TETRACHLOROETHYLENE TRICHLOROETHYLENE
Method number: | 1001 |
|
Matrix: | Air | | |
| TWA
| Ceiling (5 min)
| Peak
|
Target concentration: |
Tetrachloroethylene Trichloroethylene |
100 ppm (678 mg/m3) 100 ppm (537 mg/m3) |
200 ppm (1356 mg/m3) 200 ppm (1074 mg/m3) |
300 ppm (2034 mg/m3) 300 ppm (1612 mg/m3) |
OSHA PEL: |
Tetrachloroethylene Trichloroethylene |
100 ppm (678 mg/m3) 100 ppm (537 mg/m3) |
200 ppm (1356 mg/m3) 200 ppm (1074 mg/m3) |
300 ppm (2034 mg/m3) 300 ppm (1612 mg/m3) |
ACGIH TLV: |
Tetrachloroethylene Trichloroethylene |
25 ppm (169 mg/m3) 50 ppm (269 mg/m3) |
100 ppm (678 mg/m3)(STEL) 100 ppm (537 mg/m3)(STEL) |
none none |
|
Procedure: | Adsorbent tube samples are collected by drawing workplace air through coconut shell
charcoal tubes using personal sampling pumps. Diffusive samples are collected by
exposing SKC 575-002 Passive Samplers to workplace air. Samples are extracted
with carbon disulfide and the extracts are analyzed by gas chromatography using
flame ionization detection. |
|
Recommended sampling parameters: |
Charcoal tubes
Sampling rate: Sampling time: |
TWA Samples
50 mL/min < 240 min |
Ceiling Samples
50 mL/min > 5 min |
Peak Samples
50 mL/min > 1 min |
SKC 575-002 Samplers
Exposure time: |
< 240 min |
> 5 min |
> 5 min |
|
| Charcoal Tubes
| SKC 575-002 Samplers
|
Reliable quantitation limit: (240-min samples) | | |
Tetrachloroethylene Trichloroethylene |
14 ppb (98 µg/m3) 13 ppb (68 µg/m3) |
104 ppb (705 µg/m3) 61 ppb (325 µg/m3) |
Standard error of estimate at: |
Tetrachloroethylene
|
Trichloroethylene
|
TWA: (240 min) Ceiling: (5 min) Peak: (1 min) |
Char. Tubes
5.1% 5.0% 5.2% |
SKC 575-002
9.0%* 9.6%* ---- |
Char. Tubes
5.1% 5.0% 5.1% |
SKC 575-002
9.1%* 9.1%* ---- |
*For samples where sampling site atmospheric pressure and temperature are known. When either or both of
these values are unknown, see Table 4.4.2 for applicable standard errors of estimate. |
|
Status of method: |
Evaluated method. This method has been subjected to the established evaluation
procedures of the OSHA SLTC Methods Development Team. |
|
May 1999 | Carl J. Elskamp |
Methods Development Team
Industrial Hygiene Chemistry Division
OSHA Salt Lake Technical Center
Salt Lake City UT 84115-1802
1. General Discussion
-
1.1 Background
1.1.1 History
The determinations of tetrachloroethylene and trichloroethylene in workplace air have
consistently been among the top analyses for solvent vapors done at the OSHA Salt Lake
Technical Center (SLTC). Until this present evaluation was completed, OSHA did not have
fully validated methodologies to determine workplace exposures to vapors of either of these
solvents. This method is an extension of the evaluation work recently completed for toluene
where diffusive as well as active samplers were investigated.(1) As with toluene, exposure
limits to both of these solvents are covered in Table Z-2, thus tests were conducted at TWA,
ceiling and peak target concentrations. After the toluene method was completed, by
consensus it was decided that future evaluation work by the OSHA SLTC Methods
Development Team for solvent vapors would first address charcoal tubes for active sampling
as well as SKC 575-002 Samplers (which contain Anasorb 747) for diffusive sampling. It
was also decided that carbon disulfide (CS2) would be the first choice for the extraction
solvent for each of these samplers. Using these two samplers and CS2 as the extraction
solvent, sampling and analytical methods for both tetrachloroethylene and trichloroethylene
were fully validated at TWA, ceiling and peak target concentrations.
1.1.2 Toxic effects (This section is for information only and should not be taken as the basis of OSHA policy.)
Overexposure to tetrachloroethylene through inhalation can affect the central nervous system
and liver. Prolonged exposure to 200 ppm causes dizziness, headache, confusion, nausea,
and eye and mucous membrane irritation. These effects are intensified and include
incoordination and drunkenness when concentrations exceed 600 ppm. When
concentrations exceed 1000 ppm, anesthetic and respiratory depression effects can cause
unconsciousness and death. One single, brief exposure to greater than 6000 ppm can be
immediately dangerous to life. Prolonged exposures to greater than 200 ppm have
reportedly caused reversible changes to the liver. Alcohol consumption before or after
exposures may increase adverse effects. Repeated exposure to liquid tetrachloroethylene
may defat the skin, causing dermatitis. It can cause significant discomfort if splashed in
the eyes and can cause transient, reversible corneal damage.(2)
Exposure to high concentrations of trichloroethylene can cause headaches, vertigo, tremors,
nausea with vomiting, fatigue, intoxication, unconsciousness, and even death. It is acutely
toxic, primarily because of its anesthetic effect on the central nervous system. Exposure
by inhalation is followed by rapid absorption into the bloodstream. Concentrations of 3000
ppm can cause unconsciousness in less than 10 minutes and anesthetic effects can occur
after 20-minute exposure to 400 ppm. Skin contact may cause dermatitis.(3)
1.1.3 Workplace exposure
About 50% of the demand for tetrachloroethylene is used in the dry cleaning industry where
around 80% of all dry cleaners use it as their primary cleaning agent. It is also used as a
feedstock in the production of fluorocarbons, which accounts for 30% of the current demand.
About 12% of the demand is used as a degreaser for metals and the remaining 8% is
utilized in miscellaneous applications such as transformer insulating fluid, chemical
maskant formulations, and as a process solvent for desulfurizing coal.(4)
Around 85% of the trichloroethylene produced in the United States is used in degreasing of
fabricated metal parts with the remaining 15% divided equally between exports and use in
miscellaneous applications. Some of these applications include use as a component in
adhesive and paint-stripping formulations, a low temperature heat-transfer medium, a
nonflammable solvent carrier in industrial paint systems, and a solvent base for metal
phosphatizing systems. It is also used in the textile industry as a carrier solvent for spotting
fluids and as a solvent in waterless preparation dyeing and finishing operations. Because
of the environmental regulations that have been placed on the use of trichloroethylene, for
many of these applications it is gradually being replaced with alternative solvents such as
1,1,1-trichloroethane.(5)
1.1.4 Physical properties
Tetrachloroethylene:(6)
CAS number: | 127-18-4 |
IMIS number: | 2020 |
molecular weight: | 165.83 |
boiling point: | 121.2°C |
melting point: | -22.7°C |
appearance: | colorless liquid |
density: | 1.623 |
molecular formula: | C2Cl4 |
vapor pressure: | 5.47 kPa at 40°C |
flash point: | none |
odor: | etheral(7) |
explosive limits: | not applicable |
solubility: | 15 mg is soluble in 100 g of water at 25°C; miscible with
chlorinated solvents and most other common solvents. |
synonyms: | Perchloroethylene; ethylene tetrachloride; tetrachloroethylene;
Nema; Tetracap; Tetropil; Perclene; Ankilostin; Didakene.(8) |
structural formula: | Cl2C=CCl2 |
Trichloroethylene:(9) | |
CAS number: | 79-01-6 |
IMIS number: | 2490 |
molecular weight: | 131.39 |
boiling point: | 87.3°C |
melting point: | -86.5°C |
appearance: | colorless liquid |
density: | 1.464 |
molecular formula: | C2HCl3 |
vapor pressure: | 9.3 kPa at 25°C |
flash point: | none |
odor: | sweet smelling, chloroform-like |
explosive limits: | 8% to saturation at 25°C |
solubility: | 107 mg is soluble in 100 g of water at 20°C; miscible with many organic liquids. |
synonyms: | Trichloroethene; ethinyl trichloride; Tri-Clene; Trilene; Trichloren;
Algylen; Trimar; Triline; Trethylene; Westrosol; Chlorylen;
Gemalgene; Germalgene.(10) |
structural formula: | Cl2C=CHCl |
This method was evaluated according to the OSHA SLTC "Evaluation Guidelines for Air Sampling Methods
Utilizing Chromatographic Analyses" (11) in which the terms used and evaluation procedures followed for this
method are fully discussed and explained. The analyte air concentrations throughout this method are based
on the recommended sampling and analytical parameters. TWA target concentration samples are based on
240 minutes and ceiling samples on 5 minutes of charcoal tube sampling and diffusive sampler exposure. Peak
samples are based on 1 minute of sampling with charcoal tubes. Air concentrations listed in ppb and ppm are
referenced to 25°C and 101.3 kPa (760 mmHg).
-
1.2 Limit defining parameters
1.2.1 Detection limit of the analytical procedure
The detection limits of the analytical procedure are 16.2 pg of tetrachloroethylene and 11.2
pg of trichloroethylene. These are the amounts of analytes that will give detector responses
that are significantly different from the responses of a reagent blank. (Section 4.1)
1.2.2 Detection limit of the overall procedure
The detection limits of the overall procedure are 0.35 µg of tetrachloroethylene per sample
(4.3 ppb or 29 µg/m3) and 0.24 µg of trichloroethylene per sample (3.7 ppb or 20 µg/m3) for
charcoal tubes and 0.66 µg of tetrachloroethylene per sample (31 ppb or 211 g/m3) and
0.33 µg of trichloroethylene per sample (18 ppb or 97 µg/m3) for SKC 575-002 Samplers.
These are the amounts of analytes spiked on the respective samplers that will give detector
responses that are significantly different from the responses of respective sampler blanks.
(Section 4.2)
1.2.3 Reliable quantitation limit
The reliable quantitation limits are 1.17 µg of tetrachloroethylene per sample (14 ppb or 98
µg/m3) and 0.81 µg of trichloroethylene per sample (13 ppb or 68 µg/m3) for charcoal tubes
and 2.21 µg of tetrachloroethylene per sample (104 ppb or 705 µg/m3) and 1.11 µg of
trichloroethylene per sample (61 ppb or 325 µg/m3) for SKC 575-002 Samplers. These are
the amounts of analytes spiked on the respective samplers that will give detector responses
that are considered the lower limits for precise quantitative measurements. (Section 4.2)
1.2.4 Instrument calibration
The coefficient of determination (r2) is 0.99996 for tetrachloroethylene and 0.99993 for
trichloroethylene over the ranges of 21.1 to 16,230 µg/mL and 19.0 to 12,874 µg/mL
respectively. The coefficient of non-determination (k2) is 4×10-5 for tetrachloroethylene and
7×10-5 for trichloroethylene over the same ranges. These ranges correspond to 0.5 times
the ceiling target concentrations for SKC 575-002 Samplers to 2 times the TWA target
concentrations for charcoal tubes. (Section 4.3)
1.2.5 Precision (overall procedure)
Charcoal tubes
The precisions of the overall procedure at the 95% confidence level from the ambient
temperature storage tests for TWA, ceiling and peak samples for charcoal tubes are given
in Table 1.2.5.1. The TWA samples are 240-min samples taken from 100-ppm
atmospheres, the ceiling samples are 5-min samples taken from 200-ppm atmospheres and
the peak samples are 1-min samples taken from 300-ppm atmospheres. (Section 4.4)
Table 1.2.5.1 Precision of the Overall Procedure at the 95% Confidence Interval for Charcoal Tubes
|
analyte |
TWA samples |
ceiling samples |
peak samples |
|
tetrachloroethylene trichloroethylene |
±9.9% ±9.9% |
±9.9% ±9.9% |
±10.2% ±10.0% |
|
SKC 575-002 Samplers
The precisions of the overall procedure at the 95% confidence level from the ambient
temperature storage tests for TWA and ceiling samples for SKC 575-002 Samplers are given
in Table 1.2.5.2. The TWA samples are 240-min samples taken from 100-ppm atmospheres
and the ceiling samples are 5-min samples taken from 200-ppm atmospheres. There are
different values given for each analyte at each of the levels, depending on whether both,
either, or neither temperature or atmospheric pressure are known at the sampling site. If the
temperature is not known, it is assumed to be 22.2 ± 15°C (72 ± 27°F) and a variability of
±7.7% is included. If the atmospheric pressure is not known, it is estimated from the
sampling site elevation and a variability ±3% is included. (Section 4.4)
Table 1.2.5.2 Precision of the Overall Procedure
at the 95% Confidence Interval for SKC 575-002 Samplers When Sampling
Site Temperature (T) or Atmospheric Pressure (P) are Known or Unknown
|
analyte |
TWA samples |
ceiling samples |
|
tetrachloroethylene
both T&P known
only T known
only P known
neither T nor P known
trichloroethylene
both T&P known
only T known
only P known
neither T nor P known |
±17.7% ±18.6% ±23.3% ±23.9%
±17.9% ±18.8% ±23.3% ±24.1% |
±18.9% ±19.8% ±24.1% ±24.9%
±17.9% ±18.9% ±23.5% ±24.1% |
|
1.2.6 Recovery
The recovery of tetrachloroethylene from TWA samples in 17-day storage tests remained
above 99.5% and 98.6% for charcoal tubes and SKC 575-002 Samplers respectively when
the samples were stored at ambient temperatures. The recovery of trichloroethylene from
TWA samples in 17-day storage tests remained above 98.5% and 96.8% for charcoal tubes
and SKC 575-002 Samplers respectively when the samples were stored at ambient
temperatures. The recovery of tetrachloroethylene from ceiling samples in 17-day storage
tests remained above 100% and 97.3% for charcoal tubes and SKC 575-002 Samplers
respectively when the samples were stored at ambient temperatures. The recovery of
trichloroethylene from ceiling samples in 17-day storage tests remained above 98.8% and
97.4% for charcoal tubes and SKC 575-002 Samplers respectively when the samples were
stored at ambient temperatures. The recoveries of tetrachloroethylene and trichloroethylene
from charcoal tube peak samples in 17-day storage tests remained above 97.3% and 96.5%
respectively when the samples were stored at ambient temperatures. (Section 4.5)
1.2.7 Reproducibility
Six samples for both samplers at the TWA and ceiling target concentrations and six
charcoal tube samples at the peak target concentrations that were collected from controlled
test atmospheres, along with a draft copy of this procedure, were submitted to an SLTC
service branch for analysis. All samples were analyzed 22 days after generation and had
been stored at room temperature. No individual sample result deviated from its theoretical
value by more than the precisions reported in Section 1.2.5. (Section 4.6)
2. Sampling Procedure
All safety practices that apply to the work area being sampled should be followed. The sampling equipment
should be attached to the worker so that it will not interfere with work performance or safety.
-
2.1 Apparatus
2.1.1 Adsorbent tubes
Samples are collected with 7-cm × 4-mm i.d. × 6-mm o.d. glass sampling tubes packed
with two sections of coconut shell charcoal. The tubes contain 100 mg of adsorbent in the
front section and 50 mg in the back section. The adsorbent sections are held in place with
glass wool plugs and are separated by urethane foam plugs. The ends of the glass
sampling tubes are heat sealed. SKC Lot 2000 (SKC Inc., Fullerton, CA, Cat. No. 226-01)
charcoal tubes were used for this evaluation.
Samples are collected using personal sampling pumps that have been calibrated, with
sampling devices attached, to within ±5% at the recommended flow rate of 50 mL/min. The
charcoal tubes are contained in commercially available tube holders (SKC Cat. No. 222-3-1)
which are connected to the pumps with flexible, non-crimpable tubing.
2.1.2 Diffusive samplers
Samples are collected with SKC 575-002 Passive Samplers (SKC, Inc., Fullerton, CA). The
samplers contain 500 mg of Anasorb 747 adsorbent. Lot 263 samplers were used in this
evaluation.
A thermometer and barometer to determine the sampling site conditions.
2.2 Reagents
None required
2.3 Technique
2.3.1 Charcoal tubes
Immediately before sampling, break off the ends of the charcoal tube. All tubes should be from the same lot.
Connect the sampling tube to the sampling pump with flexible, non-crimpable tubing. It is
desirable to utilize a sampling tube holder that shields the employee from the sharp, jagged
end of the sampling tube. Position the tube so that sampled air first passes through the
larger adsorbent section.
Air being sampled should not pass through any hose or tubing before entering the sampling tube.
To avoid channeling, place the sampling tube vertically in the employee's breathing zone.
Position the sampler so it does not impede work performance or safety.
After sampling for the appropriate time, immediately remove the sampling tube and seal it
with plastic caps. Wrap each sample lengthwise with a Form OSHA-21 seal.
Submit at least one blank sampling tube with each sample set. Blanks should be handled
in the same manner as samples, except no air is drawn through them.
Record air volume (in liters), sampling time (minutes) and sampling rate (mL/min) for each sample on Form OSHA-91A.
Also list any compounds that could be considered potential interferences, especially
solvents, that are being used in the sampling area.
Ship any bulk sample(s) in a container separate from the air samples.
2.3.2 SKC 575-002 Samplers (In general, follow the manufacturer's instructions supplied with the samplers.)
Remove the sampler, which is enclosed in an airtight clear bag, from the container. Keep
the O-ring, press-on cover, cover retainer, port plugs and Teflon tube for later use.
Remove the sampler from the clear bag when ready to begin sampling. CAUTION - The
monitor immediately begins to sample when it is removed from this bag.
Record the start time on the sampler label or on a Form OSHA-91A.
Attach the sampler to the worker near his/her breathing zone with the perforations in the
sampler facing out. Assure that the area directly in front of the sampler is unobstructed
throughout the sampling period.
At the end of the sampling period, immediately detach the sampler from the worker and with
the O-ring in place, attach the cover onto the sampler securing it with the cover retainer.
Visually inspect the O-ring to be sure it is forming a proper seal around the entire
circumference of the sampler. Record the stop time on sampler label or on a Form
OSHA-91A.
Prepare a blank by removing an unused sampler from its clear bag and immediately
attaching a cover, with the O-ring in place, onto it. Secure the cover with a retainer.
Seal each sampler with an OSHA-21 form.
Verify that the sampling times are properly recorded on the Form OSHA-91A for each
sample. Also, identify blank samples on this form.
Record the room temperature and atmospheric pressure (station pressure) of the sampling
site on the Form OSHA-91A.
List any compounds that could be considered potential interferences, especially solvents,
that are being used in the sampling area.
Submit the samplers to the laboratory for analysis as soon as possible. Include all port
plugs and Teflon tubes which will be used in the laboratory analyses.
Ship any bulk sample(s) in a container separate from the air samples.
2.4 Sampler capacity (charcoal tubes) and sampler rate/capacity (SKC 575-002 Samplers)
2.4.1 Charcoal tubes
The sampling capacity of the front section of charcoal sampling tubes was tested by
sampling from a dynamically generated test atmosphere of tetrachloroethylene at 192.7 ppm
(1306 mg/m3) and from another atmosphere containing 192.0 ppm (1031 mg/m3) of
trichloroethylene. The samples were collected at nominal flow rates of 50 mL/min and the
mean relative humidity of the tetrachloroethylene atmosphere was 65% at 26.1°C and the
trichloroethylene atmosphere was 64.5% RH at 25.8°C. The breakthrough from the front
sections of charcoal tubes did not exceed 0.1% when sampling from the tetrachloroethylene
atmosphere for 9.5 hours (28.5 liters), indicating the capacity of charcoal tubes is about
37 mg of tetrachloroethylene. The mean 5% breakthrough volume was determined to be
approximately 23.3 liters (7.8 hours) when sampling from the trichloroethylene atmosphere,
indicating a capacity of about 24 mg of trichloroethylene before 5% breakthrough occurred.
(Section 4.7.1)
2.4.2 SKC 575-002 Samplers
The sampling rate and capacity of SKC 575-002 Samplers were determined by taking
samples from a dynamically generated test atmosphere of tetrachloroethylene (nominal
concentration of 200 ppm or 1356 mg/m3) and trichloroethylene (nominal concentration of
200 ppm or 1074 mg/m3) for increasing time intervals. A sampling rate of 13.06 mL/min (at
760 mmHg, 25°C) and capacity of greater than 12 mg per sample for tetrachloroethylene
and a sampling rate of 14.24 mL/min (at 760 mmHg, 25°C) and capacity of greater than 10
mg per sample for trichloroethylene were obtained from this test. (Section 4.7.2)
2.5 Extraction efficiency
It is the responsibility of each analytical laboratory to determine the extraction efficiency for each
analyte because the internal standard, adsorbent material, reagents and laboratory techniques may
influence the results and be different than the those listed in this evaluation.
2.5.1 Charcoal tubes (Section 4.8.1)
The mean respective extraction efficiencies for tetrachloroethylene and trichloroethylene from
SKC Lot 2000 charcoal tubes over the range of the RQL to 2 times the target concentrations
are 98.8% and 99.3% for TWA levels, 98.3% and 98.8% for ceiling levels and 98.5% and
99.0% for peak levels. The mean extraction efficiencies for the RQL and all of the eighteen
other concentrations are 98.5% for tetrachloroethylene and 99.1% for trichloroethylene.
The mean respective extraction efficiencies for tetrachloroethylene and trichloroethylene from
SKC Lot 2000 charcoal tubes that had humid air drawn through them before they were
spiked at the target concentrations are 98.9% and 99.5% for TWA levels, 98.6% and 98.9%
for ceiling levels and 98.6% and 98.8% for peak levels.
Extracted samples remain stable for at least 24 h.
2.5.2 SKC 575-002 Samplers (Section 4.8.2)
The mean respective extraction efficiencies for tetrachloroethylene and trichloroethylene from
SKC 575-002 Samplers over the ranges of the RQL to 2 times the target concentrations are
96.2% and 97.7% for TWA levels and 94.9% and 97.3% for ceiling levels. The mean
extraction efficiencies for the RQL and all of the twelve other concentrations are 95.4% for
tetrachloroethylene and 97.5% for trichloroethylene.
The mean respective extraction efficiencies for tetrachloroethylene and trichloroethylene from
SKC 575-002 Samplers that had been exposed to humid air before they were spiked at the
target concentrations are 93.2% and 95.3% for TWA levels and 94.0% and 96.0% for ceiling levels.
Extracted samples remain stable for at least 24 h.
2.6 Recommended air volume and sampling rate for charcoal tube samples and recommended exposure
time for SKC 575-002 Samplers
2.6.1 For TWA (long-term) samples, sample up to 12 L of air at 50 mL/min (up to 240 min) when
using charcoal tubes. When using SKC 575-002 Samplers, sample for as long as 240 minutes.
2.6.2 For ceiling samples, sample greater than 0.25 L of air at 50 mL/min (greater than 5 min)
when using charcoal tubes. When using SKC 575-002 Samplers, sample for greater than 5 minutes.
2.6.3 For peak samples, sample at least 0.05 L of air at 50 mL/min (at least 1 min) when using
charcoal tubes. When using SKC 575-002 Samplers, sample for 5 minutes or longer.
2.6.4 When short-term samples are collected, the air concentrations equivalent to the reliable
quantitation limits become larger. For example, the respective reliable quantitation limits
for tetrachloroethylene and trichloroethylene samples collected on charcoal tubes become
0.71 ppm (4.8 mg/m3) and 0.60 ppm (3.2 mg/m3) for 5-min samples and 3.6 ppm (24 mg/m3)
and 3.0 ppm (16 mg/m3) for 1-min samples.
2.7 Interferences (sampling)
2.7.1 The presence of other contaminants in the sampled air can potentially reduce the capacity
of the samplers to collect the analyte of interest. Also, the sampling rates of the SKC 575-002 Samplers could possibly be altered. Interference studies were performed by sampling
for 240 minutes from a test atmosphere (70% RH, 24°C, 659 mmHg) containing 100 ppm
of tetrachloroethylene with 83 ppm of methylene chloride, 70 ppm of isopropyl alcohol, 71
ppm of 2-butanone (MEK), 26 ppm of butyl acetate, 17 ppm of dioxane and 16 ppm of amyl
acetate. Another test was done by sampling for 240 minutes from a test atmosphere (59%
RH, 27°C, 659 mmHg) containing 100 ppm of trichloroethylene plus the aforementioned
solvent vapors. The presence of these compounds, which may represent typical substances
that may be collected with tetrachloroethylene or trichloroethylene, did not have a significant
effect on sample results using either of the samplers. (Section 4.9.1)
2.7.2 A reverse diffusion study for the SKC 575-002 Samplers and a stripping study for the
charcoal tubes was performed by sampling an atmosphere (65% RH, 26C, 651 mmHg)
containing 193 ppm of tetrachloroethylene and 194 ppm trichloroethylene for 60 minutes with
six of each samplers. Three samplers from each set were additionally subjected to 180
minutes of the same atmosphere without the analytes present to determine if any of the
collected solvents diffused off of the SKC 575-002 Samplers and also whether it was
stripped off of the charcoal tubes. Upon analysis of the samples, the mean recovery of the
removed samplers versus the mean recovery of the samplers that were additionally exposed
to the atmosphere without analyte was greater than 90% for both samplers, indicating that
reverse diffusion and stripping are not significant. (Section 4.9.2)
2.7.3 Sampling from relatively dry atmospheres was investigated by sampling from an atmosphere
(3% RH, 26°C, 650 mmHg) containing 198 ppm of tetrachloroethylene and 199 ppm of
trichloroethylene for 240 minutes with both samplers. Sampling from dry atmospheres did
not have a significant effect on the results for either of the analytes using either of the
samplers. (Section 4.9.3)
2.7.4 Sampling from atmospheres containing low concentrations of analytes was investigated by
sampling from an atmosphere (67% RH, 26°C, 649 mmHg) containing 9.0 ppm of
tetrachloroethylene and 10.7 ppm of trichloroethylene for 240 minutes with both samplers.
Sampling these low concentrations did not have a significant effect on results using either
of the samplers for either of the analytes. (Section 4.9.4)
2.7.5 Suspected interferences should be reported to the laboratory with submitted samples.
3. Analytical Procedure
Adhere to the rules set down in your Chemical Hygiene Plan, which is mandated by the OSHA laboratory
standard. Avoid skin contact and inhalation of all chemicals.
-
3.1 Apparatus
3.1.1 A GC equipped with a flame ionization detector. For this evaluation, a Hewlett-Packard 6890
Series Gas Chromatograph equipped with an Automatic Liquid Sampler was used.
3.1.2 A GC column capable of separating analyte from the extraction solvent, internal standard
and any interferences. A 60-m × 0.32-mm i.d. fused silica Stabilwax capillary column (Cat.
No. 10657) with a 1.0-µm df from Restek Corporation (Bellefonte, PA) was used in this
evaluation.
3.1.3 An electronic integrator or some other suitable means of measuring peak areas. A Waters
Millennium 2020 Networking Computer System was used in this evaluation.
3.1.4 Two-milliliter vials with Teflon-lined caps.
3.1.5 A dispenser capable of delivering 1.0 mL of extraction solvent to prepare standards and
samples. If a dispenser is not available, a 1.0-mL volumetric pipet may be used.
3.1.6 A sampler rack (SKC Cat. No. 226-04-5) and a specialized shaker (SKC Cat. No. 226D-03-1) to facilitate the extraction of SKC 575-002 Samplers.
3.2 Reagents
3.2.1 Tetrachloroethylene (CAS 127-18-4), reagent grade. Aldrich A.C.S. reagent grade (Cat. No.
44,378-6) Lot TR06115EN, was used in this evaluation.
3.2.2 Trichloroethylene (CAS 79-01-6), reagent grade. Aldrich A.C.S. reagent grade (Cat. No.
25,140-2) Lot DR0306BQ, was used in this evaluation.
3.2.3 Carbon disulfide (CS2) [CAS 75-15-0], chromatographic grade. Aldrich 99.9+% low benzene
grade (Cat. No. 34,227-0) Lot 1054JQ, was used in this evaluation.
3.2.4 A suitable internal standard, reagent grade. Supelco Neat EPA Standard, Lot LA59304,
ethylbenzene (CAS 100-41-4) was used in this evaluation.
3.2.5 The extraction solvent consists of CS2 containing 1.0 milliliter of internal standard per liter of solution (1µL/mL).
3.2.6 GC grade nitrogen, air and hydrogen.
3.3 Standard preparation
3.3.1 Prepare standards by injecting microliter amounts of analyte into vials containing 1.0 mL (for
charcoal tubes) or 2.0 mL (for SKC 575-002 Samplers) of extraction solvent delivered from
the same dispenser used to extract samples. For example, inject 5.00 µL of
tetrachloroethylene into a vial containing 1.0 mL of extraction solvent. Assuming the density
of tetrachloroethylene is 1.623 g/mL, (which is dependent on the temperature of the
tetrachloroethylene) this standard contains 8115 µg of tetrachloroethylene per milliliter.
3.3.2 Bracket sample concentrations with standard concentrations. If upon analysis, sample
concentrations fall outside the range of prepared standards, prepare and analyze additional
standards to ascertain the linearity of instrument response or dilute high samples with
extraction solvent and reanalyze the diluted samples.
3.4 Sample preparation
3.4.1 Charcoal tube samples
Transfer each section of adsorbent from the sampling tubes to separate labeled vials.
Discard the glass tubes, urethane foam plugs and glass wool plugs.
Add 1.0 mL of extraction solvent to each vial using the same dispenser as used for preparation of standards.
Immediately cap the vials.
Allow the adsorbent sections to extract for 30 minutes. Periodically apply gentle agitation
to the vials during the extraction period.
3.4.2 SKC 575-002 Samplers (In general, follow the manufacturer's instructions supplied with the samplers.)
Cut off the ends of the two protruding tubes of each sampler with a razor blade or sharp knife.
Slowly add 1.0 mL of extraction solvent through one of the protruding tubes (ports). After
about 30 seconds, slowly add another 1.0 mL of extraction solvent and then immediately
insert plugs into the ports.
Mount the samplers in the sampler rack (SKC Cat. No. 226-04-5) of a specialized shaker
(SKC Cat. No. 226D-03-1) and shake the samplers for 1 hour.
Transfer each extracted sample by removing the plugs from the sampler ports, firmly
inserting the tapered end of a supplied Teflon tube into the outer port and carefully pouring
the solution through the Teflon tube into a labeled autosampler vial.
3.5 Analysis
3.5.1 GC conditions
zone temperatures: | column- injector- detector- |
100°C (isothermal) 200°C 250°C |
gas flows: | hydrogen (carrier)- nitrogen (makeup)- hydrogen (flame)- air- |
2.5 mL/min (89 kPa head pressure) 50 mL/min 38 mL/min 450 mL/min |
injection volume: | 1.0 µL (with a 15:1 split) |
signal range: | 0 |
column: | 60-m × 0.32-mm i.d. fused silica, Stabilwax 1.0-µm df |
retention times: |
trichloroethylene- tetrachloroethylene- ethylbenzene- (CS2- 2.9 min) |
4.7 min 5.3 min 7.6 min (internal standard) |
Figure 3.5.1.1. Chromatogram of a standard near the TWA target concentrations for charcoal tubes.
Key: (1) CS2, (2) trichloroethylene, (3) tetrachloroethylene, (4) ethylbenzene. |
|
Figure 3.5.1.2. Chromatogram of a standard near the TWA target concentrations for SKC 575-002 Samplers.
Key: (1) CS2, (2) trichloroethylene, (3) tetrachloroethylene, (4) ethylbenzene. |
|
Figure 3.5.1.3. Chromatogram of a standard near the ceiling target concentrations for charcoal tubes.
Key: (1) CS2, (2) trichloroethylene, (3) tetrachloroethylene, (4) ethylbenzene. |
Figure 3.5.1.4. Chromatogram of a standard near the ceiling target concentrations for SKC 575-002 Samplers.
Key: (1) CS2, (2) trichloroethylene, (3) tetrachloroethylene, (4) ethylbenzene. |
Figure 3.5.1.5. Chromatogram of a standard near the peak target concentrations charcoal tubes.
Key: (1) CS2, (2) trichloroethylene, (3) tetrachloroethylene, (4) ethylbenzene.
3.5.2 Peak areas are measured by an integrator or other suitable means.
3.5.3 An internal standard (ISTD) calibration method is used. A calibration curve is prepared by
analyzing standards and plotting micrograms of analyte per milliliter versus ISTD-corrected
area counts of the analyte peaks. Sample concentrations must be bracketed by standards.
Figure 3.5.3.1. Calibration curve for tetrachloroethylene constructed from the data in Table 4.3.1.
The equation of the line is Y = 216.8X + 10648. |
|
Figure 3.5.3.2. Calibration curve for trichloroethylene constructed from the data in Table 4.3.2.
The equation of the line is Y = 224.8X - 827. |
-
3.6 Interferences (analytical)
3.6.1 Any compound that produces a response on a flame ionization detector and has the same
general retention time of the analyte of interest or the internal standard is a potential
interference. Possible interferences should be reported to the laboratory with submitted
samples by the industrial hygienist. These interferences should be considered before
samples are extracted.
3.6.2 GC parameters (i.e. column and column temperature) may be changed to possibly circumvent interferences.
3.6.3 The extraction efficiencies from "wet" samplers for both analytes at the various target
concentrations were investigated. The extraction efficiencies were comparable to those
obtained from samplers that had not been exposed to humid air. (Section 2.5)
3.6.4 When necessary, the identity or purity of an analyte peak may be confirmed with additional
analytical data, such as mass spectrometry. (Section 4.10)
3.7 Calculations
3.7.1 Charcoal tube samples
The analyte concentration for samples is obtained from the appropriate calibration curve in
terms of micrograms per milliliter, uncorrected for extraction efficiency. The air
concentration is calculated using the following formulae. The back (50-mg) section is
analyzed primarily to determine if there was any breakthrough from the front section during
sampling. If a significant amount of analyte is found on the back section (e.g., greater than
25% of the amount found on the front section), this fact should be reported with sample
results. If any analyte is found on the back section, it is added to the amount found on the
front section. This total amount is then corrected by subtracting the total amount (if any)
found on the blank.
mg/m3 = (µg of analyte per mL)(1.0 mL) / ((L of air sampled)(extraction efficiency))
where |
L of air sampled is [(sampling time, min)(sampling rate, mL/min)] / 1000
extraction efficiency is expressed in decimal fraction |
ppm = (mg/m3)(24.46) / (molecular weight of analyte)
where |
24.46 is the molar volume at 25°C and 101.3 kPa (760 mmHg)
MW of tetrachloroethylene is 165.83
MW of trichloroethylene is 131.39 |
3.7.2 SKC 575-002 Samplers
The analyte concentration for samples is obtained from the appropriate calibration curve in
terms of micrograms per milliliter, uncorrected for extraction efficiency. The air
concentration is calculated using the following formulae.
mg/m3 = (µg of analyte per mL)(2 mL) / ((L of air sampled)(extraction efficiency))
where |
L of air sampled is [(sampling time, min)(sampling rate, mL/min)]/1000
extraction efficiency is expressed in decimal fraction
Sampling rate for tetrachloroethylene is (13.06)(760/P)(T/298.2)1.5
Sampling rate for trichloroethylene is (14.24)(760/P)(T/298.2)1.5
P is the sampling site atmospheric pressure (mmHg)
T is the sampling site temperature (K) |
If the sampling site temperature is not provided, assume
that it is 22.2°C (295.4 K, 72°F). If the sampling site atmospheric pressure is
not given, calculate an approximate value (P) based on the sampling site elevation
(E) from the following equation. |
| P = (3.887 × 10-7)(E2) - (2.7467 × 10-2)(E) + 760 |
ppm = (mg/m3)(24.46) / (molecular weight of analyte)
where |
24.46 is the molar volume at 25°C and 101.3 kPa (760 mmHg)
MW of tetrachloroethylene is 165.83
MW of trichloroethylene is 131.39 |
4. Backup Data
This method was evaluated according to the OSHA SLTC "Evaluation Guidelines for Air Sampling Methods
Utilizing Chromatographic Analyses"(12) in which the terms used and evaluation procedures followed for this
method are fully discussed and explained.
-
4.1 Detection limit of the analytical procedure (DLAP)
The DLAP is measured as the mass of analyte introduced onto the chromatographic column. Ten
analytical standards were prepared in equal increments with the highest standard containing 2.43 µg
of tetrachloroethylene and 1.90 µg of trichloroethylene per mL. Injection of the highest standards
produced peaks approximately 10 times the baseline noise of a reagent blank when a 1-µL injection
with a 1:15 split was made onto the GC column. Standards, plus a reagent blank, were analyzed
and the data obtained were used to determine the required parameters (A and SEE) for the
calculation of the DLAP for each analyte. SEE values of 18.3 and 15.3 and A values of 3.392 and
4.089 were obtained for tetrachloroethylene and trichloroethylene respectively. The DLAP was
calculated to be 16.2 pg for tetrachloroethylene and 11.2 pg for trichloroethylene.
Table 4.1.1 DLAP for Tetrachloroethylene
|
concentration (µg/mL) |
mass on column (pg) |
peak area (µV s) |
|
0.000 0.243 0.487 0.730 0.974 1.22 1.46 1.70 1.95 2.19 2.43 |
0.0 16.2 32.5 48.7 64.9 81.3 97.3 113 130 146 162 |
0 55.9 123 178 198 276 328 364 486 571 614 |
|
| |
Figure 4.1.1 Plot of data from Table 4.1.1 to determine the DLAP. The equation of the line is Y = 3.392X + 2.79.
|
|
Table 4.1.2 DLAP for Trichloroethylene
|
concentration (µg/mL) |
mass on column (pg) |
peak area (µV s) |
|
0.000 0.190 0.380 0.571 0.761 0.951 1.14 1.33 1.52 1.71 1.90 |
0.0 12.7 25.3 38.1 50.7 63.4 76.0 88.7 101 114 127 |
0 56.9 94.5 133 170 248 320 374 409 448 512 |
|
| |
Figure 4.1.2 Plot of data from Table 4.1.2 to determine the DLAP. The equation of the line is Y = 4.089X - 7.68.
|
-
4.2 Detection limit of the overall procedure (DLOP) and reliable quantitation limit (RQL)
The DLOP is measured as mass per sample and expressed as equivalent air concentrations, based
on the recommended sampling parameters. Ten samplers of each type were spiked with equal
descending increments of tetrachloroethylene and trichloroethylene such that the highest sampler
loadings were 2.43 µg/sample and 1.90 µg/sample for the charcoal tubes and 4.87 µg/sample and
3.80 µg/sample for SKC 575-002 Samplers of tetrachloroethylene and trichloroethylene respectively.
(The SKC 575-002 Samplers were spiked with twice the amounts of analytes compared to charcoal
tubes because they are extracted with 2.0 mL of solvent versus 1.0 mL for the charcoal tubes.) These
are the amounts, when spiked on the samplers, that would produce peaks approximately 10 times
the baseline noise for sample blanks. These spiked samplers, plus blanks, were analyzed with the
recommended analytical parameters, and the data obtained used to calculate the required
parameters (A and SEE) for the calculation of the DLOPs. Respective A and SEE values of 238.1
and 27.9 for tetrachloroethylene and 264.8 and 21.5 for trichloroethylene were obtained from charcoal
tube samples to give DLOPs of 0.35 µg of tetrachloroethylene per sample (4.3 ppb or 29µg/m3) and
0.24 µg of trichloroethylene sample (3.7 ppb or 20 g/m3). Respective A and SEE values of 107.2
and 23.7 for tetrachloroethylene and 118.1 and 13.1 for trichloroethylene were obtained from SKC
575-002 Samplers to give DLOPs of 0.66 µg of tetrachloroethylene per sample (31 ppb or 211 µg/m3)
and 0.33 µg of trichloroethylene sample (18 ppb or 97 µg/m3).
Table 4.2.1 Detection Limit of the Overall Procedure for Tetrachloroethylene from Charcoal Tubes
|
mass (µg) per sample |
peak area (µV s) |
|
0.0 0.243 0.487 0.730 0.974 1.22 1.46 1.70 1.95 2.19 2.43 |
0 26.0 109 161 221 317 319 384 402 499 611 |
|
| |
Figure 4.2.1 Plot of data from Table 4.2.1 to determine the DLAP. The equation of the line is Y = 3.392X + 2.79.
|
|
Table 4.2.2 Detection Limit of the Overall Procedure for Trichloroethylene from Charcoal Tubes
|
mass (µg) per sample |
peak area (µV s) |
|
0.0 0.190 0.380 0.571 0.761 0.951 1.14 1.33 1.52 1.71 1.90 |
0 44.3 82.3 188 203 230 257 349 411 452 508 |
|
| |
Figure 4.2.2 Plot of data from Table 4.2.2 to determine the DLAP. The equation of the line is Y = 4.089X - 7.68.
|
|
Table 4.2.3 Detection Limit of the Overall Procedure for Tetrachloroethylene from SKC 575-002 Samplers
|
mass (µg) per sample |
peak area (µV s) |
|
0.0 0.487 0.974 1.46 1.95 2.43 2.92 3.41 3.90 4.38 4.87 |
0 65.5 124 175 207 239 310 374 386 462 574 |
|
| |
Figure 4.2.3. Plot of data from Table 4.2.3 to determine the DLOP/RQL for tetrachloroethylene from SKC 575-002 Samplers. The equation of the line is Y = 107.2X + 4.1.
|
|
Table 4.2.4 Detection Limit of the Overall Procedure for Trichloroethylene from SKC 575-002 Samplers
|
mass (µg) per sample |
peak area (µV s) |
|
0.0 0.380 0.761 1.14 1.52 1.90 2.28 2.66 3.04 3.42 3.80 |
0 44.9 124 134 194 243 264 329 359 425 450 |
|
| |
Figure 4.2.4. Plot of data from Table 4.2.4 to determine the DLOP/RQL for trichloroethylene from SKC 575-002 Samplers. The equation of the line is Y = 118.1X + 9.0.
|
-
The RQL is considered the lower limit for precise quantitative measurements. It is
determined from the regression line data obtained for the calculation of the DLOP. The
RQLs were calculated to be 1.17 µg per sample (14 ppb or 98 µg/m3)
and 0.81 µg per sample (13 ppb or 68 µg/m3)
respectively for tetrachloroethylene and trichloroethylene from charcoal tubes and 2.21 µg
per sample (104 ppb or 705 µg/m3) and 1.11 µg per
sample (61 ppb or 325 µg/m3) respectively for
tetrachloroethylene and trichloroethylene from SKC 575-002 Samplers. The
recoveries at these levels are 99.5%, 98.8%, 97.2% and 97.6% respectively.
Figure 4.2.5. Chromatogram near the RQLs. Key: (1) CS2,
(2) trichloroethylene, (3) tetrachloroethylene, (4) ethylbenzene.
4.3 Instrument calibration
The instrument was calibrated using standards over concentration ranges equivalent to 0.5 times the
ceiling target concentrations for SKC 575-002 Samplers to 2 times the TWA target concentrations
for charcoal tube samples. Calibration curves were constructed as shown in Section 3.5.3 from six
replicate injections of five standards equally spaced over these ranges. The coefficient of
determination (r2) is 0.99996 for tetrachloroethylene and 0.99993 for trichloroethylene. The coefficient
of non-determination (k2) is
4×10-5 for tetrachloroethylene and 7×10-5 for trichloroethylene.
Table 4.3.1 Tetrachloroethylene Standards used for Instrument Calibration |
|
standard concn (µg/mL) |
|
response in peak areas (µV s) × 10-3 |
|
21.1 4058 8115 12172 16230 |
|
4.6945 904.56 1775.2 2649.1 3529.2 |
4.6979 903.81 1770.0 2645.5 3531.8 |
4.7172 904.63 1771.2 2644.1 3524.8 |
4.6936 903.90 1772.3 2650.8 3525.0 |
4.6995 902.18 1769.5 2643.3 3527.8 |
4.6970 904.97 1772.2 2642.9 3521.7 |
|
Table 4.3.2 Trichloroethylene Standards used for Instrument Calibration |
|
standard concn (µg/mL) |
|
response in peak areas (µV s) × 10-3 |
|
19.0 3219 6437 9656 12874 |
|
4.3280 715.66 1461.8 2177.6 2891.2 |
4.3264 714.21 1451.5 2175.1 2898.6 |
4.3216 711.91 1455.5 2174.7 2887.3 |
4.2951 711.87 1457.2 2184.8 2880.9 |
4.2926 709.61 1456.7 2170.6 2879.5 |
4.2970 713.48 1455.2 2169.3 2878.0 |
|
4.4 Precision (overall procedure)
The precision of the overall procedure is determined from the storage data in Section 4.5. The SEEs
and precisions of the overall procedure are given in the following tables for charcoal tubes and SKC
575-002 Samplers. All values are based on the ambient storage tests in Section 4.5.
Table 4.4.1 SEEs and Precisions of the Overall Procedure at the 95% Confidence Interval for Charcoal Tube Samples |
|
|
TWA samples
|
ceiling samples
|
peak samples
|
analyte | SEE | precision | SEE | precision | SEE | precision |
|
tetrachloroethylene trichloroethylene |
5.06% 5.06% |
±9.9% ±9.9% |
5.05% 5.04% |
±9.9% ±9.9% |
5.21% 5.09% |
±10.2% ±10.0% |
|
Table 4.4.2 SEEs and Precisions of the Overall Procedure at the 95% Confidence Interval for SKC 575-002 Samplers when Sampling Site Temperature (T) or Atmospheric Pressure (P) are Known or Unknown |
|
|
TWA samples
|
ceiling samples
|
analyte | SEE | precision | SEE | precision |
|
tetrachloroethylene
both T&P known
only T known
only P known
neither T nor P known
trichloroethylene
both T&P known
only T known
only P known
neither T nor P known |
9.03% 9.51% 11.9% 12.2%
9.13% 9.61% 11.9% 12.3% |
±17.7% ±18.6% ±23.3% ±23.9%
±17.9% ±18.8% ±23.3% ±24.1% |
9.63% 10.1% 12.3% 12.7%
9.14% 9.62% 12.0% 12.3% |
±18.9% ±19.8% ±24.1% ±24.9%
±17.9% ±18.9% ±23.5% ±24.1% |
|
4.5 Storage tests
Charcoal tube storage samples were prepared by sampling at 50 mL/min from controlled test
atmospheres that were at approximately 80% RH and 22.2°C and at an atmospheric pressure of 655
mmHg. Storage samples for SKC 575-002 Samplers were prepared by exposing them to the same
atmospheres as were generated for the charcoal tube samples. The flow of the atmospheres through
the exposure chamber provided for face velocities of approximately 0.4 m/s on the SKC 575-002
Samplers. TWA samples were prepared by sampling from 100-ppm atmospheres for 240 minutes,
ceiling samples from 200-ppm atmospheres for 5 minutes and peak samples from 300-ppm
atmospheres for 1 minute. Six samples for each set were analyzed immediately after generation,
fifteen were stored in a refrigerator at 0°C and fifteen were stored in a closed drawer at ambient
temperatures of 20-25°C. At approximately three-day intervals, three samples were selected from
each of the storage sets and analyzed.
Table 4.5.1 TWA Storage Tests for Tetrachloroethylene (Charcoal Tubes)
|
time (days) |
refrigerated storage recovery (%) |
|
ambient storage recovery (%) |
|
0 0 6 10 13 15 17 |
98.7 99.6 101.8 99.4 99.3 99.6 99.0 |
98.2 99.9 99.4 99.8 100.5 99.6 100.7 |
100.5 99.6 100.4 100.9 100.2 99.3 99.8 |
|
100.6 102.8 101.7 99.4 99.3 99.6 100.0 |
102.0 100.4 98.7 99.3 100.0 99.0 99.3 |
99.9 101.3 100.0 100.8 99.8 99.3 98.7 |
|
Figure 4.5.1.1. Tetrachloroethylene refrigerated storage test on charcoal tubes, 240-minute samples at 100 ppm. |
Figure 4.5.1.2. Tetrachloroethylene ambient storage test on charcoal tubes, 240-minute samples at 100 ppm. |
Table 4.5.2 TWA Storage Tests for Trichloroethylene (Charcoal Tubes)
|
time (days) |
refrigerated storage recovery (%) |
|
ambient storage recovery (%) |
|
0 0 6 10 13 15 17 |
99.7 99.6 101.6 98.8 98.8 99.2 98.4 |
98.6 100.2 98.9 99.4 99.8 99.3 100.4 |
100.7 100.2 100.0 100.4 99.8 98.9 99.3 |
|
99.7 99.6 100.9 98.8 98.2 99.0 99.2 |
98.6 100.2 97.8 98.6 99.2 98.5 98.4 |
100.7 100.2 99.3 100.5 99.0 98.7 98.0 |
|
Figure 4.5.2.1. Trichloroethylene refrigerated storage test on charcoal tubes, 240-minute samples at 100 ppm. |
Figure 4.5.2.2. Trichloroethylene ambient storage test on charcoal tubes, 240-minute samples at 100 ppm. |
Table 4.5.3 Ceiling Storage Tests for Tetrachloroethylene (Charcoal Tubes)
|
time (days) |
refrigerated storage recovery (%) |
|
ambient storage recovery (%) |
|
0 0 6 10 13 15 17 |
101.8 100.9 101.8 100.9 101.1 102.4 99.8 |
100.0 101.1 100.7 102.3 99.4 100.5 100.1 |
101.2 99.7 95.1 99.3 99.1 102.2 103.1 |
|
101.8 100.9 101.0 99.8 100.6 101.0 100.9 |
100.0 101.1 100.1 100.1 99.5 99.4 100.6 |
101.2 99.7 101.7 100.2 100.0 98.8 99.9 |
|
Figure 4.5.3.1. Tetrachloroethylene refrigerated storage test on charcoal tubes, 5-minute samples at 200 ppm. |
Figure 4.5.3.2. Tetrachloroethylene ambient storage test on charcoal tubes, 5-minute samples at 200 ppm. |
Table 4.5.4 Ceiling Storage Tests for Trichloroethylene (Charcoal Tubes)
|
time (days) |
refrigerated storage recovery (%) |
|
ambient storage recovery (%) |
|
0 0 6 10 13 15 17 |
99.3 100.0 100.3 99.9 99.4 101.2 98.5 |
99.8 99.7 99.4 101.2 98.2 100.5 99.0 |
100.0 98.5 99.4 97.9 97.4 100.8 101.6 |
|
99.3 100.0 99.6 98.4 99.5 99.8 99.4 |
99.8 99.7 99.2 98.9 98.3 98.8 99.0 |
100.0 98.5 100.4 100.0 98.4 98.4 98.3 |
|
Figure 4.5.4.1. Trichloroethylene refrigerated storage test on charcoal tubes, 5-minute samples at 200 ppm. |
Figure 4.5.4.2. Trichloroethylene ambient storage test on charcoal tubes, 5-minute samples at 200 ppm. |
Table 4.5.5 Peak Storage Tests for Tetrachloroethylene (Charcoal Tubes)
|
time (days) |
refrigerated storage recovery (%) |
|
ambient storage recovery (%) |
|
0 0 6 10 13 15 17 |
99.6 96.5 97.6 95.9 95.3 100.1 98.5 |
98.6 96.9 96.0 97.0 97.8 99.2 97.5 |
96.9 97.1 96.8 97.7 96.3 99.1 95.7 |
|
99.6 96.5 97.4 97.2 95.2 99.1 97.5 |
98.6 96.9 96.4 97.6 95.4 100.0 98.7 |
96.9 97.1 96.7 97.3 94.3 96.2 97.4 |
|
Figure 4.5.5.1. Tetrachloroethylene refrigerated storage test on charcoal tubes, 1-minute samples at 300 ppm. |
Figure 4.5.5.2. Tetrachloroethylene ambient storage test on charcoal tubes, 1-minute samples at 300 ppm. |
Table 4.5.6 Peak Storage Tests for Trichloroethylene (Charcoal Tubes)
|
time (days) |
refrigerated storage recovery (%) |
|
ambient storage recovery (%) |
|
0 0 6 10 13 15 17 |
97.5 94.7 98.0 95.7 96.8 98.3 96.6 |
97.5 95.2 96.8 97.4 99.0 96.9 95.5 |
96.6 95.5 98.9 98.1 96.2 97.1 96.2 |
|
97.5 94.7 97.6 97.7 96.3 97.2 96.6 |
97.5 95.2 96.3 98.0 96.9 97.5 96.1 |
96.6 95.5 97.2 97.4 95.9 97.3 95.5 |
|
Figure 4.5.6.1. Trichloroethylene refrigerated storage test on charcoal tubes, 1-minute samples at 300 ppm. |
Figure 4.5.6.2. Trichloroethylene ambient storage test on charcoal tubes, 1-minute samples at 300 ppm. |
Table 4.5.7 TWA Storage Tests for Tetrachloroethylene (SKC 575-002 Samplers)
|
time (days) |
refrigerated storage recovery (%) |
|
ambient storage recovery (%) |
|
0 0 6 10 13 15 17 |
102.3 99.1 96.6 102.2 99.3 99.3 98.2 |
100.7 99.4 100.2 96.1 98.6 95.2 101.5 |
100.3 96.6 97.4 95.3 98.4 98.6 102.1 |
|
102.3 99.1 99.9 94.6 99.6 100.2 97.4 |
100.7 99.4 99.8 96.9 98.1 103.6 99.2 |
100.3 96.6 94.1 97.0 100.3 96.2 100.0 |
|
Figure 4.5.7.1. Tetrachloroethylene refrigerated storage test SKC 575-002, 240-minute samples at 100 ppm. |
Figure 4.5.7.2. Tetrachloroethylene ambient storage test on SKC 575-002, 240-minute samples at 100 ppm. |
Table 4.5.8 TWA Storage Tests for Trichloroethylene (SKC 575-002 Samplers)
|
time (days) |
refrigerated storage recovery (%) |
|
ambient storage recovery (%) |
|
0 0 6 10 13 15 17 |
103.0 99.1 94.7 100.7 97.4 97.7 97.0 |
100.2 98.8 98.6 94.2 96.9 94.0 100.1 |
99.2 95.1 95.0 93.3 96.4 97.2 100.6 |
|
103.0 99.1 97.4 92.9 98.2 98.6 96.8 |
100.2 98.8 97.0 95.2 96.4 102.2 97.9 |
99.2 95.1 92.0 95.2 98.0 93.8 98.3 |
|
Figure 4.5.8.1. Trichloroethylene refrigerated storage test SKC 575-002, 240-minute samples at 100 ppm. |
Figure 4.5.8.2. Trichloroethylene ambient storage test on SKC 575-002, 240-minute samples at 100 ppm. |
Table 4.5.9 Ceiling Storage Tests for Tetrachloroethylene (SKC 575-002 Samplers)
|
time (days) |
refrigerated storage recovery (%) |
|
ambient storage recovery (%) |
|
0 0 6 10 13 15 17 |
102.1 100.2 97.5 94.1 93.4 98.2 105.0 |
114.1 113.9 99.1 95.3 95.2 99.4 103.8 |
113.0 103.5 105.2 100.9 105.7 105.8 102.8 |
|
102.1 100.2 98.0 98.8 97.5 99.4 99.1 |
114.1 113.9 100.4 104.0 96.6 100.8 97.1 |
113.0 103.5 108.2 100.2 98.7 98.3 99.4 |
|
Figure 4.5.9.1. Tetrachloroethylene refrigerated storage test SKC 575-002 Samplers, 5-minute at 200 ppm. |
Figure 4.5.9.2. Tetrachloroethylene ambient storage test on SKC 575-002 Samplers, 5-minute at 200 ppm. |
Table 4.5.10 Ceiling Storage Tests for Trichloroethylene (SKC 575-002 Samplers)
|
time (days) |
refrigerated storage recovery (%) |
|
ambient storage recovery (%) |
|
0 0 6 10 13 15 17 |
102.5 102.0 99.9 94.5 97.5 99.6 106.8 |
106.1 105.6 102.1 98.0 100.1 101.4 104.4 |
102.0 97.6 96.3 99.1 104.1 105.1 105.4 |
|
102.5 102.0 100.8 95.3 98.2 99.2 98.9 |
106.1 105.6 103.7 101.0 96.7 100.4 99.4 |
102.0 97.6 96.5 94.7 94.8 97.8 99.3 |
|
Figure 4.5.10.1. Trichloroethylene refrigerated storage test SKC 575-002 Samplers, 5-minute at 200 ppm. |
Figure 4.5.10.2. Trichloroethylene ambient storage test on SKC 575-002 Samplers, 5-minute at 200 ppm. |
-
4.6 Reproducibility
Six samples for both samplers at the TWA and ceiling target concentrations and six charcoal tube
samples at the peak target concentrations that were collected from controlled test atmospheres,
along with a draft copy of this procedure, were submitted to an SLTC service branch for analysis. All
of samples were stored at room temperature and analyzed 22 days after generation. No sample
result had a deviation greater than the precisions of the overall procedure determined in Section 4.4.
Table 4.6.1 Reproducibility Data for Tetrachloroethylene Charcoal Tube Samples at the TWA Level
|
sample no. |
mg/m3 reported |
mg/m3 expected |
percent found |
deviation (%) |
|
1 2 3 4 5 6 |
652.7 647.6 649.2 647.2 634.6 649.4 |
668.5 668.5 668.5 668.5 668.5 668.5 |
97.6 96.9 97.1 96.8 94.9 97.1 |
-2.4 -3.1 -2.9 -3.2 -4.1 -2.9 |
|
| |
Table 4.6.2 Reproducibility Data for Tetrachloroethylene Charcoal Tube Samples at the Ceiling Level
|
sample no. |
mg/m3 reported |
mg/m3 expected |
percent found |
deviation (%) |
|
1 2 3 4 5 6 |
1365 1314 1324 1326 1323 1313 |
1337 1337 1337 1337 1337 1337 |
102.1 98.3 99.0 99.2 99.0 98.2 |
+2.1 -1.7 -1.0 -0.8 -1.0 -1.8 |
|
|
|
Table 4.6.3 Reproducibility Data for Tetrachloroethylene Charcoal Tube Samples at the Peak Level
|
sample no. |
mg/m3 reported |
mg/m3 expected |
percent found |
deviation (%) |
|
1 2 3 4 5 6 |
1961 1970 1957 1969 1974 1976 |
2005 2005 2005 2005 2005 2005 |
97.8 98.3 97.6 98.2 98.5 98.6 |
-2.2 -1.7 -2.4 -1.8 -1.5 -1.4 |
|
| |
Table 4.6.4 Reproducibility Data for Trichloroethylene Charcoal Tube Samples at the TWA Level
|
sample no. |
mg/m3 reported |
mg/m3 expected |
percent found |
deviation (%) |
|
1 2 3 4 5 6 |
459.1 457.4 457.4 455.4 445.5 456.5 |
479.2 479.2 479.2 479.2 479.2 479.2 |
95.8 95.5 95.5 95.0 93.0 95.3 |
-4.2 -4.5 -4.5 -5.0 -7.0 -4.7 |
|
|
|
Table 4.6.5 Reproducibility Data for Trichloroethylene Charcoal Tube Samples at the Ceiling Level
|
sample no. |
mg/m3 reported |
mg/m3 expected |
percent found |
deviation (%) |
|
1 2 3 4 5 6 |
1001 954.2 965.2 964.2 958.4 951.2 |
990.0 990.0 990.0 990.0 990.0 990.0 |
101.1 96.4 97.5 97.4 96.8 96.1 |
+1.1 -3.6 +-2.5 -2.6 -3.2 -3.9 |
|
| |
Table 4.6.6 Reproducibility Data for Trichloroethylene Charcoal Tube Samples at the Peak Level
|
sample no. |
mg/m3 reported |
mg/m3 expected |
percent found |
deviation (%) |
|
1 2 3 4 5 6 |
1385 1380 1388 1397 1404 1404 |
1463 1463 1463 1463 1463 1463 |
94.7 94.3 94.9 95.5 96.0 96.0 |
-5.3 -5.7 -5.1 -4.5 -4.0 -4.0 |
|
|
|
Table 4.6.7 Reproducibility Data for Tetrachloroethylene SKC 575-002 Samples at the TWA Level
|
sample no. |
mg/m3 reported |
mg/m3 expected |
percent found |
deviation (%) |
|
1 2 3 4 5 6 |
663.2 638.4 646.4 644.4 635.1 649.8 |
668.5 668.5 668.5 668.5 668.5 668.5 |
99.2 95.5 96.7 96.4 95.0 97.2 |
-0.8 -4.5 -3.3 -3.6 -5.0 -2.8 |
|
| |
Table 4.6.8 Reproducibility Data for Tetrachloroethylene SKC 575-002 Samples at the Ceiling Level
|
sample no. |
mg/m3 reported |
mg/m3 expected |
percent found |
deviation (%) |
|
1 2 3 4 5 6 |
1415 1326 1382 1499 1320 1360 |
1337 1337 1337 1337 1337 1337 |
105.8 99.2 103.4 112.1 98.7 101.7 |
+5.8 -0.8 +3.4 +12.1 -1.3 +1.7 |
|
|
|
Table 4.6.9 Reproducibility Data for Trichloroethylene SKC 575-002 Samples at the TWA Level
|
sample no. |
mg/m3 reported |
mg/m3 expected |
percent found |
deviation (%) |
|
1 2 3 4 5 6 |
471.5 450.4 455.7 456.2 446.6 458.1 |
479.2 479.2 479.2 479.2 479.2 479.2 |
98.4 94.0 95.1 95.2 93.2 95.6 |
-2.6 -6.0 -4.9 -4.8 -6.8 -4.4 |
|
| |
Table 4.6.10 Reproducibility Data for Trichloroethylene SKC 575-002 Samples at the Ceiling Level
|
sample no. |
mg/m3 reported |
mg/m3 expected |
percent found |
deviation (%) |
|
1 2 3 4 5 6 |
1027 969.2 1006 1062 979.1 982.1 |
990.0 990.0 990.0 990.0 990.0 990.0 |
103.7 97.9 101.6 107.3 98.9 99.2 |
+3.7 -2.1 +1.6 +7.3 -1.1 -0.8 |
|
|
-
4.7 Sampler capacity and sampling rate
4.7.1 Charcoal tubes
The sampling capacity of the front section of charcoal sampling tubes was tested by
sampling from a dynamically generated test atmosphere of tetrachloroethylene at 192.7 ppm
(1306 mg/m3) and from another atmosphere containing 192.0 ppm (1031 mg/m3) of
trichloroethylene). The samples were collected at a nominal flow rate of 50 mL/min and the
mean relative humidity of the tetrachloroethylene atmosphere was 65% at 26.1°C and the
trichloroethylene atmosphere was 64.5% RH at 25.8°C. Complete charcoal tubes were
placed in-line behind the front test sections and changed at measured intervals to check for
breakthrough. No significant breakthrough (<0.1%) occurred when sampling from the
tetrachloroethylene atmosphere for 9.5 hours (28.5 liters) which is equivalent to about 37
mg of tetrachloroethylene. The mean 5% breakthrough volume from the trichloroethylene
atmosphere was determined to be about 23.3 liters (7.8 hours), which is equivalent to
approximately 24 mg.
Table 4.7.1
Breakthrough of Trichloroethylene from
Charcoal Sampling Tubes when Sampling
from a 1031 mg/m3 Atmosphere
|
sampler tube no. |
air vol (L) |
sampling time (min) |
downstream concn (mg/m3) |
break- through (%) |
|
1
2
3 |
6.1 13.7 16.7 19.7 22.4 24.3
5.8 13.1 16.0 18.9 21.4 23.2
6.3 14.2 17.4 20.5 23.3 25.3 |
0-240 240-300 300-360 360-420 420-465 465-495
0-240 240-300 300-360 360-420 420-465 465-495
0-240 240-300 300-360 360-420 420-465 465-495 |
0 0 0.06 5.19 44.6 119.7
0 0 0 0.56 5.86 21.4
0 0.02 0.30 8.03 50.3 119.6 |
0 0 0.01 0.50 4.33 11.6
0 0 0 0.05 0.57 2.08
0 0.002 0.03 0.78 4.88 11.6 |
|
|
Figure 4.7.1. Determination of the 5% breakthrough volume for trichloroethylene collected with charcoal tubes.
|
4.7.2 SKC 575-002 Samplers
The sampling rates and capacities were determined by taking samples from a dynamically
generated test atmosphere of tetrachloroethylene (nominal concentration of 200 ppm or 1356
mg/m3) and trichloroethylene (nominal concentration of 200 ppm or 1074 mg/m3) for
increasing time intervals. The atmospheres were at approximately 70% relative humidity,
25°C and 660 mmHg. The flow of the atmospheres through the exposure chamber provided
for face velocities of approximately 0.4 m/s on the SKC 575-002 Samplers. Three samples
were taken for each sampling interval. Sampler capacity is exceeded when the sampling
rate decreases rapidly. Because this did not occur for the time period tested, the capacity
is determined to be greater than 12 mg for tetrachloroethylene and greater than 10 mg for
trichloroethylene. Sampling rates standardized to 760 mmHg and 25°C of 13.06 mL/min
for tetrachloroethylene and 14.24 mL/min for trichloroethylene were determined from the
mean sample rates from samples collected from 5.0 to 600 minutes.
Table 4.7.2.1 Sampling Rate and Capacity for Tetrachloroethylene using SKC 575-002 Samplers
|
sampling time (min) |
sampling rate (mL/min) |
RSD (%) |
|
5 10 30 60 120 180 240 360 480 600 |
12.61 12.99 12.63 12.73 13.13 12.77 13.55 13.23 13.43 13.50 |
1.29 1.01 1.02 1.71 3.04 3.06 2.53 1.38 1.16 0.94 |
|
| |
Figure 4.7.2.1 Determination of sampling rate and capacity for tetrachloroethylene using SKC 575-002 Samplers.
|
|
Table 4.7.2.2 Sampling Rate and Capacity for Trichloroethylene using SKC 575-002 Samplers
|
sampling time (min) |
sampling rate (mL/min) |
RSD (%) |
|
5 10 30 60 120 180 240 360 480 600 |
14.18 14.10 13.99 13.87 14.10 13.73 14.75 14.41 14.60 14.65 |
3.97 1.11 4.20 1.48 2.71 3.15 2.66 1.60 0.54 0.81 |
|
| |
Figure 4.7.2.2 Determination of sampling rate and capacity for trichloroethylene using SKC 575-002 Samplers.
|
-
4.8 Extraction efficiency
4.8.1 Charcoal tubes
The extraction efficiencies of tetrachloroethylene and trichloroethylene were determined by
liquid-spiking 100-mg portions of charcoal with the analytes at the RQL and 0.05 to 2 times
the target concentrations for each of three levels. These samples were stored overnight at
ambient temperature and then extracted and analyzed. The mean extraction efficiency over
the ranges of the RQL to 2 times the TWA, ceiling and peak target concentration levels are
98.8%, 98.3% and 98.5% respectively for tetrachloroethylene and 99.3%, 98.8% and 99.0%
for trichloroethylene. The mean extraction efficiencies for the RQL and all of the eighteen
other concentrations are 98.5% for tetrachloroethylene and 99.1% for trichloroethylene.
Extraction efficiencies from wet charcoal were also determined by drawing the recommended
volume (12, 0.25 and 0.05 liters for TWA, ceiling and peak levels respectively) of humid air
at approximately 80% RH, 22.2°C before the charcoal was spiked with the analytes. The
mean respective extraction efficiencies for tetrachloroethylene and trichloroethylene are
98.9% and 99.5% for TWA levels, 98.6% and 98.9% for ceiling levels and 98.6% and 98.8%
for peak levels.
Table 4.8.1.1 Extraction Efficiency of Tetrachloroethylene from Charcoal Tubes (TWA Levels)
|
level
|
sample number
|
|
× target concn |
µg per sample |
1 | 2 | 3 | 4 | 5 | 6 |
mean |
|
(RQL) 0.05 0.1 0.2 0.5 1.0 2.0
1.0 (wet) |
1.22 405.8 811.5 1623 4058 8115 16230
8115 |
97.5 97.6 98.4 98.5 98.5 98.8 99.4
98.6 |
100.1 97.9 99.0 98.5 98.5 98.4 99.4
98.8 |
97.8 97.8 98.1 98.8 98.6 98.9 99.4
98.8 |
100.0 98.0 98.6 98.9 98.5 98.7 99.7
98.8 |
101.5 97.7 98.3 98.6 98.6 98.8 99.3
98.8 |
96.8 98.2 98.5 98.7 98.5 98.7 99.5
99.1 |
99.5 97.9 98.5 98.7 98.5 98.7 99.4
98.9 |
|
Table 4.8.1.2 Extraction Efficiency of Tetrachloroethylene from Charcoal Tubes (Ceiling Levels)
|
level
|
sample number
|
× target concn |
µg per sample |
1 | 2 | 3 | 4 | 5 | 6 | mean |
|
(RQL) 0.05 0.1 0.2 0.5 1.0 2.0
1.0 (wet) |
1.22 16.88 33.76 67.52 168.8 337.6 675.2 337.6 |
97.5 97.7 98.9 97.6 97.9 98.2 98.8
98.7 |
100.1 98.0 97.7 98.5 96.7 98.6 98.9
98.4 |
97.8 97.9 99.0 98.2 97.9 98.5 98.6
98.9 |
100.0 98.7 98.6 97.9 96.7 98.7 99.3
98.7 |
101.5 96.1 99.0 94.6 97.1 98.3 99.1
98.3 |
96.8 95.6 98.8 98.4 96.4 98.5 98.8
98.5 |
99.5 97.3 98.7 98.2 97.1 98.5 98.9
98.6 |
|
Table 4.8.1.3 Extraction Efficiency of Tetrachloroethylene from Charcoal Tubes (Peak Levels)
|
level
|
sample number
|
× target concn |
µg per sample |
1 | 2 | 3 | 4 | 5 | 6 | mean |
|
(RQL) 0.05 0.1 0.2 0.5 1.0 2.0
1.0 (wet) |
1.22 5.03 10.06 20.13 50.30 100.6 201.3
100.6 |
97.5 98.1 99.4 97.2 97.6 99.0 98.5
98.8 |
100.1 101.8 96.3 97.5 98.0 98.9 99.0
98.8 |
101.3 98.0 95.8 99.3 97.4 98.8 99.2
98.4 |
100.0 97.3 96.0 98.1 98.9 98.7 99.1
98.4 |
101.5 99.6 97.6 97.9 98.3 96.6 99.2
98.6 |
96.8 100.2 98.3 98.8 98.2 98.4 98.6
98.9 |
99.5 99.2 97.2 98.1 98.4 98.4 98.9
98.6 |
|
Table 4.8.1.4 Extraction Efficiency of Trichloroethylene from Charcoal Tubes (TWA Levels)
|
level
|
sample number
|
× target concn |
µg per sample |
1 | 2 | 3 | 4 | 5 | 6 | mean |
|
(RQL) 0.05 0.1 0.2 0.5 1.0 2.0
1.0 (wet) |
0.96 321.9 643.7 1287 3219 6437 12870
6437 |
94.9 98.6 100.0 99.4 98.2 98.6 100.0
99.2 |
97.5 98.9 100.3 99.2 98.9 99.4 100.0
99.4 |
101.5 98.8 99.2 99.6 98.8 99.5 100.2
99.7 |
94.6 98.8 99.8 100.2 98.3 99.3 100.0
99.8 |
102.4 98.7 99.8 99.7 99.4 99.2 99.8
99.4 |
101.6 99.2 99.5 99.8 98.2 99.1 99.9
99.6 |
98.8 98.8 99.8 99.6 98.6 99.2 100.0
99.5 |
|
Table 4.8.1.5 Extraction Efficiency of Trichloroethylene from Charcoal Tubes (Ceiling Levels)
|
level
|
sample number
|
× target concn |
µg per sample |
1 | 2 | 3 | 4 | 5 | 6 | mean |
|
(RQL) 0.05 0.1 0.2 0.5 1.0 2.0
1.0 (wet) |
0.96 13.39 26.78 53.56 133.9 267.8 535.6 267.8 |
94.9 96.7 98.7 98.2 98.7 99.2 99.4
99.4 |
97.5 98.3 99.5 99.7 98.0 99.3 99.6
98.8 |
101.5 98.2 99.8 99.5 99.2 99.4 99.4
98.9 |
94.6 98.6 98.4 98.3 97.6 99.4 99.9
99.0 |
102.4 99.5 98.3 98.9 97.9 99.1 99.9
98.6 |
98.6 96.7 98.7 98.7 97.2 99.7 99.4
98.9 |
98.8 98.0 98.9 98.9 98.1 99.4 99.6
98.9 |
|
Table 4.8.1.6 Extraction Efficiency of Trichloroethylene from Charcoal Tubes (Peak Levels)
|
level
|
sample number
|
× target concn |
µg per sample |
1 | 2 | 3 | 4 | 5 | 6 | mean |
|
(RQL) 0.05 0.1 0.2 0.5 1.0 2.0
1.0 (wet) |
0.96 3.99 7.98 15.96 39.91 79.82 159.6
79.82 |
94.9 103.2 98.2 100.0 99.2 100.3 99.3
99.8 |
97.5 100.3 97.5 99.6 98.3 99.7 99.8
98.8 |
101.5 93.7 100.8 100.0 98.9 99.6 99.9
98.9 |
94.6 98.3 100.7 98.4 98.0 99.2 99.7
98.6 |
102.4 101.1 98.8 98.6 97.4 97.7 100.2
98.4 |
98.6 96.4 99.1 98.1 98.5 99.4 99.2
98.6 |
98.8 98.8 99.2 99.1 98.4 99.3 99.7
98.8 |
|
Stability of extracted charcoal tubes samples
The stability of extracted samples was investigated by reanalyzing the target concentration
samples 24 h after the initial analyses. After the original analyses for each target
concentration level were performed, three vials from each set were recapped with new septa,
while the remaining three retained their punctured septa. The samples were reanalyzed with
fresh standards. The mean percent change was 0.0%, -0.5% and -0.2% for TWA, ceiling
and peak tetrachloroethylene samples respectively that were resealed with new septa and
-0.7%, -1.6% and -1.4% for those that retained their punctured septa. The mean percent
change was +0.7%, -0.5% and -0.5% for TWA, ceiling and peak trichloroethylene samples
respectively that were resealed with new septa and -3.6%, -5.6% and -5.4% for those that
retained their punctured septa.
Table 4.8.1.7 Stability of Extracted Tetrachloroethylene Samples (TWA)
|
punctured septa replaced
|
|
punctured septa retained
|
initial DE (%) |
DE after one day (%) |
difference |
|
initial DE (%) |
DE after one day (%) |
difference |
|
98.8 98.4 98.9 98.7 |
98.6 98.8 98.8 (means) 98.7 |
-0.2 +0.4 -0.1 0.0 |
|
98.7 98.8 98.7 98.7 |
97.7 98.1 98.2 (means) 98.0 |
-1.0 -0.7 -0.5 -0.7 |
|
Table 4.8.1.8 Stability of Extracted Tetrachloroethylene Samples (Ceiling)
|
punctured septa replaced
|
|
punctured septa retained
|
initial DE (%) |
DE after one day (%) |
difference |
|
initial DE (%) |
DE after one day (%) |
difference |
|
98.2 98.6 98.5 98.4 |
97.9 98.0 98.0 (means) 98.0 |
-0.3 -0.6 -0.5 -0.5 |
|
98.7 98.3 98.5 98.5 |
96.8 97.0 96.8 (means) 96.9 |
-1.9 -1.3 -1.7 -1.6 |
|
Table 4.8.1.9 Stability of Extracted Tetrachloroethylene Samples (Peak)
|
punctured septa replaced
|
|
punctured septa retained
|
initial DE (%) |
DE after one day (%) |
difference |
|
initial DE (%) |
DE after one day (%) |
difference |
|
99.0 98.9 98.8 98.9 |
98.8 98.8 98.4 (means) 98.7 |
-0.2 -0.1 -0.4 -0.2 |
|
98.7 96.6 98.4 97.9 |
97.1 95.0 97.3 (means) 96.5 |
-1.6 -1.6 -1.1 -1.4 |
|
Table 4.8.1.10 Stability of Extracted Trichloroethylene Samples (TWA)
|
punctured septa replaced
|
|
punctured septa retained
|
initial DE (%) |
DE after one day (%) |
difference |
|
initial DE (%) |
DE after one day (%) |
difference |
|
98.6 99.4 99.5 99.2 |
98.8 100.6 100.2 (means) 99.9 |
+0.2 +1.2 +0.7 +0.7 |
|
99.3 99.2 99.1 99.2 |
94.8 95.6 96.4 (means) 95.6 |
-4.5 -3.6 -2.7 -3.6 |
|
Table 4.8.1.11 Stability of Extracted Trichloroethylene Samples (Ceiling)
|
punctured septa replaced
|
|
punctured septa retained
|
initial DE (%) |
DE after one day (%) |
difference |
|
initial DE (%) |
DE after one day (%) |
difference |
|
99.2 99.3 99.4 99.3 |
98.7 98.9 98.8 (means) 98.8 |
-0.5 -0.4 -0.6 -0.5 |
|
99.4 99.1 99.7 99.4 |
93.6 94.5 93.4 (means) 93.8 |
-5.8 -4.6 -6.3 -5.6 |
|
Table 4.8.1.12 Stability of Extracted Trichloroethylene Samples (Peak)
|
punctured septa replaced
|
|
punctured septa retained
|
initial DE (%) |
DE after one day (%) |
difference |
|
initial DE (%) |
DE after one day (%) |
difference |
|
100.3 99.7 99.6 99.9 |
99.4 99.4 99.4 (means) 99.4 |
-0.9 -0.3 -0.2 -0.5 |
|
99.2 97.7 99.4 98.8 |
94.4 91.8 93.9 (means) 93.4 |
-4.8 -5.9 -5.5 -5.4 |
|
4.8.2 SKC 575-002 Samplers
The extraction efficiencies of tetrachloroethylene and trichloroethylene were determined
by liquid-spiking 500-mg portions of Anasorb 747 with the analytes
at 0.05 to 2 times the target concentrations. These samples were stored overnight at ambient
temperature and then extracted and analyzed. The mean extraction efficiency over the ranges
of the RQL to 2 times the TWA and ceiling levels are 96.2% and 94.9% respectively for
tetrachloroethylene and 97.7% and 97.3% for trichloroethylene. The mean extraction efficiencies
for the RQL and all of the twelve other concentrations are 95.4% for tetrachloroethylene and
97.5% for trichloroethylene.
Extraction efficiencies from wet Anasorb 747 were also determined by exposing the
samplers for the recommended times (240 and 5 minutes for TWA and ceiling levels
respectively) to humid air at approximately 80% RH, 22.2°C before the samplers were
spiked with the analytes. The mean respective extraction efficiencies for tetrachloroethylene
and trichloroethylene are 93.2% and 95.3% for TWA levels and 94.0% and 96.0% for ceiling
levels.
Table 4.8.2.1 Extraction Efficiency of Tetrachloroethylene from SKC 575-002 Samplers (TWA Levels)
|
level
|
sample number
|
|
× target concn |
µg per sample |
1 | 2 | 3 | 4 | 5 | 6 | mean |
|
(RQL) 0.05 0.1 0.2 0.5 1.0 2.0
1.0 (wet) |
2.43 107.1 214.2 428.5 1071 2142 4285
2142 |
96.0 96.5 96.1 95.3 96.9 95.8 96.7
93.3 |
96.6 95.9 95.6 95.2 96.8 96.4 96.2
92.9 |
97.5 95.7 95.1 94.8 96.7 96.0 96.6
93.1 |
101.4 95.6 95.4 94.6 96.1 96.4 96.4
92.9 |
97.7 96.4 95.8 95.0 96.5 96.5 96.6
93.1 |
94.1 96.0 95.6 95.0 96.7 96.0 96.7
93.8 |
97.2 96.0 95.6 95.0 96.6 96.2 96.5
93.2 |
|
Table 4.8.2.2 Extraction Efficiency of Tetrachloroethylene from SKC 575-002 Samplers (Ceiling Levels)
|
level
|
sample number
|
|
× target concn |
µg per sample |
1 | 2 | 3 | 4 | 5 | 6 | mean |
|
(RQL) 0.05 0.1 0.2 0.5 1.0 2.0
1.0 (wet) |
2.43 4.38 8.76 17.5 43.8 87.6 175.2
87.6 |
96.0 94.5 96.4 96.3 90.0 94.7 92.5
94.3 |
96.6 95.6 93.3 95.3 91.1 94.6 91.5
94.4 |
97.5 98.3 94.0 98.0 91.5 95.2 90.6
93.9 |
101.4 95.0 98.2 96.4 93.2 94.9 91.0
94.2 |
97.7 97.1 92.8 98.0 91.8 93.8 92.6
93.5 |
94.1 98.0 96.6 97.1 93.3 95.1 92.4
93.6 |
97.2 96.4 95.2 96.8 91.8 94.7 91.8
94.0 |
|
Table 4.8.2.3 Extraction Efficiency of Trichloroethylene from SKC 575-002 Samplers (TWA Levels)
|
level
|
sample number
|
|
× target concn |
µg per sample |
1 | 2 | 3 | 4 | 5 | 6 | mean |
|
(RQL) 0.05 0.1 0.2 0.5 1.0 2.0
1.0 (wet) |
1.23 96.56 193.1 386.2 965.6 1931 3862
1931 |
101.2 98.1 97.9 97.6 98.7 97.6 98.0 95.4 |
94.1 97.7 97.4 97.4 98.1 98.2 97.0
94.7 |
98.6 97.7 96.7 97.3 98.2 97.4 98.4
95.2 |
99.2 97.1 97.1 96.8 97.6 98.1 97.9
95.1 |
94.5 98.1 97.7 97.3 98.0 98.1 98.2
95.4 |
98.0 98.0 97.8 97.2 98.3 97.4 98.5
95.8 |
97.6 97.9 97.4 97.3 98.2 97.8 98.0
95.3 |
|
Table 4.8.2.4 Extraction Efficiency of Trichloroethylene from SKC 575-002 Samplers (Ceiling Levels)
|
level
|
sample number
|
|
× target concn |
µg per sample |
1 | 2 | 3 | 4 | 5 | 6 | mean |
|
(RQL) 0.05 0.1 0.2 0.5 1.0 2.0
1.0 (wet) |
1.23 3.95 7.90 15.80 39.50 79.00 158.0
79.00 |
101.2 95.8 99.6 96.7 93.5 97.1 94.8
96.1 |
94.1 102.7 99.8 97.1 92.8 96.4 94.3
97.4 |
98.6 99.0 97.1 96.7 95.5 96.4 93.5
97.7 |
99.2 102.9 102.4 97.8 96.7 96.3 94.0
96.0 |
94.5 102.1 99.6 98.5 94.9 95.8 95.0
95.7 |
98.0 98.9 101.0 98.9 96.4 96.8 94.7
95.2 |
97.6 100.2 99.9 97.6 95.0 96.5 94.4
96.0 |
|
Stability of extracted SKC 575-002 Samplers
The stability of extracted samples was investigated by reanalyzing the target concentration
samples 24 h after the initial analyses. After the original analyses for each target
concentration level were performed, three vials from each set were recapped with new septa,
while the remaining three retained their punctured septa. The samples were reanalyzed with
fresh standards. The mean percent change was -0.4% and +1.0% for TWA and ceiling
tetrachloroethylene samples respectively that were resealed with new septa and -1.0% and
+0.3% for those that retained their punctured septa. The mean percent change was -0.4%
and +0.9% for TWA and ceiling trichloroethylene samples respectively that were resealed
with new septa and 4.0% and 1.6% for those that retained their punctured septa.
Table 4.8.2.5 Stability of Extracted Tetrachloroethylene Samples (TWA)
|
punctured septa replaced
|
|
punctured septa retained
|
initial EE (%) |
EE after one day (%) |
difference |
|
initial EE (%) |
EE after one day (%) |
difference |
|
95.8 96.4 96.0 96.1 |
95.1 96.2 95.8 (means) 95.7 |
-0.7 -0.2 -0.2 -0.4 |
|
96.4 96.5 96.0 96.3 |
95.0 95.5 95.4 (means) 95.3 |
-1.4 -1.0 -0.6 -1.0 |
|
Table 4.8.2.6 Stability of Extracted Tetrachloroethylene Samples (Ceiling)
|
punctured septa replaced
|
|
punctured septa retained
|
initial EE (%) |
EE after one day (%) |
difference |
|
initial EE (%) |
EE after one day (%) |
difference |
|
94.7 94.6 95.2 94.8 |
95.6 95.8 96.0 (means) 95.8 |
+0.9 +1.2 +0.8 +1.0 |
|
94.9 93.8 95.1 94.6 |
95.1 95.0 94.7 (means) 94.9 |
+0.2 +1.2 -0.4 +0.3 |
|
Table 4.8.2.7 Stability of Extracted Trichloroethylene Samples (TWA)
|
punctured septa replaced
|
|
punctured septa retained
|
initial EE (%) |
EE after one day (%) |
difference |
|
initial EE (%) |
EE after one day (%) |
difference |
|
97.6 98.2 97.4 97.7 |
97.0 97.9 97.1 (means) 97.3 |
-0.6 -0.3 -0.3 -0.4 |
|
98.1 98.1 97.4 97.9 |
92.2 94.9 94.7 (means) 93.9 |
-5.9 -3.2 -2.7 -4.0 |
|
Table 4.8.2.8 Stability of Extracted Trichloroethylene Samples (Ceiling)
|
punctured septa replaced
|
|
punctured septa retained
|
initial EE (%) |
EE after one day (%) |
difference |
|
initial EE (%) |
EE after one day (%) |
difference |
|
97.1 96.4 96.4 96.6 |
97.4 97.4 97.7 (means) 97.5 |
+0.3 +1.0 +1.3 +0.9 |
|
96.3 95.8 96.8 96.3 |
94.5 94.8 94.7 (means) 94.7 |
-1.8 -1.0 -2.1 -1.6 |
|
4.9 Interferences (sampling)
4.9.1 Interference studies were performed by sampling for 240 minutes from a test atmosphere
(70% RH, 24°C, 659 mmHg) containing 100 ppm of tetrachloroethylene with 83 ppm of
methylene chloride, 70 ppm of isopropyl alcohol, 71 ppm of 2-butanone (MEK), 26 ppm of
butyl acetate, 17 ppm of dioxane and 16 ppm of amyl acetate. Another test was done by
sampling for 240 minutes from a test atmosphere (63% RH, 27°C, 659 mmHg) containing
100 ppm of trichloroethylene plus the aforementioned solvent vapors. The presence of these
compounds, which may represent typical substances that may be collected with
tetrachloroethylene or trichloroethylene did not have a significant effect on sample results
using either of the samplers.
Table 4.9.1 Recovery (%) from the Atmosphere Described in 4.9.1 for each Analyte
|
|
tetrachloroethylene |
trichloroethylene |
|
sample no. |
charcoal tubes |
SKC 575-002 Samplers |
charcoal tubes |
SKC 575-002 Samplers |
|
1 2 3 mean |
101.6 100.0 102.0 101.2 |
104.9 108.4 105.4 106.2 |
96.8 95.8 97.7 96.8 |
98.4 100.0 100.6 99.7 |
|
4.9.2 Reverse diffusion studies for the SKC 575-002 Samplers and stripping studies for the
charcoal tubes was performed by sampling a 193-ppm atmosphere of tetrachloroethylene
and a 194-ppm atmosphere of trichloroethylene (65% RH, 26°C, 651 mmHg) for 60 minutes
with six of each samplers. Three samplers from each set were additionally subjected to 180
minutes of the same atmosphere without the analytes present to determine if any of the
collected solvents diffused off of the SKC 575-002 Samplers and also whether it was
stripped off of the charcoal tubes. Upon analysis of the samples, the mean recovery of the
removed samplers versus the mean recovery of the samplers that were additionally exposed
to the atmosphere without analyte was greater than 90% for both samplers, indicating that
reverse diffusion and stripping is insignificant. The first three samples in Table 4.9.2 for each
sampler were used to sample the atmosphere for 60 minutes, while the last three were
additionally exposed to a blank atmosphere for another 180 minutes.
Table 4.9.2 Recovery (%) for the Circumstances Described in 4.9.2 for Each Analyte
(Samples 1-3 used to sample atmosphere for 60 minutes,
samples 4-6 for an additional 180 minutes of blank atmosphere)
|
|
tetrachloroethylene |
trichloroethylene |
|
sample no. |
charcoal tubes |
SKC 575-002 Samplers |
charcoal tubes |
SKC 575-002 Samplers |
|
1 2 3 mean (1-3) 4 5 6 mean (4-6) mean (4-6) รท mean (1-3) |
100.2 99.0 100.6 99.9 101.7 101.8 101.7 101.7 1.02 |
104.6 100.1 103.2 102.6 97.3 106.8 102.7 102.3 1.00 |
100.7 99.1 101.1 100.3 101.4 101.8 101.4 101.5 1.01 |
104.8 100.2 102.3 102.4 96.7 106.3 102.6 101.9 1.00 |
|
4.9.3 The effects from sampling from relatively dry atmospheres were investigated by sampling
from an atmosphere (3% RH, 26°C, 650 mmHg) containing 198 ppm of tetrachloroethylene
and 199 ppm of trichloroethylene for 240 minutes with both samplers. The mean results for
the both analytes collected on both samplers was greater than 90% of the theoretical
amounts.
Table 4.9.3 Recovery (%) for 240-minute Samples from a
Dry Atmosphere for Each Analyte
|
|
tetrachloroethylene |
trichloroethylene |
|
sample no. |
charcoal tubes |
SKC 575-002 Samplers |
charcoal tubes |
SKC 575-002 Samplers |
|
1 2 3 mean |
100.0 99.3 100.0 99.8 |
100.4 106.8 102.4 103.2 |
98.4 98.1 97.9 98.1 |
98.5 105.3 101.0 101.6 |
|
4.9.4 The effects from sampling from atmospheres containing low concentrations of analytes were
investigated by sampling from an atmosphere (67% RH, 26°C, 649 mmHg) containing 9.0
ppm of tetrachloroethylene and 10.7 ppm of trichloroethylene for 240 minutes with both
samplers. The mean results for both analytes collected on both samplers was greater than
90% of the theoretical amounts.
Table 4.9.4 Recovery (%) for 240-minute Samples from a
Atmosphere Containing 9.0 ppm of Tetrachloroethylene
and 10.7 ppm of Trichloroethylene
|
|
tetrachloroethylene |
trichloroethylene |
|
sample no. |
charcoal tubes |
SKC 575-002 Samplers |
charcoal tubes |
SKC 575-002 Samplers |
|
1 2 3 mean |
100.2 100.2 100.8 100.4 |
101.7 101.5 100.2 101.1 |
98.5 99.3 99.5 99.1 |
100.2 100.6 98.2 99.7 |
|
4.10 Qualitative analysis
Tetrachloroethylene and trichloroethylene can easily be identified by GC/mass spectrometry. The
following figures represent typical mass spectra of tetrachloroethylene and trichloroethylene.
Figure 4.10.1 representative mass spectrum of tetrachloroethylene.
|
Figure 4.10.1 representative mass spectrum of trihloroethylene.
|
1. Elskamp, C.J. "OSHA Method No. 111, Toluene"; OSHA Salt Lake Technical Center, Salt Lake City, UT. Unpublished work, 1998.
2. Kirk-Othmer Encyclopedia of Chemical Technology, 4th ed.; Kroschwitz, J., Ed.; John Wiley & Sons: New York, 1993, Vol. 6; pp 50-59.
3. Kirk-Othmer Encyclopedia of Chemical Technology, pp 40-49.
4. Kirk-Othmer Encyclopedia of Chemical Technology, pp 50-59.
5. Kirk-Othmer Encyclopedia of Chemical Technology, pp 40-49.
6. Kirk-Othmer Encyclopedia of Chemical Technology, pp 50-59.
7. The Merck Index, 12th ed.; Budavari, S., Ed.; Merck & Co., Whitehouse Station, NJ, 1996; 9332.
8. The Merck Index, 12th ed.; Budavari, S., Ed.; Merck & Co., Whitehouse Station, NJ, 1996; 9332.
9. Kirk-Othmer Encyclopedia of Chemical Technology, pp 40-49.
10. The Merck Index, 12th ed.; Budavari, S., Ed.; Merck & Co., Whitehouse Station, NJ, 1996; 9769.
11. "Evaluation Guidelines for Air Sampling Methods Utilizing Chromatographic Analyses", OSHA Salt Lake Technical Center, Salt Lake City, UT. Unpublished work, 1999.
12. "Evaluation Guidelines for Air Sampling Methods Utilizing Chromatographic Analyses", OSHA Salt Lake Technical Center, Salt Lake City, UT. Unpublished work, 1999.
|