Market Withdrawal of LAL Reagent

DATE WITHDRAWAL INITIATED:

March 9, 2007

PRODUCT / LOT NUMBER / EXPIRATION DATE:

Pyrotell Multitest

Catalog Number: G2006 Lot Number: 206-09-395 Expiration Date: 07 Sep 2011
Catalog Number: G5006 Lot Number: 506-09-396 Expiration Date; 11 Sep 2011

MANUFACTURER:

Associate of Cape Cod
East Falmouth, MA

REASON:

This product was released with a labeled sensitivity of 0.06 EU/mL. Some samples recently tested from this lot have shown a change in sensitivity. The specifications for this product permit a twofold greater or lesser sensitivity than labeled claim. Recent results indicate that the lot is intermittently giving an endpoint of 0.25 EU/mL, a four-fold difference from label claim.


The information in this listing reflects CBER's best efforts to communicate information that has been reported to FDA. Its accuracy and comprehensiveness cannot be guaranteed.
 
Updated: March 15, 2007