Search NIOSH | NIOSH Home | NIOSH Topics | Site Index | Databases and Information Resources | NIOSH Products | Contact Us |
1.0 Purpose:The purpose of this standard is to specify minimum requirements to determine the effectiveness of self-contained escape respirators that address CBRN materials identified as inhalation hazards from possible terrorist events for use by the general working population. The respirator must meet the minimum requirements identified in the following paragraphs:
2.0 Title 42, Code of Federal Regulations (CFR), Part 84:The following paragraphs of 42 CFR, Part 84 are applicable: 2.1 42 CFR, Part 84, Subparts A, B, D, E, F, and G: Subpart A: General Provisions 2.2 42 CFR, Part 84, Subpart H: Approval under Title 42, CFR, Part 84, Subpart H, for escape only, with a minimum service time of 15 minutes. 3.0 Requirements Based on Existing National and International Standards:3.1 Field of View: The CBRN self-contained escape respirator
shall obtain a Visual Field Score (VFS) of 70 or greater when tested in
accordance with NIOSH Standard Test Procedure CET-APRS- The VFS is determined by using a VFS grid (Dots on visual field) as defined in the American Medical Association Guides to the Evaluation of Permanent Impairment, 5th Edition (2000) that is overlaid on the diagram of the visual field plot obtained using the spherical shell of EN 136 apertometer or equivalent. The VFS score is the average of three fittings of the same respirator on the specified head form. 3.2 Fogging: The CBRN self-contained escape respirator shall demonstrate an average Visual Acuity Score (VAS) of greater or equal to 70 points for all measurements for each individual. A minimum of two respirators shall be evaluated. The respirator shall be donned by the test subject in an indoor ambient temperature of approximately 72+20F at 40+5% Relative Humidity (RH) and then shall enter into a simulated outdoor extreme temperature chamber where the visual acuity tests shall be administered. The self-contained escape respirator shall be tested for fogging in the hot/humid condition of 90+20F and 60+5% RH and the cold condition of 13+2oF. 3.3 Breathing Gas:
3.4 Flammability and Heat Resistance: Self-contained escape respirators submitted for approval shall be tested for Flammability and Heat Resistance using the test equipment specified in EN 136, Respiratory Protective Devices, Full Face Masks, Requirements, testing, Marking, 1998 Edition, Class 1 facepiece. No component of the respirator shall have an after flame after 5 seconds. No component of the escape respirator shall drip, melt, or develop a visible hole. The distance between the outer surface of the escape respirator and the burner shall be adjusted to 20 ± 2 mm. The pressure reducer shall be adjusted to 2.1 ± .05 psi. The temperature of the flame positioned above the burner tip shall be 800±500 C at a point 20+2 mm above the tip. The respirator shall be rotated once through the flame at a velocity of 6 ± 0.5 cm/s. Where components of the respirator such as valves, filters, etc. are arranged on the respirator, the test shall be repeated with these components at the appropriate height of 250 mm ± 6.4 mm.
4.0 Special CBRN Requirements:4.1 Duration/Service Life: The self-contained escape respirator shall have a minimum service life of 15 minutes. 4.2 Chemical Agent Permeation and Penetration Resistance Against Distilled Sulfur Mustard (HD) and Sarin (GB) Agent Requirement: The self-contained escape respirator system, including all components and accessories shall resist the permeation and penetration of Distilled Sulfur Mustard (HD) and Sarin (GB) chemical agents when tested on an upper-torso manikin. For closed circuit devices, they will be connected to a metabolic breathing simulator, using the following protocol: For a mean VO2 = 1.67 L/min for 30 minutes (aggregate VO2 = 50 L/minTime)
For open circuit devices, a breathing machine will be used, operating at an air flow rate of 19.5 Lpm,18 respirations per minute, 1.1 Liters tidal volume. Test requirements for Distilled Sulfur Mustard (HD) are shown in Table 3. Table 3.—Vapor-Liquid Sequential Challenge of Self-contained Escape Respirator with Distilled Sulfur Mustard (HD)
†Liquid volume is applied as 25 drops of equal size. ‡ Three consecutive sequential test data points at or exceeding 0.6 mg/m3 will collectively constitute a failure where each test value is based on a detector sample time of approximately 2 minutes. § The cumulative Ct including all maximum peak excursion data points must not be exceeded for the duration of the test. ** Duration of challenge is equal to applicant’s identified duration †† Minimum Service Life is equal to twice the applicant’s identified duration. ‡‡ Respirators will be monitored in the oral/nasal and ocular regions Table 4.—Vapor Challenge of Self-contained Escape Respirator with Sarin (GB)
† The test period begins upon initial generation of vapor concentration. ‡ Three consecutive sequential test data points at or exceeding 0.087 mg/m3 will collectively constitute a failure where each test value is based on a detector sample time of approximately 2 minutes. § The cumulative Ct including all maximum peak excursion data points must not be exceeded for the duration of the test. ** Duration of challenge is equal to applicant’s identified duration †† Minimum Service Life is equal to the applicant’s identified duration. ‡‡ Respirators will be monitored in the oral/nasal and ocular regions $ Exposure will include at least two minutes at a concentration of 2000 mg-m3
4.3 Laboratory Respiratory Protection Level (LRPL) Test Requirement: The measured laboratory respiratory protection level (LRPL) for each air purifying escape respirator shall be 3000 or greater, for 95% of trials, sampled in the breathing zone of the respirator, and shall be 150, or greater, for 95% of trials, sampled outside the breathing zone (under the hood). Each trial must meet the breathing zone criteria and ‘under the hood’ criteria simultaneously for the trial to be considered passing. Test subject and replication numbers are outlined in Table 5.
The respirator is tested in an atmosphere containing 20-40 mg/m3 corn oil aerosol of a mass median aerodynamic diameter of 0.4 to 0.6 micrometers. Should a group of test subjects result in LRPL trials where less than 95% of trials have passing results, one addition run of test subjects that fills the entire anthropometric panel requirements may be performed to increase the total number of trials; the total number of trials will be the sum of trials from the first and second run of subjects. All trials shall be considered in the Practical Performance requirement criteria of paragraph 4.4. The LRPL shall be calculated using nine exercises: Normal Breathing, Deep Breathing, Turn Head Side to Side, Move Head Up and Down, Reach for the Floor and Ceiling, On Hands and Knees - Look Side to Side, Facial Grimace, Climb Stairs at a Regular Pace, and Normal Breathing. For each size category (Small, Medium, and Large), each cell corresponding to the anthropometric parameter will be tested. Cells can be either consecutively (if the test subjects only meet the requirements of a specific cell) or concurrently (if the test subjects meet the requirements of more than one cell) tested for each size category.
4.4 Practical Performance: The Practical Performance of the air-purifying escape respirator shall be evaluated as part of the test procedures of paragraphs 3.4, Breathing Gas, and 4.7, Laboratory Respirator Protection Level. The Practical Performance of the respirator shall evaluate human interface issues associated with the use of the escape respirator. As a minimum, contributing factors (if applicable based upon the respirator design) are: the use of mouth bits and nose clips; seal of the hood around the respirator wearer’s neck; seating of inner masks; position of the hood on the respirator wearer’s head; and strength required to don the respirator. Test subjects shall be trained on proper use of the escape respirator in accordance with the applicant’s instructions identified in paragraph 8.0, Training. Inability of any test subject participating in the test procedures of paragraphs 3.3, Breathing Gas, and 4.3, Laboratory Respirator Protection Level, to complete the test procedures shall constitute a failure of the Practical Performance requirement for that trial. Practical Performance trials shall be accumulated from the test procedures of paragraphs 3.3, Breathing Gas, and 4.3, Laboratory Respirator Protection Level. For the total of these accumulated trials, 95% of these trials shall exhibit acceptable Practical Performance. Should 95% of the Practical Performance test trials not be acceptable, one additional run of test trials consisting of either, or both, paragraph 3.3, Breathing Gas, or paragraph 4.3, Laboratory Respirator Protection Level, may be performed to increase the total number of trials. The total number of trials will be the sum of trials from the first and second run of subjects. 4.5 Donning: The time to don the self-contained escape respirator from the ready-to-use configuration shall be no greater than 30 seconds. The ready to use configuration is the operational packaging state prior to use such that immediately upon opening allows the user to don the respirator. 4.6 Environmental Conditioning Requirements: Environmental, vibration, and drop conditioning shall be performed on the self-contained escape respirators in the ready-to-use configuration. The ready-to-use configuration is the operational packaging state prior to use, such that immediately upon opening allows the user to don the respirator. Respirators will be visually inspected following environmental conditioning to ensure no damage or deterioration has occurred that could negatively affect the intended use of the respirator. Environmental conditioning shall be performed in accordance with Table 6.
Testing of the self-contained escape respirator shall follow Table 7.
* A total six systems tests are performed, 3 GB and 3 HD. Two systems tests, 1 GB and 1 HD, are performed prior to Para. 4.6 Environmental Conditioning. Four systems tests, 2 GB and 2 HD, are performed after Para. 4.6 Environmental Conditioning. † Breathing Gas and LRPL are performed prior to Paragraph 4.6, Environmental Conditioning
|
Start | 1st Service Date | 2nd Service Date | 3rd -- etc. | Stop |
[-----------I------------ |
-------------I---------------- |
---------------I----------------- |
--------I--------- |
--------> |
1st Service expiration date permanently visible on the unit | After a completed action on each unit stamp 2nd service date or terminal date | After a completed action on each unit stamp 3rd service date or terminal date | After a completed action, etc. | Terminal End of service life |
Note: The date on which the unit must be removed from service is to be permanently marked and clearly visible on the unit at the time of manufacture. If an incremental service life is granted, the applicant, or their authorized representative, must stamp the unit with a new date, as described by the time line model. The terminal date represents the final expiration date of the unit with no further extensions.
The applicant shall identify training requirements associated with their air-purifying escape respirator. As a minimum, the applicant shall include an instruction manual, which shall address donning procedures, respirator use, maintenance (care and useful life), and cautions and limitations. The applicant shall also provide for training aid systems, to include a training respirator that mimics the performance of the approved respirator, such as inhalation and exhalation breathing resistance that will develop user proficiency in operation of the equipment, as well as identification of periodic refresher training requirements to maintain user proficiency. The applicants’ training materials shall be used as the basis for preparing the human test subjects in the test procedures of paragraph 3.3, Breathing Gas, paragraph 4.3, Laboratory Respirator Protection Level, and paragraph 4.5, Donning.
NIOSH will authorize the use of an additional approval label on the
self-contained escape respirator that demonstrates compliance to the CBRN
criteria. This label is to be placed in a visible location. The addition
of this label will provide visible and easy identification of equipment
for its appropriate use. In accordance with the requirements of paragraph
84.33 of 42 CFR, Subpart D, approval labels shall be marked with a CBRN
Rating as determined by paragraph 4.1 Duration/Service Life Rating. For
example, respirators tested for 15 minutes are marked ESCAPE ONLY NIOSH
CBRN 15.