Home
Search
Study Topics
Glossary
|
|
|
|
|
|
Sponsored by: |
Nonin Medical, Inc |
---|---|
Information provided by: | Nonin Medical, Inc |
ClinicalTrials.gov Identifier: | NCT00788983 |
Validate Non-Motion SpO2 Accuracy from 70%-100% of various pulse oximeter/ sensor combinations and various OEM pulse oximeter modules and sensor combinations during induced hypoxia (desaturation).
Condition |
---|
Hypoxia |
Study Type: | Observational |
Study Design: | Prospective |
Official Title: | Pulse Oximeter Responses to Multiple Levels of Stable Hypoxia |
Enrollment: | 12 |
Study Start Date: | November 2008 |
Study Completion Date: | November 2008 |
Primary Completion Date: | November 2008 (Final data collection date for primary outcome measure) |
The testing is conducted on 12 healthy, consenting, non-smoking subjects in accordance with the IRB approved Protocol. The subjects shall be distributed across both genders and a range of skin tones as equally as practical.
An arterial line will be placed in the radial artery of each subject's right arm, and sensors will be attached to each subject. The subjects will be placed in a semi-supine position and allowed to breathe through a mouthpiece while the nose is blocked with a nose-clip.
Hypoxia will be induced by on each subject, as levels of oxyhemoglobin saturation (between 70% and 100%) are achieved by breathing mixtures of nitrogen, room air and carbon dioxide. Inspired O2 concentration will be adjusted breath-by-breath using a computed saturation, based on end-tidal PO2 and PCO2, as sampled by a mass spectrometer. Predicted levels of oxyhemoglobin saturations will be attained and held stable. At a minimum of sixty seconds into each plateau, a 0.5cc waste blood draw through the arterial line, followed by the first 1.0cc sample blood draw. After approximately thirty additional seconds, the second 1.0cc sample blood draw will be taken. The samples will immediately be analyzed by a Radiometer OSM-3 multi-wavelength co-oximeter and recorded. The SpO2 data from the oximeters will be collected via a laptop computer. Concurrent with the end of each blood draw, a marker will be generated on the laptop computers to identify the event.
Ages Eligible for Study: | 18 Years to 40 Years |
Genders Eligible for Study: | Both |
Accepts Healthy Volunteers: | Yes |
Sampling Method: | Non-Probability Sample |
Healthy, consenting, non-smoking subjects, 18 years or greater, both genders and a range of skin tones
Inclusion Criteria:
Exclusion Criteria:
United States, California | |
University of California, San Francisco, Induced Hypoxia Lab | |
San Francisco, California, United States, 94143 |
Principal Investigator: | Philip E Bickler, MD | University of California, San Francisco |
Responsible Party: | Nonin Medical, Inc ( L Roth, Clinical/Regulatory Specialist ) |
Study ID Numbers: | 112008 |
Study First Received: | November 7, 2008 |
Last Updated: | December 2, 2008 |
ClinicalTrials.gov Identifier: | NCT00788983 |
Health Authority: | United States: Institutional Review Board |
Induced hypoxia Pulse oximeter Nonin Medical Verification of SpO2 accuracy |
Signs and Symptoms Signs and Symptoms, Respiratory Anoxia |