Table 4. References for Applicable Containment Tests
Test CABINET TYPE BY CLASS I II III HEPA Filter Leak (F,IID)(1) (F, IID) (F, IID) Airflow Smoke Pattern No smoke shall reflux out of BSC once drawn in. (F, IIC) N/A Cabinet Integrity N/A (F,IIE) N/A Face Velocity Open Front 75-125 lfpm (F,IIB,3b) 75 lfpm - type A; 100 lfpm - type B (F,IIB) N/A Face Velocity Glove Ports/No Gloves 150 lfpm (F,IIB,3b(1)(C)) N/A N/A Water Gauge Pressure Glove Ports & Gloves N/A N/A -0.5" [p. 145](2) Velocity Profile N/A (F,IIA) N/A 1. Parenthetical references are to the NSF 49; letters and numerals indicate specific sections and subsections.
2. Bracketed references ([]) is to the Laboratory Safety Monograph; page numbers are indicated.