Program 7386.001
CHAPTER 86 MEDICAL AND RADIOLOGICAL DEVICE MONITORING AND QUALITY CONFORMANCE
Inspection and Field Testing of Radiation-Emitting Electronic Products |
IMPLEMENTATION DATE 10/31/2007 |
|
COMPLETION DATE 9/30/2010 |
||
DATA REPORTING |
||
PRODUCT CODES | PRODUCT/ASSIGNMENT CODES | |
---|---|---|
95RH-XXX (See Attachment B for detail) | 86001 | |
95RH-XXX (See Attachment C for detail) | 86002 | |
94RH-XXX (See Attachment D for detail) | 86004 |
This compliance program consolidates and supersedes the following compliance programs:
Center for Devices and Radiological Health
Office of Communication, Education and Radiation Programs
ATTN: Electronic Products Branch (HFZ-240)
1350 Piccard Dr .
Rockville MD 20850 .
This document represents the agency’s current thinking on the enforcement of the Federal Food Drug and Cosmetic Act Electronic Product Radiation Control provisions and related regulations. It does not create or confer any rights for or on any person and does not operate to bind FDA or the public. An alternative approach may be used if such approach satisfies the requirements of the applicable statute, regulations, or both.
This compliance program provides guidance to FDA field and center staff for the inspection, field test and administrative/enforcement activities related to the Electronic Product Radiation Control (EPRC) provisions of the Federal Food Drug and Cosmetic Act (FFDCA, the Act) and regulations contained in Title 21 of the Code of Federal Regulations, Parts 1000 – 1050 (21 CFR 1000 – 1050). The intent of these requirements is to protect the public from unnecessary exposure to electronic products radiation. Manufacturers are responsible for producing products that do not emit hazardous or unnecessary radiation and that comply with all applicable radiation safety performance standards. All electronic product manufacturers must comply with applicable requirements in Title 21 CFR 1000, 1002, 1003, 1004 and 1005. If a mandatory radiation safety performance standard applies to a manufacturer’s product, then the manufacturer must also comply with Title 21 CFR 1010 and the product must comply with the requirements of the specific standard found in 21 CFR 1020 – 1050. Manufacturers are required to self-certify their own products to be compliant with an applicable standard, based on a quality control testing program as described in 21 CFR 1010.2. The purpose of EPRC inspections and field tests are to verify that products comply with performance standards, and that the manufacturer’s quality control testing program ensures such product compliance and radiation safety.
This program applies to certain electronic products subject to radiation safety performance standards described in 21 CFR 1010 – 1040, including:
Diagnostic x-ray inspection and testing is conducted under Compliance Program 7386.001. Products and manufacturers subject to standards contained in 21 CFR 1020 – 1050, but are not listed above, will be subject to inspection or test on a for-cause basis only at the direction of CDRH.
The body of this program contains basic instructions for inspection, field test and administrative/enforcement activities applicable to all electronic products. Inspection and field test checklists, and additional considerations and instructions for specific products, such as laser, sunlamp, cabinet x-ray, television and microwave oven products, are covered in ATTACHMENTS B - F.
Medical devices that emit electronic product radiation are subject to EPRC requirements as well as Medical Device provisions of the Act and related regulations. Medical device inspection and enforcement activities described in Compliance Program 7382.845, Inspection of Medical Device Manufacturers, may be conducted jointly with this program at CDRH and district discretion. Examples of radiation-emitting medical devices include medical laser and sunlamp products, which could be covered by a joint EPRC/medical device inspection.
This is a continuing, non-statistical compliance program intended:
Pre-announcement of EPRC inspections conducted under this compliance program is not mandatory, although it is recommended to facilitate the inspection. Pre-announcement ensures the firm is producing electronic products for the US market on the day of inspection, gives the firm time to collect all necessary procedures and records, and ensures appropriate individuals are available during the inspection. Section 537 of the Act permits inspection of any manufacturer for good cause, grounds for which may include introduction of any noncompliant product into US commerce, failure to comply with EPRC reporting requirements, or for purposes of suspected problems with a manufacturer’s quality control testing program and product conformance with performance standards.
Inspections of radiation-emitting medical device manufacturers must be pre-announced if the inspection will cover medical device Quality Systems Regulation compliance. Refer to instructions provided in the Guide to Inspections of Quality Systems, August 1999, and IOM Section 5.2.1.1, Pre-Announcements.
Schedule an appointment with the user prior to the field test. Tell the user that the purpose of the visit is to conduct a survey of an electronic product to determine compliance with FDA’s Federal radiation safety performance standards.
Request that persons familiar with the operation of the electronic product to be tested be available to assist in the operation of the equipment.
Inspections and field testing of electronic product manufacturers should be prioritized using the following criteria:
The purpose of electronic product manufacturer inspections is to evaluate the firm’s quality control testing program to ensure product compliance with applicable performance standards and radiation safety. The inspection should also verify that EPRC requirements for reporting and recordkeeping are met by the firm.
Specific product inspection and field test checklists or forms, if available, are included in ATTACHMENTS B – F. These checklists should be used in conjunction with the above guidance to record inspection and test observations.
“This document lists observations made by the FDA representative(s) during the inspection of your facility. They are inspectional observations and do not represent a final Agency determination regarding your compliance. If you have an objection regarding an observation, or have implemented, or plan to implement, corrective actions in response to an observation, you may discuss the objection or action with FDA representative(s) during the inspection or submit this information to FDA at the address above. If you have any questions, please contact FDA at the phone number and address above."
For all medical device inspections the FDA 483 should contain the following additional statement:
“The observations noted in this form FDA 483 are not an exhaustive listing of objectionable conditions. Under the law, your firm is responsible for conducting internal self audits to identify and correct any and all violations of the quality system requirements.”
Field tests are examinations of installed electronic products and may be conducted at trade shows, manufacturing facilities or other sites where products are in use. Field tests assess the individual product’s compliance with applicable performance standard requirements alone. It can not be expected that there will be staff on site with expert knowledge of the product being field tested or that it will be possible to evaluate all aspects of product compliance.
Investigations are to be made to determine whether a suspected firm is in fact a manufacturer of one or more electronic products. The investigation may be initiated in preparation for a possible inspection, as a result of trade complaints, or from discovery via the Internet or printed materials of promotion of products that may not comply with EPRC requirements.
Physical samples of products are generally not collected under this compliance program. Samples are not required to support a letter issued to the firm or further action to include program disapproval or legal action. However, samples can be useful to support inspectional observations to demonstrate inadequacy of the quality control testing program or product noncompliance. The investigator should consult district management and CDRH to determine whether collecting physical samples would support any subsequent letter or action initiated. Documentary samples may be collected when collecting an actual physical sample is not practical and the evidence is necessary to support inspectional observations.
Collect samples according to procedures defined in the Investigations Operations Manual, Chapter 4, and coordinate any sample collection activity with CDRH and WEAC to ensure proper procedures are followed and chain of custody is observed to maintain sample integrity.
No laboratory testing will be done under this program. CDRH or WEAC testing may be required on special assignments under Compliance Program for Lab Testing CP 7386.006 or as indicated in Part III.A.5 of this program.
CDRH is generally responsible for the final review of inspections and field tests made under this program and for the issuance of letters resulting from inspections and field tests performed by field radiological health staff. Exceptions where the district has direct reference authority are noted below under section C, Regulatory Action. The intent of this program is to follow up on problems that pose a radiation safety hazard or are a flagrant violation of EPRC requirements.
Violations of EPRC requirements may include:
Appropriate regulatory or administrative actions include issuance of a notification of defect or non-compliance letter (warning or untitled letter), requiring repurchase, repair or replacement of product under an approved corrective action plan, or imposition of civil penalties and/or injunction. Appropriate follow-up actions should be determined by CDRH or in consultation with CDRH to ensure consistency in how EPRC requirements are enforced.
CDRH has classified several potential items of non-compliance that might be observed during an inspection or field test and classified those items in terms of health hazard and regulatory action. Tables are provided in Attachments A – F to provide guidance for use during the inspection or field test, while preparing FDA-483 and EIR or field test reports, and in classifying the inspection or field test and recommending follow-up.
A table of violations and their health and safety risk as well as the nature of the regulatory response has been provided for each product area.
The inspection may be classified VAI for a limited number of minor deficiencies listed in Attachments A – F and further regulatory action will be pursued at the discretion of the district and CDRH.
If it is determined that the EPRC deficiencies are of a quantity and type to conclude there is minimal probability that the firm will produce unsafe or noncompliant products, the inspection will be classified NAI and Form FDA-483 will serve to inform the firm of any objectionable findings. Deficiencies identified as violations of concern will generally not require additional follow-up but should be discussed with the firm.
Consult CDRH if additional guidance is required to classify inspection and test observations. If the inspection also covered firm compliance with medical device Quality Systems requirements, Compliance Program 7382.845, Part V, Quality System/GMP Regulatory/Administrative Follow-Up, should be consulted for appropriate regulatory and administrative follow-up.
In determining appropriate regulatory action based on inspection and test findings, the district and CDRH should consider the significance of the product, the firm’s history, whether the problem is widespread and continuing. Actions which may be considered include notification of noncompliance letters (warning and untitled letters), product repurchase, repair or replacement (recall), civil penalties and injunctions, and seizures (for radiation-emitting medical devices).
The Electronic Product Radiation Control provisions of the Federal Food Drug and Cosmetic Act (Section 535) and related regulations (21 CFR 1003) require the Agency to notify manufacturers in writing when product noncompliance with a standard is found. Manufacturers may also be advised in writing of a failure to comply with reporting and recordkeeping requirements (21 CFR 1002.31). A table classifying the severity of items of noncompliance with reporting and recordkeeping, and performance standard requirements is included in Attachments A – F.
Issuance of all letters should follow Chapter 4 of the Regulatory Procedures Manual (RPM) http://www.fda.gov/ora/compliance_ref/rpm/. Consult the Office of Enforcement’s (OE) Warning Letter page on ORA’s intranet for current instructions for obtaining Office of Chief Counsel (OCC) clearance and for current approved Warning Letter templates. Letter templates must be used to satisfy Agency notification requirements in 21 CFR 1003.11. Where approved OCC templates are not available, consult CDRH for the current version of letter templates.
Districts have DIRECT REFERENCE AUTHORITY for EPRC letters in certain areas which are described in Chapter 4 of the RPM. For example, districts have direct reference authority to issue sunlamp product warning and untitled letters, to grant exemption from notification, and approve manufacturer corrective action plans, which are further described in Attachment C of this program. CDRH is available for consult in assessing product noncompliance or developing regulatory and enforcement strategy.
For the majority of cases, where districts DO NOT have direct reference authority to issue EPRC letters, forward the report with exhibits and recommended action to CDRH for review and follow-up. CDRH will copy the accomplishing district on any letters issued and consult on regulatory and enforcement strategy when needed.
Issue a minor notification (untitled) letter when the violation of the standard does not justify further regulatory action at the time.
The Electronic Product Radiation Control provisions of the Federal Food Drug and Cosmetic Act (Section 535) and related regulations (21 CFR 1004) also provide for manufacturer repurchase, repair or replacement of the noncompliant electronic products.
Every major notification of noncompliance letter issued as a result of a major violation or several minor violations requires manufacturer repurchase, repair or replacement of the affected electronic products at no cost to the purchaser. The firm is required to address all items in the letter, and submit a corrective action plan for CDRH approval. Refer to RPM Chapter 7, Attachment E for approval of manufacturer's corrective action plans.
Manufacturers can refute the noncompliance or be granted an exemption, by making a written request to CDRH. The exemption can be granted upon request by the manufacturer or by the Agency at its own initiative, and must show that the noncompliance does not create a significant risk of injury.
Within 15 days after notification of the noncompliance/defect by FDA, a manufacturer may refute the alleged noncompliance under 21 CFR 1003.11(a)(3) or request an exemption from purchaser notification and correction as specified under 21 CFR 1003.30. If a manufacturer refutes the alleged noncompliance, or requests an exemption, the evidence presented by the manufacturer is evaluated by CDRH before granting or denying the request for exemption or responding to the refutation. Refer to RPM Chapter 7, Attachment E for information on responding to exemption requests and refutations.
Immediately notify CDRH and State and local health authorities (through RRHR) for any Class A hazard.
For all inspections and tests that may require issuance of a letter, the EIR should be provided to CDRH or the district compliance officer to allow sufficient time to review, draft, and secure approval for the letter. Timeframes for clearance of letters are provided in Chapter 4 of the RPM.
Civil penalties should be recommended for violations of Subchapter C of the Act after other actions have failed to achieve compliance, or for knowing and willful violations. More severe civil penalty assessments may be sought under Section 303(f). See CPG Sec. 390.300 and RPM Chapter 6, Civil Penalties - Electronic Product Radiation Control. Informal consultation with the Center at an early stage in the development of a regulatory action is encouraged in order to facilitate timely implementation of the action; contact Electronic Products Branch Chief or Lead CSO at (240) 276-3332.
If an establishment has a continuing pattern of significant deviations in spite of past warnings, injunction will usually be the recommended action of choice. If a serious health hazard exists, the recommendation should include a request for a temporary restraining order (TRO) to prevent the distribution of products that have been manufactured under the violative conditions documented by the inspection report per the instructions in Chapter 6 of the RPM. Civil penalties and injunctions may be recommended concurrently.
Use administrative detention and recommend seizure of a defective or noncompliant radiation-emitting medical device if all three conditions below apply:
Informal consultation with the Center at an early stage in the development of a regulatory action is encouraged; contact Electronic Products Branch Chief or Lead CSO at (240) 276-3332.
Some states have Radiation Control Programs within the State Health Department or Department of Environmental Health, which may have adopted portions of the EPRC requirements into their radiation safety regulations.
Districts should use all reasonable means available to encourage voluntary conformance of products with the performance standard regardless of the date of manufacture. It is recommended that the districts coordinate regulatory activity with appropriate state representatives through the RRHR and DFSR, particularly where local authority may assist in achieving correction of a deficiency. This may be particularly useful to address issues related to product use where the State may have regulatory authority, which extends beyond FDA authority to regulate the design, production or manufacture of the product.
Regulatory follow-up for joint EPRC/quality systems inspections can be handled separately or in combination at the discretion of the district and CDRH. Refer to Part V in Compliance Program 7382.845, Quality System/GMP Regulatory/Administrative Follow-Up, for guidance on regulatory actions related to radiation-emitting medical devices. Enforcement actions on radiation emitting medical device firms, which also include EPRC violations, require CDRH concurrence before implementation by the field. Contact CDRH for consultation when both EPRC and quality systems violations are noted during an inspection or field test.
Federal Food, Drug, and Cosmetic Act, As Amended
Electronic Product Radiation Control Provisions (formerly known as the Radiation Control for Health and Safety Act of 1968, Public Law 90-602, 0ctober 18, 1968)
http://www.fda.gov/opacom/laws/fdcact/fdctoc.htm
21 CFR 1000 – 1005, General Requirements for All Electronic Products which Emit Radiation
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfcfr/CFRSearch.cfm?CFRPartFrom=1000&CFRPartTo=1005
21 CFR 1010, Performance Standards for Electronic Products: General
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfcfr/CFRSearch.cfm?CFRPart=1010
21 CFR 1020 – 1050, Specific Performance Standards for Electronic Products
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfcfr/CFRSearch.cfm?CFRPartFrom=1020&CFRPartTo=1050
http://www.fda.gov/ora/compliance_ref/rpm/default.htm
http://www.fda.gov/ora/inspect_ref/iom/default.htm
FDA home page
http://www.fda.gov
ORA home page
http://www.fda.gov/ora/
CDRH home page
http://www.fda.gov/cdrh/
Field Accomplishments and Compliance Tracking System (FACTS)
(visit ORA’s home page, then click the FACTS icon.)
Electronic Product Radiation Control home page
http://www.fda.gov/cdrh/radhealth
Product Code Classification Database (searchable)
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpcd/pcdsimplesearch.cfm
Good Guidance Practices Database (searchable)
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfggp/search.cfm
Office of Communication, Education and Radiation Programs, Division of Mammography Quality and Radiation Programs (DMQRP)
Contact for support in planning and executing inspections and field tests, classification of items of non-compliance, and for interpretation and current policy on EPRC requirements. Send all inspection and test reports to Chief, Electronic Products Branch, FDA/CDRH Office of Communication, Education and Radiation Programs (HFZ-240), 1350 Piccard Drive, Rockville, MD 20850.
Name | Phone | Mail Stop | Position/Expertise | |
---|---|---|---|---|
Sean Boyd | (240) 276–3287 | sean.boyd@fda.hhs.gov | HFZ-240 | Chief, Diagnostic Devices Branch |
Dr. Helen Barr | (240) 276-3275 | helen.barr@fda.hhs.gov | HFZ-240 | Director, DMQRP |
Thomas Ohlhaber | (240) 276-3274 | thomas.ohlhaber@fda.hhs.gov | HFZ-240 | Deputy Director, DMQRP |
Jerome Dennis | (240) 276-3330 | jerome.dennis@fda.hhs.gov | HFZ-240 | Consumer Safety Officer, laser expert |
Daniel Hewett | (240) 276-3268 | daniel.hewett@fda.hhs.gov | HFZ-240 | Consumer Safety Officer, laser products |
Dan Kassiday | (240) 276-3280 | daniel.kassiday@fda.hhs.gov | HFZ-240 | Engineer, cabinet, industrial, analytical, security x-ray products |
George Kraus | (240) 276-3298 | george.kraus@fda.hhs.gov | HFZ-240 | Consumer Safety Officer, TV and microwave oven products, imports |
Frank Mackison | (240) 276-3264 | frank.mackison@fda.hhs.gov | HFZ-240 | Consumer Safety Officer, laser products |
L. Dale Smith | (240) 276-3295 | l.smith@fda.hhs.gov | HFZ-240 | Consumer Safety Officer, laser light show products |
Cory Tylka | (240) 276-3257 | corinne.tylka@fda.hhs.gov | HFZ-240 | Consumer Safety Officer, medical laser products |
Varsha Savalia | (240) 276-3324 | varsha.savalia@fda.hhs.gov | HFZ-240 | Consumer Safety Officer, sunlamp and UV and laser products |
Office of Science and Engineering Laboratories
Contact for assistance with identifying appropriate instrumentation for use in measuring electronic product radiation emissions.
Name | Phone | Mail Stop | Position | |
---|---|---|---|---|
Robert Faaland | (301) 827-4683 | robert.faaland@fda.hhs.gov | HFZ-130 | Engineer, laser expert |
Sharon Miller | (301) 827-4692 | sharona.miller@fda.hhs.gov | HFZ-130 | Engineer, UV expert |
Mary Walker | (301) 796-2558 | mary.walker@fda.hhs.gov | TBD | X-ray instrumentation and calibration |
Field Regional Radiological Health Representatives
Name |
Phone |
Mail Stop |
Position |
|
---|---|---|---|---|
Mike Leal |
(508) 793-0422 |
HFR-NE2570 |
NE RRHR |
|
Tom Trout |
(615) 366-7817 |
HFR-SE300 |
SE RRHR |
|
Rachel Evans |
(312) 596-6518 |
HFR-CE25 |
CE RRHR |
|
Scotty Hargrave |
(214) 253-4930 |
HFR-SW19 |
SW RRHR |
|
Suzie Kent |
(208) 334-1054 |
HFR-PA3505 |
PA RRHR |
Field Electro-Optics Specialists and laser product contacts
Name | Phone | Mail Stop | Position | |
---|---|---|---|---|
Emir Galevi | (781) 729-5700 x724 | emir.galevi@fda.hhs.gov | HFR-NE480 | Engineer, WEAC |
Leo Lagrotte | (813) 228-2671 x35 | leo.lagrotte@fda.hhs.gov | HFR-SE2585 | Electro-Optics Specialist, SER |
James E. Frye | (513) 684-2700 x149 | james.frye@fda.hhs.gov | HFR-CE400 | Electro-Optics Specialist, CER |
Don Leeseberg | (210) 541-9450 | don.leeseberg@fda.hhs.gov | HFR-SW1545 | Consumer Safety Officer, SWR |
Gary L. Zahaek | (408) 291-7548 x103 | gary.zaharek@fda.hhs.gov | HFR-PA1530 | Electro-Optics Specialist, PAR |
Frank J. Eng | (408) 291-7548 x105 | frank.eng@fda.hhs.gov | HFR-PA1530 | Electro-Optics Specialist, PAR |
Winchester Engineering and Analytical Center contacts
Name | Phone | Mail Stop | Position | |
---|---|---|---|---|
John Marzilli | (781) 729-5700 x749 | john.marzilli@fda.hhs.gov | HFR-NE400 | WEAC Director |
Jim Cherniak | (781) 729-5700 | james.cherniack@fda.hhs.gov | HFR-NE400 | Radiation Safety Officer |
Vacant | Vacant | Vacant | HFR-NE400 | Engineering Branch Director, WEAC |
Jane Driscoll | (781) 729-5700 x716 | jane.driscoll@fda.hhs.gov | HFR-NE480 | Metrology Supervisor |
Joe Matrisciano | (781) 729-5700 x736 | joseph.matrisciano@fda.hhs.gov | HFR-NE480 | Engineering Supervisor |
Headquarters contacts
Name | Phone | Mail Stop | Position | |
---|---|---|---|---|
April Kidd | (301) 827-2913 | april.kidd@fda.hhs.gov | HFC-150 | ORO, DFSR |
The following items are common to all EPRC inspections and field tests, and may be cited for any product subject to the below reporting or certification requirements. Products subject to reporting are listed in Table 1 of 1002.1, and certification requirements are applicable to all products subject to a performance standard.
Reporting requirements | |||
1002.1 | No product report | Minor | Class B |
1002.11 | No supplemental report | Minor | Class B |
1002.13 | No annual report | Minor | Class B |
1002.2 | No accidental radiation occurrence report | Minor | Class B |
Certification requirements | |||
1010.2 | No certification label | Minor | Class B |
1010.2 | Inadequate or lack of testing program | Major | Class A |
1010.2 | Incomplete testing---program exists but lacks record | Minor | Class B |
1010.2 | Incomplete testing with minor deficiencies | Concern | Class C |
1010.2 | Reference to DHEW or BRH | Concern | Class C |
1010.3 | No identification label | Concern | Class C |
1010.3 | Coded or abbreviated date | Minor | Class B |
1010.3 | Month & year in serial number on non-consumer product | Concern | Class C |
1010.3 | No manufacturer address | Concern | Class C |
1010.3 | Incomplete address | Concern | Class C |
The Laser Products Performance Standard (the standard), promulgated in August 1976, was designed to protect the public from unnecessary radiation hazards associated with the use of these products. The radiation emitted from these laser products can pose varying degrees of hazards depending upon the type, magnitude, and accessibility of the radiation and upon the particular functions or operations they perform. The standard was last amended in 1985. Since then, the CDRH has intended to harmonize the requirements of the standard with those of the international standard IEC 60825-1: 2001. As an interim step the CDRH published its Laser Notice 50 in 2001 stating that it would not object to compliance with specified requirements of the international standards in lieu of comparable requirements of the CDRH standard.
High-risk laser products and their manufacturers should be inspected or tested as a priority. Examples of high-risk laser products and manufacturers include:
Electro-optics specialists have been specifically trained in general EPRC requirements and also have specialized training in the laser product performance standards. EOS’s should perform these inspections and field tests, and may train additional field staff or accompany a medical device investigator to conduct joint EPRC/medical device inspections. If an EOS has training in both EPRC and QSIT inspections, a single EOS may conduct both portions of the inspection.
CDRH is responsible for review of laser manufacturer inspection and product field test observations and initiating administrative or regulatory follow-up.
Frequently Asked Questions about Lasers.
http://www.fda.gov/cdrh/radhealth/products/laserfaq.html
Performance Standard-Lasers and Products Incorporating Lasers
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfcfr/CFRSearch.cfm?FR=1040.10
Performance Standard-Specific Laser Products (Includes Display, Survey, and Medical Laser Products)
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfcfr/CFRSearch.cfm?FR=1040.11
Laser Compliance Guide
http://www.fda.gov/cdrh/radhlth/pdf/lasgde01.pdf
Reporting Guide-Radiation Safety Product Report for Laser Products
http://www.fda.gov/cdrh/radhlth/pdf/lasrpt0p.pdf
Reporting Guide-Radiation Safety Product Report for Laser Light Shows/Displays
http://www.fda.gov/cdrh/radhlth/pdf/llsrpt01.pdf
Laser Quality Control Guide
http://www.fda.gov/cdrh/radhlth/pdf/lasgdeqc.pdf
Refer to the laser products main page for guidance documents and additional information:
http://www.fda.gov/cdrh/radhealth/products/lasers.html.
Translation of 2-Digit Code | Product Name | Product Code | CFR | Definition | |
---|---|---|---|---|---|
Other Laser Products | Automotive Accessory, Automobile or Transport Vehicle, Laser | 95 | RDV | 1040.10 | A laser product or product containing a laser that is an automotive or other transport vehicle accessory. |
Other Laser Products | General Purpose Laser Products, Non-Medical | 95 | RDW | 1040.10 | Product, laser, general, emit beam; A laser or product containing a laser that is intended for general purpose use with no medical claims. |
Other Laser Products | Other | 95 | RZZ | Unk | A laser or product containing a laser for which its intended use is not previously defined. |
Laser Light Show/Display Products | Low-Power Laser Light Show Projector | 95 | RDZ | 1040.10; 1040.11 |
Product, laser, demo, projector, laser light show/display, Class IIIa/3R and lower; Laser projection system that incorporates a laser having a maximum radiation output of less than 5mw. |
Laser Light Show/Display Products | High-Power Laser Light Show Projector (Output > 5mW) | 95 | REA | 1040.10; 1040.11 |
Product, laser, demo, projector, laser light show/display, Class IIIb/IV/3B/4; Laser projection system that incorporates a laser having a maximum radiation output power greater than 5mw. |
Laser Light Show/Display Products | High-Power Laser Light Show | 95 | REB | 1040.10; 1040.11 |
Product, laser, demo, laser light show/display, Class IIIb/IV/3B/4; Laser light show or demonstration using laser projection equipment having an output that exceeds 5mW |
Laser Light Show/Display Products | Laser Video Projector | 95 | REC | 1040.10; 1040.11 |
Product, laser, demo, projector, display, video; A laser used in conjunction or incorporated in a video display system or projector. |
Laser Light Show/Display Products | Laser Advertising Display System | 95 | RED | 1040.10; 1040.11 |
Product, laser, demo, system, display, advertising |
Laser Light Show/Display Products | Laser Visual Display - Display Retinal Image, Non-Medical Display Product | 95 | REE | 1040.10; 1040.11 |
Product, laser, display, system, images, direct to retina |
Laser Light Show/Display Products | Other | 95 | RZZ | Unk | Other laser products used in light shows or demonstrations that are not otherwise defined. |
Medical Laser Products | Laser, Ophthalmic | 86 | HQF | 1040.10; 1040.11 |
|
Medical Laser Products | Laser Instrument, Surgical, Powered | 79 | GEX | 1040.10; 1040.11 |
|
Medical Laser Products | Laser, Surgical, Gynecologic | 85 | HHR | 1040.10; 1040.11 |
|
Medical Laser Products | Laser, ENT Microsurgical Carbon-Dioxide | 77 | EWG | 1040.10; 1040.11 |
|
Medical Laser Products | Photocoagulator and Accessories | 86 | HQB | 1040.10; 1040.11 |
|
Medical Laser Products | Lens, Surgical, Laser, Accessory, Ophthalmic Laser | 86 | LQJ | 1040.10; 1040.11 |
|
Medical Laser Products | Laser, Neurosurgical | 84 | LKW | 1040.10; 1040.11 |
|
Medical Laser Products | Laser, Neurosurgical, Argon | 84 | LLF | 1040.10; 1040.11 |
|
Medical Laser Products | Laser, Neodymium: YAG, Pulmonary Surgery | 73 | LLO | 1040.10; 1040.11 |
|
Medical Laser Products | Laser, Neodymium: YAG, for Gynecologic Use | 85 | LLW | 1040.10; 1040.11 |
|
Medical Laser Products | Laser, Neodymium: YAG, Ophthalmic for Uses Other than Posterior Capsulotomy and Cutting Pupil | 86 | LXS | 1040.10; 1040.11 |
|
Medical Laser Products | Laser, Neodymium: YAG, Ophthalmic for Uses Other than Posterior Capsulotomy | 86 | LOI | 1040.10; 1040.11 |
|
Medical Laser Products | Laser, Neodymium: YAG, Optical, Pumped Parametric Oscillator | 86 | MVQ | 1040.10; 1040.11 |
|
Medical Laser Products | Laser, Microsurgical Argon, for Use in Otology | 77 | LXR | 1040.10; 1040.11 |
|
Medical Laser Products | Laser, Microsurgical Argon, for Uses Other Than Otology | 77 | LMS | 1040.10; 1040.11 |
|
Medical Laser Products | Laser for Gastro-Urology Use | 78 | LNK | 1040.10; 1040.11 |
|
Medical Laser Products | Device, Angioplasty, Laser, Coronary | 74 | LPC | 1040.10; 1040.11 |
|
Medical Laser Products | Device, Laser Peripheral Angioplasty | 74 | LWX | 1040.10; 1040.11 |
|
Medical Laser Products | Catheter, Coronary Laser Myoplasty | 74 | MGC | 1040.10; 1040.11 |
|
Medical Laser Products | System, Laser, Transmyocardial Revascularization | 74 | MNO | 1040.10; 1040.11 |
|
Medical Laser Products | Instrument, Visual Field, Laser | 86 | HPJ | 1040.10; 1040.11 |
|
Medical Laser Products | Laser for Pain Therapy | 84 | LLP | 1040.10; 1040.11 |
|
Medical Laser Products | Laser, System, Excimer | 86 | LZS | 1040.10; 1040.11 |
|
Medical Laser Products | Laser, Dental | 76 | LYB | 1040.10; 1040.11 |
|
Medical Laser Products | Photodynamic Therapy (PDT) | 79 | MVF | 1040.10; 1040.11 |
|
Medical Laser Products | Photodynamic Therapy (PDT), Fiber Optic | 79 | MVG | 1040.10; 1040.11 |
|
Medical Laser Products | Laser, Fluorescence Caries Detection | 76 | NBL | 1040.10; 1040.11 |
|
Medical Laser Products | Laser for Wound Healing | 79 | LXU | 1040.10; 1040.11 |
|
Medical Laser Products | Ophthalmoscope, Laser Scanner | 86 | MYC | 1040.10; 1040.11 |
|
Medical Laser Products | Laser, Phacolysis | 86 | MXO | 1040.10; 1040.11 |
|
Medical Laser Products | Caries Detector, Laser, Light, Transmission | 76 | NTK | 1040.10; 1040.11 |
|
Medical Laser Products | Other | 95 | RZZ | Unk | A laser or laser product intended for medical treatment or other uses on humans, not previously defined. |
Other Demonstration Laser Products | Laser Science Education Products | 95 | REI | 1040.10; 1040.11 |
Product, laser, demo, education, illustrate science principles |
Other Demonstration Laser Products | Other | 95 | RZZ | Unk | Laser products used for demonstrations that are not otherwise defined. |
Toy, Novelty, Play Laser Products | Toy, Novelty, Play Laser Product | 95 | REJ | 1040.10; 1040.11 |
Product, laser, toy/novelty |
Research, Scientific, Laboratory Laser Products | Research Laser, Scientific, Laboratory Laser Products | 95 | REK | 1040.10 | Product, laser, research/laboratory; A laser under development in and of itself. A laser used for conducting research during development of new data or to improve a process would not be considered a research laser although it is being used in research. |
Research, Scientific, Laboratory Laser Products | Guide-Star Laser System, Research, Scientific, Laboratory Laser Products | 95 | REL | 1040.10 | Product, laser, adaptive-optics telescope focusing accessory, generate artificial star; A laser used for alignment of optical telescopes. |
Research, Scientific, Laboratory Laser Products | Spectroscopy Instrument, Laser, Research, Scientific, Laboratory Laser Products | 95 | REM | 1040.10 | Product, laser, instrument, spectroscopy; An instrument incorporating a laser for spectroscopic testing or examination with no medical claims. |
Research, Scientific, Laboratory Laser Products | Particle-Size Measuring Instrument, Laser, Scientific, Laboratory Laser Products | 95 | REN | 1040.10 | Product, laser, instrument, particle size measurement; An instrument or system incorporating a laser for determining the size or number of particles of particles a test sample. |
Research, Scientific, Laboratory Laser Products | Analytical Measuring and Detection, Research, Scientific, Laboratory Laser Products | 95 | REO | 1040.10 | Product, laser, instrument, analyze/detect chemical species |
Research, Scientific, Laboratory Laser Products | Other | 95 | RZZ | Unk | Laser products used in scientific and laboratory applications that are not otherwise defined. |
Surveying, Leveling, Alignment Laser Products | Surveying Laser Product, Leveling, Alignment Laser Products | 95 | REP | 1040.10 | Product, laser, surveying, instrument, determine position by measurement of angles |
Surveying, Leveling, Alignment Laser Products | Ranging (Geodimeter) Laser Products | 95 | REQ | 1040.10 | Product, laser, ranging, instrument, measure distance by time-of-flight |
Surveying, Leveling, Alignment Laser Products | Alignment Laser Product, Surveying, Leveling, Alignment Laser Products | 95 | RER | 1040.10; 1040.11 |
Product, laser, alignment, aid positioning or adjusting parts in relation to each other |
Surveying, Leveling, Alignment Laser Products | Laser Pointer, Surveying, Leveling, Alignment Laser Products | 95 | RES | 1040.10; 1040.11 |
Product, laser, pointer, indicate point of interest; A laser product intended specifically to define a spot or surface for drawing attention to a viewer. |
Surveying, Leveling, Alignment Laser Products | Laser Target Designator, Surveying, Leveling, Alignment Laser Products | 95 | RET | 1040.10 | Product, laser, target designator; An optical devices, using a visible beam of laser light that permits the alignment of a gun, cannon or rocket system with its target. |
Surveying, Leveling, Alignment Laser Products | Laser Aiming Product, Visible, Surveying, Leveling, Alignment Laser Products | 95 | REU | 1040.10; 1040.11 |
Product, laser, aiming, visible, attached to weapon; An optical devices, using a visible beam of laser light that permits the alignment of a gun, cannon or rocket system with its target |
Surveying, Leveling, Alignment Laser Products | Laser Aiming Product, Non-Visible, Surveying, Leveling, Alignment Laser Products | 95 | REV | 1040.10; 1040.11 |
Product, laser, aiming, infrared, attached to weapon, viewed with night-vision equipment; An optical devices, using an invisible beam of laser light that permits the alignment of a gun, cannon or rocket system with its target. |
Surveying, Leveling, Alignment Laser Products | Other | 95 | RZZ | Unk | Other laser products used for surveying, leveling and alignment that are not otherwise defined. |
Safety, Security, Surveillance Laser Products | IR Laser Illuminator with Alignment Aid/Night Vision System, Safety, Security, Surveillance Laser Products | 95 | REW | 1040.10; 1040.11 |
Product, laser, infrared, illuminator with alignment aid, viewed through night-vision equipment |
Safety, Security, Surveillance Laser Products | IR Laser Illuminator Only/Night Vision System, Safety, Security, Surveillance Laser Products | 95 | REX | 1040.10 | Product, laser, infrared, illuminator only, viewed through night-vision equipment |
Safety, Security, Surveillance Laser Products | Collision-Avoidance Laser System, Safety, Security, Surveillance Laser Products | 95 | REY | 1040.10 | Product, laser, infrared, collision-avoidance system |
Safety, Security, Surveillance Laser Products | Laser Traffic Signal, Safety, Security, Surveillance Laser Products | 95 | REZ | 1040.10 | Product, laser, traffic signal/control |
Safety, Security, Surveillance Laser Products | Laser Automotive Lighting & Signals, Safety, Security, Surveillance Laser Products | 95 | RFA | 1040.10 | Product, laser, automotive, lighting/signals |
Safety, Security, Surveillance Laser Products | IR Laser Intrusion Detection/Security System, Safety, Security, Surveillance Laser Products | 95 | RFB | 1040.10 | Product, laser, infrared, intrusion detecting, security system |
Safety, Security, Surveillance Laser Products | Laser Radar (Lidar) or Speed Measurement, Safety, Security, Surveillance Laser Products | 95 | RFC | 1040.10 | Product, laser, infrared, Doppler or time-of-flight speed measurement |
Safety, Security, Surveillance Laser Products | Other | 95 | RZZ | Unk | Laser products used in safety, security, surveillance applications not otherwise defined |
Safety, Security, Surveillance Laser Products | Laser Weapon (Military or Police), Safety, Security, Surveillance Laser Products | 95 | RFD | 1040.10 | Product, laser, weapon (military/police) |
Material Processing Laser Products | Laser Cutter, Material Processing Laser Products | 95 | RFE | 1040.10 | A high power laser intended to cut or drill a variety of materials in an industrial or commercial environment. |
Material Processing Laser Products | Laser Welder, Material Processing Laser Products | 95 | RFF | 1040.10 | A high power laser intended to weld (join) materials in an industrial or commercial environment. |
Material Processing Laser Products | Microelectronic Mask or Chip Checking/Repair, Material Processing Laser Products | 95 | RFG | 1040.10 | A laser intended to inspect and/or repair microelectronic components in an industrial or commercial environment. |
Material Processing Laser Products | UV Curing, Material Processing Laser Products | 95 | RFH | 1040.10 | An ultraviolet wavelength laser used to illuminate a material of a certain composition such that the laser "cures" or causes a chemical reaction to change the material in a desired fashion with no medical claims. Typical materials are adhesives, plastics, potting compounds, etc. |
Material Processing Laser Products | Print Industry Plate Maker, Material Processing Laser Products | 95 | RFI | 1040.10 | A laser intended to etch, engrave or otherwise create printer's plates used in an industrial or commercial environment. |
Material Processing Laser Products | Process Control, Material Processing Laser Products | 95 | RFJ | 1040.10 | A laser used for inspection, counting, or other application intended to monitor a part of the manufacturing process in an industrial or commercial environment. Often incorporated in an automated process system. |
Material Processing Laser Products | Laser Vision, Material Processing | 95 | RFK | 1040.10 | A laser used for positioning, focusing, inspection, counting, or other application in an industrial or commercial environment. Often incorporated in an automated assembly line system. |
Material Processing Laser Products | Laser Micrometer, Material Processing | 95 | RFL | 1040.10 | A laser used in high precision dimensional measurements in materials processing. |
Material Processing Laser Products | Laser-Based Material Positioning System | 95 | RFM | 1040.10 | A laser used in precision positioning of materials in manufacturing in an industrial or commercial environment. |
Material Processing Laser Products | Other | 95 | RZZ | Unk | A laser used in materials processing not otherwise defined. |
Material Processing Laser Products | General Industrial Use Material Processing Laser Products | 95 | RZN | 1040.10 | A laser used in industrial manufacturing or materials processing not otherwise defined. |
Data Measurement, Transmit, Control Laser Products | Fiber Optic Communication and Data Transfer, Laser | 95 | RFN | 1040.10 | A laser used in fiber optic communications to transmit data and information. |
Data Measurement, Transmit, Control Laser Products | IR Free-Space Data Transmit/Control, Laser | 95 | RFO | 1040.10 | A laser used in free space (open air) communications to transmit data and information. |
Data Measurement, Transmit, Control Laser Products | Remote Controller, Laser, Data Measurement, Transmit | 95 | RFP | 1040.10 | A laser used to transmit signals and/or information in order to operate equipment or machinery remotely. |
Data Measurement, Transmit, Control Laser Products | Interferometric Position Measuring Product, Laser | 95 | RFQ | 1040.10 | A laser used as an interferometer for high precision positioning and/or measurements. |
Data Measurement, Transmit, Control Laser Products | Product Incorporating Certified Class 1 Laser Data Measurement, Transmit, Control | 95 | RFR | 1040.10 | A data measurement, data transmission, or remote control product that incorporates a certified Class 1 laser. |
Data Measurement, Transmit, Control Laser Products | Other | 95 | RZZ | Unk | A data measurement, data transmission, or remote control product that incorporates a laser other than a certified Class 1 laser. |
Utility/Peripheral Laser Products | Reprographics, Laser, Utility/Peripheral Laser Products | 95 | RFS | 1040.10 | A reprographics machine that incorporates a laser utilized to expose internal sensitive components or materials for photocopying text and graphics. |
Utility/Peripheral Laser Products | Laser Printer, Utility/Peripheral Laser Products | 95 | RFT | 1040.10 | A printing machine that incorporates a laser utilized in printing images on paper with no medical claims. |
Utility/Peripheral Laser Products | Laser FAX Machine, Utility/Peripheral Laser Products | 95 | RFU | 1040.10 | A printing machine that incorporates a laser utilized in printing facsimiles of images on paper. |
Utility/Peripheral Laser Products | CD, CD-ROM Player, Laser Utility/Peripheral Laser Products | 95 | RFV | 1040.10 | A CD or CD-ROM player that utilizes a laser to read data on the compact disc. |
Utility/Peripheral Laser Products | DVD, DVD-ROM Player, Laser Utility/Peripheral Laser Products | 95 | RFW | 1040.10 | A DVD or DVD-ROM player that utilizes a laser to read data on the digitally recorded video disc. |
Utility/Peripheral Laser Products | CD-R, CD-RW Recorder, Utility/Peripheral Laser Products | 95 | RFX | 1040.10 | A CD-R or CD-RW recorder machine that utilizes a laser to read and/or write data on the compact disc. |
Utility/Peripheral Laser Products | DVD-R, DVD+R, DVD-RAM, DVD+RW, DVD-RW Recorder, Utility/Peripheral Laser Products | 95 | RFY | 1040.10 | A DVD recorder machine that utilizes a laser to read and write or read, write, and erase data on a digitally recorded video disc in any of the data formats: DVD-R, DVD+R, DVD-RAM, DVD-RW, or DVD+RW. |
Utility/Peripheral Laser Products | UPC Reader (Bar Code Reader), Utility/Peripheral Laser Products | 95 | RFZ | 1040.10 | A laser used to scan across a bar code to identify the product. Bar code readers can be hand-held accessories, under-counter components incorporated in store check-out systems, or laser scanner systems incorporated in assembly lines used for identification and inventory purposes in manufacturing facilities, warehouses and storage facilities, or other consumer, industrial, health care, or commercial locations. |
Utility/Peripheral Laser Products | Home/Office Machine Incorporating Utility/Peripheral Laser | 95 | RZP | 1040.10 | A laser utilized in the home or office environment not otherwise defined. |
Utility/Peripheral Laser Products | Product Incorporating Certified Class 1 Data Utility/Peripheral Laser Products | 95 | RGA | 1040.10 | A utility/peripheral laser product that incorporates a certified Class 1 laser. |
Utility/Peripheral Laser Products | Other | 95 | RZZ | Unk | A utility/peripheral laser product that incorporates a laser other than a certified Class 1 laser. |
In Vitro and Other Medical Laser Products | Veterinary Laser, In Vitro and Other Medical Laser Products | 95 | RGB | 1040.10; 1040.11 |
A laser used for treatment of animals other than human |
In Vitro and Other Medical Laser Products | Separator, Automated, Blood Cell, Diagnostic | 81 | GKT | 1040.10 | |
In Vitro and Other Medical Laser Products | Automated Differential Cell Counter | 81 | GKZ | 1040.10 | |
In Vitro and Other Medical Laser Products | Cell Particle Counter (Automated) | 81 | GKL | 1040.10 | |
In Vitro and Other Medical Laser Products | Urine Particle Counter | 88 | LKM | 1040.10 | |
In Vitro and Other Medical Laser Products | System, Separation, Hematopoietic Stem Cell | 81 | MZK | 1040.10 | |
In Vitro and Other Medical Laser Products | Test, Urea (Breath or Blood) for H. Pylori Test | 83 | MSQ | 1040.10 | |
In Vitro and Other Medical Laser Products | Multipurpose System for In-vitro Coagulation | 81 | JPA | 1040.10 | |
In Vitro and Other Medical Laser Products | System, Laser Assisted Hatching | 85 | MRX | 1040.10; 1040.11 |
|
In Vitro and Other Medical Laser Products | Sorter, Cell | 81 | KEX | 1040.10 | |
In Vitro and Other Medical Laser Products | Separator, Semi-Automated, Blood Component | 81 | MYY | 1040.10 | |
In Vitro and Other Medical Laser Products | Other | 95 | RZZ | Unk | A laser used for in vitro applications or other medical applications that do not expose patients to the laser radiation. |
Positioning Medical Laser Products | X-Ray Field Indicator Light (Laser), Positioning Medical Laser Products | 95 | RGC | 1020.30; 1040.10; 1040.11 |
A laser incorporated in a diagnostic x-ray system that is irradiated onto the film screen area indicating the x-radiation area. The beam is usually scanned to show a rectangular region for patient placement. |
Positioning Medical Laser Products | Monitor, Patient Position, Light Beam | 90 | IWE | 1040.10; 1040.11 |
|
Positioning Medical Laser Products | Positioning Medical Laser Product | 95 | RZS | 1040.10; 1040.11 |
A laser used for positioning in medical applications not otherwise defined. |
Performance requirements | |||
1040.10(d) | Classified in higher class | Minor, Concern | Class B, C |
1040.10(d) | Classified in lower class | Major | Class A |
1040.10(f)(1) | Protective housing allows unnecessary body access to Class IV or high IIIb radiation | Major | Class A |
1040.10(f)(1) | Protective housing allows unnecessary straight line access to interior Class IV or high IIIb radiation With high risk of exposure (IV or IIIb product) With low risk of exposure (IV or IIIb product) With any risk of exposure (I, IIa, II, or IIIa product) |
Major Minor Major |
Class A Class B Class A |
1040.10(f)(1) | Protective housing allows unnecessary body access to low Class IIIb or IIIa radiation In a Class IV or IIIb product In a Class I, IIa, II, or IIIa product |
Minor Major |
Class B Class A |
1040.10(f)(1) | Protective housing allows necessary body access to Class IIIa or IIIa radiation In a Class IV or IIIb product In a Class I, IIa, II, or IIIa product |
Concern Minor |
Class C Class B |
1040.10(f)(1) | Protective housing allows unnecessary body access to Class II radiation In a Class II product In a Class I product |
Concern Minor |
Class C Class B |
1040.10(f)(2) | Safety interlocks absent when required | Major | Class A |
1040.10(f)(2) | Single safety interlock when redundant required | Major | Class A |
1040.10(f)(2) | Single component with multiple contacts when redundant required | Minor | Class B |
1040.10(f)(2) | Defeatable safety interlocks lacks indication | Minor | Class B |
1040.10(f)(2) | Defeatable safety interlocks fails to prevent replacement of protective housing during defeat | Minor | Class B |
1040.10(f)(3) | No remote interlock connector | Major | Class A |
1040.10(f)(4) | No key control | Major | Class A |
1040.10(f)(4) | Key control removable when on | Major | Class A |
1040.10(f)(5) | No emission indicator | Major | Class A |
1040.10(f)(5) | No delay preceding radiation emission | Minor | Class B |
1040.10(f)(5) | Shorter delay than required | Minor | Class B |
1040.10(f)(5) | Remote control lacks emission indicator | Major | Class A |
1040.10(f)(6) | Beam attenuator without approvable alternate | Major | Class A |
1040.10(f)(6) | Beam attenuator with approvable alternate | Concern | Class C |
1040.10(f)(8) | Viewing optics Hazardous Non-hazardous for viewing period |
Major Concern | Class A Class C |
1040.10(f)(9) | No scanning guards | Major | Class A |
1040.10(f)(10 | No manual reset | Major | Class A |
1040.10(g)(1), (2), and (3) | Warning logotype None Classification too low Classification too high |
Major Major Minor, Concern |
Class B Class B Class B |
1040.10(g)(4) | Warning logotype output information | Minor | Class B |
1040.10(g)(5) | No aperture label | Minor | Class B |
1040.10(g)(5) | Aperture label not in close proximity to aperture | Minor | Class B |
1040.10(g)(5) | Aperture label wording incorrect | Concern | Class C |
1040.10(g)(6), (7) | No protective housing labels | Minor | Class B |
1040.10(g)(6), (7) | Protective housing placement inappropriate | Minor | Class B |
1040.10(g)(6), (7) | Protective housing wording wrong | Concern | Class C |
1040.10(g)(8) | Invisible radiation warning on labels | Minor | Class B |
1040.10(g)(9), (10) | Label positioning and legibility | Minor | Class B |
1040.10(h)(1) | User instructions (i) Promoting unsafe practices Inadequate instructions to avoid exposure (ii) Inadequate radiometric specifications (iii) Inadequate reproductions and locations (iv) Inadequate listing of controls Inadequate caution statement |
Major Minor Minor Minor Minor Concern |
Class A Class B Class B Class B Class B Class C |
1040.10(h)(2)(i) | Reproduction of warning logotype not in catalogs | Minor | Class B |
1040.10(h)(2)(ii) | Service information inadequate | Minor | Class B |
Specific product requirements | |||
1040.11(a)(1) | Means to measure medical laser output None Inaccurate |
Major Major |
Class A Class A |
1040.11(a)(2) | Inadequate calibration procedure/schedule | Major | Class A |
1040.11(a)(3) | Aperture label | Minor | Class B |
1040.11(b) | Excessive output on surveying lasers | Major | Class A |
1040.11(c) | No variance for demonstration Class IIIb or Class IV lasers | Major | Class A |
LASER PRODUCT TEST RECORD
MANUFACTURER____________________ CLASS____________
MODEL_______________________ SERIAL NUMBER_______
Status of Unit Examined (Circle one): Prototype/Production unit
Status of Assembly (circle one): Complete/Incomplete
Manufactured Date:_________________________
A. Product Description: (Include basic configuration and size of product, reference to photos and/or
diagrams, basic functions to be performed during operation and during maintenance.)
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
Yes_________ No_________
If yes, what is the Accession Number? _____________________________
Summary of Product Evaluation:
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
1. Certification label (1010.2)
a. Is the label permanently affixed? Yes___ No___ ND___ NA___
b. Is the Label readily viewable? Yes___ No___ ND___ NA___
Location:______________________
c. Is the label properly stated? Yes___ No___ ND___ NA___
(Note: Products under variance require modified certification labels 1010.4(d))
d. Remarks: ________________________________________________
__________________________________________________________
2. Identification label (1010.3)
a. Is the label permanently affixed? Yes___ No___ ND___ NA___
b. Is the label readily viewable? Yes___ No___ ND___ NA___
Location: _______________________
c. Does the label contain the full name and address?
Yes___ No___ ND___ NA___
d. Does the label contain the place of manufacture (in full or in code)?
Yes___ No___ ND___ NA___e. If coded, has CDRH been provided the code?
Yes___ No___ ND___ NA___f. Are the month and year of manufacture stated in full?*
Yes___ No___ ND___ NA___
Month and year: __________________________
g. Remarks:____________________________________________________
____________________________________________________________
*Note: Serialization is acceptable in lieu of month and year for consumer electronic products.
D. Special Purpose Products (1040.11)
1. Is the product a medical laser product?
Yes___ No___ ND___ NA___
Note: In inspecting manufacturers of not only medical laser products but also laser products that are medical devices, verify compliance with other applicable requirements including but not limited to current registration and listing, 510k market clearances, device master record or quality system, current complaint and service records, etc.
a. Does the product include a means of measurement of levels of radiation intended for irradiation of the human body?
Yes___ No___ ND___ NA___
b. How is this accomplished? Measure beam prior to delivery system and determine output levels via calibration constant _________;
Measure output of delivery system ___________;
Other ___________.c. Indication: power ________; energy________; time _______.
d. Type of indicator: energy/power select switch ___; "Test shot" display (remains constant until next best shot) ___; Real time display (displays level at all times) ___; Other ___.
e. If test shot is available only at initiation of procedure or if a select switch is used, does the product have an internal monitoring system capable of maintaining output levels to within ± 20% of displayed value?
Yes ___ No___f. Is display analog ____; or digital ____? If digital, are there sufficient significant digits to allow ± 20% accuracy? Yes ___ No ___
g. Is the total measurement error within ± 20% (see Attachment G) Yes___ No___ ND___ NA___
h. Is there a laser aiming beam? Yes ___ No ___. Is there a means to measure the level of the aiming beam if the product is ophthalmic and the aiming beam may exceed 1 mW or if the product is not ophthalmic but the aiming beam may exceed 5 mW?
Yes ___ No ___i. Remarks: _____________________________________________________
_____________________________________________________________
2. Is the product a surveying, leveling, and alignment product?
Yes___ No___ ND___ NA___
a. Is access prevented for wavelengths of 400 nm to 710 nm to radiation power in excess of 5.0 mW for any duration greater than 3.8 x 10-4 seconds?
b. Is access prevented to radiation levels in excess of Class I limits for any other combination of emission Duration and wave length range?
Yes___ No___ ND___ NA___
c. Remarks: ____________________________________________________
____________________________________________________________
3. Is the product a demonstration laser product?
Yes___ No___ ND___ NA___a. Does the product prevent human access to radiation in excess of the Class IIIa (3R) limit?
Yes___ No___ ND___ NA___b. Remarks: ____________________________________________________
____________________________________________________________
E. Label Requirements. See instruction in paragraph B.
1. Warning logotypes* (1040.10(g)(1),(2),(3),(4),(5),(9), and (10)
a. Is the logotype the correct logotype?
Yes___ No___ ND___ NA___b. Is the label properly worded for its class designation?
Yes___ No___ ND___ NA___
c. Does the label have the proper color?
Yes___ No___ ND___ NA___
d. Is the output information present and correct?
(Maximum output stated ________.) Yes___ No___ ND___ NA___
e. Is the media or wavelength information present and correct?
Yes___ No___ ND___ NA___
f. Is the label permanently affixed and clearly visible during operation, maintenance, and
service?
Yes___ No___ ND___ NA___g. Is the label positioned so as to make exposure unnecessary during reading?
Yes___ No___ ND___ NA___
Location: ____________________h. Does the label include a warning for "invisible" or "invisible and/or visible" radiation?
Yes___ No___ ND___ NA___i. Remarks: __________________________________________________________
__________________________________________________________________
__________________________________________________________________
Note: Warning labels in accordance with IEC 60825-1 including product classification are acceptable.
2. Aperture label (for Classes II, IIIa, IIIb, IV, 3R, 3B and 4) 1040.10(g)(5),(8),(9), and (10)
a. Is a label present and in proximity to each aperture?
Yes___ No___ ND___ NA___
b. Is the label properly worded? Yes___ No___ ND___ NA___
c. Is the label permanently affixed and clearly visible?
Yes___ No___ ND___ NA___
d. Is the label positioned so as to make exposure unnecessary during reading?
Yes___ No___ ND___ NA___(Location: ______________)
e. Does the label include a warning for "invisible" or "invisible and/or visible" radiation?
Yes___ No___ ND___ NA___f. Remarks: __________________________________________________________
__________________________________________________________________
__________________________________________________________________
3. Noninterlocked protective housing label (1040.10(g)(6),(8),(9), and (10)
a. Are the labels on or near all appropriate panels or covers which are removed for operation, maintenance, or service?
Yes___ No___ ND___ NA___
b. Are all labels visible prior to removal of such portions of the protective housing?
Yes___ No___ ND___ NA___c. Are all labels visible after opening?
Yes___ No___ ND___ NA___d. Are all labels correctly worded? Yes___ No___ ND___ NA___
e. Are all labels permanently affixed and clearly visible?
Yes___ No___ ND___ NA___
f. Do all labels contain a warning for "invisible" or "invisible and/or visible" radiation? Yes___ No___ ND___ NA___
g. Remarks: __________________________________________________________
__________________________________________________________________
__________________________________________________________________
4. Defeatably interlocked housing labels 1040.l0(g)(7),(8),(9), and (10)
a. Are labels provided for each defeatably interlocked panel or cover which is removed for operation, maintenance, or service?
Yes___ No___ ND___ NA___
b. Are all labels visible prior to interlock defeat?
Yes___ No___ ND___ NA___
c. Are all labels visible during interlock defeat?
Yes___ No___ ND___ NA___
d. Are all labels correctly worded?
Yes___ No___ ND___ NA___
e. Are all labels permanently affixed and clearly visible?
Yes___ No___ ND___ NA___
f. Do all labels contain a warning for "invisible" or invisible and/or visible" radiation?
Yes___ No___ ND___ NA___g. Remarks: _____________________________________________________
_____________________________________________________________
F. Performance Requirements (1040.l0(f)):
1. Protective Housing (1040.l0(f)(l))
a. Does the housing prevent access at all times to laser radiation above Class I not necessary for operation of the product?
Yes___ No___ ND___ NA___
b. Does the housing prevent access at all times to collateral optical radiation above Class I not necessary for operation of the product?
Yes___ No___ ND___ NA___c. Has x-radiation been evaluated?
Yes___ No___ ND___ NA___d. Does the housing prevent access to x-radiation levels in excess of 0.5 mR/hr at all times during operation of the product?
Yes___ No___ ND___ NA___
e. Remarks: _____________________________________________________
_____________________________________________________________
2. Safety Interlocks (1040.10(f)(2)) (Complete for each interlock. Identify the portion of
removable or displaceable housing and interlock described.)a. Do operation or maintenance functions require moving portions of the housing which could allow access to radiation?
Yes___ No___ ND___ NA___
Describe: _____________________________________________________
_____________________________________________________________
b. Class of radiation to which access could be gained?
Class ______.c. Is a fail safe or multiple interlock required (including 1040.l(f)(2)(iii)?
Yes___ No___ ND___ NA___
Where? ______________________________________________________
_____________________________________________________________
d. Are safety interlock(s) present? where?_______________
1. TYPE: Microswitch____; Mercury switch____;
male-female plug____; mechanical shutter____; other____.Describe: _____________________________________________
e. Method of limiting access: directly interrupts primary laser power____; interrupts primary laser power through relay, contactor, switching tube or transistor____; spoils the cavity____; shutter beam via solenoid____; other____.
f. Is there a multiple or fail safe interlock on each housing for which an interlock is required?
Yes___ No___ ND___ NA___
g. Is the interlock defeatable? Yes___ No___ ND___ NA___
h. Is there an indication of defeat? Yes___ No___ ND___ NA___
Describe: _______________________________________________________
i. Does the interlock preclude replacement of the housing while the interlock is defeated?
Yes___ No___ ND___ NA___j. Are non-safety interlocks present? Where?_____________
Yes___ No___ ND___ NA___
k. Remarks: _________________________________________________
_________________________________________________________
3. Remote Interlock Connector (1040.l0(f)(3), Class IIIb or IV systems only)
a. Is a remote control connector present?
Yes___ No___ ND___ NA___
b. Type? Describe: ________________________________________
c. Is the voltage across the connector less than 130 volts RMS?
Yes___ No___ ND___ NA___
d. Is the access to laser and collateral radiation prevented when the terminals are not joined?
Yes___ No___ ND___ NA___e. Method of operation: Directly interrupts laser power____; interrupts laser power through relay, etc.____; shutters beam or interrupts cavity____.
f. Does the emission delay reactivate when the remote control circuit is interrupted?
Yes___ No___ ND___ NA___g. Must the emission be manually restarted following interruption via the remote interlock connector?
Yes___ No___ ND___ NA___
h. Remarks: ____________________________________________________
____________________________________________________________
4. Key Control (1040.10(f)(4), Class IIIb, IV, 3B, or 4 systems only)
a. Is a key control present? Yes___ No___ ND___ NA___
Describe: ___________________________________________________
_____________________________________________________________
b. Is a key removable in the "on" position?
Yes___ No___ ND___ NA___
c. Is operation prevented when the key is removed?
Yes___ No___ ND___ NA___
d. How? ____________________________________________________
____________________________________________________________
e. Remarks: _________________________________________________
____________________________________________________________
5. Beam Attenuator (1040.l0(f)(6), Class IIIb, IV, 3B or 4 systems only)
a. Is a beam attenuator present? Yes___ No___ ND___ NA___
b. Type: mechanically operated shutter____; electrically operated____; aperture cap or cover____; other____.
Describe: ___________________________________________________
____________________________________________________________
c. Is the attenuator permanently attached?
Yes___ No___ ND___ NA___
d. Does the attenuator prevent access by any part of the body to radiation in excess of Class I limits?
Yes___ No___ ND___ NA___
e. If there is no beam attenuator, has the manufacturer requested and obtained approval of an alternate means of safety?
Yes___ No___ ND___ NA___
f. Remarks: _________________________________________________
_____________________________________________________________
6. Emission Indicator (1040.10(f)(5), Class, IIIb, IV, 3B or 4 Systems only)
a. Is an emission indicator present on the laser product?
Yes___ No___ ND___ NA___
Where? ________________________________________________
b. Type: tungsten lamp(s)____; neon lamp(s)____; LED(s)____; other____.
Describe: _________________________________________________
_______________________________________________________
c. If the indicator is visible, is it visible through the protective eyewear that is normally supplied or recommended?
Yes___ No___ ND___ NA___
d. Can the indicator be viewed without exposure to radiation in excess of Class I limits?
Yes___ No___ ND___ NA___
e. Is there a delay between an indication of emission and the beginning of emission?
Yes___ No___ ND___ NA___f. How is emission delay achieved? Thermal relay___; inherent in the lasing process___; delay circuit___; other___.
Describe: ________________________________________________
_________________________________________________________
g. Length of delay? ____________________
h. Is the power source or operation control separable from the laser by greater than 2 meters when assembled for use?
Yes___ No___ ND___ NA___
i. If separated greater than 2 meters, is an emission indicator present on the energy source or controller?
Yes___ No___ ND___ NA___Where? _____________________________________________________
j. Type: Tungsten lamp(s)___; neon lamp(s)___; LED(s)___; bell or buzzer___; meter or display___; mechanical flag___; other___.
Describe: ______________________________________________
______________________________________________________
k. Is there a delay between an indication of emission and the beginning of emission?
Yes___ No___ ND___ NA___l. How is emission delay achieved? Thermal relay___; inherent in the lasing process___; delay circuit___; other___.
Describe: ______________________________________________
______________________________________________________
m. Length of delay? ____________________________________________
n. Remarks: __________________________________________________
____________________________________________________________
____________________________________________________________
6. Location of Controls (1040.10(f)(7))
a. Are the controls located so that exposure is unnecessary for operation or adjustments?
Yes___ No___ ND___ NA___b. Remarks: __________________________________________________
____________________________________________________________
____________________________________________________________
7. Viewing optics (1040.l0(f)(8))
a. Are viewing optics or viewports present?
Yes___ No___ ND___ NA___
b. Type: microscope___; telescope___; viewport___; display screen___; other___.
Describe: ___________________________________________________
____________________________________________________________
____________________________________________________________
c. Where? ___________________________________________________
____________________________________________________________
d. Do the viewing optics attenuate radiation at all times during operation or maintenance to levels less than Class I limits?
Yes___ No___ ND___ NA___
e. Do the viewing optics employ a shutter or variable attenuator?
Yes___ No___ ND___ NA___
f. Upon failure of the shutter of the variable attenuator is access to radiation levels greater than the Class I limits prevented?
Yes___ No___ ND___ NA___g. Remarks: _________________________________________________
____________________________________________________________
____________________________________________________________
8. Scanning Safeguard (1040.10(f)(9))
a. Is the radiation emitted by the product scanned?
Yes___ No___ ND___ NA___
b. Is the classification of the product based on the level of scanned radiation?
Yes___ No___ ND___ NA___c. In the event of scan failure, is human access to laser radiation in excess of the product class prevented?
Yes___ No___ ND___ NA___
d. Remarks: _________________________________________________
_____________________________________________________________
9. Manual Reset Mechanism (1040.l0(f)(l0) Class IV laser systems)
Describe the operation of the Manual Reset. _______________________________________________________________________________
How is it achieved? (latching relay, etc.) ________________
____________________________________________________________
10. Removable laser system (1040.10(c)(2))
a. Does the product incorporate a laser system?
Yes___ No___ ND___ NA___
b. Is the laser system removable?
Yes___ No___ ND___ NA___c. If removable, is the laser system independently certified?
Yes___ No___ ND___ NA___
d. If not removable, specify how removability is prevented: hard wiring___; modified connector___; assembled internally from components___; other (specify). __________________________
e. Remarks: __________________________________________________
____________________________________________________________
____________________________________________________________
G. Laser Product Measurements
Model # ________________________ Serial # _______________________
Manufacturer's Claimed Classification: ___________________________________
Brief description of product: ___________________________________________
___________________________________________________________________
___________________________________________________________________
Test Instrument(s) Used: _______________________________________________
___________________________________________________________________
Circle radiometric quantity tested and specify units below (Radiance (W cm-2 sr-1), Radiant Energy (J), Power (W), etc.)
Measurement No. |
Wavelength (nm) |
Instrument reading, R (units ________) |
Calibration factor, K (units ________) |
Corrected value, R*K (units ________) |
Calculations (as needed):
__________________________________________________________________
__________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
H. Results of FDA measurements: ______________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
I. Compliance with other requirements (e.g., conditions of a variance, labeling for medical devices, etc.)
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
J. Information requirements (Directions: Complete this section only if the information and requirements are reviewed during the inspection).
1. User Information (1040.l0(h)(l))
a. Does the manual contain adequate instructions for assembly,
operation, and maintenance? Yes___ No___ ND___ NA___b. Does it contain clear warnings to avoid exposure?
Yes___ No___ ND___ NA___
c. Does it contain a statement of output parameters?
Yes___ No___ ND___ NA___
d. Does it contain legible reproductions of all labels and hazard warnings?
Yes___ No___ ND___ NA___e. Does it include the corresponding position of each label on the product?
Yes___ No___ ND___ NA___f. Does it contain listing of controls, adjustments, and procedures for operation and maintenance?
Yes___ No___ ND___ NA___g. Does it contain a schedule of maintenance?
Yes___ No___ ND___ NA___
h. Does it contain the "Caution - use of controls..." warning?
Yes___ No___ ND___ NA___
i. Does it contain a compatibility statement (laser source or laser system not supplied with the product?
Yes___ No___ ND___ NA___
j. Does it contain a calibration schedule (medical laser product)?
Yes___ No___ ND___ NA___k. Does it include a warning not to point the laser radiation at the audience (especially Class IIIa demonstration laser products)?
Yes___ No___ ND___ NA___l. Does it include information to determine nominal hazard zone(class IV multi-axis workstations)?
Yes___ No___ ND___ NA___m. Remarks: __________________________________________________________
__________________________________________________________________
__________________________________________________________________
2. Purchasing Information (1040.10(h)(2))
a. Are legible reproductions of the logotype required to be affixed to the product (including information required for positions 1, 2, and 3) contained in catalogues, specification sheets, and descriptive brochures?
Yes___ No___ ND___ NA___
3. Servicing Information (1040.l0(h)(2))
a. Are adequate instructions for service adjustments and service procedures available?
Yes___ No___ ND___ NA___b. Are clear warnings and precautions to avoid possible exposure included?
Yes___ No___ ND___ NA___c. Is a schedule of maintenance necessary to keep the product in compliance included?
Yes___ No___ ND___ NA___d. Are controls and procedures which would he used by reasons other than the manufacturer or his agent to increase accessible emission levels listed?
Yes___ No___ ND___ NA___e. Is a clear description of the locations of displaceable portions of the protective housing provided?
Yes___ No___ ND___ NA___f. Do these instructions provide legible reproductions of required labels and hazard warnings?
Yes___ No___ ND___ NA___g. Do these instructions include protective procedures for service personnel?
Yes___ No___ ND___ NA___h. Remarks: __________________________________________________________
__________________________________________________________________
__________________________________________________________________
A sunlamp product is an electronic product designed to use one or more ultraviolet lamp(s) and is intended for irradiation of any part of the living human body by ultraviolet radiation within a specified range of wavelengths to induce skin tanning. The ultraviolet lamps, subject to the performance standard, produce radiation within a prescribed range of wavelengths and are intended for use in sunlamp products.
Sunlamp products include portable home units, table top models, tanning beds and tanning booths. These units may incorporate different types of fluorescent lamps, reflector spot (RS) or High Intensity Discharge (HID) with different levels of energy output and radiation at different wavelengths.
Since sunlamp products are radiation-emitting electronic products as defined by Section 531 of Subchapter C- Electronic Product Radiation Control (EPRC) formerly the Radiation Control for Health and Safety Act (RCHSA) and medical devices as defined by Section 201(h)(3) of the Federal Food, Drug, and Cosmetic Act (FFDCA, the Act), they are regulated under both laws.
Under authority of Section 534 of the (EPRC), a performance standard for sunlamp products and ultraviolet lamps intended for use in these products was promulgated effective May 7, 1980 (21 CFR 1040.20). The standard was intended to reduce sunlamp related injuries by reducing unnecessary exposure and overexposure to sunlamp radiation by: (1) limiting shorter wavelength emissions that are not necessary and pose unreasonable risk, (2) providing for adequate warning label and user instructions containing safety information, and (3) requiring special lamp bases, protective eyewear, timers, and controls to help users limit the duration and amount of exposure.
This performance standard was promulgated when the common sunlamp product was a table-top, home portable unit incorporating one or two RS lamps having a large part of their radiation output in the wavelength range of 260 to 320 nanometers (UVB). In 1979-80, a new-wave of sunlamp products came onto the market. These products, commonly referred to as Tanning Booths, usually measured 3'x3'x7' and contained one or two fluorescent ultraviolet lamps in each corner. These products also had relatively high UVB output.
Around early 1983, another product in the shape of a bed and/or canopy entered the market with fluorescent lamps that emit radiation mainly in the 320-400 nanometer range (UVA), with usually less than 5% in-the UVB range. This type of product requires longer exposure times to achieve its intended purpose and the risk of chronic sunburn is reduced relative to the older type of products. Most manufacturers requested variance under 21 CFR 1010.4 to equip the products with timers which would allow exposure in excess of ten minutes. Since the products usually required 30 minutes to achieve their intended result, the variances were granted with two conditions: (1) the maximum timer interval shall not exceed the maximum recommended exposure time specified in the required product label, and (2) the UVB to UVA ratio shall not exceed .05 (no more than 5% UVB). The manufacturers are required to specify the variance number and effective date on the product).
Some of these products incorporate High Intensity Discharge (HID) lamps. These lamps are usually used for facial tanning, although some whole body exposure systems use such lamps exclusively. In most cases, however, these lamps are used in conjunction with ultraviolet fluorescent lamps. The HID lamps are much smaller than fluorescent lamps, (usually about 1/2" in diameter by 3" in length) and they usually incorporate an outer, clear, glass envelope.
On September 6, 1985, amendments to the performance standard were published and became effective in September 8, 1986. The purpose of the amendments is to accommodate new products employing design concepts significantly different from those for which the original standard was developed. Also, FDA experience in applying the original standard indicated that some requirements were either inappropriate for or not applicable to some products. The amendments are intended to establish a standard that is appropriate for the present technology of tanning and new sunlamp product designs. This revised program offers guidance for testing products against the original standard or revised standard, as appropriate.
Some electro-optics specialists, x-ray auditors and other radiological health specialists have been trained in general EPRC requirements and also may have specialized training in the sunlamp product performance standards. Only trained individuals should perform these inspections and field tests and may train additional field staff or accompany a medical device investigator to conduct joint EPRC/medical device inspections. If an EOS has training in both EPRC and QSIT inspections, a single EOS may conduct both portions of the inspection.
The District Offices have the authority (delegated under 21 CFR 5.37 and 5.89) to make declarations of noncompliance and/or defect for sunlamp products. The field also has the authority to approve sunlamp manufacturer corrective action plans under 21 CFR 1004 and to grant exemptions (from notification and product repair) in accordance with 21 CFR 1003.31. Consult CDRH for assistance in determining appropriate enforcement action or other support. A copy of any letter issued to a manufacturer must be sent to HFZ-240.
Sunlamp Products, Performance Standard – 21 CFR 1040.20.
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfcfr/CFRSearch.cfm?FR=1040.20
Quality Control Guide for Sunlamp Products. (Publication; FDA 84-8234)
http://www.fda.gov/cdrh/radhlth/pdf/SUNQCG.pdf
Policy on Warning Label Required on Sunlamp Products (6/25/85)
http://www.fda.gov/cdrh/radhlth/pdf/sunpol01.pdf
Policy on Maximum Timer Interval and Exposure Schedule for Sunlamp Products (8/21/86)
http://www.fda.gov/cdrh/radhlth/pdf/sunpol01.pdf
Policy on Lamp Compatibility (9/2/86).
http://www.fda.gov/cdrh/radhlth/pdf/sunpollc.pdf
Sunlamp Products Reporting Guide, (dated September, 1995).
http://www.fda.gov/cdrh/radhlth/pdf/sunrpt0p.pdf
Refer to the sunlamp products main page for additional information:
http://www.fda.gov/cdrh/radhealth/products/sunlamps.html
Translation of 2-Digit Code | Product Name | Product Code | CFR | Definition | |
---|---|---|---|---|---|
Sunlamp Products (Certified) | Suntan Booth | 79 | LEJ | 1040.20 | |
Sunlamp Products (Certified) | Suntan Bed, Sunlamp Products (Certified), Non-Medical | 95 | REF | 1040.20 | A bed or other platform that is designed to incorporate one or more ultraviolet lamps and intended for irradiation of any part of the human body, by ultraviolet radiation with wavelengths in air between 200 and 400 nanometers, to induce skin tanning with no medical claims. |
Sunlamp Products (Certified) | Suntan Lamp, Sunlamp Products (Certified), Non-Medical | 95 | REG | 1040.20 | A lamp that produces ultraviolet radiation in the wavelength range of 200 to 400 nanometers in air and that is intended for use in any sunlamp product or fixture with no medical claims. |
Sunlamp Products (Certified) | Tabletop Sunlamp System (Certified), Non-Medical | 95 | REH | 1040.20 | A sunlamp system that sits on a table, primarily intended to tan the face by ultraviolet radiation with wavelengths in air between 200 and 400 nanometers with no medical claims. |
Sunlamp Products (Certified) | Other | 95 | RZZ | Unk | Sunlamp product means any electronic product designed to incorporate one or more ultraviolet lamps and intended for irradiation of any part of the human body, by ultraviolet radiation with wavelengths in air between 200 and 400 nanometers, to induce skin tanning. |
Performance Requirements | |||
1040.20(c)(1) | Fails to comply with the irradiance ratio limits for UVC over UVB cannot exceed 0.003 | Minor | Class B |
1040.20(c)(2)(i) | Fails to incorporate a timer system with multiple timer settings adequate for recommended exposure time intervals | Major | Class A |
1040.20(c)(2)(ii) | Maximum timer interval(s) is more than 3 times greater than the manufacturer’s recommended maximum exposure time(s) as indicated on label | Major | Class A |
1040.20(c)(2)(ii) | Maximum timer interval(s) is 2 – 3 times greater than the manufacturer’s recommended maximum exposure time(s) as indicated on label | Minor | Class B |
1040.20(c)(2)(ii) | Maximum timer interval(s) is less than 2 times greater than the manufacturer’s recommended maximum exposure time(s) as indicated on label | Concern | Class C |
1040.20(c)(2)(iii) | Maximum timer interval error > 30 percent | Major | Class A |
1040.20(c)(2)(iii) | Maximum timer interval error > 20 and < 30 percent | Minor | Class B |
1040.20(c)(2)(iii) | Maximum timer interval error > 10 and < 20 percent | Concern | Class C |
1040.20(c)(2)(iv) | Timer automatically resets and causes radiation to resume. | Major | Class A |
1040.20(c)(3) | Fails to incorporate a control for termination of radiation emission (at minimum a timer system) | Major | Class A |
1040.20(c)(4)(i) | Fails to have protective eyewear | Minor | Class B |
1040.20(c)(4)(ii) | Spectral transmittance of the protective eyewear exceeds a value of 0.001 over the wavelength UVC and UVB(200nm to 320nm) | Minor | Class B |
1040.20(c)(4)(ii) | Spectral transmittance of the protective eyewear exceeds a value of 0.01 over the wavelength UVA (>320nm to 400nm) | Minor | Class B |
1040.20(c)(4)(ii) | Spectral transmittance (>400nm) of protective eyewear does not allow user to clearly see to reset the timer | Minor | Class B |
1040.20(c)(5) | UV lamp capable of insertion and operation in either the “single-contact medium screw” or the “double-contact medium screw” lamp holders. | Major | Class A |
Label Requirements for Sunlamp Products | |||
1040.20(d)(1)(i) | Fails to have warning statement “Danger UV radiation…” | Minor | Class B |
1040.20(d)(1)(ii) | Fails to have recommended exposure position(s) | Minor | Class B |
1040.20(d)(1)(iii) | Fails to have directions for recommended exposure position(s) and warning other positions may result in overexposure | Minor | Class B |
1040.20(d)(1)(iv) | Fails to have recommended exposure schedule | Minor | Class B |
1040.20(d)(1)(v) | Fails to have time before expected results statement | Concern | Class C |
1040.20(d)(1)(vi) | Fails to have ultraviolet lamp designation | Minor | Class B |
Label Requirements for Ultraviolet Lamps | |||
1040.20(d)(2)(i) | Fails to have “Sunlamp-DANGER-Ultraviolet radiation. Follow Instructions” | Minor | Class B |
1040.20(d)(2)(ii) | Fails to have model identification | Minor | Class B |
1040.20(d)(2)(iii) | Fails to have “Use ONLY in fixture equipped with timer” | Minor | Class B |
Label Specifications for Sunlamp Products and Ultraviolet Lamps | |||
1040.20(d)(3)(i) | Fails to be permanently affixed or inscribed on the exterior surface of sunlamp product when fully assembled for use so as to be legible and readily accessible to view by person being exposed immediately before use of product | Minor | Class B |
1040.20(d)(3)(ii) | Fails to be permanently affixed or inscribed on the ultraviolet lamp so as to be legible or readily accessible to view | Minor | Class B |
1040.20(d)(3)(iv) | Fails to have identification and certification labels on shelf package of ultraviolet lamps and coded mfr name and date of mfr on ultraviolet lamp | Minor | Class B |
1040.20(d)(3)(v) | Labels contain statements or illustrations that are false or misleading, diminish the impact of the required statements, or are prohibited by this chapter. | Major | Class A |
Instructions to be provided to users of Sunlamp Products | |||
1040.20(e) | Inadequate instructions for use to avoid or minimize potential injury provided to purchaser | Minor | Class B |
1040.20(e)(1)(i) | Failed to have reproduction of “Danger Ultraviolet Radiation warning statement…” | Minor | Class B |
1040.20(e)(1)(ii) | Failed to have a statement of the maximum number of users and warning that only that number of protective eyewear was provided | Concern | Class C |
1040.20(e)(1)(iii) | Failed to have instructions on the proper operations of the product including function, use, and setting of the timer and other controls , and use of the protective eyewear | Minor | Class B |
1040.20(e)(1)(iv) | Failed to have instructions determining the correct exposure time and schedule for persons according to skin type. | Minor | Class B |
1040.20(e)(1)(v) | Failed to have instructions for obtaining repairs and recommended replacement components and accessories which are compatible with the product, including compatible protective eyewear, ultraviolet lamps, timers, reflectors, and filters, if installed or used as instructed would result in continued compliance with the standard. | Minor | Class B |
1040.20(e)(2)(i) | User instructions for ultraviolet lamps not sold with sunlamp products failed to have a reproduction of the “Danger Ultraviolet Radiations…warning statement and the “Sunlamp-DANGER Ultraviolet radiation. Follow Instructions” and “Use ONLY in a fixture equipped with a timer” label | Minor | Class B |
1040.20(e)(2)(ii) | User instructions for ultraviolet lamps not sold with sunlamp products failed to have a warning that instructions should be followed to avoid or minimize potential injury | Minor | Class B |
1040.20(e)(2)(iii) | User instructions for ultraviolet lamps not sold with sunlamp products failed to have a clear identification by brand and model designation of all lamps models for which replacement lamps are promoted | Minor | Class B |
Tests for Determination of Compliance | |||
1040.20(f) | Fail to account for all errors and statistical uncertainties in the process for changes in radiation emission or degradation in radiation safety with age of the product. | Minor | Class B |
1040.20(f) | Fail to make measurements for certification under operational conditions as recommended by the manufacturer. | Minor | Class B |
1040.20(f) | Fail to position measuring instrument at recommended exposure position and oriented to result in maximum detection of the radiation | Minor | Class B |
INSPECTIONAL FIELD TEST CHECKLIST REPORT FOR SUNLAMP PRODUCTS
MANUFACTURED AFTER SEPTEMBER 8, 1986
(Including Pertinent Parts of the Regulation)
FACILITY NAME: |
_______________________________________ |
|
PERSON INTERVIEWED: |
_______________________________________ |
ADDRESS: |
_______________________________________ |
|
TELEPHONE NUMBER |
_______________________________________ |
|
_______________________________________ |
|
FIELD TEST DATE |
_______________________________________ |
WARNING LABEL [21 CFR 1040.20(d)(1)]
Accessible To View: Yes / No Legible From One Meter: Yes / No Exposure Position: Yes / No "DANGER" Statement:: Yes / No
If "NO" to any of the above, Explain:___________________________________________________________________________________
Exposure Schedule times: Minimum ____min. / Maximum ____min. Warning Label Location:_________________________________
List All Lamp Types Designated On Unit Labeling:_________________________________________________________________________
CERTIFICATION LABEL [21 CFR 1040.20(d) & 21 CFR 1010.2]
Adequate Certification: Yes / No Written In English: Yes / No Legible: Yes / No
If "NO" to any of the above, Explain:___________________________________________________________________________________
IDENTIFICATION LABEL [21 CFR 1040.20(d) & 21 CFR 1010.3] (AS APPEARS ON LABEL)
Name & Address of Manufacturer:______________________________________________________________
Model #:_______________________________ Serial #:_______________________ Date of Manufacture:___________________________
PROTECTIVE EYEWEAR [21 CFR 1040.20(C) (4)]
Maximum Number of Users for Sunlamp Product:__________
Number of pairs:___________ Model Type:_________________________ Manufacture:_______________________________________
Number of pairs:___________ Model Type:_________________________ Manufacture:_______________________________________
LAMPS IN UNIT [21 CFR 1040.20(d) (1) & (d) (2)] & LAMP COMPATIBLITY [21 CFR 1040.20 (e) 2 (iii)]
Total Number of Lamps in Unit:__________________ Lamp Compatibility Information : YES / NO / N/A
Lamp Model Designation:__________________________________ Number of Lamps: ______ Manufacture:__________________________
Lamp Model Designation:__________________________________ Number of Lamps: ______ Manufacture:__________________________
Facilities Lamp Supplier(s) (name, address, fax & phone #):_________________________________________________________________
TIMER [21 CFR 1040.20 (C)(2)]
Type of Timer: Digital / Electro-mechanical / Spring Wound / Token / Other:________________________________
Timer Capabilities:_________(Minimum Time) _________(Maximum Time) Timer Interval (i.e. 1min increments): ___________________
Timer Interval Compatible with Exposure Schedule: YES / NO, If "NO", Explain:_______________________________________________
Timer Manufacturer Name and Address:_________________________________________________________________________________
Timer Accuracy: 10%: __________min __________sec, 50%: ____________min __________sec, 100%:___________min ___________sec
(Note: Record Timer Accuracy in minutes and seconds for 10%, 50% and 100% of Maximum Timer Capability for the Sunlamp Product. Remote timers are acceptable provided all other requirements of (C)(2)/(3) are maintained.)
TERMINATION CONTROL [21 CFR 1040.20 (C)(3)]
Presence: YES / NO Description: Toggle / Push Pull / Push Button / Other:_________________________________
How is exposure re-initiated:________________________________________________________________________
USER INSTRUCTIONS [21 CFR 1040.20 (e) (1)] (i.e. owner manual / operator manual)
Provided by the Manufacturer: YES /NO, Available to Patrons: YES / NO, Contains Instructions To Determine Exposure Schedule and Skin Types: YES / NO, Contains Reproduction of "WARNING LABEL" : YES / NO, Contains Instructions for Obtaining Replacement
Parts and Repairs: YES / NO, If "NO" to any, Explain:___________________________________________________________________
_____________________________ |
|
____________________________________ |
INSPECTIONAL CHECKLIST REPORT
FOR SUNLAMP PRODUCTS MANUFACTURED PRIOR TO SEPTEMBER 8, 1986
(Including Pertinent Parts of the Regulation)
Facility Name: |
_______________________________________ |
|
Person Interviewed |
_______________________________________ |
Address: |
_______________________________________ |
|
Telephone |
( ________ ) ____________________________ |
|
_______________________________________ |
|
Field Test Date |
_______________________________________ |
Mfr. Name |
_______________________________________ |
|
Address: |
_______________________________________ |
Home District_________ CFN/FEI_________Product Type:________________________________________
Model_________Serial Number Date_________ Manufactured _________ / _________ / _________
Lamps: UV-A_____UV-B _____HID _____Properly labeled _____Mfr/Model:_______________________________________
Max Timer Setting _____Gradations_____Consistent w/exposure schedule:__________________________________
Timer Exceed Max. Recom. Exp._________Accuracy @ 10%___________ 50%___________ 100%___________
Type of Timer___________ (e.g. Token) Mfr. of Timer_______________How can user terminate exposure? ______________
How is exposure re-initiated? ____________________________________ Eyewear_________Sufficient #_________
Labeling visible w/eyewear ____ Eyewear Mfr. and Model_____________________________________________________
Certification Label: _______ (Va___________)Permanently affixed___________ Viewable_____________________
Location_____________________ Properly Worded____ Mfr. I.D. Label_______________________ Viewable_________
Full Name/Address_____________________ Date Mfrd.___________ Place Mfrd.________________
Warning Label: Readily Viewable_______ Location_____________________ Danger Statement ___________
Lamp Type___________ Min. exposure distance_______ How measured _____Warning: Min. exposure distance_________
Warning: Protective Eyewear _________ Warning: Max. exposure time_________ Exposure Schedule __________________
Time before results can be expected_______ Any misleading statements?__________________________________________
User's Instructions: Provided by the Mfr. ______ Available to patrons ____________________________________________
Contains copy of warning label_______ Instructions for replacement parts _______ Statement of # of people/eyewear _______
Equipment Recommendations: User position indicated _____ Timer error less than 10% ________ Temperature Control ________
Electrical Safety_________ Mechanical Safety_________ Protection from Lamps______ Access and Support ______
|
_____________________________ |
|
____________________________________ |
The Radiation Safety Performance Standard for Cabinet X ray Systems [Title 21 CFR § 1020.40] (performance standard) was designed to protect the public and system operators from unnecessary radiation hazards associated with the use of cabinet x-ray systems. The performance standard sets an exposure emission limit of 0.5 milliRoentgen (mR) in one hour for radiation emitted from a cabinet x ray system. Additional required safety features include interlocks, indicator lights, and warning labels. The performance standard applies to all cabinet x ray systems manufactured or assembled on or after April 10, 1975. Requirements regarding x ray systems designed primarily for the inspection of carry on airline baggage apply to systems manufactured or assembled on or after April 25, 1974.
The potential risk from a cabinet x-ray system is dependent on the maximum power that can be delivered to the x-ray tube and the environment in which the system is used. A cabinet x-ray system that can operate at higher peak tube potential and tube current will present a greater potential risk when compared with a lower power cabinet x-ray system. The following is an example of how the use environment affects the potential risk: a cabinet x-ray system used for checking circuit board quality is integrated into an automated production line and very rarely approached by anyone poses a lower potential risk than a carry on baggage security x-ray system which is loaded by members of the public and always has an operator present in close proximity.
Follow the general guidance on inspection, investigation, and field test priorities provided in section II.B.3 above and use your discretion based on the preceding discussion of potential risk. An example inspection checklist of cabinet x-ray specific issues has been included. For further guidance on compliance with specific requirements of the performance standard see the Cabinet X-Ray Compliance Guide (see reference below).
Radiological Health Specialists have been specifically trained in general EPRC requirements and also have specialized training in the cabinet x-ray product performance standards. These specialists should perform cabinet x-ray inspections and field tests, and may train additional field staff or accompany a medical device investigator to conduct joint EPRC/medical device inspections.
When conducting a cabinet x-ray system manufacturer inspection or field test all FDA personnel are required to wear a personal radiation monitor. If you do not have a personal radiation monitor badge, follow the instructions as noted in Part II of this program.
CDRH is responsible for all administrative/regulatory action, regulatory follow up, and for the issuance of all notices of violations to manufacturers of cabinet x-ray systems.
Generally cabinet x-ray field tests should be performed when requested by CDRH, in response to requests from other federal agencies, to check the validity of a trade or consumer complaint, or when it is necessary for confirmation that a manufacturer’s testing program or corrective action plan is adequate.
When performing a cabinet x-ray field test collect data in accordance with the written procedures prescribed in “Routine Compliance Testing for Cabinet X ray Systems to which 21 CFR Subchapter J is applicable, Dated March 1985” (see reference below). If it is determined that the written procedures cannot be followed, describe in detail the variance from the prescribed procedure in the comments section of the test form.
Field Test Equipment: MDH meters are not sufficiently sensitive to detect radiation emissions from a cabinet x-ray system. Use only the meters identified in the field test procedure identified below.
NOTE: Cabinet X-Ray Systems installed at airports are not to be field tested except as requested by CDRH, Transportation Security Administration (TSA), Customs and Border Protection (CBP), or Department of Agriculture (USDA). Usually there will be a manager from the relevant agency at the facility containing the system to be tested. Coordinate the test with the appropriate agency on-site manager. Where the national radiation safety contacts are known they should also be contacted. The national contacts for TSA and CBP are included below:
Name | Phone | Position | |
---|---|---|---|
Jill Segraves | (571) 227-2292 | Jill.Segraves@dhs.gov | Radiation Safety Program Manager, Transportation Security Administration |
Richard Whitman | (317)614-4843 | richard.t.whitman@dhs.gov | Radiation Safety Officer, Customs and Border Protection |
Results for all field tests of TSA or CBP cabinet x-ray systems should be sent CDRH, the appropriate contact listed above, and the on-site manager.
Frequently Asked Questions on Cabinet X-ray Systems (March 24, 2003)
http://www.fda.gov/cdrh/radhealth/products/cabinetxrayfaq.html
Compliance Guide for Cabinet X-Ray Systems: Coming soon to the web
Routine Compliance Testing for Cabinet X-ray Systems to which 21 CFR Subchapter J is applicable, Dated March 1985
http://www.fda.gov/cdrh/radhlth/pdf/cabgdeft.pdf
Refer to the Cabinet X-Ray Systems main page for additional information:
http://www.fda.gov/cdrh/radhealth/products/cabinetxray.html
Translation of 2-Digit Code |
Product Name |
Product Code |
CFR |
Definition |
|
---|---|---|---|---|---|
Cabinet X-Ray Systems, Non-Medical |
Cabinet X-Ray, Industrial, Non-Medical |
94 |
RCE |
1020.40 |
A cabinet x-ray system used for quality control, non-destructive testing, or some other industrial purpose. |
Cabinet X-Ray Systems, Non-Medical |
Explosive Detection Systems, Cabinet X-Ray Systems, Non-Medical |
94 |
RCF |
1020.40 |
A cabinet x-ray system used for detection of explosives in closed containers such as airline baggage. Usually these systems use a non-standard x-ray mode to perform this function such as computed tomography. |
Cabinet X-Ray Systems, Non-Medical |
Security X-Ray (includes Baggage X-Ray), Cabinet X-Ray Systems, Non-Medical |
94 |
RCG |
1020.40 |
A cabinet x-ray system used to examine the contents of containers such as airline baggage, brief cases, and purses to detect weapons or other contraband. |
Cabinet X-Ray Systems, Non-Medical |
Cargo X-Ray, Cabinet X-Ray Systems, Non-Medical |
94 |
RCH |
1020.40 |
A large cabinet x-ray system used to examine pallets full of cargo to find weapons or other contraband. |
Cabinet X-Ray Systems, Non-Medical |
Other |
94 |
RZZ |
1020.40 |
A cabinet x-ray system used for an unlisted specific purpose. |
Emission Limit |
|||
1020.40(c)(1)(i) |
Exceeds emission limit |
|
|
1020.40(c)(1)(i) |
Radiation emission > 10mR in one hour |
Major |
Class A |
1020.40(c)(1)(i) |
Radiation emission rate ≤ 10 mR in one hour and > 0.5 mR in one hour |
Major |
Class B |
1020.40(c)(1)(ii) |
Emission limit requirements – measurement inadequate |
Major |
See (c)(1)(i) |
Floors |
|||
1020.40(c)(2) |
Floor fails to adequately attenuate radiation emission into occupied area underneath x-ray system |
Major |
See (c)(1)(i) |
Ports and Apertures |
|||
1020.40(c)(3)(i) |
It is possible to reach the primary beam through a port Primary beam greater than 10 R per hour and beam is easy to access Primary beam greater than 10 R per hour and beam is possible but difficult to access inadvertently Primary beam less than 10 R per hour and greater than 5 R per hour Primary beam less than 5 R per hour |
Major Major
Minor Concern |
Class A Class B
Class B Class C |
1020.40(c)(3)(ii) |
Aperture allows human access to interior of cabinet Radiation exposure rate in accessed area greater than 5 R per hour Radiation exposure rate in accessed area less than 5 R per hour |
Major Minor |
Class B Class C |
Safety Interlocks |
|||
1020.40(c)(4)(i) |
Safety interlock - door does not have any interlock and emission rate with door open is > 10mR in one hour |
Major |
Class A |
1020.40(c)(4)(i) |
Safety interlock - door does not have multiple interlocks |
Major |
Class B |
1020.40(c)(4)(i) |
Neither door safety interlock causes physical disconnect |
|
|
1020.40(c)(4)(i) |
Radiation emission rate with interlock failure and door open > 2 mR per hour |
Major |
Class B |
1020.40(c)(4)(i) |
Radiation emission rate with interlock failure and door open ≤ 2 mR per hour and > 0.5 mR in any one hour |
Minor |
Class B |
1020.40(c)(4)(i) |
Safety interlocks - disconnect based on movement other than door |
|
|
1020.40(c)(4)(i) |
Radiation emission rate with interlock failure and door open > 2 mR per hour |
Major |
Class B |
1020.40(c)(4)(i) |
Radiation emission rate with interlock failure and door open ≤ 2 mR per hour and > 0.5 mR in any one hour |
Minor |
Class B |
1020.40(c)(4)(ii) |
Lack of safety interlock - access panel and emission rate with access panel open is > 10 mR in one |
Major |
Class B |
1020.40(c)(4)(iii) |
Safety interlocks - after an interruption reset of the interlock results in resumption of x-ray production |
Major |
Class B |
1020.40(c)(4)(iv) |
Safety interlocks - single component failure disables more than one interlock |
Major |
Class B |
Ground fault |
|||
1020.40(c)(5) |
Ground fault can result in x-ray initiation |
Major |
Class A |
Controls and Indicators |
|||
1020.40(c)(6)(i) |
Key control - not provided |
Major |
Class B |
1020.40(c)(6)(i) |
Key control - not functional |
Major |
Class B |
1020.40(c)(6)(ii) |
Controls to initiate and terminate x-rays other than interlocks or power control are not present |
Major |
Class B |
1020.40(c)(6)(iii) |
Two independent means of Exposure indication at initiation are not present |
Major |
Class B |
1020.40(c)(6)(iii) |
Exposure indication - other than milliammeter is not present |
Major |
Class B |
1020.40(c)(6)(iii) |
Exposure indication at initiation – is not visible from control |
Major |
Class B |
1020.40(c)(6)(iii) |
Multiple failures of exposure indication caused by a single failure |
Major |
Class B |
1020.40(c)(6)(iii) |
Exposure indication - labeling - X-RAY ON is not present |
Concern |
Class C |
1020.40(c)(6)(iii) |
Exposure indication - labeling - x-ray tube current is not present |
Concern |
Class C |
1020.40(c)(6)(iv) |
Exposure indication required to be visible from a door, panel, or port and is not present |
Major |
Class B |
1020.40(c)(6)(iv) |
Exposure indication not visible from each door, panel, or port |
Major |
Class B |
1020.40(c)(6)(iv) |
Exposure indication at door, panel, or port is not labeled - X-RAY ON |
Concern |
Class C |
Additional controls and indicators for systems designed to admit humans |
|||
1020.40(c)(7)(i) |
No means for preventing and terminating x-rays from within |
Major |
Class A |
1020.40(c)(7)(ii) |
X-rays can be initiated from within the cabinet |
Major |
Class A |
1020.40(c)(7)(iii) |
No Pre-exposure warning within cabinet |
Major |
Class A |
1020.40(c)(7)(iii) |
Pre-exposure warning within cabinet – Warning did not activate at least 10 seconds prior to exposure |
Major |
Class A |
1020.40(c)(7)(iii) |
Pre-exposure warning within cabinet - a single failure causes both audible and visual warnings to fail |
Major |
Class A |
1020.40(c)(7)(iv) |
No exposure warning within cabinet |
Major |
Class A |
1020.40(c)(7)(v) |
Lack of signs giving meaning of warning signals |
Major |
Class B |
1020.40(c)(7)(v) |
Lack of signs giving instructions for use of controls to terminate |
Major |
Class B |
1020.40(c)(7)(v) |
Signs are not legible, accessible, illuminated |
Major |
Class B |
Warning Labels |
|||
1020.40(c)(8)(i) |
Lack of Warning labels - X-rays Produced |
Concern |
Class C |
1020.40(c)(8)(ii) |
Lack of Warning labels - Human Access |
Concern |
Class C |
Information to be provided |
|||
1020.40(c)(9)(i) |
Instruction manuals - not provided |
Minor |
Class C |
1020.40(c)(9)(i) |
Instruction manuals - inadequate technical & safety information |
Minor |
Class C |
1020.40(c)(9)(i) |
Assembly instructions - required and not provided |
Major |
Class B |
1020.40(c)(9)(i) |
Assembly instructions - not adequate for compliance |
Major |
Class B |
Additional requirements for systems loaded by the public (e.g. Baggage inspection) |
|||
1020.40(c)(10) |
X-ray baggage inspection systems (public area) - No means to assure operator presence |
Major |
Class A |
1020.40(c)(10)(i) |
No means to terminate exposure |
Major |
Class B |
1020.40(c)(10)(ii) |
No means to terminate an exposure sequence |
Major |
Class B |
Modification of a certified system |
|||
1020.40(d) |
Modification – failure to re-certify and re-identify |
Major |
Class B |
This guidance is in addition to the instruction provided in Part III.A.2 of this program. Refer to the Compliance Guide for Cabinet X-Ray Systems (referenced above) for a detailed discussion of the cabinet x-ray system performance standard.
The cabinet x-ray field test procedure uses an official form to record the data. This form, FDA 2903 entitled, Cabinet X-Ray Systems Field Test Record can be found at the FDA Forms Catalog (see the FDA intranet home page under Medical Devices).
The Television Product Performance Standard (the standard) was designed to protect the public from x-radiation hazards associated with early cathode-ray-tube (CRT) television sets. The radiation emitted from these products has been dramatically reduced over the years as a result of the standard, and by improvements in technology and design. The hazards of x-ray emissions from CRT televisions and video monitors are further diminished because of a well-established and conscientious industry and the increasing market for flat panel LCD and plasma displays that do not pose a radiation hazard. A minimal, but risk-based and continued presence by FDA is needed in the television industry to ensure continued compliance with radiation safety standards so long as there is a market for CRT products. This presence is limited to for-cause manufacturer inspection and laboratory inspection. No field tests are conducted on television products.
Television product manufacturers should be inspected or tested at CDRH direction. Television product manufacturers are all located overseas, and all inspections will require foreign travel. Reasons for manufacturer inspection include:
WEAC laboratory analysts have knowledge of general EPRC requirements and also have specialized training in the television product performance standard. These analysts have experience planning and conducting foreign television manufacturer inspections. WEAC analysts should perform these inspections and field tests and may train additional field staff.
CDRH is responsible for review of television manufacturer inspection observations and initiating administrative or regulatory follow-up.
Performance Standard-Television Products
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfcfr/CFRSearch.cfm?FR=1020.10
Reporting and Compliance Guide for Television Products
http://www.fda.gov/cdrh/radhlth/pdf/tvvrptgd.pdf
Refer to the television products main page for guidance documents and additional information:
http://www.fda.gov/cdrh/radhealth/products/tvvdt.html
Translation of 2-Digit Code | Product Name | Product Code | CFR | Definition | |
---|---|---|---|---|---|
TV Receivers & Products Containing Same | Oscilloscope (Exempted), TV Receivers & Products, Non-Medical | 94 | RAY | 1020.10 | A device that depicts on a screen periodic changes in an electric quantity, as voltage or current, using a cathode ray tube and is not used in a medical application |
TV Receivers & Products Containing Same | Television Receiver, Medical Imaging, Color | 94 | RAZ | 1020.10 | A television receiver using a color cathode ray tube to display medical images in colors. |
TV Receivers & Products Containing Same | Television Receiver, Medical Imaging, Monochrome | 94 | RBA | 1020.10 | A television receiver using a monochrome cathode ray tube to display medical images in black and white with shades of gray or in different shades of one color. |
TV Receivers & Products Containing Same | Television Receiver, General Purpose, Color, Non-Medical | 94 | RBB | 1020.10 | An electronic product with no medical claims designed to receive and, using a color cathode ray tube, to display a television picture in colors from broadcast, cable, video disk player, video recorder, computer or closed circuit television signals. |
TV Receivers & Products Containing Same | Television Receiver, General Purpose, Monochrome, Non-Medical | 94 | RBC | 1020.10 | An electronic product with no medical claims designed to receive and, using a monochrome cathode ray tube, to display a television picture in black and white with shades of gray or in different shades of one color from broadcast, cable, video disk player, video recorder, computer or closed circuit television signals. |
TV Receivers & Products Containing Same | Video Monitor, Medical Imaging, Color | 94 | RBD | 1020.10 | An electronic product using a color cathode ray tube to display medical images in colors from signals from a computer or electronic medical device. |
TV Receivers & Products Containing Same | Video Monitor, General Purpose, Color | 94 | RBE | 1020.10 | An electronic product using a color cathode ray tube to display general images in colors from signals from a computer or electronic medical device. |
TV Receivers & Products Containing Same | Video Monitor, Medical Imaging, Monochrome | 94 | RBF | 1020.10 | An electronic product using a monochrome cathode ray tube to display medical images in black and white with shades of gray or in different shades of one color from signals from a computer or electronic medical device. |
TV Receivers & Products Containing Same | Video Monitor, General Purpose, Monochrome | 94 | RBG | 1020.10 | An electronic product using a monochrome cathode ray tube to display general images in black and white with shades of gray or in different shades of one color from signals from a computer or electronic medical device. |
TV Receivers & Products Containing Same | Projector, TV Receivers & Products | 94 | RBH | 1020.10 | Electronic products that use a cathode ray tube or several cathode ray tubes to generate television images which are projected on a screen either from the front or from the rear. |
TV Receivers & Products Containing Same | TV View Finder, TV Receivers and Products | 94 | RBI | 1020.10 | An electronic product using a cathode ray tube to display the image seen through the lens of a camcorder. To be exempt the cathode ray tube must operate under 5 kilovolts under the test conditions in the standard (Phase III). |
TV Receivers & Products Containing Same | Camera, Television, Surgical, Without Audio | 79 | FWB | 1020.10 | |
TV Receivers & Products Containing Same | Camera, Television, Surgical, With Audio | 79 | FWC | 1020.10 | |
TV Receivers & Products Containing Same | Camera, Television, Microsurgical, Without Audio | 79 | FWD | 1020.10 | |
TV Receivers & Products Containing Same | Camera, Television, Microsurgical, With Audio | 79 | FWE | 1020.10 | |
TV Receivers & Products Containing Same | Camera, Television, Endoscopic, Without Audio | 79 | FWF | 1020.10 | |
TV Receivers & Products Containing Same | Camera, Televsion, Endoscopic, With Audio | 79 | FWG | 1020.10 | |
TV Receivers & Products Containing Same | System, Reading, Television, Closed-Circuit | 79 | HJG | 1020.10 | |
TV Receivers & Products Containing Same | Other | 94 | RZZ | Unknown | Other electronic products using cathode ray tubes to display television images from broadcast, cable, video disk player, video recorder, computer or closed circuit television signals. |
Emission Limit |
|||
1020.10(c) | Exceeds exposure rate limit | ||
1020.10(c)(1) | Radiation emission > 10mR in one hour | Major | Class A |
1020.10(c)(3) | Test conditions are not in accordance with requirements | Minor | Class B |
1020.10(c)(4) | Critical component warning label missing or inadequate | Minor | Class B |
Manufacturer Identification
Manufacturer Name : |
|
Plant Location: |
|
Date(s) of Visit: |
|
FDA Personnel
Name |
Title |
Organization |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Organization Chart |
|
Sampling Procedures |
|
Engineering Test Plan |
|
Service Manual(s) |
|
Incoming Q. C. Test Procedures |
|
Reaction Plan Procedures |
|
Engineering Test Records |
|
Mfr's Agent agreement (21 CFR 1005.25) |
|
Instrument Calibration Control Log |
|
Labels (ID, Cert. and Crit. Comp.) |
|
Vendor Test Data |
|
Other: |
|
X-Radiation Test Record |
|
Production Line Procedures |
|
Manufacturer Distribution Records |
|
|
Specific Area Inspected |
Gen. Eval.* |
See Attach. |
Details on Page |
Specific Area Inspected |
Gen. Eval.* |
See Attach. |
Details on Page |
General Organization |
|
|
|
Production Line Q.C. Procedures |
|
|
|
Engineering Test Plan |
|
|
|
Reaction Plan |
|
|
|
Incoming Materials Testing Program |
|
|
|
X-Radiation Test Records |
|
|
|
Written Comm. Concerning Radiation Safety |
|
|
|
Sampling Procedures for Production Rad. Testing |
|
|
|
Manufacturer Distribution Records |
|
|
|
Radiation Testing Prog. for Production Receivers |
|
|
|
Instrument Calibration |
|
|
|
|
|
|
|
*Legend for Evaluation: A - Satisfactory B - Questionable C - Unsatisfactory
1. PRODUCTION SUMMARY
MAXIMUM NUMBER OF PRODUCTION LINES =
Line Name |
Model No. |
Brand |
Rate (Sets/day) |
Meets Abbr. Rep. Criteria? |
Line Name |
Model No. |
Brand |
Rate (Sets/day) |
Meets Abbr. Rep. Criteria? |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
1. Flowchart of company functions and organization available? |
|
|
|
||||||||||
|
|
Yes |
|
No SeeExhibit: |
|
|
|
|
|||||
2. Corresponding official is : |
|
|
|
|
|||||||||
|
Q.A. |
|
Q.C. |
|
Product Safety |
|
Engineering |
||||||
|
|
Production |
|
Sales |
|
Other: |
|
|
|
|
|||
3. Is the Compliance Testing Program separate from Production? |
|
Yes |
|
No |
|||||||||
4. (Foreign companies only) Does the company have a Manufacture's Agent who lives in the U.S.? (21 CFR 1005.25) |
|||||||||||||
|
|
Yes |
|
|
No |
|
|
|
|
|
1. Test Plan |
||||||||||
a) The receiver selected for the Engineering Analysis is a: |
||||||||||
|
|
Prototype |
|
Preproduction |
|
Other: |
|
|||
b) The engineering x-radiation testing is performed by: |
||||||||||
|
|
Q.C. |
|
Engineering |
|
Other: |
|
|||
c) The acceptance/rejection criteria for new design is: |
|
|
||||||||
d) The A/R decision is made by: |
|
|||||||||
e) Life test prior to mass production? |
|
Yes |
|
No |
||||||
|
|
|
|
|
||||||
2. Engineering Test Records |
||||||||||
a) Are records kept? |
||||||||||
|
|
Yes, where? |
|
|||||||
|
|
No (Explain) |
|
|||||||
b) Type of information kept on record: |
|
|||||||||
c) Is the worst tolerance chassis retained for further testing? |
|
Yes |
|
No |
|
1. Test Summary
Components |
Test Performed |
Sampling Plan |
Rejection Criteria |
Test Method |
|
Yes |
No |
||||
CRTs |
|
|
|
|
|
Capacitors |
|
|
|
|
|
H.V. Transformers |
|
|
|
|
|
Yoke |
|
|
|
|
|
Others |
|
|
|
|
|
|
|
|
|
|
|
|
Yes |
No |
|||||
2. Incoming test records on file? |
|
|
|||||
3. CRTS tested In-House? |
|
|
|||||
If yes, Registered at TEPAC? |
|
|
|||||
a) Explain the CRT test procedure: |
|
|
|
||||
b) X-Radiation Instrumentation used: |
Model |
Cal. Date |
|
||||
|
|
|
|
||||
|
|
|
|
||||
4. If CRTs are tested by vendor does the vendor provide: |
|
|
|||||
a) test data for each lot? |
|
|
|||||
b) general guarantee of Engineering X-Radiation specifications |
|
|
5. INCOMING CHECK OF REQUIRED LABELS |
|
|
1. Are the labels, which are received at the incoming area, checked for compliance with 21 CFR 1010? |
|
|
2. If yes, are the labels compared with approved labels on file? |
|
|
6. COMMUNICATIONS CONCERNING RADIATION SAFETY |
|
|
1. Are records kept? |
|
|
2. Who responds to these questions? |
|
|
7. MANUFACTURER DISTRIBUTION RECORDS |
|
|
1. Are records kept? If Yes, where are they kept?: |
|
|
2. Information kept on record: |
|
|
Dealer/Distributor name and address? |
|
|
Date distributed? |
|
|
Model and serial No.? |
|
|
3. Are records computerized? |
|
|
4. Are dealers/distributors notified of their obligation to obtain and maintain purchaser records? (for non-exempt products) |
|
|
5. Are dealers/distributors notified of the exempt products? |
|
|
|
|
|
8. INSTRUMENT CALIBRATION |
|
|
1. Is the qualitative meter given a periodic (30 day) check for proper operation? |
|
|
2. Are the actual readings for each tube recorded? |
|
|
3. The date of the CST-l source used for the thirty-day check is: |
||
4. Is it adjusted? |
|
|
5. Is the quantitativeinstrument checked to a source traceable to a NBS standard? |
|
|
6. Is there a system for reminding personnel that an instrument is due to be calibrated? |
|
|
7. Are there alternative x-radiation instruments available should the instruments in use require repair or calibration? |
|
|
9. SAMPLING PROCEDURES FOR PRODUCTION RADIATION TESTING |
Yes |
No |
||||||
1. The samples for production testing are selected by: |
|
|||||||
2. From: Each production line? |
|
|
||||||
Each shift? |
|
|
||||||
Each model? |
|
|
||||||
End of production line? |
|
|
||||||
Warehouse? |
|
|
||||||
3. Sample size: |
|
|
||||||
4. Lot size: |
|
|
||||||
5. How determined? |
|
|
||||||
6. Normal amount of production: |
|
|
||||||
7. Rejection criteria: |
||||||||
Unit: |
|
mR/hr |
Lot : |
|
mR/hr |
10. REACTION PLAN UPON REJECTION (review actual rejection cases) |
|
|
||
1. Who is notified by the test technician? |
|
|
|
|
2. Who examines the cause? |
|
|
|
|
3. Disposition of the rejected lot while examining cause: |
|
|
|
|
4. Who issues the order to stop shipment and/or production? |
|
|
|
|
5. Are other lots (previous and/or subsequent) subjected to increased testing? |
|
|
||
6. Have there been any failures? |
|
|
||
If yes,was it documented ? |
|
|
||
7. Does the Reaction plan appear to be adequate? |
|
|
11. X-RADIATION TEST RECORDS
1. Where are records kept? |
|
|
|
|
|
|
2. Are they maintained for five years? |
|
Yes |
|
No |
|
|
3. How are they filed? (model, date, etc.) |
|
|
4. What information is recorded?
|
Model/Chassis |
|
Test Date |
|
Technician |
|
Beam Current |
|
All Sides |
|
Serial # |
|
Fault |
|
High Voltage |
|
X-Radiation |
|
Background |
5. Are any records in excess of the rejection limit?
|
Yes, disposition of rejected units/lots: |
|
|
No |
|
|
|
|
12. PRODUCTION LINE PROCEDURES |
Yes |
No |
1. Shielding |
|
|
a) Is special shielding checked for proper placement? |
|
|
2. Sealed Controls |
|
|
a) Are they checked? |
|
|
b) Checking Method: |
|
Visual |
|
Mechanical |
c) Do seals appear to be permanent? |
|
|
3. Labels |
|
|
a) Is the presence of labels being checked on line? |
|
|
b) Are labels readily viewable? |
|
|
c) Are they permanently affixed? |
|
|
13. PRODUCTION LINE PROCEDURES AND OPERATIONAL SAFETY TESTS |
1) Chassis Number |
|
|
|
|||
Yes |
No |
Yes |
No |
Yes |
No |
|
2) B+ measured? |
|
|
|
|
|
|
% Checked |
% |
% |
% |
|||
Meter Calibration Current? |
|
|
|
|
|
|
Instructions Available? |
|
|
|
|
|
|
3) H.V. measured? |
|
|
|
|
|
|
% Checked |
% |
% |
% |
|||
Meter Calibration Current? |
|
|
|
|
|
|
Instructions Available? |
|
|
|
|
|
|
4) Hold Down/Safety Circuit Subassembly |
|
|
|
|
||
Finished product |
|
|
|
|
||
Instructions available? |
|
|
|
|
|
|
Comments: |
14. RADIATION TESTING PROGRAM FOR PRODUCTION SETS |
1. Test Instrumentation
Instruments |
Manufacturer |
Model |
Calibrated |
Operational Checks |
||
Last |
Due |
Yes |
No |
|||
Qualitative |
Johnson |
TVX-1 |
|
|
|
|
Quantitative |
Victoreen |
440 RF/C |
|
|
|
|
Voltmeter |
|
|
|
|
|
|
Ammeter |
|
|
|
|
|
|
H.V. Meter |
|
|
|
|
|
|
2. Demonstration Test Number 1 |
a) Identification of receiver tested:
Chassis No. |
|
|
|
Color |
|
Black and White |
||
CRT No. |
|
Model No. |
|
|||||
Serial No. |
|
|
|
|||||
Sample selected by: |
|
|||||||
Sample selected from: |
|
b) Labeling Information:
Label |
Viewable |
Obscured |
Missing |
Adhesion |
Certification |
|
|
|
|
Date of manufacturer. |
|
|
|
|
Place of Manufacturer. |
|
|
|
|
Critical Component Warning |
|
|
|
|
c) Test Conditions:
Input voltage: |
|
|
|||||||||
User controls adjusted? |
|
|
Yes |
|
No |
|
|
|
|
||
Service controls adjusted? |
|
Yes |
|
No |
|
|
|
|
|||
List adjusted controls: |
|
|
|||||||||
Describe worst-case failure: |
|
||||||||||
|
|
|
|
|
|
||||||
Usable Picture? |
Yes |
|
No |
|
|||||||
Test pattern: |
|
d) Test Results:
Max. Qualitative: |
|
counts/min at |
|
kV and |
|
A |
|
Location: |
|
Background: |
|
counts/min |
|||
Max. Quantitative: |
|
mR/hr at |
|
kV and |
|
A |
|
Location: |
|
Scan Rate: |
|
|
inches/sec |
||
Comments: |
3. Demonstration Test Number 2
|
a) Identification of receiver tested:
Chassis No. |
|
|
|
Color |
|
Black and White |
||
CRT No. |
|
Model No. |
|
|||||
Serial No. |
|
|
|
|||||
Sample selected by: |
|
|||||||
Sample selected from: |
|
b) Labeling Information:
Label |
Viewable |
Obscured |
Missing |
Adhesion |
Certification |
|
|
|
|
Date of manufacturer. |
|
|
|
|
Place of Manufacturer. |
|
|
|
|
Critical Component Warning |
|
|
|
|
c) Test Conditions:
Input voltage: |
|
|
|||||||||
User controls adjusted? |
|
|
Yes |
|
No |
|
|
|
|
||
Service controls adjusted? |
|
Yes |
|
No |
|
|
|
|
|||
List adjusted controls: |
|
|
|||||||||
Describe worst-case failure: |
|
||||||||||
|
|
|
|
|
|
||||||
Usable Picture? |
Yes |
|
No |
|
|||||||
Test pattern: |
|
d) Test Results:
Max. Qualitative: |
|
counts/min at |
|
kV and |
|
mA |
||
Location: |
|
Background: |
|
counts/min |
||||
Max. Quantitative: |
|
mR/hr at |
|
kV and |
|
mA |
||
Location: |
|
Scan Rate: |
|
|
inches/sec |
|||
Comments: |
The Microwave Oven Product Performance Standard (the standard) was designed to protect the public from unnecessary emissions from microwave ovens. A minimal, but risk-based and continued presence by FDA is needed in the microwave oven industry to ensure continued compliance with radiation safety standards. This presence is limited to for-cause manufacturer inspection and laboratory inspection. No field tests are conducted on microwave oven products.
Microwave oven product manufacturers should be inspected or tested at CDRH direction. Microwave oven product manufacturers are all located overseas, and all inspections will require foreign travel. Reasons for manufacturer inspection include:
WEAC laboratory analysts have knowledge of general EPRC requirements and also have specialized training in the microwave oven product performance standard. These analysts have experience planning and conducting foreign microwave oven manufacturer inspections. WEAC analysts should perform these inspections and field tests and may train additional field staff.
CDRH is responsible for review of microwave oven manufacturer inspection observations and initiating administrative or regulatory follow-up.
Performance Standard-Microwave Oven Products
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfcfr/CFRSearch.cfm?CFRPart=1030&showFR=1
Guide for Preparing Reports on Radiation Safety of Microwave Ovens
http://www.fda.gov/cdrh/radhlth/pdf/mworptgd.pdf
Refer to the microwave oven products main page for guidance documents and additional information:
http://www.fda.gov/cdrh/radhealth/products/microwave.html
Translation of 2-Digit Code | Product Name | Product Code | CFR | Definition | |
---|---|---|---|---|---|
Microwave Ovens (Food Prep) |
Microwave Oven, Consumer (Food Prep) |
96 | RCR | 1030.10 | A machine that utilizes microwave radiation for food preparation, designed for home use. |
Microwave Ovens (Food Prep) |
Microwave Oven, Commercial (Food Prep) |
96 | RCS | 1030.10 | A machine that utilizes microwave radiation for food preparation, designed for commercial establishments |
Microwave Ovens (Food Prep) |
Tunnel/Conveyor, Microwave Ovens (Food Prep) |
96 | RCT | 1030.10 | A machine that utilizes microwave radiation for food preparation using a conveyorized or tunnel microwave waveguide. |
Microwave Ovens (Food Prep) |
Vending Machine, Microwave Ovens (Food Prep) |
96 | RCU | 1030.10 | A machine that utilizes microwave radiation for dispensing heated foods in public areas. |
Microwave Ovens (Food Prep) |
Other | 96 | RZZ | Unknown | A machine that utilizes microwave radiation for food preparation not previously specified. |
Power density limit requirements |
|||
1030.10(c)(1) |
Leakage from door, vents, other seams > 6mW/cm 2 |
Major |
Class A |
1030.10(c)(1) |
Leakage from door, vents, other seams >1.25mW/cm 2, < 6mW/cm 2 |
Minor |
Class B |
1030.10(c)(1) |
Leakage from door, vents, etc. < 6mW/cm 2 after purchase |
Concern |
Class C |
Safety interlocks |
|||
1030.10(c)(2)(i), (iv) |
Does not incorporate two (2) independent safety interlocks or monitor |
Major |
Class A |
1030.10(c)(2)(i) |
No concealed or inaccessible interlock |
Major |
Class A |
1030.10(c)(2)(ii) |
Single mechanical/electrical failure disables interlocks |
Major |
Class A |
1030.10(c)(2)(iii) |
Secondary interlock allows leakage > 6mW/cm 2 |
Major |
Class A |
1030.10(c)(2)(iii) |
Primary interlock allows excess leakage > 6mW/cm 2 |
|
|
1030.10(c)(2)(iv) |
Insulating wire is accessible to energy-containing space Opening is obvious to user Opening is not obvious or readily accessible |
Major Minor |
Class A Class B |
User instructions |
|||
1030.10(c)(4)(ii) |
Precaution statement unclear, not located to elicit attention, not legible or durable, etc. |
Minor |
Class B |
1030.10(c)(4)(iii) |
User manual or cookbook has no precaution statement |
Minor |
Class B |
Service instructions |
|||
1030.10(c)(5)(ii) |
Safety information or precaution statement unclear, not located to elicit attention not legible or durable, etc. |
Minor |
Class B |
1030.10(c)(5)(iii) |
Service instructions have non precaution statement |
Minor |
Class B |
1030.10(c)(5)(iv) |
Service instructions have insufficient safety information |
Major |
Class A |
Warning labels |
|||
1030.10(c)(6)(i), (ii) |
No user warning label or service caution label |
Major |
Class A |
Manufacturer Identification
|
Manufacturer Name : |
|
|
Plant Location: |
|
|
Date(s) of Visit: |
|
F.D.A. Personnel
Name |
Title |
Organization |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Manufacturer Personnel
Name |
Title |
Name |
Title |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
LIST OF EXHIBITS
A |
C |
E |
G |
||||
B |
D |
F |
|
|
GENERAL INSPECTION OVERVIEW
SUMMARY OF FINDINGS (See the FDA483 in Exhibit A)
HISTORY OF BUSINESS
PERSONS INTERVIEWED AND INDIVIDUAL RESPONSIBILITY
FIRM'S TRAINING PROGRAM
RAW MATERIALS AND COMPONENTS
MANUFACTURING PROCEDURES
SAMPLES COLLECTED
Y2K ISSUES
COMPLAINTS
REFUSALS
DISCUSSION WITH MANAGEMENT
1.0 Production Summary Maximum number of production lines is:
Line Name |
Model # |
Brand |
Type* |
Rate |
Shift/Hours |
Comments |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
* CTD = Countertop/Domestic CTC = Countertop/Commercial UTC = Under‑the‑cabinet WHO = Wall hanging COM = Common cavity MOD = Module for High/Low HLO = High/Lo BDO = Built‑in‑double BSO = Built‑in single
2.0 Component Inspection
Components Test Parameters*/Sampling Rate |
|||||||||||||||
2.1 Cavities and Waveguides |
|
/ |
|
|
|
/ |
|
|
|
/ |
|
|
|
/ |
|
2.2 Interlock & Monitor Switches |
|
/ |
|
|
|
/ |
|
|
|
/ |
|
|
|
/ |
|
2.3 Wire Harnesses |
|
/ |
|
|
|
/ |
|
|
|
/ |
|
|
|
/ |
|
2.4 Door Structure, Hinges, Latches |
|
/ |
|
|
|
/ |
|
|
|
/ |
|
|
|
/ |
|
2.5 Door Chokes and Seals |
|
/ |
|
|
|
/ |
|
|
|
/ |
|
|
|
/ |
|
2.6 Door Screen Perforations |
|
/ |
|
|
|
/ |
|
|
|
/ |
|
|
|
/ |
|
2.7 Noncertified MWO Modules |
|
/ |
|
|
|
/ |
|
|
|
/ |
|
|
|
/ |
|
*Test Parameter Keys: D = dimension check, E = electrical continuity or performance, F = function check, RF = RF emission check, V = visual inspection, W = weld integrity
3.0 Component Control
3.1 Are the incoming components adequately controlled to prevent their use until quality control tests are completed and lot acceptability is determined?
Yes ____ No (Explain) ____________________________
3.2 Are the rejected lots of components adequately marked or secured so the rejected parts are not used in production unless reworked?
Yes ____ No (Explain) ____________________________
4.0 Production Line and Final Tests
General Tests |
Line Names /All Lines |
||||||||
|
|
|
|
|
|
|
|
|
|
Door installation & adjust. checks |
|
|
|
|
|
|
|
|
|
Safety interlocks & monitor continuity checks |
|
|
|
|
|
|
|
|
|
RF emission hazard waveguide, cavity seams, etc. |
|
|
|
|
|
|
|
|
|
Check door travel before sec. interlock actuation |
|
|
|
|
|
|
|
|
|
Open door (shut off‑restart) operation test |
|
|
|
|
|
|
|
|
|
Presence and content of required labels |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
RF Emission Tests |
|
|
|
|
|
|
|
|
|
Door viewing screen |
|
|
|
|
|
|
|
|
|
Door perimeter |
|
|
|
|
|
|
|
|
|
Door perimeter ~ door pulled & all interlocks operating |
|
|
|
|
|
|
|
|
|
Door perimeter ~ door pulled & only Secondary interlock operating |
|
|
|
|
|
|
|
|
|
Door hinge |
|
|
|
|
|
|
|
|
|
Control panel |
|
|
|
|
|
|
|
|
|
Vents and Louvers |
|
|
|
|
|
|
|
|
|
Underneath the oven (bottomless or exposed cavity) |
|
|
|
|
|
|
|
|
|
Automated Microwave Scanner |
|
|
|
|
|
|
|
|
|
NP = Not performed, B = Before final assembly, A = After final assembly NA = Not applicable, ND = Not determined |
4.1 Are the written procedures or diagrams available or posted in the working area for the operator performing Q.C. checks?
Yes ____ No (Explain) ____________________________
4.2 Are repaired ovens returned to the assembly line at a point prior to the test that caused their rejection?
Yes ____ No (Explain) ____________________________
4.3 Are all repaired ovens, regardless of the nature of the repair, returned to the assembly line for the open door operation test and final RF emission test?
Yes ____ No (Explain) ____________________________
5.0 Final Test Records (Check information permanently retained) |
|||
Final and highest RF value |
|
Serial no. |
|
Date of Test |
|
Secondary Interlock Only RF |
|
Safety Interlocks/Monitor Continuity |
|
Label check |
|
Scanner Start-up Test |
|
Open Door (Shut Off - Restart) Test |
6.0 Automated Microwave Oven Scanner
Line Name |
AMOS Brand/ Serial No. |
Model Family |
Model Exceptions |
Qualified |
RF Reject Limit |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
* User manual provided to person responsible for operation of AMOS?
____ Yes _____ No
* Maintenance record shows regular and adequate maintenance of the AMOS (cone checks, wires, RF absorbers, etc.)?
____ Yes _____ No
7.0 Microwave Emission Final Test
Line Name |
Number of Testers |
Scan Rate |
Meter Type |
Reject Limit |
Comments on Scan Rate or Scan Pattern |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
general instrumentation :**warm up, **reset zero, **dirty cones, **AC cover missing, **battery check, **voltage supply for AC powered meters, **barrel holding
8.0 Quality Audit
General Tests |
Line Names/ALL Lines/Lab Sampling Rate |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Magnetron/weld RF hazard test |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Continuity check: interlocks, monitor, wiring |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Check door travel before sec. interlock actuation |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Open door (shut off‑restart) operation test |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Presence and content of required labels |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Check for caution statements in User and Service manuals |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Insertion by finger or wire into concealed safety interlock(s) and cavity |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
RF Emission Tests |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Door viewing screen |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Door perimeter |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Door perimeter ~ door pulled & all interlocks operating |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Door perimeter ~ door pulled & only Secondary interlock operating |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Door hinge |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Control panel |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Vents and Louvers |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Underneath the oven (bottomless or exposed cavity) |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Automated Microwave Scanner (Audit rate - manual rescan) |
|
|
|
|
|
|
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
NP = Not performed, NA = Not applicable, ND = Not determined |
8.1 Audit Test Records (Circle information permanently retained) |
|
Final and Highest RF Value |
|
Serial No. |
|
Date of Test |
|
Secondary Interlock Only RF |
|
Safety Interlocks/Monitor Continuity |
|
Label check |
|
Daily Scanner Audit |
|
Open Door (Shut Off - Restart) Test |
8.2 Audit Size and Reaction Plan (review any actual instances of audit failures) |
|||||||
Critical Defects |
Reaction Plan |
Failures? |
Documented? |
||||
|
Excess Emission |
|
Test Entire Lot |
|
Yes |
|
Yes |
|
Interlock/Monitor |
|
Test Days Production |
|
No |
|
No |
|
Open Door Operation |
|
Tighten Sampling |
|
|
|
|
|
Missing Labels/statements |
|
|
|
|
|
|
8.3 Scanner Audit Reaction Plan
Has there been a failure in the scanner audit? (document adequate audit response)
____ No ____ Yes (Explain)________________________________
9.3 Annual Calibration |
Yes |
|
No |
Comments |
Annual calibration of LCR is performed by: |
|
|
|
|
Absolute calibration of LCR is performed annually? |
|
|
|
|
Document shows annual calibration of LCR? |
|
|
|
|
All records restarted after annual calibration of LCR? |
|
|
|
|
Are they using JMI calibration data correctly? |
|
|
|
|
Do they perform absolute. cal. of survey meters every 3 yrs.? |
|
|
|
|
9.4 Repair |
Yes |
|
No |
Comments |
Disposition of defective instruments clearly documented? |
|
|
|
|
Are broken meters segregated and labeled? |
|
|
|
|
If the Narda probe is replaced, are the meter and new probe calibrated together? |
|
|
|
|
|
|
|
|
|
10.0 Record keeping |
Yes |
|
No |
Comments |
Are the results of the quality control tests conducted on the production line kept for a minimum of 1 year after filing the annual report for these records? |
|
|
|
|
Are the quality control audit records, documentation of defective ovens found in audit, and results of audit reaction plan kept for a minimum of five years? |
|
|
|
|
Is a file maintained of all written communications from all sources concerning radiation safety including complaints, investigations, instructions, or explanations affecting the use, repair, adjustment, maintenance or testing? |
|
|
|
|
Is a file maintained of records necessary for the tracing of microwave ovens to distributors, dealers and purchasers? |
|
|
|
|
Have all the dealers and distributors been informed of their obligations to obtain the purchaser information? |
|
|
|
|
Manufacturer can trace shipment to dealers/distributors or purchasers by: |
||||
|
Model Number |
|
|
|
|
Serial No. |
|
|
|
|
Date of Manufacture |
|
|
|
|
Other (Specify): |
|
Updated November 6, 2007