Electrical/Electronic

Bob Quinn completes an air velocity measurement        Bob Quinn performs a mass flow calibration

ORNL Metrology Laboratory (ORNLML) maintains a variety of electrical standards traceable to SI through the National Institute of Standards and Technology (NIST) and other international primary laboratories. These standards generate and measure various electrical properties required to perform traceable calibrations of our customers’ instruments and equipment. ORNLML can provide accredited calibrations for any electrical parameter within the ranges detailed in our Scope of Accreditation. However, traceable non-accredited calibrations are available for electrical parameters outside of our Scope of Accreditation. The tables below list our accredited capabilities. Our entire Scope of Accreditation can be found at the following link.

Typical types of equipment the ORNLML performs electrical calibrations on are: oscilloscopes, multimeters from hand held 3.5 digit to laboratory 8.5 digit models, clamp type ammeters, resistors both fixed and decade, frequency counters and insulation testers.

If you have questions about our electrical capabilities, please contact the ORNL Metrology Laboratory for more information.

ORNL Metrology Laboratory Accredited Electrical Capabilities

AC Current


Range in A

Best Uncertainty (+/-) in ANote 1
Frequency in Hz

 

Remarks

 

10

20

40

1k

5k

10k

 

0.00022

1.16 E-06

1.16 E-06

5.78 E-07

3.05 E-07

4.19 E-07

4.19 E-07

Source/Measure

0.0022

1.16 E-05

1.16 E-05

5.78 E-06

3.03 E-06

4.19 E-06

4.19 E-06

Source/Measure

0.022

1.16 E-04

1.16 E-04

5.78 E-05

3.02 E-05

4.13 E-05

4.13 E-05

Source/Measure

0.22

1.16 E-03

1.16 E-03

6.01 E-04

4.63 E-04

9.37 E-04

9.37 E-04

Source/Measure

2.2

 

1.16 E-03

1.39 E-03

1.39 E-03

9.43 E-03

9.43 E-03

Source/Measure

11

 

 

2.91 E-02

2.95 E-02

3.59 E-02

3.67 E-02

Source/Measure

2 Wire Resistance


Range in Ω

Best Uncertainty (+/-) in Ω Note 1

 

Remarks

0

2.89 E-01

Source/Measure

1

2.89 E-01

Source/Measure

1.9

2.89 E-01

Source/Measure

10

2.89 E-01

Source/Measure

19

2.90 E-01

Source/Measure

100

2.91 E-01

Source/Measure

190

2.92 E-01

Source/Measure

1 k

3.01 E-01

Source/Measure

1.9 k

3.19 E-01

Source/Measure

10 k

4.18 E-01

Source/Measure

19 k

5.89 E-01

Source/Measure

100 k

1.63 E +00

Source/Measure

190 k

6.04 E +00

Source/Measure

1 M

2.31 E + 01

Source/Measure

1.9 M

2.27 E +02

Source/Measure

10 M

7.31 E +02

Source/Measure

19 M

1.21 E + 04

Source/Measure

100 M

5.92 E + 04

Source/Measure

 

4 Wire Resistance


Range in Ω

Best Uncertainty (+/-) in Ω Note 1

 

Remarks

0

6.56 E-05

Source/Measure

1

9.59 E-05

Source/Measure

1.9

1.40 E-04

Source/Measure

10

2.69 E-04

Source/Measure

19

9.39 E-04

Source/Measure

100

2.11 E-03

Source/Measure

190

3.11 E-03

Source/Measure

1 k

1.31 E-02

Source/Measure

1.9 k

3.10 E-02

Source/Measure

10 k

1.31 E-01

Source/Measure

19 k

2.96 E-01

Source/Measure

100 k

1.37 E +00

Source/Measure

190 k

5.76 E +00

Source/Measure

1 M

2.28 E + 01

Source/Measure

1.9 M

2.27 E +02

Source/Measure

10 M

7.31 E +02

Source/Measure

19 M

1.21 E + 04

Source/Measure

DC Voltage


Range in V

Best Uncertainty (+/-) in V Note 1

 

Remarks

0

9.33 E-07

Source/Measure

0.22

3.27 E-06

Source/Measure

2.2

2.20 E-05

Source/Measure

11

1.52 E-04

Source/Measure

22

2.92 E-04

Source/Measure

220

2.73 E-03

Source/Measure

1100

1.26 E-02

Source/Measure

 

DC Voltage Fixed Points


Range in V

Best Uncertainty (+/-) in V Note 1

 

Remarks

0.1

1.22 E-06

Source/Measure

1

3.65 E-06

Source/Measure

10

2.46 E-05

Source/Measure

100

3.33 E-04

Source/Measure

1000

4.16 E-04

Source/Measure

DC Current


Range in A

Best Uncertainty (+/-) in A Note 1

 

Remarks

0

3.47 E-09

Source/Measure

0.00022

1.38 E-08

Source/Measure

0.0022

1.14 E-07

Source/Measure

0.022

1.45 E-06

Source/Measure

0.22

3.75 E-05

Source/Measure

2.2

2.74 E-03

Source/Measure

11

4.89 E-03

Source/Measure

LF AC Voltage


Range

Best Uncertainty (+/-) in % + μVNote 1
Frequency in Hz

 

10 to 20

20 to 40

40 to 20k

20k to
50 k

50k to 100k

100k to 300k

300k to 500k

500k to 1 M

2.2 mV

0.209 +4

0.209 +4

0.18 +4

0.808 + 4

1.697 + 5

1.680 +10

1.680+10

1.683 +20

22 mV

0.036 +4

0.036 +4

0.047 +4

0.047 +4

0.105 +4

0.406 +5

1.214 +10

1.215 +20

220 mV

0.0.34+12

0.032 +7

0.032 +7

0.047 +7

0.105 +7

0.405 +17

1.214 +20

1.215 +25

2.2 V

0.034+40

0.032 +15

0.032 +15

0.046 +8

0.104 +10

0.404 +30

1.213 +80

1.214 +200

22 V

0.048 +400

0.047 +150

0.047 +150

0.052 +50

0.150 +100

0.150 +200

1.790 +600

1.791 +2000

 

Range

Best Uncertainty (+/-) in % + mVNote 1
Frequency in Hz

 

10 to 20

20 to 40

40 to 20k

20k to 50 k

50k to 100k

100k to 300k

300k to 500k

500k to 1 M

220 V

0.071 +4

0.070 +1.5

0.070 +1.5

0.150 +0.6

0.358 +1

0.358 +2.5

1.749 +16

1.777 +40

1100 V

0.067 +16

0.053 +4

0.053 +3.5

 

 

 

 

 

 

Frequency Dissemination


Range in Hz

Best Uncertainty (+/-)  (Fractional Frequency Error) Note 1

 

Remarks

10 M

1.07 x 10-11

Source/Measure

Note 1 Represents an expanded uncertainty using a coverage factor, k=2, at an approximate level of confidence of 95%.

 

 