Kansas Water Science Center
USGS Water Science Centers are located in each state. |
U.S. Geological Survey
|
Multiply | By | To obtain |
---|---|---|
gram (g) | 2.205 x 10-3 | pound |
liter (L) | 2.642 x 10-1 | gallon |
meter (m) | 3.281 | foot |
microliter (µL) | 2.642 x 10-7 | gallon |
micrometer (µm) | 3.937 x 10-5 | inch |
milligram (mg) | 3.53 x 10-5 | ounce |
millimeter (mm) | 3.937 x 10-2 | inch |
ounce (oz) | 0.02957 | liter |
pound per acre per year [(lb/acre)/yr] | 1.121 | kilogram per acre per year |
pound per square inch (lb/in²) | 6.895 | kilopascal |
Miscellaneous Abbreviations | |
---|---|
mass to charge (m/z) volt (V) cubic centimeter (cm³) |
|
Abbreviated Water-Quality Units | |
liter per minute (L/min) microgram per liter (µg/L) milliliter (mL) milliliter per minute (mL/min) molar (M) nanogram per microliter (ng/µL) |
An analytical method using high-performance liquid chromatography/mass spectrometry (HPLC/MS) was developed by the U.S. Geological Survey in 1999 for the analysis of selected chloroacetanilide herbicide degradation compounds in water. These compounds were acetochlor ethane sulfonic acid (ESA), acetochlor oxanilic acid (OXA), alachlor ESA, alachlor OXA, metolachlor ESA, and metolachlor OXA. The HPLC/MS method was updated in 2000, and the method detection limits were modified accordingly. Four other degradation compounds also were added to the list of compounds that can be analyzed using HPLC/MS; these compounds were dimethenamid ESA, dimethenamid OXA, flufenacet ESA, and flufenacet OXA.
Except for flufenacet OXA, good precision and accuracy were demonstrated for the updated HPLC/MS method in buffered reagent water, surface water, and ground water. The mean HPLC/MS recoveries of the degradation compounds from water samples spiked at 0.20 and 1.0 µg/L (microgram per liter) ranged from 75 to 114 percent, with relative standard deviations of 15.8 percent or less for all compounds except flufenacet OXA, which had relative standard deviations ranging from 11.3 to 48.9 percent. Method detection levels (MDL's) using the updated HPLC/MS method varied from 0.009 to 0.045 µg/L, with the flufenacet OXA MDL at 0.072 µg/L. The updated HPLC/MS method is valuable for acquiring information about the fate and transport of the parent chloroacetanilide herbicides in water.
The chloroacetanilide herbicides-acetochlor, alachlor, dimethenamid, flufenacet, and metolachlor are an important class of herbicides in the United States. Together with the triazine compounds, chloroacetanilide herbicides compose the majority of pesticides applied in the Midwestern United States for control of weeds in corn, soybeans, and other row crops (Gianessi and Anderson, 1995). Alachlor and metolachlor have been used extensively for more than 20 years, whereas acetochlor application is relatively recent, having been applied extensively since March 1994 (Kolpin, Nations, and others, 1996). Chloroacetanilide herbicides have been shown to degrade more rapidly in soil than other herbicides, with half-lives from 15 to 30 days. Triazine half-lives are typically 30 to 60 days (Leonard, 1988).
The herbicide dimethenamid was registered with the U.S. Environmental Protection Agency in 1993. It has a recommended maximum application rate of 1.5 (lb/acre)/yr on corn and was ranked sixth in herbicide usage during 1998 (U.S. Department of Agriculture, Agricultural Chemical Usage, 1999). It is used most extensively in Northern States, particularly Wisconsin where it was applied to 28 percent of the corn acreage in 1998 (U.S. Department of Agriculture, Agricultural Chemical Usage, 1999). The herbicide flufenacet is used to control certain annual grasses and broadleaf weeds. It has a recommended application rate of 0.78 (lb/acre)/yr (U.S. Department of Agriculture, Agricultural Chemical Usage, 1999).
Recent studies have reported the occurrence of chloroacetanilide degradation compounds in surface and ground water (Aga and others, 1996; Kolpin, Thurman, and Goolsby, 1996; Thurman and others, 1996; Kolpin and others, 1998). Kolpin and others (1998) found that degradation compound concentrations in ground water may be at similar or even higher concentrations than the parent compounds, whereas in surface water the parent compounds are more abundant in the spring after application and are replaced gradually by degradation compounds during the remaining growing season.
In understanding the fate and transport of parent compounds, reliable methods for the analysis of degradation compounds are vital. Reliable methods also are important for analytical verification of the degradation compounds in toxicological studies.
This report provides a description of a reliable, previously published method (O-2134-00) for the analysis of ethane sulfonic acid (ESA) and oxanilic acid (OXA) degradation compounds of acetochlor, alachlor, and metolachlor found in surface water and ground water using high-performance liquid chromatography/mass spectrometry (HPLC/MS) (Zimmerman and others, 2000). Since publication of the original method, several modifications have been made to achieve chromatographic separation of alachlor and acetochlor peaks. Moreover, dimethenamid ESA and OXA and flufenacet ESA and OXA have been added to the list of chloroacetanalide degradation compounds suitable for determination using the HPLC/MS method.
The original HPLC/MS method was derived from Ferrer and others (1997), with minor modification to resolve co-eluting peaks on the chromatogram as reported in Hostetler and Thurman (1999). The updated method supplements other methods of the U.S. Geological Survey (USGS) and has been implemented by the USGS Organic Geochemistry Research Group in Lawrence, Kansas.
The updated HPLC/MS method of analysis described in this report has also been assigned the method number "O-2134-00." This unique code represents the HPLC/MS automated method of analysis for organic compounds as described in this report and can be used to identify the method. This report provides a detailed description of the method, including the apparatus, reagents, instrument calibration, and the solid-phase extraction (SPE) procedure required for sample analysis. Estimated method detection limits, mean recoveries, and relative standard deviations for the six original and four additional chloroacetanilide herbicide degradation compounds determined using HPLC/MS are presented. The USGS parameter and method codes for these compounds are also given.
The updated HPLC/MS method is suitable for the determination of low concentrations (in micrograms per liter) of chloroacetanilide degradation compounds in water samples (table 1). Because suspended particulate matter is removed from the samples by filtration, this method is suitable only for dissolved-phase degradation compounds.
[ESA, ethane sulfonic acid; OXA, oxanilic acid]
Degradation compound | Molecular weight (atomic mass units) |
---|---|
Acetochlor ESA | 315.4 |
Acetochlor OXA | 265.3 |
Alachlor ESA | 315.4 |
Alachlor OXA | 265.3 |
Dimethenamid ESA | 321.4 |
Dimethenamid OXA | 271.4 |
Flufenacet ESA | 275.3 |
Flufenacet OXA | 225.3 |
Metolachlor ESA | 329.4 |
Metolachlor OXA | 279.3 |
Degradation compounds were selected for analysis because of the extensive use of the parent herbicides in the United States and their importance to current (2000) studies being conducted by the USGS. This method is applicable to concentrations from 0.05 to 5.0 µg/L without dilution.
Water samples are filtered at the collection site using glass-fiber filters with nominal 0.7-µm pore diameter to remove suspended particulate matter. In the laboratory, the filtered water sample is passed through a preconditioned C-18 (C18H37) column. The C-18 column is rinsed with ethyl acetate to remove interfering compounds. The adsorbed chloroacetanilide degradation compounds are eluted from the C-18 with methanol. The solution is spiked with an internal standard, evaporated under nitrogen, and reconstituted. The sample components are separated, identified, and measured by injecting an aliquot of the concentrated extract into an HPLC equipped with a diode array detector (DAD) and a mass spectrometer (MS) detector. Compounds eluting from the liquid chromatograph (LC) are identified by comparing the retention times of the mass spectral signals against the measurement of standards analyzed using the same conditions used for the samples. Compounds are identified further by selected fragment ions for compounds that produce fragment ions. The concentration of each identified compound is calculated by determining the ratio of the MS response produced by that compound to the MS response produced by the internal standard, which was injected into the sample, to the ratio of the MS responses of primary standards analyzed using the same method. The USGS parameter and method codes for the degradation compounds analyzed using method O-2134-00 are listed in table 2.
[ESA, ethane sulfonic acid; OXA, oxanilic acid]
Degradation compound |
Parameter code | Method code |
---|---|---|
Acetochlor ESA | 61029 | x |
Acetochlor OXA | 61030 | x |
Alachlor ESA | 50009 | x |
Alachlor OXA | 61031 | x |
Dimethenamid ESA | 61951 | x |
Dimethenamid OXA | 62482 | x |
Flufenacet ESA | 61592 | x |
Flufenacet OXA | 62483 | x |
Metolachlor ESA | 61043 | x |
Metolachlor OXA | 61044 | x |
Compounds that elute from the LC at the same time and have mass similar to the degradation compounds may interfere. Samples with high concentrations of humic materials may cause interference with the ionization of the internal standard if they elute from the LC at the same time.
Sampling methods capable of collecting water samples that accurately represent the water-quality characteristics of the surface water or ground water at a given time or location are used. Detailed descriptions of sampling methods used by the USGS for obtaining depth- and width-integrated surface-water samples are given in Edwards and Glysson (1988) and Ward and Harr (1990). Similar descriptions of sampling methods for obtaining ground-water samples are given in Hardy and others (1989).
Sample-collection equipment must be free of tubing, gaskets, and other components made of nonfluorinated plastic material that might leach interfering compounds into water samples or absorb the degradation compounds from the water. The water samples from each site are composited in a single container and filtered through a nominal 0.7-µm glass-fiber filter using a peristaltic pump. Filters are preconditioned with about 200 mL of sample prior to filtration of the sample. The filtrate for analysis is collected in baked 125-mL amber glass bottles with Teflon-lined lids. Samples are chilled immediately and shipped to the laboratory within 3 days of collection. At the laboratory, samples are logged in, assigned identification numbers, and refrigerated at 4 ±2°C until extracted and analyzed.
A calibration table and calibration curve from the analyzed extracted standards are prepared using the HP LC/MSD Chemstation software (Hewlett Packard, Wilmington, DE). Manufacture's instructions are followed for using the internal standard as a time reference and for quantitation.
RRTc = RTc/RTi ,
(1)
where
where RTc = uncorrected retention
time of the selected compound, and
where RTi = uncorrected retention
time of the internal standard (2,4-D).
See table 3 for retention times, relative retention times, and confirming ions.
[m/z, mass-to-charge ratio; ESA, ethane sulfonic acid; OXA, oxanilic acid; --, not determined]
Degradation compound | Retention time (minutes) |
Relative retention time |
Molecular ion (m/z) |
Fragment ion (m/z) |
|||
---|---|---|---|---|---|---|---|
Chloroacetanilide degradation compounds (in order of increasing retention time) | |||||||
Flufenacet OXA | 35.749 | 1.680 | 224 | 152 | |||
Dimethenamide OXA | 37.841 | 1.779 | 270 | 198 | |||
Metolachlor OXA | 45.012 | 2.116 | 278 | 206 | |||
Alachlor OXA | 57.764 | 2.715 | 264 | 160 | |||
Acetochlor OXA | 57.865 | 2.720 | 264 | 146 | |||
Flufenacet ESA | 60.173 | 2.828 | 274 | -- | |||
Dimethenamid ESA | 63.224 | 2.972 | 320 | -- | |||
Alachlor ESA | 77.870 | 3.660 | 314 | -- | |||
Metolachlor ESA | 79.269 | 3.726 | 328 | -- | |||
Acetochlor ESA | 79.855 | 3.753 | 314 | -- | |||
Internal standard | |||||||
2,4-dichlorophenoxyacetic acid | 21.275 | 1.000 | 219 | 161 |
RT = (RTTc)(RTi ),
(2)
where
where RT = expected retention time of the
selected compound,
where RTTc = relative retention time
of the selected compound, and
where RTi = uncorrected retention
time of the internal standard.
DF = (123/123 - Vnp)(123/123 - Va ),
(3)
where
where Vnp = volume not pumped =
milliliters not pumped through the SPE column, and
where Va = volume added = milliliters
of distilled water added to a sample that contains less than 123 mL.
The dilution factor is incorporated into the calculation for determining final concentrations of samples.
Extraction efficiency is determined by analyzing seven standards of the same concentrations used for extraction that are prepared for direct injection into the HPLC/MS. The extraction efficiency is the slope of the line obtained by plotting the value of the extracted standards calculated from the direct injected standards. The results are tabulated in table 4.
[ESA, ethane sulfonic acid; OXA, oxanilic acid]
Degradation compound | Extraction efficiency (slope as a percentage) |
Standard deviation (relative percentage) |
---|---|---|
Acetochlor ESA | 82.8 | 13.9 |
Acetochlor OXA | 81.2 | 13.8 |
Alachlor ESA | 81.2 | 12.4 |
Alachlor OXA | 81.6 | 14.3 |
Dimethenamid ESA | 86.8 | 16.1 |
Dimethenamid OXA | 86.5 | 21.5 |
Flufenacet ESA | 83.2 | 13.8 |
Flufenacet OXA | 74.6 | 11.0 |
Metolachlor ESA | 80.2 | 13.0 |
Metolachlor OXA | 83.2 | 13.8 |
Minimum | 74.6 | 11.0 |
Maximum | 86.8 | 21.5 |
The SPE procedure used a Tekmar six-position AutoTrace (Tekmar-Dohrmann, Cincinnati, OH). The SPE columns (C-18 Sep-Pak Vac 6 cm³) used to extract samples were obtained from Waters Corporation (Milford, MA). These vacuum cartridges contain 500 mg of 50- to 105-µm C-18 bonded to silica. The data in this report were produced using the Tekmar six-position AutoTrace procedure as listed in Appendix 1.
The HP LC/MSD Chemstation software (Hewlett Packard, Wilmington, DE) is used with the previously prepared calibration table for identification of compounds.
Alternate method (manual):
The HP LC/MSD Chemstation software (Hewlett Packard, Wilmington, DE) is used with the previously prepared calibration table for quantification of compounds.
Alternate method (manual):
C = ((Ac/Ai)(m) + y)(DF) ,
(4)
where
where C = concentration of the selected
degradation compound in the sample, in micrograms per liter;
where Ac = area of peak of the
quantitation ion for the selected degradation compound;
where Ai = area of peak of the
quantitation ion for the internal standard;
where m = slope of calibration curve using
extracted standards between the selected degradation compound and the internal
standard from the original calibration data;
where y = intercept of calibration curve between
the selected degradation compound and the internal standard from the original
calibration data; and
where DF = dilution factor calculated using
equation 3.
Chloroacetanalide herbicide degradation compounds are reported in concentrations ranging from 0.05 to 5.0 µg/L. If the concentration is greater than 5.0 mg/L, 5 mL of sample extract are reinjected and re-analyzed. If the concentration is greater than 10 µg/L, the sample is re-extracted with a 1:10 dilution (sample:distilled water) and re-analyzed for those degradation compounds that have concentrations greater than 10 µg/L.
A buffered reagent-water sample, a surface-water sample collected from Poison Creek in Valley County, Idaho, and a ground-water sample collected from a well in Valley County, Idaho, were used to test the method performance. The surface- and ground-water samples were collected in 45-L carboys and were split into 123-mL samples. One set of eight samples was spiked with 0.20 µg/L of each chloroacetanalide degradation compound, and the other set of eight samples was spiked with 1.0 µg/L of each degradation compound. In addition, unspiked samples of surface and ground water were extracted and analyzed to determine background concentrations of the pesticides. All subsamples were analyzed in one laboratory (the USGS Organic Geochemistry Research Laboratory in Lawrence, Kansas) using one HPLC/MS system. Each sample set was extracted and analyzed on different days from March through September 2000. Comparison of different matrices and concentrations included bias from day-to-day variation. Method recoveries from the analyses are listed in tables 5, 6, and 7.
Table 5. Mean recovery of chloroacetanilide herbicide degradation compounds in buffered reagent-water samples using method 0-2134-00
Eight samples spiked at 0.2 µg/L | Eight samples spiked at 1.0 µg/L | |||||||
---|---|---|---|---|---|---|---|---|
Mean recovery | Mean recovery | |||||||
Degradation compound |
(µg/L) | (percent) | Standard deviation (µg/L) |
Relative standard deviation (percent) |
(µg/L) | (percent) | Standard deviation (µg/L) |
Relative standard deviation (percent) |
Acetochlor ESA | 0.194 | 97.0 | 0.013 | 6.7 | 0.969 | 96.9 | 0.046 | 4.7 |
Acetochlor OXA | .191 | 95.5 | .013 | 6.8 | .960 | 96.0 | .048 | 5.0 |
Alachlor ESA | .183 | 91.5 | .015 | 8.2 | .932 | 93.2 | .050 | 5.4 |
Alachlor OXA | .195 | 97.5 | .017 | 8.7 | .982 | 98.2 | .064 | 6.5 |
Dimethenamid ESA | .196 | 98.0 | .018 | 9.2 | .971 | 97.1 | .069 | 7.1 |
Dimethenamid OXA | .201 | 100.5 | .018 | 9.0 | .984 | 98.4 | .075 | 7.6 |
Flufenacet ESA | .188 | 94.0 | .017 | 9.0 | .949 | 94.9 | .086 | 9.1 |
Flufenacet OXA | .169 | 84.5 | .037 | 21.9 | .758 | 75.8 | .204 | 26.9 |
Metolachlor ESA | .192 | 96.0 | .016 | 8.3 | .964 | 96.4 | .036 | 3.7 |
Metolachlor OXA | .187 | 93.5 | .020 | 10.7 | .961 | 96.1 | .045 | 4.7 |
Average | .190 | 94.8 | .018 | 9.9 | .943 | 94.3 | .072 | 8.1 |
Table 6. Mean recovery of chloroacetanilide herbicide degradation compounds in surface-water samples using method 0-2134-00
Eight samples spiked at 0.2 µg/L | Eight samples spiked at 1.0 µg/L | |||||||
---|---|---|---|---|---|---|---|---|
Mean recovery | Mean recovery | |||||||
Degradation compound |
(µg/L) | (percent) | Standard deviation (µg/L) |
Relative standard deviation (percent) |
(µg/L) | (percent) | Standard deviation (µg/L) |
Relative standard deviation (percent) |
Acetochlor ESA | 0.151 | 75.5 | 0.012 | 7.9 | 0.897 | 89.7 | 0.071 | 7.9 |
Acetochlor OXA | .188 | 94.0 | .017 | 9.0 | 1.121 | 112.1 | .118 | 10.5 |
Alachlor ESA | .150 | 75.0 | .014 | 9.3 | .898 | 89.8 | .077 | 8.6 |
Alachlor OXA | .179 | 89.5 | .020 | 11.2 | 1.017 | 101.7 | .134 | 13.2 |
Dimethenamid ESA | .165 | 82.5 | .013 | 7.9 | .865 | 86.5 | .051 | 5.9 |
Dimethenamid OXA | .204 | 102.0 | .017 | 8.3 | 1.057 | 105.7 | .094 | 9.4 |
Flufenacet ESA | .165 | 82.5 | .023 | 13.9 | 1.000 | 100.0 | .073 | 7.3 |
Flufenacet OXA | .223 | 115.5 | .109 | 48.9 | 1.135 | 113.5 | .303 | 26.7 |
Metolachlor ESA | .163 | 81.5 | .015 | 9.2 | .933 | 93.3 | .080 | 8.6 |
Metolachlor OXA | .176 | 88.0 | .017 | 9.7 | .989 | 98.9 | .112 | 11.3 |
Average | .176 | 88.2 | .026 | 13.5 | .991 | 99.1 | .111 | 10.9 |
Table 7. Mean recovery of chloroacetanilide herbicide degradation compounds in ground-water samples using method 0-2134-00
Eight samples spiked at 0.2 µg/L | Eight samples spiked at 1.0 µg/L | |||||||
---|---|---|---|---|---|---|---|---|
Mean recovery | Mean recovery | |||||||
Degradation compound |
(µg/L) | (percent) | Standard deviation (µg/L) |
Relative standard deviation (percent) |
(µg/L) | (percent) | Standard deviation (µg/L) |
Relative standard deviation (percent) |
Acetochlor ESA | 0.169 | 84.5 | 0.023 | 13.6 | 0.830 | 83.0 | 0.095 | 11.4 |
Acetochlor OXA | .185 | 92.5 | .013 | 7.0 | 1.072 | 107.2 | .134 | 12.5 |
Alachlor ESA | .155 | 77.5 | .012 | 7.7 | .831 | 83.1 | .098 | 11.8 |
Alachlor OXA | .187 | 93.5 | .014 | 7.5 | 1.000 | 100.0 | .156 | 15.6 |
Dimethenamid ESA | .170 | 85.0 | .019 | 11.2 | .839 | 83.9 | .072 | 8.6 |
Dimethenamid OXA | .198 | 99.0 | .011 | 5.6 | 1.076 | 107.6 | .126 | 11.7 |
Flufenacet ESA | .164 | 82.0 | .011 | 6.7 | .963 | 96.3 | .103 | 10.7 |
Flufenacet OXA | .159 | 79.5 | .018 | 11.3 | .980 | 98.0 | .172 | 17.6 |
Metolachlor ESA | .173 | 86.5 | .016 | 9.2 | .879 | 87.9 | .113 | 11.3 |
Metolachlor OXA | .186 | 93.0 | .014 | 7.5 | .957 | 95.7 | .151 | 15.8 |
Average | .175 | 87.3 | .015 | 8.7 | .943 | 94.3 | .122 | 12.7 |
Corrections for background concentrations-Neither surface- nor ground-water samples required correction for background concentrations of degradation compounds. All unspiked buffered reagent-water samples also had no detections of degradation compounds.
Method detection limits (MDL's)-An MDL is defined as the minimum concentration of a substance that can be identified, measured, and reported with a 99-percent confidence that the compound concentration is greater than zero. MDL's were determined according to procedures outlined by the U.S. Environmental Protection Agency (1992). Eight replicate samples of buffered reagent water spiked with 0.05 µg/L of each of the degradation compounds were analyzed to determine MDL's (table 8). Each sample was analyzed on different days from March through September 2000 so that day-to-day variation is included in the results.
[mg/L, microgram per liter; ESA, ethane sulfonic acid; OXA, oxanilic acid]
Eight samples spiked at 0.05 µg/L | |||
---|---|---|---|
Degradation compound |
Mean concentrations (µg/L) |
Standard deviation (µg/L) |
Method detection limit (µg/L) |
Acetochlor ESA | 0.053 | 0.012 | 0.036 |
Acetochlor OXA | .052 | .010 | .030 |
Alachlor ESA | .052 | .011 | .033 |
Alachlor OXA | .053 | .009 | .027 |
Dimethenamid ESA | .051 | .011 | .033 |
Dimethenamid OXA | .056 | .006 | .018 |
Flufenacet ESA | .055 | .003 | .009 |
Flufenacet OXA | .049 | .024 | .072 |
Metolachlor ESA | .052 | .011 | .033 |
Metolachlor OXA | .052 | .015 | .045 |
Minimum | .009 | ||
Maximum | .072 |
The MDL was calculated using the following equation:
MDL = (S)(t(n-1,1-a=0.99)) ,
(5)
where
where S = standard deviation of replicate
analysis, in micrograms per liter, at the spiked concentration;
where t(n-1,1-a=0.99) = Student's
t-value for the 99-percent confidence level with n-1 degrees of
freedom (U.S. Environmental Protection Agency, 1992); and
where n = number of replicate analyses.
The estimated mean MDL's ranged from 0.009 to 0.045 µg/L (table 8) for 9 of the 10 compounds, with flufenacet OXA being 0.072 µg/L (table 8). This may make low-concentration determinations of flufenacet OXA somewhat more variable than for the other nine compounds. According to the U.S. Environmental Protection Agency (1992) procedure, the spiked concentrations should be no more than five times the estimated MDL. The spiked concentrations were within five times the MDL.
Mean recovery-Mean recovery in buffered reagent-, surface-, and ground-water samples was determined by comparing the mean analyzed concentration (see "Quantitation" section) from eight replicate samples to the spiked concentration. Mean recoveries were highest overall in surface water at the 1.0-µg/L concentration (table 6) and lowest overall in ground water at 0.2 µg/L (table 7). Flufenacet OXA exhibited the greatest inconsistencies and lowest recoveries. This would indicate the extraction method is not optimized for flufenacet OXA. Alachlor ESA exhibited the lowest recoveries in all three matrices, with the lowest, 75 percent (table 6) at the 0.2-µg/L concentration, in surface water. Dimethenamid OXA exhibited consistently high recoveries in all three matrices. Relative standard deviations of the recoveries, excluding flufenacet OXA, ranged from a low of 3.7 percent to a high of 15.8 percent. Relative standard deviations for flufenacet OXA ranged from 11.3 to 48.9 percent.
An HPLC/MS method for the analysis of ethane sulfonic acids and oxanilic acids of acetochlor, alachlor, and metolachlor was reported by Ferrer and others (1997). The HPLC system described by Ferrer and others (1997) used a 5-µm, 250- x 3.0-mm C-18 column, with a mobile phase consisting of 0.3 percent acetic acid in 24 percent methanol, 36 percent distilled water, and 40 percent acetonitrile solution. With this configuration, peak resolution was not achieved for acetochlor ESA and alachlor ESA, which have the same quantitation ion (table 3). Thus, accurate quantitation of these degradation compounds was not possible. However, chromatographic separation of acetochlor ESA and alachlor ESA was achieved with the same mobile phase by coupling two 5-µm, 250- x 3.0-mm C-18 columns to one 3-µm, 150- x 2.0-mm C-18 column. The separation of the acetochlor ESA and the alachlor ESA allows quantitation of these degradation compounds. 2,4-dichlorophenoxy acid was used as the internal standard because it is amenable to negative-ion electrospray and is readily available as a commercial standard. Figure 1 shows a total ion chromatogram of a 1.0-µg/L standard in a buffered reagent-water sample. Figure 2 shows the extracted ion chromatogram for the molecular ion (314 mass-to-charge ratio) of acetochlor ESA and alachlor ESA with near baseline separation.
This report presents a method for routine analysis of 10 chloroacetanalide herbicide degradation compounds in environmental water samples. The degradation compounds are acetochlor ESA, acetochlor OXA, alachlor ESA, alachlor OXA, dimethenamid ESA, dimethenamid OXA, flufenacet ESA, flufencet OXA, metolachlor ESA, and metolachlor OXA. From the data presented in this report, solid-phase extraction and analysis using high-performance liquid chromatography/mass spectrometry (HPLC/MS) are shown to be sensitive and reliable for the determination of degradation compounds at low concentrations.
Except for flufenacet OXA, good precision and accuracy for the degradation compounds were demonstrated for the HPLC/MS method in buffered reagent water, surface water, and ground water. The extraction method as used did not optimize the recovery of flufenacet OXA. Method detection limits (MDL's) for the HPLC/MS method ranged from 0.009 to 0.045 µg/L, with the flufenacet OXA MDL at 0.072 µg/L. The mean HPLC/MS recoveries of degradation compounds from water samples spiked at 0.2 and 1.0 µg/L ranged from 75 to 114 percent, with relative standard deviations of 15.8 percent or less for all compounds except flufenacet OXA which had relative standard deviations ranging from 11.3 to 48.9 percent. The MDL for the HPLC/MS method was established at 0.05 µg/L.
Information about the fate and transport of the chloroacetanilide herbicides, acetochlor, alachlor, dimethenamid, flufenacet, and metolachlor, and their degradation compounds in water can be acquired from the analysis of surface water and ground water using the HPLC/MS method. This method also can be useful for water-quality determinations and analytical verification in toxicological studies.
APPENDIX 1. AUTOMATED SOLID-PHASE EXTRACTION PROCEDURE USING AUTOTRACE WORKSTATION MEAN CONCENTRATIONS AND METHOD DETECTION LIMITS FOR EIGHT DETERMINATIONS
[mL, milliters; mL/min, milliliters per minute; AutoTrace extraction procedure JK.123.MEOH]
Estimated time for samples : 49.1 minutes
Date : December 12, 1999
Step 1 : Process six samples using the following procedure:
Step 2 : Condition column with 3 mL methanol into SOLVENT WASTE
Step 3 : Condition column with 3 mL ethyl acetate into SOLVENT WASTE
Step 4 : Condition column with 3 mL methanol into SOLVENT WASTE
Step 5 : Condition column with 3 mL distilled water into AQUEOUS WASTE
Step 6 : Wash syringe with 5 mL ethyl acetate
Step 7 : Load 123 mL of sample onto column
Step 8 : Dry column with gas for 0.5 minute
Step 9 : Condition column with 3.2 mL ethyl acetate into SOLVENT WASTE
Step 10 : Collect 3.5-mL fraction into sample tube using methanol
Step 11 : Dry column with gas for 3 minutes
Step 12 : END
Setup Parameters | ||||
---|---|---|---|---|
FLOW RATES (mL/min) |
SOLID-PHASE EXTRACTION PARAMETERS |
|||
Condition flow: | 10.0 | Push delay: | 5 seconds | |
Load flow: | 10.0 | Air factory: | 1.0 | |
Rinse flow: | 20.0 | Autowash volume: | 1.0 mL | |
Elute flow: | 5.0 | |||
Condition air push: | 15.0 | WORKSTATION PARAMETERS | ||
Rinse air push: | 20.0 | Maximum elution volume: | 12.0 mL | |
Elute air push: | 5.0 | Exhaust fan on: | Yes | |
Beeper on: | Yes | |||
Name solvents | ||||
Solvent 1 : Ethyl acetate | ||||
Solvent 2 : Methanol | ||||
Solvent 3 : Distilled water | ||||
Solvent 4 : Not used | ||||
Solvent 5 : Not used |