PROPYLENE GLYCOL MONOMETHYL ETHERS/ACETATES
1-METHOXY-2-PROPANOL (1M2P)
2-METHOXY-1-PROPANOL (2M1P)
1-METHOXY-2-PROPYL ACETATE (1M2PA)
2-METHOXY-1-PROPYL ACETATE (2M1PA)
Method number: |
99 |
|
|
|
|
Matrix: | Air |
|
Procedure: |
Samples are collected by drawing air through standard size (6-mm o.d.,
100/50-mg sections) coconut shell charcoal tubes. Samples are desorbed with
95/5 (v/v) methylene chloride/methanol and analyzed by gas chromatography using a flame
ionization detector. Samples should be desorbed in the presence of a drying agent such as
magnesium sulfate or Drierite®.
| |
Recommended air volume and sampling rate: | 10 L at 0.1 L/min |
|
|
|
| 1M2P | 2M1P | 1M2PA | 2M1PA |
|
|
Target concentration: |
100 ppm (368 mg/m3) |
1 ppm (3.7 mg/m3) |
100 ppm (540 mg/m3) |
1 ppm (5.4 mg/m3) |
|
Reliable quantitation limit: |
20 ppb (74 µg/m3) |
20 ppb (74 µg/m3) |
20 ppb (108 µg/m3) |
20 ppb (108 µg/m3) |
|
Standard error of estimate at the target concentration: |
5.3% |
5.5% |
5.1% |
5.4% |
|
|
|
Special requirements: |
Samples for 1M2PA and 2M1PA should be refrigerated upon receipt by the laboratory to minimize hydrolysis. |
|
Status of method: |
Evaluated method. This method has been subjected to the established evaluation procedures of the Organic Methods
Evaluation Branch. |
| Date: April 1993 |
Chemist: Carl J. Elskamp |
Organic Methods Evaluation Branch
OSHA Salt Lake Technical Center
Salt Lake City, UT 84165-0200
1. General Discussion
1.1. Background
1.1.1. History
Over the past several years there has been an increase in the number of samples
submitted to the OSHA Salt Lake Technical Center for propylene glycol ethers and
their acetates. This is likely due to the increased usage of these solvents as
substitutes for ethylene glycol ethers and their acetates, which have been associated
with the potential to cause adverse reproductive effects in both male and female
workers. (Ref. 5.1.) Now there is evidence that 2M1P and 2M1PA may exhibit
analogous toxicities. (Ref. 5.2.) 2M1P and 2M1PA are present as impurities in
technical grade 1M2P and 1M2PA respectively. OSHA has adopted a TWA-PEL of
100 ppm for 1M2P, but has no PEL for the other three analytes. (Ref. 5.3.) Because
1M2P and 1M2PA have similar toxic properties, a target concentration of 100 ppm
was chosen for these two analytes. Based on their potential reproductive toxicity,
a target concentration of 1 ppm was chosen for 2M1P and 2M1PA.
There were no reported air monitoring procedures in the literature for these propylene
glycol ethers/acetates. A method has now been evaluated based on previous
evaluations done at the OSHA Salt Lake Technical Center for a number of ethylene
glycol ether/acetates, which are chemically similar to these compounds.
(Refs. 5.4.-5.5.)
Sampling is done by drawing workplace air through sampling tubes containing
coconut shell charcoal. The analytes are adsorbed by the charcoal and the tubes are
sent to the laboratory for analysis. The analytes are desorbed from the charcoal with
a 95/5 (v/v) methylene chloride/methanol mixture, and are quantitated by GC using
flame ionization detection.
As was found for the acetates of ethylene glycol ethers (Refs. 5.4.-5.5.),
the acetates of propylene glycol ethers are also prone to hydrolysis after being adsorbed
onto charcoal. The average loss due to hydrolysis was about 3% for 1M2PA (100
ppm for 10 L) and 11% for 2M1PA (1 ppm for 10 L) when generated samples were
stored for 15 days at ambient temperatures. When generated samples were
refrigerated for 15 days, the average loss was about 1% for 1M2PA and 4% for
2M1PA. The slow rate of loss due to hydrolysis indicates it is unnecessary to ship
samples under reduced temperatures, but they should be refrigerated at the lab until analyzed.
As was the case for some of the other previously evaluated ethylene glycol ethers
(Ref. 5.4.), the use of a drying agent such as magnesium sulfate is needed to
improve the desorption efficiency of 1M2P and 2M1P from charcoal. Because
MgSO4 can only be purchased as a powder, periodically the syringe used for sample
injections into the GC may be plugged by MgSO4 suspended in solution. This can
be avoided by centrifuging the samples or by allowing the powder to settle out
before analysis. Two granular drying agents, 20-40 mesh Drierite®
(anhydrous calcium sulfate) and 10-60 mesh sodium sulfate, were tested as alternatives.
Sodium sulfate was found to be ineffective. Drierite® was effective, but 400 mg
must be used versus 125 mg of MgSO4. If Drierite® is used, samples need not be
centrifuged and can be analyzed immediately after the desorption period.
Many solvent vapors collected on charcoal and analyzed at the OSHA Salt Lake
Technical Center are desorbed with 99/1 (v/v) carbon
disulfide/N,N-dimethyl-formamide (CS2/DMF)
instead of methylene chloride/methanol. Because it is possible
that these propylene glycol ether/acetates may be used with other solvents in the
workplace, CS2/DMF was investigated as an alternative desorbing solvent. For the
target concentrations studied, CS2/DMF is an acceptable alternative for 1M2PA and
2M1PA samples if MgSO4 is added to the desorption vials. Desorption with
CS2/DMF is not acceptable for 1M2P and 2M1P samples because there is
considerable differences in desorption efficiencies from wet and dry charcoal, even
if MgSO4 is used. Also, the desorption efficiency for 2M1P is less than 75% from
wet charcoal.
It is felt that there will always be some amount of corresponding alcohol present in
samples containing acetates because the acetates may be partially hydrolyzed in the
air before collection or on the charcoal after collection. For example, 1M2P would
be found in samples containing 1M2PA, and 2M1P would be found in 2M1PA
samples. Also, the alcohols could be present as contaminants in the corresponding
technical grade acetates. For these reasons, it would be wise to analyze for the
alcohols in acetate samples. Thus, unless it is absolutely necessary to desorb with
CS2/DMF for analysis of other co-collected solvents, samples for 1M2PA and 2M1PA
should also be desorbed with methylene chloride/methanol in the presence of MgSO4
or Drierite® to facilitate the analysis of the corresponding alcohols.
1.1.2. Toxic effects (This section is for information only and should not be taken as the
basis of OSHA policy.)
These propylene glycol ethers/acetates have low acute toxicity, with the main effect
at high doses being central nervous system depression. There are no reported
studies on the carcinogenicity of these compounds. 1M2P and 1M2PA appear to
lack reproductive toxicity, with the critical effect being irritation of the eyes and
mucous membranes. 2M1PA is embryotoxic and teratogenic in laboratory animals.
It is very likely that 2M1P is also teratogenic because the common metabolite of
2M1P and 2M1PA is 2-methoxypropionic acid. 3-Methoxypropionic acid, which is
a structural isomer of 2-methoxypropionic acid, causes growth retardation and
abnormalities in post-implantation rat embryo cultures. This is analogous to the
effects of methoxyacetic acid, which is a common metabolite of 2-methoxyethanol
and 2-methoxyethyl acetate. (Ref. 5.2.)
1.1.3. Workplace exposure
Propylene glycol ethers/acetates are used industrially as solvents for paints, lacquers,
resins, oils and fats. Their use has increased considerably since 1985. This is
probably because they are used as substitutes for the chemically similar ethylene
glycol ethers, which have been associated with reproductive toxicity. From NIOSH
occupational surveys done from 1981-1983, it is estimated that 329,000 people are
potentially exposed to 1M2P and about 306,000 are potentially exposed to 1M2PA.
2M1P and 2M1PA, which are also known as the beta isomers, are typically present
as impurities in varying amounts in 1M2P and 1M2PA respectively. (Ref. 5.2.)
1.1.4. Physical properties (Ref. 5.2. unless otherwise noted)
|
Property |
|
1M2P |
2M1P |
1M2PA |
2M1PA |
|
CAS number:
molecular weight:
melting point (°C):
boiling point (°C):
flash point (°C):
vapor pressure (kPa):
vapor density (25°C, air=1):
liquid density (25°C/4°C):
description:
miscibility with water: |
|
107-98-2 90.12 119.6 38 1.6 @ 25°C 3.11 0.917 colorless liquid complete |
1589-47-5 90.12 130 (Ref. 5.6.) 0.938 (Ref. 5.6.) colorless liquid complete |
108-65-6 132.16 <-67 145.8 42.2 0.5 @ 20°C 4.55 0.97 colorless liquid
19% (w/w) |
70657-70-4 132.16 colorless liquid |
|
Note: Commercial grade 1M2P contains mainly (95-99%) 1M2P, with the remainder
(1-5%) being 2M1P. Technical 1M2PA consists mainly of
(95%) 1M2PA, with
remainder largely being 2M1PA. (Ref. 5.2.)
structural formulae:
1-methoxy-2-propanol (1M2P):
2-methoxy-1-propanol (2M1P):
1-methoxy-2-propyl acetate (1M2PA):
2-methoxy-1-propyl acetate (2M1PA):
synonyms:
1-methoxy-2-propanol- 1M2P; methoxypropanol, alpha isomer; propylene glycol
monomethyl ether; propylene glycol methyl ether; PGME
2-methoxy-1-propanol- 2M1P; methoxypropanol, beta isomer; propylene glycol
monomethyl ether; propylene glycol methyl ether; bPGME
1-methoxy-2-propyl acetate- 1M2PA; methoxypropyl acetate, alpha isomer;
propylene glycol monomethyl ether acetate; propylene glycol methyl ether acetate; PGMEA
2-methoxy-1-propyl acetate- 2M1PA; methoxypropyl acetate, beta isomer;
propylene glycol monomethyl ether acetate; propylene glycol methyl ether acetate;
bPGMEA
The analyte air concentrations throughout this method are based on the recommended sampling and
analytical parameters. Air concentrations listed in ppm and ppb are referenced to 25°C and 101.3 Kpa (760 mmHg.)
1.2. Limit defining parameters
1.2.1. Detection limit of the analytical procedure
The detection limits of the analytical procedure are 48, 50, 71, and 71 pg per
injection (1.0-µL injection with a 15:1 split) for 1M2P, 2M1P, 1M2PA, and 2M1PA
respectively. These are the amounts of each analyte that will give peaks with
heights approximately 5 times the height of baseline noise. (Section 4.1.)
1.2.2. Detection limit of the overall procedure
The detection limits of the overall procedure are 0.73, 0.75, 1.1, and 1.1 µg per
sample for 1M2P, 2M1P, 1M2PA, and 2M1PA respectively. These are the amounts
of each analyte spiked on the sampling device that, upon analysis, produce a peak
similar in size to that of the respective detection limit of the analytical procedure.
These detection limits correspond to air concentrations of 20 ppb (74 µg/m³), 20 ppb
(74 µg/m³), 20 ppb (108 µg/m³), and 20 ppb (108 µg/m³) for 1M2P, 2M1P, 1M2PA,
and 2M1PA respectively. (Section 4.2.)
1.2.3. Reliable quantitation limit
The reliable quantitation limits are 0.73, 0.75, 1.1, and 1.1 µg per sample for 1M2P,
2M1P, 1M2PA, and 2M1PA respectively. These are the smallest amounts of each
analyte that can be quantitated within the requirements of recoveries of at least
75% and precisions (±1.96 SD) of ±25% or better. These reliable quantitation
limits correspond to air concentrations of 20 ppb (74 µg/m³), 20 ppb (74 µg/m³), 20
ppb (108 µg/m³), and 20 ppb (108 µg/m³) for 1M2P, 2M1P, 1M2PA, and 2M1PA
respectively. (Section 4.3.)
The reliable quantitation limits and detection limits reported in the method are based upon
optimization of the GC for the smallest possible amounts of each analyte. When the target
concentration of an analyte is exceptionally higher than these limits, they may not be
attainable at the routine operating parameters.
1.2.4. Instrument response to the analyte
The instrument response over the concentration ranges of 0.5 to 2 times the target
concentrations is linear for all four analytes. (Section 4.4.)
1.2.5. Recovery
The recovery of 1M2P, 2M1P, 1M2PA, and 2M1PA from samples used in a 15-day
storage test remained above 100%, 94%, 97%, and 90% respectively when the
samples were stored at ambient temperatures. (Section 4.5., from regression lines
shown in Figures 4.5.1.2., 4.5.2.2., 4.5.3.2. and 4.5.4.2.)
1.2.6. Precision (analytical procedure)
The pooled coefficients of variation obtained from replicate determinations of
analytical standards at 0.5, 1, and 2 times the target concentrations are 0.0025,
0.0045, 0.0025, and 0.0041 for 1M2P, 2M1P, 1M2PA, and 2M1PA respectively.
(Section 4.6.)
1.2.7. Precision (overall procedure)
The precisions at the 95% confidence level for the ambient temperature 15-day
storage tests are ±10.3, ±10.8, ±10.0, and ±10.5% for 1M2P, 2M1P, 1M2PA,
and 2M1PA respectively. These include an additional ±5% for sampling error.
(Section 4.7.)
1.2.8. Reproducibility
Six samples for each analyte collected from controlled test atmospheres and a draft
copy of this procedure were given to a chemist unassociated with this evaluation.
The samples were analyzed after nine days of refrigerated storage. No individual
sample result deviated from its theoretical value by more than the precision reported
in Section 1.2.7. (Section 4.8.)
2. Sampling Procedure
2.1. Apparatus
2.1.1. Samples are collected using a personal sampling pump calibrated to within ±5% of
the recommended flow rate with a sampling tube in line.
2.1.2. Samples are collected with solid sorbent sampling tubes containing coconut shell
charcoal. Each tube consists of two sections of charcoal separated by a urethane
foam plug. The front section contains 100 mg of charcoal and the back section, 50
mg. The sections are held in place with glass wool plugs in a glass tube 70-mm ×
4-mm i.d. × 6-mm o.d. For this evaluation, SKC Inc. (Eighty Four, PA) charcoal
tubes (catalog number 226-01, Lot 120) were used.
2.2. Reagents
None required
2.3. Technique
2.3.1. Immediately before sampling, break off the ends of the charcoal tube. All tubes
should be from the same lot.
2.3.2. Connect the sampling tube to the sampling pump with flexible, non-crimpable tubing.
It is desirable to utilize a sampling tube holder that shields the employee from the
sharp, jagged end of the sampling tube. Position the tube so that sampled air first
passes through the 100-mg section.
2.3.3. Air being sampled should not pass through any hose or tubing before entering the
sampling tube.
2.3.4. To avoid channeling, place the sampling tube vertically in the employee's breathing
zone.
2.3.5. After sampling, seal the tubes immediately with plastic caps and wrap lengthwise
with OSHA Form 21.
2.3.6. Submit at least one blank sampling tube with each sample set. Blanks should be
handled in the same manner as samples, except no air is drawn through them.
2.3.7. Record sample volumes (in liters of air) for each sample.
2.3.8. List any compounds that could be considered potential interferences, especially
solvents, that are being used in the sampling area.
2.3.9. Ship any bulk sample(s) in a container separate from the air samples.
2.4. Sampler capacity
2.4.1. Sampler capacity is determined by measuring how much air can be sampled before
breakthrough of analyte through the sampler occurs, i.e., the sampler capacity is
exceeded. Breakthrough is considered to occur when the effluent from the sampler
contains a concentration of analyte that is 5% of the upstream concentration (5%
breakthrough). Testing for 1M2P breakthrough was performed by monitoring the
effluent (with a total hydrocarbon analyzer) from sampling tubes containing only the
100-mg section of charcoal while sampling at 0.2 L/min from an atmosphere
containing 204 ppm of analyte. The atmosphere was at approximately 80% relative
humidity and 20-25°C. The average 5% breakthrough volume from three
determinations was 33.8 L (RSD=2.1%). A similar test was done for 1M2PA at
199 ppm. The average 5% breakthrough volume from three determinations was
24.9 L (RSD=6.8%).
2.4.2. Similar studies as in 2.4.1. were done for 1M2P at 2.0 ppm and 1M2PA at 2.2 ppm.
Both atmospheres were sampled for more than 6 h (>72 L) with no breakthrough
detected. 1M2P and 1M2PA were used in these studies instead of 2M1P and
2M1PA because sufficient quantities of the latter analytes were not available. It is
felt that the breakthrough volume for 2M1P would approximate that of 1M2P and
the breakthrough volume of 2M1PA would approximate that of 1M2PA.
2.5. Desorption efficiency
2.5.1. Desorption with 95/5 (v/v) methylene chloride/methanol
2.5.1.1. The average desorption efficiencies of 1M2P, 2M1P, 1M2PA, and 2M1PA
from dry Lot 120 charcoal are 100.4%, 99.7%, 101.8%, and 101.4%
respectively over the range of 0.5 to 2 times the target concentrations.
(Section 4.9.)
2.5.1.2. Desorbed samples from Section 2.5.1. remain stable for at least 24 h.
(Section 4.10.)
2.5.1.3. The desorption efficiencies at the target concentrations from wet charcoal
are essentially the same as from dry charcoal when MgSO4 is used. The
desorbed samples are stable for at least 24 h. (Section 4.11.) The use of
MgSO4 is recommended for 1M2P and 2M1P samples, but is optional for
1M2PA and 2M1PA samples.
2.5.2. Desorption with 99/1 (v/v) carbon disulfide/N,N-dimethylformamide (CS2/DMF)
2.5.2.1. The average desorption efficiencies of 1M2P, 2M1P, 1M2PA, and 2M1PA
from dry Lot 120 charcoal are 86.3%, 80.8%, 98.4%, and 98.0%
respectively over the range of 0.5 to 2 times the target concentrations.
(Section 4.12.)
2.5.2.2. All of the analytes with the exception of 2M1P were sufficiently stable for
the target concentration samples from Section 2.5.2.1. The average
desorption efficiency for 2M1P dropped from 82.2% to 75.8% in 24 h.
(Section 4.13.)
2.5.2.3. The average desorption efficiencies at the target concentrations from wet
charcoal are 79.5% for 1M2P and 70.5% for 2M1P when MgSO4 is used.
These compare to 87.4% and 82.2% respectively from dry charcoal. The
desorbed 1M2P samples are stable for at least 24 h, while the average
desorption efficiency dropped to 64.3% for the 2M1P samples. The
optional use of CS2/DMF (with MgSO4) would be acceptable for 1M2P
samples but not for 2M1P samples at or around the studied loadings.
(Section 4.14.)
The average desorption efficiencies at the target concentrations for
1M2PA and 2M1PA from wet charcoal using 99/1 CS2/DMF as the
desorbing solvent (with MgSO4) are essentially the same as from dry
charcoal. The desorbed samples are stable for at least 24 h.
(Section 4.14.)
2.6. Recommended air volume and sampling rate
2.6.1. For TWA samples, the recommended air volume is 10 L collected at 0.1 L/min
(100-min samples). The recommended air volume is significantly lower than the
breakthrough volumes, but 10 L was chosen to provide a reasonable safety margin
in case other solvent vapors are present in the sampled air. Also, 10 L is commonly
recommended for solvent vapors collected on charcoal tubes.
2.6.2. For short-term samples, the recommended air volume is 3 L collected at 0.20 L/min
(15-min samples).
2.6.3. When short-term samples are required, the reliable quantitation limits become larger.
For example, the quantitation limits are 67 ppb when 3 L is sampled.
2.7. Interferences (sampling)
2.7.1. It is not known if any compound(s) will severely interfere with the collection of any
of the four analytes on charcoal. In general, the presence of other contaminant
vapors in the air will reduce the capacity of charcoal to collect the analytes.
2.7.2. Suspected interferences should be reported to the laboratory with submitted samples.
2.8. Safety precautions (sampling)
2.8.1. Attach the sampling equipment to the employee so that it will not interfere with
work performance or safety.
2.8.2. Wear eye protection when breaking the ends of the charcoal tubes.
2.8.3. Follow all safety procedures that apply to the work area being sampled.
3. Analytical Procedure
3.1. Apparatus
3.1.1. A GC equipped with a flame ionization detector. For this evaluation, a
Hewlett-Packard 5890A Gas Chromatograph equipped with a 7673A Automatic Sampler was used.
3.1.2. A GC column capable of separating the analyte of interest from the desorbing
solvent, internal standard and any interferences. A 30-m × 0.32-mm i.d. fused
silica Stabilwax-DA® column with a 1-µm df (Cat. no. 11054, Restek Corp.,
Bellefonte, PA) was used in this evaluation.
3.1.3. An electronic integrator or some other suitable means of measuring peak areas or
heights. A Waters 860 Networking Computer System was used in this evaluation.
3.1.4. Two-milliliter vials with Teflon®-lined caps.
3.1.5. A dispenser capable of delivering 1.0 mL of desorbing solvent to prepare standards
and samples. If a dispenser is not available, a 1.0-mL volumetric pipet may be used.
3.2. Reagents
3.2.1. 1-Methoxy-2-propanol, 2-methoxy-1-propanol, 1-methoxy-2-propyl acetate, and
2-methoxy-1-propyl acetate, reagent grade or of known purity. Lot HB062777
1M2P from Aldrich Chemical Company (Milwaukee, WI), 2M1P from Dow Chemical
(Midland, MI), and mixtures of 1M2PA and 2M1PA from Aldrich (Lot 03019HY) and
Frinton Laboratories (Vineland, NJ) were used in this evaluation. The Aldrich 1M2P
contains some 2M1P as a contaminant and the Dow Chemical 2M1P contains some
1M2P. A 98.97/1.03 (w/w) 1M2P/2M1P mixture was prepared by mixing the
appropriate proportions of these two solvents. The mixture was assayed by GC.
The density of the mixture was determined to be 0.917 g/mL (RSD=0.1%, n=4)
at 24°C by weighing 10.00 mL of the mixture contained in tared volumetric flasks.
A 99.01/0.99 (w/w) 1M2PA/2M1PA mixture was similarly prepared from the Aldrich
and Frinton mixtures. The density of this mixture was determined to be 0.961 g/mL
(RSD=0.17%, n=4) at 24°C.
3.2.2. Anhydrous magnesium sulfate or 20-40 mesh Drierite® (anhydrous calcium sulfate),
reagent grade. Chempure Lot M172 KDHM magnesium sulfate was used in this evaluation.
3.2.3. Methylene chloride, chromatographic grade. Burdick and Jackson Lot BB551 was
used in this evaluation.
3.2.4. Methanol, chromatographic grade. Fisher Lot 913607 was used in this evaluation.
3.2.5. A suitable internal standard, reagent grade. "Quant Grade" 2-heptanol from
Polyscience Corporation (Niles, IL) was used in this evaluation.
3.2.6. The desorbing solvent consists of 95/5 (v/v) methylene chloride/methanol containing
an internal standard at a concentration of 1 µL/mL.
3.2.7. GC grade nitrogen, air, and hydrogen.
3.3. Standard preparation
3.3.1. Prepare standards by injecting microliter amounts of analytes into vials containing
1.0 mL of desorbing solvent delivered from the same dispenser used to desorb
samples. For example, to prepare a standard of 1M2P and 2M1P, inject 4.00 µL of
a 98.97/1.03 (w/w) 1M2P/2M1P mixture (density = 0.917) into a vial containing
1.0 mL of desorbing solvent. This standard contains 3630 µg of 1M2P and
37.78 µg of 2M1P per sample.
3.3.2. Bracket sample concentrations with working standard concentrations. If samples fall
outside of the concentration range of prepared standards, prepare and analyze
additional standards to ascertain the linearity of response.
3.4. Sample preparation
3.4.1. Transfer each section of the samples to separate vials. Discard the glass tubes and plugs.
3.4.2. For 1M2P and 2M1P samples and for 1M2PA and 2M1PA samples to be analyzed
for 1M2P and 2M1P respectively, add about 125 mg of anhydrous magnesium
sulfate or 400 mg of 20-40 mesh Drierite® to each vial.
3.4.3. Add 1.0 mL of desorbing solvent to each vial using the same dispenser as used for
preparation of standards.
3.4.4. Immediately cap the vials and shake them periodically for about 30 min.
3.4.5. If magnesium sulfate is used as the drying agent, centrifuge the vials or allow time
for the powder to settle out to avoid plugging the syringe used for GC injections.
3.5. Analysis
3.5.1. GC conditions
zone temperatures: |
column- injector- detector- |
95°C 175°C 200°C |
gas flows: |
hydrogen (carrier)- nitrogen (makeup)- hydrogen (flame)- air- |
3.0 mL/min (60 kPa head pressure) 37 mL/min 33 mL/min 390 mL/min |
signal range: | 0 |
injection volume: | 1.0 µL (with a 15:1 split) |
column: |
30-m × 0.32-mm i.d. fused silica, Stabilwax-DA®, 1-µm df |
retention times: |
1M2P- 2M1P- 1M2PA- 2M1PA- 2-heptanol- |
3.1 min 4.4 min 4.7 min 5.4 min 7.2 min (internal standard) |
chromatograms at the target concentrations: |
Chromatogram of the analytes at the target concentrations with 1M2P and 1M2PA at approximately full scale.
Key: (1) 1M2P, (2) 2M1P, (3) 1M2PA, (4) 2M1PA, (5) 2-heptanol.
|
Chromatogram of the analytes at the target concentrations with 2M1P and 2M1PA at approximately full scale.
Key: (1) 1M2P, (2) 2M1P, (3) 1M2PA, (4) 2M1PA, (5) 2-heptanol.
|
3.5.2.Peak areas (or heights) are measured by an integrator or other suitable means.
3.5.3. An internal standard (ISTD) calibration method is used. Calibration curves are
prepared by plotting micrograms of analyte per sample versus ISTD-corrected
response of standard injections. Sample concentrations must be bracketed by
standards.
3.6. Interferences (analytical)
3.6.1. Any compound that responds on a flame ionization detector and has the same
general retention time of the analyte or internal standard is a potential interference.
Possible interferences should be reported to the laboratory with submitted samples
by the industrial hygienist. These interferences should be considered before samples
are desorbed.
3.6.2. GC parameters (i.e. column and column temperature) may be changed to possibly
circumvent interferences.
3.6.3. Retention time on a single column is not considered proof of chemical identity.
Analyte identity should be confirmed by GC/mass spectrometry.
3.7. Calculations
The analyte concentration for samples is obtained from the appropriate calibration curve in
terms of micrograms of analyte per sample, uncorrected for desorption efficiency. The air
concentration is calculated using the following formulae. The back (50-mg) section is
analyzed primarily to determine if there was any breakthrough from the front (100-mg)
section during sampling. If a significant amount of analyte is found on the back section (e.g.,
greater than 25% of the amount found on the front section), this fact should be reported
with sample results. If any analyte is found on the back section, it is added to the amount
found on the front section. This total amount is then corrected by subtracting the total
amount (if any) found on the blank.
mg/m³ =
|
(µg of analyte per sample) (L of air sampled)(desorption efficiency)
|
where desorption efficiencies = |
1.00 for 1M2P, 1.00 for 2M1P, 1.02 for 1M2PA, 1.01 for 2M1PA
|
ppm = (mg/m³)(24.46) / (molecular weight of analyte)
where 24.46 is the molar volume at 25°C and 101.3 kPa (760 mmHg) and molecular weights =
90.12 for 1M2P and 2M1P,
132.16 for 1M2PA and 2M1PA
3.8. Safety precautions (analytical)
3.8.1. Avoid skin contact and inhalation of all chemicals.
3.8.2. Restrict the use of all chemicals to a fume hood when possible.
3.8.3. Wear safety glasses and a lab coat at all times while in the lab area.
4. Backup Data
4.1. Detection limit of the analytical procedure
The detection limits of 48, 50, 71, and 71 pg per injection were determined by making
injections (1.0 µL with a 15:1 split) of 726, 754, 1066, and 1066 pg/µL standards for 1M2P,
2M1P, 1M2PA, and 2M1PA respectively. These amounts were judged to produce peaks with
heights approximately 5 times the baseline noise.
4.2. Detection limit of the overall procedure
The detection limits of the overall procedure of 0.73, 0.75, 1.1, and 1.1 µg per sample were
determined by analyzing six samples for each analyte that had been spiked with 0.726,
0.754, 1.066, and 1.066 µg of 1M2P, 2M1P, 1M2PA, and 2M1PA respectively. The
detection limits of the overall procedure correspond to air concentrations of 20 ppb (74
µg/m³), 20 ppb (74 µg/m³), 20 ppb (108 µg/m³), and 20 ppb (108 µg/m³) for 1M2P, 2M1P,
1M2PA, and 2M1PA respectively.
Table 4.2.1. Detection Limit of the Overall Procedure for 1M2P
|
sample no. |
µg spiked |
µg recovered |
|
1 2 3 4 5 6 |
0.726 0.726 0.726 0.726 0.726 0.726 |
0.690 0.686 0.710 0.702 0.724 0.702 |
|
Table 4.2.3. Detection Limit of the Overall Procedure for 1M2PA
|
sample no. |
µg spiked |
µg recovered |
|
1 2 3 4 5 6 |
1.066 1.066 1.066 1.066 1.066 1.066 |
1.048 1.093 1.086 1.102 1.017 1.144 |
|
|
Table 4.2.2. Detection Limit of the Overall Procedure for 2M1P
|
sample no. |
µg spiked |
µg recovered |
|
1 2 3 4 5 6 |
0.754 0.754 0.754 0.754 0.754 0.754 |
0.713 0.756 0.742 0.753 0.720 0.760 |
|
Table 4.2.4. Detection Limit of the Overall Procedure for 2M1PA
|
sample no. |
µg spiked |
µg recovered |
|
1 2 3 4 5 6 |
1.066 1.066 1.066 1.066 1.066 1.066 |
1.079 1.140 1.098 1.075 1.099 1.069 |
|
|
4.3. Reliable quantitation limit
The reliable quantitation limits of 0.73, 0.75, 1.1, and 1.1 g per sample were determined
by analyzing six samples for each analyte that had been spiked with 0.726, 0.754, 1.066,
and 1.066 µg of 1M2P, 2M1P, 1M2PA, and 2M1PA respectively. The reliable quantitation
limits correspond to air concentrations of 20 ppb (74 µg/m³), 20 ppb (74 µg/m³), 20 ppb (108 µg/m³), and 20 ppb (108 µg/m³) for 1M2P, 2M1P, 1M2PA, and 2M1PA respectively.
Table 4.3.1. Reliable Quantitation Limit for 1M2P
(Based on samples and data of Table 4.2.1.)
|
percent recovered | statistics |
|
95.0 94.5 97.8 96.7 99.7 96.7 |
mean = SD = Precision = = |
96.7 1.9 (1.96)(±1.9) ±3.7 |
|
Table 4.3.3. Reliable Quantitation Limit for 1M2PA
(Based on samples and data of Table 4.2.3.)
|
percent recovered | statistics |
|
98.3 102.5 101.9 103.4 95.4 107.3 |
mean = SD = Precision = = |
101.5 4.1 (1.96)(±4.1) ±8.0 |
|
|
Table 4.3.2. Reliable Quantitation Limit for 2M1P
(Based on samples and data of Table 4.2.2.)
|
percent recovered | statistics |
|
94.6 100.3 98.4 99.9 95.5 100.8 |
mean = SD = Precision = = |
98.2 2.6 (1.96)(±2.6) ±5.1 |
|
Table 4.3.4. Reliable Quantitation Limit for 2M1PA
(Based on samples and data of Table 4.2.4.)
|
percent recovered | statistics |
|
101.2 106.9 103.0 100.8 103.1 100.3 |
mean = SD = Precision = = |
102.6 2.4 (1.96)(±2.4) ±4.7 |
|
|
4.4. Instrument response to the analyte
The instrument response to the analytes over the range of 0.5 to 2 times the target
concentrations was determined from multiple injections of analytical standards. The response
is linear for all four analytes with slopes (in ISTD-corrected area counts per microgram of analyte per sample) of 275, 290, 291, and 290 for 1M2P, 2M1P, 1M2PA, and 2M1PA
respectively.
Table 4.4.1. Instrument Response to 1M2P
|
× target concn µg/sample ppm |
0.5× 1815 49.3 |
1× 3630 98.5 |
2× 7260 197 |
|
area counts |
503710 501140 504620 502310 504370 503980 |
1005400 1004600 1006400 1003300 1004000 1001500 |
2005600 2001400 1991800 2002700 1992000 1997800 |
mean | 503360 | 1004200 | 1998600 |
|
| |
Figure 4.4.1. Instrument response to 1M2P. (Slope = 275)
|
Table 4.4.2. Instrument Response to 2M1P
|
× target concn µg/sample ppm |
0.5× 18.89 0.51 |
1× 37.78 1.03 |
2× 75.56 2.05 |
|
area counts |
5534 5490 5532 5534 5544 5539 |
11000 11055 11018 10987 10980 10881 |
22048 21910 21957 22031 21782 21909 |
mean | 5529 | 10987 | 21940 |
|
| |
Figure 4.4.2. Instrument response to 2M1P. (Slope = 290)
|
Table 4.4.3. Instrument Response to 1M2PA
|
× target concn µg/sample ppm |
0.5× 2664 49.3 |
1× 5328 98.6 |
2× 10657 197 |
|
area counts |
751990 757280 755840 753060 753250 755340 |
1544600 1541100 1547500 1535200 1540700 1538500 |
3086000 3073400 3089900 3076200 3083300 3084000 |
mean | 503360 | 1541300 | 3082100 |
|
| |
Figure 4.4.3. Instrument response to 1M2PA. (Slope = 291)
|
Table 4.4.4. Instrument Response to 2M1PA
|
× target concn µg/sample ppm |
0.5× 26.64 0.49 |
1× 53.28 0.99 |
2× 106.6 1.97 |
|
area counts |
7615 7575 7639 7627 7678 7646 |
15492 15425 15513 15386 15470 15314 |
30869 30848 31009 30759 30804 30871 |
mean | 7630 | 15433 | 30860 |
|
| |
Figure 4.4.4. Instrument response to 2M1PA. (Slope = 290)
|
4.5. Storage test
Storage samples were generated by sampling from atmospheres containing the analytes at
the target concentrations. 1M2P and 2M1P were generated in the same atmosphere and
1M2PA and 2M1PA were generated together in another atmosphere. For each set of 36
samples, six samples were analyzed immediately after generation, fifteen were stored in a
refrigerator at 0°C and fifteen were stored in a closed drawer at ambient temperatures of
20-25°C. Six samples from each set, three from refrigerated and three from ambient storage, were analyzed in three-day intervals over a period of fifteen days. The results are given in Tables 4.5.1.- 4.5.4. and shown graphically in Figures
4.5.1.1., 4.5.1.2.,
4.5.2.1., 4.5.2.2.,
4.5.3.1., 4.5.3.2.,
4.5.4.1., and 4.5.4.2.
Table 4.5.1. Storage Test Data for 1M2P
|
days of storage |
|
% recovery (refrigerated) |
|
% recovery (ambient) |
|
0 0 3 6 9 12 15 |
|
102.3 101.6 101.4 99.2 101.8 99.5 98.8 |
101.2 100.2 100.5 99.8 102.7 99.4 99.0 |
100.6 100.1 101.3 100.3 102.1 99.6 100.0 |
|
102.3 101.6 99.1 99.4 102.9 99.0 102.6 |
101.2 100.2 99.7 98.9 102.8 98.5
103.3 |
100.6 100.1 98.9 99.4 103.4 99.0 102.3 |
|
Table 4.5.2. Storage Test Data for 2M1P
|
days of storage |
|
% recovery (refrigerated) |
|
% recovery (ambient) |
|
0 0 3 6 9 12 15 |
|
105.1 100.9 101.8 96.4 95.0 98.5 94.5 |
98.8 103.0 95.9 98.8 95.7 96.6 92.3 |
96.7 100.4 98.0 98.4 95.7 97.3 96.9 |
|
105.1 100.9 103.2 95.9 94.8 97.4
95.8 |
98.8 103.0 100.1 98.2 93.8 97.5 94.3 |
96.7 100.4 96.8 97.2 94.6 97.2 96.0 |
|
Table 4.5.3. Storage Test Data for 1M2PA
|
days of storage |
|
% recovery (refrigerated) |
|
% recovery (ambient) |
|
0 0 3 6 9 12 15 |
|
101.1 99.2 96.4 98.4 96.6 96.9 99.2 |
100.5 98.8 97.5 97.3 96.8 97.5 99.1 |
98.4 99.8 97.6 97.7 92.8 97.0 100.0 |
|
101.1 99.2 97.2 98.8 98.7 98.8 97.8 |
100.5 98.8 97.3 98.6 98.8 98.6 97.6 |
98.4 99.8 97.2 99.1 98.1 98.7 95.6 |
|
Table 4.5.4. Storage Test Data for 2M1PA
|
days of storage |
|
% recovery (refrigerated) |
|
% recovery (ambient) |
|
0 0 3 6 9 12 15 |
|
101.3 99.0 95.8 96.1 95.0 95.3 95.0 |
100.8 101.3 96.0 96.8 94.5 95.5 95.7 |
98.4 100.0 96.9 96.6 91.2 93.8 96.6 |
|
101.3 99.0 93.8 94.6 93.8 92.5 91.2 |
100.8 101.3 92.9 93.8 94.2 93.4 89.7 |
98.4 100.0 94.0 95.5 93.3 93.3 87.0 |
|
4.6. Precision (analytical procedure)
The precision of the analytical procedure for each analyte is the pooled coefficient of
variation determined from replicate injections of standards. The coefficients of variation (CV) are calculated from the data from Tables 4.4.1.-4.4.4. The pooled coefficients of variation are 0.0025, 0.0045, 0.0025, and 0.0041 for 1M2P, 2M1P, 1M2PA, and 2M1PA respectively.
Table 4.6.1. Precision of the Analytical Method for 1M2P
(Based on the Data of Table 4.4.1.)
|
× target concn µg/sample ppm |
0.5× 1815 49.3 |
1× 3630 98.5 |
2× 7260 197 |
|
mean CV |
1352 0.00269 |
1708 0.00170 |
5728 0.00287 |
|
|
Table 4.6.2. Precision of the Analytical Method for 2M1P
(Based on the Data of Table 4.4.2.)
|
× target concn µg/sample ppm |
0.5× 18.89 0.51 |
1× 37.78 1.03 |
2× 75.56 2.05 |
|
mean CV |
19.5 0.00353 |
58.4 0.00532 |
97.0 0.00442 |
|
|
Table 4.6.3. Precision of the Analytical Method for 1M2PA
(Based on the Data of Table 4.4.3.)
|
× target concn µg/sample ppm |
0.5× 2664 49.3 |
1× 5328 98.6 |
2× 10657 197 |
|
mean CV |
2008 0.00266 |
4354 0.00282 |
6190 0.00201 |
|
|
Table 4.6.4. Precision of the Analytical Method for 2M1PA
(Based on the Data of Table 4.4.4.)
|
× target concn µg/sample ppm |
0.5× 26.64 0.49 |
1× 53.28 0.99 |
2× 106.6 1.97 |
|
mean CV |
34.4 0.00451 |
74.4 0.00482 |
84.7 0.00274 |
|
|
4.7. Precision (overall procedure)
The precision of the overall procedure is determined from the storage data. The
determination of the standard error of estimate (SEE) for a regression line plotted
through the graphed storage data allows the inclusion of storage time as one of the
factors affecting overall precision. The SEE is similar to the standard deviation,
except it is a measure of dispersion of data about a regression line instead of about a mean. It is determined with the following equation:
where |
n =
k =
k = |
total no. of data points
2 for linear regression
3 for quadratic regression |
|
Yobs = |
observed percent recovery at a given time |
|
Yest = |
estimated percent recovery from the regression line at the same given time |
An additional 5% for pump error is added to the SEE by the addition of variances. The
precision at the 95% confidence level is obtained by multiplying the SEE (with pump error
included) by 1.96 (the z-statistic from the standard normal distribution at the 95% confidence level). The 95% confidence intervals are drawn about their respective regression line in the storage graph as shown in Figure 4.5.1.1. The data for Figures 4.5.1.2., 4.5.2.2.,
4.5.3.2. and 4.5.4.2. were used to determine the SEEs of ±5.3%, ±5.5%, ±5.1%, and ±5.4% and the precisions of the overall procedure of ±10.3%, ±10.8%, ±10.0%, and ±10.5% for 1M2P, 2M1P, 1M2PA, and 2M1PA respectively.
4.8. Reproducibility
Six samples for each analyte collected from controlled test atmospheres (at about
80% R.H., 20-26°C, 86-88 kPa) were analyzed by a chemist unassociated with this evaluation. The samples were stored in a refrigerator for nine days before being analyzed.
Table 4.8.1. Reproducibility Data for 1M2P
|
µg expected |
µg found |
% found |
deviation |
|
2861 2810 2773 2766 2848 3030 |
2807 2750 2764 2750 2866 3050 |
98.1 97.9 99.7 99.4 100.6 100.7 |
-1.9 -2.1 -0.3 -0.6 +0.6 +0.7 |
|
|
Table 4.8.2. Reproducibility Data for 2M1P
|
µg expected |
µg found |
% found |
deviation |
|
29.10 28.58 28.20 28.13 28.97 30.82 |
28.18 27.31 27.70 27.66 28.83 30.52 |
96.8 95.6 98.2 98.3 99.5 99.0 |
-3.2 -4.4 -1.8 -1.7 -0.5 -1.0 |
|
|
Table 4.8.3. Reproducibility Data for 1M2PA
|
µg expected |
µg found |
% found |
deviation |
|
5824 5721 5645 5630 5798 6168 |
5583 5533 5517 5487 5711 6044 |
95.9 96.7 97.7 97.5 98.5 98.0 |
-4.1 -3.3 -2.3 -2.5 -1.5 -2.0 |
|
|
Table 4.8.4. Reproducibility Data for 2M1PA
|
µg expected |
µg found |
% found |
deviation |
|
58.24 57.21 56.45 56.30 57.98 61.68 |
54.81 54.45 54.20 53.89 55.98 59.28 |
94.1 95.2 96.0 95.7 96.6 96.1 |
-5.9 -4.8 -4.0 -4.3 -3.4 -3.9 |
|
|
4.9. Desorption efficiency [from dry charcoal using 95/5 (v/v) methylene chloride/methanol]
The average desorption efficiencies of 1M2P, 2M1P, 1M2PA, and 2M1PA are 100.4%, 99.7%, 101.8%, and 101.4% respectively over the range of 0.5 to 2 times the target concentrations. They
were determined by injecting microliter amounts of stock standards into the front section of Lot 120 charcoal tubes. Eighteen samples were prepared, six samples for each concentration level listed in the following tables. The samples were stored in a refrigerator and analyzed the next day.
Table 4.9.1. Desorption Efficiency Data for 1M2P and 2M1P
|
| desorption efficiency, % |
|
|
| 1M2P |
| 2M1P |
|
|
× target concn µg/sample ppm |
0.5× 1815 49.3 |
1× 3630 98.5 |
2× 7260 197 |
|
0.5× 18.89 0.51 |
1× 37.78 1.03 |
2× 75.56 2.05 |
|
|
100.8 100.3 99.5 100.2 101.5 100.7 |
99.6 101.5 100.0 100.7 100.4 100.1 |
99.9 100.5 100.1 100.5 99.7 100.3 |
|
99.9 98.1 97.5 97.3 99.5 101.4 |
99.7 101.2 101.8 103.1 97.4 100.6 |
99.4 100.2 97.7 100.2 98.7 100.7 |
|
mean grand mean |
100.5 |
100.4 100.4 |
100.2 |
99.0 |
100.6 99.7 |
99.5 |
|
Table 4.9.2.
Desorption Efficiency Data for 1M2PA and 2M1PA
|
| desorption efficiency, % |
|
|
| 1M2PA |
|
| 2M1PA |
|
|
× target concn µg/sample ppm |
0.5× 2664 49.3 |
1× 5328 98.6 |
2× 10657 197 |
|
0.5× 26.64 0.49 |
1× 53.28 0.99 |
2× 106.6 1.97 |
|
|
102.2 101.6 101.6 101.9 101.3 101.6 |
102.3 102.1 101.8 101.9 102.1 101.8 |
101.9 101.5 101.7 102.2 101.8 101.6 |
|
102.5 102.2 101.2 100.8 101.2 101.1 |
100.8 101.9 101.4 100.0 101.3 101.8 |
101.6 100.9 101.9 102.6 101.2 101.6 |
|
mean grand mean |
101.7 |
102.0 101.8 |
101.8 |
101.5 |
101.2 101.4 |
101.6 |
|
4.10. Stability of desorbed samples [from dry charcoal using 95/5 (v/v) methylene chloride/methanol]
The stability of desorbed samples was checked by reanalyzing the target
concentration samples from Section 4.9. one day later using fresh standards. The sample vials were resealed with new septa after the original analyses and were
allowed to stand at room temperature until reanalyzed.
| |
Table 4.10.
Stability of Desorbed Samples at the Target Concentration after 24 Hours
|
| desorption efficiency, % |
|
|
| 1M2P | 2M1P | 1M2PA | 2M1PA |
|
|
mean |
100.1 101.1 99.9 101.2 100.4 101.0 100.6 |
99.8 100.6 101.8 100.2 99.5 102.4 100.7 |
102.9 102.2 101.9 102.2 102.4 102.0 102.3 |
102.9 101.9 102.1 101.1 102.9 100.0 101.8 |
|
|
4.11. Desorption efficiency and stability of desorbed samples [from wet charcoal using 95/5 (v/v) methylene chloride/methanol]
Studies were done at the target concentrations to determine what effect the presence of
water had on the desorption efficiency for the four analytes. This was done by injecting
analytical standards into the front sections of charcoal tubes that previously had 10 L of 80% RH air drawn through them. The samples were reanalyzed 24 h later to check the stability of
desorbed samples. Finally, magnesium sulfate was added to the desorbed samples and they were
reanalyzed again.
Table 4.11.1. Desorption of 1M2P from Wet Charcoal
|
|
initial analysis |
next day analysis |
next day + MgSO4 |
|
% desorption |
96.0 96.4 96.1 96.7 96.2 96.2 |
96.1 96.9 96.6 97.4 96.9 97.0 |
103.0 101.9 101.3 102.3 102.4 101.9 |
|
mean |
96.3 |
96.8 |
102.1 |
|
|
Table 4.11.2. Desorption of 2M1P from Wet Charcoal
|
|
initial analysis |
next day analysis |
next day + MgSO4 |
|
% desorption |
92.3 94.8 94.6 93.4 93.3 93.0 |
95.3 96.8 94.7 96.2 94.8 94.8 |
102.6 103.1 103.2 102.7 102.8 101.5 |
|
mean |
93.6 |
95.4 |
102.6 |
|
|
Table 4.11.3. Desorption of 1M2PA from Wet Charcoal
|
|
initial analysis |
next day analysis |
next day + MgSO4 |
|
% desorption |
102.7 103.0 103.0 103.0 103.5 102.3 |
102.5 102.7 102.6 102.7 103.5 102.5 |
102.2 102.3 102.7 102.6 102.5 101.9 |
|
mean |
102.9 |
102.8 |
102.4 |
|
|
Table 4.11.4. Desorption of 2M1PA from Wet Charcoal
|
|
initial analysis |
next day analysis |
next day + MgSO4 |
|
% desorption |
102.4 103.2 102.5 101.9 102.7 100.7 |
103.2 102.5 103.2 103.8 102.5 103.6 |
101.8 103.4 104.0 103.0 103.1 101.4 |
|
mean |
102.2 |
103.1 |
102.8 |
|
|
4.12. Desorption efficiency [from dry charcoal using 99/1 (v/v) CS2/DMF]
The average desorption efficiencies of 1M2P, 2M1P, 1M2PA, and 2M1PA are 86.3%, 80.8%, 98.4%, and 98.0% respectively over the range of 0.5 to 2 times the target concentrations. They
were determined by injecting microliter amounts of stock standards into the front section of Lot 120 charcoal tubes. Eighteen samples were prepared, six samples for each concentration level listed in the following tables. The samples were stored in a refrigerator and analyzed the next day.
Table 4.12.1. Desorption Efficiency Data for 1M2P and 2M1P
|
| desorption efficiency, % |
|
|
| 1M2P |
| 2M1P |
|
|
× target concn µg/sample ppm |
0.5× 1815 49.3 |
1× 3630 98.5 |
2× 7260 197 |
|
0.5× 18.89 0.51 |
1× 37.78 1.03 |
2× 75.56 2.05 |
|
|
84.1 84.7 84.8 83.3 84.1 83.5 |
87.6 87.9 88.8 89.0 84.4 86.8 |
89.2 88.7 83.4 89.1 88.6 85.4 |
|
81.0 78.7 79.1 76.5 76.5 75.2 |
80.0 80.6 84.1 83.8 83.7 81.0 |
83.6 82.7 80.5 85.3 81.9 79.7 |
|
mean grand mean |
84.1 |
87.4 86.3 |
87.4 |
77.8 |
82.2 80.8 |
82.3 |
|
Table 4.12.2. Desorption Efficiency Data for 1M2PA and 2M1PA
|
| desorption efficiency, % |
|
|
| 1M2P |
| 2M1P |
|
|
× target concn µg/sample ppm |
0.5× 2664 49.3 |
1× 5328 98.6 |
2× 10657 197 |
|
0.5× 26.64 0.49 |
1× 53.28 0.99 |
2× 106.6 1.97 |
|
|
99.1 98.8 98.7 98.8 98.4 98.3 |
97.9 98.6 98.2 97.5 97.7 98.0 |
98.9 98.6 98.5 98.6 98.3 98.9 |
|
97.9 98.2 97.6 98.7 98.7 98.0 |
97.8 98.9 98.1 97.5 97.6 98.0 |
97.9 97.9 98.0 97.7 97.7 98.4 |
|
mean grand mean |
98.7 |
98.0 98.4 |
98.6 |
98.2 |
98.0 98.0 |
97.9 |
|
4.13. Stability of desorbed samples [from dry charcoal using 99/1 (v/v) CS2/DMF]
The stability of desorbed samples was checked by reanalyzing the target concentration samples from Section 4.12. one day later using fresh standards. The
sample vials were resealed with new septa after the original analyses and were
allowed to stand at room temperature until reanalyzed.
| |
Table 4.13.
Stability of Desorbed Samples at the Target Concentration after 24 Hours
|
| desorption efficiency, % |
|
|
| 1M2P | 2M1P | 1M2PA | 2M1PA |
|
|
mean |
80.9 81.4 85.2 84.9 86.8 82.7 83.7 |
72.3 72.4 78.9 77.4 78.6 75.1 75.8 |
98.2 98.1 97.1 97.6 97.6 97.6 97.7 |
97.8 97.9 96.4 97.5 96.9 97.2 97.3 |
|
|
4.14. Desorption efficiency and stability of desorbed samples [from wet charcoal using 99/1 (v/v) CS2/DMF]
Studies were done at the target concentrations to determine what effect the presence of
water had on the desorption efficiency for the four analytes. This was done by injecting
analytical standards into the front sections of charcoal tubes that previously had 10 L of 80%
RH air drawn through them. The samples were reanalyzed 24 h later to check the stability
of desorbed samples. Magnesium sulfate was added to one-half of the samples when they were desorbed.
Table 4.14.1. Desorption of 1M2P from Wet Charcoal
|
|
initial (no MgSO4) |
next day (no MgSO4) |
initial (with MgSO4) |
next day (with MgSO4) |
|
% desorption |
50.9 50.9 50.8 |
48.0 48.1 49.7 |
79.5 83.0 76.2 |
78.9 75.3 79.7 |
|
mean |
50.9 |
48.6 |
79.6 |
78.0 |
|
Table 4.14.2. Desorption of 2M1P from Wet Charcoal
|
|
initial (no MgSO4) |
next day (no MgSO4) |
initial (with MgSO4) |
next day (with MgSO4) |
|
% desorption |
40.5 40.4 40.7 |
36.9 37.7 39.0 |
70.1 74.7 66.6 |
65.8 59.1 68.1 |
|
mean |
40.5 |
37.9 |
70.5 |
64.3 |
|
Table 4.14.3. Desorption of 1M2PA from Wet Charcoal
|
|
initial (no MgSO4) |
next day (no MgSO4) |
initial (with MgSO4) |
next day (with MgSO4) |
|
% desorption |
93.2 94.5 93.0 |
94.6 94.3 93.8 |
95.8 96.6 96.7 |
96.5 95.8 95.2 |
|
mean |
93.6 |
94.2 |
96.4 |
95.8 |
|
Table 4.14.4. Desorption of 2M1PA from Wet Charcoal
|
|
initial (no MgSO4) |
next day (no MgSO4) |
initial (with MgSO4) |
next day (with MgSO4) |
|
% desorption |
91.7 93.0 91.3 |
92.8 92.5 91.3 |
95.4 96.2 96.6 |
95.7 94.9 94.7 |
|
mean |
92.0 |
92.2 |
96.1 |
95.1 |
|
5. References
5.1. "Current Intelligence Bulletin 39, Glycol Ethers"; May 2, 1983, U.S. Department of Health and Human Services, Public Health Service, Center for Disease Control, NIOSH.
5.2. "NEG and NIOSH basis for an occupational health standard: Propylene Glycol Ethers and Their Acetates"; 1991, U.S. Department of Health and Human Services, Public Health Service, Center for Disease Control, NIOSH, Publication No.
91-103.
5.3. "Table Z-1-A -- Limits for Air Contaminants", Code of Federal Regulations, Title 29; 1910.1000, U.S. Office of the Federal Register National Archives and Records Administration, Washington, DC; 1991.
5.4. "OSHA Analytical Methods Manual" U.S. Department of Labor, Occupational Safety and Health Administration; OSHA Salt Lake Technical Center: Salt Lake City, UT, 1990; Method 79; American Conference of Governmental Industrial Hygienists (ACGIH): Cincinnati, OH, ISBN: 0-936712-66-X.
5.5. Elskamp, C.J. "OSHA Method No. 83; 2-Butoxyethanol and 2-Butoxyethyl Acetate", OSHA Salt Lake Technical Center, unpublished, Salt Lake City, UT 84165-0200, May 1990.
5.6. Handbook of Chemistry and Physics, 70th ed.; CRC Press, Inc.: Boca Raton, FL, 1989-1990.
|