meet all Source Plasma eligibility criteria.

    During the inspection,
    Determine if the Source Plasma establishment has notified CBER of various pre-existing antibody collection programs. The firm may submit those changes in its annual report of minor changes.

  3. Pre-existing Disease-Associated Collection Program
  4. A Source Plasma establishment may collect Source Plasma from donors who have pre-existing, disease-associated antibodies because of a previous exposure to certain diseases or cellular antigens; e.g., IgG antibody to Cytomegalovirus or anti-hepatitis A virus. Donors must meet all Source Plasma donor eligibility criteria. The manufacturer should inform donors that their participation in a special collection program depends on the level of antibody. A donor may immediately return to Source Plasma collection if the Source Plasma establishment no longer desires to collect the antibody. The manufacturer should notify CBER of the implementation of such programs in the establishment’s annual report of minor changes.

    Consult the Guide to Inspections of Source Plasma Establishments for a partial listing of disease-associated antibodiesand the “Draft Reviewers’ Guide for Disease Associated Antibody Collection Program,” October 1995.

    http://www.fda.gov/ora/inspect_ref/igs/Source_Plasma/default.htm
    http:/www.gov/cber/gdlns/antbdyclcprgm.pdf

    During the inspection,
    Determine if the Source Plasma establishment collects Source Plasma from donors with pre-existing antibodies according to its SOPs.

  5. Disease-State Collection Program
  6. This program allows Source Plasma collection from donors who may not meet all Source Plasma eligibility requirements. These donors are generally feeling well and are not experiencing any active symptoms on the day of donation. The disease conditions under this program require physician's authorization for collection. Some examples of disease state collections are antibody to Lyme disease or antibody to coagulation factors. The plasma is used for further manufacturing into in vitro diagnostic reagents.

    Prior to implementing a disease-state collection program, the manufacturer must submit a supplement to the BLA. The supplement should include SOPs that define the donor selection criteria, labeling, quarantine procedures, and Source Plasma disposition.

    The manufacturer should collect, handle, store, and distribute reagents, samples, and Source Plasma according to current biosafety guidelines established by FDA, Centers for Disease Control and Prevention (CDC) and /or Occupational Safety and Health Administration (OSHA).

    During the inspection,
    Determine if the Source Plasma establishment follows its approved procedures and collects Source Plasma only from donors for which it has CBER approval.

  7. “High-Risk” Donor Collection Program
  8. This program allows Source Plasma establishments to collect Source Plasma from donors who have a positive test result for a communicable disease agent. The product may be used in research, or for in vitro tests, or development of therapeutic products. CBER pre-approves the SOPs and labeling for this program. Product collection, handling, storage, and disposition of samples and Source Plasma should be in accordance with current biosafety guidelines established by FDA, CDC and /or OSHA. For additional information, consult the following guidance documents.

    http://www.fda.gov/cber/bldmem/041791.txt
    http://www.fda.gov/cber/bldmem/102689.txt

    During the inspection,
    Review SOPs and determine if the Source Plasma establishment collects product according to its approved SOPs and only from "high-risk" donors for which it has CBER approval.

  9. Immunization Programs
  10. A Source Plasma manufacturer may immunize donors using licensed products; e.g., tetanus or rabies vaccines, for collection of high titer antibody. If the program is consistent with the vaccine insert instructions, the manufacturer may supplement its BLA and distribute product within 30 days of CBER notification. Immunizations using red blood cells from a source approved by CBER in a BLA license application or supplement requires an inspection prior to use. Immunizations using unlicensed vaccines are conducted under an Investigational New Drug application. An inspection of a licensed firm prior to vaccine use, however, is usually not conducted.

    During the inspection,
    1. Determine if the Source Plasma establishment has CBER approval for each immunization program, as required.
    2. Determine if Source Plasma donors meet eligibility requirements in 21 CFR 640.63 and that the immunization complies with 21 CFR 640.66.
    3. Determine if the Source Plasma establishment’s process for obtaining consent for immunization informs donors of the hazards of immunization appropriate to the immunizing agent used.
    4. Note : CBER recommends that only males or females who are incapable of bearing children participate in red blood cell immunization programs.

      CBER recommends that donors not participate in more than one immunization program at a time.

    5. For red blood cell immunizations, determine if only a licensed qualified physician selects and schedules the antigen injection and evaluates the donor’s clinical response. [21 CFR 640.66]
    6. A physician must be on the premises when immunizations are made. A PS or other trained individual under a physician’s supervision may administer immunizing agents.

      Donors who participate in a red blood cell immunization program should receive required and CBER recommended testing, initially and at periodic intervals.

      An immunized donor may return to normal Source Plasma collection if the donor fails to meet the titer requirement of the immunization program.

    7. Determine if the Source Plasma establishment uses only approved antigens or immunizing substances and that it handles and stores them appropriately.
    8. Promptly notify CBER, Division of Inspections and Surveillance (HFM‑650) at 301-827-6220 if a Source Plasma manufacturer uses red blood cells that were not qualified.

    9. If the Source Plasma establishment prepares red blood cells for immunization, review all Whole Blood donor and recipient / Source Plasma donor manufacturing records.
    10. Review the system for tracking red blood cells from the Whole Blood donor to the red blood cell recipient / Source Plasma donor. Usually a lot numbering system is devised.
    11. Review both the Whole Blood donor and recipient / Source Plasma donor communicable disease test records for any reactive or positive test results.

    Consult the following guidance documents for additional information.

    http://www.fda.gov/cber/bldmem/031495.txt
    http://www.fda.gov/cber/gdlns/imnztnsrcpl.pdf
    http://www.fda.gov/cber/bldmem/070788.txt

  11. Infrequent Plasmapheresis Collection Program
  12. A Source Plasma establishment may supplement its BLA to include Source Plasma collection from non-immunized donors, who meet whole blood donor suitability requirements, other than donation frequency, every four weeks or less frequently. The donor must weigh a minimum of 110 lbs. Note: If infrequent donors donate no more frequently than every four weeks, they do not require a physical examination or a total plasma or serum protein or immunoglobulin composition. The maximum annual number of Source Plasma collections from a donor and the volume collected at each donation are described in the following guidance documents:

    http://www.fda.gov/cber/bldmem/031095.txt
    http://www.fda.gov/cber/bldmem/110492.txt

    A donor should not participate simultaneously in other blood or plasma collection programs and should not be a frequent apheresis donor. If an infrequent donor returns for donation in less than 4 week or donates more that the maximum annual volume of Source Plasma, the Source Plasma establishment should follow all Source Plasma donor eligibility requirements, including medical examination and plasma or serum protein tests prior to considering the donor eligible for another donation. [21 CFR 640.63]

    During the inspection,
    Determine if the Source Plasma establishment is following its CBER-approved SOPs.

ATTACHMENT J
TYPES OF BLOOD ESTABLISHMENTS COVERED UNDER THIS PROGRAM

The blood establishments listed below must register with CBER and list each product that they manufacture as required by 21 CFR 607. Access the CBER Intranet to query the Blood Establishment Registration database to review registration information for active, inactive and pre-registered establishments. For additional information, consult the "Instruction for Completing the Blood Establishment Registration Form 2830," on the FDA Intranet, Forms Catalog.

  1. Source Plasma Establishment
  2. This is a facility licensed to collect Source Plasma, as defined in 21 CFR 640.60, for commercial distribution. There may be multiple Source Plasma establishments (locations) operating under a single license. The Source Plasma establishment may also collect blood and blood components for further manufacture, e.g., Source Leukocytes, Therapeutic Exchange Plasma, and/or Red Blood Cells.

  3. Plasma Broker
  4. An establishment or person that arranges the sale of Source Plasma between other entities is a broker.

    A broker that only arranges the sale or shipment of products is not required to register, but must keep appropriate records of the activities performed.

    A broker that takes possession of Source Plasma and/or engages in any manufacturing step (e.g., pooling or re-labeling products, or making aliquots) must register.

  5. Testing Laboratory
  6. A laboratory that does testing for a Source Plasma establishment; i.e., (1) required testing for evidence of a communicable disease agents, (2) donor eligibility testing including testing for re-entry, and (3) testing to support labeling claims related to product quality must register with FDA and be either certified by the Centers for Medicare and Medicaid Services (CMS) to perform such testing, or have met equivalent requirements as determined by CMS. [21 CFR 610.40]

    Note: A testing laboratory that only performs syphilis confirmatory tests is not required to register. Such testing is not required.

  7. Contractor
  8. Any person or entity that performs part or all of the steps in the manufacture of a licensed product or that performs a service for a Source Plasma manufacturer must register.

  9. Off-Site Storage Facility
  10. An off-site storage facility that performs manufacturing operations, such as culling and quarantining Source Plasma products prepared for distribution, repackaging or relabeling product, and maintaining records of those operations must register.

    An off-site facility that only stores Source Plasma under specific controlled conditions, prior to shipment to a final user(s), e.g., temporary storage pending approval of a license application / supplement and distribution, is not required to register.

  11. Other Blood Establishments
  12. A licensed blood bank may manufacture Source Plasma, Red Blood Cells, Therapeutic Exchange Plasma, and/or Source Leukocytes for further manufacture. See also Attachment K for information on Source Leukocytes and Therapeutic Exchange Plasma.


ATTACHMENT K
SOURCE LEUKOCYTES AND THERAPEUTIC EXCHANGE PLASMA

A manufacturer, blood bank, blood center, or Source Plasma establishment may collect Source Leukocytes or Therapeutic Exchange Plasma (TEP) for further manufacture. Collection of these products is subject to the licensure provisions of Section 351 (a) of the Public Health Service Act. Manufacturers must test each product for evidence of communicable disease agents as required in 21 CFR 610.40. Lookback requirements and recommendations also apply to these products.

Inspect the manufacture of Source Leukocytes and Therapeutic Exchange Plasma under this Attachment and Compliance Program 7342.001, Inspection of Licensed and Unlicensed Blood Banks, Brokers, Reference Laboratories, and Contractors.

  1. Source Leukocytes Collection Program
  2. A blood bank, blood center or a Source Plasma manufacturer may collect Source Leukocytes as a by-product of Whole Blood collection, by manual plasmapheresis or automated apheresis. Donors must meet the eligibility requirements for Whole Blood donation or Source Plasma donation, as appropriate to the collection method. [21 CFR 640.3, 640.63] Often, manufacturers ship Source Leukocytes prior to completing the testing required by 21 CFR 610.40, but must send consignees test results when testing is completed. CBER approves the standard operating procedures and labels for Source Leukocyte manufacture. Collection methods include the following:

    1. By-product of Whole Blood Collection.
      • Donors must meet Whole Blood donor eligibility requirements. [21 CFR 640.3]
      • Collection frequency - no more frequent than once every eight weeks.

    2. Manual Apheresis
      1. Single unit as a by-product of manual apheresis with no additional monitoring of the donor
      2. (1) Donors must meet Source Plasma donor eligibility requirements.
              [21 CFR 640.63]
        (2) Collection frequency - no more frequent than once every eight weeks
        (3) Manufacturers may collect Source Leukocytes only from the first unit of the 2-unit plasmapheresis collection.

      3. Single unit as a by-product of manual apheresis with additional monitoring of donor
      4. (1) Donors must meet Source Plasma donor eligibility requirements
              [21 CFR 640.63]
        (2) Collection frequency:
        (a) No more frequently than once in 48 hours or twice in a 7-day period
        (b) Total Source Leukocytes donations in one year should not exceed 32 units.

      5. Donors should have a white blood cell count of > 4000 per cubic millimeter on a blood sample tested within 7 days prior to each collection.
      6. The manufacturer may collect Source Leukocytes from one or both units of whole blood in a 2-unit plasmapheresis collection.

    3. Automated Apheresis
    4. Currently, only the Haemonetics PCS-2 is approved for automated collection of Source Leukocytes. Licensing criteria for collection of Source Leukocytes using the automated device include, but are not limited, to the following:

      • Donor must meet Source Plasma donor eligibility criteria. [21 CFR 640.63]
      • Collection frequency - should be no more frequent than once in a 7-day period and no more than 16 collections from a donor in one year.
      • Donor should have a white blood cell count of > 4000 per cubic millimeter on a blood sample tested within 7 days prior to each collection.

      Consult the following CBER guidance for additional information.
      http://www.fda.gov/cber/bldmem/012881.pdf

      During the inspection,
      1. Review the establishment’s SOPs.
      2. Determine if the firm is following its approved procedures for Source Leukocyte manufacture.

  3. Therapeutic Exchange Plasma (TEP) Collection Program
  4. Therapeutic plasmapheresis is a medical procedure for treatment of certain diseases and is carried out under a physician's order. The manufacturer removes plasma incrementally and infuses other fluids to replace the plasma. The plasma derived from these procedures is limited to the manufacture of specific in vitro diagnostic reagents for which there are no alternate source materials.

    During the inspection,
    1. 1Determine if personnel follow the establishment's CBER-approved procedures, particularly criteria for selection of donors and the initial quarantine of product until tested for agents of communicable diseases as required by 21 CFR 610.40.
    2. Donors may have a disorder that is transmissible or that is of unknown etiology. TEP is potentially hazardous; therefore, many facilities choose to quarantine it under lock and key.
    3. Determine if disposition records indicate the disposition or destruction of each container of TEP collected.

    Refer to the CBER guidance at http://www.fda.gov/cber/bldmem/121484.txt for additional Information.

 
Updated: June 25, 2008