![]() ![]() ![]() ![]() |
FDA Home Page | CDRH Home Page | Search | A-Z Index | ![]() |
|
![]() |
The draft of this document was issued on August 30, 2004
For questions regarding this document contact Jay A. Rachlin, Office of Communication, Education, and Radiation Programs, 301-594-3173, jay.rachlin@fda.hhs.gov or Joan Ferlo Todd, Office of Surveillance and Biometrics, 301-594-3174, joan.todd@fda.hhs.gov.
![]() |
U.S. Department Of Health And Human Services Division of Device User Programs and Systems Analysis (HFZ-230) Division of Postmarket Surveillance (HFZ-520) |
Written comments and suggestions may be submitted at any time for Agency consideration to the Division of Dockets Management, Food and Drug Administration, 5630 Fishers Lane, Room 1061, (HFA-305), Rockville, MD, 20852. Alternatively, electronic comments may be submitted to http://www.fda.gov/dockets/ecomments. When submitting comments, please refer to Docket No. 2004D-0343. Comments may not be acted upon by the Agency until the document is next revised or updated.
Additional copies are available from the Internet at http://www.fda.gov/cdrh/beds, or you may either send a fax request to (301) 443-8818 to receive a hard copy of the document, or send an e-mail request to GWA@CDRH.FDA.GOV to request hard or electronic copy. Please use the document number 1537 to identify the guidance you are requesting.
Introduction
The Least Burdensome Approach
Background
Organization of this Guidance
Scope
Reporting an Entrapment Adverse Event
Key Body Parts at Risk
Potential Zones of Entrapment
A Retrospective Study of Entrapment Reports to FDA
Dimensional Limits for Identified Entrapment Zones 1-4
Zone 1 - Within the Rail
Zone 2 - Under the Rail, Between the Rail Supports or Next to a Single Rail Support
Zone 3 - Between the Rail and the Mattress
Zone 4 - Under the Rail at the Ends of the Rail
Zone 5 - Between Split Bed Rails
Zone 6 - Between the End of the Rail and the Side Edge of the Head or Foot Board
Zone 7 - Between the Head or Foot Board and the End of the Mattress
APPENDIX A: List of Hospital Bed Safety Workgroup
(HBSW) Participating Organizations
APPENDIX B: Additional Information and Kit Information
APPENDIX C: References for National and International Entrapment
Standards
APPENDIX D: Anthropometric References
APPENDIX E: Drawings of Potential Entrapment in Hospital Beds
APPENDIX F: HBSW Dimensional Test Methods for Bed Systems
This guidance represents the Food and Drug Administration's (FDA's) current thinking on this topic. It does not create or confer any rights for or on any person and does not operate to bind FDA or the public. You can use an alternative approach if the approach satisfies the requirements of the applicable statutes and regulations. If you want to discuss an alternative approach, contact the FDA staff responsible for implementing this guidance. If you cannot identify the appropriate FDA staff, call the appropriate number listed on the title page of this guidance. |
This guidance provides recommendations relating to hospital beds1 and hospital bed accessories. The guidance provides recommendations intended to reduce life-threatening entrapments associated with hospital bed systems.2 It characterizes the body parts at risk for entrapment, identifies the locations of hospital bed openings that are potential entrapment areas, and recommends dimensional criteria for these devices.
Manufacturers may use this guidance when designing new beds to help ensure compliance with applicable FDA regulations such as the Quality System regulation3, to provide adequate labeling and instructions for use, and to assist in ensuring that their devices are safe when used as labeled. FDA encourages manufacturers to provide information to medical device distributors, clinicians, patients and families regarding mattress dimensions, compatible components, and issues of entrapment. In addition, the recommendations in this guidance may be useful in evaluating and reducing the entrapment risk presented by hospital beds that have been placed into use, also known as legacy beds.
Not all patients are at risk for an entrapment, and not all hospital beds pose a risk of entrapment. We suggest that facilities and manufacturers determine the level of risk for entrapment and take steps to mitigate the risk. Evaluating the dimensional limits of the gaps in hospital beds is one component of an overall assessment and mitigation strategy to reduce entrapment. As a result, healthcare facilities may use this guidance as part of a bed safety program to help identify entrapment risks that may exist with current hospital bed systems. Other educational documents are available to assist health care facilities in making decisions towards the goal of achieving a safe and comfortable sleeping environment for their patients.
FDA's guidance documents, including this guidance, do not establish legally enforceable responsibilities. Instead, guidances describe the Agency's current thinking on a topic and should be viewed only as recommendations, unless specific regulatory or statutory requirements are cited. The use of the word should in Agency guidances means that something is suggested or recommended, but not required.
We believe we should consider the least burdensome approach in all areas of medical device regulation. This guidance reflects our careful review of the relevant scientific and legal requirements and what we believe is the least burdensome way for you to comply with those requirements. However, if you believe that an alternative approach would be less burdensome, please contact us so we can consider your point of view. You may send your written comments to the contact person listed in the preface to this guidance or to the CDRH Ombudsman. Comprehensive information on CDRH's Ombudsman, including ways to contact him, can be found on the Internet at http://www.fda.gov/cdrh/ombudsman/.
For 20 years, FDA has received reports4 ,5 in which vulnerable patients6 ,7 have become entrapped in hospital beds while undergoing care and treatment in health care facilities. The term "entrapment" describes an event in which a patient/resident is caught, trapped, or entangled in the space in or about the bed rail8, mattress, or hospital bed frame. Patient entrapments may result in deaths and serious injuries.
FDA received approximately 691 entrapment reports over a period of 21 years from January 1, 1985, to January 1, 20069. In these reports, 413 people died, 120 were injured, and 158 were near-miss events with no serious injury as a result of intervention. These entrapment events have occurred in openings within the bed rails, between the bed rails and mattresses, under bed rails, between split rails, and between the bed rails and head or foot boards. The population most vulnerable to entrapment are elderly patients and residents, especially those who are frail, confused, restless, or who have uncontrolled body movement. Entrapments have occurred in a variety of patient care settings, including hospitals, nursing homes, and private homes. Long-term care facilities reported the majority of the entrapments.
In response to continued reports of patient entrapment, the FDA, in partnership with the U.S. Department of Veterans Affairs, Health Canada’s Medical Devices Bureau and representatives from national health care organizations and provider groups, patient advocacy groups, and medical bed and equipment manufacturers, formed a working group in 1999 known as the Hospital Bed Safety Workgroup (HBSW). Appendix A contains a list of HBSW participating organizations. The HBSW also worked in cooperation with the Joint Commission on Accreditation of Healthcare Organizations, the U.S. Centers for Medicare and Medicaid Services, and the U.S. Consumer Product Safety Commission to improve patient safety associated with the use of hospital beds.
The HBSW identified 7 potential entrapment zones (See Potential Zones of Entrapment, Page 13) in hospital beds. The workgroup then developed (1) educational materials regarding entrapment associated with hospital beds, (2) clinical practice guidelines to reduce the occurrence of patient entrapment, (3) evidence-based dimensional guidelines for hospital beds, (4) test tools and methods to assess gaps in hospital bed systems, and (5) information to assist in mitigating entrapment risks in currently used hospital beds. See Appendix B for information on the availability of these materials.
Consistent with 21 CFR 10.115(f), the HBSW submitted to FDA recommendations for hospital bed dimensional criteria for consideration as an FDA guidance document. Members of the HBSW have developed procedures for measuring and assessing gaps in hospital bed systems in accordance with these criteria. Healthcare personnel, through participation in HBSW testing, validated the test methods and tool. The FDA has considered these HBSW recommendations in preparing this guidance.
Standards and Future HarmonizationThe International Electrotechnical Commission (IEC) issued an internationally recognized standard that applies to a certain segment of the products addressed in this guidance, products labeled as “powered hospital beds.” This standard is the IEC 60601-2-38, Amendment 110. The current IEC standard recognizes that the bed frame, deck, and rails are the major elements involved in entrapment, but does not include the mattress as a contributor or mitigator and sets dimensions for new beds only. The standard also does not address safety issues associated with the use of non-electric hospital beds or the use of hospital beds in the home or in long-term care settings. The IEC standard is currently undergoing revision and will likely undergo significant change prior to its expected publication in 2007/2008. The recommendations in this FDA Guidance are similar to the IEC standard for some entrapment zones, but differ in other zones to include consideration of the mattress as part of the entrapment risk. The IEC test methods are not readily applicable for use by health care providers and are written primarily for test labs. Upon completion of the revised international bed standard, FDA will consider whether harmonization of the FDA guidance and the IEC standard is appropriate. See Appendix C for a list of National and International Entrapment Standards.
This guidance:
The goal of the FDA Hospital Bed System Dimensional and Assessment Guidance to Reduce Entrapment is to reduce potential life-threatening entrapments associated with hospital bed systems. Because hospital bed systems that are primarily intended for one type of care setting can be moved into other care settings, the recommendations in this guidance may be appropriate for hospital beds regardless of the healthcare setting. The dimensional limits in this guidance may not reduce entrapments in all populations, e.g. pediatrics and developmentally disabled people. See the narrative below in the exclusions section.
FDA uses the term "hospital bed" in this guidance to refer to a variety of medical devices that are classified as beds. FDA regulations classify hospital beds as Class I and Class II devices (see Table 1). These devices are used for patients in acute care, long-term care, or home care settings. FDA considers stretchers that are used for extended stay in health care facilities (because they are used like hospital beds) as hospital beds for purposes of this guidance. As stated above, the term “hospital bed system” used throughout this document encompasses the bed frame and its components, including the mattress, bed side rails, head and foot board, and any accessories added to the bed. Bed rails (see footnote 8), also called “side rails,” may be an integral part of the bed frame or they may be removable and at times are used either as a restraint, a reminder or an assistive device. Bed rails may consist of one full-length rail per side or one or more, shorter rails per side, may be a fixed height or adjustable in height, and may move as the head or foot sections of the bed are raised or lowered.
This guidance provides recommendations related to devices in the following table (Table 1). Class II devices are subject to design controls under the Quality System regulation. For those beds in class II, we recommend that manufacturers consider the recommendations in this guidance when developing their design controls. The recommendations in this guidance may also assist manufacturers of all devices listed in Table 1 below in manufacturing hospital beds that will minimize the risk for patient entrapment.
Product Code |
CFR Section |
Classification Name |
Class |
510(k) Exempt |
---|---|---|---|---|
FMR |
880.6785 |
Manual patient transfer device |
I |
Yes |
FNJ |
880.5120 |
Manual adjustable hospital bed |
I |
Yes |
FNK |
880.5110 |
Hydraulic adjustable hospital bed |
I |
Yes |
FNL |
880.5100 |
AC-powered adjustable hospital bed |
II |
Yes |
FPO |
880.6910 |
Wheeled stretcher* |
II |
Yes |
IKZ |
890.5225 |
Powered patient rotation bed |
II |
Yes |
ILK |
890.5150 |
Powered patient transport* |
II |
No |
INK |
890.3690 |
Powered wheeled stretcher* |
II |
No |
INY |
890.5180 |
Manual patient rotation bed |
I |
Yes |
IOQ |
890.5170 |
Powered flotation therapy bed |
II |
Yes |
*When labeled for extended-stay use
The recommendations in this guidance may also be useful in evaluating and reducing the risk of entrapment presented by the devices listed in Table 1 that have already been manufactured and installed (legacy beds). FDA recognizes that legacy beds have the potential for dimensional change over time through wear and tear or substitution of new mattresses and other components not contemplated in the original bed system. Because of these factors, FDA does not intend to take enforcement actions for failure to submit reports of corrections and removals under 21 CFR Part 806 for actions taken in response to this guidance that correct or improve hospital beds currently in use or held as inventory. However, manufacturers and importers should create and maintain records of these corrections and removals in accordance with 21 CFR Part 806.
Some states may have adopted special requirements for the design of hospital beds. For example, New York has adopted more stringent dimensional limits for hospital beds and twin beds used for developmentally disabled people.11 The dimensions used in the FDA guidance are not intended to replace state law requirements that may exist. Moreover, some hospital beds excluded from the scope of the recommendations in this guidance may still be subject to dimensional limits under state laws.
The movement of the bed deck is known as articulation. FDA recognizes that articulation of the bed introduces complex geometries that make applying the dimensional criteria to reduce entrapment difficult. Presently, the dimensional recommendations in this guidance apply to hospital beds in the flat deck position and rails in the fully raised position, except where noted. FDA recognizes that patient care also occurs while the bed is articulated and some articulated positions may pose a risk of entrapment. The Agency will continue to gather data and consider other approaches for assessing gaps in articulated beds in the future. We recommend that patient assessment procedures be used to assess the risk of entrapment when clinical care is provided in an articulated position.
Legacy BedsThe issue of patient entrapment in hospital beds is complex and affects manufacturers, healthcare practitioners and facilities, medical equipment suppliers, home health agencies, patients, and caregivers. Many beds currently in use may no longer have the original mattress or bed rails, and may present an entrapment hazard by increasing or creating gaps or spaces between various components of the bed system. The recommendations in this guidance therefore apply to most hospital beds, including legacy beds in various settings where health care is provided. As described above, a new hospital bed becomes a “legacy bed” as soon as it is placed into use.
Reducing the risk of entrapment involves a multi-faceted approach that includes bed design, clinical assessment and monitoring, as well as meeting patient, resident, and family needs for vulnerable patients in most health care settings - hospitals, long term care facilities, and at home. Therefore, comprehensive bed safety programs in these settings will likely involve input from manufacturers as well as facility staff. Recognizing that not all hospital beds present a risk of entrapment, and that this risk may vary depending on the patient, FDA encourages manufacturers and facilities to work together to develop bed safety programs to evaluate and, if needed, mitigate entrapment risk.
When evaluating the safe use of a hospital bed, component or accessory, manufacturers and caregivers should recognize that the risk for entrapment may increase if a hospital bed system is used for purposes, or used in a care setting, not intended by the manufacturer. Evaluating the dimensional limits of gaps in hospital beds may be one component of a bed safety program which includes a comprehensive plan for patient and bed assessment. Bed safety programs may also include plans for the reassessment of hospital bed systems. Reassessment may be appropriate when (1) there is reason to believe that some components are worn (e.g., rails wobble, rails have been damaged, mattresses are softer) and could cause increased spaces within the bed system, (2) when accessories such as mattress overlays or positioning poles are added or removed, or (3) when components of the bed system are changed or replaced (e.g., new bed rails or mattresses).
FDA recommends that healthcare facilities conduct a risk-benefit analysis to ensure that steps taken to mitigate the risk of entrapment do not create different, unintended risks or reduce clinical benefits available to patients using legacy beds. Such steps may include checking with bed system manufacturers to identify compatible mattresses, rails, and accessories. Likewise, healthcare facilities may contact their equipment suppliers for entrapment mitigating solutions that may already be available. Manufacturers should follow the appropriate FDA regulations regarding risk assessment, design controls, and documentation when developing mitigation components, attachments and other accessories to address entrapment risks in legacy hospital beds.
Additional resources are available to help caregivers and health care providers assess the individual patient's needs, consider and address entrapment risks, and recommend mitigation strategies. See Appendix B for examples of additional resources.
The dimensional criteria described on pages 15-21 are recommended for a number of products, but not all types of hospital beds. We are listing those products below for which some, or all, of the dimensional criteria are not recommended. You can find more detailed descriptions of these products in 21 CFR Parts 880 and 890. Please note that the products listed below are not free from risk of entrapment. We continue to suggest that users identify and address areas of potential entrapment for each patient or resident through a comprehensive bed safety program.
Total exclusion from the scope of this guidance:
Partial Exclusion from the scope of this guidance:
Additional caution should be taken when using these products to ensure a tight fit of the mattress to the bed system. If a powered air mattress is replacing a mattress on a bed system that meets the recommendations in the guidance with the original mattress, the resulting bed system with the new air mattress may still pose a risk of entrapment. When these products are used, we recommend that steps are taken to ensure that the therapeutic benefit outweighs the risk of entrapment.
NOTE: FDA continues to recommend the dimensional limits in this guidance for bed systems using mattress overlays. We recommend that steps be taken to assess the therapeutic benefit to the patient when applying a mattress overlay to a bed system that does not meet the recommended dimensional limits. The clinical benefit should outweigh the risk of entrapment presented by use of such a system.
FDA will continue to monitor adverse event data on entrapments in hospital beds. Your complete and descriptive reports of each entrapment or near-entrapment adverse event will help FDA to monitor the safety of hospital beds. When reporting an entrapment event, manufacturers and users often leave out details of the entrapment event that can be useful in identifying the factors or conditions that led to the event. Consequently, these reports only tell FDA that an adverse event took place. To improve the quality of entrapment adverse event reports, the following information is important and helpful to include:
More information on reporting adverse events to the FDA can be found at: http://www.fda.gov/cdrh/mdr/.
Three key body parts at risk for life-threatening entrapment in the seven zones of a hospital bed system discussed in this guidance are the head, neck, and chest. International anthropometric data references have been used to determine the relative sizes of these body parts for the population at greatest risk for entrapment and to provide a guide for the dimensional limits that would reduce their risk of entrapment. See Appendix D.
HeadTo reduce the risk of head entrapment, openings in the bed system should not allow the widest part of a small head (head breadth measured across the face from ear to ear) to be trapped. Country-specific anthropometric data show that a 1 st percentile female head breadth may be as small as 95 mm (3 3/4 inches). A dimension of 120 mm (4 3/4 inches) encompasses the 5th percentile female head breadth in all data sources used to develop these recommendations, and includes 1st percentile female head breadth as reported in some data sources.
FDA is therefore using a head breadth dimension of 120 mm (4 3/4 inches) as the basis for its dimensional limit recommendations. This dimension is consistent with the dimensions recommended by the HBSW and the IEC.
To
reduce the risk of neck entrapment, openings in the bed system should not allow
a small neck to become trapped.
Data show a 1st percentile female neck diameter of 79 mm (3 1/8 inches) [5th percentile = 83 mm (3 1/4 inches)]. Several factors, such as neck compressibility, loss of muscle mass in the neck when people age, and the asymmetrical shape of the neck, support the use of a reduced measurement. For example, one published estimate for compressibility of neck tissue is 25% of the uncompressed measure.14 Reducing the 79 mm measure by approximately 25% to account for tissue compression gives a measure of 60 mm (2 3/8 inches). Both the IEC and the HBSW recommend a dimensional limit of 60 mm (2 3/8 inches) to prevent neck entrapment. Consistent with these recommendations, FDA is recommending 60 mm (2 3/8 inches) as an appropriate dimension for neck diameter.
The concept of a wedging effect, which occurs when the neck is trapped in a V-shaped opening, recurs throughout many national and international entrapment-prevention standards (See Appendix C); however, the standards differ with respect to what is considered to be the critical angle for wedging. Some standards specify minimum angles to prevent neck entrapment based on a theoretical analysis of the forces on a cylindrical object (representing the cross-section of a neck) in an angled space. Depending on whether the wedging is considered to be caused by the total resultant forces on the neck or the horizontal components of the forces, the critical angles are identified as either 60 (See Appendix C, References 5 and 6) or 53 degrees (rounded up to 55) (See Appendix C, References 1,4 and 9), respectively.
When developing its recommendations for preventing neck entrapment, HBSW consulted a published international standard for swimming pool equipment [see EN 13451], which includes dimensional limits to prevent entrapment of various body parts of adults and children. This standard specifies that V-shaped openings should be of angles greater than 60 degrees. HBSW members performed analyses that support this limit, and the HBSW has recommended that V-shaped openings be greater than 60 degrees to avoid neck entrapment.
Given the adult population at risk for wedging entrapments in hospital beds, FDA is recommending a dimension of 60 mm (2 3/8 inches) to represent neck diameter. Additionally, to prevent wedging, a limit of greater than 60 degrees is recommended for V-shaped openings that a neck could enter. These dimensions are consistent with the dimensions recommended by the HBSW and the IEC (see IEC 60601-2-38-1).
The openings in a bed system should be wide enough not to trap a large chest through the opening between split rails. For purposes of the recommendations in this guidance, a 95th percentile male chest depth is used to represent the largest chest measure. Although one would assume that the largest chest size belongs to women, breast tissue is compressible and diminishes in size as aging occurs. Male chests, however, have less compressible tissue and do not diminish as significantly in size with aging. A 95th percentile male chest depth of 318 mm (12 1/2 inches), measured from the nipple to the back, including the pectoral muscles, is used to represent the largest chest measure.
The IEC is proposing to adopt a dimension to reduce chest entrapment of greater than 318 mm (12 1/2 inches). FDA concurs with the dimension of 318 mm (12 1/2 inches) to represent chest depth for the population vulnerable to entrapment, and has used this dimension as the basis for its recommended dimensional limits.
The body part dimensions used to develop FDA's dimensional limit recommendations are summarized in Table 2 below.
Key Body Part |
Dimension |
---|---|
Head |
120 mm (4 3/4 inches) |
Neck |
60 mm (2 3/8 inches) and an angle > 60 degree |
Chest |
318 mm (12 1/2 inches) |
This guidance describes seven zones in the hospital bed system where there is a potential for patient entrapment. Entrapment may occur in flat or articulated bed positions, with the rails fully raised or in intermediate positions. Descriptions of the seven entrapment zones appear on pages 15-21 in this guidance. Summary drawings of entrapment for all of the zones appear in Appendix E.
The
seven areas in the bed system where there is a potential for entrapment are
identified in the drawing below.
Zone 1: Within the Rail
Zone 2: Under the Rail, Between the Rail Supports or Next to a Single Rail Support
Zone 3: Between the Rail and the Mattress
Zone 4: Under the Rail, at the Ends of the Rail
Zone 5: Between Split Bed Rails
Zone 6: Between the End of the Rail and the Side Edge of the Head or Foot Board
Zone 7: Between the Head or Foot Board and the Mattress End
Entrapment at the Bed Deck or Frame
Many of the entrapment event reports FDA received involved entrapment between the rail and the bed's "frame." It is unclear from the event descriptions whether this refers to the mattress deck, the bed frame, or even the hardware attaching the bedrail to the bed system. While this guidance does not recommend dimensional limits on the space at the deck or frame locations, FDA believes that meeting the other recommended dimensional limits would reduce the possibility of entrapment at the deck or frame locations.
FDA's adverse event reporting system helps promote product safety by collecting information on products that are currently on the market. FDA’s reporting system collects reports of adverse events15 that caused or may have caused a death, a serious injury, or a malfunction. From January 1985 to March 2000, FDA received 390 entrapment event reports to its adverse events database from manufacturers, hospitals, nursing homes, and consumers. In 2000, HBSW reviewed these adverse event reports and identified entrapment areas or zones in the bed system and the body parts at risk. Based on its analysis of the reported adverse events, HBSW made recommendations for dimensional limits.
A retrospective study conducted by members of HBSW compared the HBSW recommended dimensions with dimensions of the bed models identified in the adverse events reports. For each of the entrapment adverse events in the study where the model number of the bed was reported, a participating bed manufacturer provided information on the dimensions of the identified area where an entrapment was believed to have occurred16. Four manufacturers provided this information. These data represented 215 (55%) of the 390 entrapment events. This information provided a reference range typical of hospital beds currently available for use in acute, long term care, and home settings.
The retrospective study compared the manufacturer-supplied information, in the aggregate, to the dimensions recommended by the HBSW. If the size of the openings in the reported bed models did not meet the HBSW recommended limits, i.e., the openings in the reported beds were outside the limits of the recommended gap sizes, then the HBSW dimensional limits were considered to be an appropriate limit to reduce entrapments at that area. We have considered the information from this study in developing the dimensional limit recommendations described in this guidance.
FDA is recommending dimensional limits for zones 1 through 4 at this time because we believe the majority of the entrapments reported to FDA have occurred in these zones. We based these recommended limits upon the body parts entrapped in these individual zones identified through adverse event reports and entrapment scenarios described in the reports. A summary table (Table 3) of the hospital bed dimensional limit recommendations appears on page 21 at the end of this section.
The Hospital Bed Safety Workgroup developed and validated test methods to measure and assess gaps or openings in zones 1-4 of hospital bed systems, reprinted in Appendix F. As a member of the HBSW, FDA participated in the development and validation of these test methods. FDA recommends these test methods as an acceptable approach for assessing hospital bed gap sizes in accordance with the dimensional limitations described below. The test methods and tool used to conduct these tests are available through HBSW (see Appendix B). If an alternate approach is used to assess gap sizes, FDA recommends that the dimensional limits used in any alternative approach be at least as stringent as the ones described below.
Zone 1 is any open space within the perimeter of the rail. Openings in the rail should be small enough to prevent the head from entering. A loosened bar or rail can change the size of the space. The HBSW and IEC recommend that the space be less than 120 mm (4 3/4 inches), representing head breadth.
Data from the Retrospective Study
Adverse events identified as occurring within the rail were reported in bed models where open spaces within the rail were greater than 120 mm (4 3/4 inches). Manufacturers' measurements of bed models representative of those identified in these incidents had spacing within the rail of between 177 mm (6.97 inches) and 190 mm (7.48 inches). The data suggest that nearly all of these entrapment events may have been prevented if the spaces within the rails had been less than 120 mm (4 3/4 inches), representing head breadth as described above. Consistent with HBSW’s and the IEC’s recommendations and comments received on the previously published draft FDA guidance, FDA is recommending a measure of less than 120 mm (4 3/4 inches) as the dimensional limit for any open space within the perimeter of a rail. The space should not permit an object with a circular cross-section measuring less than 120 mm (4 3/4 inches) in diameter to enter or pass through.
Zone 2 - Under the Rail, Between the Rail Supports or Next to a Single Rail Support
This space is the gap under the rail between a mattress compressed by the weight of a patient's head and the bottom edge of the rail at a location between the rail supports, or next to a single rail support. If there is a single rail support, entrapment in Zone 2 can occur anywhere along the bottom length of the rail beyond the support, up to the end of the rail. (Entrapment at the end of the rail is explained in Zone 4 .) Factors to consider are the mattress compressibility17 which may change over time due to wear, the lateral shift of the mattress or rail, and any degree of play from loosened rails or rail supports. A restless patient may enlarge the space by compressing the mattress beyond the specified dimensional limit. This space may also change with different rail height positions and as the head or foot sections of the bed are raised and lowered. The space may increase, decrease, become less accessible, or disappear entirely. In some positions, the potential for entrapment in this zone may still exist when the deck is articulated.
Preventing the head from entering under the rail would most likely prevent neck entrapment in this space. FDA recommends that this space be small enough to prevent head entrapment, less than 120 mm (4 3/4 inches). IEC recommends the same dimensions but measures the space without the mattress in place.
Data from the Retrospective Study
In the study, the manufacturers were instructed to measure diagonally from the top edge of the compressed mattress to the lowest inside edge of the rail between the rail supports. This measurement ranged from between 76 mm (3 inches) and 191 mm (7.5 inches). If the reported entrapments occurred at Zone 2,the data suggest that the HBSW recommended dimensional limit of less than 120 mm (4 3/4 inches) would have prevented only about half of the reported events at this zone. At times, the adverse event report information was not clear, and it was difficult to determine the precise location of the entrapment and to determine whether it occurred in Zone 2, 3, or 4. Most reports only stated that an entrapment occurred "between the rail and the mattress." However, given the scenarios in the reports, some of these events may have occurred at the rail end, beyond the support (Zone 4) as neck entrapments when the head entered under the rail first. 17
Because the data for reported entrapments at Zone 2 are not definitive and the most likely scenario for entrapment in this space would include a head-first entry, the dimensional limit of 120mm (4 ¾ inches) is being recommended.
Zone 3 - Between the Rail and the Mattress
This area is the space between the inside surface of the rail and the mattress compressed by the weight of a patient's head. The space should be small enough to prevent head entrapment when taking into account the mattress compressibility, any lateral shift of the mattress or rail, and degree of play from loosened rails. HBSW and IEC recommend a dimension of less than 120 mm (4 3/4 inches) because the head is presumed to enter the space before the neck. FDA is recommending a dimensional limit of less than 120 mm (4 3/4 inches) for the area between the inside surface of the rail and the compressed mattress.
Data from the Retrospective study
A review of the manufacturers' supplied measurements indicates that the horizontal gap between the rail and the uncompressed mattress for bed models involved in entrapments believed to have occurred at Zone 3 was between 38 mm (1 1/2 inches) and 127 mm (5 inches). Theoretically, entrapment would involve compression of the mattress which would have resulted in a larger gap than the manufacturer measurements without patients present and, subsequently, a larger range of measures. Further, it could not be determined from the description of entrapment events whether entrapments occurred at Zones 2, 3 or 4. FDA recommends a 120mm (4 3/4 inch) dimensional limit for this zone based on the head breadth dimension described above.
Zone 4 - Under the Rail at the Ends of the Rail
This space is the gap that forms between the mattress compressed by the patient, and the lowermost portion of the rail, at the end of the rail. Factors that may increase the gap size are: mattress compressibility, lateral shift of the mattress or rail, and degree of play from loosened rails. The space poses a risk for entrapment of a patient's neck. It may change with different rail height positions and as the head or foot sections of the bed are raised and lowered. The space may increase, decrease, become less accessible, or disappear entirely. Thus, in some positions, the potential for entrapment in this zone may still exist when the deck is articulated.
At the time of this publication, the IEC international standard recommends a dimensional limit of less than 60 mm (2 3/8 inches) measured between the mattress support platform and the lowest portion of the rail at the rail end to prevent neck entrapment. Based on the neck diameter dimension described above, FDA recommends that the dimensional limit for this space also be less than 60 mm (2 3/8 inches). To reduce the risk of neck entrapment at Zone 4, FDA recommends consideration of the combination of the gap size and the angle size (created between the mattress and the rail). Thus, FDA recommends that the V-shaped opening under the rail at its end be of an angle wide enough, i.e. greater than 60 degrees, to prevent wedging entrapment (See Neck Section on pages 11 and 12 for a description and diagram of wedging entrapments).
Data from the Retrospective Study
The retrospective study measures for Zone 4 ranged between 102 mm (4 inches) and 152 mm (6 inches) for the diagonal measure between the inside bottom edge of the rail at the end of the rail and the top of the compressed mattress. It could not be determined from the entrapment event reports whether the entrapments occurred in Zone 4, or whether events reported as neck entrapment occurred as a result of head-first entry. The HBSW recommended that a 60 mm dimension be used to represent neck diameter. Later the HBSW identified the importance of avoiding wedging of the neck at the end of the rail, and concluded that a linear measure for the Zone 4 space would not adequately address mattress compressibility and wedging forces in Zone 4. HBSW recommended that openings in Zone 4 measure both less than 60 mm in size and greater than 60 degrees in angle. FDA agrees with this recommendation.
Zones 5, 6 and 7
Although seven potential zones of entrapment have been identified by HBSW, FDA is recommending dimensional limits for zones 1-4 because these zones were most frequently reported as having entrapments. The current international standard (IEC 60601-2-38) addresses limits for zones 1, 2, 4, and 6. In addition, IEC intends to set or revise dimensional limits, for areas comparable to HBSW’s zones 1-6 in their upcoming international standard for hospital beds. FDA continues to receive entrapment reports for Zones 5 and 6, and Zone 7 remains a potential for entrapment. FDA will monitor entrapments in these zones and consider harmonization with the IEC standard once it is available.
Zone 5 - Between Split Bed Rails
This zone occurs when partial length head and foot side rails (split rails) are used on the same side of the bed. The space between the split rails may present a risk of either neck entrapment or chest entrapment between the rails if a patient attempts to, or accidentally, exits the bed at this location. In addition, any V-shaped opening between the rails may present a risk of entrapment due to wedging. FDA recognizes this area as a potential for entrapment and encourages facilities and manufacturers to report entrapment events at this zone.
Zone 6 - Between the End of the Rail and the Side Edge of the Head or Foot Board
Zone 6 is the space between the end of the rail and the side edge of the headboard or footboard. This space may present a risk of either neck entrapment or chest entrapment. In addition, any V-shaped opening between the end of the rail and the head or footboard may present a risk of entrapment due to wedging. This space may change when raising or lowering the head or foot sections of the bed. This space may increase, decrease, become less accessible, or disappear entirely. Thus, in some positions, the potential for entrapment may exist when the deck is articulated. FDA recognizes this area as a potential for entrapment and encourages facilities and manufacturers to report entrapment events at this zone.
Zone 7 - Between the Head or Foot Board and the End of the Mattress
Zone 7 is the space between the inside surface of the head board or foot board and the end of the mattress. This space may present a risk of head entrapment when taking into account the mattress compressibility, any shift of the mattress, and degree of play from loosened head or foot boards. FDA recognizes this area as a potential for entrapment and encourages facilities and manufacturers to report entrapment events at this zone.
Zone |
Dimensional Limit Recommendations |
---|---|
|
|
1 Within the rail |
<120 mm (< 4 3/4 ") |
2 Under the rail, between rail supports or next to a single rail support |
< 120 mm (< 4 3/4 ") |
3 Between rail and mattress |
<120 mm (< 4 3/4 ") |
4 Under the rail, at the ends of the rail |
<60 mm (< 2 3/8 ") AND >60° angle |
Consulting Organizations to the Hospital Bed Safety Workgroup
Websites:
Food and Drug Administration:
Bureau of Medical Devices, Health Canada:
http://www.hc-sc.gc.ca/hpfb-dgpsa/tpd-dpt/index_advisories_professionals_e.html
ECRI :
http://www.ecri.org/Patient_Information/Patient_Safety/BedSafetyClinicalGuidance.pdf
HealthSafe, Inc.:
Untie the Elderly, Kendal Corporation:
American Association of Homes and Services for the Aging:
American Health Care Association:
Kit Information:
The HBSW Bed Safety Entrapment Kit will be available through National Safety Technologies at http://www.nst-usa.com. The kit will include:
Documents
Test tool kit
Educational videotapes:
1. ASTM International. “Standard Consumer Safety Performance Specification for Playground Equipment for Public Use.” Annual Book of ASTM Standards, Vol. 15.07, Designation F 1487.
2. ASTM International. “Standard Consumer Safety Specification for Expansion Gates and Expandable Enclosures.” Annual Book of ASTM Standards, Vol. 15.07, Designation F 1004.
3. ASTM International. “Standard Consumer Safety Specification for Bunk Beds.” Annual Book of ASTM Standards, Vol. 15.07, Designation F1427.
4. U.S. Department of the Army. “ Child Development Center Play Area Inspection and Maintenance Program.” Publication No. TM 5-663.
Public domain document available at www.army.mil
5. European Committee for Standardization. “Playground Equipment – Part 1: General Safety Requirements and Test Methods.” Ref. No. EN 1176-1.
Adopted and published under various national designations by 28 member countries of CEN, including France, Great Britain, Germany, and Sweden.
6. European Committee for Standardization. “Swimming Pool Equipment – Part 1: General Safety Requirements and Test Methods.” Ref. No. EN 13451-1.
Adopted and published under various national designations by 25 member countries of CEN, including France, Great Britain, Germany, and Sweden.
7. 16 CFR Part 1213, “Safety Standard for Entrapment Hazards in Bunk Beds.”
8. 16 CFR Part 1513, “Requirements for Bunk Beds.”
9. U.S. Consumer Product Safety Commission. “Handbook for Public Playground Safety.” Publication No. 325.
Public domain document, available at www.cpsc.gov
Anthropometric References used:
Hall, Judith. Handbook of normal physical measurements. New York : Oxford
University Press, 1990.
[Note: Head width and neck circumference data for both sexes,-2SD (2.5th percentile),
from birth to age 16. Data visually extrapolated from graphs.]
Jurgens, H., Pieper, U. International data on anthropometry. Geneva,
Switzerland : International Labour Office, 1990. (Occupational safety and health
series; no. 65).
[Note: This report reviewed data for North America, Latin America (Indian population),
Latin America (European and Negroid population, Northern Europe, Central Europe,
Eastern Europe, South-Eastern Europe, France, Iberian Peninsula, North Africa,
West Africa, South-Eastern Africa, Near East, North India, South India, North
Asia, South China, South-East Asia, Australia (European Population), Japan.]
Peebles, Laura, Norris, Beverly J. Adultdata - The handbook of adult anthropometric
and strength measurements: data for design safety. London : Department of
Trade and Industry, 1998.
[Note: This handbook contains data from various sources for the following countries:
UK, Brazil, France, Germany, Italy, Japan, Poland, Sri Lanka, Sweden, Netherlands,
and USA. Data were not available from ALL these countries for EACH measurement.]
Smith, Stuart, Norris, Beverly, Peebles, Laura. Older adult data –
The handbook of measurements and capabilities of the older adult: data for design
safety. London : Department of Trade and Industry, 2000.
[Note: This handbook contains data from various sources for the following countries:
UK, Brazil, France, Germany, Italy, Japan, Poland, Sri Lanka, Sweden, Netherlands,
and USA. Data were not available from ALL these countries for EACH measurement.]
Snyder, JRG. Anthropometry of infants, children and youths to age 18 for product safety design: final report. Bethesda, MD : Consumer Product Safety Commission, 1977.
Others references consulted:
Association for the Advancement of Medical Instrumentation. Human factors engineering guidelines and preferred practices for the design of medical devices. 2nd ed. Arlington, VA : Association for the Advancement of Medical Instrumentation, 1993 ; AAMI HE-48 1993.
British Standards Institution. 1987 Ergonomics – standards and guidelines for designers. United Kingdom : British Standards Institution, 1987 ; document no. PP 7317.
Damon, Albert, Stoudt, Howard W., McFarland, Ross A. The human body in equipment design. Cambridge, MA : Harvard University Press, 1966.
Diffrient, Niels. Humanscale one-two-three. Cambridge, MA : MIT Press, 1974.
Human engineering design data digest. Washington, DC : U.S. Government Printing Office, 1975.
National Center for Health Statistics. Weight, height, and selected body dimensions of adults, United States, 1960-62. Hyattsville, MD: National Center for Health Statistics, 1980; DHHS publication no. (PHS) 80-1301. Vital and health statistics; series 11, no. 8.
Salvendy, Gavriel, ed. Handbook of human factors. New York : John Wiley & Sons, 1987.
Woodson, Wesley E. Human factors design handbook : information and guidelines for the design of systems, facilities, equipment and products for human use. New York : McGraw-Hill Book Company, 1981.
Woodson, Wesley E., Conover, Donald W. Human engineering guide for equipment designers. 2nd ed. Berkeley, CA : University of California Press, 1970.
Woodson, Wesley E., Tillman, Peggy, Tillman, Barry. Human factors design handbook. 2nd ed. New York : McGraw-Hill Professional, 1992.
Zone 1 - Entrapment within the rail
|
Zone 2 - Entrapment under the rail, between the rail supports or next to a single rail support |
Zone 3 - Entrapment between the rail and the mattress |
Zone 4 - Entrapment under the rail, at end of rail |
Zone 5 - Entrapment between split bed rails |
Zone 6 - Entrapment between the end of the rail and the side edge of the head or foot board |
Zone 7 - Entrapment between head or foot board and the mattress end |
Introduction
The Hospital Bed Safety Workgroup (HBSW) and the U.S. Food and Drug Administration (FDA) have defined seven numbered "zones" or spaces in and around hospital bed systems where patients could potentially become trapped. Actual entrapments have been reported in six of these zones, with Zones 1, 2, 3, and 4 accounting for approximately 80% of entrapment events reported to the FDA. This appendix contains instructions for testing Zones 1, 2, 3, and 4, using tools and methods developed by the HBSW.
Contents
Summary of Test Zones
Description of Test Tools
Cone and Cylinder Tool
Tool Assembly and Disassembly
Using the spring scale
General Testing Considerations
Test Methods
Zone 1 Test
Zone 2 Test
Zone 3 Test
Zone 4 Test
Appendix: Testing Tips and Frequent Questions
SAMPLE DATASHEETS
The four (4) tests in these instructions measure gaps within bed systems where
a patient could become trapped. Each test measures a different area, or zone,
where
entrapment can occur:
Zone 1: Within the rail
Zone 2: Under the rail between the rail supports or next to
a single support
Zone 3: Between the rail and the mattress
Zone 4: Under the rail at the end of the rail
Each test requires the use of simple tools, including a cone, a cylinder, and a spring scale. Your tools may look slightly different from the tools in the figures, but they will work the same way.
The cone and cylinder is a combination tool (see Figure 5). It can be easily taken apart so that the cone and cylinder can be used separately. Tests 1, 2 and 3 use only the cone. Test 4 uses the combined assembled tool.
The cone tool includes the following features:
The cylinder includes the following features:
To prevent personal injury during the measurement process, attach the strap to a secure point on the bed and shorten the length of the safety strap enough to keep the tool from dropping on your feet if it should fall during a test. Make sure the strap is long enough to not interfere with the test measurement.
Note: General procedures for tools with a screw-type connection are described here. (Your tool may have a different type of connection.) Follow the instructions supplied with your tools for more detailed information.
To take the cone and cylinder tool apart:
1. Turn the knob to loosen and remove the connection shaft.
2. Pull the cylinder from the cone.
To put the cone and cylinder tool together:
1. Align the red and green areas of the cone and cylinder.
2. Insert the pins of the cone into the cylinder.
3. Insert the connection shaft and turn the knob to tighten.
Use
the spring scale to apply 12 lbs of force to the cone when testing Zones 1 and
2. At the small end of the cone tool, insert the scale hook into the metal loop.
Pull the scale slowly until the needle points to 12 lbs.
For other types of scales (sliders, digital, etc.), and for information on scale calibration, consult the instructions supplied with your toolkit.
General Testing Considerations
This test assesses the potential for head entrapment within the perimeter of the rail.
The tools needed to do this test are the cone, the safety strap and the spring scale.
Prepare for the Zone 1 Test:
Do
the Zone 1 Test:
Interpret the Zone 1 Results:
If the large end of the cone does not enter any of the openings, this space passes the test (figure 9a).
If the large end of the cone does enter or pass through any of the openings, this space fails the test (figure 9b).
Figure 9a. Zone 1 test: Pass, large end of cone does not enter rail opening. |
Figure 9b. Zone 1 test: Fail, large end of cone does enter rail opening. |
Repeat the Zone 1 Test:
On the other rails : Repeat the test for all other rails on the bed.
This test assesses the potential for head entrapment under the rail, at a location between the rail supports or next to a single support.
The tools needed to do this test are the cone, the safety strap and the spring scale.
Prepare for the Zone 2 Test:
Do
the Zone 2 Test:
Interpret Zone 2 Test Results:
If the large end of the cone does not enter the space under the rail, or pass under the rail, this space passes the test (figures 11a and 11b).
If the large end of the cone does enter the space under the rail, or if it passes under the rail, this space fails the test (figures 11c and 11d).
Repeat the Zone 2 Test:
On the same rail: If the rail has intermediate locking positions, perform the test for every intermediate position.
On the other rails: Perform the test for all other rails on the bed, including intermediate and raised positions.
This test assesses the potential for head entrapment between the inside of the rail and the surface of the mattress (compressed by the weight of a patient’s head).
The tools needed for this test are the cone and the safety strap.
Prepare for the Zone 3 Test:
Do the Zone 3 Test:
Note: If a mattress stop, rail support, or other structure keeps the cone from sinking in the gap, put the cone tool at a different location along the rail where there is no interference.
Interpret Zone 3 Test Results:
If the line across the flat end of the cone is above the surface of the mattress, the space passes the test (figure 13a).
If the line across the flat end of the cone is at orbelow the top surface of the mattress, the space fails the test (figure 13b).
Repeat the Zone 3 Test:
On the same rail: If the rail has any intermediate positions, perform the test at every intermediate position.
On the other rails : Perform the test for all other rails on the bed, including the intermediate positions.
This test assesses the potential for neck entrapment between the top of the mattress (compressed by the patient) and the lower-most portion of the rail, at the end of the rail.
The tools needed for this test are the assembled cone and cylinder with safety strap.
Prepare for the Zone 4 Test:
Do the Zone 4 Test:
Note: If the cylinder tool cannot fit into an area between the head or footboard and the end of a rail, the space passes.
Interpret Zone 4 Test Results:
If the cylinder touches the rail in the green area, the space passes (figure 16a).
If the cylinder touches the rail in the red area, the space fails (figure 16b).
If the cylinder passes completely under the rail, the space fails (figure 16c).
Reminder: If the cylinder tool cannot fit into an area between the head or footboard and the end of a rail, the space passes.
Repeat the Zone 4 Test:
On the same rail: If the rail has any intermediate positions, perform the test at every intermediate position.
On the other rails : Perform the Zone 4 test at both ends of all other rails on the bed, including the intermediate positions.
Appendix to HBSW Dimensional Test Methods for Bed Systems
Testing Tips and Frequent Questions
Do I have to do the tests in order?
No. You may do the tests in any order you wish. You may find it convenient to do all the tests on one side of the bed, then the other (to keep from repeatedly pushing the mattress back and forth). Or, you may wish to do all of the tests that use just the cone, then all of the tests that use the cone and cylinder together (to avoid having to put the tool together and take it apart several times).
Do I have to use the safety strap?
The test tools are heavy, and the safety strap helps protect you if they fall or suddenly pull through a gap. Not using the safety strap increases your risk of being injured.
How do I keep track of my test results?
You may use any method that meets the recordkeeping requirements of your institution. HBSW has developed some sample data sheets that you may use or modify according to your needs. Sample data sheets are attached at the end of this document.
What are the most common mistakes people make when doing the tests?
In field tests conducted to evaluate the test tools and instructions, some of the most common errors made by the testers were:
When pushing the mattress, how hard do I need to push?
Some test instructions ask you to push the mattress "until it stops." Usually that means pushing until either:
Always make sure you push the mattress straight across; it should not be crooked on the bed.
When testing Zone 2, what position should the bed be in?
Unlike the other three tests, the Zone 2 test is not always done with the bed in the flat position. Before you do the Zone 2 test, you first need to find the correct testing position. With the bed in the flat position, identify the location of the Zone 2 space. Then, articulate the bed (adjust the head and knee sections) while you watch the Zone 2 space. As you articulate the bed, the Zone 2 space may get bigger, get smaller, or stay the same. Adjust the position of the bed until you find the position that makes the largest opening in Zone 2. If the size of the opening gets smaller, or does not change, return the bed to the flat position to do the test. Some examples of how Zone 2 might change with articulation are shown below.
In this example, the opening in Zone 2 gets larger when the bed is articulated (because of the bend in the mattress). Zone 2 should be tested with this bed in the position where the space is largest. |
In this example, the size of the Zone 2 opening does not change as the bed articulates (the rail and the mattress stay in the same relative position). Zone 2 should be tested with this bed in the flat position. |
In this example, the Zone 2 openings get smaller when the bed is articulated. Zone 2 should be tested with this bed in the flat position. (Note: For the rail design shown below, Zone 2 and Zone 1 are the same. See the discussion of unusual rail designs.)
|
What if one of the Zones doesn't seem to exist on a particular bed system?
If a zone doesn't exist on a particular bed, then there is no risk of a patient becoming trapped in that Zone. For example: if there are no openings in a rail, then a patient could not get his or her head trapped in Zone 1 of that rail. Your facility's policy and method of recordkeeping will determine how you should record data for Zones that do not exist. For example, your facility may use customized data sheets that simply omit zones or rails that do not exist. If you are using a generic form (such as a table or spreadsheet), it may be appropriate to record a result of either “pass,” “not applicable,” or “not tested,” or to cross out entry spaces for Zones that do not exist. (You should provide some indication that the Zone was not accidentally omitted from testing.)
I need to test a bed with an unusual rail design. The rails don't look like any of the rails shown in the instructions or video. How do I figure out where to do each test?
Occasionally, the most difficult part of a test may be figuring out where on the bed to perform the test. For some beds or rail configurations, it may be difficult to identify the locations of various Zones. The following hints and examples for each Zone may help.
Zone 1: The goal of the Zone 1 test is to see whether a patient could become trapped by putting his or her head through an opening in the rail itself.
Example: No openings in the rail
If there are no openings in the rail, then there is no risk of entrapment in Zone 1. Examples of cases where there may not be any openings in the rail are:
Zone 2: The goal of the Zone 2 test is to see whether a patient could become trapped by putting his or her headunder the rail, head first, between the rail supports (or next to a single support). (Note: Trapping the head by sliding the neck under the rail sideways is addressed by the test for Zone 4.)
In some designs, the rail supports also form part of the body of the rail. Crib-style rails, shown here, are common examples. Often, when the rail supports form part of the rail, Zones 1 and 2 overlap. Zone 1 should be tested with the bed in the flat position; if the size of the space changes with articulation, the result for Zone 2 may be different. |
Zone 3: The goal of the Zone 3 test is to see whether a patient could become trapped with his or her head in the horizontal space between the rail and the mattress. Usually, the gap in Zone 3 is easily identified, even if the rail design is unusual.
Zone 4: The goal of the Zone 4 test is to see whether a patient could become trapped by sliding or wedging his or her neck under the end of the rail.
Example: The rail end is very close to head/foot board (very little space in between)
When testing in Zone 4, do not try to force the cylinder down between the rail and the end board or insert it under the rail from behind. If the cylinder’s diameter will not fit between the end of the rail and the head or foot board, record a result of Pass for Zone 4.
An example of this rail type is shown at the right. The cylinder cannot pass between the end of the rail and the head/foot board and then under the rail -- it would stop on the support. The result for Zone 4 is Pass, because a patient would not be able to get his or her neck under the rail without the head or body going under first. The potential for head-first entrapment under the rail should be evaluated using the test for Zone 2. |
When the end of the rail is completely vertical from top to bottom, there may be no space "under" the end of the rail where the cylinder (or a patient's neck) could fit. In the example at right, the left side of the rail is vertical, and there is no space “under” the end of the rail. For the left rail end in the figure, the rating for Zone 4 should be either “pass” or “not applicable,” depending on your facility’s policy. The right end should be tested as usual for Zone 4. Note: If you are unsure about a particular rail you are testing, you should go ahead and perform the Zone 4 test to verify your answer. The cone and cylinder would give a result of Pass if they were used to physically test the area shown on the left side of the rail in the figure. |
Why did my co-worker and I get different results for the same test?
Different testers can get different test results for the same bed system. Or, the same tester may get a different result when he or she repeats a test. Some reasons this can happen are:
What if a result is too close to call?
If you are faced with a borderline pass/fail decision, use your best judgment to decide on a result. You can also try these suggestions:
Your facility may also have suggestions or procedures for handling close calls. For beds with borderline passes that result from close calls, you may wish to consider mitigation (see the next question below).
What should I do if a bed system fails one of the tests?
You should strongly consider mitigation of entrapment risks for bed systems that fail any of the four tests. You may also wish to consider mitigation for bed systems that have borderline passes. Mitigation may include patient assessment; some patients are at lower risk for entrapment than others. For more information, consult the following, available at http://www.fda.gov/cdrh/beds:
What if I have a question that wasn't answered here?
There are several resources available to help address questions. As a start, you may contact the supplier of your beds, mattresses, or rails. Your tool kit supplier is another resource. In addition, an advisory group is being established by the HBSW to address additional questions that arise. You may submit questions to that committee via the Web at [URL to be determined, www.xxx.yyy].
Start of Footnotes
1 The terms “medical bed” and “hospital bed” are used interchangeably throughout this document and include adult medical beds with siderails. See discussion in Scope, page 5.
2 As used in this guidance, “hospital bed system” encompasses the bed frame and its components, including the mattress, bed side rails, head and foot board, and any accessories added to the bed.
3 Title 21, Code of Federal Regulations, Part 820 – Quality System regulation.
4 Ferlo Todd J, Ruhl C, Gross T P, “Injury and Death Associated with Hospital Bed Side Rails: Reports to the U.S. Food and Drug Administration from 1985 to 1995.” American Journal of Public Health 1997; 87:1675-1677.
5 Hospital Bed Safety Workgroup “A Guide to Bed Safety,” http://www.fda.gov/cdrh/beds/.
6 U. S. Food and Drug Administration. FDA Safety Alert: Entrapment Hazards with Hospital Bed Side Rails (August 23, 1995), U.S. Department of Health and Human Services.
7 “Vulnerable patients” are defined in “A Guide to Bed Safety,” developed by the Hospital Bed Safety Workgroup (described on pages 3-4), as: “Patients who have problems with memory, sleeping, incontinence, pain, uncontrolled body movement or who get out of bed and walk unsafely without assistance. These patients most often have been frail, elderly or confused."
8 FDA uses the term “bed rails” frequently throughout this document. Commonly used synonymous terms are side rails, bed side rails, grab bars and safety rails. Bed rails are rigid bars that are attached to the bed and are available in a variety of sizes and configurations from full length to half, one-quarter, and one-eighth length and are used as restraints, reminders, or as assistive devices. An historical review can be found in Braun & Capezuti, "The Legal and Medial Aspects of Physical Restraints and Bed Side rails and Their Relationship to Falls and Fall-Related Injuries in Nursing Homes,” DePaul Journal of Health Care Law, vol. 4, Fall 2000.
9 FDA acknowledges several limitations of these adverse event report data. First, many adverse events may not be reported to the FDA, thus the true number of adverse events may be unknown. Second, the number of reported events does not represent incident rates for a given problem in the absence of a defined denominator—the number of individuals at risk for a given adverse event. Finally, many reports lack a complete and detailed description of the adverse event or are not verified. Despite these limitations, adverse event reports can suggest a profile of the areas or locations on a hospital bed that present a risk of entrapment, as well as the parts of the body that are at risk of entrapment.
10 International Electrotechnical Commission (IEC) 60601-2-38, Amendment 1, 1999 Medical Electrical Equipment – Part 2-38: Particular Requirements for the Safety of Electrically-Operated Hospital Beds.
11 Schatz, William M., PT, Bed Safety Consulting, Inc., Comment #C2 to Docket #2004 D-0343, Draft Guidance for Industry and FDA Staff; Hospital Bed System Dimensional Guidance to Reduce Entrapment, November 10, 2004.
12 Miles SH. Deaths between Bedrails and Air Pressure Mattresses, Journal of the American Geriatrics Society 2002; 50:1124-5.
13 Joint Commission on Accreditation of Healthcare Organizations, Issue 17 Sentinel Event Alert: Bed Rail-Related Entrapment Deaths (Sept. 6, 2002).
14 ASTM International. “Standard Consumer Safety Specification for Expansion Gates and Expandable Enclosures.” Annual Book of ASTM Standards, Vol. 15.07, Appendix X2, Designation F 1004.
15 Note: Many reports lacked a complete and detailed description of the adverse event. The beds involved in these adverse events may not have had compatible mattresses or bed rails specifically designed for the particular bed model involved in the reported entrapments. Also, information was limited regarding the condition of the beds, bed rails, and mattress at the time of the entrapments. Specific details about the exact location of the entrapments within the beds were sometimes lacking. Despite these limitations, adverse event reports can suggest a profile of the areas or locations on a hospital bed where entrapment can occur, as well as the parts of the body at risk for entrapment.
16 When manufacturers measured the gaps for the retrospective study, they used mattresses of the size, type and thickness typically recommended for use with their bed models. Mattresses involved in reporting entrapment events may have been different from the manufacturers’ recommended mattresses, which means actual gap sizes in entrapments involving the mattresses may have been different from those identified by the manufacturers in the retrospective study. The manufacturers’ measurements may have been representative of “best case measurements.” We also note that spaces in a hospital bed system may vary in size when the hospital bed system is articulated through the various ranges of motion. For the retrospective study, manufacturers measured gap sizes with the beds in the flat position. This means that if the bed was articulated in reported entrapments, the size of the gap may have been different from that provided by the manufacturers in the retrospective study.
17 "Mattress compressibility” refers to the extent to which a mattress changes in dimension as a result of the weight of a patient’s body or body part moving on, across, or off of the mattress surface. The mattress dimension may also change as a result of articulation of the deck.
Updated March 9, 2006
CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page
Center for Devices and Radiological Health / CDRH