U.S. Food and Drug Administration - CDRH Mobile
Skip NavigationFDA Logo links to FDA home pageCenter for Devices and Radiological Health, U.S. Food and Drug AdministrationHHS Logo links to Department of Health and Human Services website
FDA Home Page | CDRH Home Page | Search | A-Z Index U.S. Food and Drug AdministrationCenter for Devices and Radiological Health Questions?
horizonal rule

Medical Device Recalls
Class I Recall: Boston Scientific Express2TM (bare metal) Coronary Stent

Date Recall
Initiated:

July 16, 2004

[See also July 1, 2004 Recall for Taxus™ Express2™ (paclitaxel-eluting) Coronary Stent]

Product:
Boston Scientific Express2TM (bare metal) coronary stent system
Use:
The Express2™ stent system consists of a stent (small metal tube) which is mounted on a balloon catheter. The stent is inserted into a blood vessel and advanced within the vessel to the narrowed section of the coronary artery. When the stent is correctly positioned, the balloon is inflated, causing the stent to expand. Expansion of the stent pushes the plaque aside, opening the narrowed section of the artery restoring normal blow flow to the heart. The balloon on the stent delivery catheter is then deflated and the delivery catheter is removed from the patient. The stent remains permanently implanted supporting the newly opened section of the vessel.
Recalling Firm:
Boston Scientific Corporation
1 Scimed Place
Maple Grove, MN 55311-1565
Reason for Recall:
Characteristics in the design of this stent system resulted in failure of the balloon to deflate and impeded removal of the balloon after stent placement.
Public Contact:
Boston Scientific Corporation
Tel: 800-832-7822
FDA District:
Minneapolis
FDA Comment:
Impeded balloon deflation can result in significant patient complications, including emergency coronary artery bypass graft surgery and death. Hospitals should immediately discontinue use of any affected units. Class I recalls are the most serious type of recall and involve situations where there is a reasonable probability that use of the product will cause serious injury or death.
 

For additional information on this product recall, see the Boston Scientific press release.

Also, for additional information on this product recall, see the FDA Enforcement Report located at: http://www.fda.gov/bbs/topics/enforce/2004/ENF00856.html. Scroll down to "Recalls and Field Corrections: Devices – Class I".

Updated July 30, 2004

horizonal rule