skip navigational linksDOL Seal - Link to DOL Home Page
Photos representing the workforce - Digital Imagery© copyright 2001 PhotoDisc, Inc.
www.dol.gov

Content Last Revised: 1/19/01
---DISCLAIMER---

CFR  

Code of Federal Regulations Pertaining to U.S. Department of Labor

Title 20  

Employees' Benefits

 

Chapter VI  

Employment Standards Administration, Department of Labor

 

 

Part 718  

Standards for Determining Coal Miners' Total Disability or Death Due to Pneumoconiosis


20 CFR 718 Appendix A - Standards for Administration and Interpretation of Chest Roentgenograms (X-Rays)

  • Section Number: 718 Appendix A
  • Section Name: Standards for Administration and Interpretation of Chest Roentgenograms (X-Rays)

    The following standards are established in accordance with 
sections 402(f)(1)(D) and

413(b) of the Act. They were developed in consultation with the 
National Institute for Occupational Safety and Health. These 
standards are promulgated for the guidance of physicians and medical 
technicians to insure that uniform procedures are used in 
administering and interpreting X-rays and that the best available 
medical evidence will be submitted in connection with a claim for 
black lung benefits. If it is established that one or more standards 
have not been met, the claims adjudicator may consider such fact in 
determining the evidentiary weight to be assigned to the physician's 
report of an X-ray.
    (1) Every chest roentgenogram shall be a single postero-anterior 
projection at full inspiration on a 14 by 17 inch film. Additional 
chest films or views shall be obtained if they are necessary for 
clarification and classification. The film and cassette shall be 
capable of being positioned both vertically and horizontally so that 
the chest roentgenogram will include both apices and costophrenic 
angles. If a miner is too large to permit the above requirements, 
then a projection with minimum loss of costophrenic angle shall be 
made.
    (2) Miners shall be disrobed from the waist up at the time the 
roentgenogram is given. The facility shall provide a dressing area 
and, for those miners who wish to use one, the facility shall 
provide a clean gown. Facilities shall be heated to a comfortable 
temperature.
    (3) Roentgenograms shall be made only with a diagnostic X-ray 
machine having a rotating anode tube with a maximum of a 2 mm source 
(focal spot).
    (4) Except as provided in paragraph (5), roentgenograms shall be 
made with units having generators which comply with the following: 
(a) the generators of existing roentgenographic units acquired by 
the examining facility prior to July 27, 1973, shall have a minimum 
rating of 200 mA at 100 kVp; (b) generators of units acquired 
subsequent to that date shall have a minimum rating of 300 mA at 125 
kVp.

    Note: A generator with a rating of 150 kVp is recommended.

    (5) Roentgenograms made with battery-powered mobile or portable 
equipment shall be made with units having a minimum rating of 100 mA 
at 110 kVp at 500 Hz, or 200 mA at 110 kVp at 60 Hz.
    (6) Capacitor discharge, and field emission units may be used.
    (7) Roentgenograms shall be given only with equipment having a 
beam-limiting device which does not cause large unexposed 
boundaries. The use of such a device shall be discernible from an 
examination of the roentgenogram.
    (8) To insure high quality chest roentgenograms:
    (i) The maximum exposure time shall not exceed \1/20\ of a 
second except that with single phase units with a rating less than 
300 mA at 125 kVp and subjects with chest over 28 cm postero-
anterior, the exposure may be increased to not more than \1/10\ of a 
second;
    (ii) The source or focal spot to film distance shall be at least 
6 feet;
    (iii) Only medium-speed film and medium-speed intensifying 
screens shall be used;
    (iv) Film-screen contact shall be maintained and verified at 6-
month or shorter intervals;
    (v) Intensifying screens shall be inspected at least once a 
month and cleaned when necessary by the method recommended by the 
manufacturer;
    (vi) All intensifying screens in a cassette shall be of the same 
type and made by the same manufacturer;
    (vii) When using over 90 kV, a suitable grid or other means of 
reducing scattered radiation shall be used;
    (viii) The geometry of the radiographic system shall insure that 
the central axis (ray) of the primary beam is perpendicular to the 
plane of the film surface and impinges on the center of the film.
    (9) Radiographic processing:
    (i) Either automatic or manual film processing is acceptable. A 
constant time-temperature technique shall be meticulously employed 
for manual processing.
    (ii) If mineral or other impurities in the processing water 
introduce difficulty in obtaining a high-quality roentgenogram, a 
suitable filter or purification system shall be used.
    (10) Before the miner is advised that the examination is 
concluded, the roentgenogram shall be processed and inspected and 
accepted for quality by the physician, or if the physician is not 
available, acceptance may be made by the radiologic technologist. In 
a case of a substandard roentgenogram, another shall be made 
immediately.
    (11) An electric power supply shall be used which complies with 
the voltage, current, and regulation specified by the manufacturer 
of the machine.
    (12) A densitometric test object may be required on each 
roentgenogram for an objective evaluation of film quality at the 
discretion of the Department of Labor.
    (13) Each roentgenogram made under this Appendix shall be 
permanently and legibly marked with the name and address of the 
facility at which it is made, the miner's DOL claim number, the date 
of the roentgenogram, and left and right side of film. No other 
identifying markings shall be recorded on the roentgenogram.



Phone Numbers