U.S. Department of Labor | ||||||
Occupational Safety & Health Administration |
Standard Interpretations
02/21/1989 - Metal cyclones Supplied by SKC Inc. are not to be used for sampling respirable dust. |
Standard Interpretations - Table of Contents |
Standard Number: | 1910.1000 |
February 21, 1989 MEMORANDUM FOR: REGIONAL ADMINISTRATORS ATTN: ARA's for Technical Support THRU: LEO CAREY Director Office of Field Programs FROM: EDWARD J. BAIER Director Directorate of Technical Support SUBJECT: Cyclones Supplied by SKC Inc.The cyclones shipped with the SKC medium flow air sampling pumps for respirable dust sampling are made of metal and have been calibrated at a flow rate of 1.9 liters per minute (LPM) which is different from the 1.7 LPM required for the nylon cyclone. SKC indicated that these metal cyclones were purchased from Sweden and it was designed from the collection criteria established by the British Medical Research Council (BMRC). The nylon cyclones supplied to our CSHOs are based on the collection criteria recommended by the American Conference of Governmental Industrial Hygienists (ACGIH). Their major difference is the cut-off size. The nylon cyclone has a 50% collection efficiency at 3.5 micrometers while the SKC cyclone has a 50% collection efficiency at 5.0 micrometers (see attached table for details). Since the two collection efficiency curves are quite different and are not compatible, the SKC cyclones are not to be used for collection of respirable dust. All SKC cyclones should be returned to the OSHA Cincinnati Laboratory for replacement by the nylon cyclones. If your staff is interested in respirable dust sampling, the ACGIH publication "Air Sampling Instruments" has an excellent chapter on this subject (Size-selective Health Hazard Sampling by Morton Lippmann).
Please contact Ching-tsen Bien of my staff at FTS-523-7065 if you have any
questions.
|
Standard Interpretations - Table of Contents |
Back to Top | www.osha.gov | www.dol.gov |
Contact Us | Freedom of Information Act | Customer Survey Privacy and Security Statement | Disclaimers |
||
Occupational Safety & Health Administration 200 Constitution Avenue, NW Washington, DC 20210 |