Table 4. References for Applicable Containment Tests

Test CABINET TYPE BY CLASS
I II III
HEPA Filter Leak (F,IID)(1) (F, IID) (F, IID)
Airflow Smoke Pattern No smoke shall reflux out of BSC once drawn in. (F, IIC) N/A
Cabinet Integrity N/A (F,IIE) N/A
Face Velocity Open Front 75-125 lfpm (F,IIB,3b) 75 lfpm - type A; 100 lfpm - type B (F,IIB) N/A
Face Velocity Glove Ports/No Gloves 150 lfpm (F,IIB,3b(1)(C)) N/A N/A
Water Gauge Pressure Glove Ports & Gloves N/A N/A -0.5" [p. 145](2)
Velocity Profile N/A (F,IIA) N/A

1. Parenthetical references are to the NSF 49; letters and numerals indicate specific sections and subsections.

2. Bracketed references ([]) is to the Laboratory Safety Monograph; page numbers are indicated.