PART 493--LABORATORY REQUIREMENTS

                      Subpart A--General Provisions

Sec.
493.1  Basis and scope.
493.2  Definitions.
493.3  Applicability.
493.5  Categories of tests by complexity.
493.15  Laboratories performing waived tests.
493.17  Test categorization.
493.19  Provider-performed microscopy (PPM) procedures.
493.20  Laboratories performing tests of moderate complexity.
493.25  Laboratories performing tests of high complexity.

                    Subpart B--Certificate of Waiver

493.35  Application for a certificate of waiver.
493.37  Requirements for a certificate of waiver.
493.39  Notification requirements for laboratories issued a certificate 
          of waiver.

Subpart C--Registration Certificate, Certificate for Provider-performed 
          Microscopy Procedures, and Certificate of Compliance

493.43  Application for registration certificate, certificate for 
          provider-performed microscopy (PPM) procedures, and 
          certificate of compliance.
493.45  Requirements for a registration certificate.
493.47  Requirements for a certificate for provider-performed microscopy 
          (PPM) procedures.
493.49  Requirements for a certificate of compliance.
493.51  Notification requirements for laboratories issued a certificate 
          of compliance.
493.53  Notification requirements for laboratories issued a certificate 
          for provider-performed microscopy (PPM) procedures.

                 Subpart D--Certificate of Accreditation

493.55  Application for registration certificate and certificate of 
          accreditation.
493.57  Requirements for a registration certificate.
493.61  Requirements for a certificate of accreditation.
493.63  Notification requirements for laboratories issued a certificate 
          of accreditation.

     Subpart E--Accreditation by a Private, Nonprofit Accreditation 
  Organization or Exemption Under an Approved State Laboratory Program

493.551  General requirements for laboratories.
493.553  Approval process (application and reapplication) for 
          accreditation organizations and State licensure programs.
493.555  Federal review of laboratory requirements.
493.557  Additional submission requirements.
493.559  Publication of approval of deeming authority or CLIA exemption.
493.561  Denial of application or reapplication.
493.563  Validation inspections--Basis and focus.

[[Page 820]]

493.565  Selection for validation inspection--laboratory 
          responsibilities.
493.567  Refusal to cooperate with validation inspection.
493.569  Consequences of a finding of noncompliance as a result of a 
          validation inspection.
493.571  Disclosure of accreditation, State and HCFA validation 
          inspection results.
493.573  Continuing Federal oversight of private nonprofit accreditation 
          organizations and approved State licensure programs.
493.575  Removal of deeming authority or CLIA exemption and final 
          determination review.

                    Subpart F--General Administration

493.602  Scope of subpart.
493.606  Applicability of subpart.
493.638  Certificate fees.
493.639  Fee for revised certificate.
493.643  Fee for determination of program compliance.
493.645  Additional fee(s) applicable to approved State laboratory 
          programs and laboratories issued a certificate of 
          accreditation, certificate of waiver, or certificate for PPM 
          procedures.
493.646  Payment of fees.
493.649  Methodology for determining fee amount.

Subpart G [Reserved]

    Subpart H--Participation in Proficiency Testing for Laboratories 
  Performing Tests of Moderate Complexity (Including the Subcategory), 
           High Complexity, or Any Combination of These Tests

493.801  Condition: Enrollment and testing of samples.
493.803  Condition: Successful participation.
493.807  Condition: Reinstatement of laboratories performing tests of 
          moderate complexity (including the subcategory), high 
          complexity, or any combination of these tests, after failure 
          to participate successfully.

   Proficiency Testing by Specialty and Subspecialty for Laboratories 
  Performing Tests of Moderate Complexity (Including the Subcategory), 
           High Complexity, or Any Combination of These Tests

493.821  Condition: Microbiology.
493.823  Standard; Bacteriology.
493.825  Standard; Mycobacteriology.
493.827  Standard; Mycology.
493.829  Standard; Parasitology.
493.831  Standard; Virology.
493.833  Condition: Diagnostic immunology.
493.835  Standard; Syphilis serology.
493.837  Standard; General immunology.
493.839  Condition: Chemistry.
493.841  Standard; Routine chemistry.
493.843  Standard; Endocrinology.
493.845  Standard; Toxicology.
493.849  Condition: Hematology.
493.851  Standard; Hematology.
493.853  Condition: Pathology.
493.855  Standard; Cytology: gynecologic examinations.
493.857  Condition: Immunohematology.
493.859  Standard; ABO group and D (Rho) typing.
493.861  Standard; Unexpected antibody detection.
493.863  Standard; Compatibility testing.
493.865  Standard; Antibody identification.

Subpart I--Proficiency Testing Programs for Tests of Moderate Complexity 
  (Including the Subcategory), High Complexity, or Any Combination of 
                               These Tests

493.901  Approval of proficiency testing programs.
493.903  Administrative responsibilities.
493.905  Nonapproved proficiency testing programs.

       Proficiency Testing Programs by Specialty and Subspecialty

493.909  Microbiology.
493.911  Bacteriology.
493.913  Mycobacteriology.
493.915  Mycology.
493.917  Parasitology.
493.919  Virology.
493.921  Diagnostic immunology.
493.923  Syphilis serology.
493.927  General immunology.
493.929  Chemistry.
493.931  Routine chemistry.
493.933  Endocrinology.
493.937  Toxicology.
493.941  Hematology (including routine hematology and coagulation).
493.945  Cytology; gynecologic examinations.
493.959  Immunohematology.

 Subpart J--Patient Test Management for Moderate Complexity (Including 
  the Subcategory), High Complexity, or Any Combination of These Tests

493.1101  Condition: Patient test management; moderate complexity 
          (including the subcategory), or high complexity testing, or 
          any combination of these tests.
493.1103  Standard; Procedures for specimen submission and handling.
493.1105  Standard; Test requisition.
493.1107  Standard; Test records.
493.1109  Standard; Test report.
493.1111  Standard; Referral of specimens.

[[Page 821]]

 Subpart K--Quality Control for Tests of Moderate Complexity (Including 
  the Subcategory), High Complexity, or Any Combination of These Tests

493.1201  Condition: General quality control; moderate complexity 
          (including the subcategory) or high complexity testing, or any 
          combination of these tests.
493.1202  Standard; Moderate or high complexity testing, or both: 
          Effective from September 1, 1992 to December 31, 2000.
493.1203  Standard; Moderate or high complexity testing, or both: 
          Effective beginning December 31, 2000.
493.1204  Standard; Facilities.
493.1205  Standard; Test methods, equipment, instrumentation, reagents, 
          materials, and supplies.
493.1211  Standard; Procedure manual.
493.1213  Standard; Establishment and verification of method performance 
          specifications.
493.1215  Standard; Equipment maintenance and function checks.
493.1217  Standard; Calibration and calibration verification procedures.
493.1218  Standard; Control procedures.
493.1219  Standard; Remedial actions.
493.1221  Standard; Quality control records.
493.1223  Condition: Quality control--specialties and subspecialties for 
          tests of moderate or high complexity, or both.
493.1225  Condition: Microbiology.
493.1227  Condition: Bacteriology.
493.1229  Condition: Mycobacteriology.
493.1231  Condition: Mycology.
493.1233  Condition: Parasitology.
493.1235  Condition: Virology.
493.1237  Condition: Diagnostic immunology.
493.1239  Condition: Syphilis serology.
493.1241  Condition: General immunology.
493.1243  Condition: Chemistry.
493.1245  Condition: Routine chemistry.
493.1247  Condition: Endocrinology.
493.1249  Condition: Toxicology.
493.1251  Condition: Urinalysis.
493.1253  Condition: Hematology.
493.1255  Condition: Pathology.
493.1257  Condition: Cytology.
493.1259  Condition: Histopathology.
493.1261  Condition: Oral pathology.
493.1263  Condition: Radiobioassay.
493.1265  Condition: Histocompatibility.
493.1267  Condition: Clinical cytogenetics.
493.1269  Condition: Immunohematology.
493.1271  Condition: Transfusion services and bloodbanking.
493.1273  Standard; Immunohematological collection, processing, dating 
          periods, labeling and distribution of blood and blood 
          products.
493.1275  Standard; Blood and blood products storage facilities.
493.1277  Standard; Arrangement for services.
493.1279  Standard; Provision of testing.
493.1283  Standard; Retention of samples of transfused blood.
493.1285  Standard; Investigation of transfusion reactions.

Subpart L [Reserved]

Subpart M--Personnel for Moderate Complexity (Including the Subcategory) 
                       and High Complexity Testing

493.1351  General.

 Laboratories Performing Provider-Performed Microscopy (PPM) Procedures

493.1353  Scope.
493.1355  Condition: Laboratories performing PPM procedures; laboratory 
          director.
493.1357  Standard; laboratory director qualifications.
493.1359  Standard; PPM laboratory director responsibilities.
493.1361  Condition: Laboratories performing PPM procedures; testing 
          personnel.
493.1363  Standard; PPM testing personnel qualifications.
493.1365  Standard; PPM testing personnel responsibilities.

           Laboratories Performing Moderate Complexity Testing

493.1403  Condition: Laboratories performing moderate complexity 
          testing; laboratory director.
493.1405  Standard; Laboratory director qualifications.
493.1406  Standard; Laboratory director qualifications on or before 
          February 28, 1992.
493.1407  Standard; Laboratory director responsibilities.
493.1409  Condition: Laboratories performing moderate complexity 
          testing; technical consultant.
493.1411  Standard; Technical consultant qualifications.
493.1413  Standard; Technical consultant responsibilities.
493.1415  Condition: Laboratories performing moderate complexity 
          testing; clinical consultant.
493.1417  Standard; Clinical consultant qualifications.
493.1419  Standard; Clinical consultant responsibilities.
493.1421  Condition: Laboratories performing moderate complexity 
          testing; testing personnel.
493.1423  Standard; Testing personnel qualifications.
493.1425  Standard; Testing personnel responsibilities.

[[Page 822]]

             Laboratories Performing High Complexity Testing

493.1441  Condition: Laboratories performing high complexity testing; 
          laboratory director.
493.1443  Standard; Laboratory director qualifications.
493.1445  Standard; Laboratory director responsibilities.
493.1447  Condition: Laboratories performing high complexity testing; 
          technical supervisor.
493.1449  Standard; Technical supervisor qualifications.
493.1451  Standard; Technical supervisor responsibilities.
493.1453  Condition: Laboratories performing high complexity testing; 
          clinical consultant.
493.1455  Standard; Clinical consultant qualifications.
493.1457  Standard; Clinical consultant responsibilities.
493.1459  Condition: Laboratories performing high complexity testing; 
          general supervisor.
493.1461  Standard; General supervisor qualifications.
493.1462  General supervisor qualifications on or before February 28, 
          1992.
493.1463  Standard; General supervisor responsibilities.
493.1467  Condition: Laboratories performing high complexity testing; 
          cytology general supervisor.
493.1469  Standard; Cytology general supervisor qualifications.
493.1471  Standard; Cytology general supervisor responsibilities.
493.1481  Condition: Laboratories performing high complexity testing; 
          cytotechnologist.
493.1483  Standard; Cytotechnologist qualifications.
493.1485  Standard; Cytotechnologist responsibilities.
493.1487  Condition: Laboratories performing high complexity testing; 
          testing personnel.
493.1489  Standard; Testing personnel qualifications.
493.1491  Technologist qualifications on or before February 28, 1992.
493.1495  Standard; Testing personnel responsibilities.

Subparts N-O [Reserved]

  Subpart P--Quality Assurance for Moderate Complexity (Including the 
  Subcategory) or High Complexity Testing, or Any Combination of These 
                                  Tests

493.1701  Condition: Quality assurance; moderate complexity (including 
          the subcategory) or high complexity testing, or any 
          combination of these tests.
493.1703  Standard; Patient test management assessment.
493.1705  Standard; Quality control assessment.
493.1707  Standard; Proficiency testing assessment.
493.1709  Standard; Comparison of test results.
493.1711  Standard; Relationship of patient information to patient test 
          results.
493.1713  Standard; Personnel assessment.
493.1715  Standard; Communications.
493.1717  Standard; Complaint investigations.
493.1719  Standard; Quality assurance review with staff.
493.1721  Standard; Quality assurance records.

                          Subpart Q--Inspection

493.1771  Condition: Inspection requirements applicable to all CLIA-
          certified and CLIA-exempt laboratories.
493.1773  Standard: Basic inspection requirements for all laboratories 
          issued a CLIA certificate and CLIA-exempt laboratories.
493.1775  Standard: Inspection of laboratories issued a certificate of 
          waiver or a certificate for provider-performed microscopy 
          procedures.
493.1777  Standard: Inspection of laboratories that have requested or 
          have been issued a certificate of compliance.
493.1780  Standard: Inspection of CLIA-exempt laboratories or 
          laboratories requesting or issued a certificate of 
          accreditation.

                    Subpart R--Enforcement Procedures

493.1800  Basis and scope.
493.1804  General considerations.
493.1806  Available sanctions: All laboratories.
493.1807  Additional sanctions: Laboratories that participate in 
          Medicare.
493.1808  Adverse action on any type of CLIA certificate: Effect on 
          Medicare approval.
493.1809  Limitation on Medicaid payment.
493.1810  Imposition and lifting of alternative sanctions.
493.1812  Action when deficiencies pose immediate jeopardy.
493.1814  Action when deficiencies are at the condition level but do not 
          pose immediate jeopardy.
493.1816  Action when deficiencies are not at the condition level.
493.1820  Ensuring timely correction of deficiencies.
493.1826  Suspension of part of Medicare payments.
493.1828  Suspension of all Medicare payments.

[[Page 823]]

493.1832  Directed plan of correction and directed portion of a plan of 
          correction.
493.1834  Civil money penalty.
493.1836  State onsite monitoring.
493.1838  Training and technical assistance for unsuccessful 
          participation in proficiency testing.
493.1840  Suspension, limitation, or revocation of any type of CLIA 
          certificate.
493.1842  Cancellation of Medicare approval.
493.1844  Appeals procedures.
493.1846  Civil action.
493.1850  Laboratory registry.

Subpart S [Reserved]

                        Subpart T--Consultations

493.2001  Establishment and function of the Clinical Laboratory 
          Improvement Advisory Committee.

    Authority: Sec. 353 of the Public Health Service Act, secs. 1102, 
1861(e), the sentence following sections 1861(s)(11) through 1861(s)(16) 
of the Social Security Act (42 U.S.C. 263a, 1302, 1395x(e), the sentence 
following 1395x(s)(11) through 1395x(s)(16)).

    Source: 55 FR 9576, Mar. 14, 1990, unless otherwise noted.

                      Subpart A--General Provisions

    Source: 57 FR 7139, Feb. 28, 1992, unless otherwise noted.

Sec. 493.1  Basis and scope.

    This part sets forth the conditions that all laboratories must meet 
to be certified to perform testing on human specimens under the Clinical 
Laboratory Improvement Amendments of 1988 (CLIA). It implements sections 
1861 (e) and (j), the sentence following section 1861(s)(13), and 
1902(a)(9) of the Social Security Act, and section 353 of the Public 
Health Service Act. This part applies to all laboratories as defined 
under ``laboratory'' in Sec. 493.2 of this part. This part also applies 
to laboratories seeking payment under the Medicare and Medicaid 
programs. The requirements are the same for Medicare approval as for 
CLIA certification.

Sec. 493.2  Definitions.

    As used in this part, unless the context indicates otherwise--
    Accredited institution means a school or program which--
    (a) Admits as regular student only persons having a certificate of 
graduation from a school providing secondary education, or the 
recognized equivalent of such certificate;
    (b) Is legally authorized within the State to provide a program of 
education beyond secondary education;
    (c) Provides an educational program for which it awards a bachelor's 
degree or provides not less than a 2-year program which is acceptable 
toward such a degree, or provides an educational program for which it 
awards a master's or doctoral degree;
    (d) Is accredited by a nationally recognized accrediting agency or 
association.
    This definition includes any foreign institution of higher education 
that HHS or its designee determines meets substantially equivalent 
requirements.
    Accredited laboratory means a laboratory that has voluntarily 
applied for and been accredited by a private, nonprofit accreditation 
organization approved by HCFA in accordance with this part;
    Adverse action means the imposition of a principal or alternative 
sanction by HCFA.
    ALJ stands for Administrative Law Judge.
    Alternative sanctions means sanctions that may be imposed in lieu of 
or in addition to principal sanctions. The term is synonymous with 
``intermediate sanctions'' as used in section 1846 of the Act.
    Analyte means a substance or constituent for which the laboratory 
conducts testing.
    Approved accreditation organization for laboratories means a 
private, nonprofit accreditation organization that has formally applied 
for and received HCFA's approval based on the organization's compliance 
with this part.
    Approved State laboratory program means a licensure or other 
regulatory program for laboratories in a State, the requirements of 
which are imposed under State law, and the State laboratory program has 
received HCFA approval based on the State's compliance with this part.
    Authorized person means an individual authorized under State law to 
order tests or receive test results, or both.

[[Page 824]]

    Challenge means, for quantitative tests, an assessment of the amount 
of substance or analyte present or measured in a sample. For qualitative 
tests, a challenge means the determination of the presence or the 
absence of an analyte, organism, or substance in a sample.
    CLIA means the Clinical Laboratory Improvement Amendments of 1988.
    CLIA certificate means any of the following types of certificates 
issued by HCFA or its agent:
    (1) Certificate of compliance means a certificate issued to a 
laboratory after an inspection that finds the laboratory to be in 
compliance with all applicable condition level requirements, or reissued 
before the expiration date, pending an appeal, in accordance with 
Sec. 493.49, when an inspection has found the laboratory to be out of 
compliance with one or more condition level requirements.
    (2) Certificate for provider-performed microscopy (PPM) procedures 
means a certificate issued or reissued before the expiration date, 
pending an appeal, in accordance with Sec. 493.47, to a laboratory in 
which a physician, midlevel practitioner or dentist performs no tests 
other than PPM procedures and, if desired, waived tests listed in 
Sec. 493.15(c).
    (3) Certificate of accreditation means a certificate issued on the 
basis of the laboratory's accreditation by an accreditation organization 
approved by HCFA (indicating that the laboratory is deemed to meet 
applicable CLIA requirements) or reissued before the expiration date, 
pending an appeal, in accordance with Sec. 493.61, when a validation or 
complaint survey has found the laboratory to be noncompliant with one or 
more CLIA conditions.
    (4) Certificate of registration or registration certificate means a 
certificate issued or reissued before the expiration date, pending an 
appeal, in accordance with Sec. 493.45, that enables the entity to 
conduct moderate or high complexity laboratory testing or both until the 
entity is determined to be in compliance through a survey by HCFA or its 
agent; or in accordance with Sec. 493.57 to an entity that is accredited 
by an approved accreditation organization.
    (5) Certificate of waiver means a certificate issued or reissued 
before the expiration date, pending an appeal, in accordance with 
Sec. 493.37, to a laboratory to perform only the waived tests listed at 
Sec. 493.15(c).
    CLIA-exempt laboratory means a laboratory that has been licensed or 
approved by a State where HCFA has determined that the State has enacted 
laws relating to laboratory requirements that are equal to or more 
stringent than CLIA requirements and the State licensure program has 
been approved by HCFA in accordance with subpart E of this part.
    Condition level deficiency means noncompliance with one or more 
condition level requirements.
    Condition level requirements means any of the requirements 
identified as ``conditions'' in subparts G through Q of this part.
    Credible allegation of compliance means a statement or documentation 
that--
    (1) Is made by a representative of a laboratory that has a history 
of having maintained a commitment to compliance and of taking corrective 
action when required;
    (2) Is realistic in terms of its being possible to accomplish the 
required corrective action between the date of the exit conference and 
the date of the allegation; and
    (3) Indicates that the problem has been resolved.
    Dentist means a doctor of dental medicine or doctor of dental 
surgery licensed by the State to practice dentistry within the State in 
which the laboratory is located.
    Equivalency means that an accreditation organization's or a State 
laboratory program's requirements, taken as a whole, are equal to or 
more stringent than the CLIA requirements established by HCFA, taken as 
whole. It is acceptable for an accreditation organization's or State 
laboratory program's requirements to be organized differently or 
otherwise vary from the CLIA requirements, as long as (1) all of the 
requirements taken as a whole would provide at least the same protection 
as the CLIA requirements taken

[[Page 825]]

as a whole; and (2) a finding of noncompliance with respect to CLIA 
requirements taken as a whole would be matched by a finding of 
noncompliance with the accreditation or State requirements taken as a 
whole.
    HCFA agent means an entity with which HCFA arranges to inspect 
laboratories and assess laboratory activities against CLIA requirements 
and may be a State survey agency, a private, nonprofit organization 
other than an approved accreditation organization, a component of HHS, 
or any other governmental component HCFA approves for this purpose. In 
those instances where all of the laboratories in a State are exempt from 
CLIA requirements, based on the approval of a State's exemption request, 
the State survey agency is not the HCFA agent.
    HHS means the Department of Health and Human Services, or its 
designee.
    Immediate jeopardy means a situation in which immediate corrective 
action is necessary because the laboratory's noncompliance with one or 
more condition level requirements has already caused, is causing, or is 
likely to cause, at any time, serious injury or harm, or death, to 
individuals served by the laboratory or to the health or safety of the 
general public. This term is synonymous with imminent and serious risk 
to human health and significant hazard to the public health.
    Intentional violation means knowing and willful noncompliance with 
any CLIA condition.
    Kit means all components of a test that are packaged together.
    Laboratory means a facility for the biological, microbiological, 
serological, chemical, immunohematological, hematological, biophysical, 
cytological, pathological, or other examination of materials derived 
from the human body for the purpose of providing information for the 
diagnosis, prevention, or treatment of any disease or impairment of, or 
the assessment of the health of, human beings. These examinations also 
include procedures to determine, measure, or otherwise describe the 
presence or absence of various substances or organisms in the body. 
Facilities only collecting or preparing specimens (or both) or only 
serving as a mailing service and not performing testing are not 
considered laboratories.
    Midlevel practitioner means a nurse midwife, nurse practitioner, or 
physician assistant, licensed by the State within which the individual 
practices, if such licensing is required in the State in which the 
laboratory is located.
    Operator means the individual or group of individuals who oversee 
all facets of the operation of a laboratory and who bear primary 
responsibility for the safety and reliability of the results of all 
specimen testing performed in that laboratory. The term includes--
    (1) A director of the laboratory if he or she meets the stated 
criteria; and
    (2) The members of the board of directors and the officers of a 
laboratory that is a small corporation under subchapter S of the 
Internal Revenue Code.
    Owner means any person who owns any interest in a laboratory except 
for an interest in a laboratory whose stock and/or securities are 
publicly traded. (That is e.g., the purchase of shares of stock or 
securities on the New York Stock Exchange in a corporation owning a 
laboratory would not make a person an owner for the purpose of this 
regulation.)
    Party means a laboratory affected by any of the enforcement 
procedures set forth in this subpart, by HCFA or the OIG, as 
appropriate.
    Performance characteristic means a property of a test that is used 
to describe its quality, e.g., accuracy, precision, analytical 
sensitivity, analytical specificity, reportable range, reference range, 
etc.
    Performance specification means a value or range of values for a 
performance characteristic, established or verified by the laboratory, 
that is used to describe the quality of patient test results.
    Physician means an individual with a doctor of medicine, doctor of 
osteopathy, or doctor of podiatric medicine degree who is licensed by 
the State to practice medicine, osteopathy, or podiatry within the State 
in which the laboratory is located.
    Principal sanction means the suspension, limitation, or revocation 
of any

[[Page 826]]

type of CLIA certificate or the cancellation of the laboratory's 
approval to receive Medicare payment for its services.
    Prospective laboratory means a laboratory that is operating under a 
registration certificate or is seeking any of the three other types of 
CLIA certificates.
    Rate of disparity means the percentage of sample validation 
inspections for a specific accreditation organization or State where 
HCFA, the State survey agency or other HCFA agent finds noncompliance 
with one or more condition level requirements but no comparable 
deficiencies were cited by the accreditation organization or the State, 
and it is reasonable to conclude that the deficiencies were present at 
the time of the most recent accreditation organization or State 
licensure inspection.

    Example: Assume the State survey agency, HCFA or other HCFA agent 
performs 200 sample validation inspections for laboratories accredited 
by a single accreditation organization or licensed in an exempt State 
during a validation review period and finds that 60 of the 200 
laboratories had one or more condition level requirements out of 
compliance. HCFA reviews the validation and accreditation organization's 
or State's inspections of the validated laboratories and determines that 
the State or accreditation organization found comparable deficiencies in 
22 of the 60 laboratories and it is reasonable to conclude that 
deficiencies were present in the remaining 38 laboratories at the time 
of the accreditation organization's or State's inspection. Thirty-eight 
divided by 200 equals a 19 percent rate of disparity.

    Referee laboratory means a laboratory currently in compliance with 
applicable CLIA requirements, that has had a record of satisfactory 
proficiency testing performance for all testing events for at least one 
year for a specific test, analyte, subspecialty, or specialty and has 
been designated by an HHS approved proficiency testing program as a 
referee laboratory for analyzing proficiency testing specimens for the 
purpose of determining the correct response for the specimens in a 
testing event for that specific test, analyte, subspecialty, or 
specialty.
    Reference range means the range of test values expected for a 
designated population of individuals, e.g., 95 percent of individuals 
that are presumed to be healthy (or normal).
    Sample in proficiency testing means the material contained in a 
vial, on a slide, or other unit that contains material to be tested by 
proficiency testing program participants. When possible, samples are of 
human origin.
    State includes, for purposes of this part, each of the 50 States, 
the District of Columbia, the Commonwealth of Puerto Rico, the Virgin 
Islands and a political subdivision of a State where the State, acting 
pursuant to State law, has expressly delegated powers to the political 
subdivision sufficient to authorize the political subdivision to act for 
the State in enforcing requirements equal to or more stringent than CLIA 
requirements.
    State licensure means the issuance of a license to, or the approval 
of, a laboratory by a State laboratory program as meeting standards for 
licensing or approval established under State law.
    State licensure program means a State laboratory licensure or 
approval program.
    State survey agency means the State health agency or other 
appropriate State or local agency that has an agreement under section 
1864 of the Social Security Act and is used by HCFA to perform surveys 
and inspections.
    Substantial allegation of noncompliance means a complaint from any 
of a variety of sources (including complaints submitted in person, by 
telephone, through written correspondence, or in newspaper or magazine 
articles) that, if substantiated, would have an impact on the health and 
safety of the general public or of individuals served by a laboratory 
and raises doubts as to a laboratory's compliance with any condition 
level requirement.
    Target value for quantitative tests means either the mean of all 
participant responses after removal of outliers (those responses greater 
than 3 standard deviations from the original mean) or the mean 
established by definitive or reference methods acceptable for use in the 
National Reference System for the Clinical Laboratory (NRSCL) by the 
National Committee for the Clinical Laboratory Standards (NCCLS). In 
instances where definitive or reference methods are not available

[[Page 827]]

or a specific method's results demonstrate bias that is not observed 
with actual patient specimens, as determined by a defensible scientific 
protocol, a comparative method or a method group (``peer'' group) may be 
used. If the method group is less than 10 participants, ``target value'' 
means the overall mean after outlier removal (as defined above) unless 
acceptable scientific reasons are available to indicate that such an 
evaluation is not appropriate.
    Unsatisfactory proficiency testing performance means failure to 
attain the minimum satisfactory score for an analyte, test, 
subspecialty, or specialty for a testing event.
    Unsuccessful participation in proficiency testing means any of the 
following:
    (1) Unsatisfactory performance for the same analyte in two 
consecutive or two out of three testing events.
    (2) Repeated unsatisfactory overall testing event scores for two 
consecutive or two out of three testing events for the same specialty or 
subspecialty.
    (3) An unsatisfactory testing event score for those subspecialties 
not graded by analyte (that is, bacteriology, mycobacteriology, 
virology, parasitology, mycology, blood compatibility, immunohematology, 
or syphilis serology) for the same subspecialty for two consecutive or 
two out of three testing events.
    (4) Failure of a laboratory performing gynecologic cytology to meet 
the standard at Sec. 493.855.
    Unsuccessful proficiency testing performance means a failure to 
attain the minimum satisfactory score for an analyte, test, 
subspecialty, or specialty for two consecutive or two of three 
consecutive testing events.
    Validation review period means the one year time period during which 
HCFA conducts validation inspections and evaluates the results of the 
most recent surveys performed by an accreditation organization or State 
laboratory program.

[57 FR 7139, Feb. 28, 1992, as amended at 57 FR 7236, Feb. 28, 1992; 57 
FR 34013, July 31, 1992; 57 FR 35761, Aug. 11, 1992; 58 FR 5220, Jan. 
19, 1993; 58 FR 48323, Sept. 15, 1993; 60 FR 20043, Apr. 24, 1995; 63 FR 
26732, May 14, 1998]

Sec. 493.3  Applicability.

    (a) Basic rule. Except as specified in paragraph (b) of this 
section, a laboratory will be cited as out of compliance with section 
353 of the Public Health Service Act unless it--
    (1) Has a current, unrevoked or unsuspended certificate of waiver, 
registration certificate, certificate of compliance, certificate for PPM 
procedures, or certificate of accreditation issued by HHS applicable to 
the category of examinations or procedures performed by the laboratory; 
or
    (2) Is CLIA-exempt.
    (b) Exception. These rules do not apply to components or functions 
of--
    (1) Any facility or component of a facility that only performs 
testing for forensic purposes;
    (2) Research laboratories that test human specimens but do not 
report patient specific results for the diagnosis, prevention or 
treatment of any disease or impairment of, or the assessment of the 
health of individual patients; or
    (3) Laboratories certified by the National Institutes on Drug Abuse 
(NIDA), in which drug testing is performed which meets NIDA guidelines 
and regulations. However, all other testing conducted by a NIDA-
certified laboratory is subject to this rule.
    (c) Federal laboratories. Laboratories under the jurisdiction of an 
agency of the Federal Government are subject to the rules of this part, 
except that the Secretary may modify the application of such 
requirements as appropriate.

[57 FR 7139, Feb. 28, 1992, as amended at 58 FR 5221, Jan. 19, 1993; 60 
FR 20043, Apr. 24, 1995]

Sec. 493.5  Categories of tests by complexity.

    (a) Laboratory tests are categorized as one of the following:
    (1) Waived tests.
    (2) Tests of moderate complexity, including the subcategory of PPM 
procedures.
    (3) Tests of high complexity.
    (b) A laboratory may perform only waived tests, only tests of 
moderate complexity, only PPM procedures, only tests of high complexity 
or any combination of these tests.

[[Page 828]]

    (c) Each laboratory must be either CLIA-exempt or possess one of the 
following CLIA certificates, as defined in Sec. 493.2:
    (1) Certificate of registration or registration certificate.
    (2) Certificate of waiver.
    (3) Certificate for PPM procedures.
    (4) Certificate of compliance.
    (5) Certificate of accreditation.

[60 FR 20043, Apr. 24, 1995]

Sec. 493.15  Laboratories performing waived tests.

    (a) Requirement. Tests for certificate of waiver must meet the 
descriptive criteria specified in paragraph (b) of this section.
    (b) Criteria. Test systems are simple laboratory examinations and 
procedures which--
    (1) Are cleared by FDA for home use;
    (2) Employ methodologies that are so simple and accurate as to 
render the likelihood of erroneous results negligible; or
    (3) Pose no reasonable risk of harm to the patient if the test is 
performed incorrectly.
    (c) Certificate of waiver tests. A laboratory may qualify for a 
certificate of waiver under section 353 of the PHS Act if it restricts 
the tests that it performs to one or more of the following tests or 
examinations (or additional tests added to this list as provided under 
paragraph (d) of this section) and no others:
    (1) Dipstick or Tablet Reagent Urinalysis (non-automated) for the 
following:
    (i) Bilirubin;
    (ii) Glucose;
    (iii) Hemoglobin;
    (iv) Ketone;
    (v) Leukocytes;
    (vi) Nitrite;
    (vii) pH;
    (viii) Protein;
    (ix) Specific gravity; and
    (x) Urobilinogen.
    (2) Fecal occult blood;
    (3) Ovulation tests--visual color comparison tests for human 
luteinizing hormone;
    (4) Urine pregnancy tests--visual color comparison tests;
    (5) Erythrocyte sedimentation rate--non-automated;
    (6) Hemoglobin--copper sulfate--non-automated;
    (7) Blood glucose by glucose monitoring devices cleared by the FDA 
specifically for home use;
    (8) Spun microhematocrit; and
    (9) Hemoglobin by single analyte instruments with self-contained or 
component features to perform specimen/reagent interaction, providing 
direct measurement and readout.
    (d) Revisions to criteria for test categorization and the list of 
waived tests. HHS will determine whether a laboratory test meets the 
criteria listed under paragraph (b) of this section for a waived test. 
Revisions to the list of waived tests approved by HHS will be published 
in the Federal Register in a notice with opportunity for comment.
    (e) Laboratories eligible for a certificate of waiver must--
    (1) Follow manufacturers' instructions for performing the test; and
    (2) Meet the requirements in subpart B, Certificate of Waiver, of 
this part.

[57 FR 7139, Feb. 28, 1992, as amended at 58 FR 5221, Jan. 19, 1993]

Sec. 493.17  Test categorization.

    (a) Categorization by criteria. Notices will be published in the 
Federal Register which list each specific test system, assay, and 
examination categorized by complexity. Using the seven criteria 
specified in this paragraph for categorizing tests of moderate or high 
complexity, each specific laboratory test system, assay, and examination 
will be graded for level of complexity by assigning scores of 1, 2, or 3 
within each criteria. The score of ``1'' indicates the lowest level of 
complexity, and the score of ``3'' indicates the highest level. These 
scores will be totaled. Test systems, assays or examinations receiving 
scores of 12 or less will be categorized as moderate complexity, while 
those receiving scores above 12 will be categorized as high complexity.

    Note: A score of ``2'' will be assigned to a criteria heading when 
the characteristics for a particular test are intermediate between the 
descriptions listed for scores of ``1'' and ``3.''

    (1) Knowledge.

[[Page 829]]

    (i) Score 1. (A) Minimal scientific and technical knowledge is 
required to perform the test; and
    (B) Knowledge required to perform the test may be obtained through 
on-the-job instruction.
    (ii) Score 3. Specialized scientific and technical knowledge is 
essential to perform preanalytic, analytic or postanalytic phases of the 
testing.
    (2) Training and experience.
    (i) Score 1. (A) Minimal training is required for preanalytic, 
analytic and postanalytic phases of the testing process; and
    (B) Limited experience is required to perform the test.
    (ii) Score 3. (A) Specialized training is essential to perform the 
preanalytic, analytic or postanalytic testing process; or
    (B) Substantial experience may be necessary for analytic test 
performance.
    (3) Reagents and materials preparation.
    (i) Score 1. (A) Reagents and materials are generally stable and 
reliable; and
    (B) Reagents and materials are prepackaged, or premeasured, or 
require no special handling, precautions or storage conditions.
    (ii) Score 3. (A) Reagents and materials may be labile and may 
require special handling to assure reliability; or
    (B) Reagents and materials preparation may include manual steps such 
as gravimetric or volumetric measurements.
    (4) Characteristics of operational steps. (i) Score 1. Operational 
steps are either automatically executed (such as pipetting, temperature 
monitoring, or timing of steps), or are easily controlled.
    (ii) Score 3. Operational steps in the testing process require close 
monitoring or control, and may require special specimen preparation, 
precise temperature control or timing of procedural steps, accurate 
pipetting, or extensive calculations.
    (5) Calibration, quality control, and proficiency testing materials.
    (i) Score 1. (A) Calibration materials are stable and readily 
available;
    (B) Quality control materials are stable and readily available; and
    (C) External proficiency testing materials, when available, are 
stable.
    (ii) Score 3. (A) Calibration materials, if available, may be 
labile;
    (B) Quality control materials may be labile, or not available; or
    (C) External proficiency testing materials, if available, may be 
labile.
    (6) Test system troubleshooting and equipment maintenance.
    (i) Score 1. (A) Test system troubleshooting is automatic or self-
correcting, or clearly described or requires minimal judgment; and
    (B) Equipment maintenance is provided by the manufacturer, is seldom 
needed, or can easily be performed.
    (ii) Score 3. (A) Troubleshooting is not automatic and requires 
decision-making and direct intervention to resolve most problems; or
    (B) Maintenance requires special knowledge, skills, and abilities.
    (7) Interpretation and judgment. (i) Score 1. (A) Minimal 
interpretation and judgment are required to perform preanalytic, 
analytic and postanalytic processes; and
    (B) Resolution of problems requires limited independent 
interpretation and judgment; and
    (ii) Score 3. (A) Extensive independent interpretation and judgment 
are required to perform the preanalytic, analytic or postanalytic 
processes; and
    (B) Resolution of problems requires extensive interpretation and 
judgment.
    (b) Revisions to the criteria for categorization. The Clinical 
Laboratory Improvement Advisory Committee, as defined in subpart T of 
this part, will conduct reviews upon request of HHS and recommend to HHS 
revisions to the criteria for categorization of tests.
    (c) Process for device/test categorization utilizing the scoring 
system under Sec. 493.17(a). (1)(i) For new commercial test systems, 
assays, or examinations, the manufacturer, as part of its 510(k) and PMA 
application to FDA, will submit supporting data for device/test 
categorization. FDA will determine the complexity category, notify the 
manufacturers directly, and will simultaneously inform both HCFA and CDC 
of the device/test category. FDA will consult with CDC concerning test 
categorization in the following three situations:

[[Page 830]]

    (A) When categorizing previously uncategorized new technology;
    (B) When FDA determines it to be necessary in cases involving a 
request for a change in categorization; and
    (C) If a manufacturer requests review of a categorization decision 
by FDA in accordance with 21 CFR 10.75.
    (ii) Test categorization will be effective as of the notification to 
the applicant.
    (2) For test systems, assays, or examinations not commercially 
available, a laboratory or professional group may submit a written 
request for categorization to PHS. These requests will be forwarded to 
CDC for evaluation; CDC will determine complexity category and notify 
the applicant, HCFA, and FDA of the categorization decision. In the case 
of request for a change of category or for previously uncategorized new 
technology, PHS will receive the request application and forward it to 
CDC for categorization.
    (3) A request for recategorization will be accepted for review if it 
is based on new information not previously submitted in a request for 
categorization or recategorization by the same applicant and will not be 
considered more frequently than once per year.
    (4) If a laboratory test system, assay or examination does not 
appear on the lists of tests in the Federal Register notices, it is 
considered to be a test of high complexity until PHS, upon request, 
reviews the matter and notifies the applicant of its decision. Test 
categorization is effective as of the notification to the applicant.
    (5) PHS will publish revisions periodically to the list of moderate 
and high complexity tests in the Federal Register in a notice with 
opportunity for comment.

[57 FR 7139, Feb. 28, 1992, as amended at 58 FR 5222, Jan. 19, 1993]

Sec. 493.19  Provider-performed microscopy (PPM) procedures.

    (a) Requirement. To be categorized as a PPM procedure, the procedure 
must meet the criteria specified in paragraph (b) of this section.
    (b) Criteria. Procedures must meet the following specifications:
    (1) The examination must be personally performed by one of the 
following practitioners:
    (i) A physician during the patient's visit on a specimen obtained 
from his or her own patient or from a patient of a group medical 
practice of which the physician is a member or an employee.
    (ii) A midlevel practitioner, under the supervision of a physician 
or in independent practice only if authorized by the State, during the 
patient's visit on a specimen obtained from his or her own patient or 
from a patient of a clinic, group medical practice, or other health care 
provider of which the midlevel practitioner is a member or an employee.
    (iii) A dentist during the patient's visit on a specimen obtained 
from his or her own patient or from a patient of a group dental practice 
of which the dentist is a member or an employee.
    (2) The procedure must be categorized as moderately complex.
    (3) The primary instrument for performing the test is the 
microscope, limited to bright-field or phase-contrast microscopy.
    (4) The specimen is labile or delay in performing the test could 
compromise the accuracy of the test result.
    (5) Control materials are not available to monitor the entire 
testing process.
    (6) Limited specimen handling or processing is required.
    (c) Provider-performed microscopy (PPM) examinations. A laboratory 
may qualify to perform tests under this section if it restricts PPM 
examinations to one or more of the following procedures (or additional 
procedures added to this list as provided under paragraph (d) of this 
section), waived tests and no others:
    (1) All direct wet mount preparations for the presence or absence of 
bacteria, fungi, parasites, and human cellular elements.
    (2) All potassium hydroxide (KOH) preparations.
    (3) Pinworm examinations.
    (4) Fern tests.
    (5) Post-coital direct, qualitative examinations of vaginal or 
cervical mucous.
    (6) Urine sediment examinations.
    (7) Nasal smears for granulocytes.

[[Page 831]]

    (8) Fecal leukocyte examinations.
    (9) Qualitative semen analysis (limited to the presence or absence 
of sperm and detection of motility).
    (d) Revisions to criteria and the list of PPM procedures.
    (1) The CLIAC conducts reviews upon HHS' request and recommends to 
HHS revisions to the criteria for categorization of procedures.
    (2) HHS determines whether a laboratory procedure meets the criteria 
listed under paragraph (b) of this section for a PPM procedure. 
Revisions to the list of PPM procedures proposed by HHS are published in 
the Federal Register as a notice with an opportunity for public comment.
    (e) Laboratory requirements. Laboratories eligible to perform PPM 
examinations must--
    (1) Meet the applicable requirements in subpart C or subpart D, and 
subparts F, H, J, K, M, and P of this part.
    (2) Be subject to inspection as specified under subpart Q of this 
part.

[60 FR 20044, Apr. 24, 1995]

Sec. 493.20  Laboratories performing tests of moderate complexity.

    (a) A laboratory may qualify for a certificate to perform tests of 
moderate complexity provided that it restricts its test performance to 
waived tests or examinations and one or more tests or examinations 
meeting criteria for tests of moderate complexity including the 
subcategory of PPM procedures.
    (b) A laboratory that performs tests or examinations of moderate 
complexity must meet the applicable requirements in subpart C or subpart 
D, and subparts F, H, J, K, M, P, and Q of this part. Under a 
registration certificate or certificate of compliance, laboratories also 
performing PPM procedures must meet the inspection requirements at 
Sec. 493.1777.
    (c) If the laboratory also performs waived tests, compliance with 
subparts H, J, K, M, and P of this part is not applicable to the waived 
tests. However, the laboratory must comply with the requirements in 
Secs. 493.15(e) and 493.1775.

[60 FR 20044, Apr. 24, 1995]

Sec. 493.25  Laboratories performing tests of high complexity.

    (a) A laboratory must obtain a certificate for tests of high 
complexity if it performs one or more tests that meet the criteria for 
tests of high complexity as specified in Sec. 493.17(a).
    (b) A laboratory performing one or more tests of high complexity 
must meet the applicable requirements of subpart C or subpart D, and 
subparts F, H, J, K, M, P, and Q of this part.
    (c) If the laboratory also performs tests of moderate complexity, 
the applicable requirements of subparts H, J, K, M, P, and Q of this 
part must be met. Under a registration certificate or certificate of 
compliance, PPM procedures must meet the inspection requirements at 
Sec. 493.1777.
    (d) If the laboratory also performs waived tests, the requirements 
of subparts H, J, K, M, and P are not applicable to the waived tests. 
However, the laboratory must comply with the requirements in 
Secs. 493.15(e) and 493.1775.

[57 FR 7139, Feb. 28, 1992, as amended at 60 FR 20044, Apr. 24, 1995]

                    Subpart B--Certificate of Waiver

    Source: 57 FR 7142, Feb. 28, 1992, unless otherwise noted.

Sec. 493.35  Application for a certificate of waiver.

    (a) Filing of application. Except as specified in paragraph (b) of 
this section, a laboratory performing only one or more waived tests 
listed in Sec. 493.15 must file a separate application for each 
laboratory location.
    (b) Exceptions. (1) Laboratories that are not at a fixed location, 
that is, laboratories that move from testing site to testing site, such 
as mobile units providing laboratory testing, health screening fairs, or 
other temporary testing locations may be covered under the certificate 
of the designated primary site or home base, using its address.
    (2) Not-for-profit or Federal, State, or local government 
laboratories that engage in limited (not more than a combination of 15 
moderately complex or waived tests per certificate) public

[[Page 832]]

health testing may file a single application.
    (3) Laboratories within a hospital that are located at contiguous 
buildings on the same campus and under common direction may file a 
single application or multiple applications for the laboratory sites 
within the same physical location or street address.
    (c) Application format and contents. The application must--
    (1) Be made to HHS or its designee on a form or forms prescribed by 
HHS;
    (2) Be signed by an owner, or by an authorized representative of the 
laboratory who attests that the laboratory will be operated in 
accordance with requirements established by the Secretary under section 
353 of the PHS Act; and
    (3) Describe the characteristics of the laboratory operation and the 
examinations and other test procedures performed by the laboratory 
including--
    (i) The name and the total number of test procedures and 
examinations performed annually (excluding tests the laboratory may run 
for quality control, quality assurance or proficiency testing purposes;
    (ii) The methodologies for each laboratory test procedure or 
examination performed, or both; and
    (iii) The qualifications (educational background, training, and 
experience) of the personnel directing and supervising the laboratory 
and performing the laboratory examinations and test procedures.
    (d) Access requirements. Laboratories that perform one or more 
waived tests listed in Sec. 493.15(c) and no other tests must meet the 
following conditions:
    (1) Make records available and submit reports to HHS as HHS may 
reasonably require to determine compliance with this section and 
Sec. 493.15(e);
    (2) Agree to permit announced and unannounced inspections by HHS in 
accordance with subpart Q of this part under the following 
circumstances:
    (i) When HHS has substantive reason to believe that the laboratory 
is being operated in a manner that constitutes an imminent and serious 
risk to human health.
    (ii) To evaluate complaints from the public.
    (iii) On a random basis to determine whether the laboratory is 
performing tests not listed in Sec. 493.15.
    (iv) To collect information regarding the appropriateness of waiver 
of tests listed in Sec. 493.15.
    (e) Denial of application. If HHS determines that the application 
for a certificate of waiver is to be denied, HHS will--
    (1) Provide the laboratory with a written statement of the grounds 
on which the denial is based and an opportunity for appeal, in 
accordance with the procedures set forth in subpart R of this part;
    (2) Notify a laboratory that has its application for a certificate 
of waiver denied that it cannot operate as a laboratory under the PHS 
Act unless the denial is overturned at the conclusion of the 
administrative appeals process provided by subpart R; and
    (3) Notify the laboratory that it is not eligible for payment under 
the Medicare and Medicaid programs.

[57 FR 7142, Feb. 28, 1992, as amended at 58 FR 5222, Jan. 19, 1993; 60 
FR 20044, Apr. 24, 1995]

Sec. 493.37  Requirements for a certificate of waiver.

    (a) HHS will issue a certificate of waiver to a laboratory only if 
the laboratory meets the requirements of Sec. 493.35.
    (b) Laboratories issued a certificate of waiver--
    (1) Are subject to the requirements of this subpart and 
Sec. 493.15(e) of subpart A of this part; and
    (2) Must permit announced or unannounced inspections by HHS in 
accordance with subpart Q of this part.
    (c) Laboratories must remit the certificate of waiver fee specified 
in subpart F of this part.
    (d) In accordance with subpart R of this part, HHS will suspend or 
revoke or limit a laboratory's certificate of waiver for failure to 
comply with the requirements of this subpart. In addition, failure to 
meet the requirements of this subpart will result in suspension or 
denial of payments under Medicare and Medicaid in accordance with 
subpart R of this part.
    (e)(1) A certificate of waiver issued under this subpart is valid 
for no more

[[Page 833]]

than 2 years. In the event of a non-compliance determination resulting 
in HHS action to revoke, suspend, or limit the laboratory's certificate 
of waiver, HHS will provide the laboratory with a statement of grounds 
on which the determination of non-compliance is based and offer an 
opportunity for appeal as provided in subpart R of this part.
    (2) If the laboratory requests a hearing within the time specified 
by HHS, it retains its certificate of waiver or reissued certificate of 
waiver until a decision is made by an administrative law judge, as 
specified in subpart R of this part, except when HHS finds that 
conditions at the laboratory pose an imminent and serious risk to human 
health.
    (3) For laboratories receiving payment from the Medicare or Medicaid 
program, such payments will be suspended on the effective date specified 
in the notice to the laboratory of a non-compliance determination even 
if there has been no appeals decision issued.
    (f) A laboratory seeking to renew its certificate of waiver must--
    (1) Complete the renewal application prescribed by HHS and return it 
to HHS not less than 9 months nor more than 1 year before the expiration 
of the certificate; and
    (2) Meet the requirements of Secs. 493.35 and 493.37.
    (g) A laboratory with a certificate of waiver that wishes to perform 
examinations or tests not listed in the waiver test category must meet 
the requirements set forth in subpart C or subpart D of this part, as 
applicable.

[57 FR 7142, Feb. 28, 1992, as amended at 58 FR 5222, Jan. 19, 1993; 60 
FR 20045, Apr. 24, 1995]

Sec. 493.39  Notification requirements for laboratories issued a 
          certificate of waiver.

    Laboratories performing one or more tests listed in Sec. 493.15 and 
no others must notify HHS or its designee--
    (a) Before performing and reporting results for any test or 
examination that is not specified under Sec. 493.15 for which the 
laboratory does not have the appropriate certificate as required in 
subpart C or subpart D of this part, as applicable; and
    (b) Within 30 days of any change(s) in--
    (1) Ownership;
    (2) Name;
    (3) Location; or
    (4) Director.

[57 FR 7142, Feb. 28, 1992, as amended at 60 FR 20045, Apr. 24, 1995]

Subpart C--Registration Certificate, Certificate for Provider-performed 
          Microscopy Procedures, and Certificate of Compliance

    Source: 57 FR 7143, Feb. 28, 1992, unless otherwise noted.

Sec. 493.43  Application for registration certificate, certificate for 
          provider-performed microscopy (PPM) procedures, and 
          certificate of compliance.

    (a) Filing of application. Except as specified in paragraph (b) of 
this section, all laboratories performing tests of moderate complexity 
(including the subcategory) or high complexity, or any combination of 
these tests, must file a separate application for each laboratory 
location.
    (b) Exceptions. (1) Laboratories that are not at a fixed location, 
that is, laboratories that move from testing site to testing site, such 
as mobile units providing laboratory testing, health screening fairs, or 
other temporary testing locations may be covered under the certificate 
of the designated primary site or home base, using its address.
    (2) Not-for-profit or Federal, State, or local government 
laboratories that engage in limited (not more than a combination of 15 
moderately complex or waived tests per certificate) public health 
testing may file a single application.
    (3) Laboratories within a hospital that are located at contiguous 
buildings on the same campus and under common direction may file a 
single application or multiple applications for the laboratory sites 
within the same physical location or street address.

[[Page 834]]

    (c) Application format and contents. The application must--(1) Be 
made to HHS or its designee on a form or forms prescribed by HHS;
    (2) Be signed by an owner, or by an authorized representative of the 
laboratory who attests that the laboratory will be operated in 
accordance with the requirements established by the Secretary under 
section 353 of the Public Health Service Act; and
    (3) Describe the characteristics of the laboratory operation and the 
examinations and other test procedures performed by the laboratory 
including--
    (i) The name and total number of test procedures and examinations 
performed annually (excluding waived tests or tests for quality control, 
quality assurance or proficiency testing purposes);
    (ii) The methodologies for each laboratory test procedure or 
examination performed, or both;
    (iii) The qualifications (educational background, training, and 
experience) of the personnel directing and supervising the laboratory 
and performing the examinations and test procedures.
    (d) Access and reporting requirements. All laboratories must make 
records available and submit reports to HHS as HHS may reasonably 
require to determine compliance with this section.

[57 FR 7143, Feb. 28, 1992, as amended at 58 FR 5222, Jan. 19, 1993; 58 
FR 39155, July 22, 1993; 60 FR 20045, Apr. 24, 1995]

Sec. 493.45  Requirements for a registration certificate.

    Laboratories performing only waived tests, PPM procedures, or any 
combination of these tests, are not required to obtain a registration 
certificate.
    (a) A registration certificate is required--(1) Initially for all 
laboratories performing test procedures of moderate complexity (other 
than the subcategory of PPM procedures) or high complexity, or both; and
    (2) For all laboratories that have been issued a certificate of 
waiver or certificate for PPM procedures that intend to perform tests of 
moderate or high complexity, or both, in addition to those tests listed 
in Sec. 493.15(c) or specified as PPM procedures.
    (b) HHS will issue a registration certificate if the laboratory--
    (1) Complies with the requirements of Sec. 493.43;
    (2) Agrees to notify HHS or its designee within 30 days of any 
changes in ownership, name, location, director or technical supervisor 
(laboratories performing high complexity testing only);
    (3) Agrees to treat proficiency testing samples in the same manner 
as it treats patient specimens; and
    (4) Remits the fee for the registration certificate, as specified in 
subpart F of this part.
    (c) Prior to the expiration of the registration certificate, a 
laboratory must--
    (1) Remit the certificate fee specified in subpart F of this part;
    (2) Be inspected by HHS as specified in subpart Q of this part; and
    (3) Demonstrate compliance with the applicable requirements of this 
subpart and subparts H, J, K, M, P, and Q of this part.
    (d) In accordance with subpart R of this part, HHS will initiate 
suspension or revocation of a laboratory's registration certificate and 
will deny the laboratory's application for a certificate of compliance 
for failure to comply with the requirements set forth in this subpart. 
HHS may also impose certain alternative sanctions. In addition, failure 
to meet the requirements of this subpart will result in suspension of 
payments under Medicare and Medicaid as specified in subpart R of this 
part.
    (e) A registration certificate is--
    (1) Valid for a period of no more than two years or until such time 
as an inspection to determine program compliance can be conducted, 
whichever is shorter; and
    (2) Not renewable; however, the registration certificate may be 
reissued if compliance has not been determined by HHS prior to the 
expiration date of the registration certificate.
    (f) In the event of a noncompliance determination resulting in an 
HHS denial of a laboratory's certificate of compliance application, HHS 
will provide the laboratory with a statement of grounds on which the 
noncompliance determination is based and offer an opportunity for appeal 
as provided in subpart R.

[[Page 835]]

    (g) If the laboratory requests a hearing within the time specified 
by HHS, it retains its registration certificate or reissued registration 
certificate until a decision is made by an administrative law judge as 
provided in subpart R of this part, except when HHS finds that 
conditions at the laboratory pose an imminent and serious risk to human 
health.
    (h) For laboratories receiving payment from the Medicare or Medicaid 
program, such payments will be suspended on the effective date specified 
in the notice to the laboratory of denial of the certificate application 
even if there has been no appeals decision issued.

[57 FR 7143, Feb. 28, 1992, as amended at 58 FR 5223, Jan. 19, 1993; 60 
FR 20045, Apr. 24, 1995]

Sec. 493.47  Requirements for a certificate for provider-performed 
          microscopy (PPM) procedures.

    (a) A certificate for PPM procedures is required--
    (1) Initially for all laboratories performing test procedures 
specified as PPM procedures; and
    (2) For all certificate of waiver laboratories that intend to 
perform only test procedures specified as PPM procedures in addition to 
those tests listed in Sec. 493.15(c).
    (b) HHS will issue a certificate for PPM procedures if the 
laboratory--
    (1) Complies with the requirements of Sec. 493.43; and
    (2) Remits the fee for the certificate, as specified in subpart F of 
this part.
    (c) Laboratories issued a certificate for PPM procedures are subject 
to--
    (1) The notification requirements of Sec. 493.53;
    (2) The applicable requirements of this subpart and subparts H, J, 
K, M, and P of this part; and
    (3) Inspection only under the circumstances specified under 
Sec. 493.1776, but are not routinely inspected to determine compliance 
with the requirements specified in paragraphs (c) (1) and (2) of this 
section.
    (d) In accordance with subpart R of this part, HHS will initiate 
suspension, limitation, or revocation of a laboratory's certificate for 
PPM procedures for failure to comply with the applicable requirements 
set forth in this subpart. HHS may also impose certain alternative 
sanctions. In addition, failure to meet the requirements of this subpart 
may result in suspension of all or part of payments under Medicare and 
Medicaid, as specified in subpart R of this part.
    (e) A certificate for PPM procedures is valid for a period of no 
more than 2 years.

[58 FR 5223, Jan. 19, 1993, as amended at 60 FR 20045, Apr. 24, 1995]

Sec. 493.49  Requirements for a certificate of compliance.

    A certificate of compliance may include any combination of tests 
categorized as high complexity or moderate complexity or listed in 
Sec. 493.15(c) as waived tests. Moderate complexity tests may include 
those specified as PPM procedures.
    (a) HHS will issue a certificate of compliance to a laboratory only 
if the laboratory--
    (1) Meets the requirements of Secs. 493.43 and 493.45;
    (2) Remits the certificate fee specified in subpart F of this part; 
and
    (3) Meets the applicable requirements of this subpart and subparts 
H, J, K, M, P, and Q of this part.
    (b) Laboratories issued a certificate of compliance--
    (1) Are subject to the notification requirements of Sec. 493.51; and
    (2) Must permit announced or unannounced inspections by HHS in 
accordance with subpart Q of this part--
    (i) To determine compliance with the applicable requirements of this 
part;
    (ii) To evaluate complaints;
    (iii) When HHS has substantive reason to believe that tests are 
being performed, or the laboratory is being operated in a manner that 
constitutes an imminent and serious risk to human health; and
    (iv) To collect information regarding the appropriateness of tests 
listed in Sec. 493.15 or tests categorized as moderate complexity 
(including the subcategory) or high complexity.
    (c) Failure to comply with the requirements of this subpart will 
result in--
    (1) Suspension, revocation or limitation of a laboratory's 
certificate of

[[Page 836]]

compliance in accordance with subpart R of this part; and
    (2) Suspension or denial of payments under Medicare and Medicaid in 
accordance with subpart R of this part.
    (d) A certificate of compliance issued under this subpart is valid 
for no more than 2 years.
    (e) In the event of a noncompliance determination resulting in an 
HHS action to revoke, suspend or limit the laboratory's certificate of 
compliance, HHS will--
    (1) Provide the laboratory with a statement of grounds on which the 
determination of noncompliance is based; and
    (2) Offer an opportunity for appeal as provided in subpart R of this 
part. If the laboratory requests a hearing within 60 days of the notice 
of sanction, it retains its certificate of compliance or reissued 
certificate of compliance until a decision is made by an administrative 
law judge (ALJ) as provided in subpart R of this part, except when HHS 
finds that conditions at the laboratory pose an imminent and serious 
risk to human health or when the criteria at Sec. 493.1840(a) (4) and 
(5) are met.
    (f) For laboratories receiving payment from the Medicare or Medicaid 
program, such payments will be suspended on the effective date specified 
in the notice to the laboratory of a noncompliance determination even if 
there has been no appeals decision issued.
    (g) A laboratory seeking to renew its certificate of compliance 
must--
    (1) Complete and return the renewal application to HHS 9 to 12 
months prior to the expiration of the certificate of compliance; and
    (2) Meet the requirements of Sec. 493.43 and paragraphs (a)(2) and 
(b)(2) of this section.
    (h) If HHS determines that the application for the renewal of a 
certificate of compliance must be denied or limited, HHS will notify the 
laboratory in writing of the--
    (1) Basis for denial of the application; and
    (2) Opportunity for appeal as provided in subpart R of this part.
    (i) If the laboratory requests a hearing within the time period 
specified by HHS, the laboratory retains its certificate of compliance 
or reissued certificate of compliance until a decision is made by an ALJ 
as provided in subpart R, except when HHS finds that conditions at the 
laboratory pose an imminent and serious risk to human health.
    (j) For laboratories receiving payment from the Medicare or Medicaid 
program, such payments will be suspended on the effective date specified 
in the notice to the laboratory of nonrenewal of the certificate of 
compliance even if there has been no appeals decision issued.

[60 FR 20045, Apr. 24, 1995]

Sec. 493.51  Notification requirements for laboratories issued a 
          certificate of compliance.

    Laboratories issued a certificate of compliance must meet the 
following conditions:
    (a) Notify HHS or its designee within 30 days of any change in--
    (1) Ownership;
    (2) Name;
    (3) Location;
    (4) Director; or
    (5) Technical supervisor (laboratories performing high complexity 
only).
    (b) Notify HHS no later than 6 months after performing any test or 
examination within a specialty or subspecialty area that is not included 
on the laboratory's certificate of compliance, so that compliance with 
requirements can be determined.
    (c) Notify HHS no later than 6 months after any deletions or changes 
in test methodologies for any test or examination included in a 
specialty or subspecialty, or both, for which the laboratory has been 
issued a certificate of compliance.

[57 FR 7143, Feb. 28, 1992, as amended at 60 FR 20046, Apr. 24, 1995]

Sec. 493.53  Notification requirements for laboratories issued a 
          certificate for provider-performed microscopy (PPM) 
          procedures.

    Laboratories issued a certificate for PPM procedures must notify HHS 
or its designee--
    (a) Before performing and reporting results for any test of moderate 
or high complexity, or both, in addition to tests specified as PPM 
procedures or any test or examination that is not

[[Page 837]]

specified under Sec. 493.15(c), for which it does not have a 
registration certificate as required in subpart C or subpart D, as 
applicable, of this part; and
    (b) Within 30 days of any change in--
    (1) Ownership;
    (2) Name;
    (3) Location; or
    (4) Director.

[58 FR 5224, Jan. 19, 1993, as amended at 60 FR 20046, Apr. 24, 1995]

                 Subpart D--Certificate of Accreditation

    Source: 57 FR 7144, Feb. 28, 1992, unless otherwise noted.

Sec. 493.55  Application for registration certificate and certificate of 
          accreditation.

    (a) Filing of application. A laboratory may be issued a certificate 
of accreditation in lieu of the applicable certificate specified in 
subpart B or subpart C of this part provided the laboratory--
    (1) Meets the standards of a private non-profit accreditation 
program approved by HHS in accordance with subpart E; and
    (2) Files a separate application for each location, except as 
specified in paragraph (b) of this section.
    (b) Exceptions. (1) Laboratories that are not at fixed locations, 
that is, laboratories that move from testing site to testing site, such 
as mobile units providing laboratory testing, health screening fairs, or 
other temporary testing locations may be covered under the certificate 
of the designated primary site or home base, using its address.
    (2) Not-for-profit or Federal, State, or local government 
laboratories that engage in limited (not more than a combination of 15 
moderately complex or waived tests per certificate) public health 
testing may file a single application.
    (3) Laboratories within a hospital that are located at contiguous 
buildings on the same campus and under common direction may file a 
single application or multiple applications for the laboratory sites 
within the same physical location or street address.
    (c) Application format and contents. The application must--(1) Be 
made to HHS on a form or forms prescribed by HHS;
    (2) Be signed by an owner or authorized representative of the 
laboratory who attests that the laboratory will be operated in 
accordance with the requirements established by the Secretary under 
section 353 of the Public Health Service Act; and
    (3) Describe the characteristics of the laboratory operation and the 
examinations and other test procedures performed by the laboratory 
including--
    (i) The name and total number of tests and examinations performed 
annually (excluding waived tests and tests for quality control, quality 
assurance or proficiency testing purposes);
    (ii) The methodologies for each laboratory test procedure or 
examination performed, or both; and
    (iii) The qualifications (educational background, training, and 
experience) of the personnel directing and supervising the laboratory 
and performing the laboratory examinations and test procedures.
    (d) Access and reporting requirements. All laboratories must make 
records available and submit reports to HHS as HHS may reasonably 
require to determine compliance with this section.

[57 FR 7144, Feb. 28, 1992, as amended at 58 FR 5224, Jan. 19, 1993; 58 
FR 39155, July 22, 1993; 60 FR 20046, Apr. 24, 1995]

Sec. 493.57  Requirements for a registration certificate.

    A registration certificate is required for all laboratories seeking 
a certificate of accreditation, unless the laboratory holds a valid 
certificate of compliance issued by HHS.
    (a) HHS will issue a registration certificate if the laboratory--
    (1) Complies with the requirements of Sec. 493.55;
    (2) Agrees to notify HHS within 30 days of any changes in ownership, 
name, location, director, or supervisor (laboratories performing high 
complexity testing only);
    (3) Agrees to treat proficiency testing samples in the same manner 
as it treats patient specimens; and

[[Page 838]]

    (4) Remits the fee for the registration certificate specified in 
subpart F of this part.
    (b)(1) The laboratory must provide HHS with proof of accreditation 
by an approved accreditation program--
    (i) Within 11 months of issuance of the registration certificate; or
    (ii) Prior to the expiration of the certificate of compliance.
    (2) If such proof of accreditation is not supplied within this 
timeframe, the laboratory must meet, or continue to meet, the 
requirements of Sec. 493.49.
    (c) In accordance with subpart R of this part, HHS will initiate 
suspension, revocation, or limitation of a laboratory's registration 
certificate and will deny the laboratory's application for a certificate 
of accreditation for failure to comply with the requirements set forth 
in this subpart. In addition, failure to meet the requirements of this 
subpart will result in suspension or denial of payments under Medicare 
and Medicaid as specified in subpart R of this part.
    (d) A registration certificate is valid for a period of no more than 
2 years. However, it may be reissued if the laboratory is subject to 
subpart C of this part, as specified in Sec. 493.57(b)(2) and compliance 
has not been determined by HHS before the expiration date of the 
registration certificate.
    (e) In the event that the laboratory does not meet the requirements 
of this subpart, HHS will--
    (1) Deny a laboratory's request for certificate of accreditation;
    (2) Notify the laboratory if it must meet the requirements for a 
certificate as defined in subpart C of this part;
    (3) Provide the laboratory with a statement of grounds on which the 
application denial is based;
    (4) Offer an opportunity for appeal on the application denial as 
provided in subpart R of this part. If the laboratory requests a hearing 
within the time specified by HHS, the laboratory will retain its 
registration certificate or reissued registration certificate until a 
decision is made by an administrative law judge as provided in subpart 
R, unless HHS finds that conditions at the laboratory pose an imminent 
and serious risk to human health; and
    (5) For those laboratories receiving payment from the Medicare or 
Medicaid program, such payments will be suspended on the effective date 
specified in the notice to the laboratory of denial of the request even 
if there has been no appeals decision issued.

[57 FR 7144, Feb. 28, 1992, as amended at 60 FR 20046, Apr. 24, 1995]

Sec. 493.61  Requirements for a certificate of accreditation.

    (a) HHS will issue a certificate of accreditation to a laboratory if 
the laboratory--
    (1) Meets the requirements of Sec. 493.57 or, if applicable, 
Sec. 493.49 of subpart C of this part; and
    (2) Remits the certificate of accreditation fee specified in subpart 
F of this part.
    (b) Laboratories issued a certificate of accreditation must--
    (1) Treat proficiency testing samples in the same manner as patient 
samples;
    (2) Meet the requirements of Sec. 493.63;
    (3) Comply with the requirements of the approved accreditation 
program;
    (4) Permit random sample validation and complaint inspections as 
required in subpart Q of this part;
    (5) Permit HHS to monitor the correction of any deficiencies found 
through the inspections specified in paragraph (b)(4) of this section;
    (6) Authorize the accreditation program to release to HHS the 
laboratory's inspection findings whenever HHS conducts random sample or 
complaint inspections; and
    (7) Authorize its accreditation program to submit to HHS the results 
of the laboratory's proficiency testing.
    (c) A laboratory failing to meet the requirements of this section--
    (1) Will no longer meet the requirements of this part by virtue of 
its accreditation in an approved accreditation program;
    (2) Will be subject to full determination of compliance by HHS;
    (3) May be subject to suspension, revocation or limitation of the 
laboratory's certificate of accreditation or certain alternative 
sanctions; and
    (4) May be subject to suspension of payments under Medicare and 
Medicaid as specified in subpart R.

[[Page 839]]

    (d) A certificate of accreditation issued under this subpart is 
valid for no more than 2 years. In the event of a non-compliance 
determination as a result of a random sample validation or complaint 
inspection, a laboratory will be subject to a full review by HHS in 
accordance with Sec. 488.11 of this chapter.
    (e) Failure to meet the applicable requirements of part 493, will 
result in an action by HHS to suspend, revoke or limit the certificate 
of accreditation. HHS will--
    (1) Provide the laboratory with a statement of grounds on which the 
determination of noncompliance is based;
    (2) Notify the laboratory if it is eligible to apply for a 
certificate as defined in subpart C of this part; and
    (3) Offer an opportunity for appeal as provided in subpart R of this 
part.
    (f) If the laboratory requests a hearing within the time frame 
specified by HHS--
    (1) It retains its certificate of accreditation or reissued 
certificate of accreditation until a decision is made by an 
administrative law judge as provided in subpart R of this part, unless 
HHS finds that conditions at the laboratory pose an imminent and serious 
risk to human health; and
    (2) For those laboratories receiving payments from the Medicare or 
Medicaid program, such payments will be suspended on the effective date 
specified in the notice to the laboratory even if there has been no 
appeals decision issued.
    (g) In the event the accreditation organization's approval is 
removed by HHS, the laboratory will be subject to the applicable 
requirements of subpart C of this part or Sec. 493.57.
    (h) A laboratory seeking to renew its certificate of accreditation 
must--
    (1) Complete and return the renewal application to HHS 9 to 12 
months prior to the expiration of the certificate of accreditation;
    (2) Meet the requirements of this subpart; and
    (3) Submit the certificate of accreditation fee specified in subpart 
F of this part.
    (i) If HHS determines that the renewal application for a certificate 
of accreditation is to be denied or limited, HHS will notify the 
laboratory in writing of--
    (1) The basis for denial of the application;
    (2) Whether the laboratory is eligible for a certificate as defined 
in subpart C of this part;
    (3) The opportunity for appeal on HHS's action to deny the renewal 
application for certificate of accreditation as provided in subpart R of 
this part. If the laboratory requests a hearing within the time frame 
specified by HHS, it retains its certificate of accreditation or 
reissued certificate of accreditation until a decision is made by an 
administrative law judge as provided in subpart R of this part, unless 
HHS finds that conditions at the laboratory pose an imminent and serious 
risk to human health; and
    (4) Suspension of payments under Medicare or Medicaid for those 
laboratories receiving payments under the Medicare or Medicaid programs.

[57 FR 7144, Feb. 28, 1992, as amended at 58 FR 5224, Jan. 19, 1993]

Sec. 493.63  Notification requirements for laboratories issued a 
          certificate of accreditation.

    Laboratories issued a certificate of accreditation must:
    (a) Notify HHS and the approved accreditation program within 30 days 
of any changes in--
    (1) Ownership;
    (2) Name;
    (3) Location; or
    (4) Director.
    (b) Notify the approved accreditation program no later than 6 months 
after performing any test or examination within a specialty or 
subspecialty area that is not included in the laboratory's 
accreditation, so that the accreditation organization can determine 
compliance and a new certificate of accreditation can be issued.
    (c) Notify the accreditation program no later than 6 months after of 
any deletions or changes in test methodologies for any test or 
examination included in a specialty or subspecialty, or both, for which 
the laboratory has been issued a certificate of accreditation.

[[Page 840]]

     Subpart E--Accreditation by a Private, Nonprofit Accreditation 
  Organization or Exemption Under an Approved State Laboratory Program

    Source: 63 FR 26732, May 14, 1998, unless otherwise noted.

Sec. 493.551  General requirements for laboratories.

    (a) Applicability. HCFA may deem a laboratory to meet all applicable 
CLIA program requirements through accreditation by a private nonprofit 
accreditation program (that is, grant deemed status), or may exempt from 
CLIA program requirements all State licensed or approved laboratories in 
a State that has a State licensure program established by law, if the 
following conditions are met:
    (1) The requirements of the accreditation organization or State 
licensure program are equal to, or more stringent than, the CLIA 
condition-level requirements specified in this part, and the laboratory 
would meet the condition-level requirements if it were inspected against 
these requirements.
    (2) The accreditation program or the State licensure program meets 
the requirements of this subpart and is approved by HCFA.
    (3) The laboratory authorizes the approved accreditation 
organization or State licensure program to release to HCFA all records 
and information required and permits inspections as outlined in this 
part.
    (b) Meeting CLIA requirements by accreditation. A laboratory seeking 
to meet CLIA requirements through accreditation by an approved 
accreditation organization must do the following:
    (1) Obtain a certificate of accreditation as required in subpart D 
of this part.
    (2) Pay the applicable fees as required in subpart F of this part.
    (3) Meet the proficiency testing (PT) requirements in subpart H of 
this part.
    (4) Authorize its PT organization to furnish to its accreditation 
organization the results of the laboratory's participation in an 
approved PT program for the purpose of monitoring the laboratory's PT 
and for making the annual PT results, along with explanatory information 
required to interpret the PT results, available on a reasonable basis, 
upon request of any person. A laboratory that refuses to authorize 
release of its PT results is no longer deemed to meet the condition-
level requirements and is subject to a full review by HCFA, in 
accordance with subpart Q of this part, and may be subject to the 
suspension or revocation of its certificate of accreditation under 
Sec. 493.1840.
    (5) Authorize its accreditation organization to release to HCFA or a 
HCFA agent the laboratory's PT results that constitute unsuccessful 
participation in an approved PT program, in accordance with the 
definition of ``unsuccessful participation in an approved PT program,'' 
as specified in Sec. 493.2 of this part, when the laboratory has failed 
to achieve successful participation in an approved PT program.
    (6) Authorize its accreditation organization to release to HCFA a 
notification of the actions taken by the organization as a result of the 
unsuccessful participation in a PT program within 30 days of the 
initiation of the action. Based on this notification, HCFA may take an 
adverse action against a laboratory that fails to participate 
successfully in an approved PT program.
    (c) Withdrawal of laboratory accreditation. After an accreditation 
organization has withdrawn or revoked its accreditation of a laboratory, 
the laboratory retains its certificate of accreditation for 45 days 
after the laboratory receives notice of the withdrawal or revocation of 
the accreditation, or the effective date of any action taken by HCFA, 
whichever is earlier.

Sec. 493.553  Approval process (application and reapplication) for 
          accreditation organizations and State licensure programs.

    (a) Information required. An accreditation organization that applies 
or reapplies to HCFA for deeming authority, or a State licensure program 
that applies or reapplies to HCFA for exemption from CLIA program 
requirements of licensed or approved laboratories within the State, must 
provide the following information:

[[Page 841]]

    (1) A detailed comparison of the individual accreditation, or 
licensure or approval requirements with the comparable condition-level 
requirements; that is, a crosswalk.
    (2) A detailed description of the inspection process, including the 
following:
    (i) Frequency of inspections.
    (ii) Copies of inspection forms.
    (iii) Instructions and guidelines.
    (iv) A description of the review and decision-making process of 
inspections.
    (v) A statement concerning whether inspections are announced or 
unannounced.
    (vi) A description of the steps taken to monitor the correction of 
deficiencies.
    (3) A description of the process for monitoring PT performance, 
including action to be taken in response to unsuccessful participation 
in a HCFA-approved PT program.
    (4) Procedures for responding to and for the investigation of 
complaints against its laboratories.
    (5) A list of all its current laboratories and the expiration date 
of their accreditation or licensure, as applicable.
    (6) Procedures for making PT information available (under State 
confidentiality and disclosure requirements, if applicable) including 
explanatory information required to interpret PT results, on a 
reasonable basis, upon request of any person.
    (b) HCFA action on an application or reapplication. If HCFA receives 
an application or reapplication from an accreditation organization, or 
State licensure program, HCFA takes the following actions:
    (1) HCFA determines if additional information is necessary to make a 
determination for approval or denial of the application and notifies the 
accreditation organization or State to afford it an opportunity to 
provide the additional information.
    (2) HCFA may visit the accreditation organization or State licensure 
program offices to review and verify the policies and procedures 
represented in its application and other information, including, but not 
limited to, review and examination of documents and interviews with 
staff.
    (3) HCFA notifies the accreditation organization or State licensure 
program indicating whether HCFA approves or denies the request for 
deeming authority or exemption, respectively, and the rationale for any 
denial.
    (c) Duration of approval. HCFA approval may not exceed 6 years.
    (d) Withdrawal of application. The accreditation organization or 
State licensure program may withdraw its application at any time before 
official notification, specified at Sec. 493.553(b)(3).

Sec. 493.555  Federal review of laboratory requirements.

    HCFA's review of an accreditation organization or State licensure 
program includes, but is not limited to, an evaluation of the following:
    (a) Whether the organization's or State's requirements for 
laboratories are equal to, or more stringent than, the condition-level 
requirements for laboratories.
    (b) The organization's or State's inspection process to determine 
the comparability of the full inspection and complaint inspection 
procedures and requirements to those of HCFA, including, but not limited 
to, inspection frequency and the ability to investigate and respond to 
complaints against its laboratories.
    (c) The organization's or State's agreement with HCFA that requires 
it to do the following:
    (1) Notify HCFA within 30 days of the action taken, of any 
laboratory that has--
    (i) Had its accreditation or licensure suspended, withdrawn, 
revoked, or limited;
    (ii) In any way been sanctioned; or
    (iii) Had any adverse action taken against it.
    (2) Notify HCFA within 10 days of any deficiency identified in an 
accredited or CLIA-exempt laboratory if the deficiency poses an 
immediate jeopardy to the laboratory's patients or a hazard to the 
general public.
    (3) Notify HCFA, within 30 days, of all newly--
    (i) Accredited laboratories (or laboratories whose areas of 
specialty/subspecialty testing have changed); or

[[Page 842]]

    (ii) Licensed laboratories, including the specialty/subspecialty 
areas of testing.
    (4) Notify each accredited or licensed laboratory within 10 days of 
HCFA's withdrawal of the organization's deeming authority or State's 
exemption.
    (5) Provide HCFA with inspection schedules, as requested, for 
validation purposes.

Sec. 493.557  Additional submission requirements.

    (a) Specific requirements for accreditation organizations. In 
addition to the information specified in Secs. 493.553 and 493.555, as 
part of the approval and review process, an accreditation organization 
applying or reapplying for deeming authority must also provide the 
following:
    (1) The specialty or subspecialty areas for which the organization 
is requesting deeming authority and its mechanism for monitoring 
compliance with all requirements equivalent to condition-level 
requirements within the scope of the specialty or subspecialty areas.
    (2) A description of the organization's data management and analysis 
system with respect to its inspection and accreditation decisions, 
including the kinds of routine reports and tables generated by the 
systems.
    (3) Detailed information concerning the inspection process, 
including, but not limited to the following:
    (i) The size and composition of individual accreditation inspection 
teams.
    (ii) Qualifications, education, and experience requirements that 
inspectors must meet.
    (iii) The content and frequency of training provided to inspection 
personnel, including the ability of the organization to provide 
continuing education and training to inspectors.
    (4) Procedures for removal or withdrawal of accreditation status for 
laboratories that fail to meet the organization's standards.
    (5) A proposed agreement between HCFA and the accreditation 
organization with respect to the notification requirements specified in 
Sec. 493.555(c).
    (6) Procedures for monitoring laboratories found to be out of 
compliance with its requirements. (These monitoring procedures must be 
used only when the accreditation organization identifies noncompliance. 
If noncompliance is identified through validation inspections, HCFA or a 
HCFA agent monitors corrections, as authorized at Sec. 493.565(d)).
    (7) A demonstration of its ability to provide HCFA with electronic 
data and reports in compatible code, including the crosswalk specified 
in Sec. 493.553(a)(1), that are necessary for effective validation and 
assessment of the organization's inspection process.
    (8) A demonstration of its ability to provide HCFA with electronic 
data, in compatible code, related to the adverse actions resulting from 
PT results constituting unsuccessful participation in PT programs as 
well as data related to the PT failures, within 30 days of the 
initiation of adverse action.
    (9) A demonstration of its ability to provide HCFA with electronic 
data, in compatible code, for all accredited laboratories, including the 
area of specialty or subspecialty.
    (10) Information defining the adequacy of numbers of staff and other 
resources.
    (11) Information defining the organization's ability to provide 
adequate funding for performing required inspections.
    (12) Any facility-specific data, upon request by HCFA, which 
includes, but is not limited to, the following:
    (i) PT results that constitute unsuccessful participation in a HCFA-
approved PT program.
    (ii) Notification of the adverse actions or corrective actions 
imposed by the accreditation organization as a result of unsuccessful PT 
participation.
    (13) An agreement to provide written notification to HCFA at least 
30 days in advance of the effective date of any proposed change in its 
requirements.
    (14) An agreement to disclose any laboratory's PT results upon 
reasonable request by any person.
    (b) Specific requirements for a State licensure program. In addition 
to requirements in Secs. 493.553 and 493.555, as part of the approval 
and review process, when a State licensure program applies or reapplies 
for exemption from the CLIA program, the State must do the following:

[[Page 843]]

    (1) Demonstrate to HCFA that it has enforcement authority and 
administrative structures and resources adequate to enforce its 
laboratory requirements.
    (2) Permit HCFA or a HCFA agent to inspect laboratories in the 
State.
    (3) Require laboratories in the State to submit to inspections by 
HCFA or a HCFA agent as a condition of licensure or approval.
    (4) Agree to pay the cost of the validation program administered in 
that State as specified in Secs. 493.645(a) and 493.646(b).
    (5) Take appropriate enforcement action against laboratories found 
by HCFA not to be in compliance with requirements equivalent to CLIA 
requirements.
    (6) Submit for Medicare and Medicaid payment purposes, a list of the 
specialties and subspecialties of tests performed by each laboratory.
    (7) Submit a written presentation that demonstrates the agency's 
ability to furnish HCFA with electronic data in compatible code, 
including the crosswalk specified in Sec. 493.553(a)(1).
    (8) Submit a statement acknowledging that the State will notify HCFA 
through electronic transmission of the following:
    (i) Any laboratory that has had its licensure or approval revoked or 
withdrawn or has been in any way sanctioned by the State within 30 days 
of taking the action.
    (ii) Changes in licensure or inspection requirements.
    (iii) Changes in specialties or subspecialties under which any 
licensed laboratory in the State performs testing.
    (9) Provide information for the review of the State's enforcement 
procedures for laboratories found to be out of compliance with the 
State's requirements.
    (10) Submit information that demonstrates the ability of the State 
to provide HCFA with the following:
    (i) Electronic data and reports in compatible code with the adverse 
or corrective actions resulting from PT results that constitute 
unsuccessful participation in PT programs.
    (ii) Other data that HCFA determines are necessary for validation 
and assessment of the State's inspection process requirements.
    (11) Agree to provide HCFA with written notification of any changes 
in its licensure/approval and inspection requirements.
    (12) Agree to disclose any laboratory's PT results in accordance 
with a State's confidentiality requirements.
    (13) Agree to take the appropriate enforcement action against 
laboratories found by HCFA not to be in compliance with requirements 
comparable to condition-level requirements and report these enforcement 
actions to HCFA.
    (14) If approved, reapply to HCFA every 2 years to renew its exempt 
status and to renew its agreement to pay the cost of the HCFA-
administered validation program in that State.

Sec. 493.559  Publication of approval of deeming authority or CLIA 
          exemption.

    (a) Notice of deeming authority or exemption. HCFA publishes a 
notice in the Federal Register when it grants deeming authority to an 
accreditation organization or exemption to a State licensure program.
    (b) Contents of notice. The notice includes the following:
    (1) The name of the accreditation organization or State licensure 
program.
    (2) For an accreditation organization:
    (i) The specific specialty or subspecialty areas for which it is 
granted deeming authority.
    (ii) A description of how the accreditation organization provides 
reasonable assurance to HCFA that a laboratory accredited by the 
organization meets CLIA requirements equivalent to those in this part 
and would meet CLIA requirements if the laboratory had not been granted 
deemed status, but had been inspected against condition-level 
requirements.
    (3) For a State licensure program, a description of how the 
laboratory requirements of the State are equal to, or more stringent 
than, those specified in this part.
    (4) The basis for granting deeming authority or exemption.
    (5) The term of approval, not to exceed 6 years.

[[Page 844]]

Sec. 493.561  Denial of application or reapplication.

    (a) Reconsideration of denial. (1) If HCFA denies a request for 
approval, an accreditation organization or State licensure program may 
request, within 60 days of the notification of denial, that HCFA 
reconsider its original application or application for renewal, in 
accordance with part 488, subpart D.
    (2) If the accreditation organization or State licensure program 
requests a reconsideration of HCFA's determination to deny its request 
for approval or reapproval, it may not submit a new application until 
HCFA issues a final reconsideration determination.
    (b) Resubmittal of a request for approval-- accreditation 
organization. An accreditation organization may resubmit a request for 
approval if a final reconsideration determination is not pending and the 
accreditation program meets the following conditions:
    (1) It has revised its accreditation program to address the 
rationale for denial of its previous request.
    (2) It demonstrates that it can provide reasonable assurance that 
its accredited facilities meet condition-level requirements.
    (3) It resubmits the application in its entirety.
    (c) Resubmittal of request for approval--State licensure program. 
The State licensure program may resubmit a request for approval if a 
final reconsideration determination is not pending and it has taken the 
necessary action to address the rationale for any previous denial.

Sec. 493.563  Validation inspections--Basis and focus.

    (a) Basis for validation inspection--(1) Laboratory with a 
certificate of accreditation. (i) HCFA or a HCFA agent may conduct an 
inspection of an accredited laboratory that has been issued a 
certificate of accreditation on a representative sample basis or in 
response to a substantial allegation of noncompliance.
    (ii) HCFA uses the results of these inspections to validate the 
accreditation organization's accreditation process.
    (2) Laboratory in a State with an approved State licensure program. 
(i) HCFA or a HCFA agent may conduct an inspection of any laboratory in 
a State with an approved State licensure program on a representative 
sample basis or in response to a substantial allegation of 
noncompliance.
    (ii) The results of these inspections are used to validate the 
appropriateness of the exemption of that State's licensed or approved 
laboratories from CLIA program requirements.
    (b) Validation inspection conducted on a representative sample 
basis. (1) If HCFA or a HCFA agent conducts a validation inspection on a 
representative sample basis, the inspection is comprehensive, addressing 
all condition-level requirements, or it may be focused on a specific 
condition-level requirement.
    (2) The number of laboratories sampled is sufficient to allow a 
reasonable estimate of the performance of the accreditation organization 
or State.
    (c) Validation inspection conducted in response to a substantial 
allegation of noncompliance. (1) If HCFA or a HCFA agent conducts a 
validation inspection in response to a substantial allegation of 
noncompliance, the inspection focuses on any condition-level requirement 
that HCFA determines to be related to the allegation.
    (2) If HCFA or a HCFA agent substantiates a deficiency and 
determines that the laboratory is out of compliance with any condition-
level requirement, HCFA or a HCFA agent conducts a full CLIA inspection.
    (d) Inspection of operations and offices. As part of the validation 
review process, HCFA may conduct an onsite inspection of the operations 
and offices to verify the following:
    (1) The accreditation organization's representations and to assess 
the accreditation organization's compliance with its own policies and 
procedures.
    (2) The State's representations and to assess the State's compliance 
with its own policies and procedures, including verification of State 
enforcement actions taken on the basis of validation inspections 
performed by HCFA or a HCFA agent.
    (e) Onsite inspection of an accreditation organization. An onsite 
inspection of an accreditation organization may include, but is not 
limited to, the following:
    (1) A review of documents.

[[Page 845]]

    (2) An audit of meetings concerning the accreditation process.
    (3) Evaluation of accreditation inspection results and the 
accreditation decision-making process.
    (4) Interviews with the accreditation organization's staff.
    (f) Onsite inspection of a State licensure program. An onsite 
inspection of a State licensure program office may include, but is not 
limited to, the following:
    (1) A review of documents.
    (2) An audit of meetings concerning the licensure or approval 
process.
    (3) Evaluation of State inspection results and the licensure or 
approval decision-making process.
    (4) Interviews with State employees.

Sec. 493.565  Selection for validation inspection--laboratory 
          responsibilities.

    A laboratory selected for a validation inspection must do the 
following:
    (a) Authorize its accreditation organization or State licensure 
program, as applicable, to release to HCFA or a HCFA agent, on a 
confidential basis, a copy of the laboratory's most recent full, and any 
subsequent partial inspection.
    (b) Authorize HCFA or a HCFA agent to conduct a validation 
inspection.
    (c) Provide HCFA or a HCFA agent with access to all facilities, 
equipment, materials, records, and information that HCFA or a HCFA agent 
determines have a bearing on whether the laboratory is being operated in 
accordance with the requirements of this part, and permit HCFA or a HCFA 
agent to copy material or require the laboratory to submit material.
    (d) If the laboratory possesses a valid certificate of 
accreditation, authorize HCFA or a HCFA agent to monitor the correction 
of any deficiencies found through the validation inspection.

Sec. 493.567  Refusal to cooperate with validation inspection.

    (a) Laboratory with a certificate of accreditation. (1) A laboratory 
with a certificate of accreditation that refuses to cooperate with a 
validation inspection by failing to comply with the requirements in 
Sec. 493.565--
    (i) Is subject to full review by HCFA or a HCFA agent, in accordance 
with this part; and
    (ii) May be subject to suspension, revocation, or limitation of its 
certificate of accreditation under this part.
    (2) A laboratory with a certificate of accreditation is again deemed 
to meet the condition-level requirements by virtue of its accreditation 
when the following conditions exist:
    (i) The laboratory withdraws any prior refusal to authorize its 
accreditation organization to release a copy of the laboratory's current 
accreditation inspection, PT results, or notification of any adverse 
actions resulting from PT failure.
    (ii) The laboratory withdraws any prior refusal to allow a 
validation inspection.
    (iii) HCFA finds that the laboratory meets all the condition-level 
requirements.
    (b) CLIA-exempt laboratory. If a CLIA-exempt laboratory fails to 
comply with the requirements specified in Sec. 493.565, HCFA notifies 
the State of the laboratory's failure to meet the requirements.

Sec. 493.569  Consequences of a finding of noncompliance as a result of 
          a validation inspection.

    (a) Laboratory with a certificate of accreditation. If a validation 
inspection results in a finding that the accredited laboratory is out of 
compliance with one or more condition-level requirements, the laboratory 
is subject to--
    (1) The same requirements and survey and enforcement processes 
applied to laboratories that are not accredited and that are found out 
of compliance following an inspection under this part; and
    (2) Full review by HCFA, in accordance with this part; that is, the 
laboratory is subject to the principal and alternative sanctions in 
Sec. 493.1806.
    (b) CLIA-exempt laboratory. If a validation inspection results in a 
finding that a CLIA-exempt laboratory is out of compliance with one or 
more condition-level requirements, HCFA directs the State to take 
appropriate enforcement action.

[[Page 846]]

Sec. 493.571  Disclosure of accreditation, State and HCFA validation 
          inspection results.

    (a) Accreditation organization inspection results. HCFA may disclose 
accreditation organization inspection results to the public only if the 
results are related to an enforcement action taken by the Secretary.
    (b) State inspection results. Disclosure of State inspection results 
is the responsibility of the approved State licensure program, in 
accordance with State law.
    (c) HCFA validation inspection results. HCFA may disclose the 
results of all validation inspections conducted by HCFA or its agent.

Sec. 493.573  Continuing Federal oversight of private nonprofit 
          accreditation organizations and approved State licensure 
          programs.

    (a) Comparability review. In addition to the initial review for 
determining equivalency of specified organization or State requirements 
to the comparable condition-level requirements, HCFA reviews the 
equivalency of requirements in the following cases:
    (1) When HCFA promulgates new condition-level requirements.
    (2) When HCFA identifies an accreditation organization or a State 
licensure program whose requirements are no longer equal to, or more 
stringent than, condition-level requirements.
    (3) When an accreditation organization or State licensure program 
adopts new requirements.
    (4) When an accreditation organization or State licensure program 
adopts changes to its inspection process, as required by 
Sec. 493.575(b)(1), as applicable.
    (5) Every 6 years, or sooner if HCFA determines an earlier review is 
required.
    (b) Validation review. Following the end of a validation review 
period, HCFA evaluates the validation inspection results for each 
approved accreditation organization and State licensure program.
    (c) Reapplication procedures. (1) Every 6 years, or sooner, as 
determined by HCFA, an approved accreditation organization must reapply 
for continued approval of deeming authority and a State licensure 
program must reapply for continued approval of a CLIA exemption. HCFA 
provides notice of the materials that must be submitted as part of the 
reapplication procedure.
    (2) An accreditation organization or State licensure program that 
does not meet the requirements of this subpart, as determined through a 
comparability or validation review, must furnish HCFA, upon request, 
with the reapplication materials HCFA requests. HCFA establishes a 
deadline by which the materials must be submitted.
    (d) Notice. (1) HCFA provides written notice, as appropriate, to the 
following:
    (i) An accreditation organization indicating that its approval may 
be in jeopardy if a comparability or validation review reveals that it 
is not meeting the requirements of this subpart and HCFA is initiating a 
review of the accreditation organization's deeming authority.
    (ii) A State licensure program indicating that its CLIA exemption 
may be in jeopardy if a comparability or validation review reveals that 
it is not meeting the requirements of this subpart and that a review is 
being initiated of the CLIA exemption of the State's laboratories.
    (2) The notice contains the following information:
    (i) A statement of the discrepancies that were found as well as 
other related documentation.
    (ii) An explanation of HCFA's review process on which the final 
determination is based and a description of the possible actions, as 
specified in Sec. 493.575, that HCFA may impose based on the findings 
from the comparability or validation review.
    (iii) A description of the procedures available if the accreditation 
organization or State licensure program, as applicable, desires an 
opportunity to explain or justify the findings made during the 
comparability or validation review.
    (iv) The reapplication materials that the accreditation organization 
or State licensure program must submit and the deadline for that 
submission.

[[Page 847]]

Sec. 493.575  Removal of deeming authority or CLIA exemption and final 
          determination review.

    (a) HCFA review. HCFA conducts a review of the following:
    (1) A deeming authority review of an accreditation organization's 
program if the comparability or validation review produces findings, as 
described at Sec. 493.573. HCFA reviews, as appropriate, the criteria 
described in Secs. 493.555 and 493.557(a) to reevaluate whether the 
accreditation organization continues to meet all these criteria.
    (2) An exemption review of a State's licensure program if the 
comparability or validation review produces findings, as described at 
Sec. 493.573. HCFA reviews, as appropriate, the criteria described in 
Secs. 493.555 and 493.557(b) to reevaluate whether the licensure program 
continues to meet all these criteria.
    (3) A review of an accreditation organization or State licensure 
program, at HCFA's discretion, if validation review findings, 
irrespective of the rate of disparity, indicate widespread or systematic 
problems in the organization's accreditation or State's licensure 
process that provide evidence that the requirements, taken as a whole, 
are no longer equivalent to CLIA requirements, taken as a whole.
    (4) A review of the accreditation organization or State licensure 
program whenever validation inspection results indicate a rate of 
disparity of 20 percent or more between the findings of the organization 
or State and those of HCFA or a HCFA agent for the following periods:
    (i) One year for accreditation organizations.
    (ii) Two years for State licensure programs.
    (b) HCFA action after review. Following the review, HCFA may take 
the following action:
    (1) If HCFA determines that the accreditation organization or State 
has failed to adopt requirements equal to, or more stringent than, CLIA 
requirements, HCFA may give a conditional approval for a probationary 
period of its deeming authority to an organization 30 days following the 
date of HCFA's determination, or exempt status to a State within 30 days 
of HCFA's determination, both not to exceed 1 year, to afford the 
organization or State an opportunity to adopt equal or more stringent 
requirements.
    (2) If HCFA determines that there are widespread or systematic 
problems in the organization's or State's inspection process, HCFA may 
give conditional approval during a probationary period, not to exceed 1 
year, effective 30 days following the date of the determination.
    (c) Final determination. HCFA makes a final determination as to 
whether the organization or State continues to meet the criteria 
described in this subpart and issues a notice that includes the reasons 
for the determination to the organization or State within 60 days after 
the end of any probationary period. This determination is based on an 
evaluation of any of the following:
    (1) The most recent validation inspection and review findings. To 
continue to be approved, the organization or State must meet the 
criteria of this subpart.
    (2) Facility-specific data, as well as other related information.
    (3) The organization's or State's inspection procedures, surveyors' 
qualifications, ongoing education, training, and composition of 
inspection teams.
    (4) The organization's accreditation requirements, or the State's 
licensure or approval requirements.
    (d) Date of withdrawal of approval. HCFA may withdraw its approval 
of the accreditation organization or State licensure program, effective 
30 days from the date of written notice to the organization or State of 
this proposed action, if improvements acceptable to HCFA have not been 
made during the probationary period.
    (e) Continuation of validation inspections. The existence of any 
validation review, probationary status, or any other action, such as a 
deeming authority review, by HCFA does not affect or limit the conduct 
of any validation inspection.
    (f) Federal Register notice. HCFA publishes a notice in the Federal 
Register containing a justification for removing the deeming authority 
from an accreditation organization, or the CLIA-exempt status of a State 
licensure program.

[[Page 848]]

    (g) Withdrawal of approval-effect on laboratory status--(1) 
Accredited laboratory. After HCFA withdraws approval of an accreditation 
organization's deeming authority, the certificate of accreditation of 
each affected laboratory continues in effect for 60 days after it 
receives notification of the withdrawal of approval.
    (2) CLIA-exempt laboratory. After HCFA withdraws approval of a State 
licensure program, the exempt status of each licensed or approved 
laboratory in the State continues in effect for 60 days after a 
laboratory receives notification from the State of the withdrawal of 
HCFA's approval of the program.
    (3) Extension. After HCFA withdraws approval of an accreditation 
organization or State licensure program, HCFA may extend the period for 
an additional 60 days for a laboratory if it determines that the 
laboratory submitted an application for accreditation to an approved 
accreditation organization or an application for the appropriate 
certificate to HCFA or a HCFA agent before the initial 60-day period 
ends.
    (h) Immediate jeopardy to patients. (1) If at any time HCFA 
determines that the continued approval of deeming authority of any 
accreditation organization poses immediate jeopardy to the patients of 
the laboratories accredited by the organization, or continued approval 
otherwise constitutes a significant hazard to the public health, HCFA 
may immediately withdraw the approval of deeming authority for that 
accreditation organization.
    (2) If at any time HCFA determines that the continued approval of a 
State licensure program poses immediate jeopardy to the patients of the 
laboratories in that State, or continued approval otherwise constitutes 
a significant hazard to the public health, HCFA may immediately withdraw 
the approval of that State licensure program.
    (i) Failure to pay fees. HCFA withdraws the approval of a State 
licensure program if the State fails to pay the applicable fees, as 
specified in Secs. 493.645(a) and 493.646(b).
    (j) State refusal to take enforcement action. (1) HCFA may withdraw 
approval of a State licensure program if the State refuses to take 
enforcement action against a laboratory in that State when HCFA 
determines it to be necessary.
    (2) A laboratory that is in a State in which HCFA has withdrawn 
program approval is subject to the same requirements and survey and 
enforcement processes that are applied to a laboratory that is not 
exempt from CLIA requirements.
    (k) Request for reconsideration. Any accreditation organization or 
State that is dissatisfied with a determination to withdraw approval of 
its deeming authority or remove approval of its State licensure program, 
as applicable, may request that HCFA reconsider the determination, in 
accordance with subpart D of part 488.

                    Subpart F--General Administration

    Source: 57 FR 7138 and 7213, Feb. 28, 1992, unless otherwise noted.

Sec. 493.602  Scope of subpart.

    This subpart sets forth the methodology for determining the amount 
of the fees for issuing the appropriate certificate, and for determining 
compliance with the applicable standards of the Public Health Service 
Act (the PHS Act) and the Federal validation of accredited laboratories 
and of CLIA-exempt laboratories.

[60 FR 20047, Apr. 24, 1995]

Sec. 493.606  Applicability of subpart.

    The rules of this subpart are applicable to those laboratories 
specified in Sec. 493.3.

[58 FR 5212, Jan. 19, 1993]

Sec. 493.638  Certificate fees.

    (a) Basic rule. Laboratories must pay a fee for the issuance of a 
registration certificate, certificate for PPM procedures, certificate of 
waiver, certificate of accreditation, or a certificate of compliance, as 
applicable. Laboratories must also pay a fee to reapply for a 
certificate for PPM procedures, certificate of waiver, certificate of 
accreditation, or a certificate of compliance. The total of fees 
collected by HHS under the laboratory program must be

[[Page 849]]

sufficient to cover the general costs of administering the laboratory 
certification program under section 353 of the PHS Act.
    (1) For registration certificates and certificates of compliance, 
the costs include issuing the certificates, collecting the fees, 
evaluating and monitoring proficiency testing programs, evaluating which 
procedures, tests or examinations meet the criteria for inclusion in the 
appropriate complexity category, and implementing section 353 of the PHS 
Act.
    (2) For a certificate of waiver, the costs include issuing the 
certificate, collecting the fees, determining if a certificate of waiver 
should be issued, evaluating which tests qualify for inclusion in the 
waived category, and other direct administrative costs.
    (3) For a certificate for PPM procedures, the costs include issuing 
the certificate, collecting the fees, determining if a certificate for 
PPM procedures should be issued, evaluating which procedures meet the 
criteria for inclusion in the subcategory of PPM procedures, and other 
direct administrative costs.
    (4) For a certificate of accreditation, the costs include issuing 
the certificate, collecting the fees, evaluating the programs of 
accrediting bodies, and other direct administrative costs.
    (b) Fee amount. The fee amount is set annually by HHS on a calendar 
year basis and is based on the category of test complexity, or on the 
category of test complexity and schedules or ranges of annual laboratory 
test volume (excluding waived tests and tests performed for quality 
control, quality assurance, and proficiency testing purposes) and 
specialties tested, with the amounts of the fees in each schedule being 
a function of the costs for all aspects of general administration of 
CLIA as set forth in Sec. 493.649 (b) and (c). This fee is assessed and 
payable at least biennially. The methodology used to determine the 
amount of the fee is found in Sec. 493.649. The amount of the fee 
applicable to the issuance of the registration certificate or the 
issuance or renewal of the certificate for PPM procedures, certificate 
of waiver, certificate of accreditation, or certificate of compliance is 
the amount in effect at the time the application is received. Upon 
receipt of an application for a certificate, HHS or its designee 
notifies the laboratory of the amount of the required fee for the 
requested certificate.

[60 FR 20047, Apr. 24, 1995]

Sec. 493.639  Fee for revised certificate.

    (a) If, after a laboratory is issued a registration certificate, it 
changes its name or location, the laboratory must pay a fee to cover the 
cost of issuing a revised registration certificate. The fee for the 
revised registration certificate is based on the cost to issue the 
revised certificate to the laboratory.
    (b) A laboratory must pay a fee to cover the cost of issuing a 
revised certificate in any of the following circumstances:
    (1) The fee for issuing an appropriate revised certificate is based 
on the cost to issue the revised certificate to the laboratory as 
follows:
    (i) If a laboratory with a certificate of waiver wishes to perform 
tests in addition to those listed in Sec. 493.15(c) as waived tests, it 
must, as set forth in Sec. 493.638, pay an additional fee for the 
appropriate certificate to cover the additional testing.
    (ii) If a laboratory with a certificate for PPM procedures wishes to 
perform tests in addition to those specified as PPM procedures or listed 
in Sec. 493.15(c) as waived tests, it must, as set forth in 
Sec. 493.638, pay an additional fee for the appropriate certificate to 
cover the additional testing.
    (2) A laboratory must pay a fee to cover the cost of issuing a 
revised certificate when--
    (i) A laboratory changes its name, location, or its director; or
    (ii) A laboratory deletes services or wishes to add services and 
requests that its certificate be changed. (An additional fee is also 
required under Sec. 493.643(d) if it is necessary to determine 
compliance with additional requirements.)

[57 FR 7213, Feb. 28, 1992, as amended at 60 FR 20047, Apr, 24, 1995]

[[Page 850]]

Sec. 493.643  Fee for determination of program compliance.

    (a) Fee requirement. In addition to the fee required under 
Sec. 493.638, a laboratory subject to routine inspections must pay a fee 
to cover the cost of determining program compliance. Laboratories issued 
a certificate for PPM procedures, certificate of waiver, or a 
certificate of accreditation are not subject to this fee for routine 
inspections.
    (b) Costs included in the fee. Included in the fee for determining 
program compliance is the cost of evaluating qualifications of 
personnel; monitoring proficiency testing; conducting onsite 
inspections; documenting deficiencies; evaluating laboratories' plans to 
correct deficiencies; and necessary administrative costs. HHS sets the 
fee amounts annually on a calendar year basis. Laboratories are 
inspected biennially; therefore, fees are assessed and payable 
biennially. If additional expenses are incurred to conduct follow up 
visits to verify correction of deficiencies, to impose sanctions, and/or 
for surveyor preparation for and attendance at ALJ hearings, HHS 
assesses an additional fee to include these costs. The additional fee is 
based on the actual resources and time necessary to perform the 
activities.
    (c) Classification of laboratories that require inspection for 
purpose of determining amount of fee. (1) There are ten classifications 
(schedules) of laboratories for the purpose of determining the fee 
amount a laboratory is assessed. Each laboratory is placed into one of 
the ten following schedules based on the laboratory's scope and volume 
of testing (excluding tests performed for quality control, quality 
assurance, and proficiency testing purposes).
    (i) (A) Schedule A Low Volume. The laboratory performs not more than 
2,000 laboratory tests annually.
    (B) Schedule A. The laboratory performs tests in no more than 3 
specialties of service with a total annual volume of more than 2,000 but 
not more than 10,000 laboratory tests.
    (ii) Schedule B. The laboratory performs tests in at least 4 
specialties of service with a total annual volume of not more than 
10,000 laboratory tests.
    (iii) Schedule C. The laboratory performs tests in no more 3 
specialties of service with a total annual volume of more than 10,000 
but not more than 25,000 laboratory tests.
    (iv) Schedule D. The laboratory performs tests in at least 4 
specialties with a total annual volume of more than 10,000 but not more 
than 25,000 laboratory tests.
    (v) Schedule E. The laboratory performs more than 25,000 but not 
more than 50,000 laboratory tests annually.
    (vi) Schedule F. The laboratory performs more than 50,000 but not 
more than 75,000 laboratory tests annually.
    (vii) Schedule G. The laboratory performs more than 75,000 but not 
more than 100,000 laboratory tests annually.
    (viii) Schedule H. The laboratory performs more than 100,000 but not 
more than 500,000 laboratory tests annually.
    (ix) Schedule I. The laboratory performs more than 500,000 but not 
more than 1,000,000 laboratory tests annually.
    (x) Schedule J. The laboratory performs more than 1,000,000 
laboratory tests annually.
    (2) For purposes of determining a laboratory's classification under 
this section, a test is a procedure or examination for a single analyte. 
(Tests performed for quality control, quality assurance, and proficiency 
testing are excluded from the laboratory's total annual volume). Each 
profile (that is, group of tests) is counted as the number of separate 
procedures or examinations; for example, a chemistry profile consisting 
of 18 tests is counted as 18 separate procedures or tests.
    (3) For purposes of determining a laboratory's classification under 
this section, the specialties and subspecialties of service for 
inclusion are:
    (i) The specialty of Microbiology, which includes one or more of the 
following subspecialties:
    (A) Bacteriology.
    (B) Mycobacteriology.
    (C) Mycology.
    (D) Parasitology.
    (E) Virology.
    (ii) The specialty of Serology, which includes one or more of the 
following subspecialties:
    (A) Syphilis Serology.
    (B) General immunology

[[Page 851]]

    (iii) The specialty of Chemistry, which includes one or more of the 
following subspecialties:
    (A) Routine chemistry.
    (B) Endocrinology.
    (C) Toxicology.
    (D) Urinalysis.
    (iv) The specialty of Hematology.
    (v) The specialty of Immunohematology, which includes one or more of 
the following subspecialties:
    (A) ABO grouping and Rh typing.
    (B) Unexpected antibody detection.
    (C) Compatibility testing.
    (D) Unexpected antibody identification.
    (vi) The specialty of Pathology, which includes the following 
subspecialties:
    (A) Cytology.
    (B) Histopathology.
    (C) Oral pathology.
    (vii) The specialty of Radiobioassay.
    (viii) The specialty of Histocompatibility.
    (ix) The specialty of Cytogenetics.
    (d) Additional fees. (1) If after a certificate of compliance is 
issued, a laboratory adds services and requests that its certificate be 
upgraded, the laboratory must pay an additional fee if, in order to 
determine compliance with additional requirements, it is necessary to 
conduct an inspection, evaluate personnel, or monitor proficiency 
testing performance. The additional fee is based on the actual resources 
and time necessary to perform the activities. HHS revokes the 
laboratory's certificate for failure to pay the compliance determination 
fee.
    (2) If it is necessary to conduct a complaint investigation, impose 
sanctions, or conduct a hearing, HHS assesses the laboratory holding a 
certificate of compliance a fee to cover the cost of these activities. 
If a complaint investigation results in a complaint being 
unsubstantiated, or if an HHS adverse action is overturned at the 
conclusion of the administrative appeals process, the government's costs 
of these activities are not imposed upon the laboratory. Costs for these 
activities are based on the actual resources and time necessary to 
perform the activities and are not assessed until after the laboratory 
concedes the existence of deficiencies or an ALJ rules in favor of HHS. 
HHS revokes the laboratory's certificate of compliance for failure to 
pay the assessed costs.

[57 FR 7138 and 7213, Feb. 28, 1992, as amended at 60 FR 20047, Apr. 24, 
1995]

Sec. 493.645  Additional fee(s) applicable to approved State laboratory 
          programs and laboratories issued a certificate of 
          accreditation, certificate of waiver, or certificate for PPM 
          procedures.

    (a) Approved State laboratory programs. State laboratory programs 
approved by HHS are assessed a fee for the following:
    (1) Costs of Federal inspections of laboratories in that State (that 
is, CLIA-exempt laboratories) to verify that standards are being 
enforced in an appropriate manner.
    (2) Costs incurred for investigations of complaints against the 
State's CLIA-exempt laboratories if the complaint is substantiated.
    (3) Costs of the State's prorata share of general overhead to 
develop and implement CLIA.
    (b) Accredited laboratories. (1) In addition to the certificate fee, 
a laboratory that is issued a certificate of accreditation is also 
assessed a fee to cover the cost of evaluating individual laboratories 
to determine overall whether an accreditation organization's standards 
and inspection policies are equivalent to the Federal program. All 
accredited laboratories share in the cost of these inspections. These 
costs are the same as those that are incurred when inspecting 
nonaccredited laboratories.
    (2) If a laboratory issued a certificate of accreditation has been 
inspected and followup visits are necessary because of identified 
deficiencies, HHS assesses the laboratory a fee to cover the cost of 
these visits. The fee is based on the actual resources and time 
necessary to perform the followup visits. HHS revokes the laboratory's 
certificate of accreditation for failure to pay the assessed fee.
    (c) If, in the case of a laboratory that has been issued a 
certificate of accreditation, certificate of waiver, or certificate for 
PPM procedures, it is necessary to conduct a complaint investigation, 
impose sanctions, or conduct

[[Page 852]]

a hearing, HHS assesses that laboratory a fee to cover the cost of these 
activities. Costs are based on the actual resources and time necessary 
to perform the activities and are not assessed until after the 
laboratory concedes the existence of deficiencies or an ALJ rules in 
favor of HHS. HHS revokes the laboratory's certificate for failure to 
pay the assessed costs. If a complaint investigation results in a 
complaint being unsubstantiated, or if an HHS adverse action is 
overturned at the conclusion of the administrative appeals process, the 
costs of these activities are not imposed upon the laboratory.

[60 FR 20047, Apr. 24, 1995]

Sec. 493.646  Payment of fees.

    (a) Except for CLIA-exempt laboratories, all laboratories are 
notified in writing by HHS or its designee of the appropriate fee(s) and 
instructions for submitting the fee(s), including the due date for 
payment and where to make payment. The appropriate certificate is not 
issued until the applicable fees have been paid.
    (b) For State-exempt laboratories, HHS estimates the cost of 
conducting validation surveys within the State for a 2-year period. HHS 
or its designee notifies the State by mail of the appropriate fees, 
including the due date for payment and the address of the United States 
Department of Treasury designated commercial bank to which payment must 
be made. In addition, if complaint investigations are conducted in 
laboratories within these States and are substantiated, HHS bills the 
State(s) the costs of the complaint investigations.

[57 FR 7138 and 7213, Feb. 28, 1992, as amended at 60 FR 20048, Apr. 24, 
1995]

Sec. 493.649  Methodology for determining fee amount.

    (a) General rule. The amount of the fee in each schedule for 
compliance determination inspections is based on the average hourly rate 
(which includes the costs to perform the required activities and 
necessary administration costs) multiplied by the average number of 
hours required or, if activities are performed by more than one of the 
entities listed in paragraph (b) of this section, the sum of the 
products of the applicable hourly rates multiplied by the average number 
of hours required by the entity to perform the activity. The fee for 
issuance of the registration certificate or certificate of compliance is 
based on the laboratory's scope and volume of testing.
    (b) Determining average hourly rates used in fee schedules. Three 
different entities perform activities related to the issuance or 
reissuance of any certificate. HHS determines the average hourly rates 
for the activities of each of these entities.
    (1) State survey agencies. The following costs are included in 
determining an average hourly rate for the activities performed by State 
survey agencies:
    (i) The costs incurred by the State survey agencies in evaluating 
personnel qualifications and monitoring each laboratory's participation 
in an approved proficiency testing program. The cost of onsite 
inspections and monitoring activities is the hourly rate derived as a 
result of an annual budget negotiation process with each State. The 
hourly rate encompasses salary costs (as determined by each State's 
civil service pay scale) and fringe benefit costs to support the 
required number of State inspectors, management and direct support 
staff.
    (ii) Travel costs necessary to comply with each State's 
administrative requirements and other direct costs such as equipment, 
printing, and supplies. These costs are established based on historical 
State requirements.
    (iii) Indirect costs as negotiated by HHS.
    (2) Federal agencies. The hourly rate for activities performed by 
Federal agencies is the most recent average hourly cost to HHS to staff 
and support a full time equivalent employee. Included in this cost are 
salary and fringe benefit costs, necessary administrative costs, such as 
printing, training, postage, express mail, supplies, equipment, computer 
system and building service charges associated with support services 
provided by organizational components such as a computer center, and any 
other oversight activities necessary to support the program.

[[Page 853]]

    (3) HHS contractors. The hourly rate for activities performed by HHS 
contractors is the average hourly rate established for contractor 
assistance based on an independent government cost estimate for the 
required workload. This rate includes the cost of contractor support to 
provide proficiency testing programs to laboratories that do not 
participate in an approved proficiency testing program, provide 
specialized assistance in the evaluation of laboratory performance in an 
approved proficiency testing program, perform assessments of cytology 
testing laboratories, conduct special studies, bill and collect fees, 
issue certificates, establish accounting, monitoring and reporting 
systems, and assist with necessary surveyor training.
    (c) Determining number of hours. The average number of hours used to 
determine the overall fee in each schedule is HHS's estimate, based on 
historical experience, of the average time needed by each entity to 
perform the activities for which it is responsible.

[57 FR 7138 and 7213, Feb. 28, 1992, as amended at 60 FR 20048, Apr. 24, 
1995]

Subpart G [Reserved]

    Subpart H--Participation in Proficiency Testing for Laboratories 
  Performing Tests of Moderate Complexity (Including the Subcategory), 
           High Complexity, or Any Combination of These Tests

    Source: 57 FR 7146, Feb. 28, 1992, unless otherwise noted.

Sec. 493.801  Condition: Enrollment and testing of samples.

    Each laboratory must enroll in a proficiency testing (PT) program 
that meets the criteria in subpart I of this part and is approved by 
HHS. The laboratory must enroll in an approved program or programs for 
each of the specialties and subspecialties for which it seeks 
certification. The laboratory must test the samples in the same manner 
as patients' specimens. For laboratories subject to 42 CFR part 493 
published on March 14, 1990 (55 FR 9538) prior to September 1, 1992, the 
rules of this subpart are effective on September 1, 1992. For all other 
laboratories, the rules of this subpart are effective January 1, 1994.
    (a) Standard; Enrollment. The laboratory must--
    (1) Notify HHS of the approved program or programs in which it 
chooses to participate to meet proficiency testing requirements of this 
subpart.
    (2)(i) Designate the program(s) to be used for each specialty, 
subspecialty, and analyte or test to determine compliance with this 
subpart if the laboratory participates in more than one proficiency 
testing program approved by HCFA; and
    (ii) For those tests performed by the laboratory that are not 
included in subpart I of this part, a laboratory must establish and 
maintain the accuracy of its testing procedures, in accordance with 
Sec. 493.1709.
    (3) For each specialty, subspecialty and analyte or test, 
participate in one approved proficiency testing program or programs, for 
one year before designating a different program and must notify HCFA 
before any change in designation; and
    (4) Authorize the proficiency testing program to release to HHS all 
data required to--
    (i) Determine the laboratory's compliance with this subpart; and
    (ii) Make PT results available to the public as required in section 
353(f)(3)(F) of the Public Health Service Act.
    (b) Standard; Testing of proficiency testing samples. The laboratory 
must examine or test, as applicable, the proficiency testing samples it 
receives from the proficiency testing program in the same manner as it 
tests patient specimens.
    (1) The samples must be examined or tested with the laboratory's 
regular patient workload by personnel who routinely perform the testing 
in the laboratory, using the laboratory's routine methods. The 
individual testing or examining the samples and the laboratory director 
must attest to the routine integration of the samples into the patient 
workload using the laboratory's routine methods.

[[Page 854]]

    (2) The laboratory must test samples the same number of times that 
it routinely tests patient samples.
    (3) Laboratories that perform tests on proficiency testing samples 
must not engage in any inter-laboratory communications pertaining to the 
results of proficiency testing sample(s) until after the date by which 
the laboratory must report proficiency testing results to the program 
for the testing event in which the samples were sent. Laboratories with 
multiple testing sites or separate locations must not participate in any 
communications or discussions across sites/locations concerning 
proficiency testing sample results until after the date by which the 
laboratory must report proficiency testing results to the program.
    (4) The laboratory must not send PT samples or portions of samples 
to another laboratory for any analysis which it is certified to perform 
in its own laboratory. Any laboratory that HCFA determines intentionally 
referred its proficiency testing samples to another laboratory for 
analysis will have its certification revoked for at least one year. Any 
laboratory that receives proficiency testing samples from another 
laboratory for testing must notify HCFA of the receipt of those samples.
    (5) The laboratory must document the handling, preparation, 
processing, examination, and each step in the testing and reporting of 
results for all proficiency testing samples. The laboratory must 
maintain a copy of all records, including a copy of the proficiency 
testing program report forms used by the laboratory to record 
proficiency testing results including the attestation statement provided 
by the PT program, signed by the analyst and the laboratory director, 
documenting that proficiency testing samples were tested in the same 
manner as patient specimens, for a minimum of two years from the date of 
the proficiency testing event.
    (6) PT is required for only the test system, assay, or examination 
used as the primary method for patient testing during the PT event.

[57 FR 7146, Feb. 28, 1992, as amended at 58 FR 5228, Jan. 19, 1993]

Sec. 493.803  Condition: Successful participation.

    (a) Each laboratory performing tests of moderate complexity 
(including the subcategory) and/or high complexity must successfully 
participate in a proficiency testing program approved by HCFA, if 
applicable, as described in subpart I of this part for each specialty, 
subspecialty, and analyte or test in which the laboratory is certified 
under CLIA.
    (b) Except as specified in paragraph (c) of this section, if a 
laboratory fails to participate successfully in proficiency testing for 
a given specialty, subspecialty, analyte or test, as defined in this 
section, or fails to take remedial action when an individual fails 
gynecologic cytology, HCFA imposes sanctions, as specified in subpart R 
of this part.
    (c) If a laboratory fails to perform successfully in a HCFA-approved 
proficiency testing program, for the initial unsuccessful performance, 
HCFA may direct the laboratory to undertake training of its personnel or 
to obtain technical assistance, or both, rather than imposing 
alternative or principle sanctions except when one or more of the 
following conditions exists:
    (1) There is immediate jeopardy to patient health and safety.
    (2) The laboratory fails to provide HCFA or a HCFA agent with 
satisfactory evidence that it has taken steps to correct the problem 
identified by the unsuccessful proficiency testing performance.
    (3) The laboratory has a poor compliance history.

[57 FR 7146, Feb. 28, 1992, as amended at 60 FR 20048, Apr. 24, 1995; 63 
FR 26737, May 14, 1998]

Sec. 493.807  Condition: Reinstatement of laboratories performing tests 
          of moderate complexity (including the subcategory), high 
          complexity, or any combination of these tests, after failure 
          to participate successfully.

    (a) If a laboratory's certificate is suspended or limited or its 
Medicare or Medicaid approval is cancelled or its Medicare or Medicaid 
payments are suspended because it fails to participate successfully in 
proficiency testing

[[Page 855]]

for one or more specialties, subspecialties, analyte or test, or 
voluntarily withdraws its certification under CLIA for the failed 
specialty, subspecialty, or analyte, the laboratory must then 
demonstrate sustained satisfactory performance on two consecutive 
proficiency testing events, one of which may be on site, before HCFA 
will consider it for reinstatement for certification and Medicare or 
Medicaid approval in that specialty, subspecialty, analyte or test.
    (b) The cancellation period for Medicare and Medicaid approval or 
period for suspension of Medicare or Medicaid payments or suspension or 
limitation of certification under CLIA for the failed specialty, 
subspecialty, or analyte or test is for a period of not less than six 
months from the date of cancellation, limitation or suspension of the 
CLIA certificate.

[58 FR 5228, Jan. 19, 1993, as amended at 60 FR 20048, Apr. 24, 1995]

   Proficiency Testing by Specialty and Subspecialty for Laboratories 
  Performing Tests of Moderate Complexity (Including the Subcategory), 
           High Complexity, or Any Combination of These Tests

Sec. 493.821  Condition: Microbiology.

    The specialty of microbiology includes, for purposes of proficiency 
testing, the subspecialties of bacteriology, mycobacteriology, mycology, 
parasitology and virology.

Sec. 493.823  Standard; Bacteriology.

    (a) Failure to attain an overall testing event score of at least 80 
percent is unsatisfactory performance.
    (b) Failure to participate in a testing event is unsatisfactory 
performance and results in a score of 0 for the testing event. 
Consideration may be given to those laboratories failing to participate 
in a testing event only if--
    (1) Patient testing was suspended during the time frame allotted for 
testing and reporting proficiency testing results;
    (2) The laboratory notifies the inspecting agency and the 
proficiency testing program within the time frame for submitting 
proficiency testing results of the suspension of patient testing and the 
circumstances associated with failure to perform tests on proficiency 
testing samples; and
    (3) The laboratory participated in the previous two proficiency 
testing events.
    (c) Failure to return proficiency testing results to the proficiency 
testing program within the time frame specified by the program is 
unsatisfactory performance and results in a score of 0 for the testing 
event.
    (d)(1) For any unsatisfactory testing event for reasons other than a 
failure to participate, the laboratory must undertake appropriate 
training and employ the technical assistance necessary to correct 
problems associated with a proficiency testing failure.
    (2) Remedial action must be taken and documented, and the 
documentation must be maintained by the laboratory for two years from 
the date of participation in the proficiency testing event.
    (e) Failure to achieve an overall testing event score of 
satisfactory performance for two consecutive testing events or two out 
of three consecutive testing events is unsuccessful performance.

Sec. 493.825  Standard; Mycobacteriology.

    (a) Failure to attain an overall testing event score of at least 80 
percent is unsatisfactory performance.
    (b) Failure to participate in a testing event is unsatisfactory 
performance and results in a score of 0 for the testing event. 
Consideration may be given to those laboratories failing to participate 
in a testing event only if--
    (1) Patient testing was suspended during the time frame allotted for 
testing and reporting proficiency testing results;
    (2) The laboratory notifies the inspecting agency and the 
proficiency testing program within the time frame for submitting 
proficiency testing results of the suspension of patient testing and the 
circumstances associated with failure to perform tests on proficiency 
testing samples; and
    (3) The laboratory participated in the previous two proficiency 
testing events.
    (c) Failure to return proficiency testing results to the proficiency 
testing

[[Page 856]]

program within the time frame specified by the program is unsatisfactory 
performance and results in a score of 0 for the testing event.
    (d)(1) For any unsatisfactory testing event for reasons other than a 
failure to participate, the laboratory must undertake appropriate 
training and employ the technical assistance necessary to correct 
problems associated with a proficiency testing failure.
    (2) Remedial action must be taken and documented, and the 
documentation must be maintained by the laboratory for two years from 
the date of participation in the proficiency testing event.
    (e) Failure to achieve an overall testing event score of 
satisfactory performance for two consecutive testing events or two out 
of three consecutive testing events is unsuccessful performance.

Sec. 493.827  Standard; Mycology.

    (a) Failure to attain an overall testing event score of at least 80 
percent is unsatisfactory performance.
    (b) Failure to participate in a testing event is unsatisfactory 
performance and results in a score of 0 for the testing event. 
Consideration may be given to those laboratories failing to participate 
in a testing event only if--
    (1) Patient testing was suspended during the time frame allotted for 
testing and reporting proficiency testing results;
    (2) The laboratory notifies the inspecting agency and the 
proficiency testing program within the time frame for submitting 
proficiency testing results of the suspension of patient testing and the 
circumstances associated with failure to perform tests on proficiency 
testing samples; and
    (3) The laboratory participated in the previous two proficiency 
testing events.
    (c) Failure to return proficiency testing results to the proficiency 
testing program within the time frame specified by the program is 
unsatisfactory performance and results in a score of 0 for the testing 
event.
    (d)(1) For any unsatisfactory testing event for reasons other than a 
failure to participate, the laboratory must undertake appropriate 
training and employ the technical assistance necessary to correct 
problems associated with a proficiency testing failure.
    (2) Remedial action must be taken and documented, and the 
documentation must be maintained by the laboratory for two years from 
the date of participation in the proficiency testing event.
    (e) Failure to achieve an overall testing event score of 
satisfactory performance for two consecutive testing events or two out 
of three consecutive testing events is unsuccessful performance.

Sec. 493.829  Standard; Parasitology.

    (a) Failure to attain an overall testing event score of at least 80 
percent is unsatisfactory performance.
    (b) Failure to participate in a testing event is unsatisfactory 
performance and results in a score of 0 for the testing event. 
Consideration may be given to those laboratories failing to participate 
in a testing event only if--
    (1) Patient testing was suspended during the time frame allotted for 
testing and reporting proficiency testing results;
    (2) The laboratory notifies the inspecting agency and the 
proficiency testing program within the time frame for submitting 
proficiency testing results of the suspension of patient testing and the 
circumstances associated with failure to perform tests on proficiency 
testing samples; and
    (3) The laboratory participated in the previous two proficiency 
testing events.
    (c) Failure to return proficiency testing results to the proficiency 
testing program within the time frame specified by the program is 
unsatisfactory performance and results in a score of 0 for the testing 
event.
    (d)(1) For any unsatisfactory testing event for reasons other than a 
failure to participate, the laboratory must undertake appropriate 
training and employ the technical assistance necessary to correct 
problems associated with a proficiency testing failure.
    (2) Remedial action must be taken and documented, and the 
documentation must be maintained by the laboratory for two years from 
the date of participation in the proficiency testing event.

[[Page 857]]

    (e) Failure to achieve an overall testing event score of 
satisfactory performance for two consecutive testing events or two out 
of three consecutive testing events is unsuccessful performance.

Sec. 493.831  Standard; Virology.

    (a) Failure to attain an overall testing event score of at least 80 
percent is unsatisfactory performance.
    (b) Failure to participate in a testing event is unsatisfactory 
performance and results in a score of 0 for the testing event. 
Consideration may be given to those laboratories failing to participate 
in a testing event only if--
    (1) Patient testing was suspended during the time frame allotted for 
testing and reporting proficiency testing results;
    (2) The laboratory notifies the inspecting agency and the 
proficiency testing program within the time frame for submitting 
proficiency testing results of the suspension of patient testing and the 
circumstances associated with failure to perform tests on proficiency 
testing samples; and
    (3) The laboratory participated in the previous two proficiency 
testing events.
    (c) Failure to return proficiency testing results to the proficiency 
testing program within the time frame specified by the program is 
unsatisfactory performance and results in a score of 0 for the testing 
event.
    (d)(1) For any unsatisfactory testing event for reasons other than a 
failure to participate, the laboratory must undertake appropriate 
training and employ the technical assistance necessary to correct 
problems associated with a proficiency testing failure.
    (2) For any unsatisfactory testing events, remedial action must be 
taken and documented, and the documentation must be maintained by the 
laboratory for two years from the date of participation in the 
proficiency testing event.
    (e) Failure to achieve an overall testing event score of 
satisfactory performance for two consecutive testing events or two out 
of three consecutive testing events is unsuccessful performance.

Sec. 493.833  Condition: Diagnostic immunology.

    The specialty of diagnostic immunology includes for purposes of 
proficiency testing the subspecialties of syphilis serology and general 
immunology.

Sec. 493.835  Standard; Syphilis serology.

    (a) Failure to attain an overall testing event score of at least 80 
percent is unsatisfactory performance.
    (b) Failure to participate in a testing event is unsatisfactory 
performance and results in a score of 0 for the testing event. 
Consideration may be given to those laboratories failing to participate 
in a testing event only if--
    (1) Patient testing was suspended during the time frame allotted for 
testing and reporting proficiency testing results;
    (2) The laboratory notifies the inspecting agency and the 
proficiency testing program within the time frame for submitting 
proficiency testing results of the suspension of patient testing and the 
circumstances associated with failure to perform tests on proficiency 
testing samples; and
    (3) The laboratory participated in the previous two proficiency 
testing events.
    (c) Failure to return proficiency testing results to the proficiency 
testing program within the time frame specified by the program is 
unsatisfactory performance and results in a score of 0 for the testing 
event.
    (d)(1) For any unsatisfactory testing event for reasons other than a 
failure to participate, the laboratory must undertake appropriate 
training and employ the technical assistance necessary to correct 
problems associated with a proficiency testing failure.
    (2) For any unacceptable testing event score, remedial action must 
be taken and documented, and the documentation must be maintained by the 
laboratory for two years from the date of participation in the 
proficiency testing event.
    (e) Failure to achieve an overall testing event score of 
satisfactory performance for two consecutive testing events or two out 
of three consecutive testing events is unsuccessful performance.

[[Page 858]]

Sec. 493.837  Standard; General immunology.

    (a) Failure to attain a score of at least 80 percent of acceptable 
responses for each analyte in each testing event is unsatisfactory 
analyte performance for the testing event.
    (b) Failure to attain an overall testing event score of at least 80 
percent is unsatisfactory performance.
    (c) Failure to participate in a testing event is unsatisfactory 
performance and results in a score of 0 for the testing event. 
Consideration may be given to those laboratories failing to participate 
in a testing event only if--
    (1) Patient testing was suspended during the time frame allotted for 
testing and reporting proficiency testing results;
    (2) The laboratory notifies the inspecting agency and the 
proficiency testing program within the time frame for submitting 
proficiency testing results of the suspension of patient testing and the 
circumstances associated with failure to perform tests on proficiency 
testing samples; and
    (3) The laboratory participated in the previous two proficiency 
testing events.
    (d) Failure to return proficiency testing results to the proficiency 
testing program within the time frame specified by the program is 
unsatisfactory performance and results in a score of 0 for the testing 
event.
    (e)(1) For any unsatisfactory analyte or test performance or testing 
event for reasons other than a failure to participate, the laboratory 
must undertake appropriate training and employ the technical assistance 
necessary to correct problems associated with a proficiency testing 
failure.
    (2) For any unacceptable analyte or testing event score, remedial 
action must be taken and documented, and the documentation must be 
maintained by the laboratory for two years from the date of 
participation in the proficiency testing event.
    (f) Failure to achieve satisfactory performance for the same analyte 
or test in two consecutive testing events or two out of three 
consecutive testing events is unsuccessful performance.
    (g) Failure to achieve an overall testing event score of 
satisfactory performance for two consecutive testing events or two out 
of three consecutive testing events is unsuccessful performance.

Sec. 493.839  Condition: Chemistry.

    The specialty of chemistry includes for the purposes of proficiency 
testing the subspecialties of routine chemistry, endocrinology, and 
toxicology.

Sec. 493.841  Standard; Routine chemistry.

    (a) Failure to attain a score of at least 80 percent of acceptable 
responses for each analyte in each testing event is unsatisfactory 
analyte performance for the testing event.
    (b) Failure to attain an overall testing event score of at least 80 
percent is unsatisfactory performance.
    (c) Failure to participate in a testing event is unsatisfactory 
performance and results in a score of 0 for the testing event. 
Consideration may be given to those laboratories failing to participate 
in a testing event only if--
    (1) Patient testing was suspended during the time frame allotted for 
testing and reporting proficiency testing results;
    (2) The laboratory notifies the inspecting agency and the 
proficiency testing program within the time frame for submitting 
proficiency testing results of the suspension of patient testing and the 
circumstances associated with failure to perform tests on proficiency 
testing samples; and
    (3) The laboratory participated in the previous two proficiency 
testing events.
    (d) Failure to return proficiency testing results to the proficiency 
testing program within the time frame specified by the program is 
unsatisfactory performance and results in a score of 0 for the testing 
event.
    (e)(1) For any unsatisfactory analyte or test performance or testing 
event for reasons other than a failure to participate, the laboratory 
must undertake appropriate training and employ the technical assistance 
necessary to correct problems associated with a proficiency testing 
failure.
    (2) For any unacceptable analyte or testing event score, remedial 
action must be taken and documented, and the documentation must be 
maintained by the laboratory for two years from

[[Page 859]]

the date of participation in the proficiency testing event.
    (f) Failure to achieve satisfactory performance for the same analyte 
or test in two consecutive testing events or two out of three 
consecutive testing events is unsuccessful performance.
    (g) Failure to achieve an overall testing event score of 
satisfactory performance for two consecutive testing events or two out 
of three consecutive testing events is unsuccessful performance.

Sec. 493.843  Standard; Endocrinology.

    (a) Failure to attain a score of at least 80 percent of acceptable 
responses for each analyte in each testing event is unsatisfactory 
analyte performance for the testing event.
    (b) Failure to attain an overall testing event score of at least 80 
percent is unsatisfactory performance.
    (c) Failure to participate in a testing event is unsatisfactory 
performance and results in a score of 0 for the testing event. 
Consideration may be given to those laboratories failing to participate 
in a testing event only if--
    (1) Patient testing was suspended during the time frame allotted for 
testing and reporting proficiency testing results;
    (2) The laboratory notifies the inspecting agency and the 
proficiency testing program within the time frame for submitting 
proficiency testing results of the suspension of patient testing and the 
circumstances associated with failure to perform tests on proficiency 
testing samples; and
    (3) The laboratory participated in the previous two proficiency 
testing events.
    (d) Failure to return proficiency testing results to the proficiency 
testing program within the time frame specified by the program is 
unsatisfactory performance and results in a score of 0 for the testing 
event.
    (e)(1) For any unsatisfactory analyte or test performance or testing 
event for reasons other than a failure to participate, the laboratory 
must undertake appropriate training and employ the technical assistance 
necessary to correct problems associated with a proficiency testing 
failure.
    (2) For any unacceptable analyte or testing event score, remedial 
action must be taken and documented, and the documentation must be 
maintained by the laboratory for two years from the date of 
participation in the proficiency testing event.
    (f) Failure to achieve satisfactory performance for the same analyte 
or test in two consecutive testing events or two out of three 
consecutive testing events is unsuccessful performance.
    (g) Failure to achieve an overall testing event score of 
satisfactory performance for two consecutive testing events or two out 
of three consecutive testing events is unsuccessful performance.

Sec. 493.845  Standard; Toxicology.

    (a) Failure to attain a score of at least 80 percent of acceptable 
responses for each analyte in each testing event is unsatisfactory 
analyte performance for the testing event.
    (b) Failure to attain an overall testing event score of at least 80 
percent is unsatisfactory performance.
    (c) Failure to participate in a testing event is unsatisfactory 
performance and results in a score of 0 for the testing event. 
Consideration may be given to those laboratories failing to participate 
in a testing event only if--
    (1) Patient testing was suspended during the time frame allotted for 
testing and reporting proficiency testing results;
    (2) The laboratory notifies the inspecting agency and the 
proficiency testing program within the time frame for submitting 
proficiency testing results of the suspension of patient testing and the 
circumstances associated with failure to perform tests on proficiency 
testing samples; and
    (3) The laboratory participated in the previous two proficiency 
testing events.
    (d) Failure to return proficiency testing results to the proficiency 
testing program within the time frame specified by the program is 
unsatisfactory performance and results in a score of 0 for the testing 
event.
    (e)(1) For any unsatisfactory analyte or test performance or testing 
event for reasons other than a failure to participate, the laboratory 
must undertake appropriate training and employ the technical assistance 
necessary to

[[Page 860]]

correct problems associated with a proficiency testing failure.
    (2) For any unacceptable analyte or testing event score, remedial 
action must be taken and documented, and the documentation must be 
maintained by the laboratory for two years from the date of 
participation in the proficiency testing event.
    (f) Failure to achieve satisfactory performance for the same analyte 
or test in two consecutive testing events or two out of three 
consecutive testing events is unsuccessful performance.
    (g) Failure to achieve an overall testing event score of 
satisfactory performance for two consecutive testing events or two out 
of three consecutive testing events is unsuccessful performance.

Sec. 493.849  Condition: Hematology.

    The specialty of hematology, for the purpose of proficiency testing, 
is not subdivided into subspecialties of testing.

Sec. 493.851  Standard; Hematology.

    (a) Failure to attain a score of at least 80 percent of acceptable 
responses for each analyte in each testing event is unsatisfactory 
analyte performance for the testing event.
    (b) Failure to attain an overall testing event score of at least 80 
percent is unsatisfactory performance.
    (c) Failure to participate in a testing event is unsatisfactory 
performance and results in a score of 0 for the testing event. 
Consideration may be given to those laboratories failing to participate 
in a testing event only if--
    (1) Patient testing was suspended during the time frame allotted for 
testing and reporting proficiency testing results;
    (2) The laboratory notifies the inspecting agency and the 
proficiency testing program within the time frame for submitting 
proficiency testing results of the suspension of patient testing and the 
circumstances associated with failure to perform tests on proficiency 
testing samples; and
    (3) The laboratory participated in the previous two proficiency 
testing events.
    (d) Failure to return proficiency testing results to the proficiency 
testing program within the time frame specified by the program is 
unsatisfactory performance and results in a score of 0 for the testing 
event.
    (e)(1) For any unsatisfactory analyte or test performance or testing 
event for reasons other than a failure to participate, the laboratory 
must undertake appropriate training and employ the technical assistance 
necessary to correct problems associated with a proficiency testing 
failure.
    (2) For any unacceptable analyte or testing event score, remedial 
action must be taken and documented, and the documentation must be 
maintained by the laboratory for two years from the date of 
participation in the proficiency testing event.
    (f) Failure to achieve satisfactory performance for the same analyte 
in two consecutive events or two out of three consecutive testing events 
is unsuccessful performance.
    (g) Failure to achieve an overall testing event score of 
satisfactory performance for two consecutive testing events or two out 
of three consecutive testing events is unsuccessful performance.

Sec. 493.853  Condition: Pathology.

    The specialty of pathology includes, for purposes of proficiency 
testing, the subspecialty of cytology limited to gynecologic 
examinations.

Sec. 493.855  Standard; Cytology: gynecologic examinations.

    To participate successfully in a cytology proficiency testing 
program for gynecologic examinations (Pap smears), the laboratory must 
meet the requirements of paragraphs (a) through (c) of this section.
    (a) The laboratory must ensure that each individual engaged in the 
examination of gynecologic preparations is enrolled in a proficiency 
testing program approved by HCFA by January 1, 1995, if available in the 
State in which he or she is employed. The laboratory must ensure that 
each individual is tested at least once per year and obtains a passing 
score. To ensure this annual testing of individuals, an announced or 
unannounced testing event will be conducted on-site in each laboratory 
at least once each year. Laboratories will be notified of the time of 
each announced on-site testing event

[[Page 861]]

at least 30 days prior to each event. Additional testing events will be 
conducted as necessary in each State or region for the purpose of 
testing individuals who miss the on-site testing event and for retesting 
individuals as described in paragraph (b) of this section.
    (b) The laboratory must ensure that each individual participates in 
an annual testing event that involves the examination of a 10-slide test 
set as described in Sec. 493.945. Individuals who fail this testing 
event are retested with another 10-slide test set as described in 
paragraphs (b)(1) and (b)(2) of this section. Individuals who fail this 
second test are subsequently retested with a 20-slide test set as 
described in paragraphs (b)(2) and (b)(3) of this section. Individuals 
are given not more than 2 hours to complete a 10-slide test and not more 
than 4 hours to complete a 20-slide test. Unexcused failure to appear by 
an individual for a retest will result in test failure with resulting 
remediation and limitations on slide examinations as specified in 
(b)(1), (b)(2), and (b)(3) of this section.
    (1) An individual is determined to have failed the annual testing 
event if he or she scores less than 90 percent on a 10-slide test set. 
For an individual who fails an annual proficiency testing event, the 
laboratory must schedule a retesting event which must take place not 
more than 45 days after receipt of the notification of failure.
    (2) An individual is determined to have failed the second testing 
event if he or she scores less than 90 percent on a 10-slide test set. 
For an individual who fails a second testing event, the laboratory must 
provide him or her with documented, remedial training and education in 
the area of failure, and must assure that all gynecologic slides 
evaluated subsequent to the notice of failure are reexamined until the 
individual is again retested with a 20-slide test set and scores at 
least 90 percent. Reexamination of slides must be documented.
    (3) An individual is determined to have failed the third testing 
event if he or she scores less than 90 percent on a 20-slide test set. 
An individual who fails the third testing event must cease examining 
gynecologic slide preparations immediately upon notification of test 
failure and may not resume examining gynecologic slides until the 
laboratory assures that the individual obtains at least 35 hours of 
documented, formally structured, continuing education in diagnostic 
cytopathology that focuses on the examination of gynecologic 
preparations, and until he or she is retested with a 20-slide test set 
and scores at least 90 percent.
    (c) If a laboratory fails to ensure that individuals are tested or 
those who fail a testing event are retested, or fails to take required 
remedial actions as described in paragraphs (b)(1), (b)(2) or (b)(3) of 
this section, HCFA will initiate intermediate sanctions or limit the 
laboratory's certificate to exclude gynecologic cytology testing under 
CLIA, and, if applicable, suspend the laboratory's Medicare and Medicaid 
payments for gynecologic cytology testing in accordance with subpart R 
of this part.

[57 FR 7146, Feb. 28, 1992, as amended at 58 FR 5228, Jan. 19, 1993; 59 
FR 62609, Dec. 6, 1994]

Sec. 493.857  Condition: Immunohematology.

    The specialty of immunohematology includes four subspecialties for 
the purposes of proficiency testing: ABO group and D (Rho) typing; 
unexpected antibody detection; compatibility testing; and antibody 
identification.

Sec. 493.859  Standard; ABO group and D (Rho) typing.

    (a) Failure to attain a score of at least 100 percent of acceptable 
responses for each analyte or test in each testing event is 
unsatisfactory analyte performance for the testing event.
    (b) Failure to attain an overall testing event score of at least 100 
percent is unsatisfactory performance.
    (c) Failure to participate in a testing event is unsatisfactory 
performance and results in a score of 0 for the testing event. 
Consideration may be given to those laboratories failing to participate 
in a testing event only if--
    (1) Patient testing was suspended during the time frame allotted for 
testing and reporting proficiency testing results;

[[Page 862]]

    (2) The laboratory notifies the inspecting agency and the 
proficiency testing program within the time frame for submitting 
proficiency testing results of the suspension of patient testing and the 
circumstances associated with failure to perform tests on proficiency 
testing samples; and
    (3) The laboratory participated in the previous two proficiency 
testing events.
    (d) Failure to return proficiency testing results to the proficiency 
testing program within the time frame specified by the program is 
unsatisfactory performance and results in a score of 0 for the testing 
event.
    (e)(1) For any unsatisfactory testing event for reasons other than a 
failure to participate, the laboratory must undertake appropriate 
training and employ the technical assistance necessary to correct 
problems associated with a proficiency testing failure.
    (2) For any unacceptable analyte or unsatisfactory testing event 
score, remedial action must be taken and documented, and the 
documentation must be maintained by the laboratory for two years from 
the date of participation in the proficiency testing event.
    (f) Failure to achieve satisfactory performance for the same analyte 
in two consecutive testing events or two out of three consecutive 
testing events is unsuccessful performance.
    (g) Failure to achieve an overall testing event score of 
satisfactory for two consecutive testing events or two out of three 
consecutive testing events is unsuccessful performance.

Sec. 493.861  Standard; Unexpected antibody detection.

    (a) Failure to attain an overall testing event score of at least 80 
percent is unsatisfactory performance.
    (b) Failure to participate in a testing event is unsatisfactory 
performance and results in a score of 0 for the testing event. 
Consideration may be given to those laboratories failing to participate 
in a testing event only if--
    (1) Patient testing was suspended during the time frame allotted for 
testing and reporting proficiency testing results;
    (2) The laboratory notifies the inspecting agency and the 
proficiency testing program within the time frame for submitting 
proficiency testing results of the suspension of patient testing and the 
circumstances associated with failure to perform tests on proficiency 
testing samples; and
    (3) The laboratory participated in the previous two proficiency 
testing events.
    (c) Failure to return proficiency testing results to the proficiency 
testing program within the time frame specified by the program is 
unsatisfactory performance and results in a score of 0 for the testing 
event.
    (d)(1) For any unsatisfactory testing event for reasons other than a 
failure to participate, the laboratory must undertake appropriate 
training and employ the technical assistance necessary to correct 
problems associated with a proficiency testing failure.
    (2) For any unsatisfactory testing event score, remedial action must 
be taken and documented, and the documentation must be maintained by the 
laboratory for two years from the date of participation in the 
proficiency testing event.
    (e) Failure to achieve an overall testing event score of 
satisfactory for two consecutive testing events or two out of three 
consecutive testing events is unsuccessful performance.

Sec. 493.863  Standard; Compatibility testing.

    (a) Failure to attain an overall testing event score of at least 100 
percent is unsatisfactory performance.
    (b) Failure to participate in a testing event is unsatisfactory 
performance and results in a score of 0 for the testing event. 
Consideration may be given to those laboratories failing to participate 
in a testing event only if--
    (1) Patient testing was suspended during the time frame allotted for 
testing and reporting proficiency testing results;
    (2) The laboratory notifies the inspecting agency and the 
proficiency testing program within the time frame for submitting 
proficiency testing results of the suspension of patient testing and the 
circumstances associated with failure to perform tests on proficiency 
testing samples; and

[[Page 863]]

    (3) The laboratory participated in the previous two proficiency 
testing events.
    (c) Failure to return proficiency testing results to the proficiency 
testing program within the time frame specified by the program is 
unsatisfactory performance and results in a score of 0 for the testing 
event.
    (d)(1) For any unsatisfactory testing event for reasons other than a 
failure to participate, the laboratory must undertake appropriate 
training and employ the technical assistance necessary to correct 
problems associated with a proficiency testing failure.
    (2) For any unsatisfactory testing event score, remedial action must 
be taken and documented, and the documentation must be maintained by the 
laboratory for two years from the date of participation in the 
proficiency testing event.
    (e) Failure to achieve an overall testing event score of 
satisfactory for two consecutive testing events or two out of three 
consecutive testing events is unsuccessful performance.

Sec. 493.865  Standard; Antibody identification.

    (a) Failure to attain an overall testing event score of at least 80 
percent is unsatisfactory performance.
    (b) Failure to participate in a testing event is unsatisfactory 
performance and results in a score of 0 for the testing event. 
Consideration may be given to those laboratories failing to participate 
in a testing event only if--
    (1) Patient testing was suspended during the time frame allotted for 
testing and reporting proficiency testing results;
    (2) The laboratory notifies the inspecting agency and the 
proficiency testing program within the time frame for submitting 
proficiency testing results of the suspension of patient testing and the 
circumstances associated with failure to perform tests on proficiency 
testing samples; and
    (3) The laboratory participated in the previous two proficiency 
testing events.
    (c) Failure to return proficiency testing results to the proficiency 
testing program within the time frame specified by the program is 
unsatisfactory performance and results in a score of 0 for the testing 
event.
    (d)(1) For any unsatisfactory testing event for reasons other than a 
failure to participate, the laboratory must undertake appropriate 
training and employ the technical assistance necessary to correct 
problems associated with a proficiency testing failure.
    (2) For any unsatisfactory testing event score, remedial action must 
be taken and documented, and the documentation must be maintained by the 
laboratory for two years from the date of participation in the 
proficiency testing event.
    (e) Failure to identify the same antibody in two consecutive or two 
out of three consecutive testing events is unsuccessful performance.
    (f) Failure to achieve an overall testing event score of 
satisfactory for two consecutive testing events or two out of three 
consecutive testing events is unsuccessful performance.

Subpart I--Proficiency Testing Programs for Tests of Moderate Complexity 
  (Including the Subcategory), High Complexity, or Any Combination of 
                               These Tests

    Source: 57 FR 7151, Feb. 28, 1992, unless otherwise noted.

Sec. 493.901  Approval of proficiency testing programs.

    In order for a proficiency testing program to receive HHS approval, 
the program must be offered by a private nonprofit organization or a 
Federal or State agency, or entity acting as a designated agent for the 
State. An organization, Federal, or State program seeking approval or 
reapproval for its program for the next calendar year must submit an 
application providing the required information by July 1 of the current 
year. The organization, Federal, or State program must provide technical 
assistance to laboratories seeking to qualify under the program, and 
must, for each specialty, subspecialty, and analyte or test for which it 
provides testing--
    (a) Assure the quality of test samples, appropriately evaluate and 
score

[[Page 864]]

the testing results, and identify performance problems in a timely 
manner;
    (b) Demonstrate to HHS that it has--
    (1) The technical ability required to--
    (i) Prepare or purchase samples from manufacturers who prepare the 
samples in conformance with the appropriate good manufacturing practices 
required in 21 CFR parts 606, 640, and 820; and
    (ii) Distribute the samples, using rigorous quality control to 
assure that samples mimic actual patient specimens when possible and 
that samples are homogeneous, except for specific subspecialties such as 
cytology, and will be stable within the time frame for analysis by 
proficiency testing participants;
    (2) A scientifically defensible process for determining the correct 
result for each challenge offered by the program;
    (3) A program of sufficient annual challenge and with the frequency 
specified in Secs. 493.909 through 493.959 to establish that a 
laboratory has met minimum performance requirements;
    (4) The resources needed to provide Statewide or nationwide reports 
to regulatory agencies on individual's performance for gynecologic 
cytology and on individual laboratory performance on testing events, 
cumulative reports and scores for each laboratory or individual, and 
reports of specific laboratory failures using grading criteria 
acceptable to HHS. These reports must be provided to HHS on a timely 
basis when requested;
    (5) Provisions to include on each proficiency testing program report 
form used by the laboratory to record testing event results, an 
attestation statement that proficiency testing samples were tested in 
the same manner as patient specimens with a signature block to be 
completed by the individual performing the test as well as by the 
laboratory director;
    (6) A mechanism for notifying participants of the PT shipping 
schedule and for participants to notify the proficiency testing program 
within three days of the expected date of receipt of the shipment that 
samples have not arrived or are unacceptable for testing. The program 
must have provisions for replacement of samples that are lost in transit 
or are received in a condition that is unacceptable for testing; and
    (7) A process to resolve technical, administrative, and scientific 
problems about program operations;
    (c) Meet the specific criteria for proficiency testing programs 
listed by specialty, subspecialty, and analyte or test contained in 
Secs. 493.901 through 493.959 for initial approval and thereafter 
provide HHS, on an annual basis, with the information necessary to 
assure that the proficiency testing program meets the criteria required 
for approval; and
    (d) Comply with all applicable packaging, shipment, and notification 
requirements of 42 CFR part 72.

[57 FR 7151, Feb. 28, 1992, as amended at 58 FR 5228, Jan. 19, 1993]

Sec. 493.903  Administrative responsibilities.

    The proficiency testing program must--
    (a)(1) Provide HHS or its designees and participating laboratories 
with an electronic or a hard copy, or both, of reports of proficiency 
testing results and all scores for each laboratory's performance in a 
format as required by and approved by HCFA for each CLIA-certified 
specialty, subspecialty, and analyte or test within 60 days after the 
date by which the laboratory must report proficiency testing results to 
the proficiency testing program.
    (2) Provide HHS with reports of PT results and scores of individual 
performance in cytology and provide copies of reports to participating 
individuals, and to all laboratories that employ the individuals, within 
15 working days of the testing event;
    (b) Furnish to HHS cumulative reports on an individual laboratory's 
performance and aggregate data on CLIA-certified laboratories for the 
purpose of establishing a system to make the proficiency testing 
program's results available, on a reasonable basis, upon request of any 
person, and include such explanatory information as may be appropriate 
to assist in the interpretation of the proficiency testing program's 
results;
    (c) Provide HHS with additional information and data upon request 
and submit such information necessary for

[[Page 865]]

HHS to conduct an annual evaluation to determine whether the proficiency 
testing program continues to meet the requirements of Secs. 493.901 
through 493.959;
    (d) Maintain records of laboratories' performance for a period of 
five years or such time as may be necessary for any legal proceedings; 
and
    (e) Provide HHS with an annual report and, if needed, an interim 
report which identifies any previously unrecognized sources of 
variability in kits, instruments, methods, or PT samples, which 
adversely affect the programs' ability to evaluate laboratory 
performance.

[57 FR 7151, Feb. 28, 1992, as amended at 58 FR 5228, Jan. 19, 1993]

Sec. 493.905  Nonapproved proficiency testing programs.

    If a proficiency testing program is determined by HHS to fail to 
meet any criteria contained in Secs. 493.901 through 493.959 for 
approval of the proficiency testing program, HCFA will notify the 
program and the program must notify all laboratories enrolled of the 
nonapproval and the reasons for nonapproval within 30 days of the 
notification.

       Proficiency Testing Programs by Specialty and Subspecialty

Sec. 493.909  Microbiology.

    The subspecialties under the specialty of microbiology for which a 
program may offer proficiency testing are bacteriology, 
mycobacteriology, mycology, parasitology and virology. Specific criteria 
for these subspecialties are found at Secs. 493.911 through 493.919.

Sec. 493.911  Bacteriology.

    (a) Types of services offered by laboratories. In bacteriology, for 
proficiency testing purposes, there are five types of laboratories:
    (1) Those that interpret Gram stains or perform primary inoculation, 
or both; and refer cultures to another laboratory appropriately 
certified for the subspecialty of bacteriology for identification;
    (2) Those that use direct antigen techniques to detect an organism 
and may also interpret Gram stains or perform primary inoculation, or 
perform any combination of these;
    (3) Those that, in addition to interpreting Gram stains, performing 
primary inoculations, and using direct antigen tests, also isolate and 
identify aerobic bacteria from throat, urine, cervical, or urethral 
discharge specimens to the genus level and may also perform 
antimicrobial susceptibility tests on selected isolated microorganisms;
    (4) Those that perform the services in paragraph (a)(3) of this 
section and also isolate and identify aerobic bacteria from any source 
to the species level and may also perform antimicrobial susceptibility 
tests; and
    (5) Those that perform the services in paragraph (a)(4) of this 
section and also isolate and identify anaerobic bacteria from any 
source.
    (b) Program content and frequency of challenge. To be approved for 
proficiency testing for bacteriology, the annual program must provide a 
minimum of five samples per testing event. There must be at least three 
testing events at approximately equal intervals per year. The samples 
may be provided to the laboratory through mailed shipments or, at HHS' 
option, may be provided to HHS or its designee for on-site testing. For 
the types of laboratories specified in paragraph (a) of this section, an 
annual program must include samples that contain organisms that are 
representative of the six major groups of bacteria: anaerobes, 
Enterobacteriaceae, gram-positive bacilli, gram-positive cocci, gram-
negative cocci, and miscellaneous gram-negative bacteria, as 
appropriate. The specific organisms included in the samples may vary 
from year to year. The annual program must include samples for bacterial 
antigen detection, bacterial isolation and identification, Gram stain, 
and antimicrobial susceptibility testing.
    (1) An approved program must furnish HHS with a description of 
samples that it plans to include in its annual program no later than six 
months before each calendar year. At least 50 percent of the samples 
must be mixtures of the principal organism and appropriate normal flora. 
The program must include other important emerging

[[Page 866]]

pathogens (as determined by HHS) and either organisms commonly occurring 
in patient specimens or opportunistic pathogens. The program must 
include the following two types of samples; each type of sample must 
meet the 50 percent mixed culture criterion:
    (i) Samples that require laboratories to report only organisms that 
the testing laboratory considers to be a principal pathogen that is 
clearly responsible for a described illness (excluding immuno-
compromised patients). The program determines the reportable isolates, 
including antimicrobial susceptibility for any designated isolate; and
    (ii) Samples that require laboratories to report all organisms 
present. Samples must contain multiple organisms frequently found in 
specimens such as urine, blood, abscesses, and aspirates where multiple 
isolates are clearly significant or where specimens are derived from 
immuno-compromised patients. The program determines the reportable 
isolates.
    (2) An approved program may vary over time. For example, the types 
of organisms that might be included in an approved program over time 
are--

Anaerobes:
    Bacteroides fragilis group
    Clostridium perfringens
    Peptostreptococcus anaerobius
    Enterobacteriaceae
    Citrobacter freundii
    Enterobacter aerogenes
    Escherichia coli
    Klebsiella pneumoniae
    Proteus mirabilis
    Salmonella typhimurium
    Serratia marcescens
    Shigella sonnei
    Yersinia enterocolitica
Gram-positive bacilli:
    Listeria monocytogenes
    Corynebacterium species CDC Group JK
Gram-positive cocci:
    Staphylococcus aureus
    Streptococcus Group A
    Streptococcus Group B
    Streptococcus Group D (S. bovis and enterococcus)
    Streptococcus pneumoniae
Gram-negative cocci:
    Branhamella catarrhalis
    Neisseria gonorrhoeae
    Neisseria meningitidis
Miscellaneous Gram-negative bacteria:
    Campylobacter jejuni
    Haemophilis influenza, Type B
    Pseudomonas aeruginosa

    (3) For antimicrobial susceptibility testing, the program must 
provide at least one sample per testing event that includes gram-
positive or gram-negative strains that have a predetermined pattern of 
sensitivity or resistance to the common antimicrobial agents.
    (c) Evaluation of a laboratory's performance. HHS approves only 
those programs that assess the accuracy of a laboratory's responses in 
accordance with paragraphs (c) (1) through (7) of this section.
    (1) The program determines staining characteristics to be 
interpreted by Gram stain. The program determines the reportable 
bacteria to be detected by direct antigen techniques or isolation. To 
determine the accuracy of a laboratory's response for Gram stain 
interpretation, direct antigen detection, identification, or 
antimicrobial susceptibility testing, the program must compare the 
laboratory's response for each sample with the response which reflects 
agreement of either 90 percent of ten or more referee laboratories or 90 
percent or more of all participating laboratories.
    (2) To evaluate a laboratory's response for a particular sample, the 
program must determine a laboratory's type of service in accordance with 
paragraph (a) of this section. A laboratory must isolate and identify 
the organisms to the same extent it performs these procedures on patient 
specimens. A laboratory's performance will be evaluated on the basis of 
its final answer, for example, a laboratory specified in paragraph 
(a)(3) of this section will be evaluated on the basis of the average of 
its scores for paragraphs (c)(3) through (c)(6) as determined in 
paragraph (c)(7) of this section.
    (3) Since laboratories may incorrectly report the presence of 
organisms in addition to the correctly identified principal organism(s), 
the grading system must provide a means of deducting credit for 
additional erroneous organisms that are reported. Therefore, the total 
number of correct responses for organism isolation and identification 
submitted by the laboratory divided by the number of organisms present 
plus the number of incorrect organisms reported by the laboratory must 
be multiplied by 100 to establish a score for

[[Page 867]]

each sample in each testing event. For example, if a sample contained 
one principal organism and the laboratory reported it correctly but 
reported the presence of an additional organism, which was not 
considered reportable, the sample grade would be 1/(1+1) x 100=50 
percent.
    (4) For antimicrobial susceptibility testing, a laboratory must 
indicate which drugs are routinely included in its test panel when 
testing patient samples. A laboratory's performance will be evaluated 
for only those antibiotics for which service is offered. A correct 
response for each antibiotic will be determined as described in 
Secs. 493.911(c) (1) using criteria such as the guidelines established 
by the National Committee for Clinical Laboratory Standards. Grading is 
based on the number of correct susceptibility responses reported by the 
laboratory divided by the actual number of correct susceptibility 
responses determined by the program, multiplied by 100. For example, if 
a laboratory offers susceptibility testing for Enterobacteriaceae using 
amikacin, cephalothin, and tobramycin, and the organism in the 
proficiency testing sample is an Enterobacteriaceae, and the laboratory 
reports correct responses for two of three antimicrobial agents, the 
laboratory's grade would be 2/3 x 100=67 percent.
    (5) The performance criterion for qualitative antigen tests is the 
presence or absence of the bacterial antigen. The score for antigen 
tests is the number of correct responses divided by the number of 
samples to be tested for the antigen, multiplied by 100.
    (6) The performance criteria for Gram stain is staining reaction, 
i.e., gram positive or gram negative. The score for Gram stain is the 
number of correct responses divided by the number of challenges to be 
tested, multiplied by 100.
    (7) The score for a testing event in bacteriology is the average of 
the scores determined under paragraphs (c)(3) through (c)(6) of this 
section kbased on the type of service offered by the laboratory.

[57 FR 7151, Feb. 28, 1992, as amended at 58 FR 5228, Jan. 19, 1993]

Sec. 493.913  Mycobacteriology.

    (a) Types of services offered by laboratories. In mycobacteriology, 
there are five types of laboratories for proficiency testing purposes:
    (1) Those that interpret acid-fast stains and refer specimen to 
another laboratory appropriately certified in the subspecialty of 
mycobacteriology;
    (2) Those that interpret acid-fast stains, perform primary 
inoculation, and refer cultures to another laboratory appropriately 
certified in the subspecialty of mycobacteriology for identification;
    (3) Those that interpret acid-fast stains, isolate and perform 
identification and/or antimycobacterial susceptibility of Mycobacterium 
tuberculosis, but refer other mycobacteria species to another laboratory 
appropriately certified in the subspecialty of mycobacteriology for 
identification and/or susceptibility tests;
    (4) Those that interpret acid-fast stains, isolate and identify all 
mycobacteria to the extent required for correct clinical diagnosis, but 
refer antimycobacterial susceptibility tests to another laboratory 
appropriately certified in the subspecialty of mycobacteriology; and
    (5) Those that interpret acid-fast stains, isolate and identify all 
mycobacteria to the extent required for correct clinical diagnosis, and 
perform antimycobacterial susceptibility tests on the organisms 
isolated.
    (b) Program content and frequency of challenge. To be approved for 
proficiency testing for mycobacteriology, the annual program must 
provide a minimum of five samples per testing event. There must be at 
least two testing events per year. The samples may be provided through 
mailed shipments or, at HHS' option, provided to HHS or its designee for 
on-site testing events. For types of laboratories specified in 
paragraphs (a)(1) and (a) (3) through (5) of this section, an annual 
program must include samples that contain species that are 
representative of the 5 major groups (complexes) of mycobacteria 
encountered in human specimens. The specific mycobacteria included in 
the samples may vary from year to year.

[[Page 868]]

    (1) An approved program must furnish HHS and its agents with a 
description of samples that it plans to include in its annual program no 
later than six months before each calendar year. At least 50 percent of 
the samples must be mixtures of the principal mycobacteria and 
appropriate normal flora. The program must include mycobacteria commonly 
occurring in patient specimens and other important emerging mycobacteria 
(as determined by HHS). The program determines the reportable isolates 
and correct responses for antimycobacterial susceptibility for any 
designated isolate.
    (2) An approved program may vary over time. For example, the types 
of mycobacteria that might be included in an approved program over time 
are--

TB
    Mycobacterium tuberculosis
    Mycobacterium bovis
Group I
    Mycobacterium kansasii
Group II
    Mycobacterium szulgai
Group III
    Mycobacterium avium-intracellulare
    Mycobacterium terrae
Group IV
    Mycobacterium fortuitum

    (3) For antimycobacterial susceptibility testing, the program must 
provide at least one sample per testing event that includes 
mycobacterium tuberculosis that has a predetermined pattern of 
sensitivity or resistance to the common antimycobacterial agents.
    (4) For laboratories specified in paragraphs (a)(1) and (a)(2), the 
program must provide at least five samples per testing event that 
includes challenges that are acid-fast and challenges which do not 
contain acid-fast organisms.
    (c) Evaluation of a laboratory's performance. HHS approves only 
those programs that assess the accuracy of a laboratory's response in 
accordance with paragraphs (c)(1) through (6) of this section.
    (1) The program determines the reportable mycobacteria to be 
detected by acid-fast stain, for isolation and identification, and for 
antimycobacterial susceptibility. To determine the accuracy of a 
laboratory's response, the program must compare the laboratory's 
response for each sample with the response that reflects agreement of 
either 90 percent of ten or more referee laboratories or 90 percent or 
more of all participating laboratories.
    (2) To evaluate a laboratory's response for a particular sample, the 
program must determine a laboratory's type of service in accordance with 
paragraph (a) of this section. A laboratory must interpret acid-fast 
stains and isolate and identify the organisms to the same extent it 
performs these procedures on patient specimens. A laboratory's 
performance will be evaluated on the basis of the average of its scores 
as determined in paragraph (c)(6) of this section.
    (3) Since laboratories may incorrectly report the presence of 
organisms in addition to the correctly identified principal organism(s), 
the grading system must provide a means of deducting credit for 
additional erroneous organisms reported. Therefore, the total number of 
correct responses submitted by the laboratory divided by the number of 
organisms present plus the number of incorrect organisms reported by the 
laboratory must be multiplied by 100 to establish a score for each 
sample in each testing event. For example, if a sample contained one 
principal organism and the laboratory reported it correctly but reported 
the presence of an additional organism, which was not present, the 
sample grade would be

1/(1+1) x 100=50 percent
    (4) For antimycobacterial susceptibility testing, a laboratory must 
indicate which drugs are routinely included in its test panel when 
testing patient samples. A laboratory's performance will be evaluated 
for only those antibiotics for which susceptibility testing is routinely 
performed on patient specimens. A correct response for each antibiotic 
will be determined as described in Sec. 493.913(c)(1). Grading is based 
on the number of correct susceptibility responses reported by the 
laboratory divided by the actual number of correct susceptibility 
responses as determined by the program, multiplied by 100. For example, 
if a laboratory offers susceptibility testing using three 
antimycobacterial agents and the laboratory reports correct response for

[[Page 869]]

two of the three antimycobacterial agents, the laboratory's grade would 
be \2/3\ x 100=67 percent.
    (5) The performance criterion for qualitative tests is the presence 
or absence of acid-fast organisms. The score for acid-fast organism 
detection is the number of correct responses divided by the number of 
samples to be tested, multiplied by 100.
    (6) The score for a testing event in mycobacteriology is the average 
of the scores determined under paragraphs (c)(3) through (c)(5) of this 
section based on the type of service offered by the laboratory.

[57 FR 7151, Feb. 28, 1992, as amended at 58 FR 5228, Jan. 19, 1993]

Sec. 493.915  Mycology.

    (a) Types of services offered by laboratories. In mycology, there 
are four types of laboratories for proficiency testing purposes that may 
perform different levels of service for yeasts, dimorphic fungi, 
dermatophytes, and aerobic actinomycetes:
    (1) Those that isolate and identify only yeasts and/or dermatophytes 
to the genus level;
    (2) Those that isolate and identify yeasts and/or dermatophytes to 
the species level;
    (3) Those that isolate and perform identification of all organisms 
to the genus level; and
    (4) Those that isolate and perform identification of all organisms 
to the species level.
    (b) Program content and frequency of challenge. To be approved for 
proficiency testing for mycology, the annual program must provide a 
minimum of five samples per testing event. There must be at least three 
testing events at approximately equal intervals per year. The samples 
may be provided through mailed shipments or, at HHS' option, may be 
provided to HHS or its designee for on-site testing. An annual program 
must include samples that contain organisms that are representative of 
five major groups of fungi: Yeast or yeast-like fungi; dimorphic fungi; 
dematiaceous fungi; dermatophytes; and saprophytes, including 
opportunistic fungi. The specific fungi included in the samples may vary 
from year to year.
    (1) An approved program must, before each calendar year, furnish HHS 
with a description of samples that it plans to include in its annual 
program no later than six months before each calendar year. At least 50 
percent of the samples must be mixtures of the principal organism and 
appropriate normal background flora. Other important emerging pathogens 
(as determined by HHS) and organisms commonly occurring in patient 
specimens must be included periodically in the program.
    (2) An approved program may vary over time. As an example, the types 
of organisms that might be included in an approved program over time 
are--

Candida albicans
Candida (other species)
Cryptococcus neoformans
Sporothrix schenckii
Exophiala jeanselmei
Fonsecaea pedrosoi
Microsporum sp.
Acremonium sp.
Trichophvton sp.
Aspergillus fumigatus
Nocardia sp.
Blastomyces dermatitidis \1\
Zygomycetes sp.
    Note: \1\ Provided as a nonviable sample.

    (c) Evaluation of a laboratory's performance. HHS approves only 
those programs that assess the accuracy of a laboratory's response, in 
accordance with paragraphs (c)(1) through (5) of this section.
    (1) The program determines the reportable organisms. To determine 
the accuracy of a laboratory's response, the program must compare the 
laboratory's response for each sample with the response that reflects 
agreement of either 90 percent of ten or more referee laboratories or 90 
percent or more of all participating laboratories.
    (2) To evaluate a laboratory's response for a particular sample, the 
program must determine a laboratory's type of service in accordance with 
paragraph (a) of this section. A laboratory must isolate and identify 
the organisms to the same extent it performs these procedures on patient 
specimens.
    (3) Since laboratories may incorrectly report the presence of 
organisms in addition to the correctly identified principal organism(s), 
the grading system must deduct credit for additional

[[Page 870]]

erroneous organisms reported. Therefore, the total number of correct 
responses submitted by the laboratory divided by the number of organisms 
present plus the number of incorrect organisms reported by the 
laboratory must be multiplied by 100 to establish a score for each 
sample in each shipment or testing event. For example, if a sample 
contained one principal organism and the laboratory reported it 
correctly but reported the presence of an additional organism, which was 
not present, the sample grade would be 1/(1+1)x100=50 percent.
    (4) The score for the antigen tests is the number of correct 
responses divided by the number of samples to be tested for the antigen, 
multiplied by 100.
    (5) The score for a testing event is the average of the sample 
scores as determined under paragraph (c)(3) or (c)(4), or both, of this 
section.

[57 FR 7151, Feb. 28, 1992, as amended at 58 FR 5228, Jan. 19, 1993]

Sec. 493.917  Parasitology.

    (a) Types of services offered by laboratories. In parasitology there 
are two types of laboratories for proficiency testing purposes--
    (1) Those that determine the presence or absence of parasites by 
direct observation (wet mount) and/or pinworm preparations and, if 
necessary, refer specimens to another laboratory appropriately certified 
in the subspecialty of parasitology for identification;
    (2) Those that identify parasites using concentration preparations 
and/or permanent stains.
    (b) Program content and frequency of challenge. To be approved for 
proficiency testing in parasitology, a program must provide a minimum of 
five samples per testing event. There must be at least three testing 
events at approximately equal intervals per year. The samples may be 
provided through mailed shipments or, at HHS's option, may be provided 
to HHS or its designee for on-site testing. An annual program must 
include samples that contain parasites that are commonly encountered in 
the United States as well as those recently introduced into the United 
States. Other important emerging pathogens (as determined by HHS) and 
parasites commonly occurring in patient specimens must be included 
periodically in the program.
    (1) An approved program must, before each calendar year furnish HHS 
with a description of samples that it plans to include in its annual 
program no later than six months before each calendar year. Samples must 
include both formalinized specimens and PVA (polyvinyl alcohol) fixed 
specimens as well as blood smears, as appropriate for a particular 
parasite and stage of the parasite. The majority of samples must contain 
protozoa or helminths or a combination of parasites. Some samples must 
be devoid of parasites.
    (2) An approved program may vary over time. As an example, the types 
of parasites that might be included in an approved program over time 
are--

Enterobius vermicularis
Entamoeba histolytica
Entamoeba coli
Giardia lamblia
Endolimax nana
Dientamoeba fragilis
Iodamoeba butschli
Chilomastix mesnili
Hookworm
Ascaris lumbricoides
Strongyloides stercoralis
Trichuris trichiura
Diphyllobothrium latum
Cryptosporidium sp.
Plasmodium falciparum

    (3) For laboratories specified in paragraph (a)(1) of this section, 
the program must provide at least five samples per testing event that 
include challenges which contain parasites and challenges that are 
devoid of parasites.
    (c) Evaluation of a laboratory's performance. HHS approves only 
those programs that assess the accuracy of a laboratory's responses in 
accordance with paragraphs (c)(1) through (6) of this section.
    (1) The program must determine the reportable parasites. It may 
elect to establish a minimum number of parasites to be identified in 
samples before they are reported. Parasites found in rare numbers by 
referee laboratories are not considered in scoring a laboratory's 
performance; such findings are neutral. To determine the accuracy of a 
laboratory's response, the program must compare the laboratory's 
response with the response that reflects agreement of

[[Page 871]]

either 90 percent of ten or more referee laboratories or 90 percent or 
more of all participating laboratories.
    (2) To evaluate a laboratory's response for a particular sample, the 
program must determine a laboratory's type of service in accordance with 
paragraph (a) of this section. A laboratory must determine the presence 
or absence of a parasite(s) or concentrate and identify the parasites to 
the same extent it performs these procedures on patient specimens.
    (3) Since laboratories may incorrectly report the presence of 
parasites in addition to the correctly identified principal parasite(s), 
the grading system must deduct credit for these additional erroneous 
parasites reported and not found in rare numbers by the program's 
referencing process. Therefore, the total number of correct responses 
submitted by the laboratory divided by the number of parasites present 
plus the number of incorrect parasites reported by the laboratory must 
be multiplied by 100 to establish a score for each sample in each 
testing event. For example, if a sample contained one principal parasite 
and the laboratory reported it correctly but reported the presence of an 
additional parasite, which was not present, the sample grade would be

1/(1+1) x 100=50 percent.
    (4) The criterion for acceptable performance for qualitative 
parasitology examinations is presence or absence of a parasite(s).
    (5) The score for parasitology is the number of correct responses 
divided by the number of samples to be tested, multiplied by 100.
    (6) The score for a testing event is the average of the sample 
scores as determined under paragraphs (c)(3) through (c)(5) of this 
section.

Sec. 493.919  Virology.

    (a) Types of services offered by laboratories. In virology, there 
are two types of laboratories for proficiency testing purposes--
    (1) Those that only perform tests that directly detect viral 
antigens or structures, either in cells derived from infected tissues or 
free in fluid specimens; and
    (2) Those that are able to isolate and identify viruses and use 
direct antigen techniques.
    (b) Program content and frequency of challenge. To be approved for 
proficiency testing in virology, a program must provide a minimum of 
five samples per testing event. There must be at least three testing 
events at approximately equal intervals per year. The samples may be 
provided to the laboratory through mailed shipments or, at HHS's option, 
may be provided to HHS or its designee for on-site testing. An annual 
program must include viral species that are the more commonly identified 
viruses. The specific organisms found in the samples may vary from year 
to year. The annual program must include samples for viral antigen 
detection and viral isolation and identification.
    (1) An approved program must furnish HHS with a description of 
samples that it plans to include in its annual program no later than six 
months before each calendar year. The program must include other 
important emerging viruses (as determined by HHS) and viruses commonly 
occurring in patient specimens.
    (2) An approved program may vary over time. For example, the types 
of viruses that might be included in an approved program over time are 
the more commonly identified viruses such as Herpes simplex, respiratory 
syncytial virus, adenoviruses, enteroviruses, and cytomegaloviruses.
    (c) Evaluation of laboratory's performance. HHS approves only those 
programs that assess the accuracy of a laboratory's response in 
accordance with paragraphs (c)(1) through (5) of this section.
    (1) The program determines the reportable viruses to be detected by 
direct antigen techniques or isolated by laboratories that perform viral 
isolation procedures. To determine the accuracy of a laboratory's 
response, the program must compare the laboratory's response for each 
sample with the response that reflects agreement of either 90 percent of 
ten or more referee laboratories or 90 percent or more of all 
participating laboratories.

[[Page 872]]

    (2) To evaluate a laboratory's response for a particular sample, the 
program must determine a laboratory's type of service in accordance with 
paragraph (a) of this section. A laboratory must isolate and identify 
the viruses to the same extent it performs these procedures on patient 
specimens.
    (3) Since laboratories may incorrectly report the presence of 
viruses in addition to the correctly identified principal virus, the 
grading system must provide a means of deducting credit for additional 
erroneous viruses reported. Therefore, the total number of correct 
responses determined by virus culture techniques submitted by the 
laboratory divided by the number of viruses present plus the number of 
incorrect viruses reported by the laboratory must be multiplied by 100 
to establish a score for each sample in each testing event. For example, 
if a sample contained one principal virus and the laboratory reported it 
correctly but reported the presence of an additional virus, which was 
not present, the sample grade would be 1/(1+1) x 100=50 percent.
    (4) The performance criterion for qualitative antigen tests is 
presence or absence of the viral antigen. The score for the antigen 
tests is the number of correct responses divided by the number of 
samples to be tested for the antigen, multiplied by 100.
    (5) The score for a testing event is the average of the sample 
scores as determined under paragraph (c)(3) and (c)(4) of this section.

Sec. 493.921  Diagnostic immunology.

    The subspecialties under the specialty of immunology for which a 
program may offer proficiency testing are syphilis serology and general 
immunology. Specific criteria for these subspecialties are found at 
Secs. 493.923 and 493.927.

Sec. 493.923  Syphilis serology.

    (a) Program content and frequency of challenge. To be approved for 
proficiency testing in syphilis serology, a program must provide a 
minimum of five samples per testing event. There must be at least three 
testing events at approximately equal intervals per year. The samples 
may be provided through mailed shipments or, at HHS' option, may be 
provided to HHS or its designee for on-site testing. An annual program 
must include samples that cover the full range of reactivity from highly 
reactive to non-reactive.
    (b) Evaluation of test performance. HHS approves only those programs 
that assess the accuracy of a laboratory's responses in accordance with 
paragraphs (b)(1) through (4) of this section.
    (1) To determine the accuracy of a laboratory's response for 
qualitative and quantitative syphilis tests, the program must compare 
the laboratory's response with the response that reflects agreement of 
either 90 percent of ten or more referee laboratories or 90 percent or 
more of all participating laboratories. The proficiency testing program 
must indicate the minimum concentration, by method, that will be 
considered as indicating a positive response. The score for a sample in 
syphilis serology is the average of scores determined under paragraphs 
(b)(2) and (b)(3) of this section.
    (2) For quantitative syphilis tests, the program must determine the 
correct response for each method by the distance of the response from 
the target value. After the target value has been established for each 
response, the appropriateness of the response must be determined by 
using fixed criteria. The criterion for acceptable performance for 
quantitative syphilis serology tests is the target value <SUP>plus-minus</SUP> 
1 dilution.
    (3) The criterion for acceptable performance for qualitative 
syphilis serology tests is reactive or nonreactive.
    (4) To determine the overall testing event score, the number of 
correct responses must be averaged using the following formula:


  Number of acceptable responses for all
                challenges
-------------------------------------------   x 100=Testing event score
      Total number of all challenges
------------------------------------------------------------------------



[57 FR 7151, Feb. 28, 1992, as amended at 58 FR 5229, Jan. 19, 1993]

[[Page 873]]

Sec. 493.927  General immunology.

    (a) Program content and frequency of challenge. To be approved for 
proficiency testing for immunology, the annual program must provide a 
minimum of five samples per testing event. There must be at least three 
testing events at approximately equal intervals per year. The annual 
program must provide samples that cover the full range of reactivity 
from highly reactive to nonreactive. The samples may be provided through 
mailed shipments or, at HHS' option, may be provided to HHS or its 
designee for on-site testing.
    (b) Challenges per testing event. The minimum number of challenges 
per testing event the program must provide for each analyte or test 
procedure is five. Analytes or tests for which laboratory performance is 
to be evaluated include:

                        Analyte or Test Procedure

Alpha-l antitrypsin
Alpha-fetoprotein (tumor marker)
Antinuclear antibody
Antistreptolysin O
Anti-human immunodeficiency virus (HIV)
Complement C3
Complement C4
Hepatitis markers (HBsAg, anti-HBc, HBeAg)
IgA
IgG
IgE
IgM
Infectious mononucleosis
Rheumatoid factor
Rubella

    (c) Evaluation of a laboratory's analyte or test performance. HHS 
approves only those programs that assess the accuracy of a laboratory's 
responses in accordance with paragraphs (c)(1) through (5) of this 
section.
    (1) To determine the accuracy of a laboratory's response for 
quantitative and qualitative immunology tests or analytes, the program 
must compare the laboratory's response for each analyte with the 
response that reflects agreement of either 90 percent of ten or more 
referee laboratories or 90 percent or more of all participating 
laboratories. The proficiency testing program must indicate the minimum 
concentration that will be considered as indicating a positive response. 
The score for a sample in general immunology is either the score 
determined under paragraph (c)(2) or (3) of this section.
    (2) For quantitative immunology analytes or tests, the program must 
determine the correct response for each analyte by the distance of the 
response from the target value. After the target value has been 
established for each response, the appropriateness of the response must 
be determined by using either fixed criteria or the number of standard 
deviations (SDs) the response differs from the target value.

                   Criteria for Acceptable Performance

    The criteria for acceptable performance are--

------------------------------------------------------------------------
                                               Criteria for acceptable
              Analyte or test                        performance
------------------------------------------------------------------------
Alpha-1 antitrypsin.......................  Target value <plus-minus>3
                                             SD.
Alpha-fetoprotein (tumor marker)..........  Target value <plus-minus>3
                                             SD.
Antinuclear antibody......................  Target value +/-2 dilutions
                                             or positive or negative.
Antistreptolysin O........................  Target value +/-2 dilution
                                             or positive or negative.
Anti-Human Immunodeficiency virus.........  Reactive or nonreactive.
Complement C3.............................  Target value <plus-minus>3
                                             SD.
Complement C4.............................  Target value <plus-minus>3
                                             SD.
Hepatitis (HBsAg, anti-HBc, HBeAg)........  Reactive (positive) or
                                             nonreactive (negative).
IgA.......................................  Target value <plus-minus>3
                                             SD.
IgE.......................................  Target value <plus-minus>3
                                             SD.
IgG.......................................  Target value +/-25%.
IgM.......................................  Target value <plus-minus>3
                                             SD.
Infectious mononucleosis..................  Target value +/-2 dilutions
                                             or positive or negative.
Rheumatoid factor.........................  Target value +/-2 dilutions
                                             or positive or negative.
Rubella...................................  Target value +/-2 dilutions
                                             or immune or nonimmune or
                                             positive or negative.
------------------------------------------------------------------------

    (3) The criterion for acceptable performance for qualitative general 
immunology tests is positive or negative.
    (4) To determine the analyte testing event score, the number of 
acceptable analyte responses must be averaged using the following 
formula:


  Number of acceptable responses for the
                  analyte                    x 100=Analyte score for the
-------------------------------------------         testing event
Total number of challenges for the analyte
------------------------------------------------------------------------



[[Page 874]]

    (5) To determine the overall testing event score, the number of 
correct responses for all analytes must be averaged using the following 
formula:


  Number of acceptable responses for all
                challenges
-------------------------------------------   x 100=Testing event score
      Total number of all challenges
------------------------------------------------------------------------



[57 FR 7151, Feb. 28, 1992, as amended at 58 FR 5229, Jan. 19, 1993]

Sec. 493.929  Chemistry.

    The subspecialties under the specialty of chemistry for which a 
proficiency testing program may offer proficiency testing are routine 
chemistry, endocrinology, and toxicology. Specific criteria for these 
subspecialties are listed in Secs. 493.931 through 493.939.

Sec. 493.931  Routine chemistry.

    (a) Program content and frequency of challenge. To be approved for 
proficiency testing for routine chemistry, a program must provide a 
minimum of five samples per testing event. There must be at least three 
testing events at approximately equal intervals per year. The annual 
program must provide samples that cover the clinically relevant range of 
values that would be expected in patient specimens. The specimens may be 
provided through mailed shipments or, at HHS' option, may be provided to 
HHS or its designee for on-site testing.
    (b) Challenges per testing event. The minimum number of challenges 
per testing event a program must provide for each analyte or test 
procedure listed below is five serum, plasma or blood samples.

                        Analyte or Test Procedure

Alanine aminotransferase (ALT/SGPT)
Albumin
Alkaline phosphatase
Amylase
Aspartate aminotransferase (AST/SGOT)
Bilirubin, total
Blood gas (pH, pO2, and pCO2)
Calcium, total
Chloride
Cholesterol, total
Cholesterol, high density lipoprotein
Creatine kinase
Creatine kinase, isoenzymes
Creatinine
Glucose (Excluding measurements on devices cleared by FDA for home use)
Iron, total
Lactate dehydrogenase (LDH)
LDH isoenzymes
Magnesium
Potassium
Sodium
Total Protein
Triglycerides
Urea Nitrogen
Uric Acid

    (c) Evaluation of a laboratory's analyte or test performance. HHS 
approves only those programs that assess the accuracy of a laboratory's 
responses in accordance with paragraphs (c)(1) through (5) of this 
section.
    (1) To determine the accuracy of a laboratory's response for 
qualitative and quantitative chemistry tests or analytes, the program 
must compare the laboratory's response for each analyte with the 
response that reflects agreement of either 90 percent of ten or more 
referee laboratories or 90 percent or more of all participating 
laboratories. The score for a sample in routine chemistry is either the 
score determined under paragraph (c)(2) or (3) of this section.
    (2) For quantitative chemistry tests or analytes, the program must 
determine the correct response for each analyte by the distance of the 
response from the target value. After the target value has been 
established for each response, the appropriateness of the response must 
be determined by using either fixed criteria based on the percentage 
difference from the target value or the number of standard deviations 
(SDs) the response differs from the target value.

                   Criteria for Acceptable Performance

    The criteria for acceptable performance are--

------------------------------------------------------------------------
                                               Criteria for acceptable
              Analyte or test                        performance
------------------------------------------------------------------------
Alanine aminotransferase (ALT/SGPT).......  Target value <plus-
                                             minus>20%.
Albumin...................................  Target value <plus-
                                             minus>10%.
Alkaline phosphatase......................  Target value <plus-
                                             minus>30%.
Amylase...................................  Target value <plus-
                                             minus>30%.
Aspartate aminotransferase (AST/SGOT).....  Target value <plus-
                                             minus>20%.
Bilirubin, total..........................  Target value <plus-minus>0.4
                                             mg/dL or <plus-minus>20%
                                             (greater).

[[Page 875]]


Blood gas pO2.............................  Target value <plus-minus>3
                                             SD.
pCO2......................................  Target value <plus-minus>5
                                             mm Hg or +/-8% (greater).
pH........................................  Target value <plus-
                                             minus>0.04.
Calcium, total............................  Target value <plus-minus>1.0
                                             mg/dL.
Chloride..................................  Target value <plus-minus>5%.
Cholesterol, total........................  Target value <plus-
                                             minus>10%.
Cholesterol, high density lipoprotein.....  Target value <plus-
                                             minus>30%.
Creatine kinase...........................  Target value <plus-
                                             minus>30%.
Creatine kinase isoenzymes................  MB elevated (presence or
                                             absence) or Target value
                                             <plus-minus>3SD.
Creatinine................................  Target value <plus-minus>0.3
                                             mg/dL or <plus-minus>15%
                                             (greater).
Glucose (excluding glucose performed on     Target value <plus-minus>6
 monitoring devices cleared by FDA for       mg/dl or <plus-minus>10%
 home use.                                   (greater).
Iron, total...............................  Target value <plus-
                                             minus>20%.
Lactate dehydrogenase (LDH)...............  Target value <plus-
                                             minus>20%.
LDH isoenzymes............................  LDH1/LDH2 (+ or -) or Target
                                             value <plus-minus> 30%.
Magnesium.................................  Target value <plus-
                                             minus>25%.
Potassium.................................  Target value <plus-minus>0.5
                                             mmol/L.
Sodium....................................  Target value <plus-minus>4
                                             mmol/L.
Total Protein.............................  Target value <plus-
                                             minus>10%.
Triglycerides.............................  Target value <plus-
                                             minus>25%.
Urea nitrogen.............................  Target value <plus-minus>2
                                             mg/dL or <plus-minus>9%
                                             (greater).
Uric acid.................................  Target value <plus-
                                             minus>17%.
------------------------------------------------------------------------

    (3) The criterion for acceptable performance for qualitative routine 
chemistry tests is positive or negative.
    (4) To determine the analyte testing event score, the number of 
acceptable analyte responses must be averaged using the following 
formula:


  Number of acceptable responses for the
                  analyte                    x 100=Analyte score for the
-------------------------------------------         testing event
Total number of challenges for the analyte
------------------------------------------------------------------------


    (5) To determine the overall testing event score, the number of 
correct responses for all analytes must be averaged using the following 
formula:


  Number of acceptable responses for all
                challenges
-------------------------------------------   x 100=Testing event score
      Total number of all challenges
------------------------------------------------------------------------


Sec. 493.933  Endocrinology.

    (a) Program content and frequency of challenge. To be approved for 
proficiency testing for endocrinology, a program must provide a minimum 
of five samples per testing event. There must be at least three testing 
events at approximately equal intervals per year. The annual program 
must provide samples that cover the clinically relevant range of values 
that would be expected in patient specimens. The samples may be provided 
through mailed shipments or, at HHS' option, may be provided to HHS or 
its designee for on-site testing.
    (b) Challenges per testing event. The minimum number of challenges 
per testing event a program must provide for each analyte or test 
procedure is five serum, plasma, blood, or urine samples.

Analyte or Test
Cortisol
Free Thyroxine
Human Chorionic gonadotropin (excluding urine pregnancy tests done by 
visual color comparison categorized as waived tests)
T3 Uptake
Triiodothyronine
Thyroid-stimulating hormone
Thyroxine

    (c) Evaluation of a laboratory's analyte or test performance. HHS 
approves only those programs that assess the accuracy of a laboratory's 
responses in accordance with paragraphs (c)(1) through (5) of this 
section.
    (1) To determine the accuracy of a laboratory's response for 
qualitative and quantitative endocrinology tests or analytes, a program 
must compare the laboratory's response for each analyte with the 
response that reflects agreement of either 90 percent of ten or more 
referee laboratories or 90 percent or more of all participating 
laboratories. The score for a sample in endocrinology is either the 
score determined under paragraph (c)(2) or (c)(3) of this section.
    (2) For quantitative endocrinology tests or analytes, the program 
must determine the correct response for each analyte by the distance of 
the response from the target value. After the target value has been 
established for each response, the appropriateness of the response must 
be determined by using either fixed criteria based on the percentage 
difference from the target value or the number of standard deviations 
(SDs) the response differs from the target value.


[[Page 876]]



                   Criteria for Acceptable Performance

    The criteria for acceptable performance are--

------------------------------------------------------------------------
                                               Criteria for acceptable
              Analyte or test                        performance
------------------------------------------------------------------------
Cortisol..................................  Target value +/-25%.
Free Thyroxine............................  Target value +/-3 SD.
Human Chorionic Gonadotropin (excluding     Target value +/-3 SD
 urine pregnancy tests done by visual        positive or negative.
 color comparison categorized as waived
 tests).
T3 Uptake.................................  Target value +/-3 SD.
Triiodothyronine..........................  Target value +/-3 SD.
Thyroid-stimulating hormone...............  Target value +/-3 SD.
Thyroxine.................................  Target value +/-20% or 1.0
                                             mcg/dL (greater).
------------------------------------------------------------------------

    (3) The criterion for acceptable performance for qualitative 
endocrinology tests is positive or negative.
    (4) To determine the analyte testing event score, the number of 
acceptable analyte responses must be averaged using the following 
formula:


  Number of acceptable responses for the
                  analyte                    x 100=Analyte score for the
-------------------------------------------         testing event
Total number of challenges for the analyte
------------------------------------------------------------------------


    (5) To determine the overall testing event score, the number of 
correct responses for all analytes must be averaged using the following 
formula:


  Number of acceptable responses for all
                challenges
-------------------------------------------   x 100=Testing event score
      Total number of all challenges
------------------------------------------------------------------------



[57 FR 7151, Feb. 28, 1992, as amended at 58 FR 5229, Jan. 19, 1993]

Sec. 493.937  Toxicology.

    (a) Program content and frequency of challenge. To be approved for 
proficiency testing for toxicology, the annual program must provide a 
minimum of five samples per testing event. There must be at least three 
testing events at approximately equal intervals per year. The annual 
program must provide samples that cover the clinically relevant range of 
values that would be expected in specimens of patients on drug therapy 
and that cover the level of clinical significance for the particular 
drug. The samples may be provided through mailed shipments or, at HHS' 
option, may be provided to HHS or its designee for on-site testing.
    (b) Challenges per testing event. The minimum number of challenges 
per testing event a program must provide for each analyte or test 
procedure is five serum, plasma, or blood samples.

                        Analyte or Test Procedure

Alcohol (blood)
Blood lead
Carbamazepine
Digoxin
Ethosuximide
Gentamicin
Lithium
Phenobarbital
Phenytoin
Primidone
Procainamide
  (and metabolite)
Quinidine
Theophylline
Tobramycin
Valproic Acid

    (c) Evaluation of a laboratory's analyte or test performance. HHS 
approves only those programs that assess the accuracy of a laboratory's 
responses in accordance with paragraphs (c)(1) through (4) of this 
section.
    (1) To determine the accuracy of a laboratory's responses for 
quantitative toxicology tests or analytes, the program must compare the 
laboratory's response for each analyte with the response that reflects 
agreement of either 90 percent of ten or more referee laboratories or 90 
percent or more of all participating laboratories. The score for a 
sample in toxicology is the score determined under paragraph (c)(2) of 
this section.
    (2) For quantitative toxicology tests or analytes, the program must 
determine the correct response for each analyte by the distance of the 
response from the target value. After the target value has been 
established for each response, the appropriateness of the response must 
be determined by using fixed criteria based on the percentage difference 
from the target value

                   Criteria for Acceptable Performance

    The criteria for acceptable performance are:

------------------------------------------------------------------------
                                               Criteria for acceptable
              Analyte or test                        performance
------------------------------------------------------------------------
Alcohol, blood............................  Target Value <plus-minus>
                                             25%.
Blood lead................................  Target Value <plus-minus>10%
                                             or 4 mcg/dL (greater).

[[Page 877]]


Carbamazepine.............................  Target Value <plus-minus>
                                             25%.
Digoxin...................................  Target Value <plus-minus>
                                             20% or <plus-minus> 0.2 ng/
                                             mL (greater).
Ethosuximide..............................  Target Value <plus-minus>
                                             20%.
Gentamicin................................  Target Value <plus-minus>
                                             25%.
Lithium...................................  Target Value <plus-minus>
                                             0.3 mmol/L or <plus-minus>
                                             20% (greater).
Phenobarbital.............................  Target Value <plus-minus>
                                             20%
Phenytoin.................................  Target Value <plus-minus>
                                             25%.
Primidone.................................  Target Value <plus-minus>
                                             25%.
Procainamide (and metabolite).............  Target Value <plus-minus>
                                             25%.
Quinidine.................................  Target Value <plus-minus>
                                             25%.
Tobramycin................................  Target Value <plus-minus>
                                             25%.
Theophylline..............................  Target Value <plus-minus>
                                             25%.
Valproic Acid.............................  Target Value <plus-minus>
                                             25%.
------------------------------------------------------------------------

    (3) To determine the analyte testing event score, the number of 
acceptable analyte responses must be averaged using the following 
formula:


  Number of acceptable responses for the
                  analyte                    x 100=Analyte score for the
-------------------------------------------         testing event
Total number of challenges for the analyte
------------------------------------------------------------------------


    (4) To determine the overall testing event score, the number of 
correct responses for all analytes must be averaged using the following 
formula:


  Number of acceptable responses for all
                challenges
-------------------------------------------   x 100=Testing event score
      Total number of all challenges
------------------------------------------------------------------------



[57 FR 7151, Feb. 28, 1992, as amended at 58 FR 5229, Jan. 19, 1993]

Sec. 493.941  Hematology (including routine hematology and coagulation).

    (a) Program content and frequency of challenge. To be approved for 
proficiency testing for hematology, a program must provide a minimum of 
five samples per testing event. There must be at least three testing 
events at approximately equal intervals per year. The annual program 
must provide samples that cover the full range of values that would be 
expected in patient specimens. The samples may be provided through 
mailed shipments or, at HHS' option, may be provided to HHS and or its 
designee for on-site testing.
    (b) Challenges per testing event. The minimum number of challenges 
per testing event a program must provide for each analyte or test 
procedure is five.

                        Analyte or Test Procedure

Cell identification or white blood cell differential
Erythrocyte count
Hematocrit (excluding spun microhematocrit)
Hemoglobin
Leukocyte count
Platelet count
Fibrinogen
Partial thromboplastin time
Prothrombin time

    (1) An approved program for cell identification may vary over time. 
The types of cells that might be included in an approved program over 
time are--

Neutrophilic granulocytes
Eosinophilic granulocytes
Basophilic granulocytes
Lymphocytes
Monocytes
Major red and white blood cell abnormalities
Immature red and white blood cells

    (2) White blood cell differentials should be limited to the 
percentage distribution of cellular elements listed above.
    (c) Evaluation of a laboratory's analyte or test performance. HHS 
approves only those programs that assess the accuracy of a laboratory's 
responses in accordance with paragraphs (c) (1) through (5) of this 
section.
    (1) To determine the accuracy of a laboratory's responses for 
qualitative and quantitative hematology tests or analytes, the program 
must compare the laboratory's response for each analyte with the 
response that reflects agreement of either 90 percent of ten or more 
referee laboratories or 90 percent or more of all participating 
laboratories. The score for a sample in hematology is either the score 
determined under paragraph (c) (2) or (3) of this section.
    (2) For quantitative hematology tests or analytes, the program must 
determine the correct response for each analyte by the distance of the 
response from the target value. After the target

[[Page 878]]

value has been established for each response, the appropriateness of the 
response is determined using either fixed criteria based on the 
percentage difference from the target value or the number of standard 
deviations (SDs) the response differs from the target value.

                   Criteria for Acceptable Performance

    The criteria for acceptable performance are:

------------------------------------------------------------------------
                                               Criteria for acceptable
              Analyte or test                        performance
------------------------------------------------------------------------
Cell identification.......................  90% or greater consensus on
                                             identification.
White blood cell differential.............  Target +/- 3SD based on the
                                             percentage of different
                                             types of white blood cells
                                             in the samples.
Erythrocyte count.........................  Target +/-6%.
Hematocrit (Excluding spun hematocrits)...  Target +/-6%.
Hemoglobin................................  Target +/-7%.
Leukocyte count...........................  Target +/-15%.
Platelet count............................  Target +/-25%.
Fibrinogen................................  Target +/- 20%.
Partial thromboplastin time...............  Target +/-15%.
Prothrombin time..........................  Target +/-15%.
------------------------------------------------------------------------

    (3) The criterion for acceptable performance for the qualitative 
hematology test is correct cell identification.
    (4) To determine the analyte testing event score, the number of 
acceptable analyte responses must be averaged using the following 
formula:


  Number of acceptable responses for the
                  analyte                       x 100=Analyte score for
-------------------------------------------       the testing event
Total number of challenges for the analyte
------------------------------------------------------------------------


    (5) To determine the overall testing event score, the number of 
correct responses for all analytes must be averaged using the following 
formula:


  Number of acceptable responses for all
                challenges
-------------------------------------------   x 100=Testing event score
      Total number of all challenges
------------------------------------------------------------------------



[57 FR 7151, Feb. 28, 1992, as amended at 58 FR 5229, Jan. 19, 1993]

Sec. 493.945  Cytology; gynecologic examinations.

    (a) Program content and frequency of challenge. (1) To be approved 
for proficiency testing for gynecologic examinations (Pap smears) in 
cytology, a program must provide test sets composed of 10- and 20-glass 
slides. Proficiency testing programs may obtain slides for test sets 
from cytology laboratories, provided the slides have been retained by 
the laboratory for the required period specified in Sec. 493.1257. If 
slide preparations are still subject to retention by the laboratory, 
they may be loaned to a proficiency testing program if the program 
provides the laboratory with documentation of the loan of the slides and 
ensures that slides loaned to it are retrievable upon request. Each test 
set must include at least one slide representing each of the response 
categories described in paragraph (b)(3)(ii)(A) of this section, and 
test sets should be comparable so that equitable testing is achieved 
within and between proficiency testing providers.
    (2) To be approved for proficiency testing in gynecologic cytology, 
a program must provide announced and unannounced on-site testing for 
each individual at least once per year and must provide an initial 
retesting event for each individual within 45 days after notification of 
test failure and subsequent retesting events within 45 days after 
completion of remedial action described in Sec. 493.855.
    (b) Evaluation of an individual's performance. HHS approves only 
those programs that assess the accuracy of each individual's responses 
on both 10- and 20-slide test sets in which the slides have been 
referenced as specified in paragraph (b)(1) of this section.
    (1) To determine the accuracy of an individual's response on a 
particular challenge (slide), the program must compare the individual's 
response for each slide preparation with the response that reflects the 
predetermined consensus agreement or confirmation on the diagnostic 
category, as described in the table in paragraph (b)(3)(ii)(A) of this 
section. For all slide preparations, a 100% consensus agreement among a 
minimum of three physicians certified in anatomic pathology

[[Page 879]]

is required. In addition, for premalignant and malignant slide 
preparations, confirmation by tissue biopsy is required either by 
comparison of the reported biopsy results or reevaluation of biopsy 
slide material by a physician certified in anatomic pathology.
    (2) An individual qualified as a technical supervisor under 
Sec. 493.1449 (b) or (k) who routinely interprets gynecologic slide 
preparations only after they have been examined by a cytotechnologist 
can either be tested using a test set that has been screened by a 
cytotechnologist in the same laboratory or using a test set that has not 
been screened. A technical supervisor who screens and interprets slide 
preparations that have not been previously examined must be tested using 
a test set that has not been previously screened.
    (3) The criteria for acceptable performance are determined by using 
the scoring system in paragraphs (b)(3) (i) and (ii) of this section.
    (i) Each slide set must contain 10 or 20 slides with point values 
established for each slide preparation based on the significance of the 
relationship of the interpretation of the slide to a clinical condition 
and whether the participant in the testing event is a cytotechnologist 
qualified under Secs. 493.1469 or 493.1483 or functioning as a technical 
supervisor in cytology qualified under Sec. 493.1449 (b) or (k) of this 
part.
    (ii) The scoring system rewards or penalizes the participants in 
proportion to the distance of their answers from the correct response or 
target diagnosis and the penalty or reward is weighted in proportion to 
the severity of the lesion.
    (A) The four response categories for reporting proficiency testing 
results and their descriptions are as follows:

------------------------------------------------------------------------
             Category                            Description
------------------------------------------------------------------------
A.................................  Unsatisfactory for diagnosis due to:
                                    (1) Scant cellularity.
                                    (2) Air drying.
                                    (3) Obscuring material (blood,
                                     inflammatory cells, or lubricant).
B.................................  Normal or Benign Changes--includes:
                                    (1) Normal, negative or within
                                     normal limits.
                                    (2) Infection other than Human
                                     Papillomavirus (HPV) (e.g.,
                                     Trichomonas vaginalis, changes or
                                     morphology consistent with Candida
                                     spp., Actinomyces spp. or Herpes
                                     simplex virus).
                                    (3) Reactive and reparative changes
                                     (e.g., inflammation, effects of
                                     chemotherapy or radiation).
C.................................  Low Grade Squamous Intraepithelial
                                     Lesion--includes:
                                    (1) Cellular changes associated with
                                     HPV.
                                    (2) Mild dysplasia/CIN-1.
D.................................  High Grade Lesion and Carcinoma--
                                     includes:
                                    (1) High grade squamous
                                     intraepithelial lesions which
                                     include moderate dysplasia/CIN-2
                                     and severe dysplasia/carcinoma in-
                                     situ/CIN-3.
                                    (2) Squamous cell carcinoma.
                                    (3) Adenocarcinoma and other
                                     malignant neoplasms.
------------------------------------------------------------------------

    (B) In accordance with the criteria for the scoring system, the 
charts in paragraphs (b)(3)(ii)(C) and (D) of this section, for 
technical supervisors and cytotechnologists, respectively, provide a 
maximum of 10 points for a correct response and a maximum of minus five 
(-5) points for an incorrect response on a 10-slide test set. For 
example, if the correct response on a slide is ``high grade squamous 
intraepithelial lesion'' (category ``D'' on the scoring system chart) 
and an examinee calls it ``normal or negative'' (category ``B'' on the 
scoring system chart), then the examinee's point value on that slide is 
calculated as minus five (-5). Each slide is scored individually in the 
same manner. The individual's score for the testing event is determined 
by adding the point value achieved for each slide preparation, dividing 
by the total points for the testing event and multiplying by 100.
    (C) Criteria for scoring system for a 10-slide test set. (See table 
at (b)(3)(ii)(A) of this section for a description of the response 
categories.) For technical supervisors qualified under Sec. 493.1449(b) 
or (k):

------------------------------------------------------------------------
                Examinee's response:                   A    B    C    D
------------------------------------------------------------------------
Correct response category:
  A.................................................   10    0    0    0
  B.................................................    5   10    0    0
  C.................................................    5    0   10    5

[[Page 880]]


  D.................................................    0    5    5   10
------------------------------------------------------------------------

    (D) Criteria for scoring system for a 10-slide test set. (See table 
at paragraph (b)(3)(ii)(A) of this section for a description of the 
response categories.) For cytotechnologists qualified under 
Secs. 493.1469 or 493.1483:

------------------------------------------------------------------------
                Examinee's response:                   A    B    C    D
------------------------------------------------------------------------
Correct response category:
  A.................................................   10    0    5    5
  B.................................................    5   10    5    5
  C.................................................    5    0   10   10
  D.................................................    0   -5   10   10
------------------------------------------------------------------------

    (E) In accordance with the criteria for the scoring system, the 
charts in paragraphs (b)(3)(ii)(F) and (G) of this section, for 
technical supervisors and cytotechnologists, respectively, provide 
maximums of 5 points for a correct response and minus ten (-10) points 
for an incorrect response on a 20-slide test set.
    (F) Criteria for scoring system for a 20-slide test set. (See table 
at paragraph (b)(3)(ii)(A) of this section for a description of the 
response categories.) For technical supervisors qualified under 
Sec. 493.1449(b) or (k):

------------------------------------------------------------------------
             Examinee's response:                 A      B      C     D
------------------------------------------------------------------------
Correct response category:
  A...........................................   5         0   0     0
  B...........................................   2.5       5   0     0
  C...........................................   2.5       0   5     2.5
  D...........................................   0       -10   2.5   5
------------------------------------------------------------------------

    (G) Criteria for scoring system for a 20-slide test set. (See table 
at (b)(3)(ii)(A) of this section for a description of the response 
categories.) For cytotechnologists qualified under Secs. 493.1469 or 
493.1483:

------------------------------------------------------------------------
              Examinee's response:                 A     B      C     D
------------------------------------------------------------------------
Correct response category:
  A............................................   5        0   2.5   2.5
  B............................................   2.5      5   2.5   2.5
  C............................................   2.5      0   5     5
  D............................................   0      -10   5     5
------------------------------------------------------------------------


[57 FR 7151, Feb. 28, 1992, as amended at 58 FR 5229, Jan. 19, 1993]

Sec. 493.959  Immunohematology.

    (a) Types of services offered by laboratories. In immunohematology, 
there are four types of laboratories for proficiency testing purposes--
    (1) Those that perform ABO group and/or D (Rho) typing;
    (2) Those that perform ABO group and/or D (Rho) typing, and 
unexpected antibody detection;
    (3) Those that in addition to paragraph (a)(2) of this section 
perform compatibility testing; and
    (4) Those that perform in addition to paragraph (a)(3) of this 
section antibody identification.
    (b) Program content and frequency of challenge. To be approved for 
proficiency testing for immunohematology, a program must provide a 
minimum of five samples per testing event. There must be at least three 
testing events at approximately equal intervals per year. The annual 
program must provide samples that cover the full range of interpretation 
that would be expected in patient specimens. The samples may be provided 
through mailed shipments or, at HHS' option, may be provided to HHS or 
its designee for on-site testing.
    (c) Challenges per testing event. The minimum number of challenges 
per testing event a program must provide for each analyte or test 
procedure is five.

                        Analyte or Test Procedure

ABO group (excluding subgroups)
D (Rho) typing
Unexpected antibody detection
Compatibility testing
Antibody identification

    (d) Evaluation of a laboratory's analyte or test performance. HHS 
approves only those programs that assess the accuracy of a laboratory's 
response in accordance with paragraphs (d)(1) through (5) of this 
section.
    (1) To determine the accuracy of a laboratory's response, a program 
must compare the laboratory's response for each analyte with the 
response that reflects agreement of either 100 percent of ten or more 
referee laboratories or 95 percent or more of all participating 
laboratories except for unexpected antibody detection and antibody 
identification. To determine the accuracy of a laboratory's response for 
unexpected antibody detection and antibody

[[Page 881]]

identification, a program must compare the laboratory's response for 
each analyte with the response that reflects agreement of either 95 
percent of ten or more referee laboratories or 95 percent or more of all 
participating laboratories. The score for a sample in immunohematology 
is either the score determined under paragraph (d)(2) or (3) of this 
section.
    (2) Criteria for acceptable performance. The criteria for acceptable 
performance are--

------------------------------------------------------------------------
                                               Criteria for acceptable
              Analyte or test                        performance
------------------------------------------------------------------------
ABO group.................................  100% accuracy.
D (Rho) typing............................  100% accuracy.
Unexpected antibody detection.............  80% accuracy.
Compatibility testing.....................  100% accuracy.
Antibody identification...................  80% accuracy.
------------------------------------------------------------------------

    (3) The criterion for acceptable performance for qualitative 
immunohematology tests is positive or negative.
    (4) To determine the analyte testing event score, the number of 
acceptable analyte responses must be averaged using the following 
formula:



 Number of acceptable responses for the analyte x 100=Analyte score for
                            the testing event
-------------------------------------------------------------------------
               Total number of challenges for the analyte
-------------------------------------------------------------------------


    (5) To determine the overall testing event score, the number of 
correct responses for all analytes must be averaged using the following 
formula:



  Number of acceptable responses for all challenges x 100=Testing event
                                  score
-------------------------------------------------------------------------
                     Total number of all challenges
-------------------------------------------------------------------------


 Subpart J--Patient Test Management for Moderate Complexity (Including 
  the Subcategory), High Complexity, or Any Combination of These Tests

    Source: 57 FR 7162, Feb, 28, 1992, unless otherwise noted.

Sec. 493.1101  Condition: Patient test management; moderate complexity 
          (including the subcategory), or high complexity testing, or 
          any combination of these tests.

    Each laboratory performing moderate complexity (including the 
subcategory) or high complexity testing, or any combination of these 
tests, must employ and maintain a system that provides for proper 
patient preparation; proper specimen collection, identification, 
preservation, transportation, and processing; and accurate result 
reporting. This system must assure optimum patient specimen integrity 
and positive identification throughout the preanalytic (pre-testing), 
analytic (testing), and postanalytic (post-testing) processes and must 
meet the standards as they apply to the testing performed.

[60 FR 20048, Apr. 24, 1995]

Sec. 493.1103  Standard; Procedures for specimen submission and 
          handling.

    (a) The laboratory must have available and follow written policies 
and procedures for each of the following, if applicable: Methods used 
for the preparation of patients; specimen collection; specimen labeling; 
specimen preservation; conditions for specimen transportation; and 
specimen processing. Such policies and procedures must assure positive 
identification and optimum integrity of the patient specimens from the 
time the specimen(s) are collected until testing has been completed and 
the results reported.
    (b) If the laboratory accepts referral specimens, written 
instructions must be available to clients and must include, as 
appropriate, the information specified in paragraph (a) of this section.
    (c) Oral explanation of instructions to patients for specimen 
collection, including patient preparation, may be used as a supplement 
to written instructions where applicable.

[57 FR 7162, Feb. 28, 1992, as amended at 58 FR 5229, Jan. 19, 1993]

[[Page 882]]

Sec. 493.1105  Standard; Test requisition.

    The laboratory must perform tests only at the written or electronic 
request of an authorized person. Oral requests for laboratory tests are 
permitted only if the laboratory subsequently requests written 
authorization for testing within 30 days. The laboratory must maintain 
the written authorization or documentation of efforts made to obtain a 
written authorization. Records of test requisitions or test 
authorizations must be retained for a minimum of two years. The 
patient's chart or medical record, if used as the test requisition, must 
be retained for a minimum of two years and must be available to the 
laboratory at the time of testing and available to HHS upon request. The 
laboratory must assure that the requisition or test authorization 
includes--
    (a) The patient's name or other unique identifier;
    (b) The name and address or other suitable identifiers of the 
authorized person requesting the test and, if appropriate, the 
individual responsible for utilizing the test results or the name and 
address of the laboratory submitting the specimen, including, as 
applicable, a contact person to enable the reporting of imminent life 
threatening laboratory results or panic values;
    (c) The test(s) to be performed;
    (d) The date of specimen collection;
    (e) For Pap smears, the patient's last menstrual period, age or date 
of birth, and indication of whether the patient had a previous abnormal 
report, treatment or biopsy; and
    (f) Any additional information relevant and necessary to a specific 
test to assure accurate and timely testing and reporting of results.

[57 FR 7162, Feb. 28, 1992, as amended at 58 FR 5229, Jan. 19, 1993]

Sec. 493.1107  Standard; Test records.

    The laboratory must maintain a record system to ensure reliable 
identification of patient specimens as they are processed and tested to 
assure that accurate test results are reported. These records must 
identify the personnel performing the testing procedure. Records of 
patient testing, including, if applicable, instrument printouts, must be 
retained for at least two years. Immunohematology records and 
transfusion records must be retained for no less than five years in 
accordance with 21 CFR part 606, subpart I. In addition, records of 
blood and blood product testing must be maintained for a period not less 
than five years after processing records have been completed, or six 
months after the latest expiration date, whichever is the later date, in 
accordance with 21 CFR 606.160(d). The record system must provide 
documentation of information specified in Sec. 493.1105 (a) through (f) 
and include--
    (a) The patient identification number, accession number, or other 
unique identification of the specimen;
    (b) The date and time of specimen receipt into the laboratory;
    (c) The condition and disposition of specimens that do not meet the 
laboratory's criteria for specimen acceptability; and
    (d) The records and dates of all specimen testing, including the 
identity of the personnel who performed the test(s), which are necessary 
to assure proper identification and accurate reporting of patient test 
results.

[57 FR 7162, Feb. 28, 1992, as amended at 58 FR 5229, Jan. 19, 1993]

Sec. 493.1109  Standard; Test report.

    The laboratory report must be sent promptly to the authorized 
person, the individual responsible for using the test results or 
laboratory that initially requested the test. The original report or an 
exact duplicate of each test report, including final and preliminary 
report, must be retained by the testing laboratory for a period of at 
least two years after the date of reporting. Immunohematology reports 
and transfusion records must be retained by the laboratory for a period 
of no less than five years in accordance with 21 CFR part 606, subpart 
I. In addition, records of blood and blood product testing must be 
maintained for a period not less than five years after processing 
records have been completed, or six months after the latest expiration 
date, whichever is the later date, in accordance with 21 CFR 606.160(d). 
For pathology, test reports must be retained

[[Page 883]]

for a period of at least ten years after the date of reporting. This 
information may be maintained as part of the patient's chart or medical 
record which must be readily available to the laboratory and to HHS upon 
request.
    (a) The laboratory must have adequate systems in place to report 
results in a timely, accurate, reliable and confidential manner, and, 
ensure patient confidentiality throughout those parts of the total 
testing process that are under the laboratory's control.
    (b) The test report must indicate the name and address of the 
laboratory location at which the test was performed, the test performed, 
the test result and, if applicable, the units of measurement.
    (c) The laboratory must indicate on the test report any information 
regarding the condition and disposition of specimens that do not meet 
the laboratory's criteria for acceptability.
    (d) Pertinent ``reference'' or ``normal'' ranges, as determined by 
the laboratory performing the tests, must be available to the authorized 
person who ordered the tests or the individual responsible for utilizing 
the test results.
    (e) The results or transcripts of laboratory tests or examinations 
must be released only to authorized persons or the individual 
responsible for utilizing the test results.
    (f) The laboratory must develop and follow written procedures for 
reporting imminent life-threatening laboratory results or panic values. 
In addition, the laboratory must immediately alert the individual or 
entity requesting the test or the individual responsible for utilizing 
the test results when any test result indicates an imminent life-
threatening condition.
    (g) The laboratory must, upon request, make available to clients a 
list of test methods employed by the laboratory and, in accordance with 
Sec. 493.1213, as applicable, the performance specifications of each 
method used to test patient specimens. In addition, information that may 
affect the interpretation of test results, such as test interferences, 
must be provided upon request. Pertinent updates on testing information 
must be provided to clients whenever changes occur that affect the test 
results or interpretation of test results.
    (h) The original report or exact duplicates of test reports must be 
maintained by the laboratory in a manner that permits ready 
identification and timely accessibility.

[57 FR 7162, Feb. 28, 1992, as amended at 58 FR 5229, Jan. 19, 1993]

Sec. 493.1111  Standard; Referral of specimens.

    A laboratory must refer specimens for testing only to a laboratory 
possessing a valid certificate authorizing the performance of testing in 
the specialty or subspecialty of service for the level of complexity in 
which the referred test is categorized.
    (a) The referring laboratory must not revise results or information 
directly related to the interpretation of results provided by the 
testing laboratory.
    (b) The referring laboratory may permit each testing laboratory to 
send the test result directly to the authorized person who initially 
requested the test. The referring laboratory must retain or be able to 
produce an exact duplicate of each testing laboratory's report.
    (c) The authorized person who orders a test or procedure must be 
notified by the referring laboratory of the name and address of each 
laboratory location at which a test was performed.

 Subpart K--Quality Control for Tests of Moderate Complexity (Including 
  the Subcategory), High Complexity, or Any Combination of These Tests

    Source: 57 FR 7163, Feb. 28, 1992, unless otherwise noted.

Sec. 493.1201  Condition: General quality control; moderate complexity 
          (including the subcategory) or high complexity testing, or any 
          combination of these tests.

    (a) Applicability of subpart K of this part. Subpart K is divided 
into two sections, general quality control and quality control for 
specialties and subspecialties. The quality control requirements are 
specified in Secs. 493.1201 through 493.1285 unless--

[[Page 884]]

    (1) An alternative procedure specified in the manufacturer's 
protocol has been cleared by the Food and Drug Administration (FDA) as 
meeting certain CLIA requirements for general quality control and 
specialty/subspecialty quality control, and the manufacturer's 
instructions contain the following statement,
    Unless this device is modified by a laboratory, the laboratory's 
compliance with these quality control instructions will satisfy the 
applicable requirements of 42 CFR 493.1203(b).

or
    (2) HHS approves an equivalent procedure that is specified in 
Appendix C of the State Operations Manual (HCFA Pub. 7).
    (b) The laboratory must establish and follow written quality control 
procedures for monitoring and evaluating the quality of the analytical 
testing process of each method to assure the accuracy and reliability of 
patient test results and reports. The laboratory must meet the 
applicable standards in Secs. 493.1202 through 493.1221 of this subpart, 
unless an alternative procedure specified in the manufacturer's protocol 
has been cleared by the Food and Drug Administration (FDA) as meeting 
certain CLIA requirements for quality control or HHS approves an 
equivalent procedure specified in appendix C of the State Operations 
Manual (HCFA Pub. 7). HCFA Pub. 7 is available from the Technical 
Information Service, U.S. Department of Commerce, 5825 Port Royal Road, 
Springfield, VA 22161, telephone number (703) 487-4630.

[58 FR 5230, Jan. 19, 1993, as amended at 60 FR 20048, Apr. 24, 1995]

Sec. 493.1202  Standard; Moderate or high complexity testing, or both: 
          Effective from September 1, 1992 to December 31, 2000.

    (a) For each test of high complexity performed, the laboratory must 
meet all applicable standards of this subpart.
    (b) For each test of moderate complexity performed using a 
standardized method, or method developed in-house, a device not subject 
to clearance by the FDA (including any commercially distributed 
instrument, kit or test system subject to the Food, Drug and Cosmetic 
Act marketed prior to the Medical Device Amendments, Public Law 94-295, 
enacted on May 28, 1976, and those identified in 21 CFR parts 862, 864, 
and 866 as exempt from FDA premarket review), or using an instrument, 
kit or test system cleared by the FDA through the premarket notification 
(510(k)) or premarket approval (PMA) process for in-vitro diagnostic use 
but modified by the laboratory, the laboratory must meet all applicable 
standards of this subpart.
    (c) For all other tests of moderate complexity performed using an 
instrument, kit or test system cleared by the FDA through the premarket 
notification (510(k)) or premarket approval (PMA) process for in-vitro 
diagnostic use, the laboratory must--(1) Follow the manufacturer's 
instructions for instrument or test system operation and test 
performance;
    (2) Have a procedure manual describing the processes for testing and 
reporting patient test results;
    (3) Perform and document calibration procedures or check calibration 
at least once every six months;
    (4) Perform and document control procedures using at least two 
levels of control materials each day of testing;
    (5) Perform and document applicable specialty and subspecialty 
control procedures as specified under Sec. 493.1223;
    (6) Perform and document that remedial action has been taken when 
problems or errors are identified as specified in Sec. 493.1219; and
    (7) Maintain records of all quality control activities for two 
years. Quality control records for immunohematology and blood and blood 
products must be maintained as specified in Sec. 493.1221.

[57 FR 7163, Feb. 28, 1992, as amended at 58 FR 5230, Jan. 19, 1993]

Sec. 493.1203  Standard; Moderate or high complexity testing, or both: 
          Effective beginning December 31, 2000.

    For each moderate or high complexity test performed, the laboratory 
will be in compliance with this section if it:
    (a) Meets all applicable quality control requirements specified in 
this subpart when using a standardized method, a method developed in-
house, a device not subject to clearance by the

[[Page 885]]

FDA (including any commercially distributed instrument, kit or test 
system subject to the Food, Drug and Cosmetic Act marketed prior to the 
Medical Device Amendments, Public Law 94-295, enacted on May 28, 1976, 
and those identified in 21 CFR parts 862, 864, and 866 as exempt from 
FDA premarket review), a manufacturer's product modified by the 
laboratory, or a device (instrument, kit or test system) not cleared by 
the FDA as meeting certain CLIA quality control requirements; or
    (b) Follows manufacturer's instructions when using a device 
(instrument, kit, or test system) cleared by the FDA as meeting the CLIA 
requirements for quality control located at Secs. 493.1215, 493.1217, 
and 493.1223, and applicable parts of Secs. 493.1205, 493.1211 and 
493.1218. In addition, the laboratory must comply with the requirements 
of Secs. 493.1204, 493.1213, 493.1219, and 493.1221 and those parts of 
Secs. 493.1205, 493.1211, and 493.1218 that are unique to the laboratory 
facility and cannot be met by following manufacturer's instructions.

[58 FR 5230, Jan. 19, 1993]

Sec. 493.1204  Standard; Facilities.

    The laboratory must provide the space and environmental conditions 
necessary for conducting the services offered.
    (a) The laboratory must be constructed, arranged, and maintained to 
ensure the space, ventilation, and utilities necessary for conducting 
all phases of testing, including the preanalytic (pre-testing), analytic 
(testing), and postanalytic (post-testing), as appropriate.
    (b) Safety precautions must be established, posted, and observed to 
ensure protection from physical, chemical, biochemical and electrical 
hazards and biohazardous materials.

[57 FR 7163, Feb. 28, 1992, as amended at 58 FR 5230, Jan. 19, 1993]

Sec. 493.1205  Standard; Test methods, equipment, instrumentation, 
          reagents, materials, and supplies.

    The laboratory must utilize test methods, equipment, 
instrumentation, reagents, materials, and supplies that provide accurate 
and reliable test results and test reports.
    (a) Test methodologies and equipment must be selected and testing 
performed in a manner that provides test results within the laboratory's 
stated performance specifications for each test method as determined 
under Sec. 493.1213.
    (b) The laboratory must have appropriate and sufficient equipment, 
instruments, reagents, materials, and supplies for the type and volume 
of testing performed and for the maintenance of quality during the 
preanalytic, analytic, and postanalytic phases of testing.
    (c) The laboratory must define criteria for those conditions that 
are essential for proper storage of reagents and specimens, and accurate 
and reliable test system operation and test result reporting.
    (1) These conditions include, if applicable--
    (i) Water quality;
    (ii) Temperature;
    (iii) Humidity; and
    (iv) Protection of equipment and instrumentation from fluctuations 
and interruptions in electrical current that adversely affect patient 
test results and test reports.
    (2) Remedial actions taken to correct conditions that fail to meet 
the criteria specified in paragraph (c)(1) of this section must be 
documented.
    (d) Reagents, solutions, culture media, control materials, 
calibration materials and other supplies, as appropriate, must be 
labeled to indicate--
    (1) Identity and, when significant, titer, strength or 
concentration;
    (2) Recommended storage requirements;
    (3) Preparation and expiration date; and
    (4) Other pertinent information required for proper use.
    (e) Reagents, solutions, culture media, control materials, 
calibration materials and other supplies must be prepared, stored, and 
handled in a manner to ensure that--
    (1) Reagents, solutions, culture media, controls, calibration 
materials and other supplies are not used when they have exceeded their 
expiration date, have deteriorated or are of substandard quality. The 
laboratory must comply with the FDA product dating

[[Page 886]]

requirements of 21 CFR 610.53 for blood products and other biologicals, 
and labeling requirements, as cited in 21 CFR 809.10 for all other in 
vitro diagnostics. Any exception to the product dating requirements in 
21 CFR 610.53 will be granted by the FDA in the form of an amendment of 
the product license, in accordance with 21 CFR 610.53(d). All exceptions 
must be documented by the laboratory; and
    (2) Components of reagent kits of different lot numbers are not 
interchanged unless otherwise specified by the manufacturer.

[57 FR 7163, Feb. 28, 1992, as amended at 58 FR 5230, Jan. 19, 1993]

Sec. 493.1211  Standard; Procedure manual.

    (a) A written procedure manual for the performance of all analytical 
methods used by the laboratory must be readily available and followed by 
laboratory personnel. Textbooks may be used as supplements to these 
written descriptions but may not be used in lieu of the laboratory's 
written procedures for testing or examining specimens.
    (b) The procedure manual must include, when applicable to the test 
procedure:
    (1) Requirements for specimen collection and processing, and 
criteria for specimen rejection;
    (2) Procedures for microscopic examinations, including the detection 
of inadequately prepared slides;
    (3) Step-by-step performance of the procedure, including test 
calculations and interpretation of results;
    (4) Preparation of slides, solutions, calibrators, controls, 
reagents, stains and other materials used in testing;
    (5) Calibration and calibration verification procedures;
    (6) The reportable range for patient test results as established or 
verified in Sec. 493.1213;
    (7) Control procedures;
    (8) Remedial action to be taken when calibration or control results 
fail to meet the laboratory's criteria for acceptability;
    (9) Limitations in methodologies, including interfering substances;
    (10) Reference range (normal values);
    (11) Imminent life-threatening laboratory results or ``panic 
values'';
    (12) Pertinent literature references;
    (13) Appropriate criteria for specimen storage and preservation to 
ensure specimen integrity until testing is completed;
    (14) The laboratory's system for reporting patient results 
including, when appropriate, the protocol for reporting panic values;
    (15) Description of the course of action to be taken in the event 
that a test system becomes inoperable; and
    (16) Criteria for the referral of specimens including procedures for 
specimen submission and handling as described in Sec. 493.1103.
    (c) Manufacturers' package inserts or operator manuals may be used, 
when applicable, to meet the requirements of paragraphs (b)(1) through 
(b)(13) of this section. Any of the items under paragraphs (b)(1) 
through (b)(13) of this section not provided by the manufacturer must be 
provided by the laboratory.
    (d) Procedures must be approved, signed, and dated by the director.
    (e) Procedures must be re-approved, signed and dated if the 
directorship of the laboratory changes.
    (f) Each change in a procedure must be approved, signed, and dated 
by the current director of the laboratory.
    (g) The laboratory must maintain a copy of each procedure with the 
dates of initial use and discontinuance. These records must be retained 
for two years after a procedure has been discontinued.

Sec. 493.1213  Standard; Establishment and verification of method 
          performance specifications.

    Prior to reporting patient test results, the laboratory must verify 
or establish, for each method, the performance specifications for the 
following performance characteristics: accuracy; precision; analytical 
sensitivity and specificity, if applicable; the reportable range of 
patient test results; the reference range(s) (normal values); and any 
other applicable performance characteristic.
    (a) The provisions of this section are not retroactive. Laboratories 
are not required to verify or establish performance specifications for 
any test method

[[Page 887]]

of moderate or high complexity in use prior to September 1, 1992.
    (b)(1) Each laboratory that introduces a new procedure for patient 
testing using a device (instrument, kit, or test system) cleared by the 
FDA as meeting certain CLIA requirements for quality control, must 
demonstrate that, prior to reporting patient test results, it can obtain 
the performance specifications for accuracy, precision, and reportable 
range of patient test results, comparable to those established by the 
manufacturer. The laboratory must also verify that the manufacturer's 
reference range is appropriate for the laboratory's patient population.
    (2) Each laboratory that introduces a new method or device as 
specified in either Sec. 493.1202(a) or (b), or Sec. 493.1203(a), must, 
prior to reporting patient test results--
    (i) Verify or establish for each method the performance 
specifications for the following performance characteristics, as 
applicable:
    (A) Accuracy;
    (B) Precision;
    (C) Analytical sensitivity;
    (D) Analytical specificity to include interfering substances;
    (E) Reportable range of patient test results;
    (F) Reference range(s); and
    (G) Any other performance characteristic required for test 
performance.
    (ii) Based upon the performance specifications verified or 
established in accordance with paragraph (b)(2)(i) of this section, 
establish calibration and control procedures for patient testing as 
required under Secs. 493.1217 and 493.1218.
    (c) The laboratory must have documentation of the verification or 
establishment of all applicable test performance specifications.

[57 FR 7163, Feb. 28, 1992, as amended at 58 FR 5230, Jan. 19, 1993]

Sec. 493.1215  Standard; Equipment maintenance and function checks.

    The laboratory must perform equipment maintenance and function 
checks that include electronic, mechanical and operational checks 
necessary for the proper test performance and test result reporting of 
equipment, instruments and test systems, to assure accurate and reliable 
test results and reports.
    (a) Maintenance of equipment, instruments, and test systems. (1) For 
manufacturers' equipment, instruments or test systems cleared by the FDA 
as meeting certain CLIA requirements for quality control, the laboratory 
must--
    (i) Perform maintenance as defined by the manufacturer and with at 
least the frequency specified by the manufacturer; and
    (ii) Document all maintenance performed.
    (2) For methods or devices, as specified in either Sec. 493.1202(a) 
or (b) or Sec. 493.1203(a), the laboratory must--
    (i) Establish a maintenance protocol that ensures equipment, 
instrument, and test system performance necessary for accurate and 
reliable test results and test result reporting;
    (ii) Perform maintenance with at least the frequency specified in 
paragraph (a)(2)(i) of this section; and
    (iii) Document all maintenance performed.
    (b) Function checks of equipment, instruments, and test systems. (1) 
For manufacturers' equipment, instruments, or test systems cleared by 
the FDA as meeting certain CLIA requirements for quality control, the 
laboratory must--
    (i) Perform function checks as defined by the manufacturer and with 
at least the frequency specified by the manufacturer; and
    (ii) Document all function checks performed.
    (2) For methods or devices, as specified in either Sec. 493.1202 (a) 
or (b) or Sec. 493.1203(a), the laboratory must--
    (i) Define a function check protocol that ensures equipment, 
instrument, and test system performance necessary for accurate and 
reliable test results and test result reporting;
    (ii) Perform function checks including background or baseline checks 
specified in paragraph (b)(2)(i) of this section. Function checks must 
be within the laboratory's established limits before patient testing is 
conducted; and
    (iii) Document all function checks performed.

[57 FR 7163, Feb. 28, 1992, as amended at 58 FR 5231, Jan. 19, 1993; 58 
FR 39155, July 22, 1993]

[[Page 888]]

Sec. 493.1217  Standard; Calibration and calibration verification 
          procedures.

    Calibration and calibration verification procedures are required to 
substantiate the continued accuracy of the test method throughout the 
laboratory's reportable range for patient test rests. Calibration is the 
process of testing and adjusting an instrument, kit, or test system to 
provide a known relationship between the measurement response and the 
value of the substance that is being measured by the test procedure. 
Calibration verification is the assaying of calibration materials in the 
same manner as patient samples to confirm that the calibration of the 
instrument, kit, or test system has remained stable throughout the 
laboratory's reportable range for patient test results. The reportable 
range of patient test results is the range of test result values over 
which the laboratory can establish or verify the accuracy of the 
instrument, kit or test system measurement response. Calibration and 
calibration verification must be performed and documented as required in 
this section unless otherwise specified in Secs. 493.1223 through 
493.1285.
    (a) For laboratory test procedures that are performed using 
instruments, kits, or test systems that have been cleared by the FDA as 
meeting certain CLIA requirements for quality control, the laboratory 
must, at a minimum, follow the manufacturer's instructions for 
calibration and calibration verification procedures using calibration 
materials specified by the manufacturer.
    (b) For each method or device, as specified in either Sec. 493.1202 
(a) or (b) or Sec. 493.1203(a), the laboratory must--
    (1) Perform calibration procedures--
    (i) At a minimum, in accordance with manufacturer's instructions, if 
provided, using calibration materials provided or specified, as 
appropriate, and with at least the frequency recommended by the 
manufacturer; and
    (ii) In accordance with criteria established by the laboratory, as 
required under Sec. 493.1213(b)(2)(i)--
    (A) Including the number, type and concentration of calibration 
materials, acceptable limits for calibration, and the frequency of 
calibration; and
    (B) Using calibration materials appropriate for the methodology and, 
if possible, traceable to a reference method or reference material of 
known value; and
    (iii) Whenever calibration verification fails to meet the 
laboratory's acceptable limits for calibration verification; and
    (2) Perform calibration verification procedures--
    (i) In accordance with the manufacturer's calibration verification 
instructions when they meet or exceed the requirements specified in 
paragraph (b)(2)(ii) of this section; or
    (ii) In accordance with criteria established by the laboratory--
    (A) Including the number, type, and concentration of calibration 
materials, acceptable limits for calibration verification and frequency 
of calibration verification;
    (B) Using calibration materials appropriate for--
    (1) The methodology and, if possible, traceable to a reference 
method or reference material of known value; and
    (2) Verifying the laboratory's established reportable range of 
patient test results, which must include at least a minimal (or zero) 
value, a mid-point value, and a maximum value at the upper limit of that 
range; and
    (C) At least once every six months and whenever any of the following 
occur:
    (1) A complete change of reagents for a procedure is introduced, 
unless the laboratory can demonstrate that changing reagent lot numbers 
does not affect the range used to report patient test results, and 
control values are not adversely affected by reagent lot number changes;

    Note: If reagents are obtained from a manufacturer and all of the 
reagents for a test are packaged together, the laboratory is not 
required to perform calibration verification for each package of 
reagents, provided the packages of reagents are received in the same 
shipment and contain the same lot number.

    (2) There is major preventive maintenance or replacement of critical 
parts that may influence test performance;
    (3) Controls reflect an unusual trend or shift or are outside of the 
laboratory's acceptable limits and other

[[Page 889]]

means of assessing and correcting unacceptable control values have 
failed to identify and correct the problem; or
    (4) The laboratory's established schedule for verifying the 
reportable range for patient test results requires more frequent 
calibration verification than specified in paragraphs (b)(2)(ii)(C) (1), 
(2), or (3) of this section; and
    (3) Document all calibration and calibration verification procedures 
performed.

[58 FR 5231, Jan. 19, 1993]

Sec. 493.1218  Standard; Control procedures.

    Control procedures are performed on a routine basis to monitor the 
stability of the method or test system; control and calibration 
materials provide a means to indirectly assess the accuracy and 
precision of patient test results. Control procedures must be performed 
as defined in this section unless otherwise specified in Secs. 493.1223 
through 493.1285 of this subpart.
    (a) For each device cleared by the FDA as meeting certain CLIA 
requirements for quality control, the laboratory must, at a minimum, 
follow the manufacturer's instructions for control procedures. In 
addition, the laboratory must meet the requirements under paragraphs (c) 
through (e) of this section and, as applicable, paragraph (f) of this 
section.
    (b) For each device, as specified in either Sec. 493.1202 (a) or (b) 
or Sec. 493.1203(a), the laboratory must evaluate instrument and reagent 
stability and operator variance in determining the number, type, and 
frequency of testing calibration or control materials and establish 
criteria for acceptability used to monitor test performance during a run 
of patient specimen(s). A run is an interval within which the accuracy 
and precision of a testing system is expected to be stable, but cannot 
be greater than 24 hours or less than the frequency recommended by the 
manufacturer. For each procedure, the laboratory must monitor test 
performance using calibration materials or control materials or a 
combination thereof.
    (1) For qualitative tests, the laboratory must include a positive 
and negative control with each run of patient specimens.
    (2) For quantitative tests, the laboratory must include at least two 
samples of different concentrations of either calibration materials, 
control materials, or a combination thereof with the frequency 
determined in Sec. 493.1218(b), but not less frequently than once each 
run of patient specimens.
    (3) For electrophoretic determinations--
    (i) At least one control sample must be used in each electrophoretic 
cell; and
    (ii) The control sample must contain fractions representative of 
those routinely reported in patient specimens.
    (4) Each day of use, the laboratory must evaluate the detection 
phase of direct antigen systems using an appropriate positive and 
negative control material (organism or antigen extract). When direct 
antigen systems include an extraction phase, the system must be checked 
each day of use using a positive organism.
    (5) If calibration materials and control materials are not 
available, the laboratory must have an alternative mechanism to assure 
the validity of patient test results.
    (c) Control samples must be tested in the same manner as patient 
specimens.
    (d) When calibration or control materials are used, statistical 
parameters (e.g., mean and standard deviation) for each lot number of 
calibration material and each lot of control material must be determined 
through repetitive testing.
    (1) The stated values of an assayed control material may be used as 
the target values provided the stated values correspond to the 
methodology and instrumentation employed by the laboratory and are 
verified by the laboratory.
    (2) Statistical parameters for unassayed materials must be 
established over time by the laboratory through concurrent testing with 
calibration materials or control materials having previously determined 
statistical parameters.
    (e) Control results must meet the laboratory's criteria for 
acceptability prior to reporting patient test results.

[[Page 890]]

    (f) Reagent and supply checks. (1) The laboratory must check each 
batch or shipment of reagents, discs, stains, antisera and 
identification systems (systems using two or more substrates) when 
prepared or opened for positive and negative reactivity, as well as 
graded reactivity if applicable.
    (2) Each day of use (unless otherwise specified in this subpart), 
the laboratory must test staining materials for intended reactivity to 
ensure predictable staining characteristics.
    (3) The laboratory must check fluorescent stains for positive and 
negative reactivity each time of use (unless otherwise specified in this 
subpart).
    (4) The laboratory must check each batch or shipment of media for 
sterility, if it is intended to be sterile, and sterility is required 
for testing. Media must also be checked for its ability to support 
growth, and as appropriate, selectivity/inhibition and/or biochemical 
response. The laboratory may use manufacturer's control checks of media 
provided the manufacturer's product insert specifies that the 
manufacturer's quality control checks meet the National Committee for 
Clinical Laboratory Standards (NCCLS) for media quality control. The 
laboratory must document that the physical characteristics of the media 
are not compromised and report any deterioration in the media to the 
manufacturer. The laboratory must follow the manufacturer's 
specifications for using the media and be responsible for the test 
results.

    Note: A batch of media (solid, semi-solid, or liquid) consists of 
all tubes, plates, or containers of the same medium prepared at the same 
time and in the same laboratory; or, if received from an outside source 
or commercial supplier, consists of all of the plates, tubes or 
containers of the same medium that have the same lot numbers and are 
received in a single shipment.

[57 FR 7163, Feb. 28, 1992, as amended at 58 FR 5232, Jan. 19, 1993]

Sec. 493.1219  Standard; Remedial actions.

    Remedial action policies and procedures must be established by the 
laboratory and applied as necessary to maintain the laboratory's 
operation for testing patient specimens in a manner that assures 
accurate and reliable patient test results and reports. The laboratory 
must document all remedial actions taken when--
    (a) Test systems do not meet the laboratory's established 
performance specifications, as determined in Sec. 493.1213 of this 
section, which include but are not limited to--
    (1) Equipment or methodologies that perform outside of established 
operating parameters or performance specifications;
    (2) Patient test values that are outside of the laboratory's 
reportable range of patient test results; and
    (3) The determination that the laboratory's reference range for a 
test procedure is inappropriate for the laboratory's patient population.
    (b) Results of control and calibration materials fail to meet the 
laboratory's established criteria for acceptability. All patient test 
results obtained in the unacceptable test run or since the last 
acceptable test run must be evaluated to determine if patient test 
results have been adversely affected and the laboratory must take the 
remedial action necessary to ensure the reporting of accurate and 
reliable patient test results;
    (c) The laboratory cannot report patient test results within its 
established time frames. The laboratory must determine, based on the 
urgency of the patient test(s) requested, the need to notify the 
appropriate individual of the delayed testing; and
    (d) Errors in the reported patient test results are detected. The 
laboratory must--
    (1) Promptly notify the authorized person ordering or individual 
utilizing the test results of reporting errors;
    (2) Issue corrected reports promptly to the authorized person 
ordering the test or the individual utilizing the test results; and
    (3) Maintain exact duplicates of the original report as well as the 
corrected report for two years.

Sec. 493.1221  Standard; Quality control records.

    The laboratory must document and maintain records of all quality 
control activities specified in Secs. 493.1202 through 493.1285 of this 
subpart and retain records for at least two years.

[[Page 891]]

Immunohematology quality control records must be maintained for a period 
of no less than five years. In addition, quality control records for 
blood and blood products must be maintained for a period not less than 
five years after processing records have been completed, or six months 
after the latest expiration date, whichever is the later date, in 
accordance with 21 CFR 606.160(d).

Sec. 493.1223  Condition: Quality control--specialties and 
          subspecialties for tests of moderate or high complexity, or 
          both.

    The laboratory must establish and follow written quality control 
procedures for monitoring and evaluating the quality of the analytical 
testing process of each method to assure the accuracy and reliability of 
patient test results and reports. Except as specified in 
Sec. 493.1202(c), the laboratory must meet the applicable general 
requirements specified in Secs. 493.1201 through 493.1221. In addition, 
the laboratory must meet the applicable requirements of Secs. 493.1225 
through 493.1285 unless an alternative procedure specified in the 
manufacturer's protocol has been cleared by the Food and Drug 
Administration (FDA) as meeting certain CLIA requirements for quality 
control or HCFA approves an equivalent procedure specified in appendix C 
of the State Operations Manual (HCFA Pub. 7). Failure to meet any of the 
applicable conditions in Secs. 493.1225 through 493.1285 will result in 
intermediate sanctions, loss of Medicare or Medicaid approval, and/or 
revocation of CLIA certification for the entire specialty or 
subspecialty to which the condition applies, in accordance with subpart 
R of this part.

[58 FR 5232, Jan. 19, 1993]

Sec. 493.1225  Condition: Microbiology.

    The laboratory must meet the applicable quality control requirements 
in Secs. 493.1201 through 493.1221 and in Secs. 493.1227 through 
493.1235 of this subpart for the subspecialties for which it is 
certified under the specialty of microbiology.

Sec. 493.1227  Condition: Bacteriology.

    To meet the quality control requirements for bacteriology, the 
laboratory must comply with the applicable requirements in 
Secs. 493.1201 through 493.1221 and with paragraphs (a) through (c) of 
this section. All quality control activities must be documented.
    (a) The laboratory must check positive and negative reactivity with 
control organisms--
    (1) Each day of use for catalase, coagulase, beta-lactamase, and 
oxidase reagents and DNA probes;
    (2) Each week of use for Gram and acid-fast stains, bacitracin, 
optochin, ONPG, X, and V discs or strips; and
    (3) Each month of use for antisera.
    (b) Each week of use, the laboratory must check XV discs or strips 
with a positive control organism.
    (c) For antimicrobial susceptibility tests, the laboratory must 
check each new batch of media and each lot of antimicrobial discs 
before, or concurrent with, initial use, using approved reference 
organisms.
    (1) The laboratory's zone sizes or minimum inhibitory concentration 
for reference organisms must be within established limits before 
reporting patient results.
    (2) Each day tests are performed, the laboratory must use the 
appropriate control organism(s) to check the procedure.

Sec. 493.1229  Condition: Mycobacteriology.

    To meet the quality control requirements for mycobacteriology, the 
laboratory must comply with the applicable requirements in 
Secs. 493.1201 through 493.1221 of this subpart and with paragraphs (a) 
through (d) of this section. All quality control activities must be 
documented.
    (a) Each day of use, the laboratory must check the iron uptake test 
with at least one acid-fast organism that produces a positive reaction 
and with an organism that produces a negative reaction and check all 
other reagents or test procedures used for mycobacteria identification 
with at least one acid-fast organism that produces a positive reaction.

[[Page 892]]

    (b) The laboratory must check fluorochrome acid-fast stains for 
positive and negative reactivity each week of use.
    (c) The laboratory must check acid-fast stains each week of use with 
an acid-fast organism that produces a positive reaction.
    (d) For susceptibility tests performed on Mycobacterium tuberculosis 
isolates, the laboratory must check the procedure each week of use with 
a strain of Mycobacterium tuberculosis susceptible to all 
antimycobacterial agents tested.

Sec. 493.1231  Condition: Mycology.

    To meet the quality control requirements for mycology, the 
laboratory must comply with the applicable requirements in 
Secs. 493.1201 through 493.1221 of this subpart and with paragraphs (a) 
through (d) of this section. All quality control activities must be 
documented.
    (a) Each day of use, the laboratory using the auxanographic medium 
for nitrate assimilation must check the nitrate reagent with a peptone 
control.
    (b) Each week of use, the laboratory must check all reagents used 
with biochemical tests and other test procedures for mycological 
identification with an organism that produces a positive reaction.
    (c) Each week of use, the laboratory must check acid-fast stains for 
positive and negative reactivity.
    (d) For susceptibility tests, the laboratory must test each drug 
each day of use with at least one control strain that is susceptible to 
the drug. The laboratory must establish control limits. Criteria for 
acceptable control results must be met prior to reporting patient 
results.

Sec. 493.1233  Condition: Parasitology.

    To meet the quality control requirements for parasitology, the 
laboratory must comply with the applicable requirements of 
Secs. 493.1201 through 493.1221 of this subpart and with paragraphs (a) 
through (c) of this section. All quality control activities must be 
documented.
    (a) The laboratory must have available a reference collection of 
slides or photographs, and, if available, gross specimens for 
identification of parasites and use these references in the laboratory 
for appropriate comparison with diagnostic specimens.
    (b) The laboratory must calibrate and use the calibrated ocular 
micrometer for determining the size of ova and parasites, if size is a 
critical parameter.
    (c) Each month of use, the laboratory must check permanent stains 
using a fecal sample control that will demonstrate staining 
characteristics.

Sec. 493.1235  Condition: Virology.

    To meet the quality control requirements for virology, the 
laboratory must comply with the applicable requirements in 
Secs. 493.1201 through 493.1221 of this subpart and with paragraphs (a) 
through (c) of this section. All quality control activities must be 
documented.
    (a) The laboratory must have available host systems for the 
isolation of viruses and test methods for the identification of viruses 
that cover the entire range of viruses that are etiologically related to 
clinical diseases for which services are offered.
    (b) The laboratory must maintain records that reflect the systems 
used and the reactions observed.
    (c) In tests for the identification of viruses, the laboratory must 
simultaneously culture uninoculated cells or cell substrate controls as 
a negative control to detect erroneous identification results.

Sec. 493.1237  Condition: Diagnostic immunology.

    The laboratory must meet the applicable quality control requirements 
in Secs. 493.1201 through 493.1221 and Secs. 493.1239 through 493.1241 
of this subpart for the subspecialties for which it is certified under 
the specialty of diagnostic immunology.

Sec. 493.1239  Condition: Syphilis serology.

    To meet the quality control requirements for syphilis serology, the 
laboratory must comply with the applicable requirements in 
Secs. 493.1201 through 493.1221 of this subpart and with paragraphs (a) 
through (e) of this section. All quality control activities must be 
documented.

[[Page 893]]

    (a) For laboratories performing syphilis testing, the equipment, 
glassware, reagents, controls, and techniques for tests for syphilis 
must conform to manufacturers' specifications.
    (b) The laboratory must run serologic tests on patient specimens 
concurrently with a positive serum control of known titer or controls of 
graded reactivity plus a negative control.
    (c) The laboratory must employ positive and negative controls that 
evaluate all phases of the test system to ensure reactivity and uniform 
dosages.
    (d) The laboratory may not report test results unless the 
predetermined reactivity pattern of the controls is observed.
    (e) All facilities manufacturing blood and blood products for 
transfusion or serving as referral laboratories for these facilities 
must meet the syphilis serology testing requirements of 21 CFR 640.5(a).

Sec. 493.1241  Condition: General immunology.

    To meet the quality control requirements for general immunology, the 
laboratory must comply with the applicable requirements in 
Secs. 493.1201 through 493.1221 of this subpart and with paragraphs (a) 
through (d) of this section. All quality control activities must be 
documented.
    (a) The laboratory must run serologic tests on patient specimens 
concurrently with a positive serum control of known titer or controls of 
graded reactivity, if applicable, plus a negative control.
    (b) The laboratory must employ controls that evaluate all phases of 
the test system (antigens, complement, erythrocyte indicator systems, 
etc.) to ensure reactivity and uniform dosages when positive and 
negative controls alone are not sufficient.
    (c) The laboratory may not report test results unless the 
predetermined reactivity pattern of the controls is observed.
    (d) All facilities manufacturing blood and blood products for 
transfusion or serving as referral laboratories for these facilities 
must meet--
    (1) The HIV testing requirements of 21 CFR 610.45; and
    (2) Hepatitis testing requirements of 21 CFR 610.40.

Sec. 493.1243  Condition: Chemistry.

    The laboratory must meet the applicable quality control requirements 
in Secs. 493.1201 through 493.1221 and Secs. 493.1245 through 493.1249 
of this subpart for the subspecialties for which it is certified under 
the specialty of chemistry.

Sec. 493.1245  Condition: Routine chemistry.

    To meet the quality control requirements for routine chemistry, the 
laboratory must comply with the applicable requirements in 
Secs. 493.1201 through 493.1221. All quality control activities must be 
documented. In addition, for blood gas analyses, the laboratory must--
    (a) Calibrate or verify calibration according to the manufacturer's 
specifications and with at least the frequency recommended by the 
manufacturer;
    (b) Test one sample of control material each eight hours of testing;
    (c) Use a combination of calibrators and control materials that 
include both low and high values on each day of testing; and
    (d) Include one sample of calibration material or control material 
each time patients are tested unless automated instrumentation 
internally verifies calibration at least every thirty minutes.

Sec. 493.1247  Condition: Endocrinology.

    To meet the quality control requirements for endocrinology, the 
laboratory must comply with the applicable requirements contained in 
Secs. 493.1201 through 493.1221 of this subpart. All quality control 
activities must be documented.

Sec. 493.1249  Condition: Toxicology.

    To meet the quality control requirements for toxicology, the 
laboratory must comply with the applicable requirements in 
Secs. 493.1201 through 493.1221 of this subpart. All quality control 
activities must be documented. In addition, for drug abuse screening 
using thin layer chromatography--

[[Page 894]]

    (a) Each plate must be spotted with at least one sample of 
calibration material containing all drug groups identified by thin layer 
chromatography which the laboratory reports; and
    (b) At least one control sample must be included in each chamber, 
and the control sample must be processed through each step of patient 
testing, including extraction procedures.

Sec. 493.1251  Condition: Urinalysis.

    Except for those tests categorized as waived, to meet the quality 
control requirements for urinalysis, the laboratory must comply with the 
applicable requirements in Secs. 493.1201 through 493.1221.

[58 FR 5232, Jan. 19, 1993]

Sec. 493.1253  Condition: Hematology.

    To meet the quality control requirements for hematology, the 
laboratory must comply with the applicable requirements in 
Secs. 493.1201 through 493.1221 of this subpart and with paragraphs (a) 
through (d) of this section. All quality control activities must be 
documented.
    (a) Cell counts performed manually using a hemocytometer must be 
tested in duplicate. One control is required for each eight hours of 
operation.
    (b) For non-manual hematology testing systems, excluding 
coagulation, the laboratory must include two levels of controls each 
eight hours of operation.
    (c) For all non-manual coagulation testing systems, the laboratory 
must include two levels of control each eight hours of operation and 
each time a change in reagents occurs.
    (d) For manual coagulation tests--
    (1) Each individual performing tests must test two levels of 
controls before testing patient samples and each time a change in 
reagents occurs; and
    (2) Patient and control specimens must be tested in duplicate.

[57 FR 7163, Feb. 28, 1992, as amended at 58 FR 5232, Jan. 19, 1993]

Sec. 493.1255  Condition: Pathology.

    The laboratory must meet the applicable quality control requirements 
in Secs. 493.1201 through 493.1221 and Secs. 493.1257 through 493.1261 
of this subpart for the subspecialties for which it is certified under 
the specialty of pathology. All quality control activities must be 
documented.

Sec. 493.1257  Condition: Cytology.

    To meet the quality control requirements for cytology, the 
laboratory must comply with the applicable requirements in 
Secs. 493.1201 through 493.1221 of this subpart and paragraphs (a) 
through (g) of this section.
    (a) The laboratory must assure that--
    (1) All gynecologic smears are stained using a Papanicolaou or 
modified Papanicolaou staining method;
    (2) Effective measures are taken to prevent cross-contamination 
between gynecologic and nongynecologic specimens during the staining 
process;
    (3) Nongynecologic specimens that have a high potential for cross-
contamination are stained separately from other nongynecologic 
specimens, and the stains are filtered or changed following staining;
    (4) Diagnostic interpretations are not reported on unsatisfactory 
smears; and
    (5) All cytology slide preparations are evaluated on the premises of 
a laboratory certified to conduct testing in the subspecialty of 
cytology.
    (b) The laboratory is responsible for ensuring that--
    (1) Each individual engaged in the evaluation of cytology 
preparations by nonautomated microscopic technique examines no more than 
100 slides (one patient per slide, gynecologic or nongynecologic, or 
both) in a 24 hour period, irrespective of the site or laboratory. This 
limit represents an absolute maximum number of slides and is not to be 
employed as a performance target for each individual. Previously 
examined negative, reactive, reparative, atypical, premalignant or 
malignant gynecologic cases as defined in paragraph (c)(1) of this 
section, previously examined nongynecologic cytology preparations, and 
tissues patholoty slides examined by a technical supervisor qualified 
under Sec. 493.1449 (b) or (k) are not included in the 100 slide limit. 
(For this section, all references to technical supervisor refer to 
individuals qualified under Secs. 493.1449 (b) and (k).);

[[Page 895]]

    (2) For purposes of workload calculations, each slide preparation 
(gynecologic and nongynecologic) made using automated, semi-automated, 
or other liquid-based slide preparatory techniques which result in cell 
dispersion over one-half or less of the total available slide area and 
which is examined by nonautomated microscopic technique counts as one-
half slide.
    (3) Records are maintained of the total number of slides examined by 
each individual during each 24 hour period, irrespective of the site or 
laboratory, and the number of hours each individual spends examining 
slides in the 24 hour period;
    (i) The maximum number of 100 slides described in paragraph (b)(1) 
of this section is examined in no less than an 8 hour workday;
    (ii) For the purposes of establishing workload limits for 
individuals examining slides by nonautomated microscopic technique on 
other than an 8 hour workday basis (includes full-time employees with 
duties other than slide examination and part-time employees), a period 
of 8 hours must be used to prorate the number of slides that may be 
examined. Use the formula--


                   No. of hours examining slides x 100
-------------------------------------------------------------------------
                                    8
-------------------------------------------------------------------------



to determine maximum slide volume to be examined.

    (c) The individual qualified under Secs. 493.1449 (b) or (k) who 
provides technical supervision of cytology must ensure that--
    (1) All gynecologic smears interpreted to be showing reactive or 
reparative changes, atypical squamous or glandular cells of undetermined 
significance, or to be in the premalignant (dysplasia, cervical 
intraepithelial neoplasia or all squamous intraepithelial lesions 
including human papillomavirus-associated changes) or malignant category 
are confirmed by a technical supervisor in cytology. The report must be 
signed to reflect the review or, if a computer report is generated with 
signature, it must reflect an electronic signature authorized by the 
technical supervisor in cytology;
    (2) All nongynecologic cytologic preparations are reviewed by the 
technical supervisor in cytology. The report must be signed to reflect 
technical supervisory review or, if a computer report is generated with 
signature, it must reflect an electronic signature authorized by the 
technical supervisor;
    (3) The slide examination performance of each cytotechnologist is 
evaluated and documented, including performance evaluation through the 
re-examination of normal and negative cases and feedback on the 
reactive, reparative, atypical, malignant or premalignant cases as 
defined in paragraph (c)(1) of this section; and
    (4) A maximum number of slides, not to exceed the maximum workload 
limit described in paragraph (b) of this section is established by the 
technical supervisor for each individual examining slide preparations by 
nonautomated microscopic technique.
    (i) The actual workload limit must be documented for each individual 
and established in accordance with the individual's capability based on 
the performance evaluation as described in paragraph (c)(3) of this 
section.
    (ii) Records are available to document that each individual's 
workload limit is reassessed at least every 6 months and adjusted when 
necessary.
    (d) The laboratory must establish and follow a program designed to 
detect errors in the performance of cytologic examinations and the 
reporting of results.
    (1) The laboratory must establish a program that includes a review 
of slides from at least 10 percent of the gynecologic cases interpreted 
to be negative for reactive, reparative, atypical, premalignant or 
malignant conditions as defined in paragraph (c)(1) of this section that 
are examined by each individual not qualified under Secs. 493.1449 (b) 
or (k). This review must be done by a technical supervisor in cytology, 
a cytology general supervisor qualified under Sec. 493.1469, or a 
cytotechnologist qualified under Sec. 493.1483 who has the experience 
specified in Sec. 493.1469(b)(2).
    (i) The review must include negative cases selected at random from 
the total caseload and from patients or groups of patients that are 
identified as having a

[[Page 896]]

high probability of developing cervical cancer, based on available 
patient information;
    (ii) Records of initial examinations and rescreening results must be 
available; and
    (iii) The review must be completed before reporting patient results 
on those cases selected.
    (2) The laboratory must compare clinical information, when 
available, with cytology reports and must compare all malignant and 
premalignant (as defined in paragraph (c)(1) of this section) gynecology 
reports with the histopathology report, if available in the laboratory 
(either on-site or in storage), and determine the causes of any 
discrepancies.
    (3) For each patient with a current high grade intraepithelial 
lesion or above (moderate dysplasia or CIN-2 or above), the laboratory 
must review all normal or negative gynecologic specimens received within 
the previous five years, if available in the laboratory (either on-site 
or in storage). If significant discrepancies are found that would affect 
patient care, the laboratory must notify the patient's physician and 
issue an amended report.
    (4) The laboratory must establish and document an annual statistical 
evaluation of the number of cytology cases examined, number of specimens 
processed by specimen type, volume of patient cases reported by 
diagnosis (including the number reported as unsatisfactory for 
diagnostic interpretation), number of gynecologic cases where cytology 
and available histology are discrepant, the number of gynecologic cases 
where any rescreen of a normal or negative specimen results in 
reclassification as malignant or premalignant, as defined in paragraph 
(c)(1) of the section, and the number of gynecologic cases for which 
histology results were unavailable to compare with malignant or 
premalignant cytology cases as defined in paragraph (c)(1) of this 
section.
    (5) The laboratory must evaluate the case reviews of each individual 
examining slides against the laboratory's overall statistical values, 
document any discrepancies, including reasons for the deviation, and 
document corrective action, if appropriate.
    (e) The laboratory report must--
    (1) Clearly distinguish specimens or smears, or both, that are 
unsatisfactory for diagnostic interpretation; and
    (2) Contain narrative descriptive nomenclature for all results.
    (f) Corrected reports issued by the laboratory must indicate the 
basis for correction.
    (g) The laboratory must retain all slide preparations for five years 
from the date of examination, or slides may be loaned to proficiency 
testing programs, in lieu of maintaining them for this time period, 
provided the laboratory receives written acknowledgment of the receipt 
of slides by the proficiency testing program and maintains the 
acknowledgment to document the loan of such slides. Documentation for 
slides loaned or referred for purposes other than proficiency testing 
must also be maintained. All slides must be retrievable upon request.

[57 FR 7163, Feb. 28, 1992, as amended at 58 FR 5232, Jan. 19, 1993; 58 
FR 39155, July 22, 1993]

Sec. 493.1259  Condition: Histopathology.

    To meet the quality control requirements for histopathology, a 
laboratory must comply with the applicable requirements in 
Secs. 493.1201 through 493.1221 of this subpart and paragraphs (a) 
through (e) of this section. All quality control activities must be 
documented.
    (a) A control slide of known reactivity must be included with each 
slide or group of slides for differential or special stains. Reaction(s) 
of the control slide with each special stain must be documented.
    (b) The laboratory must retain stained slides at least ten years 
from the date of examination and retain specimen blocks at least two 
years from the date of examination.
    (c) The laboratory must retain remnants of tissue specimens in a 
manner that assures proper preservation of the tissue specimens until 
the portions submitted for microscopic examination have been examined 
and a diagnosis made by an individual qualified under Secs. 493.1449(b) 
or 493.1449(l)(1) of this part. In addition, an individual who meets the 
requirements of Secs. 493.1449(b),

[[Page 897]]

493.1449(l)(1) or 493.1449(l)(2), may examine and provide reports for 
specimens for skin pathology; an individual meeting the requirements of 
Secs. 493.1449(b) or 493.1449(l)(3) may examine and provide reports for 
ophthalmic pathology; an individual meeting the requirements of 
Secs. 493.1449(b) or 493.1449(m) may examine and provide reports for 
oral pathology specimens.
    (d) All tissue pathology reports must be signed by an individual 
qualified as specified in paragraph (c) of the section. If a computer 
report is generated with an electronic signature, it must be authorized 
by the individual qualified as specified in paragraph (c) of this 
section.
    (e) The laboratory must utilize acceptable terminology of a 
recognized system of disease nomenclature in reporting results.

Sec. 493.1261  Condition: Oral pathology.

    To meet the quality control requirements for oral pathology, the 
laboratory must comply with the applicable requirements in 
Secs. 493.1201 through 493.1221 and Sec. 493.1259 of this subpart. All 
quality control activities must be documented.

Sec. 493.1263  Condition: Radiobioassay.

    To meet quality control requirements for radiobioassay, the 
laboratory must comply with the applicable requirements of 
Secs. 493.1201 through 493.1221 of this subpart. All quality control 
activities must be documented.

Sec. 493.1265  Condition: Histocompatibility.

    In addition to meeting the applicable requirements for general 
quality control in Secs. 493.1201 through 493.1221, for quality control 
for general immunology in Sec. 493.1241 of this subpart and for 
immunohematology in Sec. 493.1269 of this subpart, the laboratory must 
comply with the applicable requirements in paragraphs (a) through (d) of 
this section. All quality control activities must be documented.
    (a) For renal allotransplantation, the laboratory must meet the 
following requirements:
    (1) The laboratory must have available and follow criteria for--
    (i) Selecting appropriate patient serum samples for crossmatching;
    (ii) The technique used in crossmatching;
    (iii) Preparation of donor lymphocytes for crossmatching; and
    (iv) Reporting crossmatch results;
    (2) The laboratory must--
    (i) Have available results of final crossmatches before an organ or 
tissue is transplanted; and
    (ii) Make a reasonable attempt and document efforts to have 
available serum specimens for all potential transplant recipients at 
initial typing, for periodic screening, for pre-transplantation 
crossmatch and following sensitizing events, such as transfusion and 
transplant loss;
    (3) The laboratory's storage and maintenance of both recipient sera 
and reagents must--
    (i) Be at an acceptable temperature range for sera and components;
    (ii) Use a temperature alarm system and have an emergency plan for 
alternate storage; and
    (iii) Ensure that all specimens are properly identified and easily 
retrievable;
    (4) The laboratory's reagent typing sera inventory (applicable only 
to locally constructed trays) must indicate source, bleeding date and 
identification number, and volume remaining;
    (5) The laboratory must properly label and store cells, complement, 
buffer, dyes, etc.;
    (6) The laboratory must--
    (i) HLA type all potential transplant recipients;
    (ii) Type cells from organ donors referred to the laboratory; and
    (iii) Have available and follow a policy that establishes when 
antigen redefinition and retyping are required;
    (7) The laboratory must have available and follow criteria for--
    (i) The preparation of lymphocytes for HLA-A, B and DR typing;
    (ii) Selecting typing reagents, whether locally or commercially 
prepared;
    (iii) The assignment of HLA antigens; and
    (iv) Assuring that reagents used for typing recipients and donors 
are adequate to define all major and International Workshop HLA-A,B and 
DR

[[Page 898]]

specificities for which reagents are readily available;
    (8) The laboratory must--
    (i) Screen potential transplant recipient sera for preformed HLA-A 
and B antibodies with a suitable lymphocyte panel on sera collected;
    (A) At the time of the recipient's initial HLA typing; and
    (B) Thereafter, following sensitizing events and upon request; and
    (ii) Use a suitable cell panel for screening patient sera (antibody 
screen), a screen that contains all the major HLA specificities and 
common splits--
    (A) If the laboratory does not use commercial panels, it must 
maintain a list of individuals for fresh panel bleeding; and
    (B) If the laboratory uses frozen panels, it must have a suitable 
storage system;
    (9) The laboratory must check--
    (i) Each typing tray using--
    (A) Positive control sera;
    (B) Negative control sera; and
    (C) Positive controls for specific cell types when applicable (i.e., 
T cells, B cells, and monocytes); and
    (ii) Each compatibility test (i.e. mixed lymphocyte cultures, 
homozygous typing cells or DNA analysis) and typing for disease-
associated antigens using controls to monitor the test components and 
each phase of the test system to ensure an acceptable performance level;
    (10) Compatibility testing for cellularly-defined antigens must 
utilize techniques such as the mixed lymphocyte culture test, homozygous 
typing cells or DNA analysis;
    (11) If the laboratory reports the recipient's or donor's, or both, 
ABO blood group and D(Rho) typing, the testing must be performed in 
accordance with Sec. 493.1269 of this subpart;
    (12) If the laboratory utilizes immunologic reagents (such as 
antibodies or complement) to remove contaminating cells during the 
isolation of lymphocytes or lymphocyte subsets, the efficacy of the 
methods must be verified with appropriate quality control procedures;
    (13) At least once each month, the laboratory must have each 
individual performing tests evaluate a previously tested specimen as an 
unknown to verify his or her ability to reproduce test results. Records 
of the results for each individual must be maintained; and
    (14) The laboratory must participate in at least one national or 
regional cell exchange program, if available, or develop an exchange 
system with another laboratory in order to validate inter-laboratory 
reproducibility.
    (b) If the laboratory performs histocompatibility testing for--
    (1) Transfusions and other non-renal transplantation, excluding bone 
marrow and living transplants, all the requirements specified in this 
section, as applicable, except for the performance of mixed lymphocyte 
cultures, must be met;
    (2) Bone marrow transplantation, all the requirements specified in 
this section, including the performance of mixed lymphocyte cultures or 
other augmented testing to evaluate class II compatibility, must be met; 
and
    (3) Non-renal solid organ transplantation, the results of final 
crossmatches must be available before transplantation when the recipient 
has demonstrated presensitization by prior serum screening except for 
emergency situations. The laboratory must document the circumstances, if 
known, under which emergency transplants are performed, and records must 
reflect any information concerning the transplant provided to the 
laboratory by the patient's physician.
    (c) Laboratories performing HLA typing for disease-associated 
studies must meet all the requirements specified in this section except 
for the performance of mixed lymphocyte cultures, antibody screening and 
crossmatching.
    (d) For laboratories performing organ donor HIV testing the 
requirements of Sec. 493.1241 of this subpart for the transfusion of 
blood and blood products must be met.

[57 FR 7163, Feb. 28, 1992, as amended at 58 FR 5233, Jan. 19, 1993]

Sec. 493.1267  Condition: Clinical cytogenetics.

    To meet the quality control requirements for clinical cytogenetics, 
the laboratory must comply with the applicable requirements of 
Secs. 493.1201

[[Page 899]]

through 493.1221 of this subpart and with paragraphs (a) through (d) of 
this section. All quality control activities must be documented.
    (a) When determination of sex is performed by X and Y chromatin 
counts, these counts must be based on an examination of an adequate 
number of cells. Confirmatory testing such as full chromosome analysis 
must be performed for all atypical results.
    (b) The laboratory must have records that reflect the media used and 
document the reactions observed, number of cells counted, the number of 
cells karyotyped, the number of chromosomes counted for each metaphase 
spread, and the quality of the banding; that the resolution is 
sufficient to support the reported results; and that an adequate number 
of karyotypes are prepared for each patient.
    (c) The laboratory also must have policies and procedures for 
assuring an accurate and reliable patient sample identification during 
the process of accessioning, cell preparation, photographing or other 
image reproduction technique, and photographic printing, and storage and 
reporting of results or photographs.
    (d) The laboratory report must include the summary and 
interpretation of the observations and number of cells counted and 
analyzed and the use of appropriate nomenclature.

Sec. 493.1269  Condition: Immunohematology.

    To meet the quality control requirements for immunohematology, the 
laboratory must comply with the applicable requirements in 
Secs. 493.1201 through 493.1221 of this subpart and with paragraphs (a) 
through (d) of this section. All quality control activities must be 
documented.
    (a) The laboratory must perform ABO group and D(Rho) typing, 
unexpected antibody detection, antibody identification and compatibility 
testing in accordance with manufacturer's instructions, if provided, and 
as applicable, with 21 CFR part 606 (with the exception of 21 CFR 
606.20a, Personnel) and 21 CFR part 640 et seq.
    (b) The laboratory must perform ABO group by concurrently testing 
unknown red cells with anti-A and anti-B grouping reagents. For 
confirmation of ABO group, the unknown serum must be tested with known 
A1 and B red cells.
    (c) The laboratory must determine the D(Rho) type by testing unknown 
red cells with anti-D (anti-Rho) blood grouping reagent.
    (d) If required in the manufacturer's package insert for anti-D 
reagents, the laboratory must employ a control system capable of 
detecting false positive D(Rho) test results.

Sec. 493.1271  Condition: Transfusion services and bloodbanking.

    If a facility provides services for the transfusion of blood and 
blood products, the facility must be under the adequate control and 
technical supervision of the pathologist or other doctor of medicine or 
osteopathy meeting the qualifications in subpart M for technical 
supervision in immunohematology. The facility must ensure that there are 
facilities for procurement, safekeeping and transfusion of blood and 
blood products and that blood and blood products must be available to 
meet the needs of the physicians responsible for the diagnosis, 
management, and treatment of patients. The facility meets this condition 
by complying with the standards in Sec. Sec. 493.1273 through 493.1285.

[58 FR 5233, Jan. 19, 1993]

Sec. 493.1273  Standard; Immunohematological collection, processing, 
          dating periods, labeling and distribution of blood and blood 
          products.

    In addition to the requirements in paragraphs (a) through (d) of 
this section, the facility must also meet the applicable quality control 
requirements in Secs. 493.1201 through 493.1221 of this part.
    (a) Blood and blood product collection, processing and distribution 
must comply with 21 CFR part 640 and 21 CFR part 606, and the testing 
laboratory must meet the applicable requirements of part 493.
    (b) Dating periods for blood and blood products must conform to 21 
CFR 610.53.

[[Page 900]]

    (c) Labeling of blood and blood products must conform to 21 CFR part 
606, subpart G.
    (d) Policies to ensure positive identification of a blood or blood 
product recipient must be established, documented, and followed.

Sec. 493.1275  Standard; Blood and blood products storage facilities.

    (a) The blood and blood products must be stored under appropriate 
conditions, which include an adequate temperature alarm system that is 
regularly inspected.
    (1) An audible alarm system must monitor proper blood and blood 
product storage temperature over a 24-hour period; and
    (2) Inspections of the alarm system must be documented.
    (b) If blood is stored or maintained for transfusion outside of a 
monitored refrigerator, the facility must ensure and document that 
storage conditions, including temperature, are appropriate to prevent 
deterioration of the blood or blood product.

Sec. 493.1277  Standard; Arrangement for services.

    In the case of services provided outside the blood bank, the 
facility must have an agreement reviewed and approved by the director 
that governs the procurement, transfer and availability of blood and 
blood products.

Sec. 493.1279  Standard; Provision of testing.

    There must be provision for prompt ABO blood group, D(Rho) type, 
unexpected antibody detection and compatibility testing in accordance 
with Sec. 493.1269 of this subpart and for laboratory investigation of 
transfusion reactions, either through the facility or under arrangement 
with an approved facility on a continuous basis, under the supervision 
of a pathologist or other doctor of medicine or osteopathy meeting the 
qualifications of Secs. 493.1449(b) or 493.1449(q).

Sec. 493.1283  Standard; Retention of samples of transfused blood.

    According to the facility's established procedures, samples of each 
unit of transfused blood must be retained for further testing in the 
event of reactions. The facility must promptly dispose of blood not 
retained for further testing that has passed its expiration date.

Sec. 493.1285  Standard; Investigation of transfusion reactions.

    The facility, according to its established procedures, must promptly 
investigate all transfusion reactions occurring in all facilities for 
which it has investigational responsibility and make recommendations to 
the medical staff regarding improvements in transfusion procedures. The 
facility must document that all necessary remedial actions are taken to 
prevent future recurrences of transfusion reactions and that all 
policies and procedures are reviewed to assure that they are adequate to 
ensure the safety of individuals being transfused within the facility.

Subpart L [Reserved]

Subpart M--Personnel for Moderate Complexity (Including the Subcategory) 
                       and High Complexity Testing

    Source: 57 FR 7172, Feb. 28, 1992, unless otherwise noted.

Sec. 493.1351  General.

    This subpart consists of the personnel requirements that must be met 
by laboratories performing moderate complexity testing, PPM procedures, 
high complexity testing, or any combination of these tests.

[60 FR 20049, Apr. 24, 1995]

 Laboratories Performing Provider-Performed Microscopy (PPM) Procedures

    Source: 60 FR 20049, Apr. 24, 1995, unless otherwise noted.

Sec. 493.1353  Scope.

    In accordance with Sec. 493.19(b), the moderate complexity 
procedures specified as PPM procedures are considered such only when 
personally performed

[[Page 901]]

by a health care provider during a patient visit in the context of a 
physical examination. PPM procedures are subject to the personnel 
requirements in Secs. 493.1355 through 493.1365.

Sec. 493.1355  Condition: Laboratories performing PPM procedures; 
          laboratory director.

    The laboratory must have a director who meets the qualification 
requirements of Sec. 493.1357 and provides overall management and 
direction in accordance with Sec. 493.1359.

Sec. 493.1357  Standard; laboratory director qualifications.

    The laboratory director must be qualified to manage and direct the 
laboratory personnel and the performance of PPM procedures as specified 
in Sec. 493.19(c) and must be eligible to be an operator of a laboratory 
within the requirements of subpart R of this part.
    (a) The laboratory director must possess a current license as a 
laboratory director issued by the State in which the laboratory is 
located, if the licensing is required.
    (b) The laboratory director must meet one of the following 
requirements:
    (1) Be a physician, as defined in Sec. 493.2.
    (2) Be a midlevel practitioner, as defined in Sec. 493.2, authorized 
by a State to practice independently in the State in which the 
laboratory is located.
    (3) Be a dentist, as defined in Sec. 493.2.

Sec. 493.1359  Standard; PPM laboratory director responsibilities.

    The laboratory director is responsible for the overall operation and 
administration of the laboratory, including the prompt, accurate, and 
proficient reporting of test results. The laboratory director must--
    (a) Direct no more than five laboratories; and
    (b) Ensure that any procedure listed under Sec. 493.19(c)--
    (1) Is personally performed by an individual who meets the 
qualification requirements in Sec. 493.1363; and
    (2) Is performed in accordance with applicable requirements in 
subparts H, J, K, M, and P of this part.

Sec. 493.1361  Condition: Laboratories performing PPM procedures; 
          testing personnel.

    The laboratory must have a sufficient number of individuals who meet 
the qualification requirements of Sec. 493.1363 to perform the functions 
specified in Sec. 493.1365 for the volume and complexity of testing 
performed.

Sec. 493.1363  Standard: PPM testing personnel qualifications.

    Each individual performing PPM procedures must--
    (a) Possess a current license issued by the State in which the 
laboratory is located if the licensing is required; and
    (b) Meet one of the following requirements:
    (1) Be a physician, as defined in Sec. 493.2.
    (2) Be a midlevel practitioner, as defined in Sec. 493.2, under the 
supervision of a physician or in independent practice if authorized by 
the State in which the laboratory is located.
    (3) Be a dentist as defined in Sec. 493.2 of this part.

Sec. 493.1365  Standard; PPM testing personnel responsibilities.

    The testing personnel are responsible for specimen processing, test 
performance, and for reporting test results. Any PPM procedure must be--
    (a) Personally performed by one of the following practitioners:
    (1) A physician during the patient's visit on a specimen obtained 
from his or her own patient or from a patient of a group medical 
practice of which the physician is a member or employee.
    (2) A midlevel practitioner, under the supervision of a physician or 
in independent practice if authorized by the State in which the 
laboratory is located, during the patient's visit on a specimen obtained 
from his or her own patient or from the patient of a clinic, group 
medical practice, or other health care provider, in which the midlevel 
practitioner is a member or an employee.
    (3) A dentist during the patient's visit on a specimen obtained from 
his or her own patient or from a patient of a group dental practice of 
which the dentist is a member or an employee; and

[[Page 902]]

    (b) Performed using a microscope limited to a brightfield or a 
phase/contrast microscope.

           Laboratories Performing Moderate Complexity Testing

Sec. 493.1403  Condition: Laboratories performing moderate complexity 
          testing; laboratory director.

    The laboratory must have a director who meets the qualification 
requirements of Sec. 493.1405 of this subpart and provides overall 
management and direction in accordance with Sec. 493.1407 of this 
subpart.

Sec. 493.1405  Standard; Laboratory director qualifications.

    The laboratory director must be qualified to manage and direct the 
laboratory personnel and the performance of moderate complexity tests 
and must be eligible to be an operator of a laboratory within the 
requirements of subpart R of this part.
    (a) The laboratory director must possess a current license as a 
laboratory director issued by the State in which the laboratory is 
located, if such licensing is required; and
    (b) The laboratory director must--
    (1) (i) Be a doctor of medicine or doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located; and
    (ii) Be certified in anatomic or clinical pathology, or both, by the 
American Board of Pathology or the American Osteopathic Board of 
Pathology or possess qualifications that are equivalent to those 
required for such certification; or
    (2)(i) Be a doctor of medicine, doctor of osteopathy, or doctor of 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located; and
    (ii) Have had laboratory training or experience consisting of:
    (A) At least one year directing or supervising non-waived laboratory 
testing; or
    (B) Beginning September 1, 1993, have at least 20 continuing medical 
education credit hours in laboratory practice commensurate with the 
director responsibilities defined in Sec. 493.1407; or
    (C) Laboratory training equivalent to paragraph (b)(2)(ii)(B) of 
this section obtained during medical residency. (For example, physicians 
certified either in hematology or hematology and medical oncology by the 
American Board of Internal Medicine); or
    (3) Hold an earned doctoral degree in a chemical, physical, 
biological, or clinical laboratory science from an accredited 
institution; and
    (i) Be certified by the American Board of Medical Microbiology, the 
American Board of Clinical Chemistry, the American Board of Bioanalysis, 
or the American Board of Medical Laboratory Immunology; or
    (ii) Have had at least one year experience directing or supervising 
non-waived laboratory testing;
    (4)(i) Have earned a master's degree in a chemical, physical, 
biological or clinical laboratory science or medical technology from an 
accredited institution;
    (ii) Have at least one year of laboratory training or experience, or 
both in non-waived testing; and
    (iii) In addition, have at least one year of supervisory laboratory 
experience in non-waived testing; or
    (5)(i) Have earned a bachelor's degree in a chemical, physical, or 
biological science or medical technology from an accredited institution;
    (ii) Have at least 2 years of laboratory training or experience, or 
both in non-waived testing; and
    (iii) In addition, have at least 2 years of supervisory laboratory 
experience in non-waived testing;
    (6) Be serving as a laboratory director and must have previously 
qualified or could have qualified as a laboratory director under 
Sec. 493.1406; or
    (7) On or before February 28, 1992, qualified under State law to 
direct a laboratory in the State in which the laboratory is located.

[57 FR 7172, Feb. 28, 1992, as amended at 58 FR 5233, Jan. 19, 1993]

Sec. 493.1406  Standard; Laboratory director qualifications on or before 
          February 28, 1992.

    The laboratory director must be qualified to manage and direct the 
laboratory personnel and test performance.
    (a) The laboratory director must possess a current license as a 
laboratory

[[Page 903]]

director issued by the State, if such licensing exists; and
    (b) The laboratory director must:
    (1) Be a physician certified in anatomical or clinical pathology (or 
both) by the American Board of Pathology or the American Osteopathic 
Board of Pathology or possess qualifications that are equivalent to 
those required for such certification;
    (2) Be a physician who:
    (i) Is certified by the American Board of Pathology or the American 
Osteopathic Board of Pathology in at least one of the laboratory 
specialties; or
    (ii) Is certified by the American Board of Medical Microbiology, the 
American Board of Clinical Chemistry, the American Board of Bioanalysis, 
or other national accrediting board in one of the laboratory 
specialties; or
    (iii) Is certified by the American Society of Cytology to practice 
cytopathology or possesses qualifications that are equivalent to those 
required for such certification; or
    (iv) Subsequent to graduation, has had 4 or more years of full-time 
general laboratory training and experience of which at least 2 years 
were spent acquiring proficiency in one of the laboratory specialties;
    (3) For the subspecialty of oral pathology only, be certified by the 
American Board of Oral Pathology, American Board of Pathology or the 
American Osteopathic Board of Pathology or possesses qualifications that 
are equivalent to those required for certification;
    (4) Hold an earned doctoral degree from an accredited institution 
with a chemical, physical, or biological science as a major subject and
    (i) Is certified by the American Board of Medical Microbiology, the 
American Board of Clinical Chemistry, the American Board of Bioanalysis, 
or other national accrediting board acceptable to HHS in one of the 
laboratory specialties; or
    (ii) Subsequent to graduation, has had 4 or more years of full-time 
general laboratory training and experience of which at least 2 years 
were spent acquiring proficiency in one of the laboratory specialties;
    (5) With respect to individuals first qualifying before July 1, 
1971, have been responsible for the direction of a laboratory for 12 
months between July 1, 1961, and January 1, 1968, and, in addition, 
either:
    (i) Was a physician and subsequent to graduation had at least 4 
years of pertinent full-time laboratory experience;
    (ii) Held a master's degree from an accredited institution with a 
chemical, physical, or biological science as a major subject and 
subsequent to graduation had at least 4 years of pertinent full-time 
laboratory experience;
    (iii) Held a bachelor's degree from an accredited institution with a 
chemical, physical, or biological science as a major subject and 
subsequent to graduation had at least 6 years of pertinent full-time 
laboratory experience; or
    (iv) Achieved a satisfactory grade through an examination conducted 
by or under the sponsorship of the U.S. Public Health Service on or 
before July 1, 1970; or
    (6) Qualify under State law to direct the laboratory in the State in 
which the laboratory is located.

    Note: The January 1, 1968 date for meeting the 12 months' laboratory 
direction requirement in paragraph (b)(5) of this section may be 
extended 1 year for each year of full-time laboratory experience 
obtained before January 1, 1958 required by State law for a laboratory 
director license. An exception to the July 1, 1971 qualifying date in 
paragraph (b)(5) of this section was made provided that the individual 
requested qualification approval by October 21, 1975 and had been 
employed in a laboratory for at least 3 years of the 5 years preceding 
the date of submission of his qualifications.

[58 FR 5233, Jan. 19, 1993]

Sec. 493.1407  Standard; Laboratory director responsibilities.

    The laboratory director is responsible for the overall operation and 
administration of the laboratory, including the employment of personnel 
who are competent to perform test procedures, and record and report test 
results promptly, accurate, and proficiently and for assuring compliance 
with the applicable regulations.
    (a) The laboratory director, if qualified, may perform the duties of 
the technical consultant, clinical consultant, and testing personnel, or 
delegate these responsibilities to personnel

[[Page 904]]

meeting the qualifications of Secs. 493.1409, 493.1415, and 493.1421, 
respectively.
    (b) If the laboratory director reapportions performance of his or 
her responsibilities, he or she remains responsible for ensuring that 
all duties are properly performed.
    (c) The laboratory director must be accessible to the laboratory to 
provide onsite, telephone or electronic consultation as needed.
    (d) Each individual may direct no more than five laboratories.
    (e) The laboratory director must--
    (1) Ensure that testing systems developed and used for each of the 
tests performed in the laboratory provide quality laboratory services 
for all aspects of test performance, which includes the preanalytic, 
analytic, and postanalytic phases of testing;
    (2) Ensure that the physical plant and environmental conditions of 
the laboratory are appropriate for the testing performed and provide a 
safe environment in which employees are protected from physical, 
chemical, and biological hazards;
    (3) Ensure that--
    (i) The test methodologies selected have the capability of providing 
the quality of results required for patient care;
    (ii) Verification procedures used are adequate to determine the 
accuracy, precision, and other pertinent performance characteristics of 
the method; and
    (iii) Laboratory personnel are performing the test methods as 
required for accurate and reliable results;
    (4) Ensure that the laboratory is enrolled in an HHS approved 
proficiency testing program for the testing performed and that--
    (i) The proficiency testing samples are tested as required under 
subpart H of this part;
    (ii) The results are returned within the timeframes established by 
the proficiency testing program;
    (iii) All proficiency testing reports received are reviewed by the 
appropriate staff to evaluate the laboratory's performance and to 
identify any problems that require corrective action; and
    (iv) An approved corrective action plan is followed when any 
proficiency testing results are found to be unacceptable or 
unsatisfactory;
    (5) Ensure that the quality control and quality assurance programs 
are established and maintained to assure the quality of laboratory 
services provided and to identify failures in quality as they occur;
    (6) Ensure the establishment and maintenance of acceptable levels of 
analytical performance for each test system;
    (7) Ensure that all necessary remedial actions are taken and 
documented whenever significant deviations from the laboratory's 
established performance specifications are identified, and that patient 
test results are reported only when the system is functioning properly;
    (8) Ensure that reports of test results include pertinent 
information required for interpretation;
    (9) Ensure that consultation is available to the laboratory's 
clients on matters relating to the quality of the test results reported 
and their interpretation concerning specific patient conditions;
    (10) Employ a sufficient number of laboratory personnel with the 
appropriate education and either experience or training to provide 
appropriate consultation, properly supervise and accurately perform 
tests and report test results in accordance with the personnel 
responsibilities described in this subpart;
    (11) Ensure that prior to testing patients' specimens, all personnel 
have the appropriate education and experience, receive the appropriate 
training for the type and complexity of the services offered, and have 
demonstrated that they can perform all testing operations reliably to 
provide and report accurate results;
    (12) Ensure that policies and procedures are established for 
monitoring individuals who conduct preanalytical, analytical, and 
postanalytical phases of testing to assure that they are competent and 
maintain their competency to process specimens, perform test procedures 
and report test results promptly and proficiently, and whenever 
necessary, identify needs for remedial training or continuing education 
to improve skills;

[[Page 905]]

    (13) Ensure that an approved procedure manual is available to all 
personnel responsible for any aspect of the testing process; and
    (14) Specify, in writing, the responsibilities and duties of each 
consultant and each person, engaged in the performance of the 
preanalytic, analytic, and postanalytic phases of testing, that 
identifies which examinations and procedures each individual is 
authorized to perform, whether supervision is required for specimen 
processing, test performance or results reporting, and whether 
consultant or director review is required prior to reporting patient 
test results.

Sec. 493.1409  Condition: Laboratories performing moderate complexity 
          testing; technical consultant.

    The laboratory must have a technical consultant who meets the 
qualification requirements of Sec. 493.1411 of this subpart and provides 
technical oversight in accordance with Sec. 493.1413 of this subpart.

Sec. 493.1411  Standard; Technical consultant qualifications.

    The laboratory must employ one or more individuals who are qualified 
by education and either training or experience to provide technical 
consultation for each of the specialties and subspecialties of service 
in which the laboratory performs moderate complexity tests or 
procedures. The director of a laboratory performing moderate complexity 
testing may function as the technical consultant provided he or she 
meets the qualifications specified in this section.
    (a) The technical consultant must possess a current license issued 
by the State in which the laboratory is located, if such licensing is 
required.
    (b) The technical consultant must--
    (1) (i) Be a doctor of medicine or doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located; and
    (ii) Be certified in anatomic or clinical pathology, or both, by the 
American Board of Pathology or the American Osteopathic Board of 
Pathology or possess qualifications that are equivalent to those 
required for such certification; or
    (2)(i) Be a doctor of medicine, doctor of osteopathy, or doctor of 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located; and
    (ii) Have at least one year of laboratory training or experience, or 
both in non-waived testing, in the designated specialty or subspecialty 
areas of service for which the technical consultant is responsible (for 
example, physicians certified either in hematology or hematology and 
medical oncology by the American Board of Internal Medicine are 
qualified to serve as the technical consultant in hematology); or
    (3)(i) Hold an earned doctoral or master's degree in a chemical, 
physical, biological or clinical laboratory science or medical 
technology from an accredited institution; and
    (ii) Have at least one year of laboratory training or experience, or 
both in non-waived testing, in the designated specialty or subspecialty 
areas of service for which the technical consultant is responsible; or
    (4)(i) Have earned a bachelor's degree in a chemical, physical or 
biological science or medical technology from an accredited institution; 
and
    (ii) Have at least 2 years of laboratory training or experience, or 
both in non-waived testing, in the designated specialty or subspecialty 
areas of service for which the technical consultant is responsible.

    Note: The technical consultant requirements for ``laboratory 
training or experience, or both'' in each specialty or subspecialty may 
be acquired concurrently in more than one of the specialties or 
subspecialties of service, excluding waived tests. For example, an 
individual who has a bachelor's degree in biology and additionally has 
documentation of 2 years of work experience performing tests of moderate 
complexity in all specialties and subspecialties of service, would be 
qualified as a technical consultant in a laboratory performing moderate 
complexity testing in all specialties and subspecialties of service.

[57 FR 7172, Feb. 28, 1992, as amended at 58 FR 5234, Jan. 19, 1993]

Sec. 493.1413  Standard; Technical consultant responsibilities.

    The technical consultant is responsible for the technical and 
scientific

[[Page 906]]

oversight of the laboratory. The technical consultant is not required to 
be onsite at all times testing is performed; however, he or she must be 
available to the laboratory on an as needed basis to provide 
consultation, as specified in paragraph (a) of this section.
    (a) The technical consultant must be accessible to the laboratory to 
provide on-site, telephone, or electronic consultation; and
    (b) The technical consultant is responsible for--
    (1) Selection of test methodology appropriate for the clinical use 
of the test results;
    (2) Verification of the test procedures performed and the 
establishment of the laboratory's test performance characteristics, 
including the precision and accuracy of each test and test system;
    (3) Enrollment and participation in an HHS approved proficiency 
testing program commensurate with the services offered;
    (4) Establishing a quality control program appropriate for the 
testing performed and establishing the parameters for acceptable levels 
of analytic performance and ensuring that these levels are maintained 
throughout the entire testing process from the initial receipt of the 
specimen, through sample analysis and reporting of test results;
    (5) Resolving technical problems and ensuring that remedial actions 
are taken whenever test systems deviate from the laboratory's 
established performance specifications;
    (6) Ensuring that patient test results are not reported until all 
corrective actions have been taken and the test system is functioning 
properly;
    (7) Identifying training needs and assuring that each individual 
performing tests receives regular in-service training and education 
appropriate for the type and complexity of the laboratory services 
performed;
    (8) Evaluating the competency of all testing personnel and assuring 
that the staff maintain their competency to perform test procedures and 
report test results promptly, accurately and proficiently. The 
procedures for evaluation of the competency of the staff must include, 
but are not limited to--
    (i) Direct observations of routine patient test performance, 
including patient preparation, if applicable, specimen handling, 
processing and testing;
    (ii) Monitoring the recording and reporting of test results;
    (iii) Review of intermediate test results or worksheets, quality 
control records, proficiency testing results, and preventive maintenance 
records;
    (iv) Direct observation of performance of instrument maintenance and 
function checks;
    (v) Assessment of test performance through testing previously 
analyzed specimens, internal blind testing samples or external 
proficiency testing samples; and
    (vi) Assessment of problem solving skills; and
    (9) Evaluating and documenting the performance of individuals 
responsible for moderate complexity testing at least semiannually during 
the first year the individual tests patient specimens. Thereafter, 
evaluations must be performed at least annually unless test methodology 
or instrumentation changes, in which case, prior to reporting patient 
test results, the individual's performance must be reevaluated to 
include the use of the new test methodology or instrumentation.

Sec. 493.1415  Condition: Laboratories performing moderate complexity 
          testing; clinical consultant.

    The laboratory must have a clinical consultant who meets the 
qualification requirements of Sec. 493.1417 of this part and provides 
clinical consultation in accordance with Sec. 493.1419 of this part.

Sec. 493.1417  Standard; Clinical consultant qualifications.

    The clinical consultant must be qualified to consult with and render 
opinions to the laboratory's clients concerning the diagnosis, treatment 
and management of patient care. The clinical consultant must--
    (a) Be qualified as a laboratory director under Sec. 493.1405(b) 
(1), (2), or (3)(i); or
    (b) Be a doctor of medicine, doctor of osteopathy or doctor of 
podiatric medicine and possess a license to practice medicine, 
osteopathy or podiatry in

[[Page 907]]

the State in which the laboratory is located.

[57 FR 7172, Feb. 28, 1992, as amended at 58 FR 5234, Jan. 19, 1993]

Sec. 493.1419  Standard; Clinical consultant responsibilities.

    The clinical consultant provides consultation regarding the 
appropriateness of the testing ordered and interpretation of test 
results. The clinical consultant must--
    (a) Be available to provide clinical consultation to the 
laboratory's clients;
    (b) Be available to assist the laboratory's clients in ensuring that 
appropriate tests are ordered to meet the clinical expectations;
    (c) Ensure that reports of test results include pertinent 
information required for specific patient interpretation; and
    (d) Ensure that consultation is available and communicated to the 
laboratory's clients on matters related to the quality of the test 
results reported and their interpretation concerning specific patient 
conditions.

Sec. 493.1421  Condition: Laboratories performing moderate complexity 
          testing; testing personnel.

    The laboratory must have a sufficient number of individuals who meet 
the qualification requirements of Sec. 493.1423, to perform the 
functions specified in Sec. 493.1425 for the volume and complexity of 
tests performed.

Sec. 493.1423  Standard; Testing personnel qualifications.

    Each individual performing moderate complexity testing must--
    (a) Possess a current license issued by the State in which the 
laboratory is located, if such licensing is required; and
    (b) Meet one of the following requirements:
    (1) Be a doctor of medicine or doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located or have earned a doctoral, master's, or bachelor's degree in a 
chemical, physical, biological or clinical laboratory science, or 
medical technology from an accredited institution; or
    (2) Have earned an associate degree in a chemical, physical or 
biological science or medical laboratory technology from an accredited 
institution; or
    (3) Be a high school graduate or equivalent and have successfully 
completed an official military medical laboratory procedures course of 
at least 50 weeks duration and have held the military enlisted 
occupational specialty of Medical Laboratory Specialist (Laboratory 
Technician); or
    (4)(i) Have earned a high school diploma or equivalent; and
    (ii) Have documentation of training appropriate for the testing 
performed prior to analyzing patient specimens. Such training must 
ensure that the individual has--
    (A) The skills required for proper specimen collection, including 
patient preparation, if applicable, labeling, handling, preservation or 
fixation, processing or preparation, transportation and storage of 
specimens;
    (B) The skills required for implementing all standard laboratory 
procedures;
    (C) The skills required for performing each test method and for 
proper instrument use;
    (D) The skills required for performing preventive maintenance, 
troubleshooting and calibration procedures related to each test 
performed;
    (E) A working knowledge of reagent stability and storage;
    (F) The skills required to implement the quality control policies 
and procedures of the laboratory;
    (G) An awareness of the factors that influence test results; and
    (H) The skills required to assess and verify the validity of patient 
test results through the evaluation of quality control sample values 
prior to reporting patient test results.

[57 FR 7172, Feb. 28, 1992, as amended at 58 FR 5234, Jan. 19, 1993]

Sec. 493.1425  Standard; Testing personnel responsibilities.

    The testing personnel are responsible for specimen processing, test 
performance, and for reporting test results.
    (a) Each individual performs only those moderate complexity tests 
that

[[Page 908]]

are authorized by the laboratory director and require a degree of skill 
commensurate with the individual's education, training or experience, 
and technical abilities.
    (b) Each individual performing moderate complexity testing must--
    (1) Follow the laboratory's procedures for specimen handling and 
processing, test analyses, reporting and maintaining records of patient 
test results;
    (2) Maintain records that demonstrate that proficiency testing 
samples are tested in the same manner as patient samples;
    (3) Adhere to the laboratory's quality control policies, document 
all quality control activities, instrument and procedural calibrations 
and maintenance performed;
    (4) Follow the laboratory's established corrective action policies 
and procedures whenever test systems are not within the laboratory's 
established acceptable levels of performance;
    (5) Be capable of identifying problems that may adversely affect 
test performance or reporting of test results and either must correct 
the problems or immediately notify the technical consultant, clinical 
consultant or director; and
    (6) Document all corrective actions taken when test systems deviate 
from the laboratory's established performance specifications.

             Laboratories Performing High Complexity Testing

Sec. 493.1441  Condition: Laboratories performing high complexity 
          testing; laboratory director.

    The laboratory must have a director who meets the qualification 
requirements of Sec. 493.1443 of this subpart and provides overall 
management and direction in accordance with Sec. 493.1445 of this 
subpart.

Sec. 493.1443  Standard; Laboratory director qualifications.

    The laboratory director must be qualified to manage and direct the 
laboratory personnel and performance of high complexity tests and must 
be eligible to be an operator of a laboratory within the requirements of 
subpart R.
    (a) The laboratory director must possess a current license as a 
laboratory director issued by the State in which the laboratory is 
located, if such licensing is required; and
    (b) The laboratory director must--
    (1)(i) Be a doctor of medicine or doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located; and
    (ii) Be certified in anatomic or clinical pathology, or both, by the 
American Board of Pathology or the American Osteopathic Board of 
Pathology or possess qualifications that are equivalent to those 
required for such certification; or
    (2) Be a doctor of medicine, a doctor of osteopathy or doctor of 
podiatric medicine licensed to practice medicine, osteopathy or podiatry 
in the State in which the laboratory is located; and
    (i) Have at least one year of laboratory training during medical 
residency (for example, physicians certified either in hematology or 
hematology and medical oncology by the American Board of Internal 
Medicine); or
    (ii) Have at least 2 years of experience directing or supervising 
high complexity testing; or
    (3) Hold an earned doctoral degree in a chemical, physical, 
biological or clinical laboratory science from an accredited institution 
and--
    (i) Be certified by the American Board of Medical Microbiology, the 
American Board of Clinical Chemistry, the American Board of Bioanalysis, 
the American Board of Medical Laboratory Immunology or other board 
deemed comparable by HHS; or
    (ii) Until December 31, 2000, must have at least--
    (A) Two years of laboratory training or experience, or both;
    (B) Two years of experience directing or supervising high complexity 
testing; and
    (C) On December 31, 2000, individuals must meet the qualifications 
specified in paragraph (b)(3)(i) of this section;
    (4) Be serving as a laboratory director and must have previously 
qualified or could have qualified as a laboratory director under 
regulations at 42 CFR 493.1415, published March 14, 1990 at 55 FR 9538, 
on or before February 28, 1992; or

[[Page 909]]

    (5) On or before February 28, 1992, be qualified under State law to 
direct a laboratory in the State in which the laboratory is located; or
    (6) For the subspecialty of oral pathology, be certified by the 
American Board of Oral Pathology, American Board of Pathology, the 
American Osteopathic Board of Pathology, or possess qualifications that 
are equivalent to those required for certification.

[57 FR 7172, Feb. 28, 1992, as amended at 58 FR 5234, Jan. 19, 1993; 59 
FR 62609, Dec. 6, 1994; 62 FR 25858, May 12, 1997; 63 FR 55034, Oct. 14, 
1998]

Sec. 493.1445  Standard; Laboratory director responsibilities.

    The laboratory director is responsible for the overall operation and 
administration of the laboratory, including the employment of personnel 
who are competent to perform test procedures, record and report test 
results promptly, accurately and proficiently, and for assuring 
compliance with the applicable regulations.
    (a) The laboratory director, if qualified, may perform the duties of 
the technical supervisor, clinical consultant, general supervisor, and 
testing personnel, or delegate these responsibilities to personnel 
meeting the qualifications under Secs. 493.1447, 493.1453, 493.1459, and 
493.1487, respectively.
    (b) If the laboratory director reapportions performance of his or 
her responsibilities, he or she remains responsible for ensuring that 
all duties are properly performed.
    (c) The laboratory director must be accessible to the laboratory to 
provide onsite, telephone or electronic consultation as needed.
    (d) Each individual may direct no more than five laboratories.
    (e) The laboratory director must--
    (1) Ensure that testing systems developed and used for each of the 
tests performed in the laboratory provide quality laboratory services 
for all aspects of test performance, which includes the preanalytic, 
analytic, and postanalytic phases of testing;
    (2) Ensure that the physical plant and environmental conditions of 
the laboratory are appropriate for the testing performed and provide a 
safe environment in which employees are protected from physical, 
chemical, and biological hazards;
    (3) Ensure that--
    (i) The test methodologies selected have the capability of providing 
the quality of results required for patient care;
    (ii) Verification procedures used are adequate to determine the 
accuracy, precision, and other pertinent performance characteristics of 
the method; and
    (iii) Laboratory personnel are performing the test methods as 
required for accurate and reliable results;
    (4) Ensure that the laboratory is enrolled in an HHS-approved 
proficiency testing program for the testing performed and that--
    (i) The proficiency testing samples are tested as required under 
subpart H of this part;
    (ii) The results are returned within the timeframes established by 
the proficiency testing program;
    (iii) All proficiency testing reports received are reviewed by the 
appropriate staff to evaluate the laboratory's performance and to 
identify any problems that require corrective action; and
    (iv) An approved corrective action plan is followed when any 
proficiency testing result is found to be unacceptable or 
unsatisfactory;
    (5) Ensure that the quality control and quality assurance programs 
are established and maintained to assure the quality of laboratory 
services provided and to identify failures in quality as they occur;
    (6) Ensure the establishment and maintenance of acceptable levels of 
analytical performance for each test system;
    (7) Ensure that all necessary remedial actions are taken and 
documented whenever significant deviations from the laboratory's 
established performance characteristics are identified, and that patient 
test results are reported only when the system is functioning properly;
    (8) Ensure that reports of test results include pertinent 
information required for interpretation;
    (9) Ensure that consultation is available to the laboratory's 
clients on matters relating to the quality of the test

[[Page 910]]

results reported and their interpretation concerning specific patient 
conditions;
    (10) Ensure that a general supervisor provides on-site supervision 
of high complexity test performance by testing personnel qualified under 
Sec. 493.1489(b)(4);
    (11) Employ a sufficient number of laboratory personnel with the 
appropriate education and either experience or training to provide 
appropriate consultation, properly supervise and accurately perform 
tests and report test results in accordance with the personnel 
responsibilities described in this subpart;
    (12) Ensure that prior to testing patients' specimens, all personnel 
have the appropriate education and experience, receive the appropriate 
training for the type and complexity of the services offered, and have 
demonstrated that they can perform all testing operations reliably to 
provide and report accurate results;
    (13) Ensure that policies and procedures are established for 
monitoring individuals who conduct preanalytical, analytical, and 
postanalytical phases of testing to assure that they are competent and 
maintain their competency to process specimens, perform test procedures 
and report test results promptly and proficiently, and whenever 
necessary, identify needs for remedial training or continuing education 
to improve skills;
    (14) Ensure that an approved procedure manual is available to all 
personnel responsible for any aspect of the testing process; and
    (15) Specify, in writing, the responsibilities and duties of each 
consultant and each supervisor, as well as each person engaged in the 
performance of the preanalytic, analytic, and postanalytic phases of 
testing, that identifies which examinations and procedures each 
individual is authorized to perform, whether supervision is required for 
specimen processing, test performance or result reporting and whether 
supervisory or director review is required prior to reporting patient 
test results.

Sec. 493.1447  Condition: Laboratories performing high complexity 
          testing; technical supervisor.

    The laboratory must have a technical supervisor who meets the 
qualification requirements of Sec. 493.1449 of this subpart and provides 
technical supervision in accordance with Sec. 493.1451 of this subpart.

Sec. 493.1449  Standard; Technical supervisor qualifications.

    The laboratory must employ one or more individuals who are qualified 
by education and either training or experience to provide technical 
supervision for each of the specialties and subspecialties of service in 
which the laboratory performs high complexity tests or procedures. The 
director of a laboratory performing high complexity testing may function 
as the technical supervisor provided he or she meets the qualifications 
specified in this section.
    (a) The technical supervisor must possess a current license issued 
by the State in which the laboratory is located, if such licensing is 
required; and
    (b) The laboratory may perform anatomic and clinical laboratory 
procedures and tests in all specialties and subspecialties of services 
except histocompatibility and clinical cytogenetics services provided 
the individual functioning as the technical supervisor--
    (1) Is a doctor of medicine or doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located; and
    (2) Is certified in both anatomic and clinical pathology by the 
American Board of Pathology or the American Osteopathic Board of 
Pathology or Possesses qualifications that are equivalent to those 
required for such certification.
    (c) If the requirements of paragraph (b) of this section are not met 
and the laboratory performs tests in the subspecialty of bacteriology, 
the individual functioning as the technical supervisor must--
    (1)(i) Be a doctor of medicine or doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located; and
    (ii) Be certified in clinical pathology by the American Board of 
Pathology or

[[Page 911]]

the American Osteopathic Board of Pathology or possess qualifications 
that are equivalent to those required for such certification; or
    (2)(i) Be a doctor of medicine, doctor of osteopathy, or doctor of 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located; and
    (ii) Have at least one year of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of bacteriology; or
    (3)(i) Have an earned doctoral degree in a chemical, physical, 
biological or clinical laboratory science from an accredited 
institution; and
    (ii) Have at least 1 year of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of bacteriology; or
    (4)(i) Have earned a master's degree in a chemical, physical, 
biological or clinical laboratory science or medical technology from an 
accredited institution; and
    (ii) Have at least 2 years of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of bacteriology; or
    (5)(i) Have earned a bachelor's degree in a chemical, physical, or 
biological science or medical technology from an accredited institution; 
and
    (ii) Have at least 4 years of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of bacteriology.
    (d) If the requirements of paragraph (b) of this section are not met 
and the laboratory performs tests in the subspecialty of 
mycobacteriology, the individual functioning as the technical supervisor 
must--
    (1)(i) Be a doctor of medicine or doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located; and
    (ii) Be certified in clinical pathology by the American Board of 
Pathology or the American Osteopathic Board of Pathology or possess 
qualifications that are equivalent to those required for such 
certification; or
    (2)(i) Be a doctor of medicine, doctor of osteopathy, or doctor or 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located; and
    (ii) Have at least 1 year of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of mycobacteriology; or
    (3)(i) Have an earned doctoral degree in a chemical, physical, 
biological or clinical laboratory science from an accredited 
institution; and
    (ii) Have at least 1 year of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of mycobacteriology; or
    (4)(i) Have earned a master's degree in a chemical, physical, 
biological or clinical laboratory science or medical technology from an 
accredited institution; and
    (ii) Have at least 2 years of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of mycobacteriology; or
    (5)(i) Have earned a bachelor's degree in a chemical, physical or 
biological science or medical technology from an accredited institution; 
and
    (ii) Have at least 4 years of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of mycobacteriology.
    (e) If the requirements of paragraph (b) of this section are not met 
and the laboratory performs tests in the subspecialty of mycology, the 
individual

[[Page 912]]

functioning as the technical supervisor must--
    (1)(i) Be a doctor of medicine or doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located; and
    (ii) Be certified in clinical pathology by the American Board of 
Pathology or the American osteopathic Board of Pathology or possess 
qualifications that are equivalent to those required for such 
certification; or
    (2)(i) Be a doctor of medicine, doctor of osteopathy, or doctor of 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located; and
    (ii) Have at least 1 year of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of mycology; or
    (3)(i) Have an earned doctoral degree in a chemical, physical, 
biological or clinical laboratory science from an accredited 
institution; and
    (ii) Have at least 1 year of laboratory training or experience, or 
both in high complexity testing within the speciality of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of mycology; or
    (4)(i) Have earned a master's degree in a chemical, physical, 
biological or clinical laboratory science or medical technology from an 
accredited institution; and
    (ii) Have at least 2 years of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of mycology; or
    (5)(i) Have earned a bachelor's degree in a chemical, physical or 
biological science or medical technology from an accredited institution; 
and
    (ii) Have at least 4 years of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of mycology.
    (f) If the requirements of paragraph (b) of this section are not met 
and the laboratory performs tests in the subspecialty of parasitology, 
the individual functioning as the technical supervisor must--
    (1)(i) Be a doctor of medicine or a doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located; and
    (ii) Be certified in clinical pathology by the American Board of 
Pathology or the American Osteopathic Board of Pathology or possess 
qualifications that are equivalent to those required for such 
certification; or
    (2)(i) Be a doctor of medicine, doctor of osteopathy, or doctor of 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located; and
    (ii) Have at least one year of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of parasitology;
    (3)(i) Have an earned doctoral degree in a chemical, physical, 
biological or clinical laboratory science from an accredited 
institution; and
    (ii) Have at least 1 year of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of parasitology; or
    (4)(i) Have earned a master's degree in a chemical, physical, 
biological or clinical laboratory science or medical technology from an 
accredited institution; and
    (ii) Have at least 2 years of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of parasitology; or
    (5)(i) Have earned a bachelor's degree in a chemical, physical or 
biological science or medical technology from an accredited institution; 
and
    (ii) Have at least 4 years of laboratory training or experience, or 
both, in

[[Page 913]]

high complexity testing within the specialty of microbiology with a 
minimum of 6 months experience in high complexity testing within the 
subspecialty of parasitology.
    (g) If the requirements of paragraph (b) of this section are not met 
and the laboratory performs tests in the subspecialty of virology, the 
individual functioning as the technical supervisor must--
    (1)(i) Be a doctor of medicine or doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located; and
    (ii) Be certified in clinical pathology by the American Board of 
Pathology or the American Osteopathic Board of Pathology or possess 
qualifications that are equivalent to those required for such 
certification; or
    (2)(i) Be a doctor of medicine, doctor of osteopathy, or doctor of 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located; and
    (ii) Have at least 1 year of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of virology; or
    (3)(i) Have an earned doctoral degree in a chemical, physical, 
biological or clinical laboratory science from an accredited 
institution; and
    (ii) Have at least 1 year of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of virology; or
    (4)(i) Have earned a master's degree in a chemical, physical, 
biological or clinical laboratory science or medical technology from an 
accredited institution; and
    (ii) Have at least 2 years of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of virology; or
    (5)(i) Have earned a bachelor's degree in a chemical, physical or 
biological science or medical technology from an accredited institution; 
and
    (ii) Have at least 4 years of laboratory training or experience, or 
both, in high complexity testing within the specialty of microbiology 
with a minimum of 6 months experience in high complexity testing within 
the subspecialty of virology.
    (h) If the requirements of paragraph (b) of this section are not met 
and the laboratory performs tests in the specialty of diagnostic 
immunology, the individual functioning as the technical supervisor 
must--
    (1)(i) Be a doctor of medicine or a doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located; and
    (ii) Be certified in clinical pathology by the American Board of 
Pathology or the American Osteopathic Board of Pathology or possess 
qualifications that are equivalent to those required for such 
certification; or
    (2)(i) Be a doctor of medicine, doctor of osteopathy, or doctor of 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located; and
    (ii) Have at least 1 year of laboratory training or experience, or 
both, in high complexity testing for the specialty of diagnostic 
immunology; or
    (3)(i) Have an earned doctoral degree in a chemical, physical, 
biological or clinical laboratory science from an accredited 
institution; and
    (ii) Have at least 1 year of laboratory training or experience, or 
both, in high complexity testing within the specialty of diagnostic 
immunology; or
    (4)(i) Have earned a master's degree in a chemical, physical, 
biological or clinical laboratory science or medical technology from an 
accredited institution; and
    (ii) Have at least 2 years of laboratory training or experience, or 
both, in high complexity testing for the specialty of diagnostic 
immunology; or
    (5) (i) Have earned a bachelor's degree in a chemical, physical or 
biological science or medical technology from an accredited institution; 
and
    (ii) Have at least 4 years of laboratory training or experience, or 
both, in

[[Page 914]]

high complexity testing for the specialty of diagnostic immunology.
    (i) If the requirements of paragraph (b) of this section are not met 
and the laboratory performs tests in the specialty of chemistry, the 
individual functioning as the technical supervisor must--
    (1)(i) Be a doctor of medicine or doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located; and
    (ii) Be certified in clinical pathology by the American Board of 
Pathology or the American Osteopathic Board of Pathology or possess 
qualifications that are equivalent to those required for such 
certification; or
    (2)(i) Be a doctor of medicine, doctor of osteopathy, or doctor of 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located; and
    (ii) Have at least 1 year of laboratory training or experience, or 
both, in high complexity testing for the specialty of chemistry; or
    (3)(i) Have an earned doctoral degree in a chemical, physical, 
biological or clinical laboratory science from an accredited 
institution; and
    (ii) Have at least 1 year of laboratory training or experience, or 
both, in high complexity testing within the specialty of chemistry; or
    (4)(i) Have earned a master's degree in a chemical, physical, 
biological or clinical laboratory science or medical technology from an 
accredited institution; and
    (ii) Have at least 2 years of laboratory training or experience, or 
both, in high complexity testing for the specialty of chemistry; or
    (5)(i) Have earned a bachelor's degree in a chemical, physical or 
biological science or medical technology from an accredited institution; 
and
    (ii) Have at least 4 years of laboratory training or experience, or 
both, in high complexity testing for the specialty of chemistry.
    (j) If the requirements of paragraph (b) of this section are not met 
and the laboratory performs tests in the specialty of hematology, the 
individual functioning as the technical supervisor must--
    (1)(i) Be a doctor of medicine or a doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located; and
    (ii) Be certified in clinical pathology by the American Board of 
Pathology or the American Osteopathic Board of Pathology or possess 
qualifications that are equivalent to those required for such 
certification; or
    (2)(i) Be a doctor of medicine, doctor of osteopathy, or doctor of 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located; and
    (ii) Have at least one year of laboratory training or experience, or 
both, in high complexity testing for the specialty of hematology (for 
example, physicians certified either in hematology or hematology and 
medical oncology by the American Board of Internal Medicine); or
    (3)(i) Have an earned doctoral degree in a chemical, physical, 
biological or clinical laboratory science from an accredited 
institution; and
    (ii) Have at least 1 year of laboratory training or experience, or 
both, in high complexity testing within the specialty of hematology; or
    (4)(i) Have earned a master's degree in a chemical, physical, 
biological or clinical laboratory science or medical technology from an 
accredited institution; and
    (ii) Have at least 2 years of laboratory training or experience, or 
both, in high complexity testing for the specialty of hematology; or
    (5)(i) Have earned a bachelor's degree in a chemical, physical or 
biological science or medical technology from an accredited institution; 
and
    (ii) Have at least 4 years of laboratory training or experience, or 
both, in high complexity testing for the specialty of hematology.
    (k)(1) If the requirements of paragraph (b) of this section are not 
met and the laboratory performs tests in the subspecialty of cytology, 
the individual functioning as the technical supervisor must--
    (i) Be a doctor of medicine or a doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located; and

[[Page 915]]

    (ii) Meet one of the following requirements--
    (A) Be certified in anatomic pathology by the American Board of 
Pathology or the American Osteopathic Board of Pathology or possess 
qualifications that are equivalent to those required for such 
certification; or
    (B) Be certified by the American Society of Cytology to practice 
cytopathology or possess qualifications that are equivalent to those 
required for such certification;
    (2) An individual qualified under Sec. 493.1449(b) or paragraph 
(k)(1) of this section may delegate some of the cytology technical 
supervisor responsibilities to an individual who is in the final year of 
full-time training leading to certification specified in paragraphs (b) 
or (k)(1)(ii)(A) of this section provided the technical supervisor 
qualified under Sec. 493.1449(b) or paragraph (k)(1) of this section 
remains ultimately responsible for ensuring that all of the 
responsibilities of the cytology technical supervisor are met.
    (l) If the requirements of paragraph (b) of this section are not met 
and the laboratory performs tests in the subspecialty of histopathology, 
the individual functioning as the technical supervisor must--
    (1) Meet one of the following requirements:
    (i) (A) Be a doctor of medicine or a doctor of osteopathy licensed 
to practice medicine or osteopathy in the State in which the laboratory 
is located; and
    (B) Be certified in anatomic pathology by the American Board of 
Pathology or the American Osteopathic Board of Pathology or possess 
qualifications that are equivalent to those required for such 
certification;
    (ii) An individual qualified under Sec. 493.1449(b) or paragraph 
(l)(1) of this section may delegate to an individual who is a resident 
in a training program leading to certification specified in paragraph 
(b) or (l)(1)(i)(B) of this section, the responsibility for examination 
and interpretation of histopathology specimens.
    (2) For tests in dermatopathology, meet one of the following 
requirements:
    (i) (A) Be a doctor of medicine or doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located and--
    (B) Meet one of the following requirements:
    (1) Be certified in anatomic pathology by the American Board of 
Pathology or the American Osteopathic Board of Pathology or possess 
qualifications that are equivalent to those required for such 
certification; or
    (2) Be certified in dermatopathology by the American Board of 
Dermatology and the American Board of Pathology or possess 
qualifications that are equivalent to those required for such 
certification; or
    (3) Be certified in dermatology by the American Board of Dermatology 
or possess qualifications that are equivalent to those required for such 
certification; or
    (ii) An individual qualified under Sec. 493.1449(b) or paragraph 
(l)(2)(i) of this section may delegate to an individual who is a 
resident in a training program leading to certification specified in 
paragraphs (b) or (l)(2)(i)(B) of this section, the responsibility for 
examination and interpretation of dermatopathology specimens.
    (3) For tests in ophthalmic pathology, meet one of the following 
requirements:
    (i)(A) Be a doctor of medicine or doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located and--
    (B) Must meet one of the following requirements:
    (1) Be certified in anatomic pathology by the American Board of 
Pathology or the American Osteopathic Board of Pathology or possess 
qualifications that are equivalent to those required for such 
certification; or
    (2) Be certified by the American Board of Ophthalmology or possess 
qualifications that are equivalent to those required for such 
certitication and have successfully completed at least 1 year of formal 
post-residency fellowship training in ophthalmic pathology; or
    (ii) An individual qualified under Sec. 493.1449(b) or paragraph 
(1)(3)(i) of this section may delegate to an individual who is a 
resident in a training program

[[Page 916]]

leading to certification specified in paragraphs (b) or (1)(3)(i)(B) of 
this section, the responsibility for examination and interpretation of 
ophthalmic specimens; or
    (m) If the requirements of paragraph (b) of this section are not met 
and the laboratory performs tests in the subspecialty of oral pathology, 
the individual functioning as the technical supervisor must meet one of 
the following requirements:
    (1)(i) Be a doctor of medicine or a doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located and--
    (ii) Be certified in anatomic pathology by the American Board of 
Pathology or the American Osteopathic Board of Pathology or possess 
qualifications that are equivalent to those required for such 
certification; or
    (2) Be certified in oral pathology by the American Board of Oral 
Pathology or possess qualifications for such certification; or
    (3) An individual qualified under Sec. 493.1449(b) or paragraph (m) 
(1) or (2) of this section may delegate to an individual who is a 
resident in a training program leading to certification specified in 
paragraphs (b) or (m) (1) or (2) of this section, the responsibility for 
examination and interpretation of oral pathology specimens.
    (n) If the requirements of paragraph (b) of this section are not met 
and the laboratory performs tests in the specialty of radiobioassay, the 
individual functioning as the technical supervisor must--
    (1)(i) Be a doctor of medicine or a doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located; and
    (ii) Be certified in clinical pathology by the American Board of 
Pathology or the American Osteopathic Board of Pathology or possess 
qualifications that are equivalent to those required for such 
certification; or
    (2)(i) Be a doctor of medicine, doctor of osteopathy, or doctor of 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located; and
    (ii) Have at least 1 year of laboratory training or experience, or 
both, in high complexity testing for the specialty of radiobioassay; or
    (3)(i) Have an earned doctoral degree in a chemical, physical, 
biological or clinical laboratory science from an accredited 
institution; and
    (ii) Have at least 1 year of laboratory training or experience, or 
both, in high complexity testing within the specialty of radiobioassay; 
or
    (4)(i) Have earned a master's degree in a chemical, physical, 
biological or clinical laboratory science or medical technology from an 
accredited institution; and
    (ii) Have at least 2 years of laboratory training or experience, or 
both, in high complexity testing for the specialty of radiobioassay; or
    (5)(i) Have earned a bachelor's degree in a chemical, physical or 
biological science or medical technology from an accredited institution; 
and
    (ii) Have at least 4 years of laboratory training or experience, or 
both, in high complexity testing for the specialty of radiobioassay.
    (o) If the laboratory performs tests in the specialty of 
histocompatibility, the individual functioning as the technical 
supervisor must either--
    (1)(i) Be a doctor of medicine, doctor of osteopathy, or doctor of 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located; and
    (ii) Have training or experience that meets one of the following 
requirements:
    (A) Have 4 years of laboratory training or experience, or both, 
within the specialty of histocompatibility; or
    (B)(1) Have 2 years of laboratory training or experience, or both, 
in the specialty of general immunology; and
    (2) Have 2 years of laboratory training or experience, or both, in 
the specialty of histocompatibility; or
    (2)(i) Have an earned doctoral degree in a biological or clinical 
laboratory science from an accredited institution; and
    (ii) Have training or experience that meets one of the following 
requirements:

[[Page 917]]

    (A) Have 4 years of laboratory training or experience, or both, 
within the specialty of histocompatibility; or
    (B)(1) Have 2 years of laboratory training or experience, or both, 
in the specialty of general immunology; and
    (2) Have 2 years of laboratory training or experience, or both, in 
the specialty of histocompatibility.
    (p) If the laboratory performs tests in the specialty of clinical 
cytogenetics, the individual functioning as the technical supervisor 
must--
    (1)(i) Be a doctor of medicine, doctor of osteopathy, or doctor of 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located; and
    (ii) Have 4 years of training or experience, or both, in genetics, 2 
of which have been in clinical cytogenetics; or
    (2)(i) Hold an earned doctoral degree in a biological science, 
including biochemistry, or clinical laboratory science from an 
accredited institution; and
    (ii) Have 4 years of training or experience, or both, in genetics, 2 
of which have been in clinical cytogenetics.
    (q) If the requirements of paragraph (b) of this section are not met 
and the laboratory performs tests in the specialty of immunohematology, 
the individual functioning as the technical supervisor must--
    (1)(i) Be a doctor of medicine or a doctor of osteopathy licensed to 
practice medicine or osteopathy in the State in which the laboratory is 
located; and
    (ii) Be certified in clinical pathology by the American Board of 
Pathology or the American Osteopathic Board of Pathology or possess 
qualifications that are equivalent to those required for such 
certification; or
    (2)(i) Be a doctor of medicine, doctor of osteopathy, or doctor of 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located; and
    (ii) Have at least one year of laboratory training or experience, or 
both, in high complexity testing for the specialty of immunohematology.

    Note: The technical supervisor requirements for ``laboratory 
training or experience, or both'' in each specialty or subspecialty may 
be acquired concurrently in more than one of the specialties or 
subspecialties of service. For example, an individual, who has a 
doctoral degree in chemistry and additionally has documentation of 1 
year of laboratory experience working concurrently in high complexity 
testing in the specialties of microbiology and chemistry and 6 months of 
that work experience included high complexity testing in bacteriology, 
mycology, and mycobacteriology, would qualify as the technical 
supervisor for the specialty of chemistry and the subspecialties of 
bacteriology, mycology, and mycobacteriology.

[57 FR 7172, Feb. 28, 1992, as amended at 58 FR 5234, Jan. 19, 1993]

Sec. 493.1451  Standard: Technical supervisor responsibilities.

    The technical supervisor is responsible for the technical and 
scientific oversight of the laboratory. The technical supervisor is not 
required to be on site at all times testing is performed; however, he or 
she must be available to the laboratory on an as needed basis to provide 
supervision as specified in (a) of this section.
    (a) The technical supervisor must be accessible to the laboratory to 
provide on-site, telephone, or electronic consultation; and
    (b) The technical supervisor is responsible for--
    (1) Selection of the test methodology that is appropriate for the 
clinical use of the test results;
    (2) Verification of the test procedures performed and establishment 
of the laboratory's test performance characteristics, including the 
precision and accuracy of each test and test system;
    (3) Enrollment and participation in an HHS approved proficiency 
testing program commensurate with the services offered;
    (4) Establishing a quality control program appropriate for the 
testing performed and establishing the parameters for acceptable levels 
of analytic performance and ensuring that these levels are maintained 
throughout the entire testing process from the initial receipt of the 
specimen, through sample analysis and reporting of test results;
    (5) Resolving technical problems and ensuring that remedial actions 
are taken whenever test systems deviate from the laboratory's 
established performance specifications;

[[Page 918]]

    (6) Ensuring that patient test results are not reported until all 
corrective actions have been taken and the test system is functioning 
properly;
    (7) Identifying training needs and assuring that each individual 
performing tests receives regular in-service training and education 
appropriate for the type and complexity of the laboratory services 
performed;
    (8) Evaluating the competency of all testing personnel and assuring 
that the staff maintain their competency to perform test procedures and 
report test results promptly, accurately and proficiently. The 
procedures for evaluation of the competency of the staff must include, 
but are not limited to--
    (i) Direct observations of routine patient test performance, 
including patient preparation, if applicable, specimen handling, 
processing and testing;
    (ii) Monitoring the recording and reporting of test results;
    (iii) Review of intermediate test results or worksheets, quality 
control records, proficiency testing results, and preventive maintenance 
records;
    (iv) Direct observation of performance of instrument maintenance and 
function checks;
    (v) Assessment of test performance through testing previously 
analyzed specimens, internal blind testing samples or external 
proficiency testing samples; and
    (vi) Assessment of problem solving skills; and
    (9) Evaluating and documenting the performance of individuals 
responsible for high complexity testing at least semiannually during the 
first year the individual tests patient specimens. Thereafter, 
evaluations must be performed at least annually unless test methodology 
or instrumentation changes, in which case, prior to reporting patient 
test results, the individual's performance must be reevaluated to 
include the use of the new test methodology or instrumentation.
    (c) In cytology, the technical supervisor or the individual 
qualified under Sec. 493.1449(k)(2)--
    (1) May perform the duties of the cytology general supervisor and 
the cytotechnologist, as specified in Secs. 493.1471 and 493.1485, 
respectively;
    (2) Must establish the workload limit for each individual examining 
slides;
    (3) Must reassess the workload limit for each individual examining 
slides at least every 6 months and adjust as necessary;
    (4) Must perform the functions specified in Sec. 493.1257(c);
    (5) Must ensure that each individual examining gynecologic 
preparations participates in an HHS approved cytology proficiency 
testing program, as specified in Sec. 493.945 and achieves a passing 
score, as specified in Sec. 493.855; and
    (6) If responsible for screening cytology slide preparations, must 
document the number of cytology slides screened in 24 hours and the 
number of hours devoted during each 24-hour period to screening cytology 
slides.

[57 FR 7172, Feb. 28, 1992, as amended at 58 FR 5235, Jan. 19, 1993]

Sec. 493.1453  Condition: Laboratories performing high complexity 
          testing; clinical consultant.

    The laboratory must have a clinical consultant who meets the 
requirements of Sec. 493.1455 of this subpart and provides clinical 
consultation in accordance with Sec. 493.1457 of this subpart.

Sec. 493.1455  Standard; Clinical consultant qualifications.

    The clinical consultant must be qualified to consult with and render 
opinions to the laboratory's clients concerning the diagnosis, treatment 
and management of patient care. The clinical consultant must--
    (a) Be qualified as a laboratory director under Sec. 493.1443(b)(1), 
(2), or (3)(i) or, for the subspecialty of oral pathology, 
Sec. 493.1443(b)(6); or
    (b) Be a doctor of medicine, doctor of osteopathy, doctor of 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located.

[57 FR 7172, Feb. 28, 1992, as amended at 58 FR 5235, Jan. 19, 1993]

[[Page 919]]

Sec. 493.1457  Standard; Clinical consultant responsibilities.

    The clinical consultant provides consultation regarding the 
appropriateness of the testing ordered and interpretation of test 
results. The clinical consultant must--
    (a) Be available to provide consultation to the laboratory's 
clients;
    (b) Be available to assist the laboratory's clients in ensuring that 
appropriate tests are ordered to meet the clinical expectations;
    (c) Ensure that reports of test results include pertinent 
information required for specific patient interpretation; and
    (d) Ensure that consultation is available and communicated to the 
laboratory's clients on matters related to the quality of the test 
results reported and their interpretation concerning specific patient 
conditions.

Sec. 493.1459  Condition: Laboratories performing high complexity 
          testing; general supervisor.

    The laboratory must have one or more general supervisors who are 
qualified under Sec. 493.1461 of this subpart to provide general 
supervision in accordance with Sec. 493.1463 of this subpart.

Sec. 493.1461  Standard: General supervisor qualifications.

    The laboratory must have one or more general supervisors who, under 
the direction of the laboratory director and supervision of the 
technical supervisor, provides day-to-day supervision of testing 
personnel and reporting of test results. In the absence of the director 
and technical supervisor, the general supervisor must be responsible for 
the proper performance of all laboratory procedures and reporting of 
test results.
    (a) The general supervisor must possess a current license issued by 
the State in which the laboratory is located, if such licensing is 
required; and
    (b) The general supervisor must be qualified as a--
    (1) Laboratory director under Sec. 493.1443; or
    (2) Technical supervisor under Sec. 493.1449.
    (c) If the requirements of paragraph (b)(1) or paragraph (b)(2) of 
this section are not met, the individual functioning as the general 
supervisor must--
    (1)(i) Be a doctor of medicine, doctor of osteopathy, or doctor of 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located or have earned 
a doctoral, master's, or bachelor's degree in a chemical, physical, 
biological or clinical laboratory science, or medical technology from an 
accredited institution; and
    (ii) Have at least 1 year of laboratory training or experience, or 
both, in high complexity testing; or
    (2)(i) Qualify as testing personnel under Sec. 493.1489(b)(2); and
    (ii) Have at least 2 years of laboratory training or experience, or 
both, in high complexity testing; or
    (3)(i) Except as specified in paragraph (3)(ii) of this section, 
have previously qualified as a general supervisor under Sec. 493.1462 on 
or before February 28, 1992.
    (ii) Exception. An individual who achieved a satisfactory grade in a 
proficiency examination for technologist given by HHS between March 1, 
1986 and December 31, 1987, qualifies as a general supervisor if he or 
she meets the requirements of Sec. 493.1462 on or before January 1, 
1994.''
    (4) On or before September 1, 1992, have served as a general 
supervisor of high complexity testing and as of April 24, 1995--
    (i) Meet one of the following requirements:
    (A) Have graduated from a medical laboratory or clinical laboratory 
training program approved or accredited by the Accrediting Bureau of 
Health Education Schools (ABHES), the Commission on Allied Health 
Education Accreditation (CAHEA), or other organization approved by HHS.
    (B) Be a high school graduate or equivalent and have successfully 
completed an official U.S. military medical laboratory procedures course 
of at least 50 weeks duration and have held the military enlisted 
occupational specialty of Medical Laboratory Specialist (Laboratory 
Technician).
    (ii) Have at least 2 years of clinical laboratory training, or 
experience, or both, in high complexity testing; or

[[Page 920]]

    (5) On or before September 1, 1992, have served as a general 
supervisor of high complexity testing and--
    (i) Be a high school graduate or equivalent; and
    (ii) Have had at least 10 years of laboratory training or 
experience, or both, in high complexity testing, including at least 6 
years of supervisory experience between September 1, 1982 and September 
1, 1992.
    (d) For blood gas analysis, the individual providing general 
supervision must--
    (1) Be qualified under Secs. 493.1461(b) (1) or (2), or 493.1461(c); 
or
    (2)(i) Have earned a bachelor's degree in respiratory therapy or 
cardiovascular technology from an accredited institution; and
    (ii) Have at least one year of laboratory training or experience, or 
both, in blood gas analysis; or
    (3)(i) Have earned an associate degree related to pulmonary function 
from an accredited institution; and
    (ii) Have at least two years of training or experience, or both in 
blood gas analysis.
    (e) The general supervisor requirement is met in histopathology, 
oral pathology, dermatopathology, and ophthalmic pathology because all 
tests and examinations, must be performed:
    (1) In histopathology, by an individual who is qualified as a 
technical supervisor under Secs. 493.1449(b) or 493.1449(l)(1);
    (2) In dermatopathology, by an individual who is qualified as a 
technical supervisor under Secs. 493.1449(b) or 493.1449(l) or (2);
    (3) In ophthalmic pathology, by an individual who is qualified as a 
technical supervisor under Secs. 493.1449(b) or 493.1449(1)(3); and
    (4) In oral pathology, by an individual who is qualified as a 
technical supervisor under Secs. 493.1449(b) or 493.1449(m).

[57 FR 7172, Feb. 28, 1992, as amended at 58 FR 5235, Jan. 19, 1993; 58 
FR 39155, July 22, 1993; 60 FR 20049, Apr. 24, 1995]

Sec. 493.1462  General supervisor qualifications on or before February 
          28, 1992.

    To qualify as a general supervisor under Sec. 493.1461(c)(3), an 
individual must have met or could have met the following qualifications 
as they were in effect on or before February 28, 1992.
    (a) Each supervisor possesses a current license as a laboratory 
supervisor issued by the State, if such licensing exists; and
    (b) The laboratory supervisor--
    (1) Who qualifies as a laboratory director under 
Sec. 493.1406(b)(1), (2), (4), or (5) is also qualified as a general 
supervisor; therefore, depending upon the size and functions of the 
laboratory, the laboratory director may also serve as the laboratory 
supervisor; or
    (2)(i) Is a physician or has earned a doctoral degree from an 
accredited institution with a major in one of the chemical, physical, or 
biological sciences; and
    (ii) Subsequent to graduation, has had at least 2 years of 
experience in one of the laboratory specialties in a laboratory; or
    (3)(i) Holds a master's degree from an accredited institution with a 
major in one of the chemical, physical, or biological sciences; and
    (ii) Subsequent to graduation has had at least 4 years of pertinent 
full-time laboratory experience of which not less than 2 years have been 
spent working in the designated specialty in a laboratory; or
    (4)(i) Is qualified as a laboratory technologist under 
Sec. 493.1491; and
    (ii) After qualifying as a laboratory technologist, has had at least 
6 years of pertinent full-time laboratory experience of which not less 
than 2 years have been spent working in the designated laboratory 
specialty in a laboratory; or
    (5) With respect to individuals first qualifying before July 1, 
1971, has had at least 15 years of pertinent full-time laboratory 
experience before January 1, 1968; this required experience may be met 
by the substitution of education for experience.

[58 FR 39155, July 22, 1993]

Sec. 493.1463  Standard: General supervisor responsibilities.

    The general supervisor is responsible for day-to-day supervision or 
oversight of the laboratory operation and personnel performing testing 
and reporting test results.

[[Page 921]]

    (a) The general supervisor--(1) Must be accessible to testing 
personnel at all times testing is performed to provide on-site, 
telephone or electronic consultation to resolve technical problems in 
accordance with policies and procedures established either by the 
laboratory director or technical supervisor;
    (2) Is responsible for providing day-to-day supervision of high 
complexity test performance by a testing personnel qualified under 
Sec. 493.1489;
    (3) Except as specified in paragraph (c) of this section, must be 
onsite to provide direct supervision when high complexity testing is 
performed by any individuals qualified under Sec. 493.1489(b)(5); and
    (4) Is responsible for monitoring test analyses and specimen 
examinations to ensure that acceptable levels of analytic performance 
are maintained.
    (b) The director or technical supervisor may delegate to the general 
supervisor the responsibility for--
    (1) Assuring that all remedial actions are taken whenever test 
systems deviate from the laboratory's established performance 
specifications;
    (2) Ensuring that patient test results are not reported until all 
corrective actions have been taken and the test system is properly 
functioning;
    (3) Providing orientation to all testing personnel; and
    (4) Annually evaluating and documenting the performance of all 
testing personnel.
    (c) Exception. For individuals qualified under Sec. 493.1489(b)(5), 
who were performing high complexity testing on or before January 19, 
1993, the requirements of paragraph (a)(3) of this section are not 
effective, provided that all high complexity testing performed by the 
individual in the absence of a general supervisor is reviewed within 24 
hours by a general supervisor qualified under Sec. 493.1461.

[57 FR 7172, Feb. 28, 1992, as amended at 58 FR 5235, Jan. 19, 1993; 60 
FR 20050, Apr. 24, 1995]

Sec. 493.1467  Condition: Laboratories performing high complexity 
          testing; cytology general supervisor.

    For the subspecialty of cytology, the laboratory must have a general 
supervisor who meets the qualification requirements of Sec. 493.1469 of 
this subpart, and provides supervision in accordance with Sec. 493.1471 
of this subpart.

Sec. 493.1469  Standard: Cytology general supervisor qualifications.

    The cytology general supervisor must be qualified to supervise 
cytology services. The general supervisor in cytology must possess a 
current license issued by the State in which the laboratory is located, 
if such licensing is required, and must--
    (a) Be qualified as a technical supervisor under Sec. 493.1449 (b) 
or (k); or
    (b)(1) Be qualified as a cytotechnologist under Sec. 493.1483; and
    (2) Have at least 3 years of full-time (2,080 hours per year) 
experience as a cytotechnologist within the preceding 10 years.

Sec. 493.1471  Standard: Cytology general supervisor responsibilities.

    The technical supervisor of cytology may perform the duties of the 
cytology general supervisor or delegate the responsibilities to an 
individual qualified under Sec. 493.1469.
    (a) The cytology general supervisor is responsible for the day-to-
day supervision or oversight of the laboratory operation and personnel 
performing testing and reporting test results.
    (b) The cytology general supervisor must--
    (1) Be accessible to provide on-site, telephone, or electronic 
consultation to resolve technical problems in accordance with policies 
and procedures established by the technical supervisor of cytology;
    (2) Document the slide interpretation results of each gynecologic 
and nongynecologic cytology case he or she examined or reviewed (as 
specified under Sec. 493.1257(d));
    (3) For each 24-hour period, document the total number of slides he 
or she examined or reviewed in the laboratory as well as the total 
number of slides examined or reviewed in any other laboratory or for any 
other employer; and
    (4) Document the number of hours spent examining slides in each 24-
hour period.

[[Page 922]]

Sec. 493.1481  Condition: Laboratories performing high complexity 
          testing; cytotechnologist.

    For the subspecialty of cytology, the laboratory must have a 
sufficient number of cytotechnologists who meet the qualifications 
specified in Sec. 493.1483 to perform the functions specified in 
Sec. 493.1485.

Sec. 493.1483  Standard: Cytotechnologist qualifications.

    Each person examining cytology slide preparations must meet the 
qualifications of Sec. 493.1449 (b) or (k), or--
    (a) Possess a current license as a cytotechnologist issued by the 
State in which the laboratory is located, if such licensing is required; 
and
    (b) Meet one of the following requirements:
    (1) Have graduated from a school of cytotechnology accredited by the 
Committee on Allied Health Education and Accreditation or other 
organization approved by HHS; or
    (2) Be certified in cytotechnology by a certifying agency approved 
by HHS; or
    (3) Before September 1, 1992--
    (i) Have successfully completed 2 years in an accredited institution 
with at least 12 semester hours in science, 8 hours of which are in 
biology; and
    (A) Have had 12 months of training in a school of cytotechnology 
accredited by an accrediting agency approved by HHS; or
    (B) Have received 6 months of formal training in a school of 
cytotechnology accredited by an accrediting agency approved by HHS and 6 
months of full-time experience in cytotechnology in a laboratory 
acceptable to the pathologist who directed the formal 6 months of 
training; or
    (ii) Have achieved a satisfactory grade to qualify as a 
cytotechnologist in a proficiency examination approved by HHS and 
designed to qualify persons as cytotechnologists; or
    (4) Before September 1, 1994, have full-time experience of at least 
2 years or equivalent within the preceding 5 years examining slide 
preparations under the supervision of a physician qualified under 
Sec. 493.1449(b) or (k)(1), and before January 1, 1969, must have--
    (i) Graduated from high school;
    (ii) Completed 6 months of training in cytotechnology in a 
laboratory directed by a pathologist or other physician providing 
cytology services; and
    (iii) Completed 2 years of full-time supervised experience in 
cytotechnology; or
    (5)(i) On or before September 1, 1994, have full-time experience of 
at least 2 years or equivalent examining cytology slide preparations 
within the preceding 5 years in the United States under the supervision 
of a physician qualified under Sec. 493.1449(b) or (k)(1); and
    (ii) On or before September 1, 1995, have met the requirements in 
either paragraph (b)(1) or (2) of this section.

[57 FR 7172, Feb. 28, 1992, as amended at 59 FR 685, Jan. 6, 1994]

Sec. 493.1485  Standard; Cytotechnologist responsibilities.

    The cytotechnologist is responsible for documenting--
    (a) The slide interpretation results of each gynecologic and 
nongynecologic cytology case he or she examined or reviewed (as 
specified in Sec. 493.1257(d));
    (b) For each 24-hour period, the total number of slides examined or 
reviewed in the laboratory as well as the total number of slides 
examined or reviewed in any other laboratory or for any other employer; 
and
    (c) The number of hours spent examining slides in each 24-hour 
period.

Sec. 493.1487  Condition: Laboratories performing high complexity 
          testing; testing personnel.

    The laboratory has a sufficient number of individuals who meet the 
qualification requirements of Sec. 493.1489 of this subpart to perform 
the functions specified in Sec. 493.1495 of this subpart for the volume 
and complexity of testing performed.

Sec. 493.1489  Standard; Testing personnel qualifications.

    Each individual performing high complexity testing must--
    (a) Possess a current license issued by the State in which the 
laboratory is located, if such licensing is required; and
    (b) Meet one of the following requirements:

[[Page 923]]

    (1) Be a doctor of medicine, doctor of osteopathy, or doctor of 
podiatric medicine licensed to practice medicine, osteopathy, or 
podiatry in the State in which the laboratory is located or have earned 
a doctoral, master's or bachelor's degree in a chemical, physical, 
biological or clinical laboratory science, or medical technology from an 
accredited institution;
    (2)(i) Have earned an associate degree in a laboratory science, or 
medical laboratory technology from an accredited institution or--
    (ii) Have education and training equivalent to that specified in 
paragraph (b)(2)(i) of this section that includes--
    (A) At least 60 semester hours, or equivalent, from an accredited 
institution that, at a minimum, include either--
    (1) 24 semester hours of medical laboratory technology courses; or
    (2) 24 semester hours of science courses that include--
    (i) Six semester hours of chemistry;
    (ii) Six semester hours of biology; and
    (iii) Twelve semester hours of chemistry, biology, or medical 
laboratory technology in any combination; and
    (B) Have laboratory training that includes either of the following:
    (1) Completion of a clinical laboratory training program approved or 
accredited by the ABHES, the CAHEA, or other organization approved by 
HHS. (This training may be included in the 60 semester hours listed in 
paragraph (b)(2)(ii)(A) of this section.)
    (2) At least 3 months documented laboratory training in each 
specialty in which the individual performs high complexity testing.
    (3) Have previously qualified or could have qualified as a 
technologist under Sec. 493.1491 on or before February 28, 1992;
    (4) On or before April 24, 1995 be a high school graduate or 
equivalent and have either--
    (i) Graduated from a medical laboratory or clinical laboratory 
training program approved or accredited by ABHES, CAHEA, or other 
organization approved by HHS; or
    (ii) Successfully completed an official U.S. military medical 
laboratory procedures training course of at least 50 weeks duration and 
have held the military enlisted occupational specialty of Medical 
Laboratory Specialist (Laboratory Technician);
    (5)(i) Until September 1, 1997--
    (A) Have earned a high school diploma or equivalent; and
    (B) Have documentation of training appropriate for the testing 
performed before analyzing patient specimens. Such training must ensure 
that the individual has--
    (1) The skills required for proper specimen collection, including 
patient preparation, if applicable, labeling, handling, preservation or 
fixation, processing or preparation, transportation and storage of 
specimens;
    (2) The skills required for implementing all standard laboratory 
procedures;
    (3) The skills required for performing each test method and for 
proper instrument use;
    (4) The skills required for performing preventive maintenance, 
troubleshooting, and calibration procedures related to each test 
performed;
    (5) A working knowledge of reagent stability and storage;
    (6) The skills required to implement the quality control policies 
and procedures of the laboratory;
    (7) An awareness of the factors that influence test results; and
    (8) The skills required to assess and verify the validity of patient 
test results through the evaluation of quality control values before 
reporting patient test results; and
    (ii) As of September 1, 1997, be qualified under 
Sec. 493.1489(b)(1), (b)(2), or (b)(4), except for those individuals 
qualified under paragraph (b)(5)(i) of this section who were performing 
high complexity testing on or before April 24, 1995;
    (6) For blood gas analysis--
    (i) Be qualified under Sec. 493.1489(b)(1), (b)(2), (b)(3), (b)(4), 
or (b)(5);
    (ii) Have earned a bachelor's degree in respiratory therapy or 
cardiovascular technology from an accredited institution; or
    (iii) Have earned an associate degree related to pulmonary function 
from an accredited institution; or

[[Page 924]]

    (7) For histopathology, meet the qualifications of Sec. 493.1449 (b) 
or (l) to perform tissue examinations.

[57 FR 7172, Feb. 28, 1992, as amended at 58 FR 5236, Jan. 19, 1993; 58 
FR 39155, July 22, 1993; 60 FR 20050, Apr. 24, 1995]

Sec. 493.1491  Technologist qualifications on or before February 28, 
          1992.

    In order to qualify as high complexity testing personnel under 
Sec. 493.1489(b)(3), the individual must have met or could have met the 
following qualifications for technologist as they were in effect on or 
before February 28, 1992. Each technologist must--
    (a) Possess a current license as a laboratory technologist issued by 
the State, if such licensing exists; and
    (b)(1) Have earned a bachelor's degree in medical technology from an 
accredited university; or
    (2) Have successfully completed 3 years of academic study (a minimum 
of 90 semester hours or equivalent) in an accredited college or 
university, which met the specific requirements for entrance into a 
school of medical technology accredited by an accrediting agency 
approved by the Secretary, and has successfully completed a course of 
training of at least 12 months in such a school; or
    (3) Have earned a bachelor's degree in one of the chemical, 
physical, or biological sciences and, in addition, has at least 1 year 
of pertinent full-time laboratory experience or training, or both, in 
the specialty or subspecialty in which the individual performs tests; or
    (4)(i) Have successfully completed 3 years (90 semester hours or 
equivalent) in an accredited college or university with the following 
distribution of courses--
    (A) For those whose training was completed before September 15, 
1963. At least 24 semester hours in chemistry and biology courses of 
which--
    (1) At least 6 semester hours were in inorganic chemistry and at 
least 3 semester hours were in other chemistry courses; and
    (2) At least 12 semester hours in biology courses pertinent to the 
medical sciences; or
    (B) For those whose training was completed after September 14, 1963.
    (1) 16 semester hours in chemistry courses that included at least 6 
semester hours in inorganic chemistry and that are acceptable toward a 
major in chemistry;
    (2) 16 semester hours in biology courses that are pertinent to the 
medical sciences and are acceptable toward a major in the biological 
sciences; and
    (3) 3 semester hours of mathematics; and
    (ii) Has experience, training, or both, covering several fields of 
medical laboratory work of at least 1 year and of such quality as to 
provide him or her with education and training in medical technology 
equivalent to that described in paragraphs (b)(1) and (2) of this 
section; or
    (5) With respect to individuals first qualifying before July 1, 
1971, the technologist--
    (i) Was performing the duties of a laboratory technologist at any 
time between July 1, 1961, and January 1, 1968, and
    (ii) Has had at least 10 years of pertinent laboratory experience 
prior to January 1, 1968. (This required experience may be met by the 
substitution of education for experience); or
    (6) Achieves a satisfactory grade in a proficiency examination 
approved by HHS.

[58 FR 39155, July 22, 1993]

Sec. 493.1495  Standard; Testing personnel responsibilities.

    The testing personnel are responsible for specimen processing, test 
performance and for reporting test results.
    (a) Each individual performs only those high complexity tests that 
are authorized by the laboratory director and require a degree of skill 
commensurate with the individual's education, training or experience, 
and technical abilities.
    (b) Each individual performing high complexity testing must--
    (1) Follow the laboratory's procedures for specimen handling and 
processing, test analyses, reporting and maintaining records of patient 
test results;
    (2) Maintain records that demonstrate that proficiency testing 
samples are tested in the same manner as patient specimens;

[[Page 925]]

    (3) Adhere to the laboratory's quality control policies, document 
all quality control activities, instrument and procedural calibrations 
and maintenance performed;
    (4) Follow the laboratory's established policies and procedures 
whenever test systems are not within the laboratory's established 
acceptable levels of performance;
    (5) Be capable of identifying problems that may adversely affect 
test performance or reporting of test results and either must correct 
the problems or immediately notify the general supervisor, technical 
supervisor, clinical consultant, or director;
    (6) Document all corrective actions taken when test systems deviate 
from the laboratory's established performance specifications; and
    (7) Except as specified in paragraph (c) of this section, if 
qualified under Sec. 493.1489(b)(5), perform high complexity testing 
only under the onsite, direct supervision of a general supervisor 
qualified under Sec. 493.1461.
    (c) Exception. For individuals qualified under Sec. 493.1489(b)(5), 
who were performing high complexity testing on or before January 19, 
1993, the requirements of paragraph (b)(7) of this section are not 
effective, provided that all high complexity testing performed by the 
individual in the absence of a general supervisor is reviewed within 24 
hours by a general supervisor qualified under Sec. 493.1461.

[57 FR 7172, Feb. 28, 1992, as amended at 58 FR 5236, Jan. 19, 1993; 60 
FR 20050, Apr. 24, 1995]

Subparts N-O [Reserved]

  Subpart P--Quality Assurance for Moderate Complexity (Including the 
  Subcategory) or High Complexity Testing, or Any Combination of These 
                                  Tests

    Source: 57 FR 7183, Feb. 28, 1992, unless otherwise noted.

Sec. 493.1701  Condition: Quality assurance; moderate complexity 
          (including the subcategory) or high complexity testing, or any 
          combination of these tests.

    Each laboratory performing moderate complexity (including the 
subcategory) or high complexity testing, or any combination of these 
tests, must establish and follow written policies and procedures for a 
comprehensive quality assurance program that is designed to monitor and 
evaluate the ongoing and overall quality of the total testing process 
(preanalytic, analytic, postanalytic). The laboratory's quality 
assurance program must evaluate the effectiveness of its policies and 
procedures; identify and correct problems; assure the accurate, reliable 
and prompt reporting of test results; and assure the adequacy and 
competency of the staff. As necessary, the laboratory must revise 
policies and procedures based upon the results of those evaluations. The 
laboratory must meet the standards as they apply to the services 
offered, complexity of testing performed and test results reported, and 
the unique practices of each testing entity. All quality assurance 
activities must be documented.

[60 FR 20050, Apr. 24, 1995]

Sec. 493.1703  Standard; Patient test management assessment.

    The laboratory must have an ongoing mechanism for monitoring and 
evaluating the systems required under subpart J, Patient Test 
Management. The laboratory must monitor, evaluate, and revise, if 
necessary, based on the results of its evaluations, the following:
    (a) The criteria established for patient preparation, specimen 
collection, labeling, preservation and transportation;
    (b) The information solicited and obtained on the laboratory's test 
requisition for its completeness, relevance, and necessity for the 
testing of patient specimens;
    (c) The use and appropriateness of the criteria established for 
specimen rejection;

[[Page 926]]

    (d) The completeness, usefulness, and accuracy of the test report 
information necessary for the interpretation or utilization of test 
results;
    (e) The timely reporting of test results based on testing priorities 
(STAT, routine, etc.); and
    (f) The accuracy and reliability of test reporting systems, 
appropriate storage of records and retrieval of test results.

Sec. 493.1705  Standard; Quality control assessment.

    The laboratory must have an ongoing mechanism to evaluate the 
corrective actions taken under Sec. 493.1219, Remedial actions. 
Ineffective policies and procedures must be revised based on the outcome 
of the evaluation. The mechanism must evaluate and review the 
effectiveness of corrective actions taken for--
    (a) Problems identified during the evaluation of calibration and 
control data for each test method;
    (b) Problems identified during the evaluation of patient test values 
for the purpose of verifying the reference range of a test method; and
    (c) Errors detected in reported results.

Sec. 493.1707  Standard; Proficiency testing assessment.

    Under subpart H of this part, Proficiency Testing, the corrective 
actions taken for any unacceptable, unsatisfactory, or unsuccessful 
proficiency testing result(s) must be evaluated for effectiveness.

Sec. 493.1709  Standard; Comparison of test results.

    (a) If a laboratory performs the same test using different 
methodologies or instruments, or performs the same test at multiple 
testing sites, the laboratory must have a system that twice a year 
evaluates and defines the relationship between test results using the 
different methodologies, instruments, or testing sites.
    (b) If a laboratory performs tests that are not included under 
subpart I of this part, Proficiency Testing Programs, the laboratory 
must have a system for verifying the accuracy of its test results at 
least twice a year.

[58 FR 5236, Jan. 19, 1993]

Sec. 493.1711  Standard; Relationship of patient information to patient 
          test results.

    For internal quality assurance, the laboratory must have a mechanism 
to identify and evaluate patient test results that appear inconsistent 
with relevant criteria such as--
    (a) Patient age;
    (b) Sex;
    (c) Diagnosis or pertinent clinical data, when provided;
    (d) Distribution of patient test results when available; and
    (e) Relationship with other test parameters, when available within 
the laboratory.

Sec. 493.1713  Standard; Personnel assessment.

    The laboratory must have an ongoing mechanism to evaluate the 
effectiveness of its policies and procedures for assuring employee 
competence and, if applicable, consultant competence.

Sec. 493.1715  Standard; Communications.

    The laboratory must have a system in place to document problems that 
occur as a result of breakdowns in communication between the laboratory 
and the authorized individual who orders or receives the results of test 
procedures or examinations. Corrective actions must be taken, as 
necessary, to resolve the problems and minimize communication 
breakdowns.

[58 FR 5236, Jan. 19, 1993]

Sec. 493.1717  Standard; Complaint investigations.

    The laboratory must have a system in place to assure that all 
complaints and problems reported to the laboratory are documented. 
Investigations of complaints must be made, when appropriate, and, as 
necessary, corrective actions are instituted.

Sec. 493.1719  Standard; Quality assurance review with staff.

    The laboratory must have a mechanism for documenting and assessing

[[Page 927]]

problems identified during quality assurance reviews and discussing them 
with the staff. The laboratory must take corrective actions that are 
necessary to prevent recurrences.

Sec. 493.1721  Standard; Quality assurance records.

    The laboratory must maintain documentation of all quality assurance 
activities including problems identified and corrective actions taken. 
All quality assurance records must be available to HHS and maintained 
for a period of 2 years.

[58 FR 5236, Jan. 19, 1993]

                          Subpart Q--Inspection

    Source: 57 FR 7184, Feb. 28, 1992, unless otherwise noted.

Sec. 493.1771  Condition: Inspection requirements applicable to all 
          CLIA-certified and CLIA-exempt laboratories.

    (a) Each laboratory issued a CLIA certificate must meet the 
requirements in Sec. 493.1773 and the specific requirements for its 
certificate type, as specified in Secs. 493.1775 through 493.1780.
    (b) All CLIA-exempt laboratories must comply with the inspection 
requirements in Secs. 493.1773 and 493.1780, when applicable.

[63 FR 26737, May 14, 1998]

Sec. 493.1773  Standard: Basic inspection requirements for all 
          laboratories issued a CLIA certificate and CLIA-exempt 
          laboratories.

    (a) A laboratory issued a certificate must permit HCFA or a HCFA 
agent to conduct an inspection to assess the laboratory's compliance 
with the requirements of this part. A CLIA-exempt laboratory and a 
laboratory that requests, or is issued a certificate of accreditation, 
must permit HCFA or a HCFA agent to conduct validation and complaint 
inspections.
    (b) General requirements. As part of the inspection process, HCFA or 
a HCFA agent may require the laboratory to do the following:
    (1) Test samples, including proficiency testing samples, or perform 
procedures.
    (2) Permit interviews of all personnel concerning the laboratory's 
compliance with the applicable requirements of this part.
    (3) Permit laboratory personnel to be observed performing all phases 
of the total testing process (preanalytic, analytic, and postanalytic).
    (4) Permit HCFA or a HCFA agent access to all areas encompassed 
under the certificate including, but not limited to, the following:
    (i) Specimen procurement and processing areas.
    (ii) Storage facilities for specimens, reagents, supplies, records, 
and reports.
    (iii) Testing and reporting areas.
    (5) Provide HCFA or a HCFA agent with copies or exact duplicates of 
all records and data it requires.
    (c) Accessible records and data. A laboratory must have all records 
and data accessible and retrievable within a reasonable time frame 
during the course of the inspection.
    (d) Requirement to provide information and data. A laboratory must 
provide, upon request, all information and data needed by HCFA or a HCFA 
agent to make a determination of the laboratory's compliance with the 
applicable requirements of this part.
    (e) Reinspection. HCFA or a HCFA agent may reinspect a laboratory at 
any time to evaluate the ability of the laboratory to provide accurate 
and reliable test results.
    (f) Complaint inspection. HCFA or a HCFA agent may conduct an 
inspection when there are complaints alleging noncompliance with any of 
the requirements of this part.
    (g) Failure to permit an inspection or reinspection. Failure to 
permit HCFA or a HCFA agent to conduct an inspection or reinspection 
results in the suspension or cancellation of the laboratory's 
participation in Medicare and Medicaid for payment, and suspension or 
limitation of, or action to revoke the laboratory's CLIA certificate, in 
accordance with subpart R of this part.

[63 FR 26737, May 14, 1998; 63 FR 32699, June 15, 1998]

[[Page 928]]

Sec. 493.1775  Standard: Inspection of laboratories issued a certificate 
          of waiver or a certificate for provider-performed microscopy 
          procedures.

    (a) A laboratory that has been issued a certificate of waiver or a 
certificate for provider-performed microscopy procedures is not subject 
to biennial inspections.
    (b) If necessary, HCFA or a HCFA agent may conduct an inspection of 
a laboratory issued a certificate of waiver or a certificate for 
provider-performed microscopy procedures at any time during the 
laboratory's hours of operation to do the following:
    (1) Determine if the laboratory is operated and testing is performed 
in a manner that does not constitute an imminent and serious risk to 
public health.
    (2) Evaluate a complaint from the public.
    (3) Determine whether the laboratory is performing tests beyond the 
scope of the certificate held by the laboratory.
    (4) Collect information regarding the appropriateness of tests 
specified as waived tests or provider-performed microscopy procedures.
    (c) The laboratory must comply with the basic inspection 
requirements of Sec. 493.1773.

[63 FR 26737, May 14, 1998]

Sec. 493.1777  Standard: Inspection of laboratories that have requested 
          or have been issued a certificate of compliance.

    (a) Initial inspection. (1) A laboratory issued a registration 
certificate must permit an initial inspection to assess the laboratory's 
compliance with the requirements of this part before HCFA issues a 
certificate of compliance.
    (2) The inspection may occur at any time during the laboratory's 
hours of operation.
    (b) Subsequent inspections. (1) HCFA or a HCFA agent may conduct 
subsequent inspections on a biennial basis or with such other frequency 
as HCFA determines to be necessary to ensure compliance with the 
requirements of this part.
    (2) HCFA bases the nature of subsequent inspections on the 
laboratory's compliance history.
    (c) Provider-performed microscopy procedures. The inspection sample 
for review may include testing in the subcategory of provider-performed 
microscopy procedures.
    (d) Compliance with basic inspection requirements. The laboratory 
must comply with the basic inspection requirements of Sec. 493.1773.

[63 FR 26738, May 14, 1998]

Sec. 493.1780  Standard: Inspection of CLIA-exempt laboratories or 
          laboratories requesting or issued a certificate of 
          accreditation.

    (a) Validation inspection. HCFA or a HCFA agent may conduct a 
validation inspection of any accredited or CLIA-exempt laboratory at any 
time during its hours of operation.
    (b) Complaint inspection. HCFA or a HCFA agent may conduct a 
complaint inspection of a CLIA-exempt laboratory or a laboratory 
requesting or issued a certificate of accreditation at any time during 
its hours of operation upon receiving a complaint applicable to the 
requirements of this part.
    (c) Noncompliance determination. If a validation or complaint 
inspection results in a finding that the laboratory is not in compliance 
with one or more condition-level requirements, the following actions 
occur:
    (1) A laboratory issued a certificate of accreditation is subject to 
a full review by HCFA, in accordance with subpart E of this part and 
Sec. 488.11 of this chapter.
    (2) A CLIA-exempt laboratory is subject to appropriate enforcement 
actions under the approved State licensure program.
    (d) Compliance with basic inspection requirements. CLIA-exempt 
laboratories and laboratories requesting or issued a certificate of 
accreditation must comply with the basic inspection requirements in 
Sec. 493.1773.

[63 FR 26738, May 14, 1998]

                    Subpart R--Enforcement Procedures

    Source: 57 FR 7237, Feb. 28, 1992, unless otherwise noted.


[[Page 929]]



Sec. 493.1800  Basis and scope.

    (a) Statutory basis. (1) Section 1846 of the Act--
    (i) Provides for intermediate sanctions that may be imposed on 
laboratories that perform clinical diagnostic tests on human specimens 
when those laboratories are found to be out of compliance with one or 
more of the conditions for Medicare coverage of their services; and
    (ii) Requires the Secretary to develop and implement a range of such 
sanctions, including four that are specified in the statute.
    (2) The Clinical Laboratories Improvement Act of 1967 (section 353 
of the Public Health Service Act) as amended by CLIA '88--
    (i) Establishes requirements for all laboratories that perform 
clinical diagnostic tests on human specimens;
    (ii) Requires a Federal certification scheme to be applied to all 
such laboratories; and
    (iii) Grants the Secretary broad enforcement authority, including--
    (A) Use of intermediate sanctions;
    (B) Suspension, limitation, or revocation of the certificate of a 
laboratory that is out of compliance with one or more requirements for a 
certificate; and
    (C) Civil suit to enjoin any laboratory activity that constitutes a 
significant hazard to the public health.
    (3) Section 353 also--
    (i) Provides for imprisonment or fine for any person convicted of 
intentional violation of CLIA requirements;
    (ii) Specifies the administrative hearing and judicial review rights 
of a laboratory that is sanctioned under CLIA; and
    (iii) Requires the Secretary to publish annually a list of all 
laboratories that have been sanctioned during the preceding year.
    (b) Scope and applicability. This subpart sets forth--
    (1) The policies and procedures that HCFA follows to enforce the 
requirements applicable to laboratories under CLIA and under section 
1846 of the Act; and
    (2) The appeal rights of laboratories on which HCFA imposes 
sanctions.

Sec. 493.1804  General considerations.

    (a) Purpose. The enforcement mechanisms set forth in this subpart 
have the following purposes:
    (1) To protect all individuals served by laboratories against 
substandard testing of specimens.
    (2) To safeguard the general public against health and safety 
hazards that might result from laboratory activities.
    (3) To motivate laboratories to comply with CLIA requirements so 
that they can provide accurate and reliable test results.
    (b) Basis for decision to impose sanctions. (1) HCFA's decision to 
impose sanctions is based on one or more of the following:
    (i) Deficiencies found by HCFA or its agents in the conduct of 
inspections to certify or validate compliance with Federal requirements, 
or through review of materials submitted by the laboratory (e.g., 
personnel qualifications).
    (ii) Unsuccessful participation in proficiency testing.
    (2) HCFA imposes one or more of the alternative or principal 
sanctions specified in Secs. 493.1806 and 493.1807 when HCFA or HCFA's 
agent finds that a laboratory has condition-level deficiencies.
    (c) Imposition of alternative sanctions. (1) HCFA may impose 
alternative sanctions in lieu of, or in addition to principal sanctions, 
(HCFA does not impose alternative sanctions on laboratories that have 
certificates of waiver because those laboratories are not inspected for 
compliance with condition-level requirements.)
    (2) HCFA may impose alternative sanctions other than a civil money 
penalty after the laboratory has had an opportunity to respond, but 
before the hearing specified in Sec. 493.1844.
    (d) Choice of sanction: Factors considered. HCFA bases its choice of 
sanction or sanctions on consideration of one or more factors that 
include, but are not limited to, the following, as assessed by the State 
or by HCFA, or its agents:
    (1) Whether the deficiencies pose immediate jeopardy.
    (2) The nature, incidence, severity, and duration of the 
deficiencies or noncompliance.

[[Page 930]]

    (3) Whether the same condition level deficiencies have been 
identified repeatedly.
    (4) The accuracy and extent of laboratory records (e.g., of remedial 
action) in regard to the noncompliance, and their availability to the 
State, to other HCFA agents, and to HCFA.
    (5) The relationship of one deficiency or group of deficiencies to 
other deficiencies.
    (6) The overall compliance history of the laboratory including but 
not limited to any period of noncompliance that occurred between 
certifications of compliance.
    (7) The corrective and long-term compliance outcomes that HCFA hopes 
to achieve through application of the sanction.
    (8) Whether the laboratory has made any progress toward improvement 
following a reasonable opportunity to correct deficiencies.
    (9) Any recommendation by the State agency as to which sanction 
would be appropriate.
    (e) Number of alternative sanctions. HCFA may impose a separate 
sanction for each condition level deficiency or a single sanction for 
all condition level deficiencies that are interrelated and subject to 
correction by a single course of action.
    (f) Appeal rights. The appeal rights of laboratories dissatisfied 
with the imposition of a sanction are set forth in Sec. 493.1844.

[57 FR 7237, Feb. 28, 1992; 57 FR 35761, Aug. 11, 1992, as amended at 60 
FR 20051, Apr. 24, 1995]

Sec. 493.1806  Available sanctions: All laboratories.

    (a) Applicability. HCFA may impose one or more of the sanctions 
specified in this section on a laboratory that is out of compliance with 
one or more CLIA conditions.
    (b) Principal sanction. HCFA may impose any of the three principal 
CLIA sanctions, which are suspension, limitation, or revocation of any 
type of CLIA certificate.
    (c) Alternative sanctions. HCFA may impose one or more of the 
following alternative sanctions in lieu of or in addition to imposing a 
principal sanction, except on a laboratory that has a certificate of 
waiver.
    (1) Directed plan of correction, as set forth at Sec. 493.1832.
    (2) State onsite monitoring as set forth at Sec. 493.1836.
    (3) Civil money penalty, as set forth at Sec. 493.1834.
    (d) Civil suit. HCFA may bring suit in the appropriate U.S. District 
Court to enjoin continuation of any activity of any laboratory 
(including a CLIA-exempt laboratory that has been found with 
deficiencies during a validation survey), if HCFA has reason to believe 
that continuation of the activity would constitute a significant hazard 
to the public health.
    (e) Criminal sanctions. Under section 353(1) of the PHS Act, an 
individual who is convicted of intentionally violating any CLIA 
requirement may be imprisoned or fined.

[57 FR 7237, Feb. 28, 1992, as amended at 58 FR 5237, Jan. 19, 1993]

Sec. 493.1807  Additional sanctions: Laboratories that participate in 
          Medicare.

    The following additional sanctions are available for laboratories 
that are out of compliance with one or more CLIA conditions and that 
have approval to receive Medicare payment for their services.
    (a) Principal sanction. Cancellation of the laboratory's approval to 
receive Medicare payment for its services.
    (b) Alternative sanctions. (1) Suspension of payment for tests in 
one or more specific specialties or subspecialties, performed on or 
after the effective date of sanction.
    (2) Suspension of payment for all tests in all specialties and 
subspecialties performed on or after the effective date of sanction.

Sec. 493.1808  Adverse action on any type of CLIA certificate: Effect on 
          Medicare approval.

    (a) Suspension or revocation of any type of CLIA certificate. When 
HCFA suspends or revokes any type of CLIA certificate, HCFA concurrently 
cancels the laboratory's approval to receive Medicare payment for its 
services.
    (b) Limitation of any type of CLIA certificate. When HCFA limits any 
type of CLIA certificate, HCFA concurrently limits Medicare approval to 
only those

[[Page 931]]

specialties or subspecialties that are authorized by the laboratory's 
limited certificate.

Sec. 493.1809  Limitation on Medicaid payment.

    As provided in section 1902(a)(9)(C) of the Act, payment for 
laboratory services may be made under the State plan only if those 
services are furnished by a laboratory that has a CLIA certificate or is 
licensed by a State whose licensure program has been approved by the 
Secretary under this part.

[57 FR 7237, Feb. 28, 1992; 57 FR 35761, Aug. 11, 1992]

Sec. 493.1810  Imposition and lifting of alternative sanctions.

    (a) Notice of noncompliance and of proposed sanction: Content. If 
HCFA or its agency identifies condition level noncompliance in a 
laboratory, HCFA or its agent gives the laboratory written notice of the 
following:
    (1) The condition level noncompliance that it has identified.
    (2) The sanction or sanctions that HCFA or its agent proposes to 
impose against the laboratory.
    (3) The rationale for the proposed sanction or sanctions.
    (4) The projected effective date and duration of the proposed 
sanction or sanctions.
    (5) The authority for the proposed sanction or sanctions.
    (6) The time allowed (at least 10 days) for the laboratory to 
respond to the notice.
    (b) Opportunity to respond. During the period specified in paragraph 
(a)(6) of this section, the laboratory may submit to HCFA or its agent 
written evidence or other information against the imposition of the 
proposed sanction or sanctions.
    (c) Notice of imposition of sanction--(1) Content. HCFA gives the 
laboratory written notice that acknowledges any evidence or information 
received from the laboratory and specifies the following:
    (i) The sanction or sanctions to be imposed against the laboratory.
    (ii) The authority and rationale for the imposing sanction or 
sanctions.
    (iii) The effective date and duration of sanction.
    (2) Timing. (i) If HCFA or its agent determines that the 
deficiencies pose immediate jeopardy, HCFA provides notice at least 5 
days before the effective date of sanction.
    (ii) If HCFA or its agent determines that the deficiencies do not 
pose immediate jeopardy, HCFA provides notice at least 15 days before 
the effective date of the sanction.
    (d) Duration of alternative sanctions. An alternative sanction 
continues until the earlier of the following occurs:
    (1) The laboratory corrects all condition level deficiencies.
    (2) HCFA's suspension, limitation, or revocation of the laboratory's 
CLIA certificate becomes effective.
    (e) Lifting of alternative sanctions--(1) General rule. Alternative 
sanctions are not lifted until a laboratory's compliance with all 
condition level requirements is verified.
    (2) Credible allegation of compliance. When a sanctioned laboratory 
submits a credible allegation of compliance, HCFA's agent determines 
whether--
    (i) It can certify compliance on the basis of the evidence presented 
by the laboratory in its allegation; or
    (ii) It must revisit to verify whether the laboratory has, in fact, 
achieved compliance.
    (3) Compliance achieved before the date of revisit. If during a 
revisit, the laboratory presents credible evidence (as determined by 
HCFA or its agent) that it achieved compliance before the date of 
revisit, sanctions are lifted as of that earlier date.

Sec. 493.1812  Action when deficiencies pose immediate jeopardy.

    If a laboratory's deficiencies pose immediate jeopardy, the 
following rules apply:
    (a) HCFA requires the laboratory to take immediate action to remove 
the jeopardy and may impose one or more alternative sanctions to help 
bring the laboratory into compliance.
    (b) If the findings of a revisit indicate that a laboratory has not 
eliminated the jeopardy, HCFA suspends or limits the laboratory's CLIA 
certificate no earlier than 5 days after the date of notice of 
suspension or limitation. HCFA may later revoke the certificate.

[[Page 932]]

    (c) In addition, if HCFA has reason to believe that the continuation 
of any activity by any laboratory (either the entire laboratory 
operation or any specialty or subspecialty of testing) would constitute 
a significant hazard to the public health, HCFA may bring suit and seek 
a temporary injunction or restraining order against continuation of that 
activity by the laboratory, regardless of the type of CLIA certificate 
the laboratory has and of whether it is State-exempt.

Sec. 493.1814  Action when deficiencies are at the condition level but 
          do not pose immediate jeopardy.

    If a laboratory has condition level deficiencies that do not pose 
immediate jeopardy, the following rules apply:
    (a) Initial action. (1) HCFA may cancel the laboratory's approval to 
receive Medicare payment for its services.
    (2) HCFA may suspend, limit, or revoke the laboratory's CLIA 
certificate.
    (3) If HCFA does not impose a principal sanction under paragraph 
(a)(1) or (a)(2) of this section, it imposes one or more alternative 
sanctions. In the case of unsuccessful participation in proficiency 
testing, HCFA may impose the training and technical assistance 
requirement set forth at Sec. 493.1838 in lieu of, or in addition to, 
one or more alternative sanctions.
    (b) Failure to correct condition level deficiencies. If HCFA imposes 
alternative sanctions for condition level deficiencies that do not pose 
immediate jeopardy, and the laboratory does not correct the condition 
level deficiencies within 12 months after the last day of inspection, 
HCFA--
    (1) Cancels the laboratory's approval to receive Medicare payment 
for its services, and discontinues the Medicare payment sanctions as of 
the day cancellation is effective.
    (2) Following a revisit which indicates that the laboratory has not 
corrected its condition level deficiencies, notifies the laboratory that 
it proposes to suspend, limit, or revoke the certificate, as specified 
in Sec. 493.1816(b), and the laboratory's right to hearing; and
    (3) May impose (or continue, if already imposed) any alternative 
sanctions that do not pertain to Medicare payments. (Sanctions imposed 
under the authority of section 353 of the PHS Act may continue for more 
than 12 months from the last date of inspection, while a hearing on the 
proposed suspension, limitation, or revocation of the certificate of 
compliance, registration certificate, certificate of accreditation, or 
certificate for PPM procedures is pending.)
    (c) Action after hearing. If a hearing decision upholds a proposed 
suspension, limitation, or revocation of a laboratory's CLIA 
certificate, HCFA discontinues any alternative sanctions as of the day 
it makes the suspension, limitation, or revocation effective.

[57 FR 7237, Feb. 28, 1992, as amended at 60 FR 20051, Apr. 24, 1995]

Sec. 493.1816  Action when deficiencies are not at the condition level.

    If a laboratory has deficiencies, that are not at the condition 
level, the following rules apply:
    (a) Initial action. The laboratory must submit a plan of correction 
that is acceptable to HCFA in content and time frames.
    (b) Failure to correct deficiencies. If, on revisit, it is found 
that the laboratory has not corrected the deficiencies within 12 months 
after the last day of inspection, the following rules apply:
    (1) HCFA cancels the laboratory's approval to receive Medicare 
payment for its services.
    (2) HCFA notifies the laboratory of its intent to suspend, limit, or 
revoke the laboratory's CLIA certificate and of the laboratory's right 
to a hearing.

Sec. 493.1820  Ensuring timely correction of deficiencies.

    (a) Timing of visits. HCFA, the State survey agency or other HCFA 
agent may visit the laboratory at any time to evaluate progress, and at 
the end of the period to determine whether all corrections have been 
made.
    (b) Deficiencies corrected before a visit. If during a visit, a 
laboratory produces credible evidence that it achieved compliance before 
the visit, the sanctions are lifted as of that earlier date.

[[Page 933]]

    (c) Failure to correct deficiencies. If during a visit it is found 
that the laboratory has not corrected its deficiencies, HCFA may propose 
to suspend, limit, or revoke the laboratory's CLIA certificate.
    (d) Additional time for correcting lower level deficiencies not at 
the condition level. If at the end of the plan of correction period all 
condition level deficiencies have been corrected, and there are 
deficiencies, that are not at the condition level, HCFA may request a 
revised plan of correction. The revised plan may not extend beyond 12 
months from the last day of the inspection that originally identified 
the cited deficiencies.
    (e) Persistence of deficiencies. If at the end of the period covered 
by the plan of correction, the laboratory still has deficiencies, the 
rules of Secs. 493.1814 and 493.1816 apply.

Sec. 493.1826  Suspension of part of Medicare payments.

    (a) Application. (1) HCFA may impose this sanction if a laboratory--
    (i) Is found to have condition level deficiencies with respect to 
one or more specialties or subspecialties of tests; and
    (ii) Agrees (in return for not having its Medicare approval 
cancelled immediately) not to charge Medicare beneficiaries or their 
private insurance carriers for the services for which Medicare payment 
is suspended.
    (2) HCFA suspends Medicare payment for those specialities or 
subspecialties of tests for which the laboratory is out of compliance 
with Federal requirements.
    (b) Procedures. Before imposing this sanction, HCFA provides notice 
of sanction and opportunity to respond in accordance with Sec. 493.1810.
    (c) Duration and effect of sanction. This sanction continues until 
the laboratory corrects the condition level deficiencies or HCFA cancels 
the laboratory's approval to receive Medicare payment for its services, 
but in no event longer than 12 months.
    (1) If the laboratory corrects all condition level deficiencies, 
HCFA resumes Medicare payment effective for all services furnished on or 
after the date the deficiencies are corrected.
    (2) [Reserved]

[57 FR 7237, Feb. 28, 1992; 57 FR 35761, Aug. 11, 1992]

Sec. 493.1828  Suspension of all Medicare payments.

    (a) Application. (1) HCFA may suspend payment for all Medicare-
approved laboratory services when the laboratory has condition level 
deficiencies.
    (2) HCFA suspends payment for all Medicare covered laboratory 
services when the following conditions are met:
    (i) Either--
    (A) The laboratory has not corrected its condition level 
deficiencies included in the plan of correction within 3 months from the 
last date of inspection; or
    (B) The laboratory has been found to have the same condition level 
deficiencies during three consecutive inspections; and
    (ii) The laboratory has chosen (in return for not having its 
Medicare approval immediately cancelled), to not charge Medicare 
beneficiaries or their private insurance carriers for services for which 
Medicare payment is suspended.
    (3) HCFA suspends payment for services furnished on and after the 
effective date of sanction.
    (b) Procedures. Before imposing this sanction, HCFA provides notice 
of sanction and opportunity to respond in accordance with Sec. 493.1810.
    (c) Duration and effect of sanction. (1) Suspension of payment 
continues until all condition level deficiencies are corrected, but 
never beyond twelve months.
    (2) If all the deficiencies are not corrected by the end of the 12 
month period, HCFA cancels the laboratory's approval to receive Medicare 
payment for its services.

Sec. 493.1832  Directed plan of correction and directed portion of a 
          plan of correction.

    (a) Application. HCFA may impose a directed plan of correction as an 
alternative sanction for any laboratory that has condition level 
deficiencies. If HCFA does not impose a directed plan of correction as 
an alternative sanction for a laboratory that has condition

[[Page 934]]

level deficiencies, it at least imposes a directed portion of a plan of 
correction when it imposes any of the following alternative sanctions:
    (1) State onsite monitoring.
    (2) Civil money penalty.
    (3) Suspension of all or part of Medicare payments.
    (b) Procedures--(1) Directed plan of correction. When imposing this 
sanction, HCFA--
    (i) Gives the laboratory prior notice of the sanction and 
opportunity to respond in accordance with Sec. 493.1810;
    (ii) Directs the laboratory to take specific corrective action 
within specific time frames in order to achieve compliance; and
    (iii) May direct the laboratory to submit the names of laboratory 
clients for notification purposes, as specified in paragraph (b)(3) of 
this section.
    (2) Directed portion of a plan of correction. HCFA may decide to 
notify clients of a sanctioned laboratory, because of the seriousness of 
the noncompliance (e.g., the existence of immediate jeopardy) or for 
other reasons. When imposing this sanction, HCFA takes the following 
steps--
    (i) Directs the laboratory to submit to HCFA, the State survey 
agency, or other HCFA agent, within 10 calendar days after the notice of 
the alternative sanction, a list of names and addresses of all 
physicians, providers, suppliers, and other clients who have used some 
or all of the services of the laboratory since the last certification 
inspection or within any other timeframe specified by HCFA.
    (ii) Within 30 calendar days of receipt of the information, may send 
to each laboratory client, via the State survey agency, a notice 
containing the name and address of the laboratory, the nature of the 
laboratory's noncompliance, and the kind and effective date of the 
alternative sanction.
    (iii) Sends to each laboratory client, via the State survey agency, 
notice of the recission of an adverse action within 30 days of the 
rescission.
    (3) Notice of imposition of a principal sanction following the 
imposition of an alternative sanction. If HCFA imposes a principal 
sanction following the imposition of an alternative sanction, and for 
which HCFA has already obtained a list of laboratory clients, HCFA may 
use that list to notify the clients of the imposition of the principal 
sanction.
    (c) Duration of a directed plan of correction. If HCFA imposes a 
directed plan of correction, and on revisit it is found that the 
laboratory has not corrected the deficiencies within 12 months from the 
last day of inspection, the following rules apply:
    (1) HCFA cancels the laboratory's approval for Medicare payment of 
its services, and notifies the laboratory of HCFA's intent to suspend, 
limit, or revoke the laboratory's CLIA certificate.
    (2) The directed plan of correction continues in effect until the 
day suspension, limitation, or revocation of the laboratory's CLIA 
certificate.

Sec. 493.1834  Civil money penalty.

    (a) Statutory basis. Sections 1846 of the Act and 353(h)(2)(B) of 
the PHS Act authorize the Secretary to impose civil money penalties on 
laboratories. Section 1846(b)(3) of the Act specifically provides that 
incrementally more severe fines may be imposed for repeated or 
uncorrected deficiencies.
    (b) Scope. This section sets forth the procedures that HCFA follows 
to impose a civil money penalty in lieu of, or in addition to, 
suspending, limiting, or revoking the certificate of compliance, 
registration certificate, certificate of accreditation, or certificate 
for PPM procedures of a laboratory that is found to have condition level 
deficiencies.
    (c) Basis for imposing a civil money penalty. HCFA may impose a 
civil money penalty against any laboratory determined to have condition 
level deficiencies regardless of whether those deficiencies pose 
immediate jeopardy.
    (d) Amount of penalty--(1) Factors considered. In determining the 
amount of the penalty, HCFA takes into account the following factors:
    (i) The nature, scope, severity, and duration of the noncompliance.
    (ii) Whether the same condition level deficiencies have been 
identified during three consecutive inspections.
    (iii) The laboratory's overall compliance history including but not 
limited to any period of noncompliance that occurred between 
certifications of compliance.

[[Page 935]]

    (iv) The laboratory's intent or reason for noncompliance.
    (v) The accuracy and extent of laboratory records and their 
availability to HCFA, the State survey agency, or other HCFA agent.
    (2) Range of penalty amount.
    (i) For a condition level deficiency that poses immediate jeopardy, 
the range is $3,050-$10,000 per day of noncompliance or per violation.
    (ii) For a condition level deficiency that does not pose immediate 
jeopardy, the range is $50-$3,000 per day of noncompliance or per 
violation.
    (3) Decreased penalty amounts. If the immediate jeopardy is removed, 
but the deficiency continues, HCFA shifts the penalty amount to the 
lower range.
    (4) Increased penalty amounts. HCFA may, before the hearing, propose 
to increase the penalty amount for a laboratory that has deficiencies 
which, after imposition of a lower level penalty amount, become 
sufficiently serious to pose immediate jeopardy.
    (e) Procedures for imposition of civil money penalty--(1) Notice of 
intent. (i) HCFA sends the laboratory written notice, of HCFA's intent 
to impose a civil money penalty.
    (ii) The notice includes the following information:
    (A) The statutory basis for the penalty.
    (B) The proposed daily or per violation amount of the penalty.
    (C) The factors (as described in paragraph (d)(1) of this section) 
that HCFA considered.
    (D) The opportunity for responding to the notice in accordance with 
Sec. 493.1810(c).
    (E) A specific statement regarding the laboratory's appeal rights.
    (2) Appeal rights. (i) The laboratory has 60 days from the date of 
receipt of the notice of intent to impose a civil money penalty to 
request a hearing in accordance with Sec. 493.1844(g).
    (ii) If the laboratory requests a hearing, all other pertinent 
provisions of Sec. 493.1844 apply.
    (iii) If the laboratory does not request a hearing, HCFA may reduce 
the proposed penalty amount by 35 percent.
    (f) Accrual and duration of penalty--(1) Accrual of penalty. The 
civil money penalty begins accruing as follows:
    (i) 5 days after notice of intent if there is immediate jeopardy.
    (ii) 15 days after notice of intent if there is not immediate 
jeopardy.
    (2) Duration of penalty. The civil money penalty continues to accrue 
until the earliest of the following occurs:
    (i) The laboratory's compliance with condition level requirements is 
verified on the basis of the evidence presented by the laboratory in its 
credible allegation of compliance or at the time or revisit.
    (ii) Based on credible evidence presented by the laboratory at the 
time of revisit, HCFA determines that compliance was achieved before the 
revisit. (In this situation, the money penalty stops accruing as of the 
date of compliance.)
    (iii) HCFA suspends, limits, or revokes the laboratory's certificate 
of compliance, registration certificate, certificate of accreditation, 
or certificate for PPM procedures.
    (g) Computation and notice of total penalty amount--(1) Computation. 
HCFA computes the total penalty amount after the laboratory's compliance 
is verified or HCFA suspends, limits, or revokes the laboratory's CLIA 
certificate but in no event before--
    (i) The 60 day period for requesting a hearing has expired without a 
request or the laboratory has explicitly waived its right to a hearing; 
or
    (ii) Following a hearing requested by the laboratory, the ALJ issues 
a decision that upholds imposition of the penalty.
    (2) Notice of penalty amount and due date of penalty. The notice 
includes the following information:
    (i) Daily or per violation penalty amount.
    (ii) Number of days or violations for which the penalty is imposed.
    (iii) Total penalty amount.
    (iv) Due date for payment of the penalty.
    (h) Due date for payment of penalty. (1) Payment of a civil money 
penalty is due 15 days from the date of the notice specified in 
paragraph (g)(2) of this section.

[[Page 936]]

    (2) HCFA may approve a plan for a laboratory to pay a civil money 
penalty, plus interest, over a period of up to one year from the 
original due date.
    (i) Collection and settlement--(1) Collection of penalty amounts. 
(i) The determined penalty amount may be deducted from any sums then or 
later owing by the United States to the laboratory subject to the 
penalty.
    (ii) Interest accrues on the unpaid balance of the penalty, 
beginning on the due date. Interest is computed at the rate specified in 
Sec. 405.378(d) of this chapter.
    (2) Settlement. HCFA has authority to settle any case at any time 
before the ALJ issues a hearing decision.

[57 FR 7237, Feb. 28, 1992, as amended at 60 FR 20051, Apr. 24, 1995; 61 
FR 63749, Dec. 2, 1996]

Sec. 493.1836  State onsite monitoring.

    (a) Application. (1) HCFA may require continuous or intermittent 
monitoring of a plan of correction by the State survey agency to ensure 
that the laboratory makes the improvements necessary to bring it into 
compliance with the condition level requirements. (The State monitor 
does not have management authority, that is, cannot hire or fire staff, 
obligate funds, or otherwise dictate how the laboratory operates. The 
monitor's responsibility is to oversee whether corrections are made.)
    (2) The laboratory must pay the costs of onsite monitoring by the 
State survey agency.
    (i) The costs are computed by multiplying the number of hours of 
onsite monitoring in the laboratory by the hourly rate negotiated by 
HCFA and the State.
    (ii) The hourly rate includes salary, fringe benefits, travel, and 
other direct and indirect costs approved by HCFA.
    (b) Procedures. Before imposing this sanction, HCFA provides notice 
of sanction and opportunity to respond in accordance with Sec. 493.1810.
    (c) Duration of sanction. (1) If HCFA imposes onsite monitoring, the 
sanction continues until HCFA determines that the laboratory has the 
capability to ensure compliance with all condition level requirements.
    (2) If the laboratory does not correct all deficiencies within 12 
months, and a revisit indicates that deficiencies remain, HCFA cancels 
the laboratory's approval for Medicare payment for its services and 
notifies the laboratory of its intent to suspend, limit, or revoke the 
laboratory's certificate of compliance, registration certificate, 
certificate of accreditation, or certificate for PPM procedures.
    (3) If the laboratory still does not correct its deficiencies, the 
Medicare sanction continues until the suspension, limitation, or 
revocation of the laboratory's certificate of compliance, registration 
certificate, certificate of accreditation, or certificate for PPM 
procedures is effective.

[57 FR 7237, Feb. 28, 1992, as amended at 60 FR 20051, Apr. 24, 1995]

Sec. 493.1838  Training and technical assistance for unsuccessful 
          participation in proficiency testing.

    If a laboratory's participation in proficiency testing is 
unsuccessful, HCFA may require the laboratory to undertake training of 
its personnel, or to obtain necessary technical assistance, or both, in 
order to meet the requirements of the proficiency testing program. This 
requirement is separate from the principal and alternative sanctions set 
forth in Secs. 493.1806 and 493.1807.

Sec. 493.1840  Suspension, limitation, or revocation of any type of CLIA 
          certificate.

    (a) Adverse action based on actions of the laboratory's owner, 
operator or employees. HCFA may initiate adverse action to suspend, 
limit or revoke any CLIA certificate if HCFA finds that a laboratory's 
owner or operator or one of its employees has--
    (1) Been guilty of misrepresentation in obtaining a CLIA 
certificate;
    (2) Performed, or represented the laboratory as entitled to perform, 
a laboratory examination or other procedure that is not within a 
category of laboratory examinations or other procedures authorized by 
its CLIA certificate;
    (3) Failed to comply with the certificate requirements and 
performance standards;
    (4) Failed to comply with reasonable requests by HCFA for any 
information

[[Page 937]]

or work on materials that HCFA concludes is necessary to determine the 
laboratory's continued eligibility for its CLIA certificate or continued 
compliance with performance standards set by HCFA;
    (5) Refused a reasonable request by HCFA or its agent for permission 
to inspect the laboratory and its operation and pertinent records during 
the hours that the laboratory is in operation;
    (6) Violated or aided and abetted in the violation of any provisions 
of CLIA and its implementing regulations;
    (7) Failed to comply with an alternative sanction imposed under this 
subpart; or
    (8) Within the preceding two-year period, owned or operated a 
laboratory that had its CLIA certificate revoked. (This provision 
applies only to the owner or operator, not to all of the laboratory's 
employees.)
    (b) Adverse action based on improper referrals in proficiency 
testing. If HCFA determines that a laboratory has intentionally referred 
its proficiency testing samples to another laboratory for analysis, HCFA 
revokes the laboratory's CLIA certificate for at least one year, and may 
also impose a civil money penalty.
    (c) Adverse action based on exclusion from Medicare. If the OIG 
excludes a laboratory from participation in Medicare, HCFA suspends the 
laboratory's CLIA certificate for the period during which the laboratory 
is excluded.
    (d) Procedures for suspension or limitation--(1) Basic rule. Except 
as provided in paragraph (d)(2) of this section, HCFA does not suspend 
or limit a CLIA certificate until after an ALJ hearing decision (as 
provided in Sec. 493.1844) that upholds suspension or limitation.
    (2) Exceptions. HCFA may suspend or limit a CLIA certificate before 
the ALJ hearing in any of the following circumstances:
    (i) The laboratory's deficiencies pose immediate jeopardy.
    (ii) The laboratory has refused a reasonable request for information 
or work on materials.
    (iii) The laboratory has refused permission for HCFA or a HCFA agent 
to inspect the laboratory or its operation.
    (e) Procedures for revocation. (1) HCFA does not revoke any type of 
CLIA certificate until after an ALJ hearing that upholds revocation.
    (2) HCFA may revoke a CLIA certificate after the hearing decision 
even if it had not previously suspended or limited that certificate.
    (f) Notice to the OIG. HCFA notifies the OIG of any violations under 
paragraphs (a)(1), (a)(2), (a)(6), and (b) of this section within 30 
days of the determination of the violation.

Sec. 493.1842  Cancellation of Medicare approval.

    (a) Basis for cancellation. (1) HCFA always cancels a laboratory's 
approval to receive Medicare payment for its services if HCFA suspends 
or revokes the laboratory's CLIA certificate.
    (2) HCFA may cancel the laboratory's approval under any of the 
following circumstances:
    (i) The laboratory is out of compliance with a condition level 
requirement.
    (ii) The laboratory fails to submit a plan of correction 
satisfactory to HCFA.
    (iii) The laboratory fails to correct all its deficiencies within 
the time frames specified in the plan of correction.
    (b) Notice and opportunity to respond. Before canceling a 
laboratory's approval to receive Medicare payment for its services, HCFA 
gives the laboratory--
    (1) Written notice of the rationale for, effective date, and effect 
of, cancellation;
    (2) Opportunity to submit written evidence or other information 
against cancellation of the laboratory's approval.
    This sanction may be imposed before the hearing that may be 
requested by a laboratory, in accordance with the appeals procedures set 
forth in Sec. 493.1844.
    (c) Effect of cancellation. Cancellation of Medicare approval 
terminates any Medicare payment sanctions regardless of the time frames 
originally specified.

Sec. 493.1844  Appeals procedures.

    (a) General rules. (1) The provisions of this section apply to all 
laboratories and prospective laboratories that are

[[Page 938]]

dissatisfied with any initial determination under paragraph (b) of this 
section.
    (2) Hearings are conducted in accordance with procedures set forth 
in subpart D of part 498 of this chapter, except that the authority to 
conduct hearings and issue decisions may be exercised by ALJs assigned 
to, or detailed to, the Departmental Appeals Board.
    (3) Any party dissatisfied with a hearing decision is entitled to 
request review of the decision as specified in subpart E of part 498 of 
this chapter, except that the authority to review the decision may be 
exercised by the Departmental Appeals Board.
    (4) When more than one of the actions specified in paragraph (b) of 
this section are carried out concurrently, the laboratory has a right to 
only one hearing on all matters at issue.
    (b) Actions that are initial determinations. The following actions 
are initial determinations and therefore are subject to appeal in 
accordance with this section:
    (1) The suspension, limitation, or revocation of the laboratory's 
CLIA certificate by HCFA because of noncompliance with CLIA 
requirements.
    (2) The denial of a CLIA certificate.
    (3) The imposition of alternative sanctions under this subpart (but 
not the determination as to which alternative sanction or sanctions to 
impose).
    (4) The denial or cancellation of the laboratory's approval to 
receive Medicare payment for its services.
    (c) Actions that are not initial determinations. Actions that are 
not listed in paragraph (b) of this section are not initial 
determinations and therefore are not subject to appeal under this 
section. They include, but are not necessarily limited to, the 
following:
    (1) The finding that a laboratory accredited by a HCFA-approved 
accreditation organization is no longer deemed to meet the conditions 
set forth in subparts H, J, K, M, P, and Q of this part. However, the 
suspension, limitation or revocation of a certificate of accreditation 
is an initial determination and is appealable.
    (2) The finding that a laboratory determined to be in compliance 
with condition-level requirements but has deficiencies that are not at 
the condition level.
    (3) The determination not to reinstate a suspended CLIA certificate 
because the reason for the suspension has not been removed or there is 
insufficient assurance that the reason will not recur.
    (4) The determination as to which alternative sanction or sanctions 
to impose, including the amount of a civil money penalty to impose per 
day or per violation.
    (5) The denial of approval for Medicare payment for the services of 
a laboratory that does not have in effect a valid CLIA certificate.
    (6) The determination that a laboratory's deficiencies pose 
immediate jeopardy.
    (7) The amount of the civil money penalty assessed per day or for 
each violation of Federal requirements.
    (d) Effect of pending appeals--(1) Alternative sanctions. The 
effective date of an alternative sanction (other than a civil money 
penalty) is not delayed because the laboratory has appealed and the 
hearing or the hearing decision is pending.
    (2) Suspension, limitation, or revocation of a laboratory's CLIA 
certificate--(i) General rule. Except as provided in paragraph 
(d)(2)(ii) of this section, suspension, limitation, or revocation of a 
CLIA certificate is not effective until after a hearing decision by an 
ALJ is issued.
    (ii) Exceptions. (A) If HCFA determines that conditions at a 
laboratory pose immediate jeopardy, the effective date of the suspension 
or limitation of a CLIA certificate is not delayed because the 
laboratory has appealed and the hearing or the hearing decision is 
pending.
    (B) HCFA may suspend or limit a laboratory's CLIA certificate before 
an ALJ hearing or hearing decision if the laboratory has refused a 
reasonable request for information (including but not limited to billing 
information), or for work on materials, or has refused permission for 
HCFA or a HCFA agent to inspect the laboratory or its operation.
    (3) Cancellation of Medicare approval. The effective date of the 
cancellation

[[Page 939]]

of a laboratory's approval to receive Medicare payment for its services 
is not delayed because the laboratory has appealed and the hearing or 
hearing decision is pending.
    (4) Effect of ALJ decision. (i) An ALJ decision is final unless, as 
provided in paragraph (a)(3) of this section, one of the parties 
requests review by the Departmental Appeals Board within 60 days, and 
the Board reviews the case and issues a revised decision.
    (ii) If an ALJ decision upholds a suspension imposed because of 
immediate jeopardy, that suspension becomes a revocation.
    (e) Appeal rights for prospective laboratories--(1) Reconsideration. 
Any prospective laboratory dissatisfied with a denial of a CLIA 
certificate, or of approval for Medicare payment for its services, may 
initiate the appeals process by requesting reconsideration in accordance 
with Secs. 498.22 through 498.25 of this chapter.
    (2) Notice of reopening. If HCFA reopens an initial or reconsidered 
determination, HCFA gives the prospective laboratory notice of the 
revised determination in accordance with Sec. 498.32 of this chapter.
    (3) ALJ hearing. Any prospective laboratory dissatisfied with a 
reconsidered determination under paragraph (e)(1) of this section or a 
revised reconsidered determination under Sec. 498.30 of this chapter is 
entitled to a hearing before an ALJ, as specified in paragraph (a)(2) of 
this section.
    (4) Review of ALJ hearing decisions. Any prospective laboratory that 
is dissatisfied with an ALJ's hearing decision or dismissal of a request 
for hearing may file a written request for review by the Departmental 
Appeals Board as provided in paragraph (a)(3) of this section.
    (f) Appeal rights of laboratories--(1) ALJ hearing. Any laboratory 
dissatisfied with the suspension, limitation, or revocation of its CLIA 
certificate, with the imposition of an alternative sanction under this 
subpart, or with cancellation of the approval to receive Medicare 
payment for its services, is entitled to a hearing before an ALJ as 
specified in paragraph (a)(2) of this section and has 60 days from the 
notice of sanction to request a hearing.
    (2) Review of ALJ hearing decisions. Any laboratory that is 
dissatisfied with an ALJ's hearing decision or dismissal of a request 
for hearing may file a written request for review by the Departmental 
Appeals Board, as provided in paragraph (a)(3) of this section.
    (3) Judicial review. Any laboratory dissatisfied with the decision 
to impose a civil money penalty or to suspend, limit, or revoke its CLIA 
certificate may, within 60 days after the decision becomes final, file 
with the U.S. Court of Appeals of the circuit in which the laboratory 
has its principal place of business, a petition for judicial review.
    (g) Notice of adverse action. (1) If HCFA suspends, limits, or 
revokes a laboratory's CLIA certificate or cancels the approval to 
receive Medicare payment for its services, HCFA gives notice to the 
laboratory, and may give notice to physicians, providers, suppliers, and 
other laboratory clients, according to the procedures set forth at 
Sec. 493.1832. In addition, HCFA notifies the general public each time 
one of these principal sanctions is imposed.
    (2) The notice to the laboratory--
    (i) Sets forth the reasons for the adverse action, the effective 
date and effect of that action, and the appeal rights if any; and
    (ii) When the certificate is limited, specifies the specialties or 
subspecialties of tests that the laboratory is no longer authorized to 
perform, and that are no longer covered under Medicare.
    (3) The notice to other entities includes the same information 
except the information about the laboratory's appeal rights.
    (h) Effective date of adverse action. (1) When the laboratory's 
deficiencies pose immediate jeopardy, the effective date of the adverse 
action is at least 5 days after the date of the notice.
    (2) When HCFA determines that the laboratory's deficiencies do not 
pose immediate jeopardy, the effective date of the adverse action is at 
least 15 days after the date of the notice.

[57 FR 7237, Feb. 28, 1992; 57 FR 35761, Aug. 11, 1992]

Sec. 493.1846  Civil action.

    If HCFA has reason to believe that continuation of the activities of 
any laboratory, including a State-exempt

[[Page 940]]

laboratory, would constitute a significant hazard to the public health, 
HCFA may bring suit in a U.S. District Court to enjoin continuation of 
the specific activity that is causing the hazard or to enjoin the 
continued operation of the laboratory if HCFA deems it necessary. Upon 
proper showing, the court shall issue a temporary injunction or 
restraining order without bond against continuation of the activity.

Sec. 493.1850  Laboratory registry.

    (a) Once a year HCFA makes available to physicians and to the 
general public specific information (including information provided to 
HCFA by the OIG) that is useful in evaluating the performance of 
laboratories, including the following:
    (1) A list of laboratories that have been convicted, under Federal 
or State laws relating to fraud and abuse, false billing, or kickbacks.
    (2) A list of laboratories that have had their CLIA certificates 
suspended, limited, or revoked, and the reason for the adverse actions.
    (3) A list of persons who have been convicted of violating CLIA 
requirements, as specified in section 353(1) of the PHS Act, together 
with the circumstances of each case and the penalties imposed.
    (4) A list of laboratories on which alternative sanctions have been 
imposed, showing--
    (i) The effective date of the sanctions;
    (ii) The reasons for imposing them;
    (iii) Any corrective action taken by the laboratory; and
    (iv) If the laboratory has achieved compliance, the verified date of 
compliance.
    (5) A list of laboratories whose accreditation has been withdrawn or 
revoked and the reasons for the withdrawal or revocation.
    (6) All appeals and hearing decisions.
    (7) A list of laboratories against which HCFA has brought suit under 
Sec. 493.1846 and the reasons for those actions.
    (8) A list of laboratories that have been excluded from 
participation in Medicare or Medicaid and the reasons for the exclusion.
    (b) The laboratory registry is compiled for the calendar year 
preceding the date the information is made available and includes 
appropriate explanatory information to aid in the interpretation of the 
data. It also contains corrections of any erroneous statements or 
information that appeared in the previous registry.

Subpart S [Reserved]

                        Subpart T--Consultations

    Source: 57 FR 7185, Feb. 28, 1992, unless otherwise noted.

Sec. 493.2001  Establishment and function of the Clinical Laboratory 
          Improvement Advisory Committee.

    (a) HHS will establish a Clinical Laboratory Improvement Advisory 
Committee to advise and make recommendations on technical and scientific 
aspects of the provisions of this part 493.
    (b) The Clinical Laboratory Improvement Advisory Committee will be 
comprised of individuals involved in the provision of laboratory 
services, utilization of laboratory services, development of laboratory 
testing or methodology, and others as approved by HHS.
    (c) HHS will designate specialized subcommittees as necessary.
    (d) The Clinical Laboratory Improvement Advisory Committee or any 
designated subcommittees will meet as needed, but not less than once 
each year.
    (e) The Clinical Laboratory Improvement Advisory Committee or 
subcommittee, at the request of HHS, will review and make 
recommendations concerning:
    (1) Criteria for categorizing tests and examinations of moderate 
complexity (including the subcategory) and high complexity;
    (2) Determination of waived tests;
    (3) Personnel standards;
    (4) Patient test management, quality control, quality assurance 
standards;
    (5) Proficiency testing standards;
    (6) Applicability to the standards of new technology; and
    (7) Other issues relevant to part 493, if requested by HHS.

[[Page 941]]

    (f) HHS will be responsible for providing the data and information, 
as necessary, to the members of the Clinical Laboratory Improvement 
Advisory Committee.

[57 FR 7185, Feb. 28, 1992, as amended at 58 FR 5237, Jan. 19, 1993; 60 
FR 20051, Apr. 24, 1995]

                          PART 494  [RESERVED]