Comments and suggestions may be submitted at any time for Agency consideration to Heather Rosecrans, Center for Devices and Radiological Health (HFZ-404), Food and Drug Administration, 9200 Corporate Blvd., Rockville, MD 20850. Comments may not be acted upon by the Agency until the document is next revised or updated. For questions regarding the use or interpretation of this guidance contact Heather Rosecrans at 240-276-4021.
Additional Copies:
World Wide Web/CDRH home page at http://www.fda.gov/cdrh or CDRH Facts on Demand at 1-800-899-0381 or 301-827-0111, specify number 499 when prompted for the document shelf number.
Guidance for Industry on
General/Specific Intended Use
Introduction
This guidance1 document identifies the general principles that will be considered by the Food and Drug Administration (FDA) in determining when a specific indication for use is reasonably included within a general indication for use of a medical device2 for purposes of determining substantial equivalence under Section 513(f) or Section 520(l) of the Federal Food, Drug and Cosmetic Act (the Act). This guidance is issued in accordance with new Section 513(i)(1)(F) of the Act, which was added by Section 206 of the Food and Drug Administration Modernization Act of 1997 (FDAMA).
There are a number of reasons medical device manufacturers may seek to add a specific indication for use to a general use of a legally marketed predicate device. In some cases, technology may drive a manufacturers decision to request the addition of a specific indication for use; "minor" technological changes to a device may make it more applicable to one specific indication for use and less applicable to other uses. Alternatively, a new competing device may enter the market with a specific claim resulting in a potential loss of market share for the device without that claim. Sometimes the identification of a specific intended use is the result of the evolution of medical practice once a device is marketed. When the medical community adopts a specific indication for use as routine practice, manufacturers and physicians want that specific indication for use to appear on the labeling for both liability and reimbursement purposes.
2 Please note that the addition of a specific use to a device may result in a product that is considered a combination product or otherwise requires input from other FDA Centers as presented in the intercenter agreements effective October 31, 1991. In such cases, regulatory issues not addressed in this document may apply
Purpose
The purpose of this document is to help medical device manufacturers understand the principles used by FDA to determine whether the addition of a specific indication for use to a medical device cleared for marketing with a general indication for use could trigger the need for a PMA. The guidance is intended to help manufacturers answer the following questions: Under what circumstances is a device with a new, specific indication for use likely to be found to be substantially equivalent to a device legally marketed for a general indication for use? Conversely, when does a specific indication for use become a new intended use that requires submission of a PMA to establish the safety and effectiveness of the device? 3
This guidance does not offer a bright line rule to answer these questions. The agency believes it could not formulate such a rule without compromising the ability of FDA reviewers to factor in the important public health and regulatory considerations that are essential to making appropriate classification determinations. Thus, the purpose of this guidance is to describe that decision-making process and its basis in the law and in agency practice.
Nor does this guidance construct a new or separate SE/NSE decision-making process. That process is addressed in other agency guidance (see Blue Book Memorandum #K86-3, "510(k) Substantial Equivalence Decision-Making Process"). Instead, this document provides guidance to sponsors by describing the criteria that FDA considers in deciding whether the addition of a specific indication for use alters the intended use of a product that is already marketed with a general indication for use, requiring approval of a PMA.
Item #4 of the 510(k) flowchart (K86-3), which FDA reviewers have used for years in 510(k) evaluations, asks "Do the differences [ in indications ] alter the intended therapeutic/diagnostic effect" and directs FDA reviewers to "consider impact on safety and effectiveness ." This general/specific guidance, therefore, does not add a new level of scrutiny to the review process; rather, it articulates the factors which are currently used by FDA in assessing the impact of a change from general to specific use on safety and effectiveness.
Definitions
For the purpose of this guidance, the definitions for "general to specific" and "level of specificity" listed below are used.
General to Specific
A change from a general to a specific indication for use is defined as: Any proposed increase in the level of specificity of the indication for use of a medical device. A change in a devices indication for use from general to specific usually results in an indication for use that is narrower than the approved or cleared general use. Such a change or additional indication generally will narrow the indication for use with respect to function, target population, organ or organ system, tissue type, disease entity, or analyte.
Levels of Specificity
The level of specificity is defined as: a qualitative ranking of the proposed indications for use of a medical device. Levels of specificity for diagnostic and therapeutic devices in order of increasing specificity from general to specific can be categorized as follows:
Levels of Specificity for diagnostic medical devices:
Levels of Specificity for therapeutic (including preventive) medical devices:
Regulatory Background
For products not requiring a Premarket Approval that are not exempt from premarket notification ( 510(k) ), a 510(k) submission is required whenever a medical device is introduced into commercial distribution in the United States (21 CFR 807.81). In addition, a 510(k) is required when a legally marketed device is to be significantly modified in design, components, method of manufacture, or intended use ( 21 CFR 807.81(a)(3) ). In either situation, review of the 510(k) submission is how the agencys experts determine whether the device is substantially equivalent (SE) to the predicate device to which it is being compared.
The 510(k) review process requires the agency to determine the proper classification for a medical device. Devices that are determined to be substantially equivalent to legally marketed devices that are not Class III devices subject to premarket approval, are placed in the same regulatory class as the predicate and may go to market upon receiving clearance from FDA. Devices determined to be substantially equivalent to a Class III device subject to premarket approval requirements, as well as devices determined to be not substantially equivalent (NSE) to a predicate device, are placed in Class III. Unless reclassified into Class I or Class II, these devices cannot go to market without an approved PMA or a completed product development protocol. Under new 513(f)(2) of the act, sponsors of devices declared NSE may seek FDA evaluation of their devices automatic Class III designation.
FDA has issued guidance documents that describe the process by which substantial equivalence decisions are rendered. Guidance about the Agencys 510(k) decision making process is contained in Blue Book memorandum #K86-3 (http://www.fda.gov/cdrh/k863.html). Guidance about Agency decisions with respect to the requirement for 510(k) clearance when modifications are made to legally marketed devices are contained in Blue Book memorandum #K97-1 (http://www.fda.gov/cdrh/ode/510kmod.pdf or 510kmod.html). While these guidance documents provide information relevant to FDAs decision-making processes with respect to general/specific use, Congress indicated through FDAMA ยง 206 that FDA should provide additional guidance on the approach that the agency takes when evaluating whether a new indication for use, which appears to fall within the scope of the intended use of a legally marketed predicate device, is a new intended use that would require a PMA.
Blue Book memorandum #K86-3 further states that the Center assesses any differences in indications for use in terms of the safety and effectiveness questions those differences may raise. This guidance also indicates that some modifications in indications will be considered a new use, "even though the intended effect of the new device is very similar to that of the predicate device." The Blue Book memorandum notes that slight modifications in indications for use can significantly change the intended use of the predicate device.
It should also be noted that the agency has received previous Congressional guidance which bears directly on the issue of substantial equivalence in the Report of the Committee on Interstate and Foreign Commerce on the Medical Device Amendments of 1976 (Senate Report):
The committee believes that the term, substantial equivalence, should be construed narrowly where necessary to assure the safety and effectiveness of a device but not narrowly where differences between a new device and a marketed device do not relate to safety and effectiveness.
FDAMA
The Senate committee report which preceded the final FDAMA bill stated that "this clarification [with respect to general/specific use] is important because FDA has not established a consistent pattern upon which persons who submit premarket notifications may rely." (S. Rept. 105-43, at 48 (1997) ). Two specific examples may help us to understand the need to address the general/specific use issue.
The first example, cited in the same 1997 Senate report, relates to a substantial equivalence determination made for condoms, using the general indication for use of prevention of sexually transmitted disease as a predicate for condoms labeled to prevent the transmission of HIV. This was a situation where an overriding potential public health benefit, an established safety profile, and an identical mechanism of action were weighed against concerns regarding the level of available effectiveness data in deciding that 510(k) was the appropriate regulatory pathway for this indication for use. The Senate report concluded that, "[T]his determination made perfect public health sense, despite the fact that the general use labeling pre-dated the Medical Device Amendments of 1976 and HIV was unknown at that time." (S. Rept. 105-43, at 48).
A second example, cited in Blue Book Memorandum #K86-3, relates to powered suction-aspiration devices, which were initially cleared to remove tissue and fluid from the body during surgery. New versions of these devices were presented to the agency to be cleared for use in "suction lipectomy" for body contouring. In this instance, FDA determined that this was a new intended use requiring submission of a PMA. Key factors that led the agency to this decision included:
These major differences in risk, benefit, and clinical endpoints led FDA to conclude that submission of a PMA was necessary to establish whether there was reasonable assurance that the device was safe and effective for this intended use.
Decision-Making Criteria
The criteria that follow are provided as guidance on the Agencys decision-making process for determining substantial equivalence or non-equivalence for general/specific uses. The list of criteria should not be considered to be all-inclusive. Nor should the list be viewed as a scale which can be used to calculate a particular outcome. Rather, these criteria should be seen as important contributing factors, which, when used appropriately, can help the agency consistently arrive at reasonable regulatory decisions that relate to the safety and effectiveness of medical devices. These criteria should be evaluated in connection with the Levels of Specificity4described earlier in this document.
Specific uses that ordinarily fall within a general use for the purpose of determining that the device with the specific indication for use is substantially equivalent to the general use device include:
Specific indications for use that ordinarily fall outside a general use for the purpose of determining substantial equivalence include:
Examples
The following are examples of Agency determinations with respect to 510(k) submissions made for purposes of establishing the substantial equivalence of a new device labeled with the specific indication for use to the legally marketed device labeled for the general use. The Agency believes these examples will help illustrate its consideration and application of the criteria and levels of specificity described above in this memorandum.
Diagnostic Devices
In vitro diagnostics
1. Factor VII Assay
2. IgG Assay for H. pylori
In vivo diagnostics
3. Diagnostic Ultrasound
4. Diagnostic Ultrasound
Therapeutic Devices
5. Cryosurgery in gynecology
6. Radiofrequency devices in urology
7. Percutaneous vascular catheters
8. Excimer Laser
9. CO2 Laser
It should be noted that the vast majority of submissions for 510(k)s for specific indications for use are cleared as substantially equivalent to a legally marketed predicate for the general use. Some of these submissions require clinical data before clearance. The list of examples cited above include some of the relatively few cases that have not been found substantially equivalent in order to illustrate a spectrum of situations where FDA may make a NSE determination.
Conclusion
Determinations of substantial equivalence related to specific versus general indications are often difficult and complex. There are multiple factors that influence those determinations. Those factors vary from device to device and also may change over time with respect to a particular device. This document attempts to elucidate some of those factors in an effort to assist manufacturers in their research and development efforts. Manufacturers may obtain further guidance by communicating directly with the relevant ODE division either through submission of a 510(k) or prior to making a submission.