Market Withdrawal of LAL Reagent
DATE WITHDRAWAL INITIATED:
March 9, 2007
PRODUCT / LOT NUMBER / EXPIRATION DATE:
Pyrotell Multitest
Catalog Number: G2006 Lot Number: 206-09-395 Expiration Date: 07 Sep 2011
Catalog Number: G5006 Lot Number: 506-09-396 Expiration Date; 11 Sep 2011
MANUFACTURER:
Associate of Cape Cod
East Falmouth, MA
REASON:
This product was released with a labeled sensitivity of 0.06 EU/mL. Some samples recently tested from this lot have shown a change in sensitivity. The specifications for this product permit a twofold greater or lesser sensitivity than labeled claim. Recent results indicate that the lot is intermittently giving an endpoint of 0.25 EU/mL, a four-fold difference from label claim.
The information in this listing reflects CBER's best efforts to communicate information that has been reported to FDA. Its accuracy and comprehensiveness cannot be guaranteed.