U.S. Food and Drug Administration - CDRH Mobile
Skip NavigationFDA Logo links to FDA home pageCenter for Devices and Radiological Health, U.S. Food and Drug AdministrationHHS Logo links to Department of Health and Human Services website
FDA Home Page | CDRH Home Page | Search | A-Z Index U.S. Food and Drug AdministrationCenter for Devices and Radiological Health Questions?
horizonal rule

Public Health Notification from FDA: Vail Products Enclosed Bed Systems

Updated: December 4, 2007 & June 24, 2005

Original Date: March 25, 2005

On June 24, 2005, we updated a Preliminary Public Health Notification warning users of Vail Enclosed Bed Systems that they pose a health risk because patients can become entrapped in them and suffocate. This is an update of that notification, providing final recommendations for users as well as access to the revised user manuals that are no longer available from the company.

Product status and user manuals

Vail Products, Inc. of Toledo, Ohio, publicly stated on June 16, 2005, that it permanently ceased the manufacture, sale and distribution of all Vail enclosed bed systems. The company is no longer available to provide accessories, replacement parts, or retrofit kits. On June 23 and 24, 2005, revised instruction manuals and warning labels were mailed to customers with Vail 500, Vail 1000 or Vail 2000 enclosed bed systems. The revised manuals include new warnings, precautions, and instructions for use. These revised manuals are available online at http://www.fda.gov/cdrh/safety/vail-instructions500.pdf, http://www.fda.gov/cdrh/safety/vail-instructions1000.pdfand http://www.fda.gov/cdrh/safety/vail-instructions2000.pdf.

Recommendation for Users

Because of the suffocation risk, FDA advises hospitals, nursing homes and consumers who have a Vail enclosed bed system to stop using it and move the patient to an alternate bed. Consumers who are using Vail beds at home can consult with their physicians about other options.

If continued use of the Vail bed is the only option, the following safety precautions recommended by Vail should be followed:

Additional recommendations for the Vail 1000 and 2000 models:

Background Information

Vail enclosed bed systems are canopy-like padded beds covered with nylon netting that is zipped into place. They are used for at risk patients, both adults and children, with cognitive impairment, unpredictable behavior, spasms, seizures, and other disorders. The beds are used as an alternative to a physical or drug restraint to reduce falls from a bed and prevent patients from wandering.

On March 22, 2005, in response to ongoing concerns about manufacturing quality and labeling, the FDA and the U.S. Department of Justice initiated a seizure of all Vail model 500, 1000, and 2000 enclosed bed systems present at Vail’s facility at that time. On June 24, 2005, Vail Products, Inc. announced a voluntary Class I recall of all Vail model 500, 1000, and 2000 enclosed bed systems.

Approximately 5,000 of these beds were distributed nationwide. FDA is aware of approximately 30 adverse event reports, including at least 8 deaths, resulting from entrapments, falls, and other incidents. More than half of the 30 incidents involved children aged 16 and under.

The Vail 500, 1000, and 2000 beds can be identified by a Vail label containing a model number. The label is on the front of the bed or on the leg of the bed.

Reporting Adverse Events to FDA

FDA requires hospitals and other user facilities to report deaths and serious injuries associated with the use of medical devices. If you suspect that a reportable adverse event was related to the use of a Vail enclosed bed system, you should follow the reporting procedure established by your facility.

We also encourage you to report adverse events related to Vail enclosed bed systems that do not meet the requirements for mandatory reporting. You can report these directly to MedWatch, the FDA’s voluntary reporting program.

You may submit reports to MedWatch by phone at 1-800-FDA-1088; by FAX at 1-800-FDA-0178; by mail to MedWatch, Food and Drug Administration, 5600 Fishers Lane, Rockville, MD 20857-9787; or online at http://www.fda.gov/medwatch/report.htm. Consumers can also report directly to MedWatch.

Getting More Information

For more information on the recall, see the recall notice at: http://www.fda.gov/cdrh/recalls/recall-062405.html. For further details on the seizure, see FDA’s 3/22/05 Talk Paper at: www.fda.gov/bbs/topics/ANSWERS/2005/ANS01347.html.

If you have questions for FDA, please contact the Office of Surveillance and Biometrics (HFZ-510), 1350 Piccard Drive, Rockville, Maryland, 20850, Fax at 240-276-3356, or by e-mail at phann@cdrh.fda.gov. You may also leave a voice mail message at 240-276-3357 and we will return your call as soon as possible.

FDA medical device Public Health Notifications are available on the Internet at http://www.fda.gov/cdrh/safety.html. You can also be notified through email on the day the safety notification is released by subscribing to our list server. To subscribe, visit: http://service.govdelivery.com/service/subscribe.html?code=USFDACDRH_10.

Sincerely yours,


Daniel G. Schultz, MD
Director
Center for Devices and Radiological Health
Food and Drug Administration

Updated December 4, 2007

horizonal rule

CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page

Center for Devices and Radiological Health / CDRH