U.S. Food and Drug Administration - CDRH Mobile
Skip NavigationFDA Logo links to FDA home pageCenter for Devices and Radiological Health, U.S. Food and Drug AdministrationHHS Logo links to Department of Health and Human Services website
FDA Home Page | CDRH Home Page | Search | A-Z Index U.S. Food and Drug AdministrationCenter for Devices and Radiological Health Questions?
horizonal rule

PDF Printer VersionThe HDE Checklist for Filing Decision

Identification:

HDE Number: Date Received: ____________

HUDNUMBER (from OOPD): ___________ Date Received: ___________

Sponsor: ________________________

Device: ________________________

Procode: ____________

Division/Branch: ________________________

Decision:

Recommendation: ___ File ___ Not File

Administrative Reviewer Signature: ________________________ Date: ____________

Supervisory Signature: ________________________ Date: ____________


THE HDA CHECKLIST FOR FILING DECISION

PART A - DEFICIENCIES TO BE INCLUDED AS REASONS FOR NOT-FILING THE HDE

Filing Review Elements

Yes
Present
Omission
Justified
No
Inadequate
Omitted
I. Screening Information    
A. Applicant Information    
  1. Is the name and address of the applicant provided? (21 CFR 814.20(b)(1)) Checkbox-Yes Checkbox-No
  2. Has the applicant or authorized representative signed the HDE? (If applicant does not reside or maintain a place of business in the United States, the application must be countersigned by an authorized representative who does live or maintains a place of business in the U.S. Also the name and address of this person should be included) (21 CFR 814.104(a)) Checkbox-Yes Checkbox-No
B. Certification Information    
 

1. Did the submission include a completed FORM FDA 3674, Certification of Compliance with Requirements of ClinicalTrials.gov Data Bank? (If not, then contact sponsor and request completed form.)

Checkbox-Yes Checkbox-No
  2. Data from Element 9 (check one):
____ No clinical trials referenced in submission.
____ Requirements are not applicable to referenced clinical trials.
____ Requirements are applicable and have been met.
Checkbox-Yes Checkbox-No
C. Are you aware of the applicant being the subject of an integrity investigation? If yes, consult the ODE integrity officer. Has the ODE Integrity Officer given permission to proceed with the review?
(Blue Book Memo #I91-2, 21 CFR 814.42(e)(4) and Federal Register 90N-0332, September 10, 1991)
Checkbox-Yes Checkbox-No
D. Is there a prior history with this device? For example, has a previously submitted PMA/HDE for this device been withdrawn? If yes, does the current HDE address any historical issues related to the reasons for the withdrawal such as fraud or safety? Checkbox-Yes Checkbox-No
II. Organizational and Administrative Elements
(21 CFR 814.20, 21 CFR 814.104)
   
A. Is the HDE sufficiently well organized to permit substantive review? (21 CFR 814.20(b)(2), table of contents, pages numbered, sections identified, six copies, and one copy which identifies trade secret or confidential information) Checkbox-Yes Checkbox-No
B. Is the device appropriate for review as an HDE?    
  1. Is there a copy or reference to the determination made by FDA’s Office of Orphan Products Development that the device qualifies as a HUD? (21 CFR 814.104(c)(1)) Checkbox-Yes Checkbox-No
 

2. Is there an explanation of why the device would not be available unless the HDE were granted?
(21 CFR 814.104(c)(2))

The applicant should provide information such as an estimate of the number of patients who would be required to generate data to support a full PMA and explain why such a study is not feasible or why the cost of conducting such a study could not reasonably be expected to be recovered (64 FR 33234).

 

Checkbox-Yes Checkbox-No
  3. Is there a statement that no other comparable device (other than another HUD approved under this subpart or under an approved IDE) is available to treat or diagnose the disease or condition? (21 CFR 814.104(c)(2)) Checkbox-Yes Checkbox-No
C. Does the HDE contain an explanation of why the probable benefit to health from the use of the device outweighs the risk of injury or illness from its use? (21 CFR 814.104(c)(3)) Checkbox-Yes Checkbox-No
  1. Does the explanation take into account the probable risks and benefits of the currently available devices or alternative forms of treatment? Checkbox-Yes Checkbox-No
  2. Does the explanation include a description, explanation, or theory of the underlying disease process or condition, and known or postulated mechanism(s) of action of the device in relation to the disease process or condition? Checkbox-Yes Checkbox-No
III. Summary of Safety and Probable Benefit
(Blue Book Memo #P86-1, remember an HDE is exempt from the effectiveness requirements of sections 514 and 515 of the act and 21 CFR Part 814)
   
A. Are indications for use provided?
(21 CFR 814.20(b)(3)(i))
Checkbox-Yes Checkbox-No
B. Is an abbreviated device description provided?
(21 CFR 814.20(b)(3)(ii))
Checkbox-Yes Checkbox-No
C. Have alternative practices and procedures been included and described? (21 CFR 814.20(b)(3)(iii)) Checkbox-Yes Checkbox-No
D. Is a description of the prior marketing history provided? (814.20(b)(3)(iv)) Checkbox-Yes Checkbox-No
E. Is a summary of any study(ies) that may have been conducted provided? (21 CFR 814.20(b)(3)(v) and (vi) and 21 CFR 814.104(c)(4)(i))    
1. Is a summary of the non-clinical laboratory studies and results provided? (21 CFR 814.20(b)(3)(v)(A)) Checkbox-Yes Checkbox-No
2. Does the HDE include summaries, conclusions, and results of all clinical experience or investigations (whether adverse or supportive) reasonably obtainable by the applicant that are relevant to an assessment of the risks and probable benefits of the device? (21 CFR 814.104(c)(4)(i)) Checkbox-Yes Checkbox-No
IV. Technical Information
A. Device Characteristics and Manufacturing Section    
  1. Is a description of the device, pictorial representations, and materials specifications present, including the functional component(s) or ingredients? (21 CFR 814.20(b)(4)(i) and (ii)) Checkbox-Yes Checkbox-No
  2. Is a description of the principle of operation of the device (including components) and properties of the device relevant to clinical function present? (21 CFR 814.20(b)(4)(iii) and (iv)) Checkbox-Yes Checkbox-No
  3. The Manufacturing Section may be waived for filing purposes and submitted later during the substantive review period; OCS reviews for GMP issues; ODE reviews for safety issues.    
  Has a description of the methods, facilities, and controls used in the manufacture, processing, packing, storage, and installation of the device been provided? (21 CFR 814.20(b)(4)(v)and Guidance for the Preparation of PMA Manufacturing Information) Checkbox-Yes Checkbox-No
  4. Performance Standards and FDA Guidance/ Guidelines: Has the applicant provided documentation to establish conformance with applicable standards and/or FDA guidance/ guidelines that are relevant to the safety of the device and that is known to or should reasonably be known to the applicant?
(21 CFR 814.20(b)(5))
   
B. Nonclinical Laboratory Studies - as appropriate, are the following provided? (21 CFR 814.20(b)(6)(i))    
  1. Biocompatibility Checkbox-Yes Checkbox-No
  2. Microbiological Checkbox-Yes Checkbox-No
  3. Mechanical (wear, stress, etc.) Checkbox-Yes Checkbox-No
  4. Shelf life Checkbox-Yes Checkbox-No
  5. Analytical (for IVDs) Checkbox-Yes Checkbox-No
  6. Animal testing for safety-if appropriate, material specific and/or device specific Checkbox-Yes Checkbox-No
  7. A statement whether the nonclinical laboratory tests comply with the GLP regulation (Part 58), and if not, a brief statement of the reason for the noncompliance Checkbox-Yes Checkbox-No
C. Clinical Experience and/or Investigations (21 CFR 814.104(c)(4)(i))
  1. Summaries, conclusions, and results of all clinical experience or investigations (whether adverse or supportive), reasonably obtainable by the applicant, that are relevant to an assessment of the risks and probable benefits of the device Checkbox-Yes Checkbox-No
  2. If a clinical investigation has been conducted to demonstrate safety, have the following been provided? (Note: Under 21 CFR 814.104(c)(4)(i), an HDE application is only required to contain the summaries, conclusion, and results listed above. Therefore, if the information listed below is not included, this is not a basis for refusing to file the application.)    
 
a. a description of the protocol, subject inclusion/exclusion criteria, study period, and clinically significant safety endpoints
Checkbox-Yes Checkbox-No
 
b. a description of study population including number of patients, device design used (if more than one), follow-up period, and demographics including gender considerations
Checkbox-Yes Checkbox-No
 
c. a summary or data from subject evaluations and description of adverse events, e.g. adverse reactions, complications, discontinuations, complaints, failures, replacements, etc.
Checkbox-Yes Checkbox-No
 
d. if needed, a statistical analyses of the safety data (does OST Statistician recommend filing?)
Checkbox-Yes Checkbox-No
 
e. report forms for patients who died or were discontinued
Checkbox-Yes Checkbox-No
  3. If there is an IDE for the device, has the data presented in the HDE taken into account any safety concerns raised in the IDE? Checkbox-Yes Checkbox-No
  4. If foreign clinical data are included, are data acceptable?
(21 CFR 814.15(b) and 814.15(d))
Checkbox-Yes Checkbox-No
  5. Has the applicant submitted the financial certification and/or disclosure statements as required by 21 CFR Part 54? Checkbox-Yes Checkbox-No
D. Labeling (21 CFR 814.20(b)(10); 21 CFR 814.104(c)(4)(ii) and Blue Book #P91-4)    
  1. Has appropriate draft labeling been submitted (e.g., Physician, Patient, Technical, etc.)? Checkbox-Yes Checkbox-No
  2. Does the labeling include the statement: "Humanitarian Device. Authorized by Federal law for use in the [treatment or diagnosis] of [specify disease or condition]. The effectiveness of this device for this use has not been demonstrated"? Checkbox-Yes Checkbox-No
E. Amount Charged - Does the HDE include the amount to be charged for the device and, if the amount charged is more than $250.00, a report by an independent certified public accountant or an attestation by a responsible individual of the organization, verifying that the amount charged does not exceed the costs of the device’s research, development, fabrication, and distribution? (21 CFR 814.104(c)(5)) Checkbox-Yes Checkbox-No
V.

Is there any other reason not addressed above which should be identified as a reason for not filing the HDE? If so, briefly explain.

_______________________________________________________________

_______________________________________________________________

Checkbox-Yes Checkbox-No

 

PART B - DEFICIENCIES TO BE INCLUDED IN THE "MINOR" SECTION OF THE NOT-FILING OR FILING BOILERPLATE LETTER

Additional Filing Review Elements

Yes No
VI.. Additional Administrative,Organizational and Regulatory Elements
A. Do we need a device sample? If yes, has it been provided? (21 CFR 814.20(b)(9)) Checkbox-Yes Checkbox-No
B. Is a bibliography provided? (21 CFR 814.20(b)(8)(i))

Have copies of key articles been provided and are English translations included, if appropriate?

Checkbox-Yes

Checkbox-Yes

Checkbox-No

Checkbox-No

C. If there are color additive considerations, has an attempt been made to document them?
(21 CFR 814.20(f))
Checkbox-Yes Checkbox-No
D. Is reference to applicable IDE(s) given?
IDE# ___________________
Checkbox-Yes Checkbox-No
 

PART C – ADDITIONAL CONSIDERATIONS

VII. Additional Considerations Yes No
A.

Are there any special administrative issues? If so, explain. _______________________________

__________________________________________________________

Checkbox-Yes Checkbox-No
B.

Are there any precedent setting substantive issues? If so, explain. ___________________________

________________________________________________________

Checkbox-Yes Checkbox-No

Draft Date: 24-APR-2000
Rev: 20-DEC-2007

Updated December 26, 2007

horizonal rule

CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page

Center for Devices and Radiological Health / CDRH